Abstract
4D flow MRI enables quantitative assessment of helical flow. Current methods are susceptible to noise. To evaluate helical flow patterns in healthy volunteers and patients with bicuspid aortic valves (BAV) at 1.5 T and 3 T using pressure-based helix-extraction in 4D flow MRI. Two intraindividual 4D flow MRI examinations were performed at 1.5 T and 3 T in ten healthy volunteers (5 females, 32 ± 3 years) and 2 patients with BAV using different acceleration techniques (kt-GRAPPA and centra). Several new quantitative parameters for the evaluation of volumes [ml], lengths [mm] as well as temporal parameters [ms] of helical flow were introduced and analyzed using the software tool Bloodline. We found good correlations between measurements in volunteers at 1.5 T and 3 T regarding helical flow volumes (R = 0.98) and temporal existence (R = 0.99) of helices in the ascending aorta. Furthermore, we found significantly larger (11.7 vs. 77.6 ml) and longer lasting (317 vs. 769 ms) helices in patients with BAV than in volunteers. The assessed parameters do not depend on the magnetic field strength used for the acquisition. The technique of pressure-based extraction of 4D flow MRI pattern is suitable for differentiation of normal and pathological flow.
Similar content being viewed by others
Introduction
Time-resolved, 3-directional, 3-dimensional (3D) phase-contrast cardiac magnetic resonance (CMR) imaging (4D flow MRI) is an innovative technique for the assessment of cardiovascular disease. Beside absolute flow quantification, 4D flow MRI allows measurements of helical blood flow and may therefore provide deeper insights into cardiovascular pathologies1. Visual quantitative analysis of helical flow patterns can be performed already with appropriate reproducibility2. However, for more complex analyses, follow-up examinations or the assessment of therapeutic success, absolute quantification of helical flow patterns, e.g. of volumes or velocities within a helix, might be beneficial.
Numerous new parameters for the assessment of cardiovascular diseases with 4D flow MRI were already defined3,4,5,6 and different techniques for the semi-automatic assessment of helical flow have been described: recent studies utilized the λ2 criterion, vorticity, helicity or normalized helicity for evaluation of complex flow patterns in the ascending aorta7,8,9,10. However, it has been shown that most of those parameters are susceptible to noise and highly depend on sufficient image quality11. Therefore, we recently introduced a more stable and precise method: a pressure-based technique for helix extraction12, which is less susceptible to noise and will be used in this study.
The aim of our study was to utilize 4D flow MRI for the quantitative, pressure-based extraction and analysis of flow patterns in the ascending aorta in healthy volunteers and two patients with bicuspid aortic valves (BAV) as a proof of concept with different fast sequences at 1.5 T and 3 T.
Methods
Study cohort
Ten healthy volunteers with no history of cardiovascular diseases (CVD) (5 females, 22–52 years, mean age 32 ± 3 years) and two patients with BAV (1 female, age 39 and 45 years, referred to as BAV_1 and BAV_2) were included in this prospective, comparative study. A local ethics board approved the study and written informed consent was obtained from all participants. All experiments were performed in accordance with relevant guidelines and regulations.
Magnetic resonance (MR) image acquisition
For the volunteers, two intraindividual CMR examinations were performed at two different days within a six month time span (mean 99 ± 61 days). The first datasets were acquired at a 3 Tesla (T) scanner using a 16-channel surface coil in combination with a 12-element spine coil (Magnetom Verio Dot, Siemens Healthcare GmbH, Erlangen, Germany) using a kt-GRAPPA accelerated sequence: TE 1.8 ms; TR 32 ms. This sequence has been evaluated before13,14. The second examinations were performed at a 1.5 T scanner using a five-element phased-array surface coil (Achieva, Philips Healthcare, Best, Netherlands): TE 2.3 ms, TR 37.6 ms. 4D flow MRI data were acquired in a sagittal oblique 3D volume covering the whole thoracic aorta. The flow-encoded velocity (Venc) was adjusted to a maximum flow velocity of 150 cm/s along all three flow directions. Variable parameters like field of view (280 × 260 × 52), Flip Angle (5°) and encoded phases (25) were kept constant in both examinations. Patients underwent CMR at 1.5 T, only. Data acquisition was performed in line with recently published recommendations15.
Cardiac magnetic resonance (CMR) data analysis
Vessel segmentation, Blood flow visualization and Pre-processing
All processing, segmentation and measurement steps were carried out using Bloodline, a custom-made software tool for guided analysis of 4D flow MRI datasets, which has been introduced before16,17,18 (Department of Simulation and Graphics, University of Magdeburg, Germany). A centerline was drawn through the whole ascending aorta semi-automatically, beginning at the level of the aortic root. Anatomically, the ascending aorta was defined as the volume of the aorta between the aortic valve and the origin of the brachiocephalic trunk.
Bloodline enables the analysis of defined regions of interest, e.g. the ascending aorta, and provides the volume of the segmented vessel in ml as well as the length of the centerline in mm. Aortic blood flow was visualized using time-resolved pathlines. We corrected for phase wraps, eddy currents and background noise, as previously described4,19.
Measurements and flow quantifications
All segmentations and measurements were carried out by two observers with >5 years and 2 years of experience in CMR. Readers were blinded for which datasets were from which participant. A measuring plane was positioned at the mid-ascending aorta for 4D flow MRI volume quantification (ml/cycle) and peak velocity (m/sec). The automatic identification of complex flow patterns used by the software Bloodline utilizes a recently described pressure-based technique12. Both helical and vortical flow patterns produce pressure low-points, thus local variations of pressure are indicative of a helix or vortex. The calculation of the relative pressure map is based on an iterative method introduced by Tyszka et al.20 with additional post-processing steps. The resulting pressure map is smoothed using one iteration of a 3 × 3 × 3 binomial filter, clamped to the 1–99% quantile range to remove outliers. Lastly, to ensure that the pressure map is still relative to the average pressure after filtering, all values are shifted by subtracting the mean of the filtered pressure values. In the resulting pressure map, areas with a relative pressure P < 0 mmHg are considered to be vortex regions and used as a filter mask for densely seeded pathlines in order to visually represent the helical or vortical flow. Therefore, the relative pressure of P < 0 mmHg was used as a threshold to define regions with helical flow versus regions with non-helical flow. As this technique does not require any user interaction beyond the vessel segmentation, it might be used in a standardized manner and yields highly reproducible results.
However, the differentiation between helical and vortical flow might be difficult, because there is a smooth transition between both phenomena. Often, neither a visual qualitative nor quantitative differentiation between both is possible, because definite cut-offs do not exist. Visually, helical flow describes a rotational, corkscrew-like or helical flow pattern with a usually antegrade forward motion, whereas vortical flow is defined as a flow pattern with a revolving flow volume on the spot around a common axis without any forward motion21.
Since we could not quantitatively discriminate between both similar flow patterns and the assumption that a real vortex without any motion component in the direction of flow does not exist, we will use the common term “helix” or “helical” for both helical and vortical flow.
We introduced several new quantitative parameters to better describe helical flow patterns:
- 1.
Temporal Helical Existence (THEx) to describe the duration of helical flow throughout the cardiac cycle. This parameter is calculated from the time-resolved relative pressure map by measuring over how many time steps helical flow (i.e. negative pressure) is present, regardless of the size or number of detected helices.
THEx can only produce a single value, thus it is a global parameter for each dataset. The extracted duration is the sum of the duration of all helical flow according to the pressure map during the cardiac cycle.
The calculation of the pressure map takes into account the entirety of the segmented vessel and thus, only depends on the segmentation quality. Defining a specific region of interest, e.g. the aortic arch, is independent of the pressure map calculation
It was given as an absolute value (absTHEx) in milliseconds (ms) and as a relative value (relTHEx) in percent (%) of the cardiac cycle.
- 2.
Different parameters were used to asses the size of helices:
- 2.1
The maximum Helical Volume (HVmax) as an absolute value in milliliters [ml] and the relative value in percent [%] of the volume of the ascending aorta at a definite time point of the cardiac cycle.
- 2.2
Furthermore, the accumulated Helical Volume (HVacc) of the helix was defined as a summation of all volumes that contributed to the helix during the cardiac cycle. This parameter can be described as a temporal maximum intensity projection of the helix. It is given as an absolute value in milliliters [ml] or relatively in percent [%] based on the volume of the ascending aorta (see Fig. 1).
- 2.3
According to the HVacc we assessed the accumulated length of a helix with the accumulated Helical Volume Length (HVLacc). This parameter describes the length of a temporal maximum intensity projection of the helix. It is given as an absolute value in millimeters [mm] or relatively in percent [%] based on the length of the ascending aorta (see Fig. 1).
In addition, all volumetric parameters were given as an index in ml/m2, which describes the volume of the helix normalized to the body surface area (BSA in m2) as HV(I)max and HV(I)acc.
The rotation direction (RD) was subdivided into left-handed and right-handed. If a helix rotates in flow direction clockwise around the centerline, it is considered as right-handed (RD+), otherwise left-handed (RD−). RD was assessed as described by Meuschke et al.22.
For all pathline segments, each consisting of a point and its successor, the RD is calculated separately. Each position of the line is projected onto the vessel cross-section plane at the nearest centerline point. This plane is defined by its center c (the centerline point) and two axes (x and y). The x-axis nx is determined by using the cross-product of the centerline tangent n at c and the word-space x-axis 1,0,0 T.
Similarly, the y-axis ny is the cross-product of nx and n. For the projected points, we identified the angle to the x-axis of the plane. For example, if the successive points are located in the first and second quadrant, they present an RD+ segment (Fig. 2a). If the successive points are in the third and fourth quadrant, a left-handed or RD− segment exists (Fig. 2b). If they are located in diagonal quadrants, the intersection between the connecting line and the y-axis is used to define a RD (Fig. 2c). Depending on whether a path line has more right or left rotating segments, it is classified as RD+ or RD−.
Intra- and interobserver variability
The intraobserver variablility was assessed by one investigator with >5 years of experience in post-processing. This observer performed the previously described segmentations and measurements in all 12 datasets and repeated them in random order two months after the first evaluation. The interobserver variablility was analyzed by the second investigator, who repeated all segmentation and measurements as described above. The segmentation and measurements were done twice.
Statistical analysis
All analyses were performed using MedCalc Statistical Software V15.11.4 (MedCalc Software, Ostend, Belgium). Quantitative variables were expressed as mean values and standard deviations (SD). A Wilcoxon test for comparison of temporal helical existence (THEx), accumulated helical volume length (HVLacc), helical volume (HVmax, HVacc) of volunteers at 1.5 T and 3 T was performed. A p-value < 0.05 was considered statistically significant. Additionally, correlation analyses were performed using paired correlations. Bland-Altman analysis provided the mean differences between measurements (bias) and the limits of agreement (LOA) used for the different methods of flow analysis. Intra- and interobserver variability was assessed using linear regression analysis, scatter plots and interclass correlation (ICC).
Ethics approval and consent to participate
A local ethics board approved this study: Ethik-Kommission an der Medizinischen Fakultät der Universität Leipzig AZ 443/16-ek.
Results
Volunteer characteristics
Ten healthy volunteers and two patients with BAV with an equal gender distribution and age were chosen for this comparative study. The volunteer’s mean resting heart rate during the examination at 1.5 T and at 3 T did not differ significantly (p = 0.198) with 61/min ± 19 (1.5 T) and 60/min ± 18 (3 T), respectively, as well as the mean cardiac output (p = 0.303) during the two different time points with 4.47 l/min ± 1.45 (1.5 T) and 4.28 l/min ± 1.39 (3 T). In patients, the resting heart rate was 63 (BAV_1) and 67/min (BAV_2) and the cardiac output was 4.0 (BAV_1) and 4.8 l/min (BAV_2). The mean maximum diameter and volume of the ascending aorta was 30 mm ± 5 and 53.27 ml ± 15.91 in volunteers. The maximum diameter and volume of the aorta ascendens in patients were 33 mm and 71.91 ml (BAV_1) and 57 mm and 290.02 ml (BAV_2). The mean length of the ascending aorta was 92.4 mm ± 14.6 in volunteers and 88.3 (BAV_1) and 139.6 mm (BAV_2) in patients.
CMR acquisition
Mean acquisition time was 8:09 ± 2:28 min at 1.5 T and 9:40 ± 3:15 at 3 T.
Flow quantification
We found no significant differences between 3 T and 1.5 T regarding net forward flow [ml/cycle] and peak velocity [m/sec] in volunteers (see Table 1).
Presence of helices in the ascending aorta
In all healthy participants helical flow could be detected only in the ascending aorta, at the same location at 1.5 T and 3 T. The automated helical flow detection of the software tool Bloodline detected one helix within the ascending aorta in every subject. The intraindividual comparison between helix visualizations at 1.5 T compared to 3 T in the same volunteer showed no significant differences (Fig. 3A,B). Visually, a clear corkscrew-like flow pattern could be observed. In every case the helix started close downstream from the aortic valve and evolved through the ascending aorta towards the aortic arch. The flow pattern at the very proximal parts of the observed helix could be further characterized as vortical flow in every participant, always located in the region of the coronary sinuses (Fig. 3C). Vortices were larger in patients than in volunteers. This was the only spot where vortical flow in the narrow sense occurred. The other parts of the helix could be classified as pure helical flow.
Duration of helical flow in the ascending aorta throughout the cardiac cycle – the temporal helical existence (THEx)
In volunteers the mean THEx of the helices was 316.8 ms (±95.6) at 1.5 T and 320.7 ms (±92.8) at 3 T, respectively, without significant differences between the measurements (p = 0.448) (Fig. 4A,B). The correlation coefficient (R) was 0.99 and limits of agreement (LOA) were −31.1–21.2 ms (see Table 2). The helical flow started in early systole and faded away in the mid or late systole in all ten volunteers, independently from field strength. The THEx in BAV_1 was 769.0 and 567.4 ms in BAV_2 and thus longer than in volunteers. Volunteer’s mean relative THEx was 33.6% (±11.9) at 1.5 T and 34.0% (±11) at 3 T of the cardiac cycle, without significant differences between the measurements (p = 0.401, R = 0.99, LOA −4.1–3.5%) (see Table 2). The relative THEx in patients was 100% in both cases and therefore more than 2-fold longer as compared to volunteers. In other words, in BAV patients helical lasted the whole cardiac cycle (see Fig. 4C,D).
Maximum helical (HVmax) and accumulated helical (HVacc) volumes in the ascending aorta
The volunteers’ absolute and relative HVmax demonstrated a broad range of volumes from 2.9–19.5 ml (9.9–34.7%) at 1.5 T and 2.6–19.4 ml (10.0–35.2%) at 3 T. The mean absolute or relative HVmax was 11.7 ml (±7.5) or 23.8% (±17.4) at 1.5 T and 12.3 ml (±7.1) or 24.9% (±16.4) at 3 T. We found no significant differences between the measurements at 1.5 and 3 T (p = 0.628 and p = 0.640) (Fig. 5A,B). The correlation R was 0.98 and the LAO were −2.4–3.3 ml and −5.2–7.1%, respectively (see Table 2). The absolute HVmax in patients was larger with 41.8 ml (BAV_1) and 77.6 ml (BAV_2), but not the relative HVmax with 17.4% (BAV_1) and 32.3% (BAV_2) (Fig. 5C,D). In all cases the helices reached their maximum volume within the first 30% of the cardiac cycle.
The mean absolute or relative HVacc in volunteers was 19.9 ± 9.0 ml or 35.5 ± 19.9% at 1.5 T and 19.7 ± 8.9 ml or 37.9 ± 15.8% at 3 T without significant differences between measurements (p = 0.501 and p = 0.417) (Fig. 6A,B). The correlation coefficient (R) was 0.99 and the LOA were −1.8–2.2 ml and −15.9–12.1% (see Table 2 and Fig. 6C,D)).
The mean absolute and relative HVacc in patients was larger with 55.9 ml in BAV_1 and 236.1 ml in BAV_2 (77.7 and 81.1%) (Fig. 6E,F).
Additionally, we introduced the accumulated and maximum helical volume index, (HV(I)acc and HV(I)max), indicating the accumulated and maximum volume of a helix normalized to the BSA. With the introduction of these parameters we wanted to reduce the broad range of HVacc and HVmax, assuming that these parameters were highly depending on the shape and weight of each individual. The HV(I)acc and (HV(I)max) ranged from 3.5 to 15.3 (0.6 to 13.6) ml in volunteers and up to 100.2 (72.2) ml/m2 in patients. The volunteers’ mean calculated HV(I)acc was 9.9 ± 3.8 at 1.5 T and 9.8 ± 4.1 ml/m2 at 3 T without significant differences (p = 0.558), LOA were −1.0–1.2 (Fig. 7A,B). HV(I)max was 6.1 ± 4.0 ml/m2 at 1.5 T and 6.2 ± 4.1 at 3 T without significant differences between the measurements (p = 0.631, R = 0.99). Compared to HVacc and HVmax the SD of these indeces was lower. In patients the HVIacc was 37.4 (BAV_1) and 100.2 ml/m2 (BAV_2) and the HVImax in BAV_1 was 41.8 ml and in BAV_2 77.6 ml (Fig. 7C).
In conclusion, all absolute and relative temporal and volumetric helical parameters were larger in patients than in volunteers, except the relative maximum helical volume (HVmax).
Accumulated helical volume length (HVLacc) of helices in the ascending aorta
The volunteers’ mean absolute or relative HVLacc was 73 ± 12 mm or 78.8 ± 13.7% at 1.5 T and 77 ± 11 mm or 79.3 ± 12.9% at 3 T without significant differences between the measurements (p = 0.411 and p = 0.491) (Fig. 8A,B). The correlation coefficient (R) was 0.99 and the LOA were −3–4 (absolute HVLacc [mm]) and −3.2–4.1 (relative HVLacc [%]) (see Table 2 and Fig. 8C,D).
The absolute (relative) HVLacc in BV_1 was 88 mm and in BAV_2 140 mm (100%) (Fig. 8E,F). In all cases HV Lacc was larger in patients than in volunteers.
Rotational direction (RD) of helices in the ascending aorta
In all cases helical flow was right-handed (RD+); we found no left-handed helical flow in healthy volunteers as well as in our two patients.
Intra- and interobserver variability
Analysis of the intraobserver variability showed for the measurements of absolute THEx, HVacc, HVLacc and HVmax R = 0.99, 0.99, 1.0 and 1.0 (p < 0.0001), ICC was 0.99, 1.0, 1.0 and 0.99, respectively.
Analysis of the interobserver variability showed for the measurements of absolute THEx, HVacc, HVLacc and HVmax R = 0.96, 0.97, 0.98 and 0.99 (p < 0.0001), ICC was 0.97, 0.95, 0.98 and 0.97, respectively.
Discussion
The software Bloodline utilizes a pressure-based technique for the assessment of flow patterns. The major finding of this study is the consistency of quantitative 4D flow MRI parameters for the assessment of helical flow across platforms and field strengths using new software for helix extraction. The results demonstrated that these parameters neither depend on the magnetic field strength nor on the scanner platform used for the acquisition. In addition, we demonstrated as a proof of concept that there are significant differences between healthy volunteers and patients with BAV.
For establishing normal values or for planning multi-center trials, scanner- and center-independent parameters are mandatory. If this prerequisite is not fulfilled, each center has to establish its own normal values. Therefore, the Society for Cardiovascular Magnetic Resonance (SCMR) recommends to establish normal values for every particular set-up (scanner, vendor, field strength, contrast regime, sequences)23 when using different T1- and T2-mapping techniques.
We performed 4D flow MRI measurements at different scanners using different sequences, acceleration techniques and different field strengths and found no significant differences. Thus, we state that those parameters could serve for establishing normal values or for follow-up examinations in a clinical setting with a possible multi-scanner application.
There are different methods for identifying flow patterns, which can be divided into local and global methods24. For example, the λ2 criterion by Jeong and Hussain and the Q criterion by Hunt are local methods25,26,27. Recent studies utilized vorticity, helicity and normalized helicity for assessment of complex flow patterns, which are local methods as well. In 2013 Köhler et al. found that those methods depend highly on sufficient image quality, otherwise they could misinterpret noisy regions for helices11. Compared to local methods, global methods give more precise results at the expense of higher computation times11. One example of a global method is a technique based on local pressure minima. Köhler et al.12 validated this technique by visual inspection of the helix regions and comparison with the commonly used and accepted λ2-based helix detection on 10 4D flow MRI datasets (1 healthy volunteer, 3 aortic aneurysms, 2 BAV, 1 aortic arch prosthesis, 1 aortic bypass, 1 left ventricle and atrium with valve defect, 1 insufficient pulmonary valve). In most cases, pathlines filtered using the pressure-based method produced a more satisfying visual representation of the helix compared to a λ2-based filtering, e.g. as a result of increased line density and less fragmented lines, whereas in all other cases the visual quality was at least on par. The analysis of the inter- and intraobserver variability showed that the introduced technique and parameters yield highly reproducible results.
Although there was a six month time span between the two volunteer examinations, we found no significant differences between the volunteer groups regarding net forward flow, peak velocity and helical flow volumes in the ascending aorta and cardiac output, which underlines the stability of measurements in 4D flow MRI across platforms and even with different magnetic field strength. These findings fit well with the literature: In 2012 Strecker et al. found identical performance of 4D flow MRI at 1.5 T and 3 T regarding flow velocities and volumes28.
Contrarily, we found significant differences between healthy volunteers and patients with BAV. More importantly, for all assessed parameters we found significantly higher results in patients than in volunteers, indicating that the used method of pressure-based analysis enables a distinction between normal and pathological flow. The technique of pressure-based helix analysis could therefore serve for detailed assessment of pathologies of the cardiovascular system. This should be addressed in further studies.
With the introduced technique of pressure-based helix extraction we were able to detect one helix in the ascending aorta in every participant. These findings partially match with the results of previous studies indicating that helical flow in the ascending aorta is a physiological phenomenon. Moreover, helical flow is thought to protect against atherosclerosis8,29. However, in 2011 Morbiducci et al. analyzed flow patterns in the thoracic aorta in five healthy volunteers and found no helical flow in the ascending aorta or the proximal aortic arch. These analyses have been carried out by visual pathline tracking only, with possible misinterpretation due to an overlay of the pathlines with helical flow and pathlines with non-helical flow. In contrast, by utilizing a technique enabling semiautomatic extraction, visualization and analysis of helical flow separately from the pathlines with non-helical flow, we found one helix in the ascending aorta close downstream from the aortic valve in each volunteer.
Recently, the evolution of helical flow has been investigated and it has been stated that helical flow emerges mainly in peak systole8. Contrarily, our results indicate that helical flow usually starts already in early-systole and ceases in late-systole in healthy volunteers or even last the whole cardiac cycle in BAV patients. Additionally, we found that helices in the early systole are rather small, while they reach their maximum volume and length around mid-systole in both patients and volunteers.
One possible explanation for those differences could be that in visual analyses small helices could be missed due to the overlay of pathlines with helical flow and pathlines with non-helical flow, while (semi)automatic analyses enable detection of even small helices.
In 2012 Bürk et al. e.g. assessed 3D blood flow patterns like the helical volume, but they evaluated the helical volume only visually. They discriminated whether a helix partially or fully filled the aortic lumen and found rather small helices in healthy participants21. These findings fit with our results: In the study presented here, we found a partial filling of the aortic lumen in all volunteers and a complete filling in patients. More importantly, we were able to absolutely quantify helical flow. Furthermore, we introduced additional parameters for the quantitative assessment of flow patterns in the ascending aorta like the absolute or relative accumulated Helical Volume (HVacc) and the accumulated Helical Volume Length (HVLacc).
We showed that these parameters could help to identify and precisely quantify helical flow. We found that the temporal existence (ms), volume (ml) and length (mm) as absolute values strongly depend on individual characteristics of the volunteers. Therefore, they might not be suitable for inter-individual comparisons. It seems to be reasonable to prefer normalized values like the helix’ temporal existence normalized to the duration of the cardiac cycle, its volume normalized to the volume of the ascending aorta or on the BSA and its length normalized to the length of the ascending aorta.
For the assessment of the dimensions of the right and left ventricle it is recommended to normalize the volumes of the ventricles to the subject’s BSA30. Analogous to this normalization we introduced the maximum Helical Volume index (HVImax), which represents the maximum helical volume of the helix normalized to the BSA. We found the HVImax to be significantly higher in patients with BAV than in volunteers, indicating that this parameter could serve for the assessment of congenital malformations.
Additionally, we introduced the accumulated Helical Volume (HVacc) and Helical Volume Length (HVLacc) and found those parameters to be significantly higher in BAV patients compared to healthy volunteers.
It is known that somehow bigger and longer helices with a “mismatch” between the volume of the helix and aorta could lead to aortic dilatation or elevated wall shear stress21,31,32. However, further studies are needed to evaluate which parameters are best for the assessment of pathologic flow patterns.
In previous studies, three-dimensional helical parameters have mainly been assessed quantitatively only in several two-dimensional cross-sections distributed over the aorta2,33. These methods enabled e.g. precise measurements of the rotational direction of helices. However, since these measurements were performed in two-dimensional cross-sections of the aorta, they could only cover a small part of the helix and thus e.g. precise measurements of the duration, volume and length of a helix could not be performed. In our study, we introduced a comprehensive technique enabling 3D analysis of helices with full coverage.
Previous studies utilizing only a visual assessment of 4D flow MRI patterns in the aorta showed that vortical flow is a common phenomenon at the coronary sinuses and seems therefore to be a physiological phenomenon34. This is well in line with our findings: we semi-automatically detected vortical flow at the coronary sinuses and helical flow in the ascending aorta in every subject.
The main limitation of this study is that only 10 healthy individuals and two patients (one with a dilated aorta) were included. For further validation of the results a larger study cohort is needed, however, this study was designed as a proof of concept and an introduction of new parameters.
In conclusion, the technique of pressure-based extraction of 4D flow MRI pattern is suitable for quantitative analysis of helical flow in the ascending aorta and for differentiation of normal and pathological flow. It yields results that are independent from the used sequences, strengths of the magnetic field and manufacturers. With this technique, it might be possible to perform follow-up examinations and to quantitatively assess therapeutic success. These parameters could possibly give deeper insights and lead to a better understanding of arterial hemodynamic and pathology.
Data availability
Please contact author for data requests.
References
van der Geest, R. J. & Garg, P. Advanced Analysis Techniques for Intra-cardiac Flow Evaluation from 4D Flow MRI. Curr. Radiol. Rep. 4, 38 (2016).
Lorenz, R. et al. 4D Flow MRI in aortic valve disease demonstrates altered distribution of aortic blood flow helicity. Magn Reson Med. 31, 2012 (2012).
Markl, M., Frydrychowicz, A., Kozerke, S., Hope, M. & Wieben, O. 4D flow MRI. J. Magn. Reson. Imaging 36, 1015–1036 (2012).
Dyverfeldt, P. et al. 4D flow cardiovascular magnetic resonance consensus statement. J. Cardiovasc. Magn. Reson. 17, 72 (2015).
Stankovic, Z., Allen, B. D., Garcia, J., Jarvis, K. B. & Markl, M. 4D flow imaging with MRI. Cardiovasc. Diagn. Ther. 4, 173–92 (2014).
Stalder, A. F. et al. Quantitative 2D and 3D phase contrast MRI: Optimized analysis of blood flow and vessel wall parameters. Magn. Reson. Med. 60, 1218–1231 (2008).
Callaghan, F. M. et al. Age-Related Changes of Shape and Flow Dynamics in Healthy Adult Aortas: A 4D Flow MRI Study. J. Magn. Reson. Imaging 1–11 https://doi.org/10.1002/jmri.26210 (2018).
Morbiducci, U. et al. Mechanistic insight into the physiological relevance of helical blood flow in the human aorta: An in vivo study. Biomech. Model. Mechanobiol. 10, 339–355 (2011).
Valverde, I. et al. Three-Dimensional Quantification of Vorticity and Helicity From 3D Cine PC-MRI Using Finite-Element Interpolations. 553, 541–553 (2018).
Spiczak, J. V, Crelier, G., Giese, D., Kozerke, S. & Maintz, D. Quantitative Analysis of Vortical Blood Flow in the Thoracic Aorta Using 4D Phase Contrast MRI. 1–19, https://doi.org/10.1371/journal.pone.0139025 (2015).
Koehler, B. et al. Semi-automatic vortex extraction in 4D PC-MRI cardiac blood flow data using line predicates. IEEE Trans. Vis. Comput. Graph. 19, 2773–2782 (2013).
Köhler, B., Grothoff, M., Gutberlet, M. & Preim, B. Pressure-based vortex extraction in cardiac 4D PC-MRI blood flow data. EUROVIS 0–4, https://doi.org/10.2312/eurovisshort.20181071 (2018).
Ebel, S. et al. Validation of two accelerated 4D flow MRI sequences at 3 T: a phantom study. Eur. radiolgy Exp. 3, 12 (2019).
Ebel, S. et al. Comparison of two accelerated 4D-flow sequences for aortic flow quantification. Sci. Rep. 9, 1–10 (2019).
Zhong, L. et al. Intracardiac 4D Flow MRI in Congenital Heart Disease: Recommendations on Behalf of the ISMRM Flow & Motion Study Group. J. Magn. Reson. Imaging 50, 677–681 (2019).
Köhler, B., Grothoff, M., Gutberlet, M. & Preim, B. Visual and quantitative analysis of great arteries’ blood flow jets in cardiac 4D PC-MRI data. Comput. Graph. Forum 37, 195–204 (2018).
Köhler, B. et al. Guided Analysis of Cardiac 4D PC-MRI Blood Flow Data. Eurographics Dirk Bartz Prize. 2015, 2–5 (2015).
Köhler, B., Grothoff, M., Gutberlet, M. & Preim, B. Bloodline: A system for the guided analysis of cardiac 4D PC-MRI data. Comput. Graph. 82, 32–43 (2019).
Bock, J., Kreher, B. W., Hennig, J. & Markl, M. Optimized pre-processing of time-resolved 2D and 3D Phase Contrast MRI data. Proc. 15th Annu. Meet. ISMRM 15, 3138 (2007).
Tyszka, J. M., Laidlaw, D. H., Asa, J. W. & Silverman, J. M. Three-dimensional, time-resolved (4D) relative pressure mapping using magnetic resonance imaging. J. Magn. Reson. Imaging 12, 321–329 (2000).
Bürk, J. et al. Evaluation of 3D blood flow patterns and wall shear stress in the normal and dilated thoracic aorta using flow-sensitive 4D CMR. J. Cardiovasc. Magn. Reson. 14, 1–11 (2012).
Meuschke, M., Köhler, B., Preim, U., Preim, B. & Lawonn, K. Semi-automatic Vortex Flow Classification in 4D PC-MRI Data of the Aorta. Comput. Graph. Forum 35, 351–360 (2016).
Moon, J. C. et al. Myocardial T1 mapping and extracellular volume quantification: A Society for Cardiovascular Magnetic Resonance (SCMR) and CMR Working Group of the European Society of Cardiology consensus statement. J. Cardiovasc. Magn. Reson. 15, 1–12 (2013).
Gharib, M., Rambod, E., Kheradvar, A., Sahn, D. J. & Dabiri, J. O. Optimal vortex formation as an index of cardiac health. Proc. Natl. Acad. Sci., https://doi.org/10.1073/pnas.0600520103 (2006).
Jeong, J. & Hussain, F. On the identification of a vortex. J. Fluid Mech., https://doi.org/10.1017/S0022112095000462 (1995).
Hunt, J. C. R. ‘Vorticity and vortex dynamics in complex turbulent flows.’ Trans. Can. Soc. Mech. Eng. 11.1, 21–35 (1987).
Schafhitzel, T. Topology-Preserving λ -based Vortex Core Line Detection for Flow Visualization. Eurographics/ IEEE-VGTC Symp. Vis. 27 (2008).
Strecker, C., Harloff, A., Wallis, W. & Markl, M. Flow-sensitive 4D MRI of the thoracic aorta: Comparison of image quality, quantitative flow, and wall parameters at 1.5 T and 3 T. J. Magn. Reson. Imaging 36, 1097–1103 (2012).
Liu, X. et al. A numerical study on the flow of blood and the transport of LDL in the human aorta: the physiological significance of the helical flow in the aortic arch. AJP Hear. Circ. Physiol. 297, H163–H170 (2009).
Alfakih, K. et al. Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences. J. Magn. Reson. Imaging https://doi.org/10.1002/jmri.10262 (2003).
Hope, M. D. et al. Post-stenotic dilation: Evaluation of ascending aortic dilation with 4D flow MR imaging. Int. J. Cardiol. 156, e40–e42 (2012).
Kauhanen, S. P. et al. Aortic dilatation associates with flow displacement and increased circumferential wall shear stress in patients without aortic stenosis: A prospective clinical study. J. Magn. Reson. Imaging 50, 136–145 (2019).
Bissell, M. M. et al. Aortic Dilation in Bicuspid Aortic Valve Disease: Flow Pattern Is a Major Contributor and Differs with Valve Fusion Type. Circ Cardiovasc Imaging 6 (2013).
Markl, M. et al. Time-Resolved 3-Dimensional Velocity Mapping in the Thoracic Aorta. J. Comput. Assist. Tomogr. 28, 459–468 (2004).
Acknowledgements
The authors like to thank Stefanie Quade for lecturing this manuscript. The study was supported by a DFG grant (GU 777/4-1 AOBJ 629068 and GR 4617/2-1 AOBJ 629069).
Author information
Authors and Affiliations
Contributions
S.E.: Conceptualization and methodology of the study, data acquisition, measurements, visualization, evaluation of the software Bloodline, statistical analysis, writing the paper, revision of the paper, corresponding author. J.D.: data acquisition, measurements. B.K.: Invention and developing of the software Bloodline. B.P.: Review of the paper, invention and developing of the software Bloodline. B.B.: Review of the paper and revision of the methods section. Support regarding the software Bloodline. B.R.: Review of the paper, post-processing. B.J.: Development of the k-t GRAPPA sequence. C.S.: Support regarding the CENTRA sequence, revision of the methods section, review of the paper. S.K.: Statistical analysis. M.Gro: Review of the paper, conceptualization and methodology of the study, data acquisition, senior author. M.Gut: Review of the paper, conceptualization and methodology of the study, senior author.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ebel, S., Dufke, J., Köhler, B. et al. Automated Quantitative Extraction and Analysis of 4D flow Patterns in the Ascending Aorta: An intraindividual comparison at 1.5 T and 3 T. Sci Rep 10, 2949 (2020). https://doi.org/10.1038/s41598-020-59826-2
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-020-59826-2
This article is cited by
-
Restoration of flow in the aorta: a novel therapeutic target in aortic valve intervention
Nature Reviews Cardiology (2024)
-
Quantitative normal values of helical flow, flow jets and wall shear stress of healthy volunteers in the ascending aorta