Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

Analysis of remaining motion using one innovative upper airway opening cervical collar and two traditional cervical collars

Abstract

The aim of this study was to compare the remaining motion of an immobilized cervical spine using an innovative cervical collar as well as two traditional cervical collars. The study was performed on eight fresh human cadavers. The cervical spine was immobilized with one innovative (Lubo Airway Collar) and two traditional cervical collars (Stifneck and Perfit ACE). The flexion and lateral bending of the cervical spine were measured using a wireless motion tracker (Xsens). With the Weinman Lubo Airway Collar attached, the mean remaining flexion was 20.0 ± 9.0°. The mean remaining flexion was lowest with the Laerdal Stifneck (13.1 ± 6.6°) or Ambu Perfit ACE (10.8 ± 5.8°) applied. Compared to that of the innovative Weinmann Lubo Airway Collar, the remaining cervical spine flexion was significantly decreased with the Ambu Perfit ACE. There was no significant difference in lateral bending between the three examined collars. The most effective immobilization of the cervical spine was achieved when traditional cervical collars were implemented. However, all tested cervical collars showed remaining motion of the cervical spine. Thus, alternative immobilization techniques should be considered.

Introduction

Although evidence of the effectiveness of cervical collars is rare1,2,3, the use of rigid cervical collars in preclinical trauma care is known to be a common immobilization technique. The importance of immobilization of the cervical spine is recognized in national and international recommendations for initial trauma care. The application of the cervical collar to avoid secondary injuries of the cervical spine4,5 is recommended as one of the first emergency measures6,7.

Nevertheless, immobilization of the cervical spine by rigid collars is controversially discussed in the literature8,9,10,11,12. Cervical spine immobilization with collars can be accompanied by disadvantages, such as pain, discomfort13,14 or pressure ulceration13,15,16. Furthermore, the application of a cervical collar can also lead to considerably serious and even life-threatening complications. In particular, restricted respiratory function17 and difficult airway management18 have been described, which can quickly lead to severe complications or even death, especially in trauma patients7. To ease airway management, an innovative cervical collar has been developed that should combine cervical spine immobilization and airway protection. This innovative cervical collar is equipped with a flexible belt for the mandible19. While this innovative collar seems to succeed in airway protection, the quality of cervical spine immobilization has not been proven, so far. However, the effectiveness of a cervical collar depends on its perfect fit20,21,22.

The aim of this study was to compare the remaining motion of the cervical spine using an innovative cervical collar as well as two traditional cervical collars on fresh human cadavers.

Materials and methods

Study design

The study was performed on fresh human cadavers. The body donors were given detailed information before death and had to give their written informed consent to the body donation. After death, the body was made available for scientific research purposes. Fresh human cadavers were briefly frozen after death. For the experiments, the bodies were thawed to room temperature. This method enabled the simulation of the joint elasticity and soft tissue situation of a living body. Current biomechanical studies do not describe any significant difference in cervical spine motion between fresh human cadavers and patients4,23,24.

The inclusion criteria for the current study were as follows: (1) existing written informed consent for body donation for scientific research; (2) no injuries, diseases or operations on the cervical spine; and (3) complete medical records. The complete medical history and the existing medical data of all examined patients were analyzed. Patients with diseases such as tumors, thyroid diseases or similar conditions were excluded from the study.

The local ethics committee reviewed and approved the present study (Ethics Committee, Mainz, Rhineland-Palatinate, Germany, ID: 837.156.16). The study was registered in the German Clinical Trials Register (ID: DRKS00010499). All experiments were performed in accordance with the relevant guidelines and regulations.

Cervical collars

In the present study, three different cervical collars were tested on fresh human cadavers. First, the innovative Weinmann Lubo Airway Collar (Weinmann Emergency Medical Technology GmBh and Co. KG, Hamburg, Germany, Fig. 1A) was tested. This collar was newly developed in 2007 to stabilize the cervical spine and to avoid supraglottic airway obstruction at the same time (Fig. 1D–F) 19. Second, two traditional collars were tested, namely, the Laerdal Stifneck (Laerdal Medical GmbH, Puchheim, Germany, Fig. 1B) and the Ambu Perfit ACE (Ambu GmbH, Bad Nauheim, Germany, Fig. 1C). These traditional rigid collars with one-belt fixation have been in use nearly worldwide for many decades.

Figure 1
figure 1

In the present study, three different cervical collars were tested on fresh human cadavers: Weinmann Lubo Airway Collar (A), Laerdal Stifneck (B) and Ambu Perfit ACE (C). Shown is the innovative mechanism (D,E,F). The red angle (black dot) pushes the mandible ventrally (black arrow) for opening the upper airway.

All cervical collars were adjusted according to the manufacturers' instructions. All different sizes of cervical collars were available during the study. The three cervical collars (Fig. 1A–C) were applied by one experienced emergency medical service (EMS) person to fresh human cadavers. Likewise, the repetitive measurements were performed only by one person experienced in rescue medicine.

Cervical spine motion measurement

In the present study, the remaining flexion and the remaining lateral bending of the immobilized cervical spine were measured with a wireless human motion tracker system (Xsens Technologies, Enschede, Netherlands). Compared to other methods, this measurement method with a motion tracker has been proven25,26,27 and guarantees exact measurement results28. The endpoints of the measurements were maximal flexion and maximal lateral bending.

In the standardized motion protocol, the rotation was not included separately since combination motions always occur here. The extension was also not recorded since the fresh human cadavers were lying in the supine position and thus no extension could be measured (Fig. 2).

Figure 2
figure 2

A fresh human cadaver is positioned supine on a spine board with a cervical collar; the motion trackers are fixed to the forehead and to the sternum.

In the experimental setup, fresh human cadavers were placed on a spineboard (Laerdal BaXstrap, Stavanger, Norway) in supine position and fixed by a harness fixation system (MIH-Medical Spiderstrap, Georgsmarienhütte, Germany). One motion tracker was attached to the forehead, and one motion tracker was attached to the thorax of the fresh human cadaver (Fig. 2). The 3D data were synchronized every 20 microseconds by the Xsens recording tool. However, with the help of the motion trackers, rotation and extension were recorded in the respective measured direction of motion (Fig. 3A).

Figure 3
figure 3

Remaining flexion of the cervical spine if an innovative cervical collar is applied (A). Range of remaining flexion of the cervical spine (B) and range of remaining lateral bending (C) of the three tested cervical collars (B,C: Only significant differences are marked).

The head of the fresh human cadaver was passively moved ventrally for flexion (Fig. 3A) and to the right for lateral bending. The traction force was 100 N. This force is in the range and in the direction as it may be applied during intubation or patient transport and may therefore initiate or aggravate an assumed spinal injury3,29,30. The force was measured with an electronic spring balance (LENI, Fa. Korona, Sundern, Germany).

In the standardized motion protocol, flexion and lateral bending were first analyzed for each fresh human cadaver without the cervical collar (control group). Subsequently, the motion analysis was performed with the three cervical collars in place.

Statistical data analysis

In the present study, statistical calculations were performed to determine possible differences between the three tested cervical collars. With the D'Agostino and Pearson test, the three groups were tested for normal distribution. The statistical comparison between the four groups was performed using analysis of variance (ANOVA). A p-value of < 0.05 was considered statistically significant.

An exploratory analysis of the measured values was performed. For all measurements, descriptive data with the mean, standard deviation and ranges are given. The statistical data analysis was performed with GraphPad PRISM software (Version 8.2.1, San Diego, California, USA).

Results

Study collective

Eight fresh human cadavers were included in the present study. The mean age of the cadavers at death was 79.0 ± 12.8 years (range: 56–94 years). The study included five female and three male fresh human cadavers.

Motion measurement of the cervical spine

The highest remaining flexion of the cervical spine with a cervical collar applied was 34.0° (Fig. 3A). In this case, Weinman Lubo Airway Collar (Fig. 1A) was attached. Overall, the fresh human cadavers with the Weinman Lubo Airway Collar in place had a mean flexion of 20.0 ± 9.0° (range: 10.0–34.0°; Fig. 3B; Table 1). A lower remaining flexion at the cervical spine could be achieved using the two traditional cervical collars (Fig. 1B,C). With the Laerdal Stifneck applied, the mean flexion was 13.1 ± 6.6° (range: 8.0–24.0°). The mean flexion was lowest with the Ambu Perfit ACE (10.8 ± 5.8°; range: 5.0–19.0°; Fig. 3B). Without cervical collar, the fresh human cadavers had a mean cervical flexion of 25 ± 5.0° (range: 18.0°–34.0°, Fig. 3B).

Table 1 Results of the remaining motion examination of the three cervical collars and the control group.

The mean flexion of the cervical spine with the Weinmann Lubo Airway Collar on was almost twice as high as that with the Ambu Perfit ACE applied. Compared to the Weinmann Lubo Airway Collar, the remaining cervical spine flexion was significantly (p = 0.0474) decreased with the Ambu Perfit ACE (Fig. 3B). The differences in remaining cervical spine flexion between the Laerdal Stifneck and the Weinmann Lubo Airway Collar (p = 0.1655) and between the Laerdal Stifneck and the Ambu Perfit ACE (p = 0.7923) were not significant. There was no significant difference in cervical flexion with Weinmann Lubo Airway Collar applied or no collar applied (p = 0.3611). Between the control examination and the Laerdal Stiffneck (p = 0.0049) and the Ambu Perfit ACE (p = 0.0008) there was a significant reduction in remaining cervical spine flexion.

The mean remaining lateral bending of the Weinmann Lubo Airway Collar was 17.8 ± 10.2° (range: 7.0–35.0°; Table 1). With the Laerdal Stifneck attached, the mean remaining lateral bending was 12.0 ± 4.8° (range: 7.0–19.0°). The Ambu Perfit ACE allowed for a mean remaining lateral bending of 12.1 ± 4.3° (range: 6.0–17.0°; Figs. 1C and 3C). Without cervical collar, the lateral bending of the cervical spine was 37.5 ± 8.7° (range: 29.0°–55.0°, Fig. 3C).

There was no significant difference in lateral bending between the three examined collars (Weinmann Lubo Airway Collar vs. Ambu Perfit ACE: p = 0.2600; Weinmann Lubo Airway Collar vs. Laerdal Stifneck: p = 0.2458; Laerdal Stifneck vs. Ambu Perfit ACE: p = 0.9993; Fig. 3C). Lateral bending of the cervical spine was significantly reduced by all three cervical collars (p < 0.0001).

Discussion

The aim of this study was to measure and to evaluate quantitatively the remaining motion of the cervical spine during immobilization using one innovative upper airway opening cervical collar and two traditional cervical collars. The present study showed that the attached innovative and traditional cervical collars resulted in significantly different remaining flexion of the immobilized cervical spine. The design of the cervical collars seems to have an influence on their fit and thus on the effectiveness of the immobilization.

The data of the present study showed that compared to traditional cervical collars, the innovative cervical collar allows for twofold increased flexion of the cervical spine. There was no significant difference in the cervical spine flexion whether the innovative cervical collar was applied or not. The novel upper airway opening design is probably decisive for the poor result. The innovative cervical collar has a one-belt system; whereby the mandible is addressed with an extra flexible belt (Fig. 1A). In contrast, both traditional cervical collars address the mandible as well as the occiput, sternum, clavicle, shoulders and upper back with their rigid form (Fig. 1B,C)19,21,31. The novel fixation system therefore seems to perform worse than the classic cervical collars in the important flexion motion. The Weinmann Lubo Airway Collar can support airway clearance, but a classic cervical collar should be used in unconscious and intubated patients with suspected cervical spine injury.

According to a study by Lubovsky et al., the innovative Weinmann Lubo Airway Collar, which combines an external airway protector in combination with a cervical collar, appears to be safe and effective in opening and maintaining open airways in anesthetized supine patients19. In addition, the authors recommend that further clinical trials are indicated to evaluate the safety and utility of this device under preclinical conditions19. To our knowledge, the efficacy of the collar toward cervical spine immobilization has not been tested so far. According to the data of the current study, efficiency of the innovative cervical collar by the means of cervical spine immobilization is worse compared to traditional cervical collars.

The study by Karason et al. also compared widely used cervical collars32. The endpoints of the remaining motion were analyzed using a gonimeter. In addition, the intracranial pressure was extrapolated, and the wearing comfort was evaluated subjectively. As in the present study, the classic rigid cervical collars were shown to achieve safe immobilization. Karason et al. were able to determine a remaining motion of immobilization of the cervical spine of 18.0° ± 7.0°. The present study showed significantly better immobilization values. The extent to which this depends on the measurement method used or is even due to the test subjects should be analyzed in further studies.

The motion values that have been measured for the traditional cervical collars were in the same range as in other studies. Hostler et al. found a mean flexion of 9.2° ± 5.0° with the Ambu Perfit ACE applied33. Thus, complete immobilization of the cervical spine could not be achieved by a cervical collar alone. These findings are supported by various other studies3,34. In addition, many cervical collars are not applied correctly35; therefore, their effectiveness is further reduced.

Since a secondary injury or a worsening of an existing injury cannot be completely excluded in cases of remaining cervical spine motion4,36, alternative immobilization techniques should be considered. Cervical spine immobilization by a collar could be substantially improved by additionally fixing the patient to a spine board34. However, in this case, the main immobilization will derive from fixation on the spineboard, and the cervical collar is dispensable37,38. The full body immobilization technique is especially recommended during patient transport25.

Limitations

The present study is limited to some extent. Only human cadavers with an intact cervical spine were analyzed. For patients with an already injured cervical spine, different results can certainly be expected. Furthermore, the cervical spine force was applied in a standardized manner only for flexion and lateral bending. Combined movements were not performed. There is no literature that provides any indication of the magnitude and direction of the force required to cause or even to aggravate a cervical spine injury. Previous studies have shown that after an initial trauma, no further injury or aggravation can occur through further manipulation39. However, recent biomechanical studies have shown a direct correlation between manipulation of the severely injured cervical spine and the width of the dural sac29,30,40. Furthermore, the application of a cervical collar itself may aggravate a spinal injury29. Thus, in future studies, not only the remaining cervical spine motion but also the manipulation that is caused by the collar’s application should be analyzed. Here, significant differences have been previously shown41. In addition, it should always be remembered that the cervical support immobilizes best when other immobilization tools are applied26. Three cervical collars were examined in the current study; and, according to the literature, many additional cervical collars are available33. This efficiency of these other collars should be addressed in further studies.

Conclusions

The results of the present study indicate that the best immobilization of the cervical spine could be achieved if the cervical collar has a nonmobile, rigid design that is closely attached to the mandible, occiput, sternum, clavicle, shoulders and upper back. However, all tested cervical collars showed remaining motion of the cervical spine. Thus, alternative immobilization techniques should be considered.

References

  1. Hood, N. & Considine, J. Spinal immobilisaton in pre-hospital and emergency care: A systematic review of the literature. Australas. Emerg. Nurs. J. 18, 118–137. https://doi.org/10.1016/j.aenj.2015.03.003 (2015).

    Article  PubMed  Google Scholar 

  2. Peck, G. E., Shipway, D. J. H., Tsang, K. & Fertleman, M. Cervical spine immobilisation in the elderly: A literature review. Br. J. Neurosurg. 32, 286–290. https://doi.org/10.1080/02688697.2018.1445828 (2018).

    Article  PubMed  Google Scholar 

  3. Horodyski, M., DiPaola, C. P., Conrad, B. P. & Rechtine, G. R. Cervical collars are insufficient for immobilizing an unstable cervical spine injury. J. Emerg. Med. 41, 513–519. https://doi.org/10.1016/j.jemermed.2011.02.001 (2011).

    Article  PubMed  Google Scholar 

  4. Hindman, B. J. et al. Intubation biomechanics: Laryngoscope force and cervical spine motion during intubation in cadavers-cadavers versus patients, the effect of repeated intubations, and the effect of type II odontoid fracture on C1–C2 motion. Anesthesiology 123, 1042–1058. https://doi.org/10.1097/aln.0000000000000830 (2015).

    Article  PubMed  Google Scholar 

  5. Benger, J. & Blackham, J. Why do we put cervical collars on conscious trauma patients?. Scand. J. Trauma Resusc. Emerg. Med. 17, 44. https://doi.org/10.1186/1757-7241-17-44 (2009).

    Article  PubMed  PubMed Central  Google Scholar 

  6. NAEMT. Prehospital Trauma Life Support (PHTLS) 7th edn. (Prehospital Trauma Life Support Committee of the National Association of Emergency Medical Technicians in Cooperation with the Committee on Trauma of The American College of Suregons, 2010).

    Google Scholar 

  7. Surgeons, A. C. O. Advanced Trauma Life Support (ATLS®): The Ninth Edition. 2013/04/24 edn (2013).

  8. Hess, T. et al. Pre-hospital care management of acute spinal cord injury. Anasthesiol. Intensivmed. Notfallmed. Schmerzther. 51, 226–237. https://doi.org/10.1055/s-0042-105016 (2016).

    Article  PubMed  Google Scholar 

  9. Walters, B. C. et al. Guidelines for the management of acute cervical spine and spinal cord injuries: 2013 update. Neurosurgery 60, 82–91. https://doi.org/10.1227/01.neu.0000430319.32247.7f (2013).

    Article  PubMed  Google Scholar 

  10. Unfallchirurgie, D. G. F. S3 Guideline on Treatment of Patients with Severe and Multiple Injuries. English Version of the German Guideline S3 Leitlinie Polytrauma/Schwerverletzten-Behandlung AWMF Register-Nr. 012/019. https://www.awmf.org/uploads/tx_szleitlinien/012-019l_S3_Polytrauma_Schwerverletzten-Behandlung_2017-08.pdf (2016).

  11. Centre, N. C. G. Spinal Injury: Assessment and Initial Management. NICE Guideline, No. 41 (2016).

  12. Kwan, I., Bunn, F. & Roberts, I. Spinal immobilisation for trauma patients. Cochrane Database Syst. Rev. 2001, Cd002803. https://doi.org/10.1002/14651858.Cd002803 (2001).

    Article  PubMed Central  Google Scholar 

  13. Cordell, W. H., Hollingsworth, J. C., Olinger, M. L., Stroman, S. J. & Nelson, D. R. Pain and tissue-interface pressures during spine-board immobilization. Ann. Emerg. Med. 26, 31–36. https://doi.org/10.1016/s0196-0644(95)70234-2 (1995).

    CAS  Article  PubMed  Google Scholar 

  14. Swartz, E. E. et al. Prehospital cervical spine motion: Immobilization versus spine motion restriction. Prehosp. Emerg. Care 22, 630–636. https://doi.org/10.1080/10903127.2018.1431341 (2018).

    Article  PubMed  Google Scholar 

  15. Ham, W., Schoonhoven, L., Schuurmans, M. J. & Leenen, L. P. Pressure ulcers from spinal immobilization in trauma patients: A systematic review. J. Trauma Acute Care Surg. 76, 1131–1141. https://doi.org/10.1097/ta.0000000000000153 (2014).

    Article  PubMed  Google Scholar 

  16. Pernik, M. N. et al. Comparison of tissue-interface pressure in healthy subjects lying on two trauma splinting devices: The vacuum mattress splint and long spine board. Injury 47, 1801–1805. https://doi.org/10.1016/j.injury.2016.05.018 (2016).

    Article  PubMed  Google Scholar 

  17. Totten, V. Y. & Sugarman, D. B. Respiratory effects of spinal immobilization. Prehosp. Emerg. Care 3, 347–352. https://doi.org/10.1080/10903129908958967 (1999).

    CAS  Article  PubMed  Google Scholar 

  18. Goutcher, C. M. & Lochhead, V. Reduction in mouth opening with semi-rigid cervical collars. Br. J. Anaesth. 95, 344–348. https://doi.org/10.1093/bja/aei190 (2005).

    CAS  Article  PubMed  Google Scholar 

  19. Lubovsky, O., Liebergall, M., Weissman, C. & Yuval, M. A new external upper airway opening device combined with a cervical collar. Resuscitation 81, 817–821. https://doi.org/10.1016/j.resuscitation.2010.02.013 (2010).

    Article  PubMed  Google Scholar 

  20. White, A. A. & Panjabi, M. M. Clinical Biomechanics of the Spine 2nd edn. (J. B. Lippincott & Co., 1990).

    Google Scholar 

  21. Bell, K. M. et al. Assessing range of motion to evaluate the adverse effects of ill-fitting cervical orthoses. Spine J. 9, 225–231. https://doi.org/10.1016/j.spinee.2008.03.010 (2009).

    Article  PubMed  Google Scholar 

  22. Ben-Galim, P. et al. Extrication collars can result in abnormal separation between vertebrae in the presence of a dissociative injury. J. Trauma 69, 447–450. https://doi.org/10.1097/TA.0b013e3181be785a (2010).

    Article  PubMed  Google Scholar 

  23. Lennarson, P. J. et al. Segmental cervical spine motion during orotracheal intubation of the intact and injured spine with and without external stabilization. J. Neurosurg. 92, 201–206. https://doi.org/10.3171/spi.2000.92.2.0201 (2000).

    CAS  Article  PubMed  Google Scholar 

  24. McCahon, R. A. et al. Cadaveric study of movement of an unstable atlanto-axial (C1/C2) cervical segment during laryngoscopy and intubation using the Airtraq(®), Macintosh and McCoy laryngoscopes. Anaesthesia 70, 452–461. https://doi.org/10.1111/anae.12956 (2015).

    CAS  Article  PubMed  Google Scholar 

  25. Nolte, P. C. et al. Analysis of cervical spine immobilization during patient transport in emergency medical services. Eur. J. Trauma Emerg. Surg. https://doi.org/10.1007/s00068-019-01143-z (2019).

    Article  PubMed  Google Scholar 

  26. Uzun, D. D. et al. Remaining cervical spine movement under different immobilization techniques. Prehosp. Disaster Med. https://doi.org/10.1017/s1049023x2000059x (2020).

    Article  PubMed  Google Scholar 

  27. Haske, D. et al. An explorative, biomechanical analysis of spine motion during out-of-hospital extrication procedures. Injury 51, 185–192. https://doi.org/10.1016/j.injury.2019.10.079 (2020).

    Article  PubMed  Google Scholar 

  28. Weerts, J. O. N., Schier, L., Schmidt, H. & Kreinest, M. Review of existing measurement tools to assess spinal motion during prehospital immobilization. Eur. J. Emerg. Med. 25, 161–168. https://doi.org/10.1097/mej.0000000000000467 (2018).

    Article  PubMed  Google Scholar 

  29. Liao, S. et al. Motion and dural sac compression in the upper cervical spine during the application of a cervical collar in case of unstable craniocervical junction—A study in two new cadaveric trauma models. PLoS ONE 13, e0195215. https://doi.org/10.1371/journal.pone.0195215 (2018).

    CAS  Article  PubMed  PubMed Central  Google Scholar 

  30. Liao, S. et al. Spinal movement and dural sac compression during airway management in a cadaveric model with atlanto-occipital instability. Eur. Spine J. 27, 1295–1302. https://doi.org/10.1007/s00586-017-5416-9 (2018).

    Article  PubMed  Google Scholar 

  31. Worsley, P. R., Stanger, N. D., Horrell, A. K. & Bader, D. L. Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin. Med. Dev. (Auckl.) 11, 87–94. https://doi.org/10.2147/mder.S149419 (2018).

    CAS  Article  Google Scholar 

  32. Karason, S., Reynisson, K., Sigvaldason, K. & Sigurdsson, G. H. Evaluation of clinical efficacy and safety of cervical trauma collars: Differences in immobilization, effect on jugular venous pressure and patient comfort. Scand. J. Trauma Resusc. Emerg. Med. 22, 37. https://doi.org/10.1186/1757-7241-22-37 (2014).

    Article  PubMed  PubMed Central  Google Scholar 

  33. Hostler, D., Colburn, D. & Seitz, S. R. A comparison of three cervical immobilization devices. Prehosp. Emerg. Care: Off. J. Natl. Assoc. EMS Phys. Natl. Assoc. State EMS Dir. 13, 256–260. https://doi.org/10.1080/10903120802706195 (2009).

    Article  Google Scholar 

  34. Chandler, D. R., Nemejc, C., Adkins, R. H. & Waters, R. L. Emergency cervical-spine immobilization. Ann. Emerg. Med. 21, 1185–1188 (1992).

    CAS  Article  Google Scholar 

  35. Kreinest, M. et al. Application of cervical collars—An analysis of practical skills of professional emergency medical care providers. PLoS ONE 10, e0143409. https://doi.org/10.1371/journal.pone.0143409 (2015).

    CAS  Article  PubMed  PubMed Central  Google Scholar 

  36. Donaldson, W. F., Heil, B. V., Donaldson, V. P. & Silvaggio, V. J. The effect of airway maneuvers on the unstable C1–C2 segment. A cadaver study. Spine (Phila Pa 1976) 22, 1215–1218. https://doi.org/10.1097/00007632-199706010-00008 (1997).

    Article  Google Scholar 

  37. Holla, M. Value of a rigid collar in addition to head blocks: A proof of principle study. Emerg. Med. J. 29, 104–107. https://doi.org/10.1136/emj.2010.092973 (2012).

    CAS  Article  PubMed  Google Scholar 

  38. Butler, J. & Bates, D. Towards evidence based emergency medicine: Best BETs from the Manchester Royal Infirmary. Cervical collars in patients requiring spinal immobilisation. Emerg. Med. J.: EMJ 18, 275 (2001).

    CAS  Article  Google Scholar 

  39. Hauswald, M. & Braude, D. Spinal immobilization in trauma patients: Is it really necessary?. Curr. Opin. Crit. Care 8, 566–570. https://doi.org/10.1097/00075198-200212000-00014 (2002).

    Article  PubMed  Google Scholar 

  40. Liao, S. et al. Cadaveric study of movement in the unstable upper cervical spine during emergency management: Tracheal intubation and cervical spine immobilisation-a study protocol for a prospective randomised crossover trial. BMJ Open 7, e015307. https://doi.org/10.1136/bmjopen-2016-015307 (2017).

    Article  PubMed  PubMed Central  Google Scholar 

  41. James, C. Y., Riemann, B. L., Munkasy, B. A. & Joyner, A. B. Comparison of cervical spine motion during application among 4 rigid immobilization collars. J. Athl. Train. 39, 138–145 (2004).

    PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Professor Dr. Erik Popp for his technical support in the preparation of the fresh human cadavers. We would like to thank the staff of the Anatomical Institute of the University of Heidelberg for providing the premises and for their support in handling fresh human cadavers. We would also like to thank Mr. Geir Dillan for taking the photographs.

Funding

There was no funding for this study.

Author information

Authors and Affiliations

Authors

Contributions

Conceptualization, M.K.J., N.R.E.S. and M.K.; Data curation, M.K.J., G.V. R.E. and M.K.; Formal analysis, H.K.; Investigation, G.V.R.E. and M.K.; Methodology, M.K.J., P.A.G. and M.K.; Project administration, M.K.J. and M.K.; Resources, H.K., N.R.E.S. and M.K.; Supervision, P.A.G. and M.K.; Validation, M.K.J., G.V. R.E. and M.K.; Visualization, M.K.J. and N.R.E.S.; Writing—original draft, M.K.J.; Writing—review and editing, M.K.J., G.V. R.E., P.A.G., H.K., N.R.E.S. and M.K.

Corresponding author

Correspondence to Michael Kreinest.

Ethics declarations

Competing interests

The authors declare no competing interests.

Additional information

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and Permissions

About this article

Verify currency and authenticity via CrossMark

Cite this article

Jung, M.K., von Ehrlich-Treuenstätt, G.V.R., Keil, H. et al. Analysis of remaining motion using one innovative upper airway opening cervical collar and two traditional cervical collars. Sci Rep 11, 20619 (2021). https://doi.org/10.1038/s41598-021-00194-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1038/s41598-021-00194-w

Further reading

Comments

By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.

Search

Quick links

Nature Briefing

Sign up for the Nature Briefing newsletter — what matters in science, free to your inbox daily.

Get the most important science stories of the day, free in your inbox. Sign up for Nature Briefing