Abstract
The effects of gradual vascular occlusion on the blood supply of perfused areas are poorly described. Information relating to the comparison of flap monitoring techniques is lacking. Varying stenotic conditions (0%, 25%, 50%, 75% and 100%) were generated on purpose at the A. and V. femoralis in the rat model. Analyses included flowmeter, simultaneous laser-Doppler flowmetry and tissue spectrophotometry (O2C) and indocyanine green- (ICG-) videoangiography with integrated FLOW 800 tool. A Random Forests prediction model was used to analyse the importance of each method to diagnose the stenotic conditions. The ability to discriminate and to accurately estimate the probability of stenosis was assessed by Receiver Operating Characteristic (ROC) curves and calibration plots. Blood flow changes for all modalities were described in detail. Flowmeter displayed earliest a linear decrease as a result of increasing stenosis. A stenosis of 50% degrees was most difficult to detect correctly. The combination of flowmeter and ICG-videoangiography showed high diagnostic power for each stenotic situation (area under the ROC > 0.79). Flowmeter and ICG-videoangiography showed to be most relevant in detection of varying stenotic conditions and may change the clinical outcome. The O2C showed less effect on varying stenotic situations as the only surface monitoring device.
Similar content being viewed by others
Introduction
The microvascular free tissue transfer has revolutionised reconstructive microsurgery and has improved the patients’ functional outcome1. Despite the well documented co-morbidities of patients undergoing head and neck surgery, free flap surgery has contributed to a continuous improvement in the patient-reported and clinical outcomes2,3,4. Due to the significant patient co-morbidities and the complexity of the operations, this patient cohort has a high free flap failure rate. A successful outcome necessitates a detailed pre-, intra- and postoperative patient screening in addition to the free flap monitoring, for at least the first 72 hours, postoperatively. The reasons for free flap failures are commonly vascular (thrombosis, endothelial damage) or extravascular, (hematoma, kinking, twisting) in origin. Flap necrosis invariably leads to a prolonged hospital stay and may result in mortality due to poor physiological reserves. There is a definitive window of surgical intervention that must not be missed, for the salvage operation to be successful5,6. For this reason, the routine postoperative free flap monitoring has to be effective, reproducible and reliable. Free flap monitoring techniques vary and may include the clinical observation, pinprick testing and handheld Doppler7. More sophisticated, objective methods include simultaneous laser-Doppler flowmetry and tissue spectrophotometry, tissue oximetry monitoring, infrared thermography and indocyanine green- (ICG-) videoangiography8,9,10,11,12,13,14,15.
The changing nature of surgical training, with restricted resident work hours, may affect the continuity of care and render essential the need for close monitoring16. Apart from predictable and valid, the monitoring technique must be cost-effective to the clinical team. ICG-videoangiography was introduced into reconstructive free flap surgery in the last 20 years. More recently, this technique was also applied for the immediate analysis of patency following microvascular anastomosis or to predict the incidence of necrosis on the basis of only one measurement as an integrated system in the operating microscope15,17,18.
The purpose of this study was to evaluate the outcome of varying stenotic vascular situations, on the local and distal flap perfusion, with three established modalities in a well described rat model19.
Results
Descriptive analysis
A total of 20 rats were operated successfully on both hind legs (left and right) and all animals survived the operation. All animals tolerated the intravenous injection of ICG well and no allergic reactions occurred. The median values with range (minimum-maximum) of the assessed parameters are given in Supplementary Tables S1–4. For illustration, the results are displayed as Boxplots in Figs. 1 and 2. A total of 520 flowmeter measurements were performed and 520 ROIs in ICG-videoangiography with corresponding FLOW 800 analyses have been set and analysed in 40 A. and V. femoralis and hind legs.
Flowmeter analyses with varying arterial or venous stenotic situations of the A. or V. femoralis showed at all three areas of measurements (pre-, poststenotic and accompanying A. or V. femoralis) an almost linear decrease of blood flow (Supplementary Table S1 and Fig. 1A,B). Median (range) arterial prestenotic values for 0%, 25%, 50%, 75% and 100% degrees of stenosis at the A. femoralis were 1.86 ml/min (0.42–3.24), 1.39 ml/min (0.20–2.94), 0.92 ml/min (0.22–2.74), 0.36 ml/min (0.08–1.66) and 0.00 ml/min (0.00–0.34), respectively. Median (range) arterial poststenotic values for 25%, 50%, 75% and 100% degrees of stenosis at the A. femoralis were 1.18 ml/min (0.26–2.76), 0.70 ml/min (0.14–2.44), 0.20 ml/min (0.04–1.40) and 0.00 ml/min (0.00–0.30), respectively. Median (range) venous prestenotic values for 0%, 25%, 50%, 75% and 100% degrees of stenosis at the V. femoralis were 1.58 ml/min (0.46–2.86), 1.22 ml/min (0.20–2.94), 0.85 ml/min (0.14–2.62), 0.29 ml/min (0.06–1.32) and 0.01 ml/min (0.0–0.7), respectively. Median (range) venous poststenotic values for 25%, 50%, 75% and 100% degrees of stenosis at the V. femoralis were 1.05 ml/min (0.20–2.78), 0.69 ml/min (0.10–2.48), 0.21 ml/min (0.04–1.28) and 0.01 ml/min (0.00–0.36), respectively.
Simultaneous laser-Doppler flowmetry and tissue spectrophotometry (O2C) measurements showed a decrease of all variables in superficial (S) and deep (D) detection planes in cases of increasing stenosis of the A. femoralis. SO2 D and Hb D showed an increase of values in cases of stenosis of the V. femoralis (58.0 AU (21.0–88.0) to 64.0 AU (0.0–96.0) and (70.5 AU (36.0–118.0) to 76.0 AU (32.0–124.0), respectively). All other variables showed a decrease in superficial and deep detection planes (Supplementary Table S2 and Fig. 1C–F).
The quantitative analysis of ICG-videoangiography showed an acute decrease of all variables (1st and 2nd maximum, maximal in- and decrease, and area under the curve (AUC)) at 75% and 100% degrees of stenosis at the A. femoralis in the prestenotic ROIs. In the poststenotic ROIs a valley of all variables was recognized at 25% degrees of stenosis at the A. femoralis. After 75% degrees of stenosis at the A. femoralis and more a continuous decrease of all variables was seen, but 0 arbitrary units [AU] was never reached at any degree of stenosis. At 25% degrees of stenosis at the A. femoralis the values for all variables showed an increase, followed by a nearly linear decrease in the corresponding V. femoralis (Supplementary Table S3 and Fig. 2).
All pre- and poststenotic variables showed a continuous decrease at 50% and 75% degrees of stenosis at the V. femoralis after a slight increase at 25% degrees of stenosis. At 100% degrees of stenosis at the V. femoralis an increase of 1st and 2nd maximum was seen. ICG-videoangiography AUC showed a continuous decrease at 50–100% stenosis of the V. femoralis. All variables in the corresponding A. femoralis were comparable to the control measurement without any stenosis (0%) at any degree of stenosis of the V. femoralis (Supplementary Table S4 and Fig. 2).
In summary, Tables 1 and 2 show the correlation coefficients according to Spearman in an analysis between the arterial and venous degree of stenosis and the analysed parameter for each modality.
Diagnosis of stenotic condition
The variable importance measures of all three modalities used in the Random Forests prediction model showed a predominant importance of flowmeter and ICG-videoangiography analyses, thereby suggesting the exclusion of O2C measurements from the full model (Fig. 3A). The ranking of importance measures of flowmeter and ICG-videoangiography analyses stayed mainly unaffected in such a reduced model (Fig. 3B). The discrimination and calibration, measured by the area under the ROC curves (AUROC) and calibration plots, even slightly improved for the reduced model by the omission of O2C measurements from the full model (Fig. 4A,B). In summary, no stenosis and 100% degrees of stenosis were best predicted (AUROC 0.999 and 0.980 respectively) by the reduced model. Most difficult prediction remained a stenosis of 50%, which still resulted in a high AUROC of 0.801 (Fig. 4A). The calibration plots show a considerable agreement between predicted probabilities and relative frequencies of stenotic conditions (Fig. 4B). The Random Forest prediction model takes the flowmeter, the ICG-videoangiography and the O2C measurements as input and returns the likelihood that a stenosis is of degree 0%, 25%, …, or 100% as output. The distributions of these estimated likelihoods are presented in Fig. 4C,D, grouped by the actually observed level of stenosis. With a true level of 0%, for example, the stenoses are most likely classified as 0% by the prediction model. With a true level of 25% or 50% however, the prediction model returns equally high likelihoods for a classification into both of these levels. These results provide a more in-depth exploration of the discriminatory power of the models than the ROC analysis. The latter produced high AUC values as the classes 25% and 50% could still be very well discriminated from the other classes. The differentiation between 25% and 50% is more difficult though, which is shown by Fig. 4C (full model) and 4D (reduced model). Cohen’s (weighted) Kappa was computed in an additional analysis to measure agreement between observed and predicted stenotic conditions. It reached values of Κ = 0.815 and Κ = 0.823 for the full and reduced model, respectively (Table 3).
Discussion
The current study analysed the quantitative and qualitative results, as well as the importance and validity of three established techniques in perfusion analyses for detection of different degrees of venous and arterial stenoses of the nourishing vessel in an established rat hind leg model19. Vascular compromise is likely to progress from partial to full obstruction (e.g. thrombosis and hematoma formation). In this process the time course may differ. In times of increasingly elderly and ill (ASA > II) patients needing local or free flap surgery2, early, reliable, reproducible and correct monitoring techniques are needed. Nevertheless, only a minority of studies have addressed the effect of partial to full occlusion of vascular supply and on the effectiveness of its detection in a methodically comparative study.
The technique for ICG-videoangiography as well as for O2C measurements are described elsewhere in detail8,20,21,22. The technical and clinical advantages of objective monitoring methods are generally obvious and well described, but most studies are focussing on the qualitative value of visualization of patency of microvascular anastomosis or tissue perfusion alone14,23. While most other studies focus on qualitative observations, we included a highly sophisticated statistical analysis based on machine learning (Random Forest prediction model) to compare the importance of each individual parameter that was quantitatively registered by each analysed modality. What is more, not two but three different modalities were analysed in this study. Interestingly, our results suggest that applying all three modalities for monitoring and detection of varying stenotic conditions was not superior to the combination of flowmeter and ICG-videoangiography alone (Figs. 3 and 4). While O2C data correctly predict the trend towards an increasing grade of stenosis, especially venous stenosis showed less significant correlation between degree of stenosis and registered parameter.
In detail, the simple flowmeter analyses showed interestingly the highest importance in the Random Forest prediction model. As described earlier, the combination with ICG-videoangiography can increase the validity of the assessment17. But as stated by Nasser et al. the detection of stenotic conditions is difficult. They reported significant fluorescence changes at 85% stenosis using the LifeCell SPY-Elite (LifeCell) system19. In contrast, we used the integrated software tool FLOW 800. With that, an initial drop of 1st and 2nd maxima were registered already with stenoses of 50% degrees. Our ROC curve analyses gave interesting insights, how validly each stenotic condition was detected. The results revealed that it was most difficult to correctly differentiate a stenosis of 50% degrees from the other stenotic degrees (Fig. 4A,C,D). Nevertheless, area under the ROC curve (AUROC) was 0.801 (without O2C), representing a high diagnostic power. AUROC for 25% and 75% degrees of stenosis were even higher (without O2C 0.892 and 0.943, respectively). So, in contrast to Nasser et al., even 25% degree of stenosis were correctly diagnosed with a high diagnostic power using a combination of flowmeter and ICG-videoangiography. Clinically ≥75% degrees of stenosis may begin to become relevant for distal organ perfusion, function and survival. Yang et al. described a more turbulent blood flow with a consecutively increased risk of distal embolization in a stenotic middle cerebral artery model24. Schoenberg et al. described a mean flow reduction of 50% at a stenotic degree of 90% at the renal artery25. A decrease of blood flow was also described by Nasser et al. In the presented study, the results of the flowmeter had the highest importance in the Random Forest prediction model. Further, also the parameters “O2CVelo or O2CFlow” had the highest importance in the Random Forest prediction model within the O2C measurements. These observations made by us and others reflect the significance of blood flow behavior itself in the difficult assessment of the (stenotic) vessel. Therefore, its valid detection is of great importance. Long-term effects were not analysed in this study, since the immediate detection and effects on vascular supply and flap physiology were of main interest.
Further, ICG-videoangiography in venous conditions showed also good results with good importance in the Random Forest prediction model, indicated by the values “ICGVenPre1Max, ICGVenPre2Max and ICGVenPreMaxdecr”. This is clinically of great importance, because venous congestion remains to be the leading reason for flap failure26. Early detection of flaps prone to fail is mandatory for successful flap salvage within a tissue/flap-dependent critical time of congestion, since blood flow changes occur earlier than clinical signs appear6,27,28. What is more, flap failure does not seem to be an “all or none phenomenon” as described by Weinzweig et al.29. They described the occurrence of slow flap failure as a result of microcirculatory disturbances due to persistent microthrombembolism. In this setting, the correct and valid detection of stenotic disease, that might facilitate microthrombembolism, is as crucial as a technically correct performed microvascular anastomosis.
In this context, the direct methods analysing blood flow changes in the pre- and post-stenotic area showed to be superior, in comparison to methods that are applied distally at the transplant´s surface (as seen in the O2C measurements in our study). Further, a single measurement is not enough and multiple measurements are needed to estimate the course more correctly while applying O2C30. In conclusion of the results for O2C measurements, the physiologic reaction on blood flow compromising situations results in delayed and hardly recognizable, and distinguishable changes in oxygen saturation, haemoglobin level, velocity and blood flow. Therefore, prediction of the clinical course may be limited though31. On the other hand, reliable and valid surface-based methods would be easier to use on the ward or intensive care unit and add an important, objective method to common clinical observation and tests including recapilarisation time, pinprick testing and surgical APGAR score7,32.
Akita et al. described a promising approach to increase the predictability of vascular nourishing of free flaps on the ward. They combined regional oxygen saturation monitoring with ICG-angiography. The regional oxygen saturation sensor was placed onto an intraoperatively localised ICG angiographically early-staining area33. With that Akita et al. described an AUROC of 1.0 for their developed oxygen saturation index. This result was further superior to simultaneously registered blood glucose measurement index.
On the basis of our results we will advocate a combination of methods as the safest way for reliable monitoring of critical and/or buried flaps. We are aware that our good results were reduced by the O2C measurements, but clinically the application of invasive methods as ICG-videoangiography or flowmeter analysis at the anastomotic region is not always immediately possible. Implantable Cook-Schwartz Doppler probes or flow coupler devices might be therefore the regional methods of choice in daily usage and improve salvage rates9,34. Moreover, a combination of two established methods is enough and an additional third method is not always beneficial.
The presented study is based on a small animal model. Potential translation to clinical care is controversial and therefore our results have to be interpreted with caution. Nevertheless, rats remain the gold standard in the field of microsurgical research35 and during a time of recurrent ethical questions and surveys, preclinical studies still have to be performed.
The main disadvantage of ICG (video-) angiography is its cost, especially in the case of an integrated system in an operating microscope, as previously described by others36. Based on our experience, the possibility to decrease free flap failure and to increase salvage rates outweigh the perceived costs. Further, we only described values for the OPMI Pentero integrated near-infrared videoangiography detection system with the integrated FLOW 800 tool (INFRARED 800; Carl Zeiss Meditec AG; Oberkochen, Germany); for other systems, results may differ. The next logical step from this work, will be the analysis of the degree of vascular stenosis. A study in a larger animal model with different types of composite flaps would be necessary to increase the clinical relevance and translation to clinical care.
Methods
Ethical statement
The study conformed with current German regulations, with guidelines for animal welfare, and to the international principles of laboratory animal care. The local government approved the animal experiments (Regierung Oberbayern; AZ.: 55.2-1-54-2532-85-15). The animals were housed in filter-top cages under hygienic conditions according to the guidelines of the Federation of Laboratory Animal Science Associations (FELASA). Water and standard rodent diet (Altromin; Altromin Spezialfutter GmbH & Co. KG; Lage, Germany) were given ad libitum. The operations were final experiments and no postoperative visits were necessary.
Operation
Male Wistar rats weighing between 300 and 350 g were initially anesthetized with intraperitoneal administration of ketamine 10% (1 mL/kg/weight) and xylazine 2% (0.25 mL/kg/weight) supplemented by intravenous application of one-eighth doses 10% ketamine when needed using an inserted venous microcatheter (Premicath; VYGON GmbH & Co. KG; Aachen, Germany) in the external jugular vein according to the femoral vein access as previously described37,38.
All 20 rats were operated in a standardized fashion. The inguinal region and the hind legs on both sides were shaved. After a sharp inguinal incision of the skin and subcutaneous tissue, the femoral neurovascular sheath (femoral artery, vein and nerve) was exposed and the vessels were circumferentially prepared. All vascular branches between the inguinal ligament and epigastric vessels were ligated. Following increasing degrees of arterial and venous stenosis (each 0%, 25%, 50%, 75% and 100%) were applied separately by tightening a perivascular suture under visual control with millimetre gauge (Fig. 5)17. After each step of vascular stenosis different flow measurements and analyses (flowmeter, simultaneous laser-Doppler flowmetry and tissue spectrophotometry and ICG-videoangiography with integrated FLOW 800 tool) were performed.
Analyses
After arterial or venous stenosis (0%, 25%, 50%, 75% or 100%) has been set, ten minutes were waited to establish a regular and adapted flow. Following that, blood flow was measured pre-, post-stenotic and at the accompanying arterial or venous vessel for 30 seconds with a flowmeter (TS-420; Transonic System Inc.; Ithaca, NY, USA). An average of five measurements within these 30 seconds was taken as the result, in order to reduce a bias due to movement or spasm during measurement. After flowmetry, intraoperative ICG-videoangiography of the femoral vein and artery were performed using the operating microscope type OPMI Pentero with integrated near-infrared videoangiography detection system and FLOW 800 tool (INFRARED 800; Carl Zeiss Meditec AG; Oberkochen, Germany)17,39. The ICG (ICG-PULSION; Pulsion Medical System AG; Munich, Germany) was injected (0.3 mg/kg body weight, 25 mg dissolved in 5 ml sterile water) as a bolus into the external jugular vein using the inserted venous microcatheter (Premicath; VYGON GmbH & Co.KG; Aachen, Germany). The ICG-videoangiography was conducted at a fixed working distance of 300 mm and with a 5-fold magnification. Registration of the emission signal was started immediately after injection of the dye and was recorded for a period of 120 seconds with 25 images per second. All data were immediately analysed with a mathematical software tool FLOW 800 (FLOW 800; Carl Zeiss AG; Oberkochen, Germany) and colour encoded with respect to fluorescence over time. The fluorescence intensity was recorded as arbitrary units [AU]21. In cases of 0% stenosis only two regions of interest (ROI) were positioned, one in the femoral vein and one in the femoral artery (number of ROIs for 0% stenosis n = 2). In all other stenotic degrees (25%, 50%, 75% and 100%) three ROIs were positioned: pre-, poststenotic and at the accompanying artery or vein. For each ROI 5 defined parameter were registered: 1. and 2. Maximum, maximal increase and decrease, area under the curve for a period of 120 seconds.
Finally, simultaneous laser-Doppler flowmetry and tissue spectrophotometry of the ipsilateral hind leg (O2C; LEA Medizintechnik GmbH; Giessen, Deutschland) was performed8. For the measurement, the skin was cleaned and dried and the O2C sensor was positioned in the centrally of the ipsilateral hind leg (Fig. 5). The measurement was performed for 30 seconds to reduce bias due to movements. With one measurement oxygen saturation (SO2, in arbitrary units, [AU]), haemoglobin level (Hb, in [AU]), velocity (velo, in [AU]) and blood flow (flow, in [AU]) were non-invasively calculated superficially at the skin surface (S) and at a depth (D) of 8 mm. The rat was sacrificed in deep anaesthesia with intracardiac injection of pentobarbital 60 mg/kg body weight (Narcoren, Fa. Rhone Merieux GmbH, Laupheim) following a standard protocol40 after all measurements on both hind legs have been performed.
Following coding key was used for illustration in Fig. 3 and Tables 2 and 3: modality (flow, O2C or ICG) + vessel (stenosis at femoral artery or vein) + localization of flowmeter probe or ROIs in ICG-videoangiography (pre-, poststenotic or accompanying) + parameter.
Statistical methods
Study data were prospectively collected and analysed. The distribution of continuous data is presented by median (range) and illustrated by boxplots. Computations have been performed using IBM SPSS Statistics for Windows, version 23.0 (IBM Corp., Armonk, New York, USA). Machine Learning was used to enable a flexible modelling of the potentially complex and interacting relations between flowmeter, ICG-videoangiography and O2C measurements to stenotic conditions. Random Forests based on conditional inference trees provide a corresponding model structure, ensure unbiased variable selection and are well-known for their outstanding prediction performance41. As each tree in a Random Forest is fit to a random sample of the data, the left-out observations can be used for an unbiased assessment of performance measures and model diagnostics. Additional test data for validation was therefore not necessary in the present investigations. Variable importance measures were computed to explore the relevance of flowmeter, ICG-videoangiography and O2C measurements for the diagnosis of stenotic conditions42. The discriminatory and diagnostic ability of the prediction model was assessed by the area under the curve (AUROC) of the Receiver Operating Characteristic (ROC) curve. Calibration plots were used to investigate the agreement of predicted probabilities and observed relative frequencies of stenotic conditions. In similarity to that, the agreement of observed and predicted stenotic conditions was quantified by Cohen’s Kappa. The Random Forests model consisted of 5000 trees to achieve stability of results and to increase prediction performance. It was fit to the data using R 3.5.0 (R Foundation for Statistical Computing, Vienna, Austria) and the function cforest(), leaving all meta parameters of the function, except for the number of trees, unchanged41. The degrees of stenosis served as multinomial outcome. The predictor variables of the model are listed in Fig. 3.
Data availability
All data generated or analyzed during this study are included in this published article (and its Supplementary Information files).
Change history
28 February 2020
An amendment to this paper has been published and can be accessed via a link at the top of the paper.
References
Clancy, K. et al. Outcomes of microvascular free tissue transfer in twice-irradiated patients. Microsurgery, https://doi.org/10.1002/micr.30154 (2017).
Kesting, M. R. et al. Use of microvascular flap technique in older adults with head and neck cancer: a persisting dilemma in reconstructive surgery? J Am Geriatr Soc 59, 398–405, https://doi.org/10.1111/j.1532-5415.2011.03315.x (2011).
Ayala, C. & Blackwell, K. E. Protein C deficiency in microvascular head and neck reconstruction. Laryngoscope 109, 259–265, https://doi.org/10.1097/00005537-199902000-00016 (1999).
Srikanthan, K., Viswanathan, N. & Yuen, J. C. Free-flap failure in thrombophilia: case report and systematic review of the literature. Annals of plastic surgery 71, 675–681, https://doi.org/10.1097/SAP.0b013e3182586b47 (2013).
Farwell, D. G. et al. Predictors of perioperative complications in head and neck patients. Arch Otolaryngol Head Neck Surg 128, 505–511, https://doi.org/10.1001/archotol.128.5.505 (2002).
Mücke, T., Schmidt, L. H., Fichter, A. M., Wolff, K.-D. & Ritschl, L. M. Influence of venous stasis on survival of epigastric flaps in rats. The British journal of oral & maxillofacial surgery 56, 310–314, https://doi.org/10.1016/j.bjoms.2018.01.019 (2018).
Disa, J. J., Cordeiro, P. G. & Hidalgo, D. A. Efficacy of conventional monitoring techniques in free tissue transfer: an 11-year experience in 750 consecutive cases. Plastic and reconstructive surgery 104, 97–101, https://doi.org/10.1097/00006534-199907000-00013 (1999).
Mücke, T. et al. Identification of perioperative risk factor by laser-doppler spectroscopy after free flap perfusion in the head and neck: a prospective clinical study. Microsurgery 34, 345–351, https://doi.org/10.1002/micr.22206 (2014).
Um, G. T. et al. Implantable Cook-Swartz Doppler probe versus Synovis Flow Coupler for the post-operative monitoring of free flap breast reconstruction. Journal of plastic, reconstructive & aesthetic surgery: JPRAS 67, 960–966, https://doi.org/10.1016/j.bjps.2014.03.034 (2014).
Mericli, A. F. et al. A prospective clinical trial comparing visible light spectroscopy to handheld Doppler for postoperative free tissue transfer monitoring. Plastic and reconstructive surgery, https://doi.org/10.1097/PRS.0000000000003600 (2017).
Koolen, P. G. et al. Does Increased Experience with Tissue Oximetry Monitoring in Microsurgical Breast Reconstruction Lead to Decreased Flap Loss? The Learning Effect. Plastic and reconstructive surgery 137, 1093–1101, https://doi.org/10.1097/01.prs.0000481071.59025.82 (2016).
Just, M. et al. Monitoring of microvascular free flaps following oropharyngeal reconstruction using infrared thermography: first clinical experiences. Eur Arch Otorhinolaryngol 273, 2659–2667, https://doi.org/10.1007/s00405-015-3780-9 (2016).
Teven, C. M., Ooi, A. S. H., Inbal, A. & Chang, D. W. Implantable Doppler monitoring of buried free flaps during vascularized lymph node transfer. J Surg Oncol 116, 371–377, https://doi.org/10.1002/jso.24655 (2017).
Hitier, M., Cracowski, J. L., Hamou, C., Righini, C. & Bettega, G. Indocyanine green fluorescence angiography for free flap monitoring: A pilot study. J Craniomaxillofac Surg 44, 1833–1841, https://doi.org/10.1016/j.jcms.2016.09.001 (2016).
Mücke, T. et al. Indocyanine green videoangiography-assisted prediction of flap necrosis in the rat epigastric flap using the flow(R) 800 tool. Microsurgery 37, 235–242, https://doi.org/10.1002/micr.30072 (2017).
Patel, U. A. et al. Free Flap Reconstruction Monitoring Techniques and Frequency in the Era of Restricted Resident Work Hours. JAMA Otolaryngol Head Neck Surg 143, 803–809, https://doi.org/10.1001/jamaoto.2017.0304 (2017).
Mücke, T., Wolff, K. D., Wagenpfeil, S., Hölzle, F. & Scholz, M. Reliability of near-infrared angiography and micro-Doppler sonography for evaluating microvascular anastomoses. Plastic and reconstructive surgery 126, 1506–1514, https://doi.org/10.1097/PRS.0b013e3181f0215a 00006534-201011000-00011 [pii] (2010).
Holm, C., Mayr, M., Hofter, E., Dornseifer, U. & Ninkovic, M. Assessment of the patency of microvascular anastomoses using microscope-integrated near-infrared angiography: a preliminary study. Microsurgery 29, 509–514, https://doi.org/10.1002/micr.20645 (2009).
Nasser, A. et al. Utilizing Indocyanine Green Dye Angiography to Detect Simulated Flap Venous Congestion in a Novel Experimental Rat Model. Journal of reconstructive microsurgery 31, 590–596, https://doi.org/10.1055/s-0035-1558869 (2015).
Holm, C. et al. Intraoperative evaluation of skin-flap viability using laser-induced fluorescence of indocyanine green. Br J Plast Surg 55, 635–644, https://doi.org/10.1054/bjps.2002.3969 (2002).
Mücke, T. et al. Objective qualitative and quantitative assessment of blood flow with near-infrared angiography in microvascular anastomoses in the rat model. Microsurgery 33, 287–296, https://doi.org/10.1002/micr.22095 (2013).
Giunta, R. E. et al. Prediction of flap necrosis with laser induced indocyanine green fluorescence in a rat model. Br J Plast Surg 58, 695–701, https://doi.org/10.1016/j.bjps.2005.02.018 (2005).
Bigdeli, A. K. et al. Indocyanine Green Fluorescence for Free-Flap Perfusion Imaging Revisited: Advanced Decision Making by Virtual Perfusion Reality in Visionsense Fusion Imaging Angiography. Surg Innov 23, 249–260, https://doi.org/10.1177/1553350615610651 (2016).
Yang, Y., Grosset, D. G., Li, Q., Shuaib, A. & Lees, K. R. Turbulence and circulating cerebral emboli detectable at Doppler ultrasonography: a differentiation study in a stenotic middle cerebral artery model. AJNR Am J Neuroradiol 23, 1229–1236 (2002).
Schoenberg, S. O., Bock, M., Kallinowski, F. & Just, A. Correlation of hemodynamic impact and morphologic degree of renal artery stenosis in a canine model. J Am Soc Nephrol 11, 2190–2198, 1046-6673/1112-2190 (2000).
Chen, K. T. et al. Timing of presentation of the first signs of vascular compromise dictates the salvage outcome of free flap transfers. Plastic and reconstructive surgery 120, 187–195, https://doi.org/10.1097/01.prs.0000264077.07779.50 (2007).
Mirzabeigi, M. N. et al. Free flap take-back following postoperative microvascular compromise: predicting salvage versus failure. Plastic and reconstructive surgery 130, 579–589, https://doi.org/10.1097/PRS.0b013e31825dbfb7 (2012).
Gimbel, M. L., Rollins, M. D., Fukaya, E. & Hopf, H. W. Monitoring partial and full venous outflow compromise in a rabbit skin flap model. Plastic and reconstructive surgery 124, 796–803, https://doi.org/10.1097/PRS.0b013e3181b03768 (2009).
Weinzweig, N. & Gonzalez, M. Free tissue failure is not an all-or-none phenomenon. Plastic and reconstructive surgery 96, 648–660, https://doi.org/10.1097/00006534-199509000-00018 (1995).
Mücke, T. et al. Changes of perfusion of microvascular free flaps in the head and neck: a prospective clinical study. The British journal of oral & maxillofacial surgery 52, 810–815, https://doi.org/10.1016/j.bjoms.2014.07.001 (2014).
Ritschl, L. M. et al. Multimodal analysis using flowmeter analysis, laser-Doppler spectrophotometry, and indocyanine green videoangiography for the detection of venous compromise in flaps in rats. J Craniomaxillofac Surg 46, 905–915, https://doi.org/10.1016/j.jcms.2018.03.016 (2018).
Ettinger, K. S. et al. Application of the Surgical Apgar Score to Microvascular Head and Neck Reconstruction. J Oral Maxillofac Surg 74, 1668–1677, https://doi.org/10.1016/j.joms.2016.02.013 (2016).
Akita, S. et al. Regional Oxygen Saturation Index: A Novel Criterion for Free Flap Assessment Using Tissue Oximetry. Plastic and reconstructive surgery 138, 510e–518e, https://doi.org/10.1097/PRS.0000000000002498 (2016).
Lenz, Y. et al. Evaluation of the Implantable Doppler Probe for Free Flap Monitoring in Lower Limb Reconstruction. Journal of reconstructive microsurgery 34, 218–226, https://doi.org/10.1055/s-0037-1608628 (2018).
Zhang, F., Sones, W. D. & Lineaweaver, W. C. Microsurgical flap models in the rat. Journal of reconstructive microsurgery 17, 211–221, https://doi.org/10.1055/s-2001-14353 (2001).
Yeoh, M. S., Kim, D. D. & Ghali, G. E. Fluorescence angiography in the assessment of flap perfusion and vitality. Oral Maxillofac Surg Clin North Am 25, 61–66, vi, https://doi.org/10.1016/j.coms.2012.11.004 (2013).
Ritschl, L. M. et al. Ketamine-Xylazine Anesthesia in Rats: Intraperitoneal versus Intravenous Administration Using a Microsurgical Femoral Vein Access. Journal of reconstructive microsurgery 31, 343–347, https://doi.org/10.1055/s-0035-1546291 (2015).
Petry, J. J. & Wortham, K. A. The anatomy of the epigastric flap in the experimental rat. Plastic and reconstructive surgery 74, 410–413, https://doi.org/10.1097/00006534-198409000-00014 (1984).
Raabe, A., Beck, J., Gerlach, R., Zimmermann, M. & Seifert, V. Near-infrared indocyanine green video angiography: a new method for intraoperative assessment of vascular flow. Neurosurgery 52, 132–139; discussion 139, https://doi.org/10.1097/00006123-200301000-00017 (2003).
Close, B. et al. Recommendations for euthanasia of experimental animals: Part 1. DGXI of the European Commission. Lab Anim 30, 293–316, https://doi.org/10.1258/002367796780739871 (1996).
Hothorn, T., Hornik, K. & Zeileis, A. Unbiased Recursive Partitioning: A Conditional Inference Framework. Journal of Computational and Graphical Statistics 15, 651–674, https://doi.org/10.1198/106186006x133933 (2006).
Hapfelmeier, A., Hothorn, T., Ulm, K. & Strobl, C. A New Variable Importance Measure for Random Forests with Missing Data. Stat Comput 24, 21–34, https://doi.org/10.1007/s11222-012-9349-1 (2014).
Acknowledgements
This work was supported by the German Research Foundation (DFG) and the Technical University of Munich within the funding program Open Access Publishing.
Author information
Authors and Affiliations
Contributions
T.M. study design/conduction, Data acquisition and interpretation, contributor in writing the manuscript, A.H. Statistical analyses, contributor in writing the manuscript, L.H.S. operations, Data acquisition and interpretation, revising the manuscript, A.M.F. study design, contributor in writing the manuscript and figure designing, A.K. contributor in writing the manuscript, revising the manuscript, K.-D.W. Study conception, revision of manuscript, L.M.R. study design/conduction, operations, analyses and major contributor in writing the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests. None of the authors has a financial interest in any of the products, devices, or drugs mentioned in this manuscript.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Mücke, T., Hapfelmeier, A., Schmidt, L.H. et al. A comparative analysis using flowmeter, laser-Doppler spectrophotometry, and indocyanine green-videoangiography for detection of vascular stenosis in free flaps. Sci Rep 10, 939 (2020). https://doi.org/10.1038/s41598-020-57777-2
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-020-57777-2
This article is cited by
-
Methods of ex vivo analysis of tissue status in vascularized composite allografts
Journal of Translational Medicine (2023)
-
Effects of endothelial defects and venous interposition grafts on the acute incidence of thrombus formation within microvascular procedures
Scientific Reports (2021)
-
Usefulness of a finger-mounted tissue oximeter with near-infrared spectroscopy for evaluating the intestinal oxygenation and viability in rats
Surgery Today (2021)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.