Abstract
We introduce a new approach for a continuous loop tendon-graft preparation, benchmarking it against established graft preparation techniques widely used in conjunction with non-adjustable interference screw fixation. A four-strand bovine tendon graft was prepared using the following graft preparation techniques: standard graft using the baseball stitch technique (M-tech group); continuous loop graft using the GraftLinkTM technique (Arthrex-tech group); continuous loop graft using the Kessler anastomosis technique (Kessler-tech group); and continuous loop graft using a Double-Z anastomosis technique (Double Z-tech group). Each group of eight specimens underwent cyclic loading followed by a load-to-failure test. The M–technique yielded a smaller graft diameter (8.4 ± 0.5 mm) compared to the statistically equivalent diameters of the three continuous loop techniques (8.9 ± 0.6 mm of Arthrex-tech group, 9.1 ± 0.4 mm of Kessler-tech group and 9.2 ± 0.6 mm of Double Z-Tech group). The continuous loop grafts formed by the Double Z-Technique showed outstanding performance among the tested techniques in terms of ultimate failure load (982 ± 121 N) and cyclic elongation (3.7 ± 1.0 mm). There was no significant difference between the four groups in cyclic stiffness. Of the assessed techniques, the Arthrex technique resulted in the lowest ultimate elongation (2.0 ± 0.7 mm), followed by the Double Z-tech (4.5 ± 1.8 mm), the M-tech (5.2 ± 3.9 mm), and the Kessler-tech (5.3 ± 2.4 mm). The Arthrex-tech group (5.98 ± 0.38 min) displayed the shortest graft preparation time, followed by the M-Tech (7.94 ± 0.58 min), Kessler-tech (9.03 ± 0.39 min) and Double Z-Tech (13.29 ± 1.14 min). Double Z-Tech can improve the construct of continuous loop tendon graft with regard to mechanical performance.
Similar content being viewed by others
Introduction
The anterior cruciate ligament (ACL) has an important stabilization function for tibiofemoral stability and is one of the most frequently injured structures of the knee joint1.
Untreated ACL rupture can be associated with the occurrence of knee osteoarthritis and knee instability2. Therapeutic strategies of ACL ruptures have received extensive attention from scholars and with the development of sports medicine and arthroscopic techniques, arthroscopic anterior cruciate ligament reconstruction (ACLR) has become the most common surgical strategy for the treatment of ACL rupture3.
Currently, mainstream ACLR techniques utilize autogenous lower limb tendon grafts (i.e. hamstring tendons, quadriceps tendon) and bone patellar tendon bone (BPTB) grafts with tendon grafts being the most common4,5. Fixation with these techniques is usually achieved with suspensory or interference screw fixation with suspensory fixation being preferred by many surgeons6. Whereas conventional cortical fixation devices provide fixed length suspension, adjustable loop fixation7 introduced the possibility of individual graft length adjustment and maximization of graft-bone contact8. Depending on the surgeon’s preference, adjustable loop fixation devices can be used for femoral-only as well as for femoral and tibial graft fixation. In the latter case, the tendon explant is folded twice over the two adjustable loop fixation devices and the two tendon ends are connected to form a continuous double-loop tendon graft. This graft configuration is in contrast to the tendon-graft configuration used for more conventional fixation techniques such as interference screw tibial fixation where the two tendon ends are “baseball- stitched” and most commonly folded twice to create a tendon graft folded in M-configuration. Whereas adjustable loop fixation devices may provide sufficient fixation strength and rigidity the biomechanical behavior of continuous double-loop tendon grafts has not been reported.
In this biomechanical study, we compared a frequently applied hamstring graft preparation technique used for tibial adjustable length suspensory graft fixation7 (Arthrex-Tech: GraftLinkTM, Arthrex, Naples, FL) to a conventional M-configuration technique (M-Tech) used in tibial interference screw fixation. Meanwhile, we also explored modified graft suturing techniques (Kessler-tech and Double Z-Tech) designed to improve the construct with regard to mechanical performance in isolated biomechanical testing. It was hypothesized that continuous loop tendon grafts were superior to the M-Tech graft in mechanical properties. The second hypothesis was that a graft using Kessler-tech or Double Z-Tech can improve the construct of continuous loop tendon graft with regard to mechanical performance.
Materials and Methods
Specimen preparation
This study was performed in accordance with the National Institutes of Health guidelines for the use of experimental animals, and all animal protocols were approved by the Institutional Animal Care and Use Committee of University of Zurich. The Ethics Committee of Canton of Zurich approved the study protocol, including any relevant details. Testing was performed using 16 bovine calf tendons purchased from a local slaughterhouse, with structural and biomechanical properties similar to those of human hamstringtendons9. Specimens were preserved at −20 °C and defrosted at room temperature for 1 hour before graft preparation. During graft preparation and biomechanical testing, phosphate buffered saline (PBS; pH 7.4) was used to keep the tendon hydrated per 10 minutes. Excess soft tissues were removed from the surfaces of tendons before graft preparation. The tendons were cut in half along their long axis for the 4-folded grafts to have a diameter of approximately 8 mm.
Grouping
The 32 tendon specimens were randomly divided into four groups: 4-strand graft using baseball stitching at both tendon ends to form a double-folded tendon graft in M-configuration (M-tech), continuous loop graft using the GraftLinkTM graft preparation technique (Arthrex-tech), continuous loop graft using Kessler technique (Kessler-tech), and continuous loop graft using Double-Z technique group (Double Z-tech).
M-Technique graft preparation
As shown in Fig. 1, for the preparation of M-Tech grafts 20 mm of both tendon ends were whip stitched, the tendon was folded once over size 5 polyethylene suture (Meister & Cie AG, Bern, Switzerland), inserted into the femoral suspension loop, folded a second time to form a four-strand M-configuration tendon bundle and tied up using size 2 polyethylene suture (Orthocord, DePuy Mitek, Raynham, Massachusetts, USA).
Continuous loop grafts
Continuous loop grafts were prepared by inserting the tendon into two looped size 5 polyethylene loading sutures (Meister & Cie AG, Bern, Switzerland) twice to form a double-loop continuous tendon graft (Fig. 1). The free tendon ends were then connected using one of the three tested suturing techniques (Arthrex-tech, Kessler-tech, Double Z-tech, Fig. 2). Once stitching was complete, the sutured tendon was moved to the inside of the construct and tensioned before applying the final circular contraction suture.
GraftLink (Arthrex)technique
GraftLink (Arthrex)Technique was aided by use of guides for the continuous loop GraftLink (Arthrex, Naples, USA)10. Briefly, the graft was quadrupled symmetrically over two looped size 5 polyethylene loading sutures, and the free ends were whip-stitched and then sutured with a buried-knot technique7 (Fig. 2A).
Kessler technique
To connect the free tendon ends the Kessler technique was applied11. The suture was hereby inserted longitudinally into the tendon core for a length of 10 mm. Then the suture was passed to the side of the tendon and reinserted into its core and brought back to the end of the tendon (Fig. 2B). After repeating these steps on the second free tendon end, the two suture ends were connected using a surgeon’s knot and three alternating overhand knots.
Double-Z technique
For Double-Z graft preparation two Z-shaped stitches were used on each of the two tendon ends as shown in Figs. 2C and 3. The four suture ends were then connected with a surgeon’s knot and three alternating overhand knots.
Circular contraction suturing
After connecting the free tendon ends using one of the above described techniques, the doubled continuous tendon loop was tied up using a slight modification of the technique described by Lubowitz et al.7 with the purpose of making the tendon graft more compact and increasing its stability. The connected tendon ends were hereby first moved to the inside of the graft. Size 2 non-resorbable suture was passed through the graft four times at a distance of 10 mm from the connection of the free tendon ends. First the suture was stitched from the inside of the graft to the surface, then it was brought through all tendon strands twice, wrapped around the graft twice and brought back from the outside in, back to the other suture end. The two suture ends were connected using surgeon’s knots and the buried knot technique7. This was repeated 10 mm from the connection of the free tendon ends in the opposite direction as well as on the opposing end of the tendon graft (Fig. 4A,B).
To determine the best location of three-contraction suture, a baseline biomechanical study was performed to analyze three different locations for the stitch end. GraftLink (Arthrex) technique was selected to connect the free tendon ends. One stitch location was near the loading suture at the outside the circular contraction sutures (group A); one stitch location was under that of the circular contraction suture (Group B) and in the middle of the two circular sutures (Group C). Each locations type contains 6 samples and was tested according to the following protocol.
Biomechanical testing
Before mechanical testing, the diameter of the graft was measured using a graft sizing device (KARL STORZ SE & Co. KG, Tuttlingen, Germany). The tendon grafts were then placed into a universal testing machine (Zwick GmbH, Ennepetal, Germany) suspended with size 5 suture on both sides (Fig. 4C). After applying a preload of 10 N for 30 seconds and a subsequent preconditioning phase of 10 cycles from 10 to 100 N the graft was loaded between 50 and 250 N at a maximum rate of 1.5 mm/s for 5000 cycles12,13. The preconditioning protocol was aimed to simulate the loading regime the graft undergoes during surgical insertion. Then the tendon grafts were ramp loaded until failure at a machine speed of 20 mm/min14. Table 1 summarizes the specifics of the testing protocols.
Biomechanical analysis
Force-displacement data was recorded using TestXpert 10 software (Zwick GmbH, Ennepetal, Germany) and processed in MATLAB (Matlab R2016a, Mathworks, Natrick, USA). Ultimate failure load (N) and ultimate elongation (mm) were defined as maximum force achieved and respective elongation with elongation zeroed at the end of preconditioning. Cyclic elongation (mm) describes total graft elongation during the 5000 test cycles. Cyclic stiffness (N/mm) was measured by averaging the slope of the linear curve connecting minimum and maximum force-position of each cycle (Fig. 4D,E).
Failure mode
Sample testing was monitored by Dr. Chen Zhang and Tobias Götschi alternately and documented using a digital camera (EO Edmund, Tokyo, Japan). Four modes of sample failure were documented: Baseball stitch failure, loading ropes cut through tendon, continuous loop anastomosis (whipstitch, Kessler anastomosis and Double Z anastomosis) rupture, and continuous loop anastomosis loosening and subsequent structural failure during cyclic testing.
Graft preparation time evaluation
Time required for the preparation of the graft using either of the techniques was recorded.
Statistical analysis
Statistical results were generated by using a standard statistical software (GraphPad Prism6, CA, USA). Measurement data were expressed as the mean ± SD. Kolmogorov-Smirnov test was used to test if the values were adequately fit by Gaussian distributions. The measurement data were statistically analyzed using one-way analysis of variance for multiple group comparisons and the Dunnett t test for 2 group comparisons. The test level of α was 0.05.
Ethics approval
The Ethics Committee of Canton of Zurich approved the study protocol.
Results
The best location for the stitch end
After the baseline biomechanical assessments, we found most of the stiches loosened and failed due to the pulling force of the loading loop when the graft securing sutures were located near the loading suture (6 of 6 in group A) and under the circular contraction sutures (5 of 6 in group B). No visible gap nor anastomosis loose was observed during cyclic testing in group C. Stitch location which was in the middle of the two circular sutures was thus chosen for the experiments in this study.
Each group of 8 graft specimens was mechanically tested and analyzed (Table 2, Fig. 5). The M-tech group and Double Z-tech group each lost one specimen due to data output failure from the test machine. In the Kessler-tech group, one graft experienced anastomosis loosening and failure during cycling testing rendering its mechanical parameters invalid for further analysis. Consequently, there were 29 specimens included in final analysis.
Diameter
Each group of 8 graft specimens was mechanically tested and analyzed (Table 2, Fig. 5). The M-tech group and Double Z-tech group each lost one specimen due to data output failure from the test machine. In the Kessler-tech group, one graft experienced anastomosis loosening and failure during cycling testing rendering its mechanical parameters invalid for further analysis. Consequently, there were 29 specimens included in final analysis.
Analysis indicated that M-tech (M) group has the smaller diameter (8.43 ± 0.53 mm) than the three continuous loop graft groups (Fig. 5A). The mean diameter of Kessler-tech (K) group (9.07 ± 0.35 mm) and Double Z-tech (Z) group (9.21 ± 0.57 mm) were significantly greater than in the M-tech group (tK/M = 2.673, p = 0.0203; tZ/M = 2.668, p = 0.0205), whereas the diameter of Arthrex-tech (A) group (8.94 ± 0.56 mm) was not significantly different (tA/M = 1.788, p = 0.0971). However, there was no significant difference among the three continuous loop graft groups (tK/A = 0.5448, p = 0.5951; tZ/A = 2.09, p = 0.0585; tZ/K = 0.5695, p = 0.5795).
Ultimate failure load
Optimized continuous loop grafts of Double Z –tech group (982 ± 121 N) had significantly higher ultimate failure load than conventional hamstring graft (777 ± 175 N) and continues loop graft using Arthrex-tech (762.2 ± 103 N) and Kessler-tech (712 ± 187 N) (tZ/M = 2.542, p = 0.0259; tZ/A = 3.794, p = 0.0022; tZ/K = 3.196, p = 0.0077) (Fig. 5B). Failure strength of the other two continuous loop grafts was comparable to the conventional hamstring graft. No differences were found between Arthrex-tech group, the Kessler-tech group and the M -tech group (tA/M = 0.2023, p = 0.8428; tK/M = 0.6674, p = 0.5171; tK/A = 0.6519, p = 0.5259).
Cyclic elongation
The continuous loop grafts showed outstanding performance than hamstring graft in terms of cyclic elongation (Fig. 5C). The Double Z-tech group (3.72 ± 0.97 mm) displayed the smallest cyclic elongation compared to the M-tech group (8.88 ± 2.25 mm), Arthrex-tech group (4.94 ± 0.69 mm) and Kessler-tech group (7.23 ± 2.17 mm) (tZ/M = 5.561, p = 0.0001; tZ/A = 2.838, p = 0.014; tZ/K = 3.899, p = 0.0021). The Arthrex-tech group also showed significantly better performance than M-tech group and Kessler-tech group (tA/M = 4.711, p = 0.0004; tA/K = 2.824, p = 0.0143). While cyclic elongation of the Kessler-tech group was the largest among the continuous loop groups, it was not statistically different to the conventional hamstring graft. No difference was observed between Kessler-tech group and M-tech group (tK/M = 1.397, p = 0.1878).
Ultimate elongation
The ultimate elongation of continuous loop grafts was not inferior to conventional hamstring graft preparation (Fig. 5D). The Arthrex-tech group (2.04 ± 0.68 mm) showed the smallest ultimate elongation compared to the M-tech group (5.23 ± 3.94 mm), the Kessler-tech group (5.32 ± 2.40 mm) and the Double Z-tech group (4.48 ± 1.78 mm) (tA/M = 2.265, p = 0.0413; tA/K = 3.721, p = 0.0026; tA/Z = 3.612, p = 0.0032). No difference was found between Kessler-tech group, Double Z-tech group and M-tech group (tK/M = 0.0503, p = 0.9607; tZ/M = 0.4612, p = 0.6529; tK/Z = 0.746, p = 0.47).
Cyclic stiffness
Cyclic stiffness was highly variable in all groups, and no intergroup differences were observed (F = 1.416, p = 0.2428). The measured cyclic stiffness of Arthrex-tech group was 2219 ± 862 N/mm, the Kessler-tech group was 1758 ± 1051 N/mm, the Double Z-tech group was 1927 ± 1436 N/mm, and the M-tech group was 1405 ± 433 N/mm (Fig. 5E).
Failure mode
Of the eight grafts that were tested in M-tech group, one was excluded due to data output failure, five failed in the form of baseball stitch failure and two of loading rope cut through. In the Arthrex-tech group, all of 8 samples failed by whipstitch rupture. Interestingly, 3 of 8 samples non-critically damaged with a visible gap forming at the 115th, 214th and 2750th cycle of cyclic testing. However, the graft structures were stable and survived to the end of the testing protocol. The Kessler-tech group showed the weaker anastomosis strength compared to the Arthrex group. Here, 4 of 8 samples similarly showed a visible gap within the free ends at early stages of cyclic loading (20th, 150th, 272th and 963th cycles), with one of the four grafts failing completely at the 2200th cycle due to continuous loop anastomosis loose and structural failure. The remaining 7 samples completed the cyclic testing protocol failed as Kessler anastomosis rupture during ramp loading. Of the eight tested grafts in the Double Z-tech group, one was excluded due to data output failure, five failed in the form of Double Z anastomosis rupture and two from loading rope cut through. No visible gap nor anastomosis loose was observed during cyclic testing in Double Z-tech group.
Graft preparation time
The Arthrex-tech group (5.98 ± 0.38 min) displayed the shortest graft preparation time compared to the M-tech group (7.94 ± 0.58 min), Kessler-tech group (9.03 ± 0.39 min) and Double Z-tech group (13.29 ± 1.14 min) (tA/M = 7.952, p < 0.001; tA/K = 15.52, p < 0.001; tA/Z = 17.19, P < 0.001). While graft preparation time of the Double Z-tech group was the longest among the continuous loop groups, the mean graft preparation time of continuous loop groups (9.32 ± 0.68 min) was not statistically different to the M-tech graft (t = 1.153, p = 0.2581).
Discussion
This study compared a frequently applied hamstring graft preparation technique used for tibial adjustable length suspensory graft fixation (Arthrex-tech) to a conventional M-configuration technique (M-tech) used in tibial interference screw fixation. Arthrex-tech grafts outperformed the more traditional M-tech in terms of cyclic- and ultimate elongation. In this work, we also explored modified graft suturing techniques designed to improve the construct with regard to mechanical performance in isolated biomechanical testing. Using a newly introduced Double-Z graft preparation technique (Double Z-tech) we could significantly increase ultimate failure load (by 28.9% compared to the Arthrex-tech) and reduce cyclic elongation of the continuous-loop construct (by 24.7% compared to the Arthrex-tech). To the authors’ knowledge, the current investigation is the first to functionally assess and compare these graft configuration techniques.
It was originally hypothesized that the diameters of continuous loop grafts and hamstring grafts would show no significant difference. However, the diameters of all three grafts were larger than hamstring graft, although there was no statistical difference between Arthrex-tech group and hamstring graft group. The reasons may be due to the continuous loop technology itself. As is known, the suture configuration ultimately determines the diameter of the entire graft. In fact, Arthrex-tech’s whip stitch suture is a side-by-side suture between the free ends of the tendon, which potentially increases the diameter of the suture configuration. However, due to the applied compression of the graft, the diameter of the tendon could be reduced by a circular contraction suture, leading to a lack of any significant differences in diameter between the Arthrex-tech group and the M-tech group. Both the Kessler-tech and the Double-tech Z group differ from the Arthrex-tech in that the suture passes through the interior of the tendon entirely, restricting compression of the tendon and consequently increasing the end diameter. Considerations for decreasing the diameter of the graft will be pursued in follow-up studies.
Elongation of the ACL graft should be kept to a minimum in order to avoid postoperative laxity. Our data indicated that continuous loop grafts showed better performance in terms of cyclic elongation and ultimate elongation compared to other grafts. In this respect, the Double Z-tech group and Arthrex-tech group outperformed the Kessler-tech group and M-tech groups. Notably, cyclic elongation in Double Z-tech group was superior to the Arthrex-tech group. This aspect may be useful in minimizing the widely-discussed windshield-wiper/bungee cord phenomenon and could help to potentially avoid postoperative laxity. Petre et al.15 reported a biomechanical result of cyclic elongation using the Arthrex technique with a tendon–button suspensory device named TightRope RT (Arthrex, Naples, USA). They reported a total elongation of 4.5 ± 0.7 mm after 1000 cycles (50 and 250 N). Our results using the Double Z-tech group and Arthrex-tech group are comparable - indicating that our implementation of the Arthrex technique can be considered as standard, and demonstrates that the Double Z-tech can provide superior technical performance in terms of ultimate failure load and cyclic elongation.
For the purpose of making the tendon graft more compact and increasing its stability, the doubled continuous tendon loop was secured using a three-suture technique. The actual location of three stitches were realized to be extremely important to the ultimate stabilization of the graft structure. We first performed a baseline biomechanical study analyzing three different locations for the stitch end. One stitch location was near the loading suture at the outside the circular contraction sutures; one stitch location was under that of the circular contraction suture and in the middle of the two circular sutures. In these biomechanical assessments we found most of the stiches loosened and failed due to the pulling force of the loading loop when the graft securing sutures were located near the loading suture and outside the circular contraction sutures. It should be noted that locating stitches beneath the circular contraction, while stabilizing the construct, can substantially enlarge the graft diameter. Locating the stitches in the middle of the two circular sutures yielded both good biomechanical performance and smaller diameters than other configurations, and was thus chosen for the experiments described in this study.
In this study, the Arthrex-tech group displayed the shortest graft preparation time while graft preparation time of the Double Z-tech group was the longest among the continuous loop groups. The reasons were due to the continuous loop technology and Double Z-tech themselves. As we introduced above, Arthrex-tech, M-tech and Kessler-tech were more mature and easy suture surgical techniques. In the sample preparation process, they were easier to implement. Because of the innovation and complexity of Double Z technology, it will take longer preparation time. However, the preparation time of these four surgical techniques were completed within 15 minutes, which was sufficient and reasonable for the preparation of the tendon for ACLR surgery.
During this time, the surgeon can complete surgical procedures such as joint cleaning, ligament exploration and bone bed preparation, without affecting the implementation of the operation. Therefore, although Double Z technology has the longest preparation time, it can still meet clinical needs. Considerations for decreasing the preparation time of the grafts will be pursued in follow-up studies.
Among the limitations to the present study, is the fact that bovine tendons rather than human hamstring tendons were used as a testing model. Bovine tendons yield a homogeneous graft size that allows the suturing method effects on ultimate implant size to be more accurately compared. Further, since no tendon rupture and no differences in stiffness were observed in the present study, it can be assumed that the differences between the groups were not affected by differences in tendon properties. Of course, such uniformly high-quality tendon does not reflect actual clinical practice. Third, as we discussed above, the diameter in continuous groups were unsatisfyingly large, and additional work will be required to refine the methods in a manner that minimizes increases in graft diameter. Finally, it is important to note that this in vitro study focuses only on mechanics of graft and neglects biological factors such as healing and incorporation of the graft into the surrounding bone tunnel. Additional intra-articular biomechanical experiments are planned to clarify some of these aspects.
Altogether, we hypothesized that continuous loop tendon grafts were superior to the M-Tech graft in mechanical properties and the Double Z-Tech can improve the construct of continuous loop tendon graft with regard to mechanical performance. Our data indicated that continuous loop grafts showed better performance in terms of cyclic elongation and ultimate elongation compared to other grafts. In this respect, the Double Z-tech group outperformed the Kessler-tech group and M-tech groups. Our findings confirm our hypothesis and oriented a new approach for continuous loop tendon-graft preparation. Future work is warranted to exploit potential performance increases that can be achieved by improved graft design.
Conclusions
This study shows that continuous loop tendon grafts perform equivalently to the standard M-folding technique in cyclic and ramp-to-failure testing. We introduce a novel “Double Z” technique that biomechanically outperforms other preparation techniques, albeit with the drawbacks of being more time-consuming and yielding slightly larger overall graft diameters.
Data availability
The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.
References
Dargel, J. et al. Biomechanics of the anterior cruciate ligament and implications for surgical reconstruction. Strategies in Trauma and Limb Reconstruction 2, 1–12 (2007).
Smith, H. C. et al. Risk factors for anterior cruciate ligament injury: a review of the literature-part 2: hormonal, genetic, cognitive function, previous injury, and extrinsic risk factors. Sports Health 4, 155–161 (2012).
Fayard, J. M. et al. Factors affecting outcome of ACL reconstruction in over-50-year-olds. Orthopaedics & Traumatology: Surgery & Research 105, S247–S251 (2019).
Persson, A. et al. Increased risk of revision with hamstring tendon grafts compared with patellar tendon grafts after anterior cruciate ligament reconstruction: a study of 12,643 patients from the Norwegian Cruciate Ligament Registry, 2004–2012. The American Journal of Sports Medicine 42, 285–291 (2014).
Pinczewski, L. A. et al. A 10-year comparison of anterior cruciate ligament reconstructions with hamstring tendon and patellar tendon autograft: a controlled, prospective trial. The American Journal of Sports Medicine 35, 564–574 (2007).
Chechik, O. et al. An international survey on anterior cruciate ligament reconstruction practices. International Orthopaedics 37, 201–206 (2013).
Lubowitz, J. H., Ahmad, C. S. & Anderson, K. All-inside anterior cruciate ligament graft-link technique: second-generation, no-incision anterior cruciate ligament reconstruction. Arthroscopy 27, 717–727 (2011).
Ahmad, S. S. et al. Adjustable loop ACL suspension devices demonstrate less reliability in terms of reproducibility and irreversible displacement. Knee Surgery Sports Traumatology Arthroscopy 26, 1392–1398 (2018).
Monaco, E. et al. Biomechanical evaluation of different anterior cruciate ligament fixation techniques for hamstring graft. Journal of Orthopaedic Science 15, 125–131 (2010).
Lubowitz, J. H. All-inside anterior cruciate ligament graft link: graft preparation technique. Arthroscopy Techniques 1, e165–168 (2012).
Zobitz, M. E., Zhao, C., Amadio, P. C. & An, K. N. Comparison of mechanical properties of various suture repair techniques in a partially lacerated tendon. Journal of Biomechanical Engineering 122, 604–607 (2000).
Walsh, M. P. et al. The 1:1 versus the 2:2 tunnel-drilling technique: optimization of fixation strength and stiffness in an all-inside double-bundle anterior cruciate ligament reconstruction–a biomechanical study. The American Journal of Sports Medicine 37, 1539–1547 (2009).
Walsh, M. P., Wijdicks, C. A., Parker, J. B., Hapa, O. & LaPrade, R. F. A comparison between a retrograde interference screw, suture button, and combined fixation on the tibial side in an all-inside anterior cruciate ligament reconstruction: a biomechanical study in a porcine model. The American Journal of Sports Medicine 37, 160–167 (2009).
Smith, P. A. & DeBerardino, T. M. Tibial Fixation Properties of a Continuous-Loop ACL Hamstring Graft Construct with Suspensory Fixation in Porcine Bone. The Journal of Knee Surgery 28, 506–512 (2015).
Petre, B. M. et al. Femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction: a comparative biomechanical study. The American Journal of Sports Medicine 41, 416–422 (2013).
Acknowledgements
We are grateful to China Scholarship Council and the Second Affiliated Hospital of Xi’an Jiaotong University for the supporting in Dr. Zhang’s academic visiting in Swiss Federal Institute of Technology (ETH). This work was supported by the National Natural Science Foundation of China (No. 81772346), the China Postdoctoral Science Foundation (No. 2017M613151) and the Postdoctoral Science Foundation of Shaanxi province (No. 2017BSHYDZZ41).
Author information
Authors and Affiliations
Contributions
All of the authors contributed to the conception and design of this study. C.Z. and T.G. were responsible for experiment performing, data analysis and the drafting of the manuscript. J.S. aided in design of the study and participated in its coordination, critical revision of the manuscript, and final approval of the version to be published. All authors have read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zhang, C., Götschi, T., Li, X. et al. Biomechanical comparison of the use of different surgical suture techniques for continuous loop tendon grafts preparation. Sci Rep 10, 538 (2020). https://doi.org/10.1038/s41598-019-57332-8
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-019-57332-8
This article is cited by
-
Transversal calcaneal anchored Achilles tendon reconstruction with free semitendinosus tendon autograft for acute rupture of Achilles tendon: clinical evaluation
Scientific Reports (2024)
-
Study on Preparation of Core-Spun Yarn Surgical Sutures by Compositing Drug-Loaded Nanofiber Membrane with PLA and Its Controllable Drug Release Performance
Fibers and Polymers (2023)