Abstract
Congenital Hypertrophic Pyloric Stenosis (CHPS) is a disease condition that is caused as a result of pylorus wall hypertrophy and hyperplasia. In this study, we used color Doppler flow imaging (CDFI) and Doppler artifacts technique to observe the blood flow of hypertrophic pylorus tissue and the dynamic imaging of liquid passing through the pyloric canal in CHPS infants. 65 cases of CHPS infants and 50 infants without CHPS served as control group. We found that there were statistically significant differences between the blood flow grade of muscular layer and mucosal layer between CHPS and control infants, but no significant differences were observed in the same group. Doppler artifacts technique demonstrated the whole process of contrast agent flow through pyloric canal was directly observed in 35 of 65 subjects, and the internal diameter of the pyloric canal was 1.93 ± 0.33mm.Conclusion that CDFI combined with color Doppler artifacts technique was proved to be effective to observe the distribution feature and blood flow grade in each layer of pyloric canal in CHPS patients. This method provides the evidence for judging the degree of pyloric stenosis clinically, and furnishes the basis of therapy along with its clinical significance and good application value.
Similar content being viewed by others
Introduction
Congenital Hypertrophic Pyloric Stenosis (CHPS) is a disease condition that is caused as a result of pylorus wall hypertrophy and hyperplasia, and incomplete mechanical obstruction. This has been the most common etiology of vomiting in infants1. The incidence rate was approximately 2 to 5 per 1000 live births, with infants of affected parents at greater risk (2.5–20%). The risk is four times higher in boys than girls2, 3. The pyloric hypertrophies after birth causes progressive gastric outlet obstruction and most commonly observed in infants between 2 and 6 weeks of age4. The main clinical symptoms include frequent vomiting, usually 10 to 30 min after feeding, palpable in the middle of the abdomen and olive-shaped lumps, as well as feeding visible peristaltic waves5. For a better prognosis results, early diagnosis and early treatment are required. If it is not timely diagnosed and treated, it causes severe malnutrition and death6. The disease can be diagnosed mainly by imaging examination, the upper gastroenterography being the most successful with the longest history7. However, upper gastroenterography cannot directly display the hypertrophic pyloric and the damage caused by the gastroenterography to the baby cannot be ignored, the barium and aspiration risk8, 9.
In recent years, with the growing popularity of ultrasound and the development of ultrasonic technology, it directly reflects the situation of Helicobacter pylori, easy to measure the thickness of the muscle and the length of the tube, and has no damage, safe, simple and can be easily performed repeatedly. Therefore, at present, ultrasound examination has been an important screening tool for most of the children with clinical CHPS10,11,12.
Color Doppler flow imaging (CDFI) is a technique for examining blood flow using Doppler’s principle. Compared to other techniques, CDFI provides a noninvasive, real-time information flow signal of the lesion area. In the color sampling frame of CDFI, the color signal of the non-real blood flow is color Doppler artifacts image13. At present, the international literatures on ultrasonic examination of CHPS are mostly two-dimensional ultrasound images, few literatures reported that CDFI was used to analyze the blood flow signals of hypertrophic pyloric in CHPS14,15,16. In this study, the blood flow distribution characteristics of each layer of the mucosal layer during CHPS were observed and examined by CDFI, and provided objective information clinically. This helped the doctors to prevent the losses of large blood vessel during surgery, and in turn reduces the risk of surgery. Hemanz-schulman17 in his study graded the blood flow of pyloric hypertrophy and in this paper, we used this method to evaluate the abundant blood supply of pyloric hypertrophy, which also provides effective information clinically. At the same time we used color Doppler artifacts technology, which is more intuitive and dynamic to observe the whole process of liquid passage through the pyloric canal. Similarly, this in turn can make up with the two-dimensional ultrasound conditions for liquid echo and pyloric canal echo contrast which is not strong and can be easily missed. The purpose of this paper is to discuss how the application of CDFI combined artifacts imaging technique innovatively in the diagnosis of CHPS, and provide more objective basis for clinical evaluation of pyloric stenosis.
Results
Clinical manifestations between CHPS and control groups
There were 59 males and 6 females, aged 15 to 103 days in the CHPS group, with male prevalence ratio of 9.8:1,with a mean age of 42.48 days ±17.21 (SD), a mean weight of 3.77 kg ± 0.56 (SD). And there were 27 males and 23 females, aged 20 to 88 days in the control group, with male prevalence ratio of 1.2:1, with a mean age of 41.58 days ± 12.70 (SD), a mean weight of 4.90 kg ± 1.18 (SD) (Table 1). Compared with the control group (N = 50) infants who did not have CHPS, the differences in male prevalence ratio and the mean weight were significant.
Comparison of pyloric parameters between CHPS and control group
CHPS was present in 65 infants, who satisfied the ultrasound diagnostic criteria, the characteristics of the ultrasound showed that the longitudinal section of the pyloric canal was like a cervix, and the transverse section was like a target ring. Compared with the control group, the differences in muscular layer thickness, the length of pyloric canal and the diameter of pylorus were highly significant (P = 7.44E-44, 3.01E-73, 5.77E-29), and the thickness of mucosal layer remained significant (P = 6.34E-18) (Table 2).
The distribution feature of blood flow in the CHPS group
The distribution feature of blood flow in each layer of pyloric canal mucosa was observed clearly by CDFI examination in the CHPS group: Longitudinal section of pylorus showed that the blood flow in the pyloric muscle was thick and a short rod was arranged in parallel and perpendicular to the pyloric diameter, the blood flow in the serosal layer and mucosal layer was continuously stripped and was parallel to the pyloric canal diameter (Fig. 1a). On the transverse section of pylorus showed that the blood flow in the pyloric muscle was radiated and the blood flow direction was annular blood flow, but some segments could not be displayed because of parallel arrangement with sound beam (Fig. 1b). In the control group the pyloric canal mucosa of each layer was small with poor blood flow.
The difference of the grade of flow between CHPS and control group
In the control group, the mean grade of flow in the muscular layer was 1.58 ± 0.57 and in the mucosal layer 1.42 ± 1.50. In the CHPS group, the mean grade of flow in the muscular layer was 2.83 ± 0.38 and in the mucosal layer 2.91 ± 0.30. Among them, we identified 11 infants whose inflow was grade II and 54 infants inflow was grade III in the muscular layer, 6 infants showed an inflow of grade II and 59 infants with an inflow of grade III in the mucosal layer. The differences in the muscle and mucosal flow grade between the CHPS and control groups was highly significant (t value is 13.33, 18.77, both p < 0.01, Table 3).There was no difference observed in the blood flow between muscle and mucosal flow grade in both control (t value is 1.49, p > 0.05) as well as CHPS group (t value is 1.30, p > 0.05).
Time and frequency of liquid passage through the pyloric canal and the pyloric diameter
Contrast agent was injected through the stomach tube into the stomach to observe the flow condition in the pyloric canal (Fig. 2). By using color Doppler artifacts technique, the moving condition of contrast agent in the pyloric canal was observed clearly. When the liquid passed from the gastric antrum to the duodenum, color Doppler artifacts showed a mixed stack effect which was obvious and the signal was bright with a colorful flow signal (Fig. 3).
During 20 min observation by CDFI, 10 cases of 65 CHPS infants for the first through time was 5 min, 5–20 min in 25 cases, 30 cases had no obvious liquid through after 20 min. During the observation, liquid passing in 7 cases was observed for more than 3 times, 28 cases for 1–3 times. The mean pyloric diameter was about 1.98 ± 0.33 mm (Table 4).
Discussion
The structure of pylorus tissue was similar to the gastric wall, from inside to outside, and was divided into four layers, mucosa, submucosa, muscular layer and placenta percreta. The inferior pylorus artery, which mainly originated from the gastroduodenum and right gastroepiploic artery and from their bifurcation, was the main region of blood supply to the pylorus18. Currently, ultrasound examination of CHPS are mostly in two-dimensional state of observation, few literatures reported the demonstration of blood flow distribution characteristics of hypertrophic pyloric tissues in the CHPS using CDFI19. In our study, the blood flow distribution characteristics of each layer of CHPS were observed clearly, which showed that the blood in the pyloric muscular layer was thick and was short rod arranged in parallel on longitudinally while it was radiating on the transverse section, and perpendicular to the pyloric diameter. The direction of blood flow in the serous layer and submucosal layer was a continuous strip and the pyloric tube diameter was parallel to the pyloric diameter on the longitudinal section, which provided some objective information clinically, which avoided the loss of larger blood vessels in surgery in order to reduce the blood loss and the risk of infection after operation. It is of great significance for surgical treatment. This study found that compared to the control group, the difference of blood flow grade in muscular layer and mucosal layer was significant. But in CHPS group, the difference of blood flow grade in muscular layer and mucosal layer was no statistical significance. However, we found that 31 cases with blood flow of pyloric submucosal distribution was grade II, 34 cases was grade III; 36 cases of pyloric muscular layer flow was grade II and 29 cases was grade III, which showed that the blood flow in the submucosal layer was more abundant than that in the muscular layer. Mucosal layer is composed of connective tissues, blood vessels and nerves, which is full of blood vessels, so blood flow is rich, which can be explained from the normal morphological structure and physiological function20.
Our study used color Doppler imaging to observe the condition of contrast agents through the pyloric canal. Together with the experience of domestic and foreign scholars21, 22, color Doppler artifacts can be broadly divided into the following categories: Firstly, there is no color or few color signals were observed in the area of blood flow. Secondly, too many color signals were observed in the blood flow. Thirdly, color signals appeared in the non-blood flow area. Fourthly, color signals or its shades of color change caused the misunderstanding in the direction and velocity of blood flow. In our study, the color signals appeared in the non-blood flow regions. The principle behind this is that it was similar to that of the “flame sign” caused by the spray of the ureteral orifice. It is an artifact that was generated by the rapid flow of liquid, which may be related to the presence of micro materials in the flow of liquid causing ultrasound scattering23, 24. In the planar ultrasonic condition, it is sometimes difficult to catch the movement of contrast agents through the pyloric canal because of the contrast echoes produced by the liquid as well as pyloric canal were not strong, which cannot reflect the pyloric stenosis objectively and may be misleading clinically. Using color Doppler artifacts, we can dynamically observe the whole process of liquid passing through the pyloric canal more intuitively, the time and frequency of liquid that passes into the stomach through the pyloric canal, and also measure the pyloric length, which in turn provides an objective basis for the clinical evaluation of hypertrophic pyloric stenosis.
Ultrasound examination was performed based on the subjective and objective factors. Subjective factors refer to the operator’s skill and experience, objective factors refers to the resolution and sensitivity of the blood flow by the equipment. In order to reduce or avoid the influence of the above factors, we tried to let the experienced physicians operate with higher grade ultrasound instrument, and the same person to check on the same instrument to reduce the error. In addition, because of the small study, the statistical results might show possible errors which were different between the two groups in age and gender, and may be considered as random which were not good enough.
In summary, ultrasound along with color Doppler imaging technology not only diagnose CHPS but also examine the distribution of blood flow in different mucosal layers and blood flow grade accurately. Also, examine if the contrast agent can pass through the pyloric canal successfully and pass through the condition. This provides an objective basis for the clinical evaluation of hypertrophic pyloric stenosis, also a selection for the treatment of infants, and provides clinical evidence to avoid rich blood vessels further extending its significance.
Methods
Study design and patients
Sixty-five infants were examined due to clinical suspicion of CHPS and were prospectively recruited in the study. From August 2006 to December 2015, 65 infants diagnosed with CHPS by ultrasonic and upper gastroenterography and confirmed by gastroscopy in the neonatal department of Guangzhou No.1 People’s Hospital, China and all the cases were in accordance with the diagnostic criteria of CHPS25,26,27 were enrolled in the study. At present, the more generally accepted CHPS ultrasound diagnostic criteria include the muscular layer thickness of at least 3mm and/or the length of the tube with at least 17mm28,29,30. All the cases had typical vomiting, stomach type and gastric peristalsis, abdominal palpable pyloric olive-shaped lumps were included, and the exclusion criteria were pylorospasm and hiatal hernia in children associated with vomiting. There were 59 males and 6 females, aged 15 to 103 days, weighing 2.2 to 6.6 kg in the CHPS group. The control group consisted of 50 infants (27 males and 23 females), aged 20 to 88 days, weighing 3.7 to 10.8 kg, hospitalized in the same period for other diseases such as pneumonia, urological problems or abdominal hernias, with no other clinical manifestations of pyloric stenosis. The presence of CHPS was confirmed by gastroscope; the absence of CHPS was determined by reviewing the medical records. The protocol for this study was approved by the Institutional Review Board of Guangzhou First People’s Hospital. All the patient’s family members provided written informed consent. All methods were performed in accordance with the relevant guidelines and regulations.
Indicators and measurement
Ultrasonic instrument using Logiq company GE 9/E Vivid 9 ultrasound diagnostic instrument and Philips company IU22 ultrasound diagnostic instrument, the probe using high frequency linear array transducer and with a center frequency of 7.0~10.0 MHz and 7.0~12.0 MHz were used. Infants were fasted for 4 hours and gastric lavage was performed with tube reserve, by administering 10% chloral hydrate sedation (0.5 ml/kg). They are placed in supine position and scanning was performed in the longitudinal plane where the cardia and lower esophagus can be observed. Then the infant was turned towards left to observe the stomach cavity, stomach wall and the motility situation of the stomach. Then scanning was performed by turning the infant to the right, where the transverse of the pyloric canal can be found between the top of the right kidney and the lower of the gall bladder, turned at 90° to observe the long axis of the pyloric canal, the morphology and structure of the pylorus, and measure the thickness of the pyloric muscular, pylorus diameter and internal diameter, the length of the pyloric canal. The pylorus was then subsequently evaluated with color Doppler US. The presence of color signal in the muscular layer or mucosal layer was documented, and the number of signals (i.e. degree of flow) was classified into three-point scale17: Grade 1 meant no flow signal; grade 2 meant two to five signals (i.e. moderate flow); and grade 3 with extensive or continuous flow. Flow was confirmed with Doppler spectrum analysis. Then infuse the stomach with warm water or condensed milk as a comparison to the contrast agents, observed for 20 min, and then used the Doppler artifacts technique to examine the flowing condition of the contrast agents, and measure the diameter of pylorus.
Statistics
Data was expressed using frequencies, percentages or means and standard deviation (SD). The t test was used to compare different parameters of the control and the CHPS groups, the blood flow of the muscular layer and the mucosal layer between the two groups, the condition of the liquid passed through the pyloric canal by using the contrast agent alone and combined with Doppler artifacts image technology in the CHPS group. A P value < 0.05 was defined as statistically significant. Statistical analysis was performed using SPSS 13.0 software package for windows (SPSS Inc, IL, USA). All data generated or analysed during this study are included in this published article.
References
Peeters, B., Benninga, M. A. & Hennekam, R. C. Infantile hypertrophic pyloric stenosis-genetics and syndromes. Nature Reviews Gastroenterology & Hepatology. 9, 646–660 (2012).
Feng, Z. P. et al. The Clinical Features of Infantile Hypertrophic Pyloric Stenosis in Chinese Han Population: Analysis from 1998 to 2010. PLoS One. 9, 88–92 (2014).
George, E. M. III et al. Feasibility of Emergency Physician Diagnosis of Hypertrophic Pyloric Stenosis Using Point-of-care Ultrasound: A Multi-center case series. The Journal of Emergency Medicine. 37, 283–286 (2009).
AlMaramhy, H. H. Is There a Relation Between Pyloric Muscle Thickness and Clinical and Laboratory Data in Infants with Hypertrophic Pyloric Stenosis? Indian J Surg. 77, 827–830 (2015).
Feng, Z. Q. et al. Clinical features of congenital hypertrophic pyloric. Modern Digestion & Intervention. 15, 218–222 (2010).
Gotley, L. M. et al. Pyloric stenosis: a retrospective study of an Australian population. Emerg Med Australas. 21, 407–413 (2009).
Shaoul, R. Clinical presentation of pyloric stenosis: the change is in our hands. Isr. Med. Assoc. J. 6, 134–136 (2004).
Gupta, A. K. & Guglani, B. Imaging of congenital anomalies of the gastrointestinal tract. Indian J. Pediatr. 72, 403–414 (2005).
Doyle, D., O’neill, M. & Kelly, D. Changing trends in the mangement of infantile hypertrophic pyloric stenosis-an audit over 11 years. Ir. J. Med. Sci. 174, 33–35 (2005).
McVay, M. R. et al. Surgeon-performed ultrasound for diagnosis of pyloric stenosis is accurate, reproducible and clinically valuable. J. Pediatr Surg. 44, 169–171 (2009).
Mehboob, H. Sonographic Diagnosis of Infantile Hypertrophic Pyloric stenosis- Use of Simultaneous Grey-scale &Colour Doppler Examination. Int. J. Health Sci. 2, 134–140 (2008).
Sílvia, C. D. et al. Hypertrophic pyloric stenosis: tips and tricks for ultrasound diagnosis. Insights Imaging. 3, 247–250 (2012).
Jenssen, C. et al. Ultrasound artifacts and their diagnostic significance in internal medicine and gastroenterology - part 2: color and spectral Doppler artifacts. ŽGastroenterol. 54, 569–578 (2016).
Ayaz, Ü. Y., Döğen, M. E., Dilli, A., Ayaz, S. & Api, A. The use of ultrasonography in infantile hypertrophic pyloric stenosis: does the patient’s age and weight affect pyloric size and pyloric ratio? Med. Ultrason. 17, 28–33 (2015).
Khan, A. A., Yousaf, M. A. & Ashraf, M. Role of ultrasonography in early diagnosis of infantile hypertrophic pyloric stenosis. J. Ayub Med. Coll. Abbottabad. 26, 316–319 (2014).
Costa, D. S. et al. Hypertrophic pyloric stenosis: tips and tricks for ultrasound diagnosis. Insights Imaging. 3, 247–250 (2012).
Hemanz, S. M. et al. Hypertrophic Pyloric Stenosis in Infants: US Evaluation of Vascularity of the Pyloric Canal. Radiology. 229, 389–393 (2003).
Hemanz, S. M. Infantile Hypertrophic Pyloric Stenosis. Radiology. 227, 319–331 (2003).
He, J. B. et al. Value of Color Doppler Ultrasonography in Diagnosis of Congenital Hypertrophic Pyloric Stenosis. Journal Of Clinical Pediatric Surgery. 13, 182–184 (2004).
Christopher, M. S. et al. Association of prematurity with the development of infantile hypertrophic pyloric stenosis. Pediatric Research. 78, 218–222 (2015).
Prabhu, S. J. et al. Ultrasound artifacts: classification, applied physics with illustrations, and imaging appearances. Ultrasound Q. 30, 145–157 (2014).
Zhang, W. & Wang, J. R. Distinction and its significance of artifacts from color Doppler flow imaging. Chin. J. Ultrasonogr. 4, 303–306 (2004).
Yang, Y. H., Zhou, Z. G. & Xie, Y. M. Identification and application of color Doppler artifacts in liquid lesions. National Medical Frontiers of China. 5, 1467–1468 (2010).
Gao, H. G., Su, G. M. & Zhou, B. C. The value of color Doppler image artifacts in monitoring the localization of the ureteral catheter. Clin. J. Med.Offic. 40, 935–937 (2012).
Singh, U. K. & Kumar, R. Congenital hypertrophic pyloric stenosis. Indian J Pediatr. 69, 713–715 (2002).
Ranells, J. D., Carver, J. D. & Kirby, R. S. Infantile hypertrophic pyloric stenosis: epidemiology, genetics, and clinical update. Adv. Pediatr. 58, 195–206 (2011).
Udassin, R. New insights in infantile hypertrophic pyloric stenosis. Isr. Med. Assoc. J. 6, 160–161 (2004).
Jerzy, N. et al. Accuracy of sonographic criteria in the decision for surgical treatment in infantile hypertrophic pyloric stenosis. Arch. Med. Sci. 7, 508–511 (2011).
Maheshwari, P., Abograra, A. & Shamam, O. Sonographic evaluation of gastrointestinal obstruction in infants: a pictorial essay. J. Pediatr. Surg. 44, 2037–2042 (2009).
Meena, S. et al. Ultrasound Measurements in Hypertrophic Pyloric Stenosis: Don’t Let the Numbers Fool You. Perm. J. 16, 25–27 (2012).
Acknowledgements
The authors thank Professor Liu, doctoral tutor of Guangzhou First People’s Hospital, for timely assistance with formatting the numerous figures in this article. The color Doppler flow imaging (CDFI) and Doppler artifacts technique was performed at the Department of Medical Ultrasound, Guangzhou First People’s Hospital, Guangzhou Medical University. This work was supported by Guangdong Special Fund Project of Fundamental and Applied Research, China (Grant No. 2016A030313461) and supported by Guangdong Planned Project of Science and Technology (Grant No. 2014A020212014, 2014A020212374), China.
Author information
Authors and Affiliations
Contributions
S.M., J.L. conceived and designed the experiments and wrote the paper; S.M. performed the experiments and analyzed the data; S.M., Y.Y., H.J., X.M., H.W. contributed reagents/materials/analysis tools; J.L., Y.Z. provided therapy and clinical data. All authors gave final approval for publication.
Corresponding author
Ethics declarations
Competing Interests
The authors declare that they have no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ma, S., Liu, J., Zhang, Y. et al. Application of color Doppler ultrasound combined with Doppler imaging artifacts in the diagnosis and estimate of congenital hypertrophic pyloric stenosis. Sci Rep 7, 9527 (2017). https://doi.org/10.1038/s41598-017-10264-7
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-10264-7
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.