Abstract
Magnetic resonance fingerprinting (MRF) is a rapidly developing fast quantitative mapping technique able to produce multiple property maps with reduced sensitivity to motion. MRF has shown promise in improving the diagnosis of clinically significant prostate cancer but requires further validation as part of a prostate multiparametric (mp) MRI protocol. mpMRI protocol mandates the inclusion of dynamic contrast enhanced (DCE) imaging, known for its significant T1 shortening effect. MRF could be used to measure both pre- and post-contrast T1 values, but its utility must be assessed. In this proof-of-concept study, we sought to evaluate the variation in MRF T1 measurements post gadolinium-based contrast agent (GBCA) injection and the utility of such T1 measurements to differentiate peripheral and transition zone tumours from normal prostatic tissue. We found that the T1 variation in all tissues increased considerably post-GBCA following the expected significant T1 shortening effect, compromising the ability of MRF T1 to identify transition zone lesions. We, therefore, recommend performing MRF T1 prior to DCE imaging to maintain its benefit for improving detection of both peripheral and transition zone lesions while reducing additional scanning time. Demonstrating the effect of GBCA on MRF T1 relaxometry in patients also paves the way for future clinical studies investigating the added value of post-GBCA MRF in PCa, including its dynamic analysis as in DCE-MRF.
Similar content being viewed by others
Introduction
Prostate cancer (PCa) is the second commonest male malignancy worldwide with multiparametric (mp) MRI now recommended by major European and American guidelines as the first-line investigation for patients with suspected early stage disease1,2,3,4. The current Prostate Imaging Reporting and Data System (PI-RADS) guidelines only incorporate qualitative measures for interpretation, however, quantitative metrics have been suggested as a means of reducing the considerable interobserver variation of PI-RADS evaluation, shortening the learning curve of mpMRI, and improving diagnostic performance5,6,7,8.
Magnetic resonance fingerprinting (MRF) is a quantitative technique able to simultaneously generate multiple inherently spatially registered property maps (e.g. T1, T2, apparent proton-density). Quantitative mapping provided by MRF has demonstrated high reproducibility between centers over standard T1 or T2 mapping techniques that can have a system dependence9,10. These maps can be obtained in the presence of motion while being acquired within imaging times comparable to or faster than conventional mapping techniques11,12. The described features of MRF present a particular interest for cancer imaging, where fast and robust quantitative characterization of tissue biology can add value to routinely used qualitative measures for image assessment12.
MRF has shown promise for identifying both peripheral zone (PZ) and transition zone (TZ) prostate lesions, demonstrating added value to standard mpMRI sequences for differentiating between indolent and clinically significant disease13,14,15. Further prospective validation of MRF requires additional evaluation of how it can be incorporated into a standard clinical prostate mpMRI protocol, which includes dynamic contrast enhanced (DCE) imaging3. Gadolinium is known to have a preferential T1 shortening effect at low doses, which was confirmed for MRF in pre-clinical studies involving a murine glioblastoma model16,17. However, no attempts have been made to investigate the impact of gadolinium-based contrast agents (GBCA) on MRF-based T1 relaxometry within patients in the clinical setting.
For prostate imaging, understanding the potential added value and robustness of post-GBCA MRF is of considerable practical interest for two major reasons. Firstly, should MRF be incorporated into the clinical mpMRI protocol, the decision on its running order in relation to DCE-MRI should be evidenced and balanced against the increasing trend towards reducing scanning time18,19,20,21. Secondly, post-GBCA MRF T1 mapping may improve the performance of DCE-MRI in TZ lesions, where it is currently of limited use in the context of adequate T2-weighted imaging and diffusion-weighted imaging22,23. Finally, investigating the robustness of MRF T1 relaxometry post GBCA in the clinical setting would also be relevant to cancers located in other anatomical regions where MRF has also shown promising results and the use of GBCA is routine.
Therefore, in this proof-of-concept study we sought to evaluate the variation in MRF T1 measurements post GBCA administration and evaluate its impact on the technique’s ability to differentiate between tumor and normal tissue in patients with biopsy-proven PZ and TZ prostate lesions. To ensure the robustness of the MRF technique used, we also validated it against “gold standard” quantitative mapping techniques as part of a phantom study.
Results
Phantom results
Figure 1a,b shows the mean MRF T1 values obtained from the ISMRM/NIST phantom plotted against T1 values obtained using the “gold standard” inversion recovery fast spin echo (IR-FSE) imaging and 3D variable flip angle (VFA) T1 mapping. The results show a strong linear correlation between MRF and IR-FSE (R2 = 0.996, p ≤ 0.0001, slope = 0.991) and a slightly weaker correlation between MRF and VFA (R2 = 0.975, p ≤ 0.0001, slope = 0.732). The comparison of slopes of the linear fits (presented alongside y-intercepts in Fig. 1a,b) suggests better performance of MRF compared to VFA in the phantom setting.
Figure 1c demonstrates the mean phantom MRF T2 values plotted against the values obtained using the “gold standard” multiple spin echo (MSE) T2 mapping; slopes and y-intercepts are also shown. Although still significant, the correlation between the values is considerably lower than for both T1 method comparisons presenting as parabolic rather than linear relationship (R2 = 0.9129, p ≤ 0.0001), suggesting lower reliability of MRF T2 mapping used in this study.
Bland–Altman plots were constructed to evaluate the agreement between the aforementioned techniques and are presented in Fig. 1d–f. The mean bias and the 95% limits of agreement (LOA) for T1 values obtained using IR-FSE and MRF and VFA and MRF are presented in Fig. 1d,e, respectively, while the same parameters for T2 values obtained using MSE and MRF are shown in Fig. 1f. One data point with the longest T1 was outside the LOA for IR-FSE vs MRF.
Patient characteristics
The study included 14 patients with biopsy-proven PCa with mean age 70 years (IQR, 67.3–73.5 years), mean PSA 6.29 ng/mL (IQR, 3.8–8.7 ng/mL), with mean time since last biopsy being 16 months (range 4–48 months). A total of 19 MR-visible prostate lesions were included in the analysis, 10 of which were located in the peripheral zone (PZ) and 9 in the transition zone (TZ). Three lesions exhibited intermediate-grade Gleason score 3 + 4 = 7 disease (grade group 2) while other lesions harboured low-grade disease with Gleason score of 3 + 3 = 6 (grade group 1).
In vivo agreement between MRF-, VFA- and MSE-based T1 and T2 relaxometry
Summary pre- and post-GBCA MRF- and pre-GBCA VFA- and MSE-based T1 and T2 along with ADC values obtained from all prostate lesions combined (n = 19), PZ (n = 10) and TZ (n = 9) lesions, pooled nPZ and nTZ, internal obturator muscle and subcutaneous fat (n = 14 for all) are presented in Table 1. Bland–Altman analysis showed lower agreement between pre-GBCA MRF- and VFA-based T1 relaxation times compared to the phantom experiment, which is to be expected given the physiological motion in vivo, with the mean bias being 410.5 ms and the 95% LOA ranging between − 1171.0 and 1192.0 ms as demonstrated in Fig. 2a. A similar trend was revealed when the Bland–Altman analysis was used to evaluate the agreement between pre-GBCA MRF- and MSE-based T2 relaxation times, with the mean bias being − 281.2 ms and 95% LOA ranging between − 727.5 ms and 165.1 ms (Fig. 2b).
MRF T1 variation post GBCA
Table 2 presents coefficients of variation (CVs) calculated for all acquired values to compare their variation in different tissues. Pre-GBCA MRF T1 demonstrated low variation in all tissues except for nTZ, where it reached 25.4%. (Fig. 3a) Post-GBCA, MRF T1 variation was above 25% in all tissue types except nPZ, muscle and fat, where CVs remained in the same category as pre-GBCA. A particularly marked, almost five-fold increase in data heterogeneity was observed for MRF T1 values obtained from TZ lesions and, to a lesser extent nTZ, whereas PZ lesions demonstrated only a two-fold increase in CV and only a marginal change in variation was noted in nPZ. (Fig. 3b; Table 2).
MRF T2 variation post GBCA
In line with the outcomes of the Bland–Altman analysis indicating lower robustness of MRF T2 compared to “gold standard” T2 mapping, pre-GBCA MRF T2 values were considerably more variable compared to pre-GBCA MRF T1 values in all tissues except fat, further suggesting low reliability of in vivo MRF T2 mapping used in this study. The variation of MRF T2 also increased considerably post-GBCA reaching 113.7% in nTZ (Table 2).
MRF-based T1 relaxometry for differentiating tumour and normal tissue
Prior to GBCA administration, a paired t test revealed significantly shorter MRF T1 values for both peripheral and transition zone lesions when compared to corresponding nPZ and nTZ in the same patients (1640.0 ms ± 368.1 ms vs 2200.0 ms ± 776.5 ms for PZ and 1696.0 ms ± 200.5 ms vs 1966.0 ms ± 315.1 ms for TZ; p = 0.03 and 0.013, respectively) (Fig. 4a). In pooled nPZ, MRF T1 relaxation time was significantly longer than in nTZ (2521.0 ms ± 405.9 ms vs 1753.0 ms ± 444.7 ms; p < 0.0001). Post-GBCA MRF T1 remained significantly shorter within peripheral zone lesions compared to the corresponding normal PZ (678.4 ms ± 287.9 ms vs 1317.0 ms ± 219.6 ms; p < 0.0001), however, there was no longer a significant difference in TZ tumours compared to corresponding nTZ (723.8 ms ± 407.3 ms vs 966.4 ms ± 635.5 ms, p = 0.207) (Fig. 4b). Pooled nTZ T1 relaxation time was again significantly shorter than those of nPZ (1270 ms ± 224.6 ms vs 723.8 ms ± 407.3 ms; p < 0.0001). Paired t test showed a significant MRF T1 shortening effect of GBCA in all tissues (Fig. 5).
The information about the diagnostic utility of MRF- and MSE-based T2, VFA-based T1 and ADC mapping is provided in the Supplementary Information S1.
Discussion
This prospective, proof-of-concept study demonstrates the effect of gadolinium-based contrast agent administration on MRF-based T1 relaxometry in the clinical setting. We have shown that GBCA considerably increases MRF T1 variation in both normal and malignant prostate tissues and compromises its diagnostic utility in the transition zone. To our knowledge, this is the first study reporting both pre- and post-GBCA MRF T1 and T2 values as well as a combination of MRF-, VFA- and MSE-based T1 and T2 values obtained from the same patients with prostate cancer. These results will help inform future studies, when MRF may be incorporated into prostate mpMRI protocols as an additional sequence.
The observed MRF T1 shortening effect post-GBCA is expected, as gadolinium facilitates both longitudinal and transverse magnetic relaxation, thereby shortening both T1 and T2 of tissues24,25. Other authors also observed a similar trend when measuring gadolinium and dysprosium concentrations in murine glioma models using dual contrast-MRF. 16,26 The observed GBCA-induced increase in MRF T1 heterogeneity varied between tissues. Normal TZ exhibited the greatest variation pre-GBCA, which is expected in this age group, given the high prevalence of benign prostatic hyperplasia (BPH). Marked hypervascularity within BPH nodules may additionally explain the more marked (two-fold) increase in MRF T1 heterogeneity in nTZ tissue following GBCA administration22. The finding of increased MRF T1 variation was even more marked for TZ lesions (almost five-fold), which likely reflects previously reported large variation in their microvascular parameters and explains the inability of MRF T1 to identify TZ lesions post-GBCA27. Although a study by Panda et al. has shown the added value of pre-GBCA MRF T2WI, which is considered the primary sequence for assessment of the TZ28,29,30, the reduced performance of post-GBCA MRF T1 in assessing TZ lesions should be considered when deciding on its timing in relation to DCE. Conversely, normal PZ, fat and muscle, which can be considered as “control” type tissues given their relatively low vascularity and morphological homogeneity in this age group, maintained low MRF T1 heterogeneity post-GBCA. Hence, understanding the rationale for acquiring MRF T1 prior to GBCA administration would not only ensure its optimal performance for both TZ and PZ assessment but also enable evidenced planning of the overall scanning time, which is critical due to the growing demand on imaging services.
We also demonstrated that pre-GBCA MRF T1 relaxation times were significantly lower in cancers compared to normal tissue in both the TZ and PZ of the prostate. These findings align well with previous studies where a combination of MRF and ADC maps worked best for identifying PZ and TZ lesions; our MRF T1 and ADC absolute values are similar to these reported values13,14,15. Our pre-GBCA T1 values obtained from PZ and TZ lesions are comparable to those reported previously in clinically significant PCa (1628.0 ms ± 344.0 ms vs 2247.0 ms ± 450 ms for PZ lesions and nPZ; 1450 ms ± 110 ms vs 1800 ms ± 150 ms for TZ lesions and nTZ)13,14 and consistent with those reported by Yu et al.15 for low-grade tumours that dominated the sample size in our study (1679.0 ms ± 422 ms). Although we showed low variation of ADC values, significant motion and susceptibility artefacts were not observed in our cohort, which is less representative of real-life clinical practice where motion often hinders assessment of the peripheral zone31,32,33, for which DWI is the key sequence and where 75–80% of clinically significant lesions are located34,35. Acceptable pre-GBCA variation of MRF T1 coupled with the technique’s intrinsic robustness to motion therefore further support the need for investigating the added value of MRF in prostate imaging, particularly when DWI fails due to artefact.
As expected, MRF had considerably shorter maximum scanning time of 3 min 40 s compared to standard T1 and T2 mapping at 4 min 20 s and almost 6 min, respectively. The additional benefit of MRF for prostate imaging is that it can be sensitive to a wider range of T1 values (in this work, 1000–2000 ms) than VFA, owing to MRF’s extended range of measured flip angles. Conventional T1 mapping techniques, in turn, can be confounded by T2 effects, the choice of flip angles, and field (B1+) non-uniformity36,37,38. This reduction of T1 sensitivity to the aforementioned confounders is illustrated by our lower measured heterogeneity for all tissue types when compared with VFA T1 measurements, which may explain superior performance of pre-GBCA MRF T1 in vivo for differentiating tumour versus normal prostatic tissue in both PZ and TZ; better performance of MRF T1 was also demonstrated in the phantom study. Lower heterogeneity was noted for MSE T2 mapping, which is more well-studied in prostate cancer39.
This study has several limitations. Firstly, the small sample size may have artificially increased data variation leading to the inability of post-GBCA MRF T1 and pre-GBCA VFA T1 and MSE T2 to identify TZ and PZ tumours, respectively, and did not allow us to quantify grade-dependent variation of MRF T1 post GBCA. Secondly, only patients with low- and intermediate-grade disease were included in this study, which may also have had an impact on the diagnostic utility of both MRF and conventional mapping techniques. However, the inability of post-GBCA MRF T1 to detect any TZ lesions regardless of their Gleason grade indicates a reduction in robustness of the technique, while the added value of pre-GBCA MRF in detecting both TZ and PZ lesions has been demonstrated previously. VFA- and MSE-based T1 and T2 maps were not acquired post-GBCA, however, a head-to-head comparison of post-GBCA MRF and standard mapping techniques was not the purpose of this study. In vivo MRF-derived T2 values were unreliable in this study showing parabolic relationship with MSE-based T2 values, the reliability of which was also undermined by the longest TE being four-fold shorter than the calculated T2. However, gadolinium is known to have a more prominent T1 shortening effect at low doses, which underpinned the focus of this work on MRF T1 relaxometry. Moreover, as evidenced by the comparison with the “gold standard” IR-FSE, MRF T1 mapping used in this study was considered robust. In future work investigating T2 with MRF, we would give strong consideration for methods that increase the signal-to-noise ratio by increasing the slice thickness, voxel sizes, or by performing averaging with additional acquisitions. While averaging for MRF T2 mapping would remain sensitive to motion, we hypothesize that this effect would be reduced due to the pattern matching algorithm of MRF, therefore, supporting the clinical feasibility of both MRF T1 and T2 mapping. Acquiring MSE-based T2 maps with longer TEs to match the known T2 values observed in the prostate could also be helpful to improve their reliability and address the reported parabolic relationship with MRF, which may, however, represent a consistent trend worth further investigation.
In conclusion, GBCA administration leads to a considerable increase in MRF T1 variation following the expected significant T1 shortening effect and compromises its ability to detect TZ lesions. Therefore, performing MRF T1 prior to DCE imaging as part of a prostate mpMRI protocol should be considered as a preferred option to retain the technique’s added value for both PZ and TZ lesions and reduce the additional scanning time.
Methods
Phantom study
To evaluate the accuracy of T1 and T2 measurements, MRF and standard relaxation mapping data were obtained from the ISMRM/NIST phantom40. Phantom data were obtained on a 3 T MR750 scanner (GE Healthcare, Waukesha, WI, USA) using a 32-channel receiver coil. Regions-of-interest (ROI) were created from the spheres in either the T1 or T2 layer of the phantom.
Conventional T1 maps were obtained with an inversion recovery (IR) and variable flip angle (VFA) techniques. A conventional T2 map was obtained with multiple spin echo (MSE) measurements. The field-of-view (FOV) = 260 × 260 mm2, matrix = 256 × 256 and slice thickness = 3 mm matched in all techniques. Inversion recovery fast spin echo (IR-FSE) T1 maps from the T1 layer were obtained with inversion times (TI) = 50 ms, 100 ms, 200 ms, 400 ms, 800 ms, 1600 ms, 2400 ms, repetition time (TR) = 8000 ms and echo time (TE) = 13 ms. Multiple 3D GRE sequences were used for T1 mapping using the VFA method with flip angles of 2°, 5°, 8°, 12°, 15°, 18°, 22°, 26°, TR = 10 ms, TE = 1.6 ms. MSE T2 maps were obtained with TR = 600 ms and TEs = 8.2, 16.3, 24.5, 32.6, 40.8, 49.0, 57.1, 65.3 ms. Fitting was performed using a non-linear least squares fit to the IR, VFA and MSE signal equations in Python.
MR fingerprinting was also performed as further described in the MR fingerprinting protocol subsection.
Patient study
All elements of this prospective study were carried out in accordance with the Declaration of Helsinki and were approved by the institutional ethics board (NRES Committee East of England, UK), with written informed consent obtained from all participants. All methods were performed in accordance with the relevant guidelines and regulations. Patients on active surveillance, with MR-visible, biopsy-proven prostate cancer were included in this study. Exclusion criteria included prostate biopsy within the preceding 3 months, presence of pelvic metalwork, or any previous treatment for PCa.
Biopsy technique
Depending on clinical recommendation, biopsy was performed by either a transrectal or transperineal approach, using MRI/ultrasound fusion. All biopsy procedures were performed by experienced urologists and included 12–24 systematic cores, with 2–4 separate target cores acquired from the MRI defined lesion/s. All targets were defined by radiologists pre-procedure using T2-weighted imaging as the primary and diffusion-weighted imaging as the secondary source images, using the DynaCAD system (InVivo Corp, Orlando, FL) for transrectal and Biopsee software (Oncology Systems Limited, Shrewsbury, UK) for transperineal approaches as previously described41.
Multiparametric MRI
Patients underwent prostate MRI on a 3 T MR750 scanner (GE Healthcare, Waukesha, WI, USA) using a 32-channel receiver coil. Intravenous injection of hyoscine butylbromide (Buscopan, 20 mg/mL; Boehringer, Ingelheim am Rhein, Germany) was administered prior to imaging to reduce peristaltic movement, unless clinically contraindicated. Multiparametric MRI protocol included Axial T1 and multiplanar high-resolution T2-weighted 2D fast recovery FSE (field of view (FOV) 18 × 18 cm; voxel size 0.35 × 0.35 mm2; slice thickness 3 mm; gap 0 mm). Diffusion-weighted imaging (DWI) was performed using a spin-echo echo-planar imaging pulse sequence (FOV 28 cm; slice thickness 3 mm; gap 0 mm; b-values: b-150, b-750, and b-1400 s/mm2) and an additional small FOV (24 cm) b-2000 s/mm2 DWI sequence; apparent diffusion coefficient (ADC) maps were calculated automatically. T1 and T2 mapping were performed with VFA, MSE and MRF prior to dynamic contrast enhancement (DCE). DCE was performed using a standard sequence (FOV 24 cm; slice thickness and gap 3 mm and 0 mm, respectively; temporal resolution 7 s) following a bolus of Gadobutrol (Gadovist, 0.1 mmol/kg, Bayer) at 28 s via a power injector, at a rate of 3 ml/s (dose 0.1 mmol/kg). Post-GBCA MRF was performed immediately after DCE in all patients.
In vivo variable flip angle (VFA) T1 mapping
Multiple 3D GRE sequences were used for T1 mapping using the variable flip angle method with flip angles of 2°, 5°, 12°, 20°, and 32°. Each flip angle was acquired in 52 s, for a total duration of 4 min 20 s. Other parameters included: FOV = 36 cm, matrix = 256 × 256, slices = 52, slice thickness = 3 mm, echo time (TE) = 2.0 ms, repetition time (TR) = 15 ms, with 70% sampling.
In vivo multiple spin echo (MSE) T2 mapping
Multiple echo 2D FSE images were acquired for T2 mapping; FOV 36 cm, matrix 256 × 256, voxel size 1.4 × 1.4 mm2, slice thickness 2.5 mm, TR 2.6 s, TEs = 8.5 ms, 16.9 ms, 25.4 ms, 42.3 ms, 50.8 ms, 59.2 ms, 67.7 ms, averages = 0.5 (partial k-space), acquisition time 359 s (5 min 59 s).
MR fingerprinting protocol
A 2D steady-state-free-precession (SSFP) MRF sequence with inversion preparation was used for T1 and T2 mapping with 979 under-sampled, interleaved spirals for k-space sampling. The maximum gradient strength per spiral was 28 mT/m and the maximum slew rate was 108 T/m/s. The imaging parameters were: FOV = 260 × 260 mm2, matrix = 256 × 256, slices = 15–22, slice thickness = 3.0 mm, spacing 1.0 mm, sampling bandwidth = ± 250 kHz, slice dephasing = 8π, echo time (TE) = 2.5 ms, repetition time (TR) = 10 ms, acquisition time = 9.79 s/slice (maximum scanning time 3 min 40 s). We used similar flip angle lists to those used in previous works in order to more accurately assess the utility of MRF with standard lists36,42. Axial images were acquired to match the standard acquisition planes for other prostate image assessments. MRF image reconstruction, dictionary simulation and pattern matching parameters are listed in the Supplementary information S1.
Image analysis
MRF T1 values for MR-visible lesions (PI-RADS scores 4/5)6, normal peripheral zone (nPZ), normal transition zone (nTZ), subcutaneous abdominal fat and normal internal obturator muscle were calculated from ROIs originally drawn on anatomical T2-weighted images with reference to ADC maps (as presented in Supplementary Fig. S1) by a single fellowship-trained uro-radiologist with 12 years’ experience of reporting prostate MRI using the open-source segmentation software ITK-SNAP43. ROIs were then transposed on to the MRF T1 and T2, VFA- and MSE-based T1 and T2 and ADC maps with their size and location being matched to the appropriate FOV parameters and anatomical position of the outlined structures using in-house software developed within Python using the PyQtGraph and PyDicom libraries44. nPZ and nTZ ROIs were drawn in regions with biopsy-confirmed healthy tissue.
Statistics
In the phantom study, simple linear regression was used to evaluate the relationship between T1 and T2 values obtained using IR-FSE, VFA, MSE and MRF mapping techniques with their agreement assessed using the Bland–Altman analysis. In the patient study, the Shapiro–Wilk test was applied to assess the normality of imaging values with their intergroup comparison performed using either paired or unpaired t test as appropriate. Planned independent paired comparisons between MRF T1 and T2 values obtained from PZ and TZ lesions versus corresponding nPZ and nTZ were not adjusted for multiplicity; all other post hoc comparisons were adjusted using the Holm–Šidak method with alpha set at 0.05 as advised by an expert biostatistician. The variation of imaging values was evaluated using coefficient of variation (CV); CV of less than 25% indicated acceptable heterogeneity. Agreement between the conventional VFA- and MRF-based T1 relaxation times, treated as single paired measurements, was assessed using the Bland–Altman analysis. All plots and figures were created in Prism 8 (GraphPad Software, San Diego, CA, USA).
Data availability
The primary research data is available at https://data.mendeley.com/datasets/g3k7xwjpd3/draft?a=da4b46df-b5b3-4dac-aa8d-84b8c75ee526 (https://doi.org/10.17632/g3k7xwjpd3.1).
References
Rawla, P. & Rawla, P. Epidemiology of prostate cancer. World J. Oncol. 10, 63–89 (2019).
Kasivisvanathan, V. et al. MRI-targeted or standard biopsy for prostate-cancer diagnosis. N. Engl. J. Med. 378, 1767–1777 (2018).
EAU Guidelines: Prostate Cancer|Uroweb. https://uroweb.org/guideline/prostate-cancer/.
Standard operating procedure for multiparametric magnetic resonance imaging in the diagnosis, staging and management of prostate cancer—American Urological Association. https://www.auanet.org/guidelines/mri-of-the-prostate-sop.
Gaziev, G. et al. Defining the learning curve for multiparametric magnetic resonance imaging (MRI) of the prostate using MRI-transrectal ultrasonography (TRUS) fusion-guided transperineal prostate biopsies as a validation tool. BJU Int. 117, 80–86 (2016).
Turkbey, B. et al. Prostate imaging reporting and data system version 2.1: 2019 update of prostate imaging reporting and data system version 2. Eur. Urol. 76, 340–351 (2019).
Greer, M. D. et al. Interreader variability of prostate imaging reporting and data system version 2 in detecting and assessing prostate cancer lesions at prostate MRI. Am. J. Roentgenol. 212, 1197–1205 (2019).
de Rooij, M. et al. ESUR/ESUI consensus statements on multi-parametric MRI for the detection of clinically significant prostate cancer: Quality requirements for image acquisition, interpretation and radiologists’ training. Eur. Radiol. https://doi.org/10.1007/s00330-020-06929-z (2020).
Bipin Mehta, B. et al. Magnetic resonance fingerprinting: A technical review. Magn. Reson. Med. 81, 25–46 (2019).
McGivney, D. F. et al. Magnetic resonance fingerprinting review part 2: Technique and directions. J. Magn. Reson. Imaging 51, 993–1007 (2020).
Ma, D. et al. Magnetic resonance fingerprinting. Nature 495, 187–192 (2013).
Poorman, M. E. et al. Magnetic resonance fingerprinting Part 1: Potential uses, current challenges, and recommendations. J. Magn. Reson. Imaging 51, 675–692 (2020).
Panda, A. et al. Targeted biopsy validation of peripheral zone prostate cancer characterization with magnetic resonance fingerprinting and diffusion mapping. Invest. Radiol. 54, 485–493 (2019).
Panda, A. et al. MR fingerprinting and ADC mapping for characterization of lesions in the transition zone of the prostate gland. Radiology 292, 685–694 (2019).
Yu, A. C. et al. Development of a combined MR fingerprinting and diffusion examination for prostate cancer. Radiology 283, 729–738 (2017).
Gu, Y. et al. Fast magnetic resonance fingerprinting for dynamic contrast-enhanced studies in mice. Magn. Reson. Med. 80, 2681–2690 (2018).
Xiao, Y. D. et al. MRI contrast agents: Classification and application (Review). Int. J. Mol. Med. 38, 1319–1326 (2016).
Sushentsev, N. et al. The effect of capped biparametric magnetic resonance imaging slots on weekly prostate cancer imaging workload. Br. J. Radiol. 93, 20190929 (2020).
Kuhl, C. K. et al. Abbreviated biparametric prostate MR imaging in men with elevated prostate-specific antigen. Radiology 285, 493–505 (2017).
van der Leest, M. et al. High diagnostic performance of short magnetic resonance imaging protocols for prostate cancer detection in biopsy-naïve men: The next step in magnetic resonance imaging accessibility. Eur. Urol. 76, 574–581 (2019).
Zawaideh, J. P. et al. Diagnostic accuracy of biparametric versus multiparametric prostate MRI: Assessment of contrast benefit in clinical practice. Eur. Radiol. 30, 20 (2020).
Rosenkrantz, A. B. et al. Transition zone prostate cancer: Revisiting the role of multiparametric MRI at 3 T. Am. J. Roentgenol. 204, W266–W272 (2015).
Barrett, T., Rajesh, A., Rosenkrantz, A. B., Choyke, P. L. & Turkbey, B. PI-RADS version 2.1: One small step for prostate MRI. Clin. Radiol. 74, 841–852 (2019).
Caravan, P., Ellison, J. J., McMurry, T. J. & Lauffer, R. B. Gadolinium(III) chelates as MRI contrast agents: Structure, dynamics, and applications. Chem. Rev. 99, 2293–2352 (1999).
De Leõn-Rodríguez, L. M., Martins, A. F., Pinho, M. C., Rofsky, N. M. & Sherry, A. D. Basic MR relaxation mechanisms and contrast agent design. J. Magn. Reson. Imaging 42, 545–565 (2015).
Anderson, C. E. et al. Dynamic, simultaneous concentration mapping of multiple MRI contrast agents with dual contrast—magnetic resonance fingerprinting. Sci. Rep. 9, 20 (2019).
Van Niekerk, C. G., Witjes, J. A., Barentsz, J. O., Van Der Laak, J. A. W. M. & Hulsbergen-Van De Kaa, C. A. Microvascularity in transition zone prostate tumors resembles normal prostatic tissue. Prostate 73, 467–475 (2013).
Padhani, A. R. et al. Dynamic contrast enhanced MRI of prostate cancer: Correlation with morphology and tumour stage, histological grade and PSA. Clin. Radiol. 55, 99–109 (2000).
Hansford, B. G. et al. Dynamic contrast-enhanced MR imaging curve-type analysis: Is it helpful in the differentiation of prostate cancer from healthy peripheral zone?. Radiology 275, 448–457 (2015).
Greer, M. D. et al. Validation of the dominant sequence paradigm and role of dynamic contrast-enhanced imaging in Pi-RADS version 2. Radiology 285, 859–869 (2017).
Mazaheri, Y., Vargas, H. A., Nyman, G., Akin, O. & Hricak, H. Image artifacts on prostate diffusion-weighted magnetic resonance imaging: Trade-offs at 1.5 tesla and 3.0 tesla. Acad. Radiol. 20, 1041–1047 (2013).
Caglic, I., Hansen, N. L., Slough, R. A., Patterson, A. J. & Barrett, T. Evaluating the effect of rectal distension on prostate multiparametric MRI image quality. Eur. J. Radiol. 90, 174–180 (2017).
Slough, R. A., Caglic, I., Hansen, N. L., Patterson, A. J. & Barrett, T. Effect of hyoscine butylbromide on prostate multiparametric MRI anatomical and functional image quality. Clin. Radiol. 73(216), e9-216.e14 (2018).
McNeal, J. E., Redwine, E. A., Freiha, F. S. & Stamey, T. A. Zonal distribution of prostatic adenocarcinoma. Correlation with histologic pattern and direction of spread. Am. J. Surg. Pathol. 12, 897–906 (1988).
McNeal, J. E. Cancer volume and site of origin of adenocarcinoma in the prostate: Relationship to local and distant spread. Hum. Pathol. 23, 258–266 (1992).
Buonincontri, G. et al. Multi-site repeatability and reproducibility of MR fingerprinting of the healthy brain at 1.5 and 3.0 T. Neuroimage 195, 362–372 (2019).
Heule, R., Ganter, C. & Bieri, O. Variable flip angle T1 mapping in the human brain with reduced t2 sensitivity using fast radiofrequency-spoiled gradient echo imaging. Magn. Reson. Med. 75, 1413–1422 (2016).
Cheng, H. L. M. & Wright, G. A. Rapid high-rsolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity. Magn. Reson. Med. 55, 566–574 (2006).
Hoang Dinh, A. et al. Characterization of prostate cancer using T2 mapping at 3 T: A multi-scanner study. Diagn. Interv. Imaging 96, 365–372 (2015).
Jiang, Y. et al. Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom. Magn. Reson. Med. 78, 1452–1457 (2017).
Barrett, T. et al. Three-year experience of a dedicated prostate mpMRI pre-biopsy programme and effect on timed cancer diagnostic pathways. Clin. Radiol. 74, 20 (2019).
Jiang, Y., Ma, D., Seiberlich, N., Gulani, V. & Griswold, M. A. MR fingerprinting using fast imaging with steady state precession (FISP) with spiral readout. Magn. Reson. Med. 74, 1621–1631 (2015).
Yushkevich, P. A. et al. User-guided 3D active contour segmentation of anatomical structures: Significantly improved efficiency and reliability. Neuroimage 31, 1116–1128 (2006).
Mason, D. SU-E-T-33: Pydicom: An open source DICOM library. Med. Phys. 38, 3493–3493 (2011).
Acknowledgements
The authors acknowledge research support from National Institute of Health Research Cambridge Biomedical Research Centre, Cancer Research UK (Cambridge Imaging Centre Grant number C197/A16465), Cancer Research UK and the Engineering and Physical Sciences Research Council Imaging Centre in Cambridge and Manchester and the Cambridge Experimental Cancer Medicine Centre.
Author information
Authors and Affiliations
Contributions
N.S., J.D.K., and T.B. formulated the research idea, developed the study design and methodology. N.S. performed data collection, statistical and computational analysis, wrote the initial draft of the article and coordinated its subsequent revision. J.D.K. set up and validated the imaging techniques used in this study, implemented the computing code and supporting algorithms and oversaw research data production and curation. N.S. and J.D.K. jointly conducted the phantom study. G.B. and R.F.S. provided technical support to J.D.K. in setting up the imaging techniques at the core institution thereby serving as external mentors for the core team. M.J.G. facilitated the data collection and provided the required instrumentation and computing resources for both phantom and clinical components of this study. T.B. supervised and coordinated both the phantom and clinical components of this study and was instrumental in revising the manuscript. V.J.G. curated patient inclusion in this study and coordinated communication with the clinical urology team. All authors contributed considerably to reviewing the presentation of the submitted work.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Sushentsev, N., Kaggie, J.D., Buonincontri, G. et al. The effect of gadolinium-based contrast agent administration on magnetic resonance fingerprinting-based T1 relaxometry in patients with prostate cancer. Sci Rep 10, 20475 (2020). https://doi.org/10.1038/s41598-020-77331-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-020-77331-4
This article is cited by
-
Synthetic MRI in breast cancer: differentiating benign from malignant lesions and predicting immunohistochemical expression status
Scientific Reports (2023)
-
MR fingerprinting of the prostate
Magnetic Resonance Materials in Physics, Biology and Medicine (2022)
-
MRI-derived radiomics model for baseline prediction of prostate cancer progression on active surveillance
Scientific Reports (2021)
-
Accelerated 3D whole-brain T1, T2, and proton density mapping: feasibility for clinical glioma MR imaging
Neuroradiology (2021)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.