Abstract
Light’s polarisation contains information about its source and interactions, from distant stars to biological samples. Polarimeters can recover this information, but reliance on birefringent or rotating optical elements limits their wavelength range and stability. Here we present a static, single-shot polarimeter based on a Fresnel cone - the direct spatial analogue to the popular rotating quarter-wave plate approach. We measure the average angular accuracy to be 2.9° (3.6°) for elliptical (linear) polarization states across the visible spectrum, with the degree of polarisation determined to within 0.12 (0.08). Our broadband full Stokes polarimeter is robust, cost-effective, and could find applications in hyper-spectral polarimetry and scanning microscopy.
Similar content being viewed by others
Introduction
Polarimeters identify the polarisation state of light and are an important tool in astronomy1,2,3, material characterisation4,5,6, remote sensing7,8,9 and medicine10,11,12. Polarisation states describe two orthogonal complex field components and are commonly expressed as four real numbers in a Stokes vector13. Characterisation therefore requires at least four measurements, typically performed by either spatial splitting or temporal modulation.
Spatial splitting techniques divide incident light into several sub-beams, each requiring its own polarisation analysing optics. More recently this has been implemented using “division of focal plane” spatial modulation, where the detector is divided with components made using advanced fabrication techniques14,15,16. Spatial splitting allows measurements to be obtained simultaneously, however often at the cost of complex or expensive setups. Recent devices address some of these issues17,18,19.
Temporal modulation techniques require taking sequential measurements, such as the rotating wave plate technique20 typically found in commercial polarimeters, liquid crystal variable retarders21 or photo-elastic modulators22,23. These techniques become ineffective if the initial polarisation is varying on time-scales comparable to the modulation time. Moreover, devices with moving parts are prone to instability and broadband operation remains a challenge.
In this paper, we demonstrate a polarimeter based on the back-reflection from a Fresnel cone which is a direct spatial analogue to the popular rotating quarter-wave plate device20,24,25. We show that our polarimeter is intrinsically broadband and stable due to its lack of both birefringent elements and moving parts.
Fresnel cones - solid glass cones with a 90° apex angle - act like azimuthally varying wave plates26. Fresnel’s equations predict that total internal reflection (TIR) produces phase shifts between the s and p polarisation components which are dependent on the angle of incidence and the refractive index of the cone material. The conical surface leads to an azimuthally varying decomposition of the input polarisation state into s and p components, and the resulting azimuthally varying phase shifts produce polarisation structures. Fresnel cones are simple to implement, intrinsically broadband and cost-effective.
Every spatially uniform initial polarisation state is mapped onto a unique polarisation structure by the Fresnel cone. Detection of this polarisation structure therefore allows us to identify the initial polarisation state. In the following we present the theoretical treatment of the Fresnel cone polarimeter as the spatial analogue of the rotating quarter-wave plate polarimeter and report on the performance and calibration of the device, before discussing further optimisation.
Polarimeter Theory
One of the most wide-spread and simple polarimeters uses a rotating quarter-wave plate and a linear polariser followed by an intensity measurement on a photodiode, which we summarise in the following. Written in the familiar Stokes formalism this is
where S = [s0, s1, s2, s3]T and S′ = [s′0, s′1, s′2, s′3]T are the initial and final Stokes vectors respectively, the M matrices are Mueller matrices corresponding to a horizontally aligned linear polariser (Mpol), a quarter-wave plate with horizontal fast-axis (Mqwp), and R(θ) is the Mueller rotation matrix by an angle θ. The terms R(−θ)MqwpR(θ) represent a quarter-wave plate rotated by an angle θ from horizontal (see Supplement 1). Here we are using a right-handed Cartesian coordinate system, where the light propagates in the +z direction and the observer is looking towards the light source. The intensity I(θ) measured on the photodiode is then determined by the first element of the vector S′ in Eq. 1 as
Measurements for different θ values are taken by spinning the quarter-wave plate and recording the intensity as a function of time I(t), which is mapped to I(θ) through knowledge of the rotational frequency. Eq. 2 can be converted to a truncated Fourier series in θ and allows us to express the Fourier coefficients in terms of the components of the initial Stokes vector S. Measurement of these coefficients then allows recovery of S through matrix inversion27,28,29.
Our polarimeter approach is analogous to the above rotating quarter-wave plate technique. Our system consists of a beam-splitter, Fresnel cone, linear polariser and camera (shown in Fig. 1), the equation for this system is
where R(−θ)MwedgeR(θ) are the Mueller matrices representing the Fresnel cone, comprised of double total internal reflection from a rotated (R(θ)) glass wedge (Mwedge). The Jones matrix for a glass wedge is found in26 and given as
where
rs and rp are the Fresnel reflection coefficients for s and p polarisation components and n = nair/ncone. We note that in our case we are operating above the critical angle for total internal reflection, and hence only the phase is affected. The Mueller matrix is determined by following the method described in30. For an appropriately chosen refractive index of the cone, formally Mwedge = Mqwp and Eq. 3 becomes Eq. 1 where θ is now the azimuthal spatial angle with the origin of the polar coordinate system at the cone apex. The cone system is therefore revealed to be a direct spatial analogue to the rotating quarter-wave plate system.
Polarisation State Recovery
For a polarimeter with optimally designed, ideal optical components, we can recover the initial Stokes vector S by solving Eq. 2 evaluated for at least four θi. For this we re-express Eq. 2 as a truncated Fourier series in θi:
where the discrete Fourier coefficients for N discrete angles are
The values on the right hand side can be experimentally determined from Fourier analysis of the signal I(θ), easily achieved by fast Fourier transform (FFT). The matrix form of Eqs 7–10 is
and the polarisation state S is then found by matrix inversion via
This analysis holds equally for an ideal quarter-wave plate polarimeter and our Fresnel cone polarimeter. In both cases optical elements may introduce unwanted polarisation shifts, however these can be accounted for with suitable analysis, producing optimal polarimeter results even for imperfect devices.
We find that the main source of unwanted polarisation shifts in the Fresnel cone polarimeter arises from the beam-splitter. To counteract these we include the Mueller matrices of the beam-splitter transmission (Btrans) and reflection (Brefl):
To recover S from an experimental system, the procedure outlined in equations 6–12 is followed, replacing Eq. 3 with Eq. 13, where the fundamental difference to the ideal case is the appearance of a cos2θ term in the truncated Fourier series (see Supplement 1). The linear system of equations then becomes over-determined as we have five equations in only four unknowns. We therefore cannot use the direct matrix inverse to solve for S but instead use the matrix pseudo-inverse. The quality of the pseudo-inverse matrix can be evaluated using the condition number31, which is a measure of noise amplification to the final result. For the ideal system represented by Eq. 12 the condition number is 2.618 (to 3 decimal places), while the condition number for the experimental system represented by Eq. 13 is 2.619 (to 3 decimal places) for each the red, green and blue channels.
Experimental Method
In a proof of principle experiment shown in Fig. 1, we demonstrate the operation of our Fresnel cone polarimeter for broadband light. The system consists of a polarisation state generator (PSG), producing the initial polarisation states, and the actual polarimeter. The light source is a white LED, collimated by a lens (f = 25.4 mm) and apertured to control the spatial coherence. The PSG consists of a linear polariser (Thorlabs, LPVISE100-A) and an optional λ/4 retardance Fresnel rhomb (Thorlabs, FR600QM) to generate broadband initial polarisation states. Half of the light then transmits through a non-polarising beam-splitter (Thorlabs, BS013) and reflects from a Fresnel cone (Edmund Optics 45–939, with aluminium coating removed). After back-reflection from the cone, half of the remaining light reflects from the non-polarising beam-splitter into the measurement arm, which consists of a horizontally aligned linear polariser (Thorlabs, LPVISE100-A) and lenses which re-image the plane of the cone tip onto a camera (Thorlabs, DCC1645C). As the light travels twice through a non-polarising beam-splitter the nominal efficiency of this configuration is limited to 25%, however, optimised designs are possible with up to 100% nominal efficiency (see26 supplementary material). The camera records the full profile of the reflected light from which we can extract I(θ). We note, however, that carefully placed individual detectors could be used instead. Resolving the zeroth, second and fourth azimuthal frequency components requires at least 5 detectors equally spaced in θ, and distributed around half of the output beam. Using a camera allows us to obtain the intensity for quasi-continuous θ values (we use 499), and reduce the effect of noise in the system by averaging over the radial parameter. In addition, the camera provides 24-bit colour images from which we can extract three 8-bit colour channels (RGB) thus monitoring the operation across the visible spectrum.
Light generated in different polarisation states produces different intensity patterns measured on the camera (an example for horizontal initial polarisation is shown in Fig. 2a). We calibrate the intensity of our images by recording a background image with the LED off, and two normalisation images with the PSG set to produce horizontal and vertical polarisation respectively. The intensity patterns used for our Fourier analysis are background corrected and normalised to the sum of the two normalisation images. We unwrap the calibrated images from a Cartesian (x-y) to a polar (θ-r) coordinate frame, relative to the cone centre. The noisy data at the centre and edges of the intensity pattern is removed by selecting a region of interest, delimited by red lines in Fig. 2a,b, leaving the cleaned data shown in Fig. 2c.
The resulting averaged 1D intensity profile (see Fig. 2d) is I(θ), first element of the vector S′ in Eq. 13, and from this we obtain the Fourier coefficients C using a fast Fourier transform (FFT) as discussed in the section above. Calculation of M requires determination of Brefl and Btrans, which we achieve following the method outlined in30,32. Systematic errors in these measurements can lead to non-physical Mueller matrices, which can be avoided by converting Brefl and Btrans into the form shown in Equation (5) of33 (see Supplement 1). This conversion assumes that there is no loss of polarisation in the beam-splitter and that the Mueller matrix can be parameterised by 3 angles, namely a phase-shift induced between s and p polarisation components, a rotation of the linear polarisation components, and an orientation of the beam-splitter from horizontal. We achieve this normalisation by numerical minimisation of the difference between our measured and normalised matrix while varying all 3 angles. The pseudo-inverse of M is calculated using the singular value decomposition (SVD) method and we finally recover S using Eq. 12.
The imperfections of the optical elements, described by the Mueller matrices Brefl and Btrans, are in general frequency dependent. We identify these matrices for the red, green and blue colour channels of our camera independently. These are applied to the individual intensity patterns for the three colour channels, obtaining S for red, green and blue frequency ranges. Our final white light polarisation state is taken to be their average.
Results
To assess our polarimeter quantitatively, we consider its performance for measurement of the polarised and unpolarised components of the light separately. The polarised component is assessed by measuring the so-called angular accuracy, which can be thought of as the angle between initial and measured polarisation vectors on the Poincaré sphere as demonstrated in Fig. 3a. Angular accuracy is defined as
where S123 = [s1, s2, s3]T and S′123 = [s′1, s′2, s′3]T. The unpolarised component of the state can be quantified using the degree of polarisation (DOP) accuracy, where the DOP is defined as
and its accuracy is given by the magnitude of the difference between the initial polarisation state DOP and measured DOP. We assess the angular accuracy and DOP accuracy for a range of elliptical and linear initial states. The linear polarisation states are generated by rotating a linear polariser in 5° increments over a range of 180°. Elliptical polarisation states are generated in the same way, with an additional quarter-wave Fresnel rhomb (horizontal fast-axis) after the linear polariser (see Fig. 3b) for a Poincaré representation of these polarisation states).
To visualise the angular accuracy only, we set DOP = 1. The remaining Stokes parameters for a range of elliptic input states are shown in Fig. 4a, finding excellent agreement between the initial (solid lines) and measured (data points) Stokes parameters. Similar results are obtained for the linear polarisation initial states (not shown). The overall performance can be gauged by taking the average angular accuracy for all elliptical (linear) initial states, which we find to be 2.9° (3.6°).
In addition to the overall angular accuracy of the polarimeter we also investigate its frequency dependence, which is obtained by performing the data analysis on each colour channel of the camera individually. Figure 4b shows the angular accuracy in the red, green and blue colour bands. We see that the polarimeter has similar performance across the three frequency bands for elliptical (linear) initial states, with average angular accuracies of 4.1° (2.1°), 3.0° (2.7°) and 5.2° (6.4°) in the red, green and blue colour bands respectively. These results confirm that our polarimeter performs well across the full visible spectrum.
In addition to fully polarised states, our polarimeter also allows measurement of the full Stokes vector. We measure the DOP accuracy for the same elliptical (linear) initial polarisation states as before and calculate the average error to be 0.12 (0.08) across all initial polarisation states and colour channels. Individual colour channel performance is 0.11 (0.10), 0.13 (0.07) and 0.11 (0.08) for the red, green and blue colour channels respectively, showing good performance across the visible spectrum.
Discussion
The formal equivalence between our polarimeter and the canonical rotating quarter-wave plate polarimeter, enables us to apply the findings of previous optimisation studies. The number of measurements used for Stokes vector determination has been discussed in31, where it was shown that performance is increased when using more than the minimum required number of θ measurements. For temporally modulated polarimeters this requires a longer acquisition time, however here we record measurements for many θ angles simultaneously. In agreement with these studies we find best performance when we use as many θ values as possible (499). The optimum number of r values however changes between angular accuracy and DOP accuracy measurements. The best angular accuracy results are obtained by averaging all r values from 25 pixels to 350 pixels, avoiding the noisy central region. Optimal DOP accuracy results are instead found when averaging over only 10 r values at the edge of the image. We attribute this difference in optimal pixel range to the background noise level. At small r there are fewer unique θ values due to pixellation, reducing contrast and increasing background noise. This effects the DOP measurements through its reliance on a0, while leaving angular accuracy measurements relatively unaffected.
Performance can also be improved by adjusting the retardance, as discussed in34, with an optimal retardance of 132°. The glass used in our Fresnel cone (BK7) has an average refractive index of ~1.52 for white light, resulting in a retardance of approximately 79°. By engineering the Fresnel cone with refractive index of 1.86 (LaSF9 glass), a retardance of 132° can be achieved, which is predicted to reduce the effect of measurement noise on the polarisation results. Compensation for errors in retardance value and azimuthal misalignment error of the linear polariser have also been discussed in28 and could be applied to the Fresnel cone polarimeter.
A potential advantage of our Fresnel cone polarimeter compared to conventional polarimeters is the increased acquisition rate. In our camera-based system, the data acquisition rate is set by the frame rate, allowing Stokes measurements in the kHz range using a suitable camera. Using individual detectors, as outlined in the method section, this could be increased into the MHz or GHz range. We can make an estimate of the performance of a setup with individual detectors by taking a subset of our camera data. Initial tests taking only 9 equally azimuthally spaced camera pixels (at r = 300) around the cone shows an average angular accuracy of 5.4° (5.0°) and average DOP error of 0.12 (0.09) for elliptical (linear) input polarisation states, confirming that the system still performs well even with few measurements. Although alignment of the cone centre is important, this can be resolved using fibre-coupling in the final device.
Conclusions
We have demonstrated a full Stokes polarimeter based on the back-reflection from a Fresnel cone, and show that it is the spatial analogue to the ubiquitous rotating quarter-wave plate technique. We characterise the performance of our polarimeter by measuring the angular accuracy, experimentally demonstrating this to have an average of 2.9° and 3.6°, with average DOP errors of 0.12 and 0.08 for elliptical and linear polarisation states respectively. We have also shown that the polarimeter performs well across the visible spectrum, by measuring the angular accuracy and DOP error in the red, green and blue colour bands respectively. This proof of principle experiment uses off-the-shelf polarisation components, and specialised optics could improve performance while reducing size. Previous research on polarimeter optimisation can also be applied to our polarimeter, improving performance through optimisation of refractive index and measurement strategy.
Not only does our device record a full Stokes vector in a single shot for broadband light, but Fresnel cone polarimeters are robust, stable and low-cost (~£500). Our polarimeter could increase reliability and lower costs in a wide range of applications, for example, polychromatic spectro-polarimeters for astronomy and remote sensing which provide broadband Stokes measurements3,35,36,37. Scanning microscopy techniques such as confocal or 2-photon microscopy could also be equipped with our Fresnel cone polarimeter, which would remove the need for rotating wave plates in applications such as orientational order measurements38,39, or to provide Mueller analysis of samples when coupled with a broadband polarisation state generator.
References
Costa, E. et al. An efficient photoelectric X-ray polarimeter for the study of black holes and neutron stars. Nat. 411, 662–665, https://doi.org/10.1038/35079508 (2001).
Lites, B. W. et al. The Horizontal Magnetic Flux of the Quiet-Sun Internetwork as Observed with the Hinode Spectro-Polarimeter. The Astrophys. J. 672, 1237–1253, https://doi.org/10.1086/522922 (2008).
Sterzik, M. F., Bagnulo, S. & Palle, E. Biosignatures as revealed by spectropolarimetry of Earthshine. Nat. 483, 64–66, https://doi.org/10.1038/nature10778 (2012).
Johnson, P. M. et al. Structure of the liquid-crystal ferrielectric phases as determined by ellipsometry. Phys. Rev. Lett. 84, 4870–4873, https://doi.org/10.1103/PhysRevLett.84.4870 (2000).
Kovaleva, N. N. et al. Spin-Controlled Mott-Hubbard Bands in LaMnO3 Probed by Optical Ellipsometry. Phys. Rev. Lett. 93, 147204, https://doi.org/10.1103/PhysRevLett.93.147204 (2004).
Losurdo, M. et al. Spectroscopic ellipsometry and polarimetry for materials and systems analysis at the nanometer scale: State-of-the-art, potential, and perspectives. J. Nanoparticle Res. 11, 1521–1554, https://doi.org/10.1007/s11051-009-9662-6 arXiv:1011.1669v3 (2009).
Cloude, S. R. & Pettier, E. A review of target decomposition theorems in radar polarimetry. IEEE Transactions on Geosci. Remote. Sens. 34, 498–518, https://doi.org/10.1109/36.485127 (1996).
Diner, D. J. et al. Dual-photoelastic-modulator-based polarimetric imaging concept for aerosol remote sensing. Appl. optics 46, 8428–8445, https://doi.org/10.1364/AO.46.008428 (2007).
Imazawa, R., Kawano, Y., Ono, T. & Itami, K. Development of real-time rotating waveplate Stokes polarimeter using multi-order retardation for ITER poloidal polarimeter. Rev. Sci. Instruments 87, https://doi.org/10.1063/1.4939444 (2016).
Greenfield, D. S., Knighton, R. W. & Huang, X. R. Effect of corneal polarization axis on assessment of retinal nerve fiber layer thickness by scanning laser polarimetry. Am. J. Ophthalmol. 129, 715–722, https://doi.org/10.1016/S0002-9394 (2000).
Ghosh, N. Tissue polarimetry: concepts, challenges, applications, and outlook. J. Biomed. Opt. 16, 110801, https://doi.org/10.1117/1.3652896 (2011).
Vizet, J. et al. In vivo imaging of uterine cervix with a Mueller polarimetric colposcope. Sci. Reports 7, 1–12, https://doi.org/10.1038/s41598-017-02645-9 (2017).
Bickel, W. S. & Bailey, W. M. Stokes vectors, Mueller matrices, and polarized scattered light. Am. J. Phys. 53, 468–478, https://doi.org/10.1119/1.14202 (1985).
Nordin, G. P., Meier, J. T., Deguzman, P. C. & Jones, M. W. Micropolarizer array for infrared imaging polarimetry. J. Opt. Soc. Am. A 16, 1168, https://doi.org/10.1364/JOSAA.16.001168 (1999).
Myhre, G. et al. Liquid crystal polymer full-stokes division of focal plane polarimeter. Opt. Express 20, 27393, https://doi.org/10.1364/OE.20.027393 (2012).
Deng, P., Jin, R. B., Zhao, X., Cao, Y. & Yu, J. Novel “quasi-Bayer” micro-polarizer patterns for the division-of-focal-plane polarimeters. Optik 132, 216–222, https://doi.org/10.1016/j.ijleo.2016.12.003 (2017).
Chang, J., Zeng, N., He, H., He, Y. & Ma, H. Single-shot spatially modulated Stokes polarimeter based on a GRIN lens. Opt. Lett. 39, 2656, https://doi.org/10.1364/OL.39.002656 (2014).
Balthasar Mueller, J. P., Leosson, K. & Capasso, F. Ultracompact metasurface in-line polarimeter. Opt. 3, 42, https://doi.org/10.1364/OPTICA.3.000042 (2016).
Estévez, I., Sopo, V., Lizana, A., Turpin, A. & Campos, J. Complete snapshot Stokes polarimeter based on a single biaxial crystal. Opt. Lett. 41, 4566–4569, https://doi.org/10.1364/OL.41.004566 (2016).
Berry, H. G., Gabrielse, G. & Livingston, A. E. Measurement of the Stokes parameters of light. Appl. Opt. 16, 3200, https://doi.org/10.1364/AO.16.003200 (1977).
Bueno, J. M. Polarimetry using liquid-crystal variable retarders: Theory and calibration. J. Opt. A: Pure Appl. Opt. 2, 216–222, https://doi.org/10.1088/1464-4258/2/3/308 (2000).
Liu, Y., Jones, G. A., Peng, Y. & Shen, T. H. Generalized theory and application of Stokes parameter measurements made with a single photoelastic modulator. J. Appl. Phys. 100, https://doi.org/10.1063/1.2353894 (2006).
Arteaga, O., Freudenthal, J., Wang, B. & Kahr, B. Mueller matrix polarimetry with four photoelastic modulators: theory and calibration. Appl. Opt. 51, 6805, https://doi.org/10.1364/AO.51.006805 (2012).
Pelizzari, S., Rovati, L. & Angelis, C. D. Rotating polarizer and rotating retarder plate polarimeters: comparison of performances. Proc. SPIE 4285, 235, https://doi.org/10.1117/12.426894 (2001).
Lin, J. F. & Lee, M. Z. Concurrent measurement of linear birefringence and dichroism in ferrofluids using rotating-waveplate Stokes polarimeter. Opt. Commun. 285, 1669–1674, https://doi.org/10.1016/j.optcom.2011.12.032 (2012).
Radwell, N., Hawley, R. D., Götte, J. B. & Franke-Arnold, S. Achromatic vector vortex beams from a glass cone. Nat. Commun. 7, 10564, https://doi.org/10.1038/ncomms10564 (2016).
Dlugunovich, V. A., Snopko, V. N. & Tsaryuk, O. V. Analysis of a method for measuring polarization characteristics with a Stokes polarimeter having a rotating phase plate. J. Opt. Technol. 68, 269–273 (2001).
Flueraru, C., Latoui, S., Besse, J. & Legendre, P. Error analysis of a rotating quarter-wave plate stokes’ polarimeter. IEEE Transactions on Instrumentation Meas. 57, 731–735, https://doi.org/10.1109/TIM.2007.913752 (2008).
Romerein, M. J., Philippson, J. N., Brooks, R. L. & Shiell, R. C. Calibration method using a single retarder to simultaneously measure polarization and fully characterize a polarimeter over a broad range of wavelengths. Appl. Opt. 50, 5382, https://doi.org/10.1364/AO.50.005382 (2011).
Azzam, R. M. A. & Bashara, N. M. Ellipsometry and polarized light. (North-Holland, 1987).
Tyo, J. S. S.-T.-N. Ratio and Minimization of Systematic Error. Appl. Opt. 41, 619–630 (2002).
Fujiwara, H. Spectroscopic ellipsometry: Principles and applications. (Wiley-Blackwell, 2007).
Hauge, P. S. Mueller matrix ellipsometry with imperfect compensators. J. Opt. Soc. Am. 68, 1519, https://doi.org/10.1364/JOSA.68.001519 (1978).
Sabatke, D. S. et al. Optimization of retardance for a complete Stokes polarimeter. Opt. Lett. 25, 802–804, https://doi.org/10.1364/OL.25.000802 (2000).
Snik, F. et al. An overview of polarimetric sensing techniques and technology with applications to different research fields. Proc. SPIE 9099, 90990B, https://doi.org/10.1117/12.2053245 (2014).
Snik, F., van Harten, G., Alenin, A. S., Vaughn, I. J. & Tyo, J. S. A multi-domain full-Stokes polarization modulator that is efficient for 300–2500 nm spectropolarimetry. Proc. SPIE 9613, 96130G, https://doi.org/10.1117/12.2187459 (2015).
Avilés, A., García, P., Chávez, C. & Pérez-Tijerina, E. On the orbital period of the cataclysmic variable V767 CYG. Revista Mexicana de Astron. y Astrofisica 54, 389–396, https://doi.org/10.1051/0004-6361 arXiv:0803.1210v1 (2018).
Mojzisova, H. et al. Polarization-sensitive two-photon microscopy study of the organization of liquid-crystalline dna. Biophys. journal 97, 2348–2357 (2009).
Smalyukh, I. I., Shiyanovskii, S. & Lavrentovich, O. Three-dimensional imaging of orientational order by fluorescence confocal polarizing microscopy. Chem. Phys. Lett. 336, 88–96 (2001).
Acknowledgements
We would like to thank Gergely Ferenczi and Jonathan Taylor for useful and interesting discussions. This work has been supported by EPSRC Quantum Technology Program grant number EP/M01326X/1. R. D. Hawley’s work was supported by the EPSRC CDT in Intelligent Sensing and Measurement, Grant Number EP/L016753/1.
Author information
Authors and Affiliations
Contributions
R.D.H. and J.C. performed the experiment, N.R. and R.D.H. analysed the data, N.R. and S.F.-A. devised the experiment and N.R., R.D.H. and S.F.-A wrote the manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Hawley, R.D., Cork, J., Radwell, N. et al. Passive broadband full Stokes polarimeter using a Fresnel cone. Sci Rep 9, 2688 (2019). https://doi.org/10.1038/s41598-019-39118-0
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-019-39118-0
This article is cited by
-
Revealing the invariance of vectorial structured light in complex media
Nature Photonics (2022)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.