Abstract
We evaluated bleb fluid images taken after Baerveldt glaucoma implantation. T2-weighted images of bleb fluid were scanned with 3 Tesla magnetic resonance imaging in 52 patients who had undergone tube-shunt surgery using the 350-mm2 endplate Baerveldt glaucoma implant; three-dimensional images were constructed from these images. Bleb fluid images were classified into either a layer of bleb fluid on either side of the endplate (double bleb layer group; n = 24) or one layer outside the endplate (single bleb layer group; n = 28). Despite there being no correlation between the bleb volume and the postoperative IOP (r = −0.080; P = 0.57), the double bleb layer group had significantly lower postoperative IOPs than the single bleb layer group (12.3 ± 3.8 mmHg vs. 14.7 ± 4.1 mmHg, respectively; P = 0.033). The single bleb layer was significantly related to higher numbers of prior intraocular surgeries (relative risk = 2.85; P = 0.0014). Formation of a layer of bleb fluid on either side of the endplate may have resulted in the lower postoperative IOPs after Baerveldt glaucoma implantation. Repeated intraocular surgery adversely affects formation of the double bleb layer.
Similar content being viewed by others
Introduction
Tube-shunt surgery using glaucoma drainage implants has become increasingly popular in refractory glaucoma in patients who have undergone one or more intraocular surgeries1. The Tube Versus Trabeculectomy Study showed that tube-shunt surgery using the Baerveldt glaucoma implant with a 350-mm2 endplate offers a higher success rate in patients that have undergone prior trabeculectomy and/or cataract surgery than those who underwent trabeculectomy alone2. Because the intraocular pressure (IOP)-lowering effect of trabeculectomy depends on the formation of a filtering bleb in the conjunctiva near the corneal limbus, surgical success is greatly affected by any prior surgical incisions in the conjunctiva3. Furthermore, because the Baerveldt glaucoma implant has a tube that is inserted into the eye and an endplate placed in the posterior region of the eye, the filtering bleb is formed around the endplate4. The surgical outcome of this procedure appears to be less affected by conjunctival surgical incisions than trabeculectomies, because the bleb forms in the posterior region of the eye compared to the bleb formation after trabeculectomy.
Unlike the filtering bleb that results from trabeculectomies, the bleb fluid structure created by Baerveldt glaucoma implantation cannot be assessed by the instruments used in standard ophthalmic examinations such as anterior-segment optical coherence tomography or ultrasound biomicroscopy because of its posterior location. Magnetic resonance imaging (MRI) is a useful tool for noninvasively evaluating bleb fluid structure after tube-shunt surgery5,6,7,8. T2-weighted images can be used to visualize and quantify the bleb fluid around the endplate, and the bleb image is shown as a dark circumlunar band (endplate) surrounded by a pocket of water density (bleb fluid)7. A recent cross-sectional study using 3 Tesla MRI showed that the bleb structure consists of two fluid layers divided by the endplate8. These previous case series have demonstrated that filtering bleb volume is negatively correlated with postoperative IOP7, 8. However, one study included eight cases with Ahmed glaucoma valve and Molteno implants7, whereas another study included 27 cases with Ahmed glaucoma valve and Baerveldt glaucoma implant with a 350-mm2 or 250-mm2 endplate8. IOP-lowering effect of these implants is affected by the size9,10,11, material12, and internal valve of the endplate10, 11. Although a larger endplate appears to be associated with a larger bleb volume, comparisons between the 350-mm2 and 500-mm2 endplate Baerveldt glaucoma implants and between the 350-mm2 and 250-mm2 endplate Baerveldt glaucoma implants showed a higher success rate for the 350-mm2 endplate than that for the 500-mm2 endplate13 and a comparable success rate between the 350-mm2 and 250-mm2 endplates14. Moreover, 11 eyes with the Ahmed glaucoma valve showed no correlation between the filtering bleb volume and postoperative IOP15. To assess these controversies, it is necessary to visualize bleb images and quantify bleb volumes in a large sample population implanted with the same glaucoma drainage implants having same size endplates using MRI.
The purpose of the present study was to assess the bleb structure after tube-shunt surgery using a 350-mm2 endplate Baerveldt glaucoma implant, and to determine the parameters of bleb images that correlated with postoperative IOP in a large sample size.
Methods
Patient Selection
This cross-sectional case series was approved by the institutional review board of Fukui University Hospital, Fukui, Japan. The protocol followed the guidelines of the Declaration of Helsinki. Written informed consent was obtained from all patients after detailed explanation of the procedures involved.
Patients previously treated with Baerveldt glaucoma implantation at the Fukui University Hospital were recruited. Orbital MRIs were obtained more than 6 months after surgery between December 1, 2014 and March 31, 2016. Patients aged 20 years or older who underwent tube-shunt surgery with the Baerveldt glaucoma implant BG101-350 or BG102-350 (Abbott Medical Optics, Abbott Park, IL) were included. Patients treated with other endplate types of the Baerveldt glaucoma implant (BG103-250); those who underwent Baerveldt glaucoma implantation within 6 months or less; those who had undergone other tube-shunt surgeries; patients with contraindications for MRI including patient intolerance, cardiac pacemakers, or presence of tattoos or metal in their bodies; or those who showed postoperative complications such as a bleb leak or hypotony at the time of MRI were excluded. The three-dimensional images were constructed using a volume-rendering tool software (Advantage Workstation VolumeShare7 Ext5; GE Healthcare, Milwaukee, WI).
Surgical Procedures
One surgeon (MI) performed all surgeries, and the surgical procedure was performed as described in this section. In brief, all tube-shunt surgeries were performed using a 350-mm2 Baerveldt glaucoma implant (BG101-350 or BG102-350). The silicone tube was completely occluded with a 8-0 absorbable vicryl (Coated VICRYL; Ethicon, Somerville, OH) suture to minimize the risk of early postoperative hypotony. A fornix-based conjunctival flap was created after the administration of subconjunctival and sub-Tenon’s xylocaine anesthesia. The silicone plate was preferentially placed in the superotemporal scleral quadrant. If the superotemporal quadrant had intensive surgical conjunctival scarring, the plate was placed in one of the other quadrants. After the endplate was placed under the rectus muscles, it was fixed on the scleral surface with a 7-0 nylon about 10 mm from the corneal limbus. A scleral tunnel was created into the anterior chamber with a 23-gauge needle to insert the tube into the anterior chamber. As for tube insertion into the vitreous cavity, a 20-gauge needle was used to penetrate the vitreous space. Then, the tube with the Hoffmann elbow (BG102-350) was trans-sclerally inserted into the vitreous space and sutured to the sclera with a 9-0 nylon. To reduce the frequency of early postoperative IOP elevation, Sherwood slits were created in the tube with the needle of the 9-0 nylon. The tube was covered with a patch graft of preserved sclera supplied by the eye bank. The scleral patch and the conjunctival flap were sutured using a 9-0 nylon and 8-0 absorbable vicryl suture, respectively.
All patients received similar postoperative topical medication, with 0.5% levofloxacin for 3 weeks and 0.1% betamethasone sodium phosphate for 6 months.
Data Collection
Patient data including sex, age, type of glaucoma, axial length, preoperative IOP, the endplate position, postoperative IOP, history of previous intraocular surgery, and the number of glaucoma medications were collected.
Magnetic Resonance Imaging
All scans were obtained more than 6 months after surgery. High-resolution orbital images were obtained using a 3 Tesla scanner (Discovery MR 750 3.0 T; GE Healthcare, Milwaukee, WI) in combination with a 32-element phased-array head coil with fast imaging sequences employing steady-state acquisition cycled phases (FIESTA-C). The imaging parameters were as follows: repetition time, 5.6 ms; echo time, 2.7 ms; field-of-view, 180 mm; matrix size, 320 × 288; and a slice thickness of 0.6 mm with an overlap thickness of 0.3 mm, resulting in an in-plane resolution of 0.56 mm × 0.63 mm8. Images of the coronal plane were obtained perpendicular to the plane containing both optic nerves. Acquisition time for the whole scan was approximately 3 min. Subjects were repeatedly asked to avoid unnecessary movements during scanning. Baerveldt glaucoma implants are not ferromagnetic and are therefore MRI safe. The specific absorption rate was calculated to be less than 2.0 W/kg, which is classified as “less invasive” for patients. None of the patients felt warmth from the implants during the scans.
Measurement of Bleb Volume
MRI scans obtained were analyzed using the ImageJ software (available at http://rsb.info.nih.gov/ij/; developed by Wayne Rasband, MD, PhD, National Institutes of Health, Bethesda, MD) on a personal computer8. Image data were saved as DICOM files and imported into the ImageJ software. A threshold was applied to identify individual blebs in each slice, and bleb area was calculated using the ImageJ software by two independent observers (KI, SA). For an automatic threshold selection, Otsu thresholding16 was chosen. Interobserver reliability test was obtained by an intraclass correlation coefficient (ICC) from a 2-way mixed effect model. The ICC was 0.984 (95% confidence interval, 0.973 – 0.991). The mean values of bleb areas quantified by two observers were integrated to calculate the bleb volume.
Primary Outcome Measure
The primary outcome measure was the classification of bleb layer formation.
Secondary Outcome Measures
The secondary outcome measures included relationships between the parameters of bleb images (bleb volume or classification based on the structure) and preoperative patient data, endplate position, and postoperative IOP.
Statistical Analysis
Univariate comparisons between groups were performed using the Student t-test, Welch t-test, and the Mann–Whitney U nonparametric test. The correlation between bleb volume and postoperative IOP was analyzed using the Spearman rank correlation coefficient. P values < 0.05 were considered statistically significant. Multivariate analysis was performed to determine the prognostic factors for bleb structures using the Logistic regression model.
Results
Patient Characteristics
Sixty-two patients were enrolled in the present study. Two patients withdrew from the study because of poor physical condition at the time of MRI. In eight patients, the quality of MRI scans was determined as insufficient for the quantification of bleb areas and volumes because of body motion. Therefore, their data were excluded from the study. In total, 52 blebs were evaluated. The mean ± standard deviation length of time between surgery and MRI was 15.2 ± 10.1 months. Table 1 summarizes the baseline characteristics of patients and the position of the Baerveldt glaucoma implant. In 40 out of 52 eyes, the endplate was placed in the superotemporal quadrant.
Primary Outcome Measure
In all cases, a curvilinear low-intensity band that corresponded to the signal intensity of the silicone material was observed at the location of the endplate. Areas with the same high intensity as the intraocular fluid, corresponding to the signal intensity of the aqueous humor in the filtering bleb, were located adjacent to the endplate. In 28 out of 52 eyes (53.8%), the signal corresponding to the aqueous humor in the filtering bleb was detected on the outside of the endplate. The remainder of the eyes (46.2%) showed aqueous humor signals on both sides of the endplate. Based on this information, we classified the eyes into two groups: the single bleb layer group and the double bleb layer group (Fig. 1). Three-dimensional images were obtained in all cases. In both the single and double bleb layer groups, four signal-free holes were detected in each bleb layer. Because the endplate has four fenestrations, each signal-free hole was located on both sides of these fenestrations in the endplate (Fig. 2; Supplement Movie 1S).
Three hundred sixty degree panoramas of magnetic resonance imaging scans of eyes with the Baerveldt glaucoma implant
(A) A 360-degree panorama of an eye in the single bleb layer group. (B) A 360-degree panorama of an eye in the double bleb layer group.
Secondary Outcome Measures
Relationships between the parameters of bleb images and postoperative IOP
The mean ± standard deviation bleb volume was 0.23 ± 0.15 cm3 (n = 52). There was no significant correlation between bleb volume and postoperative IOP (r = –0.080; P = 0.57) (Fig. 3). There were no significant correlations between bleb volume and postoperative IOP in subgroup analyzes in both the single bleb layer group (r = 0.024; P = 0.90) and the double bleb layer group (r = 0.016; P = 0.94).
Comparisons of postoperative IOPs between the single bleb layer group and the double bleb layer group revealed significantly lower postoperative IOPs in the double bleb layer group than the single bleb layer group (12.3 ± 3.8 mmHg vs. 14.7 ± 4.1 mmHg, respectively; P = 0.033). The number of postoperative glaucoma medications taken did not differ significantly between the two groups (P = 0.87). Comparison of bleb volumes between the two groups showed a significantly larger mean bleb volume in the double bleb layer group than that of the single bleb layer group (0.28 ± 0.2 cm3 vs. 0.19 ± 0.1 cm3, respectively; P = 0.024).
Comparison of patient characteristics between the single bleb layer and double bleb layer groups
Table 2 shows patient data of the single bleb layer and double bleb layer groups. Among the preoperative data, there was a significant difference in the number of intraocular surgeries before Baerveldt glaucoma implantation between the two groups. The single bleb layer group had undergone more intraocular surgeries before Baerveldt glaucoma implantation than the double bleb layer group (2.8 ± 1.4 vs. 1.9 ± 0.8, respectively; P = 0.0068). However, there was no significant difference about details of prior surgeries between the groups (Supplement Table 1S). No significant differences were found in age, preoperative IOP, number of preoperative and postoperative glaucoma medications, axial length, length of time between surgery and MRI, type of glaucoma, tube insertion position, or the endplate position between the groups.
Determinants for the single/double bleb layer
Patient characteristics including age, preoperative IOP, number of preoperative and postoperative glaucoma medications, axial length, length of time between surgery and MRI, number of previous intraocular surgeries, type of glaucoma, tube insertion position, and endplate position, were evaluated as possible determinants for the formation of the single/double bleb layer. Multivariate analyzes using Logistic regression models showed that a higher number of previous intraocular surgeries was significantly associated with the formation of a single bleb layer (relative risk [RR], 2.85; P = 0.0014; Table 3). Conversely, the other factors were not significantly associated with the formation of a single/double bleb layer.
Discussion
The primary outcome of the present study was the characteristics of the bleb fluid structure in eyes that had undergone tube-shunt surgery using the 350-mm2 endplate Baerveldt glaucoma implant. The bleb fluid structure was characterized into two classes based on T2-weighted MRI scans: a single bleb fluid layer adjacent to the outer surface of the endplate (single bleb layer group) and a bleb fluid layer on either side of the endplate (double bleb layer group). Three-dimensional images demonstrated that each bleb fluid structure has four holes, which correspond to the four fenestrations of the endplate. While no significant correlation was found between the bleb volume and postoperative IOP (r = −0.080; P = 0.57), the double bleb layer group had a significantly lower mean postoperative IOP than the single bleb layer group (12.3 ± 3.8 mmHg vs. 14.7 ± 4.1 mmHg, respectively; P = 0.033). A higher number of prior intraocular surgeries before the Baerveldt glaucoma implantation was significantly associated with the formation of a single bleb layer (RR = 2.85; P = 0.0014).
Three previous studies had characterized and quantified bleb structure after tube-shunt surgery. Detorakis et al.7 analyzed bleb images of four eyes with the FP-7 Ahmed glaucoma valve and four eyes with the single-plate Molteno implant. Sano et al.8 analyzed bleb images of eight eyes with the FP-7 Ahmed glaucoma valve, 16 eyes with the 350-mm2 endplate Baerveldt glaucoma implant, and three eyes with the 250-mm2 endplate Baerveldt glaucoma implant. Additionally, Ferreira et al.15 analyzed 11 eyes with the FP-7 Ahmed glaucoma valve using MRI less than 6 months after surgery; however, it is important to note that postoperative IOP fluctuates within the 6 months after tube-shunt surgery17. Our present study is unique because the bleb images were evaluated in 52 eyes that had undergone tube-shunt surgery using the 350-mm2 endplate Baerveldt glaucoma implant more than 6 months after surgery. Moreover, we obtained bleb fluid images with thinner slices (0.6 mm) using a 3 Tesla scanner in combination with FIESTA-C. Our MRI conditions offer more detailed bleb images than the images from previous reports.
MRI data from the present study allowed distinct classification of the eyes into the single bleb layer and double bleb layer groups. Previous reports have indicated the existence of single and double bleb layers. In MRI scans from a previous study of eyes with the Ahmed glaucoma valve or the Baerveldt glaucoma valve, all eyes had the double bleb layer8. However, in another study, in MRI scans of 11 eyes with the Ahmed glaucoma valve, five eyes had bleb fluid between the sclera and the endplate as well as bleb fluid adjacent to the outside of the endplate15. The results of the present study suggest that the bleb structure in eyes with the Baerveldt glaucoma implant is similar to that in eyes with the Ahmed glaucoma valve. In the present study, the bleb fluid in all eyes with a single bleb layer was located adjacent to the outside of the endplate. Aqueous humor exits through the silicone tube from the tube opening on the outside of the endplate. The implant design may explain why bleb formation was more frequently found outside rather than inside the endplate.
Each bleb fluid has four holes, which correspond to the four fenestrations in the endplate. The old type of endplate did not have fenestrations18. The fenestrations are designed to allow the growth of fibrous columns through the bleb fluid. These fibrous columns connect the sclera and the bleb roof through the plate. It is believed that this design prevents the formation of oversized blebs, which could result in diplopia caused by ocular motility disturbance4. In fact, in a recent study using the Baerveldt glaucoma implant19, the frequency of diplopia seemed to be markedly lower than that in a previous report20. The results of the present study suggest that fenestrations contribute to the formation of fibrous columns.
We did not find a significantly negative correlation between postoperative IOP and bleb volume (r = −0.080; P = 0.57). Two previous studies showed significantly negative correlations between postoperative IOP and bleb volume7, 8. In contrast, another study did not find a negative correlation between postoperative IOP and bleb volume in eyes with the Ahmed glaucoma valve; in fact, they found positive correlations between postoperative IOP and the bleb height and length15. Two previous studies included more than two types of endplates, and both studies demonstrated a stronger IOP-lowering effect of the Baerveldt glaucoma implant than that of the Ahmed glaucoma valve10, 11. The IOP-lowering effect of the double-plate Molteno implant, which is more effective than the single-plate Molteno implant9, is comparable to that of the 350-mm2 Baerveldt glaucoma valve21. Because the endplate sizes of the Ahmed glaucoma implant and the single-plate Molteno implant are smaller than that of the 350-mm2 Baerveldt glaucoma valve, endplate size may be a confounding factor that is involved in the significantly negative correlation between postoperative IOP and bleb volume.
Eyes with double bleb layers exhibit lower IOPs than eyes with single bleb layers. At present, we cannot determine the reason why double bleb layers contribute to lower IOPs. A possible explanation is that the space between the endplate and the sclera in the eyes with the single bleb layer may be filled with fibrous tissue. Intense fibrosis causes bleb failure in filtering surgery22,23,24. The relatively high IOPs in the eyes with single bleb layers may reflect intense postoperative fibrosis in the filtering bleb. In fact, we found that a higher number of prior intraocular surgeries is a prognostic factor for the single bleb layer. Intraoperative conjunctival incisions cause fibroblast activation in the subconjunctival tissue25, and repeated ocular surgery may cause active fibrosis in the subconjunctiva, resulting in high postoperative IOP because of the single bleb layer.
The present study has some limitations, which are mostly related to the cross-sectional nature of the study. First, the endplate position was not standardized. Twelve out of 52 eyes had the endplate in the superonasal, inferotemporal, or inferonasal quadrants. It is possible that the choice of the endplate position depended on conjunctival scarring caused by prior surgical incisions. Having the endplate in quadrants other than the superotemporal quadrant appears to be associated with severe conjunctival scarring. However, our multivariate analyzes showed that having the endplate positioned in the superotemporal quadrant is not an independent factor for the formation of the double bleb layer. Second, we cannot discern whether the double bleb layer changes to a single bleb layer postoperatively, whether the double bleb layer is formed from the single bleb layer postoperatively, or whether the two types of bleb structures do not change during the postoperative follow-up period. A longitudinal study for each patient is required to clarify this limitation.
In conclusion, Baerveldt glaucoma implantation results in either a layer of bleb fluid on each side of the endplate or one layer outside the endplate. Each bleb fluid structure has four holes, which correspond to the fenestrations of the endplate. Lower postoperative IOPs are associated with the formation of the double bleb layer, and the larger bleb is likely associated with the lower IOP seen in eyes with the double bleb layer. Repeated intraocular surgery may cause the formation of the double bleb layer to fail, resulting in a higher IOP after Baerveldt glaucoma implantation.
References
Arora, K. S., Robin, A. L., Corcoran, K. J., Corcoran, S. L. & Ramulu, P. Y. Use of various glaucoma surgeries and procedures in medicare beneficiaries from 1994 to 2012. Ophthalmology 122, 1615–1624 (2015).
Gedde, S. J. et al. Treatment outcomes in the tube versus trabeculectomy (TVT) study after five years of follow-up. Am J Ophthalmol 153, 789–803 (2012).
Fontana, H., Nouri-Mahdavi, K. & Caprioli, J. Trabeculectomy with mitomycin C in pseudophakic patients with open-angle glaucoma: outcomes and risk factors for failure. Am J Ophthalmol 141, 652–659 (2006).
Schwartz, K. S., Lee, R. K. & Gedde, S. J. Glaucoma drainage implants: a critical comparison of types. Curr Opin Ophthalmol 17, 181–189 (2006).
Danesh-Meyer, H. V., Spaeth, G. L. & Maus, M. Cosmetically significant proptosis following a tube shunt procedure. Arch Ophthalmol 120, 846–847 (2002).
Jeon, T. Y., Kim, H. J., Kim, S. T., Chung, T. Y. & Kee, C. MR imaging features of giant reservoir formation in the orbit: an unusual complication of Ahmed glaucoma valve implantation. AJNR Am J Neuroradiol 28, 1565–1566 (2007).
Detorakis, E. T. et al. Evaluation of the position and function of aqueous drainage implants with magnetic resonance imaging. J Glaucoma 18, 453–459 (2009).
Sano, I. et al. Assessment of filtration bleb and endplate positioning using magnetic resonance imaging in eyes implanted with long-tube glaucoma drainage devices. PloS one 10, e0144595 (2015).
Heuer, D. K., Lloyd, M. A. & Abrams, D. A. et al. Which is better? One or two? A randomized clinical trial of single-plate versus double-plate molteno implantation for glaucomas in aphakia and pseudophakia. Ophthalmology 99, 1512–1519 (1992).
Budenz, D. L., Barton, K. & Gedde, S. J. et al. Five-year treatment outcomes in the Ahmed Baerveldt comparison study. Ophthalmology 122, 308–316 (2015).
Christakis, P. G., Kalenak, J. W. & Tsai, J. C. et al. The Ahmed versus Baerveldt Study: Five-Year Treatment Outcomes. Ophthalmology 123, 2093–2102 (2016).
Ishida, K. et al. Comparison of polypropylene and silicone Ahmed Glaucoma Valves. Ophthalmology 113, 1320–1326 (2006).
Britt, M. T., LaBree, L. D. & Lloyd, M. A. et al. Randomized clinical trial of the 350-mm2 versus the 500-mm2 Baerveldt implant: longer term results: is bigger better? Ophthalmology 106, 2312–2318 (1999).
Allan, E. J., Khaimi, M. A., Jones, J. M., Ding, K. & Skuta, G. L. Long-term efficacy of the Baerveldt 250 mm2 compared with the Baerveldt 350 mm2 implant. Ophthalmology 122, 486–493 (2015).
Ferreira, J., Fernandes, F. & Patricio, M. et al. Magnetic resonance imaging study on Blebs Morphology of Ahmed Valves. J Curr Glaucoma Pract 9, 1–5 (2015).
Otsu, N. A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern 9, 62–66 (1979).
Gedde, S. J. et al. The tube versus trabeculectomy study: design and baseline characteristics of study patients. Am J Ophthalmol 140, 275–287 (2005).
Lloyd, M. A., Baerveldt, G., Heuer, D. K., Minckler, D. S. & Martone, J. F. Initial clinical experience with the baerveldt implant in complicated glaucomas. Ophthalmology 101, 640–650 (1994).
Rauscher, F. M. et al. Motility disturbances in the tube versus trabeculectomy study during the first year of follow-up. Am J Ophthalmol 147, 458–466 (2009).
Smith, S. L., Starita, R. J., Fellman, R. L. & Lynn, J. R. Early clinical experience with the Baerveldt 350-mm2 glaucoma implant and associated extraocular muscle imbalance. Ophthalmology 100, 914–918 (1993).
Smith, M. F., Doyle, J. W. & Sherwood, M. B. Comparison of the Baerveldt glaucoma implant with the double-plate Molteno drainage implant. Arch Ophthalmol 113, 444–447 (1995).
Georgoulas, S., Dahlmann-Noor, A., Brocchini, S. & Khaw, P. T. Modulation of wound healing during and after glaucoma surgery. Prog Brain Res 173, 237–254 (2008).
Yamanaka, O., Kitano-Izutani, A., Tomoyose, K. & Reinach, P. S. Pathobiology of wound healing after glaucoma filtration surgery. BMC Ophthalmol 15, 157 (2015).
Schlunck, G., Meyer-ter-Vehn, T., Klink, T. & Grehn, F. Conjunctival fibrosis following filtering glaucoma surgery. Exp Eye Res 142, 76–82 (2016).
Gozawa, M., Takamura, Y. & Miyake, S. et al. Prospective observational study of conjunctival scarring after phacoemulsification. Acta ophthalmologica 94, e541–e549 (2016).
Acknowledgements
This work was supported by a grant from Abbott Medical Optics, Japan.
Author information
Authors and Affiliations
Contributions
M.I. had full access to all the data in the present study and takes responsibility for the integrity of the data and the accuracy of the data analysis. K.I., H.K. and M.I. designed study concept. Acquisition, analysis and interpretation of data were performed by K.I., M.K., Yu.T., S.A., Yo,T., H.K. and M.I. K.I. and M.I. drafted the manuscript. All authors reviewed and approved the manuscript.
Corresponding author
Ethics declarations
Competing Interests
M.I. is responsible for submitting a competing financial interest statement on behalf of all authors of the paper. M.I.’s work has been partially funded by HOYA, Novartis, Pfizer, Alcon, Senju, Santen, Kowa, Otsuka, NIDEK, Wakamoto, and AMO. The other authors declare no potential conflict.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Iwasaki, K., Kanamoto, M., Takihara, Y. et al. Evaluation of Bleb Fluid After Baerveldt Glaucoma Implantation Using Magnetic Resonance Imaging. Sci Rep 7, 11345 (2017). https://doi.org/10.1038/s41598-017-11054-x
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-11054-x
This article is cited by
-
Long-term outcomes of Baerveldt glaucoma implant surgery in Japanese patients
Scientific Reports (2023)
-
Surgical safety and efficacy of a multi-stent system, the comet stent, in Baerveldt glaucoma implant surgery: a prospective pilot study
Japanese Journal of Ophthalmology (2023)
-
Prognostic factors for successful Baerveldt glaucoma implant surgery for refractory glaucoma after multiple surgeries
Japanese Journal of Ophthalmology (2021)
-
Ex-Press® versus Baerveldt implant surgery for primary open-angle glaucoma and pseudo-exfoliation glaucoma
International Ophthalmology (2021)
-
Evaluation of intraocular gas using magnetic resonance imaging after pars plana vitrectomy with gas tamponade for rhegmatogenous retinal detachment
Scientific Reports (2020)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.