Introduction

The description of strongly correlated electron systems by means of elementary excitations has been a useful concept in modern condensed-matter physics1. Magnetic excitations or magnons and their associated quasi-particles are traditionally studied with inelastic neutron scattering. It is a recent finding that resonant inelastic x-ray scattering (RIXS) can actually probe magnons as a result of the core-hole spin-orbit coupling of transition metal L-edges2,3. Nowadays, RIXS is routinely used to measure magnetic excitations in various different types of materials including cuprates4, iridates5,6,7 and osmates8, with the highest energy resolution being ~25 meV at the Ir L3 absorption edge7,9.

There is great interest in improving the energy resolution of the RIXS spectrometer to sub-10 meV level to fully extract information necessary to understand the low-lying magnetic dynamics of iridates which have been irresolvable within the current best energy resolution. Of particular recent interests are such elusive systems as quantum spin liquids, topological insulator, Axion insulator and Weyl semimetal, which are emergent quantum phases in correlated spin-orbit coupled materials10,11,12.

The x-ray optics of state-of-the-art, high-resolution RIXS spectrometers are based on diced, spherical crystal analyzers in a near back-scattering geometry13,14,15. Due to the resonant character of the technique, the incident energy is determined by the absorption edge of a given magnetic ion. However, the highly symmetric crystal structure of Si provides very few choices for near-backscattering reflections, thereby limiting the highest resolution achievable. Other crystals of lower symmetry such as quartz and sapphire in principle offer many more reflection choices, some of which have much smaller intrinsic energy width than Si with very good throughput16. However, the technology to process these materials is still in its infancy17.

In this article, a novel RIXS spectrometer with sub-10 meV energy resolution at the Ir L3 edge is introduced, based on flat-quartz crystal optics18. The unique advantage of the flat-crystal analyzer is that its resolution no longer depends on any geometrical factors, such as the focusing of the mirrors, the divergences of the X-ray beams, or the exact shape of the spherical analyzer. As a new important feature, efficient polarization analysis without loss of energy resolution is also demonstrated. The principle of the instrument design is described, the performance of its implementation is characterized and representative measurements of phonon dispersion in diamond and magnetic excitations in Sr3Ir2O7 are provided.

Principle of the flat crystal RIXS analyzer

Flat-crystal x-ray optics are superior in performance to curved crystal systems, because they are free from figure errors and strains that degrade the performance18,19,20,21,22. However, the angular acceptance of flat crystal systems is so small that they are not practical in situations where weak, radially scattered radiations from a sample needs to be collected, as is the case of a RIXS spectrometer. Recent advances in multilayer optics provide a way to overcome this impediment23,24,25,26 through the use of curved, laterally-graded mirrors (called Montel mirror) that can easily accept radiations in a large solid angle.

Figure 1 shows a prototype of the flat crystal RIXS analyzer implemented for the operation at the Ir L3 absorption edge (11.215 keV) at the RIXS beamline 27-ID of the Advanced Photon Source (see Supplementary Figure 1 for a photograph of the installed spectrometer). The Montel mirror is located at a distance of 200 mm from the sample to serve as initial collimator27. It accepts more than 10 × 10 mrad2 of scattered radiation from the sample in the vertical and horizontal directions, and collimates it to less than 100 × 100 μrad2. In the installed spectrometer, a Si(111) asymmetrically cut collimator (C-crystal) is placed at a distance of ~300 mm from the Montel mirror center to avoid collisions with the Montel mirror housing, the 6-circle spectrometer and the Plexiglass box. Angular acceptance of the C-crystal matches the emittance of the Montel mirror and further reduces the beam divergence to ~6 μrad in the vertical direction. For the analyzer (A-crystal), the availability of nearly perfect α-Quartz provides a near-backscattering reflection (309) with an intrinsic energy width of 3.7 meV, ~4 times smaller than the highest resolving Si(844) reflection (14.6 meV) at this incident energy. The A-crystal is placed at a distance of about 1000 mm from the C-crystal to secure a proper space for the detector which is positioned below the C-cryatal and collects the reflected beam from the A-crystal. The acceptance of the A-crystal (11.5 μrad) is large enough to accept the collimated x-rays from the C-crystal. The X-ray beam from the A-crystal is collected by a position-sensitive detector (Mythen by Dectris), with an effective length of 64 mm, divided into 1280 pixels of 50 μm width each. The energy resolution of the flat crystal RIXS analyzer is solely determined by the intrinsic width of the A-crystal because the bandpasses of the Montel mirror and the C-crystal are 145 eV and 6.3 eV, respectively. In the following, this RIXS analyzer system will be referred to as a CA-analyzer.

Figure 1
figure 1

Schematic of the flat-crystal RIXS spectrometer. This spectrometer, designed for the Ir L3 edge at 11.215 keV, consists of a nested set (Montel) of two parabolic, laterally-graded Ru/C multilayer mirrors, followed by an asymmetrically cut Si(111) collimator (C-crystal) and a symmetric Quartz(309) analyzer (A-crystal). The beam from the A-crystal is collected by a position-sensitive (strip-) detector. For polarization analysis, a Si(444) polarizer (P-crystal) is placed in the beam from the A-crystal, with the detector relocated to receive the diffracted beam from the P-crystal.

In an ideal RIXS measurement, the energy, momentum and polarization of scattered radiation are measured. To date, polarization measurements using curved crystal optics have been inefficient and tend to compromise the energy resolution28,29. A remarkable, additional advantage of the CA analyzer is to analyze polarization with high efficiency and without loss of energy resolution. For this purpose, the beam from the A crystal is further diffracted by a Si(444) polarizer P-crystal at a Bragg angle near 45°, with the detector relocated to receive the beam from the P-crystal.

Results

Performance of the flat-crystal RIXS spectrometer

For alignment and characterization, elastically scattered radiation from a tape sample (3 M Scotch Magic tape) is used. The incident beam is monochromatized to a band pass of 8.9 meV. The CA analyzer is positioned at a scattering angle 2θ = 15°. In order to assess the performance of the CA-analyzer, static diffraction characteristics as well as dynamic rocking curves and energy scans were extensively simulated. Extended DuMond diagrams30 based on two-beam dynamical diffraction theory31 were used to compute 3-dimensional reflectivity profiles as function of both the angle and the energy at various points along the chain of the crystals. In order to calculate the energy- or angle-scan profiles, the corresponding profiles are convoluted along the angle- or energy-axis.

Figure 2a shows the measured rocking curve of the asymmetric Si (111) C-crystal. The measured full-width-half-maximum (FWHM) value is 0.0076°, which could be reproduced by the simulation (solid curve) that are based on a model with an incidence divergence of 92 μrad from the Montel mirror and an incident energy bandwidth of 8.9 meV. Figure 2b shows the rocking curve for the symmetric quartz (309) A-crystal. The FWHM is about 0.0018°, which also could be reproduced by the simulations. These results indicate the near-ideal quality of the C- and A-crystals, as well as their strain-free mountings and an effective degree of collimation of the Montel mirror.

Figure 2
figure 2

Measured angle scans and x-ray intensity profiles. (a) The FWHM of the C-crystal angle scan is 0.0076°, which agrees with simulations based on its asymmetry angle and assuming 92 μrad incident divergence from the Montel mirror. (b) The FWHM of the A-crystal scan is 0.0018°, in agreement with simulations based on a divergence of 6.1 μrad from the C-crystal and a 8.9 meV bandpass. This angular width corresponds to an energy spread of 9.7 meV. (c) X-ray intensity profiles as measured by the positon-sensitive detector. Blue trace, top-axis, corresponds to downstream of Montel mirror, while black and green traces, bottom axis, correspond to downstream of the C- and A-crystal respectively. The width of the blue profile indicates that the degree of collimation by the Montel mirror is as expected. The preservation of the same distinct pattern before and after the C-crystal indicates that collimation by the C crystal performs as designed.

Regarding the measured efficiency, the C-crystal reflects 70% of x-rays coming from the Montel mirror, while the A-crystal passes 30% of x-rays from the C-crystal to the detector, amounting to an overall efficiency of about 21%. This is in very good agreement with the simulations that give an overall efficiency of 22%, which indicates that the successive collimation by the Montel mirror and the C-crystal works as designed. This is also borne out by more detailed observations of the vertical beam profiles measured at various points along the beam path using the position-sensitive detector. Figure 2c shows the profiles downstream of the Montel mirror, the C-crystal and the A-crystal. Firstly, the full size of the beam emerging from the Montel mirror was measured to be about 2.6 mm as shown in the blue line (top and right axes). The angular acceptance corresponding to this measured size is about 14.5 mrad. The intensity modulations are attributed to the figure errors of the Montel mirror. The black trace in Fig. 2c shows the beam profile downstream of the C-crystal. The measured beam size is now 40.8 mm, consistent with the designed asymmetry factor of b = 15.6, indicating that the full beam from the Montel mirror is accepted and passed along. The incident intensity fluctuations are preserved. Finally, the green trace shows the beam profile downstream of the symmetric A-crystal. Again, the intensity fluctuations are preserved, and the reduction in intensity is consistent with a reduced reflectivity for quartz and its narrow energy acceptance.

The spectral resolution function of the CA-analyzer can be obtained in two ways: by scanning the analyzer itself or by scanning the incident energy. For the first scan mode, a single angle scan of the A-crystal is sufficient, because the energy bandpass of the C crystal is very large (6.3 eV) compared to that of the A-crystal. Thus, the A-crystal rocking curve shown in Fig. 2b, when the angle is converted to the energy, gives the spectral resolution function with a width of just under 10 meV. For the second mode, the spectral resolution function is obtained by scanning the high-resolution monochromator. Figure 3a shows an image plot of such an incident energy scan. The x-axis represents the energy loss, defined as the difference between the incident and scattered energy. The y-axis is the position on the position-sensitive detector. It can be seen that all elastic intensities are well aligned along the zero-energy loss. The spectral resolution function is obtained by integrating spectral intensities at all the positions as a function of energy loss. Figure 3b shows the spectral resolution function obtained in this way. It has a FWHM of 9.7 meV, consistent with the corresponding simulation (red solid curve) and the result from Fig. 2b. Considering the energy bandpass of the incident beam of 8.9 meV, the intrinsic energy resolution of the CA-analyzer can be approximated by \(\sqrt{{9.72}^{2}-{8.92}^{2}}\) = 3.9 meV, very close to the theoretical value of 3.7 meV. For comparison, the spectral resolution function of a spherical diced Si(844) analyzer is also plotted.

Figure 3
figure 3

Spectral resolution functions. (a) Image plot of elastic scattering as a function of incident energy and position along the detector. (b) Spectral resolution function of the flat-crystal spectrometer (open circles), as obtained by integrating intensity along position. FWHM is 9.7 meV. Its profile agrees well with simulation (red line). For comparison, the resolution function of a Si(844) spherical analyzer (25 meV) (blue line) is shown.

Longitudinal acoustic and optical phonons in Diamond crystal

For the purpose of characterizing the performance of the flat-crystal RIXS spectrometer and calibrating its energy scale, we have reproduced the well-known phonon spectra from diamond, which were investigated previously by various high resolution spectroscopies such as inelastic neutron scattering (INS), Raman scattering, and inelastic x-ray scattering (IXS)32. Diamond phonons are well suited for the purpose because their energies span a sufficiently large energy window, and the specific balance factor precisely defines the intensity ratio between Stokes and anti-Stokes scattering, allowing judgment of the uniform response of the instrument.

For these measurements, a flat Diamond(111) crystal was used with the azimuthal angle chosen such that the scattering (1 0 −1) direction was contained in the scattering plane. Longitudinal acoustic (LA) and optical (LO) phonons were measured along Q = (H H H), with 1 < H < 1.5. The momentum resolution corresponding to the angular acceptance of the mirror of 14.5 mrad is about 0.022 rlu, or 0.39 nm−1.

The image plot on the left of Fig. 4a shows the LA phonon spectrum at H = 1.02 recorded by the CA-analyzer. The strong intensity near zero energy loss is from the (1 1 1) Bragg peak. The two intensity streaks at positive and negative energy loss correspond to Stokes and anti-Stokes scattering, respectively. The image plot in the center shows the LA phonon spectrum at H = 1.04. Here, elastic scatterings are suppressed and the separation between the Stokes and anti-Stokes scattering becomes larger. For H = 1.08, as shown in the image plot on the right, anti-Stokes scattering is out of the measured range and the Stokes scattering appears at a higher energy loss. The phonon spectra in Fig. 4b and c were obtained by integrating the intensities as a function of the energy loss. Here, we restrict the integration position range to half the length of the detector (20 mm) as indicated by the yellow dashed lines in Fig. 4a to have better momentum resolution (0.011 rlu). Note that because of a highly collimated nature of the collected x-rays, each x-ray at each position of the position-sensitive detector preserves its information of the momentum transfer and one can have better momentum resolution by restricting the region of interests. For the case of H = 1.01, 1.02, and 1.03, the areas with strong elastic intensity around the −2 mm position are excluded. This is one of the powerful features of the flat-crystal RIXS spectrometer in combination with a position-sensitive detector; the integration ranges can be chosen a posteriori to set the momentum resolution or mask off unwanted areas.

Figure 4
figure 4

Longitudinal acoustic (LA) and optical (LO) phonons in diamond. (a) Image plots of the LA phonon spectra for H = 1.02 (left), 1.04 (center), and 1.08 (right). The two streaks at positive and negative energy loss correspond to Stokes and anti-Stokes scattering respectively. The strong intensity at zero energy loss for H = 1.02 comes from the (1 1 1) Bragg peak. The yellow dashed lines indicate the integration range for extracting phonon spectra, restricted for better momentum resolution. (b) Phonon spectra for H = 1.01–1.10. (c) Phonon spectra for H = 1.05, 1.2, 1.3, and 1.4 with extended energy range. The LO phonons are now visible around 160 meV. The red lines in (b) and (c) are the Voigt function fits that satisfy the detailed balance factor. (d) Dispersion curves for both longitudinal modes. Measured data agree well with IXS results32.

Figure 4b shows the LA phonon spectra. The measured spectra are fitted using the Voigt functions that satisfy the detailed balance factor. Figure 4c shows the spectra for H = 1.05, 1.2, 1.3, and 1.4 over a wider energy range. The LO phonon is seen at around 160 meV, along with the highly dispersing LA phonon. Figure 4d shows the resulting dispersion of the two longitudinal phonon modes. The results from the CA-analyzer agree very well with the IXS results32, indicating that the high-resolution monochromator is properly calibrated. Alternatively, the same spectra can be obtained by scanning the angle of the A-crystal. Figure 5a shows an image plot for H = 1.04 taken by such an angle scan, which is directly comparable to the center plot in Fig. 4a. The phonon spectra obtained in both scan modes shown in Fig. 5b agree well with each other.

Figure 5
figure 5

Phonon spectrum for H = 1.04 obtained by A-crystal angle scan. (a) Image plot of the A-crystal angle scan, equivalent to the center panel of Fig. 4a and b. The phonon spectrum obtained with both scan modes agree well with each other. The red line is a Voigt fit.

Magnon spectrum in Sr3Ir2O7 at the magnetic zone center

The bilayer iridate Sr3Ir2O7 (lattice constants a = b = 0.55 and c = 2.09 nm) is a magnetic insulator driven by spin-orbit coupling with a small charge gap33. Previous RIXS study found an exceptionally large magnon gap of ~90 meV for both acoustic and optical branches, which is interpreted as resulting from bond-directional, pseudodipolar interactions that are strongly enhanced near the metal-insulator transition boundary6.

The magnon spectrum in Sr3Ir2O7 was measured at the magnetic zone center (Q = (1 0 26.5)) by the CA-analyzer as a demonstration of a measurement in a real Ir sample. Figure 6a shows the measured full spectrum. The red line is the Voigt fit. The elastic line fit indicates that the elastic scattering at zero energy loss has a FWHM of 9.7 meV, confirming the overall energy resolution determined earlier. Magnon peak measured by a standard RIXS spectrometer with an overall energy resolution of 25 meV6 is also plotted in Fig. 6a for comparison. Figure 6b shows the magnon peak at ~90 meV, which is not resolution-limited. The magnon peak could be fit with two resolution function peaks (broken red lines) where the lower (higher) energy peak corresponds to the acoustic (optical) branch. Note that the previous RIXS study could identify two magnon branches by carrying out the c-axis momentum transfer dependence6.

Figure 6
figure 6

The magnon spectrum of Sr3Ir2O7 at the magnetic zone center. (a) The spectrum measured by the flat-crystal RIXS spectrometer is shown as filled black circles (left axis). The red line is the Voigt fit. The elastic line fit (resolution function) indicates the overall energy resolution of 9.7 meV. A sharp magnon feature is seen at ~90 meV. For comparison, a spectrum obtained with a standard RIXS spectrometer (25 meV resolution) is also shown (open blue circles, right axis). (b) The measured magnon peak is fit with two resolution function peaks, which represent acoustic and optical magnon modes in the bilayer Sr3Ir2O7.

Here, we compare the spectral efficiency of the CA-analyzer to the spherical analyzer using the 90 meV magnon spectra in Sr3Ir2O7. The magnon peak of the 25 meV RIXS spectrometer in Fig. 6a has a FWHM of 28 meV, a peak amplitude of ~90 counts per second (cps) and an integrated intensity of ~2500. In the CA-analyzer measurement, the incident flux is lower (23%) due to an additional Si(844) channel cut and the solid angle captured by the Montel mirror is smaller (11%). In the same condition, the 25 meV RIXS spectrometer is estimated to have a peak amplitude of ~2 cps and an integrated intensity of ~63. In Fig. 6b, the magnon peak of the CA-analyzer has a peak amplitude of ~3 cps and an integrated intensity of ~39. We can see that the spectral efficiency of the CA-analyzer is in a comparable range to that of the spherical analyzer.

Polarization analysis with high efficiency and high energy resolution

Scattered radiation typically contains both σ− and π− polarizations4. Equally efficient detection of both polarizations is essential. The CA-analyzer presented in this work has a calculated π/σ detection ratio of 0.88, suitable for a practical instrument.

Polarization analysis can then be accomplished by including a polarizer between the A-crystal and the detector. A polarizer (P) crystal should have a scattering angle close to 90°. Si(444) provides 89.68° at 11.215 keV, and is therefore suitable. A symmetric Si(444) crystal was used for this proof-of principle study. As shown in Fig. 1, the P-crystal is placed in the beam emerging from the A-crystal and the detector is relocated to intercept the diffracted beam from the P-crystal. In order to determine the energy resolution in the polarization analyzing mode, elastic scatterings from a tape sample were collected at scattering angles of 2θ = 0° and 13°. Figure 7a shows the spectra for both scattering angles, obtained by scanning the incident energy. Two spectra are virtually indistinguishable from each other, with a FWHM of 9.7 meV. These measurements clearly show that the overall energy resolution is unaffected by adding a polarization analyzer.

Figure 7
figure 7

Polarization analysis performance (polarizer setup as shown in Fig. 1). (a) Elastic spectra at the scattering angle of 2θ = 15° with the polarizer and without the polarizer, obtained by scanning the incident energy. Two spectra are virtually indistinguishable from each other, with a FWHM of 9.7 meV. (b) Intensity profiles of the A- and P-crystal as a function of position along the detector. The low overall efficiency is due to poor reflectivity of the P-crystal along its length, except for the region around the center (−5 mm < x < 5 mm) indicated by the blue lines.

Regarding the efficiency, however, the P-crystal used for this demonstration was found to be somewhat strained, not diffracting uniformly over its length and rendering the measured overall efficiency (35%) lower than expected. Figure 7b shows the intensity profiles of the A- and P-crystal as a function of position along the detector, recorded at peak intensity. It can be seen that the low overall efficiency is due to poor reflectivity of the P-crystal along its length, except for a small region around the center (−5 mm < x < 5 mm). Considering only this region, the reflection efficiency is as good as 76%, comparable to what was expected from simulations.

This measurement shows that adding a polarization analyzer to the flat-crystal spectrometer can be achieved without compromising the energy resolution and with high efficiency, and that a less-than-perfect P-crystal lowers the overall efficiency but not the overall energy resolution. We note that the reported efficiency of a polarization analyzer for a standard (spherical analyzer based) RIXS spectrometer is only 1~2% and an energy resolution better than 100 meV has not yet been achieved28,29.

Discussion and Outlook

In this work, a novel, flat-crystal RIXS spectrometer has been designed and implemented, achieving an unprecedented energy resolution better than 10 meV at the Ir-L3 edge, with spectral efficiency comparable to that of a state-of-the-art instrument based on diced, spherical analyzer optics. Besides its superior energy resolution, the flat-crystal system has the capability to analyze polarization with no loss of resolution and good efficiency. There are additional features: convenient post-measurement selection of the momentum resolution, and a large sample space for complex sample environments such as high-pressure and magnetic-field. The CA RIXS spectrometer can effectively augments the RIXS spectroscopic capability in terms of its high resolution and scattered x-ray polarization analysis.

Improvements in performance and extensions to other absorption edge energies can be realized in future implementation of the instrument. The best energy resolution at optimal throughput is achieved when the incident band pass is matched to the intrinsic energy width of the A-crystal. For the present prototype, according to simulations, reducing the incident energy bandpass from the current 8.9 meV to 3.7 meV would improve the overall energy resolution from 9.7 to 5.2 meV (with the same throughput) and raise the spectral efficiency from 21% to 49%.

In order to extend the concept of the flat-crystal spectrometer to other incident energies, its components should be selected following these guidelines: (1) The choice of A-crystal determines the ultimate energy resolution and therefore a reflection with an appropriately small intrinsic width and high reflectivity should be selected. Compared with silicon that has only a few tens of Bragg reflections in the medium energy range (<16 keV), quartz has hundreds to thousands of reflections. This enables us to implement flat-quartz analyzers at virtually any photon energies with resolution up to 1 meV. (2) The C-crystal should be of low reflection order, to attain a large angular acceptance at moderate asymmetry. Si(111) is a good choice for most situations. The Montel mirror needs to be designed to achieve appropriate collection, collimation and throughput at a chosen working distance. (3) Finally, the P-crystal needs to provide a near 90 scattering angle at the desired energy with good reflectivity.

Methods

Upstream configuration

The diamond(111) high-heat-load monochromator reflects x-rays from two in-line undulators into a four-bounce high-resolution monochromator, which consists of two monolithic Si(844) channel-cut crystals. The resulting energy bandpass is 8.9 meV, determined by a simulation based on dynamic diffraction theory. The beam is then focused by a set of Kirkpatrick-Baez mirrors, yielding a typical spot size of 10 × 40 μm2 FWHM (v × h) at the sample. The sample scattering plane is horizontal, while the scattering plane of the CA crystals is vertical.

Collimating (Montel) mirror

A laterally-graded parabolic collimating nested (Montel) mirror, where two one-dimensionally figured multilayer mirrors are attached orthogonally, was chosen for its superior surface figuring, compact design, and easy instrumentation and stability control, compared with traditional designs such as a Kirkpatrick-Baez (KB) system. It is designed to collimate scattered x-ray beams around the Ir-L3 absorption photon energy (11.215 keV) in both vertical and horizontal directions. Mirror parameters (parabolic shape, overall dimensions, choices of multilayer reflector and spacer elements) were determined by the input requirements of the CA RIXS spectrometer. The distance between the scattering source and the mirror center is 200 mm and the mirror length is 150 mm. The Bragg angle is 1.02° and the angular acceptance is as large as 14.5 mrad. The multilayer reflector and spacer are Ru and C, respectively. Detailed design parameters and its performance can be found in ref.25. A small vertical spot size is important to obtain the required x-ray collimation (<100 μrad) after the Montel mirror. Previous study showed that the x-ray collimation in the horizontal direction is rather poor (~200 μrad) due to the large horizontal spot size27. However, this does not significantly affect the efficiency or energy resolution of the CA analyzer because the horizontal angular acceptances of both C and A crystals are larger than 200 μrad.

Simulations

In designing the CA-RIXS spectrometer, care was taken to maximize the incident solid-angle acceptance, optimize the throughput, maintain the best energy resolution and provide efficient polarization analysis without sacrificing resolution. This was accomplished by carefully selecting suitable crystal reflection and asymmetry angles. For guidance in this selection process, diffraction conditions and crystal scans were simulated in extended 3-dimensional DuMond diagrams30, based on two-beam dynamical diffraction theory31 and representing intensities after successive reflections and asymmetry transformation as a function of angle and energy.