Introduction

Understanding the viability of the ischemic myocardium is a critical issue for surgical treatment options for the failing heart after myocardial infarction (MI). Various functional evaluations of the heart are conducted preoperatively by, e.g., computed tomography, magnetic resonance imaging, radioisotope imaging, and echocardiography1,2,3,4, to determine whether the ischemic myocardium has a potential to recover contractile functions after reestablishment of the coronary circulation5,6. However, these diagnostic modalities are inadequate to detect precise, regional myocardial dysfunction in surgical situations, because of their relatively poor regional correlation with the real heart under direct vision. During heart surgery, viability of the myocardium is evaluated merely from the appearance of the heart in situ by reference to the indirect preoperative assessments. Thus, deeply desired is a useful means for intraoperative evaluation of the myocardial viability under direct vision to obtain better outcome of cardiac surgery.

Raman spectroscopy is expected to be a valuable analytical tool in the biomedical research field, allowing label-free, functional imaging of biological samples via molecular vibrations with no need for fixation or staining7,8,9,10,11,12,13. For over two decades, this modality has been applied experimentally to a variety of human tissues, e.g., the brain14, coronary artery15, skin16, breast17,18, and peripheral nerves19,20,21. In heart tissue, however, the great challenge for clinical application of the conventional Raman spectroscopy is because of the weak signals. We previously reported Raman spectroscopic analysis of old myocardial infarct of rats created by coronary artery ligation by resonant scattering of cytochromes22,23,24, which provides higher sensitivity and selectivity for the Raman signals as compared with the conventional Raman spectroscopy. The proper excitation wavelength, which is electronically resonant with the electronic transition of target molecules, allowed us to detect Raman spectra of cytochrome b5 and c in non-infarcted cardiac muscle of rat heart, having signal intensity 103–105 times stronger than that of the non-resonance signal9,25. In contrast to the rat experimental model, where infarcted tissues are distinctly replaced by fibrosis consisting mainly of type I collagen with no other major components26, accurate Raman spectroscopic evaluation is quite difficult in human old MI because of the spectral complexity of the tissue constituents and lack of relevant characteristic molecules exhibiting the resonant Raman scattering in the infarcted tissues. Analogously, the human myocardium may more or less present different Raman spectra between non-infarcted and infarcted myocardium; however, still unclear are the definitive Raman spectral fingerprints and spectroscopic criteria for evaluation of human old MI.

In this study, we sought to identify Raman spectral fingerprints and a prediction model for the evaluation of human old MI. To elucidate definitive Raman spectral fingerprints and a reliable prediction model for the evaluation of myocardial viability under the presence of noise in Raman spectrum, we employed a multivariate spectral analysis method, partial least squares regression-discriminant analysis (PLS-DA). PLS-DA employs a wide region of Raman spectra to construct a prediction model, enabling the reduction of noise effect that appears in the whole Raman spectrum on the prediction model23,27. Furthermore, the PLS-DA calculates latent variables (LVs) and scores, which gives us important spectral information for the prediction of tissue species23,27. We successfully obtained Raman spectra of non-infarcted and infarcted regions of human myocardium resected for volume reduction from five patients with chronic ischemic cardiomyopathy under left ventricular reconstruction (LVR). Our novel application of Raman spectroscopy to the human heart would be a useful means for the detection of myocardial viability during surgery.

Results

Raman spectra of the myocardial tissue samples excised surgically from human heart

The Raman spectra of the non-infarcted and infarcted myocardium showed their own distinct features (Fig. 1). In all the specimens excised from 5 patients, the non-infarcted myocardium, which predominantly consists of cardiomyocytes, exhibited four evident Raman peaks at 755, 1133, 1318, and 1590 cm−1 that were all assigned to heme proteins23 with no obvious differences among the patients (Fig. 1b). In contrast, the Raman spectra of the infarcted myocardium were large in variation among the patients (Fig. 1c). In addition, the infarcted myocardium had a spectral peak at 2942 cm−1, which is distinct from the peak at 2934 cm−1 in the non-infarcted myocardium. The spectral shapes at 1248, 1453, 1661, and 2942 cm−1 were nearly identical to those of collagens (Supplemental Fig. 1), a major constituent of the fibrotic tissue of old myocardial infarct, suggesting the presence of collagens in the infarcted myocardium as identified in our previous study on the rat heart23. Similar to the non-infarcted myocardium, the four evident peaks assigned to heme proteins at 755, 1133, 1318, and 1590 cm−1 were also identified in the infarcted myocardium.

Figure 1
figure 1

Raman spectra of the border zone myocardium obtained from 5 patients. (a) A representative HE-stained image of the boundary between the infarcted region (dashed area A) and non-infarcted region (dashed area B) of the heart tissue excised from patient 3. Scale bar represents 1 mm. Representative Raman spectra of non-infarcted myocardium (b) and infarcted myocardium (c) of each patient. Each spectrum was averaged to obtain sufficient signal-to-noise ratio for the detail identification of Raman bands, and normalized by the integrated intensity of each Raman spectrum.

We compared peak intensities of the four representative Raman bands at 755, 1133, 1318, and 1590 cm−1 of the non-infarcted and infarcted myocardium (Fig. 2). The intensities of non-infarcted myocardium were significantly larger than those of the infarcted myocardium, indicating that the relative contribution of molecular compositions assigned to these bands was much larger in the non-infarcted myocardium than in the infarcted myocardium. Predominant chemical components assigned to these Raman bands might be different in these two regions, i.e., cytochrome c in the non-infarcted myocardium and heme proteins other than cytochrome c in the infarcted myocardium as described in the Discussion section. As a result, the Raman bands of cytochrome c were significantly emphasized in the non-infarcted myocardium, and the intensity variance of non-infarcted myocardium among individual patients might indicate the interpatient variability of reduced cytochrome c in residual cardiomyocytes, which is an important index of viability of the myocardium.

Figure 2
figure 2

Peak intensities of representative Raman bands of the non-infarcted and infarcted myocardium obtained from 5 patients at 755 cm−1 (a), 1133 cm−1 (b), 1318 cm−1 (c), and 1590 cm−1 (d). Asterisks indicate significant differences (p < 0.01) by Student’s t test.

Prediction of the myocardial tissue characteristics from single Raman bands

We sought to predict tissue species by a simple approach using a single representative Raman band of non-infarcted and infarcted myocardium under the reasonable condition of 10-s exposure of 532-nm excitation at about 250 µW/µm2. We evaluated three representative Raman spectral features, i.e., peak intensity at 687 cm−1, peak intensity at 755 cm−1, and peak position ranging from 2930 to 2950 cm−1.

For the Raman band at 755 cm−1, prediction performance evaluated by using ROC analysis indicates that the sensitivities of non-infarcted and infarcted myocardium attained were 83.0% and 87.6%, respectively (Fig. 3a). Although high prediction performance was obtained with the area under the curve (AUC) of 94.0% by using the peak intensity at 755 cm−1, this Raman band reflected not only cytochrome c, but also the other heme proteins such as cytochrome b5, myoglobin, and hemoglobin23.

Figure 3
figure 3

ROC analysis for the prediction of non-infarcted and infarcted myocardium of 5 patients by using representative Raman bands. Peak intensity of 755 cm−1 (a), peak intensity of 687 cm−1 (b), and peak position around 2930–2950 cm−1 (c) were used for the ROC curve calculation.

To confirm the contribution of cytochrome c in the peak intensity at 755 cm−1, we obtained the peak intensity at 687 cm−1, which is one of the signatures of the presence of cytochrome c23. However, prediction performance of the peak intensity at 687 cm−1 was quite low with the AUC of 62.8%, and the sensitivities of non-infarcted and infarcted myocardium were only 65.6% and 55.0%, respectively (Fig. 3b). This indicates difficulty in confirming the presence of cytochrome c under the experimental conditions, i.e., 10-s exposure of 532-nm excitation at 235 to 300 µW/µm2. Typical normalized Raman spectra of infracted and non-infarcted regions obtained under the condition are shown in Supplementary Figure 4. Higher signal-to-noise ratio detection with longer exposure time or higher excitation laser power would improve the prediction performance; however, long exposure time might perturb the procedures of cardiac surgery, and higher excitation laser power might induce photo-induced damage to the specimen.

The peak position ranging from 2930 to 2950 cm−1 is also the representative signature of the difference between non-infarcted and infarcted myocardium, indicating the presence of collagen in the infarcted myocardium. The ROC analysis of the peak position provided the AUC of 88.1%. The sensitivities of non-infarcted and infarcted myocardium were 91.8% and 80.8%, respectively.

These results indicated that the prediction of non-infarcted and infarcted myocardium was realized by using single Raman bands. The Raman bands representing the contribution of heme proteins (755 cm−1) and collagen (peak position around 2930 to 2950 cm−1) would be key signatures for the prediction of tissue species. However, prediction performance using a single band was limited in terms of signal-to-noise ratio of signal intensity.

Reliable prediction for the discrimination of the infarcted myocardium from non-infarcted myocardium by PLS-DA

To construct a more reliable prediction model for the evaluation of myocardial viability under the presence of noise in Raman spectrum, we employed a multivariate spectral analysis, PLS-DA. The PLS-DA provides LVs and scores that reflect spectral features of non-infarcted and infarcted myocardium. The LVs provide common spectral features of a Raman spectral data set employed in PLS-DA in terms of the difference of tissue species. The scores on individual LVs represent the contribution of the LVs on each observed Raman spectrum. The LVs thus have implications for assignment of Raman bands in terms of the statistical differences between non-infarcted and infarcted myocardium when combined with scores of PLS-DA. Furthermore, the distribution of score plot indicates a similarity of the spectral data set of each tissue species and patient in terms of LVs, showing predictability of non-infarcted and infarcted myocardium. LVs and scores of PLS-DA calculated with a total of 10,000 Raman spectra of non-infarcted and infarcted myocardium in five patients are shown in Fig. 4.

Figure 4
figure 4

PLS-DA of non-infarcted and infarcted myocardium. (a) Score plot of LV1, LV2, and LV3. Triangle, non-infarcted cardiac tissue; circle, infarcted cardiac tissue; red, patient 1; green, patient 2; blue, patient 3; yellow, patient 4; and purple, patient 5. Contributions of LV1 (b), LV2 (c), and LV3 (d) to Raman spectra are 78.65%, 2.56%, and 0.41%, respectively.

Both scores of LV1 of the non-infarcted and infarcted myocardium of all patients were distributed in the negative value region (Fig. 4a and Supplemental Movie 1). The scores of LV1 of the non-infarcted myocardium were smaller than those of the infarcted myocardium. The spectral peaks of LV1 included 1160, 1248, and 2942 cm−1, which represented specific Raman bands of infarcted myocardium (Fig. 4b). Although a few specific spectral features of non-infarcted myocardium, such as 687, 1177, and 1366 cm−1, also appeared in LV1, the other specific spectral features of non-infarcted myocardium, such as 648, 2861, and 2934 cm−1, were not included in LV1. In LV2, the Raman bands included 648, 687, 1177, 1366, 2861, and 2934 cm−1, which were all specific in the non-infarcted myocardium, while no spectral feature of the infarcted myocardium appeared in LV2 (Fig. 4c). The scores of LV2 of non-infarcted and infarcted myocardium were distributed in the positive region and negative region, respectively (Fig. 4a and Supplemental Movie 1). As a result, LV1 and LV2 predominantly represented the important spectral features for the discrimination of non-infarcted and infarcted myocardium. Raman bands appearing in both the negative direction of LV1 and the positive direction of LV2 seemed to indicate the spectral features of non-infarcted myocardium. In contrast, Raman bands of LV1 in the negative direction with the subtraction of LV2 in the positive direction represented the spectral features of infarcted myocardium.

In LV3, a derivative-like spectrum was obtained, especially representative Raman bands of heme proteins, such as 755, 1133, 1318, and 1590 cm−1 (Fig. 4d). The scores of LV3 were roughly distributed depending on patients, not on tissue species (Fig. 4a and Supplemental Movie 1). This result indicated that LV3 represented the difference among patients.

By using the prediction model derived by the PLS-DA, we sought to predict tissue species of non-infarcted and infarcted myocardium. Cross-validated detection accuracy of non-infarcted and infarcted myocardium is shown in Table 1. High leave-one-out cross-validated sensitivities and specificities of 99.98% and 99.92%, respectively, were obtained for the discrimination of non-infarcted and infarcted myocardium with the detection accuracy of 99.95%. We also confirmed that the high leave-one-patient-out cross-validated sensitivities and specificities of 99.66% and 98.72% with the detection accuracy of 96.80%, in which the tissue species of a patient were predicted by a prediction model constructed with other four patients data (Supplementary Tables 16). This highly accurate detection capacity for tissue species supported the notion that our findings for the spectral features in the LVs of PLS-DA provided important information about differences of the Raman spectra of tissue species. Furthermore, these results also demonstrated the fundamental feasibility of the Raman spectroscopy for the evaluation of old MI of patients on the basis of the constituent molecules.

Table 1 Detection power of non-infarcted and infarcted myocardium. Detection accuracy for the prediction of non-infarcted and infarcted myocardium was 99.95%.

Raman imaging of myocardium of marginal region of myocardial infarction

Raman imaging of non-infarcted and infarcted myocardium was performed by using PLS-DA. For reconstruction of Raman images, we obtained PLS-DA-predicted values defined as the scalar product of regression vectors of each tissue species at each pixel. Representative Raman images of the marginal area of non-infarcted and infarcted myocardium obtained from patient 3 revealed that the PLS-DA-predicted values of non-infarcted and infarcted myocardium highlighted these tissue species, and visualized the boundary of non-infarcted and infarcted myocardium (Fig. 5). These tissue distributions were consistent with the histology confirmed by HE- and Azan-stained serial section. Furthermore, fibrotic tissues that were present among cardiomyocytes were also visualized. Similar images were obtained for all the tissues from the other 4 patients (data not shown). These results indicated that the Raman spectroscopy with the PLS-DA has a capability to visualize non-infarcted and infarcted myocardium of human myocardium without staining.

Figure 5
figure 5

Representative 2D Raman images of a marginal region of non-infarcted and infarcted myocardium. (a,b) Two representative marginal regions of non-infarcted and infarcted myocardium of patient 3. Raman images at the dashed squares indicated in the HE-stained images were reconstructed by using PLS-DA-based prediction of tissue species. Scale bars in the HE-stained and Raman images in (a) and (b) are 200 µm and 100 µm, respectively.

Discussion

Our previous studies provided a proof-of-principle demonstration of Raman spectroscopy for the label-free evaluation of MI by using rat samples22,23. Here, we extended Raman spectroscopy to the human myocardium, and showed its feasibility for old MI by means of PLS-DA. We found that the Raman spectra of the human infarcted myocardium comprised various spectral components that did not appear in our previous rat study, while the spectral features of non-infarcted myocardium were almost identical to those in the rat22,23. Such varieties of Raman spectra of human infarcted myocardium hampered the discrimination of infarcted myocardium from non-infarcted myocardium. In the present study, our multivariate analytical method enables the prediction of non-infarcted and infarcted myocardium with high prediction accuracy.

In our previous study, we confirmed that strong Raman peaks at 755, 1133, and 1590 cm−1 of rat heart were assigned to the specific vibrational modes originated from a porphyrin ring in the center of heme proteins. In addition to these bands, the Raman bands at 648, 687, and 1366 cm−1 indicated the presence of reduced cytochrome c23. Thus, the Raman spectra obtained from human non-infarcted myocardium that included these Raman bands also seemed to have arisen from reduced forms of cytochromes c in cardiomyocytes (Fig. 1a). Reduced cytochrome c is generally present in viable cardiomyocytes. In contrast, there were no Raman bands at 648 and 687 cm−1 in the infarcted myocardium, indicating the Raman bands at 755, 1133, and 1590 cm−1 were assigned to heme proteins other than cytochrome c, such as hemoglobin, myoglobin, and other cytochromes23. Heme proteins are not common molecules in infarcted myocardium, which might appear by blood deposition during surgery or other contaminations following repair process of human old MI. As a result, patient variability of Raman spectra appeared in infarcted myocardium, but barely appeared in non-infarcted myocardium.

The intensities of the Raman bands at 755, 1133, 1318, and 1590 cm−1 in non-infarcted myocardium tend to depend on the intensities of these Raman bands of non-infarcted myocardium (Fig. 1). Especially in patient 4, large contribution of these Raman bands was obtained in the infarcted myocardium, and the intensities of these Raman bands in the non-infarcted myocardium were relatively larger than that in the other patients. This result indicated that the contributions of heme proteins appearing in the infarcted myocardium seemed to also appear in the non-infarcted myocardium; in other words, homogeneous background of the Raman spectrum of heme proteins might be present. The Raman spectra of non-infarcted and infarcted myocardium of each patient were obtained at near borders of MI, in which the distances between these observation points in non-infarcted and infarcted myocardium were within a few millimeters. One possible case for these homogeneous contributions of heme proteins in the non-infarcted and infarcted myocardium is the deposition of hemoglobin of blood. The myocardium was excised following LVR, thus blood must be exposed in the myocardium by hemorrhage following cardiac surgery. No red blood cells were observed in terms of HE-stained sections, indicating that the deposition of hemoglobin of hemolyzed blood might contribute to the Raman spectra in some cases.

In Raman spectroscopy, visible or near-infrared excitation light is irradiated to the specimen and observed inelastic scattering light reflects molecular information about the myocardium via Raman spectrum. Since contrast agents and any preparation are not needed, Raman spectroscopy has potential for intraoperative observation with minimal invasiveness. Indeed, several groups initiated clinical trials of Raman spectroscopy for label-free in vivo diagnosis in the fields of breast surgery28, gastrointestinal surgery29, and skin surgery30. Although further confirmation of the Raman spectroscopic technique under in vivo condition is required for the application to cardiac surgery, we anticipate that label-free discrimination of non-infarcted and infarcted myocardium will allow us intraoperative evaluation of myocardial viability of human patients in the future.

If our proposed approach could be applicable as an aid of cardiac surgery, some important insights can be provided for surgical treatments. Importantly, if this method is applied in vivo during surgery, intraoperative evaluation of myocardial viability under direct vision becomes possible, which cannot be realized by the conventional methods. The intraoperative evaluation of myocardial viability under direct vision can clarify the more accurate state and region of myocardial dysfunction. One possible application is LVR, in which the infarcted myocardium is excluded and the ventricular volume is reduced. The intraoperative evaluation of myocardial viability can accurately identify the region of viable myocardium, enabling proper resection of non-viable myocardium. Another possible application is coronary artery bypass grafting (CABG), in which the coronary circulation is reestablished by creating new routes around narrowed or blocked coronary arteries from large arteries. In CABG, the recoverable region of myocardium after reestablishing the coronary circulation can be estimated by accurate and in situ evaluation of myocardial viability. Furthermore, the accurate understanding of the state and region of myocardial dysfunction is important for treatment management of myocardial dysfunction not only during but also after cardiac surgery. Although further studies are required with large number of cases, we believe that our feasibility study will serve as a basis for a new medical device that enables the evaluation of myocardial viability under direct vision during cardiac surgery.

Conclusions

This is the first report demonstrating the evaluation of old MI in patients undergoing cardiac surgery by means of Raman spectroscopy employing PLS-DA. Our results revealed the feasibility of the Raman spectroscopy for the label-free assessment of myocardial viability of human myocardium. Our proposed approach may be a promising technique for direct evaluation of myocardial viability during surgery in the future.

Methods

Patients and sample preparation

All experiments were performed in accordance with the relevant guidelines and regulations. The study was approved by the Ethics Committee of Kyoto Prefectural University of Medicine (Permission No. RBMR-C-966). For all patients, fully informed consent was obtained after a full explanation of the study design and before the surgery. Myocardium was resected from five patients with old myocardial infarct under LVR at University Hospital, Kyoto Prefectural University of Medicine, conducted between August 2012 and January 2014 during cardiac surgery. The excised myocardium kept cold on ice was quickly transferred to the laboratory, embedded in frozen section compound (FSC 22; Leica, Wetzlar, Germany), snap-frozen in a freezing mixture (frozen carbon dioxide with acetone), and stored at −80 °C until experiment. The frozen samples were sliced into 5-µm-thick sections with a cryostat microtome (CM1900; Leica). Three serial sections were obtained from each tissue for Raman and histological analysis. The sections for the Raman analysis were mounted on a 0.17-mm-thick cover glass without fixation. The sections for the histological analysis were subjected to hematoxylin and eosin (HE) staining or Azan staining to confirm histology.

Raman spectroscopy

Raman spectra and the reflecting spectral images were acquired with a slit-scanning confocal Raman microscope (RAMAN-11; Nanophoton, Osaka, Japan) as described previously19,22,23,31. A frequency doubled Nd:YAG laser (532 nm) was employed for excitation. The excitation laser beam was focused into a line on a sample through a cylindrical lens and a high-magnitude water-immersion objective lens (UPLSAPO x60, NA = 1.2; Olympus, Tokyo, Japan) or a low-magnitude objective lens (UPlanFL N, x10, NA = 0.3; Olympus, Tokyo, Japan). The high-magnitude water-immersion objective lens was used for the acquisition of Raman spectra. These Raman spectra were also used for the construction of a prediction model with PLS-DA analysis in two-dimensional imaging. The low-magnitude objective lens was used for the acquisition of two-dimensional images.

Raman scattered light was collected with the same objective lens. The Raman scattered light was focused onto the input slit with 70 µm width of the spectrometer with a 600 grooves/mm grating, and detected with a two-dimensional image sensor (Pixis 400BR, −70 °C, 1340 × 400 pixels; Princeton Instruments, Trenton, NJ, USA). Owing to the line shaped focus of the excitation beam, a one-dimensional spectral image (one-dimensional space and Raman spectrum) from the line-illuminated site was simultaneously obtained. Each pixel data in space was used for a single data point in all spectral analysis. Two-dimensional Raman spectral images were obtained by scanning the line-shaped laser focus in a single direction. The excitation laser power and the exposure time was set at from 235 to 300 µW/µm2 on the sample plane, and 10 s, respectively.

Preprocessing of spectral analysis

Before spectral analysis, all raw Raman spectra were calibrated using the known lines of ethanol. The broad fluorescence background was fitted with the modified least-squares polynomial curve32 and subtracted it from calibrated Raman spectra with Nanophoton software (Nanophoton, Osaka, Japan). The background-removed Raman spectra were normalized with the integrated intensity of each Raman spectrum.

Statistical analysis of prediction capacity by means of a single band Raman spectrum

The prediction capacity by means of a single band Raman spectrum was evaluated by using the peak intensity of 687 cm−1, peak intensity of 755 cm−1, and peak position of 2930 to 2950 cm−1. The peak intensities of 687 and 755 cm−1 were obtained in terms of area intensity as shown in Supplementary Figure 2 to extract proper intensity of each Raman bands. To extract peak position of 2930 to 2950 cm−1 against noise contribution, Raman spectrum around 2940 cm−1 was fitted by Gaussian function in IGOR Pro software (WaveMetrics, OR, USA). Peak position comprised in the fitted Gaussian function was obtained for the statistical analysis. These spectral analyses were performed with programs written in programing language in MATLAB (Mathworks, Natick, MA, USA) and IGOR Pro software (WaveMetrics, OR, USA).

Student’s t test was performed with MATLAB (Mathworks, Natick, MA, USA) with the assumption of independent samples.

Receiver operating characteristic (ROC) analysis was employed to predict tissue species by using single band of Raman spectrum. ROC analysis was performed with programs written in the IGOR programing language in IGOR Pro software (WaveMetrics, OR, USA). Optimal sensitivities of non-infarcted and infarcted myocardium were defined as those yielding the minimal value for (1-sensitivitynon-infarcted)2 + (1-sensitivityinfarcted)2, where sensitivitynon-infarcted and sensitivityinfarcted respectively indicated the sensitivities of non-infarcted and infarcted myocardium.

PLS-DA analysis for the prediction of tissue species

PLS-DA analysis was employed for the construction of a prediction model of old MI. PLS-DA analysis is widely applied to construct quantitative prediction model for classification in spectral analysis by projecting the predicted and observable variables into a new space to maximize the covariance between the response and independent variables. In PLS-DA algorithm, observed Raman spectral data set of Stissue can be decomposed to LV matrix P and score matrix T:

$${S}_{tissue}=T{P}^{t}.$$
(1)

By using LV matrix P and score matrix T, we analyzed common spectral features of a Raman spectral data set, and constructed a prediction model. All calculations in PLS-DA were performed with PLS toolbox (eigenvector Research, WA, USA) in MATLAB (Mathworks, Natick, MA, USA).

Randomly obtained 1000 Raman spectra in each tissue species of each patient were analyzed. Since the cover glass retaining frozen sections exhibited Raman spectra from around 850 to 1020 cm−1, and Raman-silent region from around 2000 to 2800 cm−1 exhibited no significant Raman spectra in myocardium, we extracted Raman spectra from 596 to 836 cm−1, from 1024 to 1723 cm−1, and from 2818 to 2980 cm−1 of each individual spectrum for PLS-DA analysis. We constructed a prediction model with orthogonal condition and calculated LVs and scores. Leave-one-out cross validation analysis was employed for the verification of the predictive accuracy of the prediction model using PLS-DA33,34,35. The optimal number of LVs used in the prediction model was determined as first minimum value of number of misclassification rate to obtain optimum misclassification rate only with LVs involving low noise contribution as shown in Supplementary Figure 3.

For reconstruction of two-dimensional Raman images, we obtained the PLS-DA-predicted values defined as the scalar product of a regression vector with the Raman spectrum at each pixel. Original Raman spectral-images were obtained with 400 × 100 pixels in spatial dimension, and were binning

The regression vector was derived by PLS-DA analysis in the construction process of a prediction model as shown above. The Raman spectra of 596 to 836 cm−1, from 1024 to 1723 cm−1, and from 2818 to 2980 cm−1 were used for the reconstruction of two-dimensional prediction map of tissue species.

Raman spectroscopy of pure chemical of collagen

Powdered collagen type I derived from bovine Achilles tendon (C9879, Sigma-Aldrich, MO, USA) was placed on a slide glass, and observed a Raman spectrum using the slit-scanning confocal Raman microscope as shown above.