Abstract
The detection and quantification of protein biomarkers in interstitial fluid is hampered by challenges in its sampling and analysis. Here we report the use of a microneedle patch for fast in vivo sampling and on-needle quantification of target protein biomarkers in interstitial fluid. We used plasmonic fluor—an ultrabright fluorescent label—to improve the limit of detection of various interstitial fluid protein biomarkers by nearly 800-fold compared with conventional fluorophores, and a magnetic backing layer to implement conventional immunoassay procedures on the patch and thus improve measurement consistency. We used the microneedle patch in mice for minimally invasive evaluation of the efficiency of a cocaine vaccine, for longitudinal monitoring of the levels of inflammatory biomarkers, and for efficient sampling of the calvarial periosteum—a challenging site for biomarker detection—and the quantification of its levels of the matricellular protein periostin, which cannot be accurately inferred from blood or other systemic biofluids. Microneedle patches for the minimally invasive collection and analysis of biomarkers in interstitial fluid might facilitate point-of-care diagnostics and longitudinal monitoring.
Similar content being viewed by others
Main
Interstitial fluid (ISF)—compared with other peripheral biofluids such as saliva, sweat and tears—is a particularly rich source of soluble bioanalytes including proteins, peptides, metabolites and nucleic acids that exhibits close correlation with blood1,2,3,4,5,6. It also represents the locoregional biomolecular composition of specific tissues of interest, such as within the tumour microenvironment7. Simple and effective methods that enable comprehensive analysis of ISF can lead to transformative advances in biodiagnostic technologies that are not only minimally invasive and pain free, but also ideally suited for point-of-care and resource-limited settings8,9,10. Extraction of ISF followed by ex vivo analysis11 has not been widely embraced in preclinical or clinical applications, largely because of (1) difficulty in extracting ISF, which is time consuming and requires bulky instruments12,13,14 and (2) the small amount of ISF that can be extracted using current technology, making comprehensive analysis challenging15. For example, microneedle-assisted extraction of ISF yields about 2 μl of biofluid from 4 cm2 of human skin even after 20 min of vacuum suction, which is insufficient for comprehensive proteomic and metabolomic analysis15. Indeed, in preclinical settings (for example, when using small animal models), to measure the concentration of target biomarkers, it is common to pool ISF from multiple subjects to obtain sufficient material, which inevitably masks the subject-to-subject biological variability16.
In contrast to ISF extraction, microneedles functionalized with biorecognition elements can specifically capture target biomarkers in ISF, which can be followed by ex vivo analysis17,18. Direct exposure of microneedles to ISF allows the biorecognition elements on the microneedle to capture target biomarkers in situ, thus offering a promising technology for simple and efficient biodetection. However, physiological concentrations of the protein biomarkers in the ISF are usually lower compared to those in blood4,19. Moreover, analyte–antibody binding kinetics are deteriorated due to the dense tissue environment, which results in slower diffusion of target biomolecules to the sensor surface (that is, the microneedle surface), further lowering the probability of analyte capture and consequent signal intensity corresponding to the analyte. These challenges exacerbate the difficulty of detection of protein biomarkers in interstitial fluid. Despite the recent advances in multiplexed detection of biomarkers20, the sensitivities of existing microneedle-based analytical methods are insufficient to detect (or quantify) most ISF protein biomarkers, which limits the development potential for diagnostic tests based on ISF biomarker levels. Most previous reports are limited to mice that have been intravenously injected with high concentrations of recombinant target markers as pseudo models, or to biomolecules present at relatively high levels (micrograms per millilitre in blood)17. Finally, existing microneedle-based in vivo sampling and detection methods are limited to qualitative analysis in which the target biomarker concentration is represented as relative fluorescence intensity, absorbance value or normalized relative quantity18,20,21. This limitation precludes quantitative comparisons of the biomarker concentrations across different experiments and across different laboratories for biomedical research and decreases opportunities for standardization of the cut-off values for clinical biomarkers.
Here we demonstrate ultrasensitive and quantitative measurement of target protein biomarkers in ISF through microneedle-based in vivo sampling and subsequent on-needle analysis. To improve the sensitivity of our microneedle-based immunoassay, we used an ultrabright fluorescent nanolabel, termed plasmonic fluor, which improved the limit of-detection of various ISF protein biomarkers by nearly 800-fold compared with conventional fluorophores and markedly decreased the sampling time. Moreover, by harnessing the bilayer design of the microneedle, we were able to replicate conventional immunoassay procedures on microneedle patches, including a calibration curve based on ‘standard micropatches’. Using a series of mouse models, we demonstrate that the microneedle patch can be used for ultrasensitive and quantitative monitoring of various protein biomarkers using a simple stick-and-peel process. First, we probed the efficiency of a cocaine vaccine by monitoring cocaine-specific antibodies in dermal ISF. Second, we demonstrated sensitive detection and longitudinal monitoring of inflammatory biomarker levels in mice after induction of endotoxin-mediated shock. Last, we validated the application of the microneedle patch in the detection and quantification of the matricellular protein periostin (POSTN) in the calvarial periosteum—a challenging site for detection—using both control wild–type (WT) and periostin knockout (PostnKO) mice. The minimally invasive microneedle patch obviates the need for destruction of target tissues and repeated drawing of blood over a short period, which can cause poor patient compliance or potential death of experimental mice in preclinical settings. This ultrasensitive biodetection technology may enable the broad use of microneedle patches for biomarker capture and analysis.
Results
Design and fabrication of the microneedle patch
The biodetection technology introduced in this study relies on microneedles functionalized with biorecognition elements (such as antibodies) that penetrate the stratum corneum (or periosteum) and selectively capture protein biomarkers in the local ISF in a concentration-dependent manner. Subsequently, the microneedle patch is peeled off from the skin and the protein biomarkers bound on the microneedles are quantified by an ultrasensitive fluorescent immunoassay implemented ex vivo (Fig. 1a). For efficient capture of target biomarkers in vivo, microneedles are required to exhibit high protein–antibody binding ability, high mechanical strength and biocompatibility. Owing to its low cost, easy processability and hydrophobic nature, polystyrene is widely used for microtitre plates in biomedical research and clinical diagnostics. Affinity reagents such as capture antibodies and blocking proteins can be efficiently immobilized on the polystyrene surface owing to the hydrophobic interactions between polystyrene and the nonpolar residues on the proteins. Thus, we used polystyrene for the fabrication of the microneedles, which were subsequently coated with capture antibodies to enable specific binding of the target biomarkers.
The microneedle patch was fabricated using a silicone mould via two successive drop-casting steps (Fig. 1b). Polystyrene solution (25% w/v in dichloromethane) was first cast on a silicone mould and the solvent was allowed to evaporate slowly under ambient conditions. Subsequently, a backing layer comprised of the mixture of polystyrene and magnetic (Fe3O4) nanoparticles was formed on top of the pristine polystyrene layer. Incorporation of the magnetic nanoparticles in the backing layer is important to ensure that the microneedle patches stay at the bottom of a microtitre plate that is placed over a magnet during subsequent immunoassay procedures (Fig. 1b,c and Supplementary Fig. 1). This bilayer design overcomes the low throughput and poor reproducibility of patch-by-patch handling (that is, incubation and wash steps performed manually one patch at a time), making the microneedle-based assay highly consistent, fast and reproducible (Fig. 1d and Supplementary Fig. 2). The patch is comprised of an array of microneedles with a centre-to-centre distance of 600 μm (Fig. 1e). Each microneedle is conical in shape with a 4 μm radius of curvature at the tip, a diameter of 300 μm at the base and around 600 μm in height (Fig. 1e and Supplementary Fig. 3). To assess the density and uniformity of antibody coating on polystyrene microneedles, we coated the microneedles with biotinylated anti-mouse IgG followed by blocking with bovine serum albumin (BSA). Subsequently, the microneedle patches were exposed to dye (LT680)-labelled streptavidin, resulting in a strong and uniform fluorescent signal along the entire length of the microneedle, indicating the uniform coating of the antibodies on the polystyrene microneedle surface (Fig. 1f). By contrast, microneedles coated with BSA which was subsequently exposed to LT680–streptavidin exhibited extremely weak fluorescence intensity (155-fold lower, signal close to that of a pristine microneedle) (Supplementary Fig. 4), suggesting low auto-fluorescence of the BSA and polystyrene microneedle surface.
Plasmonic fluor-linked immunosorbent assay on the microneedle patch
Conventional sandwich enzyme-linked immunosorbent assay (ELISA) involves an enzymatic reaction that results in the formation of a soluble coloured product in an analyte concentration-dependent manner. While highly standardized and routinely implemented in microtitre plates comprised of identical sampling wells, this approach is unsuitable for the microneedle patches due to (1) the relatively low sensitivity stemming from the limited sampling surface area (analyte is present only on the micro-sized needles in real sampling situation), making the quantification of low-abundant analytes challenging; and (2) the soluble nature of the coloured product, which masks spatial variations in the amount of analyte bound across the patch, eliminating the possible spatial multiplexing capability. Therefore, existing approaches are limited to pseudo mouse models involving high amounts of target analytes, which do not represent their true pathological and physiological concentrations. To overcome these challenges, we used a fluorophore-linked immunosorbent assay (FLISA) that relies on plasmonic fluor as an ultrabright and highly specific fluorescent nanolabel. Plasmonic fluor comprises a gold nanorod (AuNR) coated with fluorophores (800CW) and a universal biological recognition element (for example, biotin)22. BSA is used as a scaffold to assemble all of these functional elements, as well as to resist non-specific binding (Fig. 2a and Supplementary Fig. 5; a detailed description of plasmonic fluor is provided in the Supplementary Information). Siloxane copolymer is used as a spacer layer between the AuNR and the fluorophores to avoid metal-induced fluorescence quenching23. Transmission electron microscopy (TEM) images of the plasmonic fluor confirmed the presence of a thin organic layer around the AuNRs (polymer and BSA conjugate) with an overall thickness of about 6.3 ± 1.1 nm (all values are expressed as mean ± s.d.) (Fig. 2b and Supplementary Fig. 6). Binding of plasmonic fluor–800CW to strepdavidin–800CW coated onto the bottom of a microtitre well resulted in a nearly 1,400-fold enhancement of the ensemble fluorescence intensity (Fig. 2c and Supplementary Fig. 7).
To test the applicability of plasmonic fluor as an ultrabright biolabel for use on a microneedle patch, we used mouse interleukin 6 (IL-6), a pro-inflammatory cytokine, as a representative protein biomarker. Conventional FLISA involves a standard sandwich immunoassay format, where the capture antibody is immobilized on the surface of the microneedle, recognition and capture of analyte (IL-6), binding of biotinylated detection antibody to the captured analyte, and exposure to streptavidin–fluorophore (800CW in this study), which binds to the biotin on the detection antibody with very high affinity. In contrast to conventional FLISA, plasmonic fluor-linked immunosorbent assay (p-FLISA) involves the use of plasmonic fluor instead of a conventional fluorophore as the fluorescent label. To determine the improvement in sensitivity and limit of-detection (LOD, defined as mean + 3σ of the blank) of p-FLISA compared with FLISA, serial dilutions of IL-6 of known concentration (2.5 ng ml−1 to 2.5 pg ml−1) were used as standards. Fluorescence signals obtained after applying the plasmonic fluor–800CW revealed nearly 530-fold enhancement in the ensemble fluorescence intensity compared with conventional FLISA at the highest IL-6 concentration tested (2.5 ng ml−1) (Fig. 2d,e). The LOD of the IL-6 p-FLISA was found to be around 0.33 pg ml−1, 790-fold lower than that of conventional FLISA (261 pg ml−1) (Fig. 2g,h). Scanning electron microscopy (SEM) images revealed the presence of plasmonic fluors on the surface of the microneedle (Fig. 2f, a few plasmonic fluors are highlighted by yellow arrows). Conversely, ELISA performed on the microneedle exhibited weak colorimetric signal and a large standard deviation (Fig. 2i). The optical density corresponding to the highest IL-6 concentration of ELISA on microneedle patch is nearly tenfold lower than the standard ELISA implemented on a microtitre plate, approaching the background noise level (Fig. 2 and Supplementary Fig. 8). The LOD of ELISA on the microneedle patch was measured at around 53 pg ml−1, 160-fold higher than that of p-FLISA (0.33 pg ml−1) (Fig. 2h,i).
Biophysicochemical properties of microneedle patches
Before deploying the microneedle patches for in vivo transdermal biodetection, we investigated their biophysicochemical properties, such as (1) mechanical strength for penetration of dermal tissue; (2) biocompatibility; and (3) biosafety and potential adverse effects. To determine whether the polystyrene microneedles possess sufficient mechanical strength to penetrate the skin under compression, we performed a micro-compression test on a microneedle patch comprised of an 11 × 11 array of microneedles. The microneedle patch tolerated compression force greater than 0.4 N per microneedle, which is sufficiently high to puncture the skin without causing the microneedles to mechanically yield24 (Fig. 3a). The microneedles penetrated the mouse skin, as evidenced by Trypan blue staining (Fig. 3b). Haematoxylin and eosin (H&E) staining of the extracted mouse skin tissue further confirmed that the microneedles had penetrated the stratum corneum and perforated into the epidermal layer (Fig. 3c and Supplementary Fig. 9). The penetration depth was around 140–170 μm (Supplementary Fig. 9). SEM images indicated that the microneedles maintained their conical shape and sharp tips after removal from the mouse skin, further confirming their mechanical integrity during administration (Supplementary Fig. 10). Furthermore, micro-indents on skin caused by the application of the microneedle patch quickly became invisible as the skin recovered to its normal state within 15 min (Supplementary Fig. 11). The rapid recovery of the skin reaffirms the minimally invasive nature of the procedure compared with conventional techniques such as blood collection.
Considering that the antibodies are immobilized on the microneedles, it is important to understand the stability of these proteins during the penetration of the microneedles into the dermal tissue, residence in the dermal tissue and subsequent withdrawal (Supplementary Fig. 12a). To test this, we coated microneedles with fluorescently labelled BSA (used as a model protein) before application onto mouse skin for different durations (15 s, 10 min, 20 min, 30 min and 60 min). Upon withdrawal, microneedles exhibited a loss of around 20% in fluorescence intensity, indicating that a small fraction of proteins had desorbed from their surface, possibly due to the shear forces between the microneedle surface and the epidermis (Supplementary Fig. 12b). The decrease in the fluorescence intensity (representing the amount of protein lost) did not depend on the residence time of the microneedles in the dermal tissue, indicating minimal loss from desorption or proteolytic degradation of these proteins in the dermal tissue, even for a residence time of 60 min.
To evaluate the biocompatibility of the polystyrene microneedle patches, human dermal fibroblasts (HDFs), one of the major cell types in the dermis layer, were cultured in the presence of microneedles and BSA-coated microneedles (since the in vivo sampling involves microneedles blocked with a dense layer of BSA). Notably, no change in cell viability was observed after either 1 h or 16 h of in situ culture with the microneedle patches (Fig. 3d). Systemic toxicity of the microneedle patch was also investigated by H&E staining. Negligible changes in cell states between the control group and the microneedle-administered group demonstrated the excellent biocompatibility of the microneedle (Fig. 3e). In addition, we measured the endotoxin level of the microneedle patch to investigate the possibility of an unwanted immune response after administration. Microneedles coated with BSA exhibited negligible endotoxin levels compared with the limit set by the US Food & Drug Administration (20 EU per device) (Supplementary Fig. 13 and Supplementary Table 1). Together, these findings indicate that the biofunctionalized microneedle patch represents a safe approach for in vivo biodetection.
Detection of cocaine-specific antibody in an immunized mouse model
Cocaine overdose and cocaine-use disorder, for which there is currently no FDA-approved medication, remain a global medical and social problem25,26. While vaccines do not directly address the underlying neurobiological mechanism behind cocaine-use disorder, cocaine vaccines that produce antibodies reduce the rate and quantity of drug entry into the brain and inhibit the psychoactive effects of the drug27. Unlike other vaccines that confer prolonged protection, currently designed antidrug vaccines require frequent boosting to maintain the effective antibody levels28,29. A recent study revealed that lack of pre-vaccination screening assays that predict the most effective vaccines or subjects amenable to vaccination is the major obstacle for clinical translation30. Therefore, technologies that enable rapid and reliable testing of hapten-specific antibody titres in vaccinated subjects may aid rational vaccine design and provide screening tools to predict vaccine clinical efficacy against drugs of abuse. Here, we set out to demonstrate a simple, rapid, and non-invasive method for evaluating the efficacy of the cocaine vaccine using a microneedle patch.
Mice were subcutaneously immunized and boosted with BSA–cocaine combined with adjuvants, lipopolysaccharide (LPS) and alum, as depicted in Fig. 4a. This vaccine was expected to result in the generation of two types of santibody: anti-BSA and anti-cocaine. As a standard way to determine antibody titre and immunization efficacy, serum collected on the 7th week was serially diluted and the concentrations of the two antibodies (anti-BSA and anti-cocaine) were tested by the ELISA using BSA and BSA–cocaine as recognition elements, respectively (Fig. 4b,c). Since BSA–cocaine is expected to bind with both types of santibody simultaneously, the difference in the antibody concentrations obtained using BSA–cocaine and BSA represents the concentration of anti-cocaine antibody. Mice with higher anti-BSA–cocaine titre than the anti-BSA titre are considered as vaccine responders (Fig. 4d and Supplementary Fig. 14). These responder mice were subjected to microneedle administration to test the ability of this method to detect the presence of anti-cocaine antibody in ISF.
To determine the shortest timescale for effective capture of antibodies, we administered the BSA-coated microneedle patch on the mouse dorsal skin for different durations. We found that 30 s of microneedle patch administration was sufficient to capture BSA-specific antibodies (Fig. 4e). Longer administration of the microneedle patch did not alter the fluorescence signals corresponding to the antibodies. To assess the site-to-site variations in the measured concentrations, we administered four microneedle patches coated with BSA on the ventral and on the dorsal skin of a responder mouse. The microneedle patches were subsequently probed with anti-mouse IgG and plasmonic fluor ex vivo. We noted very small differences (relative s.d. <8%) in the fluorescence intensity among the four microneedle patches administered on the same side of the mouse. However, the 4 microneedle patches administered on the ventral side exhibited 30–40% higher fluorescence intensity compared with those on the dorsal side (Fig. 4f). This difference possibly stems from the differences in the thickness of skin and blood flow between ventral and dorsal sides of the mice31,32. This observation underscores the importance of accounting for site-to-site variation and consistent administration of the microneedle patches to ensure reliable comparison across different subjects and experiments.
To validate the applicability of the plasmonic fluor-enhanced microneedle patch in detecting cocaine-specific antibodies in a minimally invasive manner, we used five unimmunized mice and five responder mice. Each mouse was administered with two microneedle patches coated with BSA and BSA–cocaine, and the patches were left on the skin for only 30 s. In the five responder mice, we observed high fluorescence signal and significant difference in the fluorescence intensity between the BSA and BSA–cocaine microneedle patches. However, in the case of the microneedle patches from five unimmunized mice, we noted a much lower fluorescence signal and negligible difference between the BSA and BSA–cocaine microneedle patches (Fig. 4g). Together, these results indicate that the plasmonic fluor-enhanced microneedle patch is a promising tool to determine and evaluate the vaccine response efficiency in a simple and non-invasive manner. In contrast to conventional blood drawing, the microneedle method enables frequent and easy monitoring of vaccine efficiency in individuals and can potentially accelerate vaccine development.
Detection and quantification of cytokines in a mouse model of endotoxin-induced shock
Next, we set out to demonstrate the detection of cytokines in dermal ISF in a mouse model of LPS-induced endotoxin shock using the microneedle patch. LPS, a pathogen-associated molecular pattern, triggers innate immunity and induces secretion of multiple pro-inflammatory cytokines (Fig. 5a). Measurement and monitoring of these pro-inflammatory cytokines in blood has been an established method for evaluating the ability of the immune system to mount an innate inflammatory immune response33.
BALB/C mice were injected intraperitoneally with LPS (1 μg g−1) to induce an acute phase response and systemic inflammation. Using functionalized microneedle patches, we measured the longitudinal concentrations of mouse IL-6 in ISF, which is known to increase in serum over 4 h after LPS administration34. Microneedle patches, pre-functionalized with IL-6-capture antibodies, were administered at different time points on mouse ventral skin and left undisturbed for 20 min (as indicated in Fig. 5b). After removing the microneedle patches from the skin, p-FLISA was performed ex vivo to measure the concentration of IL-6. Fluorescence intensity corresponding to the plasmonic fluors on the microneedle patches exhibited a gradual increase from 1 h to 4 h after LPS injection (Fig. 5c), whereas no significant change was observed in mice injected with saline (negative control group) (Fig. 5c and Supplementary Fig. 15). On the basis of the standard curve (obtained using microneedle patches exposed to known concentrations of IL-6), the concentrations of IL-6 in mouse ISF were determined to be equivalent to 2.6 ± 1.9, 12.3 ± 8.6, 120.4 ± 73.4 and 1271.9 ± 393.4 pg ml−1 at 0, 1, 2.5 and 4 h after LPS injection, respectively (Fig. 5d). Specificity tests also validated that other cytokines or chemokines, whose levels might also increase with LPS injection, do not interfere with the detection and quantification of IL-6 on microneedles (Supplementary Fig. 17 and Supplementary Table 2). The IL-6 concentration in sera collected 4 h after LPS or saline injection exhibited good qualitative correlation with that measured in ISF using the microneedle method at the same time point (Fig. 5e). However, the serum IL-6 concentration was found to be nearly 22-fold higher compared to that in the ISF. The absolute concentration of the protein biomarkers in ISF was lower than the concentration in blood, which can partly be ascribed to the difference between microneedle-based analyte sampling method and solution-based standard curve, as well as inherent variation of proteins in body fluids4. The p-FLISA standard curve, which is used for estimating the concentration of the analyte, was obtained by exposing the microneedle patches to known concentrations of IL-6 in standard dilution buffer. By contrast, microneedle-based ISF sampling occurs in a dense tissue matrix, resulting in slower diffusion kinetics and, consequently, a lower apparent concentration of the analyte. Nevertheless, the analyte concentration determined using the microneedle-based method also exhibited excellent qualitative agreement with the measured concentrations in serum samples (Fig. 5e and Supplementary Fig. 16).
Frequent and timely measurement of protein biomarkers is critical for disease monitoring and diagnostics in both biomedical research and clinical applications. Unfortunately, conventional longitudinal measurements require frequent blood draws in a short period, which may cause iatrogenic anaemia and increase morbidity of patients. Moreover, it is often impossible to repeatedly draw blood from small experimental animals, which can result in their death. The minimally invasive microneedle method represents a transformative approach to perform frequent, sensitive and accurate measurements of protein biomarkers in a longitudinal manner in the same mouse.
As opposed to in vitro diagnostics that involve sample acquisition (such as blood or urine) at a specific time point along the disease progression and analysis at a later point in time, in vivo diagnostics involving capture of target analytes from a dynamically varying matrix (for example, dermal ISF in the present case) is inherently a non-equilibrium condition. This is particularly true for the cases in which the target analyte concentration varies within the sampling timescales, for example, IL-6 level in LPS-stimulated mice. In such cases, the concentration determined using microneedle patches represents a time-average concentration of the analyte in ISF over the sampling period. From a diagnostic translation standpoint, this time-averaged concentration can be standardized by setting rigorous guidelines for microneedle administration (for example, administration time, location) and ex vivo analysis (for example, standard curve conditions).
Detection and quantification of endogenous matricellular protein in periosteum
The quantification of endogenous biomarkers at specific sites or inner tissues of interest is highly desirable for biomedical and clinical research. Conventional biofluids, such as blood, are not able to reflect local concentrations of biomarkers. Furthermore, such local detection and monitoring of relevant biomolecules remains challenging due to the difficulty in collecting an adequate amount of biofluid in a minimally invasive manner. While invasive procedures are sometimes inevitable for detection of protein biomarkers in inner tissues or organs, the microneedle-assisted technique will be able to minimize the damage to the target sites and allow repeated sampling. Here we hypothesized that the microneedle patch would efficiently sample and measure protein analytes at specific tissue or membrane locations with high sensitivity and specificity in a minimally destructive manner (Fig. 6a).
The periosteum is a fibro-cellular membrane, which covers the outer surface of bones and has key roles in bone growth, fracture healing and skeletal regeneration35,36. In rodents and humans, the average thickness of the periosteum is 40 μm and 100 μm for tibia and femur, respectively37,38. In our study, the average periosteal thickness in the mouse calvaria was 29 ± 15 μm (Fig. 6b). The extracellular matrix of the periosteum contains a high local concentration of the matricellular protein periostin—encoded by the Postn gene—which has important roles in bone regeneration and bone tumour metastasis39,40. In addition, a circulating isoform of periostin has been identified as a potential biomarker of bone density loss, osteoporosis, tumour metastasis and airway disease41,42. We set out to test the applicability of our microneedle patch for the detection and quantification of periostin in the calvarial periosteum and in dermal ISF.
We used PostnKO and WT mice as experimental and control groups, respectively. WT mice demonstrated high expression of periostin protein in the calvarial periosteum by immunohistochemistry, which was absent in PostnKO mice (Fig. 6b). To detect periostin, microneedles were pre-functionalized with anti-periostin antibodies and administered to the calvarial periosteum layer for 10 min (Fig. 6c). Simultaneously, another pre-functionalized patch was administered on the dorsal skin of the same mouse for 10 min. Subsequently, p-FLISA assay was performed ex vivo to measure the local concentrations of periostin. Fluorescence intensity corresponding to the plasmonic fluors on the microneedle patches (both on periosteum and dorsal skin) exhibited a strong signal from WT mice, whereas no significant signal was observed in PostnKO mice (Fig. 6d). On the basis of the standard curve, concentrations of periostin in periosteum and skin were around 2,000 pg ml−1 in WT mice, compared with a negligible amount in PostnKO mice.
Compared to conventional methods, the plasmonic fluor-enhanced microneedle method did not require extraction or isolation of periosteum during analysis, indicating that multi-round testing and long-term monitoring is achievable (Fig. 6a). Notably, the concentration of periostin in periosteum and skin measured by microneedle is nearly 450-fold and 330-fold lower, respectively, than the concentration in mouse serum (around 950 ng ml−1) (Fig. 6e). This phenomenon is possibly caused by binding and retention of periostin within the extracellular matrix, reducing the amount of available perostin that can be captured by the microneedle. Since the sampling area for skin and periosteum is different, absolute comparison of the values between these two tissues is precluded. It is also worth mentioning that blood periostin represents a systemic level of periostin while local periostin plays important roles in stem cell recruitment. Understanding its specific level is critical especially when there is a change introduced to the local environment of periosteum or bone. For example, if fracture occurred on the bone and the regeneration process has started, the local level of periostin at defined locations around this fracture can be revealed without substantially disturbing the regeneration process. In this way, it is easier to track the progress of bone regeneration and better understand how periostin contributes to this process by looking at changes in local periostin levels. The same concept could be applied in diverse fields to analyse desired target biomarkers and understand disease progression. In general, these results indicate that the microneedle patch can be used to evaluate the amount of an endogenous matricellular protein biomarker at specific tissue locations in a minimally invasive manner.
Discussion
In summary, we have designed a minimally invasive, ultrasensitive and quantitative biodetection technology based on a bilayered microneedle patch with plasmonic-enhanced fluoroimmunoassay. Using a series of mouse models, we demonstrate that this technology enables simple and timely detection of biomarkers of interest, including longitudinal monitoring of inflammatory immune responses, evaluation of vaccine efficiency in a minimally invasive manner, as well as quantification of localized changes in protein content within the tissue microenvironment. While largely retaining or enhancing the sensitivity and convenient workflow of conventional immunoassays, the plasmonic-enhanced microneedle method overcomes the need for tedious sample collection (for example, blood draw or ISF extraction), making it highly attractive for clinical diagnostics. Although we focused on the detection of antibodies, cytokines and matricellular proteins in this work, the biodetection method may be broadly applicable to a wide range of other diseases for which rapid and timely diagnosis and personalized treatment are effective interventions. For example, the diagnosis of myocardial infarction could be expedited by applying the microneedle patch (specific to cardiac troponin) in the ambulance as the patient is being moved to the emergency room after experiencing chest pain. Moreover, this technology could enable at-home sampling and centralized detection of biomarker levels through self-administration of microneedle patch, which can ultimately enable timely monitoring of therapy, such as certain antibiotic treatments, chemotherapy or checkpoint inhibitor immunotherapy. Therefore, this sensing technique can potentially be employed in point-of-care and resource-limited settings to facilitate rapid disease diagnosis and efficient therapeutic intervention in a more patient-friendly manner.
We note that various protein biomarker levels in the ISF are not well characterized and understood, which is an important bottleneck in the clinical translation of this technology4. Extensive efforts need to be dedicated to understanding the biomarker levels in ISF, their correlation with blood and, more importantly, the associated lag time correlated with blood, before this technology can be employed in clinical diagnostics. The technology demonstrated here serves as an efficient tool for biomedical and clinical researchers to fill this knowledge gap and propel the technology towards clinical applications.
Methods
Animals
All procedures have been approved by the Institutional Animal Care and Use Committee (IACUC) at Washington University in St Louis. Mice were housed in the housing facility at a constant temperature (21–23 °C) and humidity (45–50%) on a 12 h light–dark cycle (lights on 07:00–19:00 h), with food and water available ad libitum throughout the studies.
Synthesis of magnetic nanoparticles
Fe3O4 nanoparticles were synthesized via previously reported polyol synthesis method43. In brief, 0.15 g of iron (III) chloride hexahydrate (Sigma Aldrich, 236489) was dissolved in a mixture of 3 ml ethylene glycol (Sigma Aldrich, 102466) and 1 ml ethanolamine (Sigma Aldrich, 398136) to form a stable light-brown solution. After 30 min, 167.5 mg of polyethylene glycol (Sigma Aldrich, P3640) and 663.6 mg of sodium acetate trihydrate (Sigma Aldrich, S8625) was added under vigorous stirring. Subsequently, the solution was transferred to a Teflon-lined stainless-steel autoclave and heated to 200 °C for 8 h. The Fe3O4 nanoparticles were collected and washed three times with ethanol and water, and re-dispersed in dichloromethane for further use.
Fabrication of microneedles
Microneedles were prepared using silicone moulds with conical holes (Blueacre Technology). Each microneedle was 600 μm in length with a diameter of 300 μm at the base. The centre-to-centre spacing between the microneedles was 600 μm. Polystyrene–dichloromethane solution (25% w/v) was deposited on the silicone mould and left under vacuum at the bottom to create a negative pressure, followed by drying at room temperature for at least 4 h, allowing the evaporation of the solvent and the formation of the polystyrene microneedle. Fe3O4 nanoparticles dispersed in polystyrene–dichloromethane solution (final concentration: 15% w/v of polystyrene and 8 mg ml−1 Fe3O4 nanoparticles in dichloromethane) were deposited on the back of the polystyrene microneedle patch, followed by vacuum. After drying at room temperature for 4 h, the microneedle patch was carefully separated from the mould and tailored into small square pieces. The preparation and storage of microneedle patches were performed in sterile conditions.
Synthesis of plasmonic fluor
Synthesis of plasmonic fluor was performed according to a procedure described previously22.
Synthesis of AuNR
The wavelength of gold nanorods can be easily tuned to couple with dye molecules to achieve the best enhancement factor44. To prepare plasmonic fluor–800CW, AuNR-760 (localized surface-plasmon resonance wavelength ~760 nm) was prepared by the seed-mediated method45,46. To prepare seed solution, 0.6 ml of 10 mM ice-cold NaBH4 solution (Sigma Aldrich, 71321) was added into a solution containing 9.75 ml 0.1 M hexadecyltrimethylammonium bromide (CTAB) (Sigma Aldrich, H5882) and 0.25 ml 10 mM HAuCl4 (Sigma Aldrich, 520918) under vigorous stirring at room temperature for 10 min. The solution changed from yellow to brown, indicating the formation of Au seed. To synthesize gold nanorods, the growth solution was prepared by the sequential addition of 2 ml 0.01 M HAuCl4 aqueous solution, 38 ml 0.1 M CTAB, 0.55 ml 0.01 M AgNO3 (Sigma Aldrich, 20439 0), 0.8 ml 1 M HCI (Sigma Aldrich, H9892) and 0.22 ml 0.1 M ascorbic acid (Sigma Aldrich, A92902) followed by gentle homogenization. Subsequently, 5 µl of the seed solution was added into the growth solution and left undisturbed in dark for 24 h. AuNR solution was collected by centrifugation at 6,000 rpm for 40 min to remove the supernatant. AuNR was then re-dispersed into nanopure water for further use.
Conjugation procedures
BSA was first conjugated with biotin and 800CW sequentially through EDC–NHS chemistry. Specifically, 2 mg NHS–PEG4–biotin (Thermo Scientific, 21329) was added to 2.2 ml of 5 mg ml−1 BSA (Sigma Aldrich, A7030) in 1× PBS and incubated at room temperature for 1 h. BSA–biotin conjugation was purified by a desalting column (Thermo Scientific, 21329, 7,000 MWCO). Next, 800CW was conjugated to BSA–biotin. One-hundred microlitres of 1 M K2HPO4 buffer pH9 was added into 1 ml purified BSA–biotin solution to raise the pH. Next, 25 µl of 4 mg ml−1 NHS–800CW (LI-COR, 929–70020) was added to the mixture, and the solution was incubated at room temperature for 2.5 h. BSA–biotin–800CW was purified using a Zeba desalting column pre-equilibrated with nanopure water.
Synthesis of plasmonic fluor
To prepare plasmonic fluor–800CW, AuNR (wavelength around 760 nm) was employed as the nanoantenna. One microlitre (3-mercaptopropyl)trimethoxysilane (MPTMS) (Sigma Aldrich, 175617) was added to 1 ml AuNR (extinction ~2) and the mixture was shaken on rocking bed for 1 h. Subsequently, MPTMS-modified AuNR was collected by centrifugation at 6,000 rpm for 10 min and was further mixed with 2 µl APTMS (Sigma Aldrich, 281778) and 2 µl trimethoxy(propyl)silane (TMPS) (Sigma Aldrich, 662275) to form the polymer spacer layer. Finally, AuNR–polymer solution was collected by two centrifugations at 6,000 rpm for 10 min to remove the free monomer and concentrated into a final volume of 10 µl.
Next, BSA–biotin–800CW conjugate was coated around AuNR–polymer modified using a previously reported method47. Specifically, pH of 100 µl 4 mg ml−1 BSA–biotin–800CW was first lowered by adding 1 µl of 20 mg ml−1 citric acid (Alfa Aesar, 36664). Subsequently, concentrated AuNR–polymer solution was added into BSA–biotin–800CW solution and sonicated for 20 min under dark condition. The coated nanostructures were then collected by centrifugation at 5,000 rpm for 5 min and subsequently incubated with 0.5 ml of 0.4 mg ml−1 BSA–biotin–800CW (pH 10) for 3 d at 4 °C. Finally, the nanostructures were washed 4 times using alkaline nanopure water (pH 10) by centrifugation at 6,000 rpm and re-dispersed in 1% BSA in 1× PBS solution for further use.
Fluorescence enhancement using plasmonic fluor
The test procedure is schematically illustrated in Supplementary Fig. 7. Specifically, BSA–biotin was first immobilized on a 96-well plate by incubating the well with 50 ng ml−1 BSA–biotin in 1× PBS at room temperature for 15 min. The plate was washed by three times using PBST (0.05% Tween 20 in 1× PBS) and then blocked using Odyssey Blocking Buffer (PBS) (LI-COR, 927-40100). Streptavidin–800CW (1 µg ml−1) was subsequently added and incubated for 10 min. Next, the plate was washed three times using PBST and then incubated with ~76 pM plasmonic fluor–800CW (in 1% BSA). Finally, after washing, 200 µl of PBST was added to each well and the fluorescence signals before and after the addition of plasmonic fluor were recorded using a LI-COR CLX fluorescence imager with the following scanning parameters: laser power ~L2; resolution ~169 μm; channel: 800; height: 4 mm.
Material characterization
TEM images were obtained using a JEOL JEM-2100F field emission instrument. A drop of aqueous solution was dried on a carbon-coated grid, which had been made hydrophilic by glow discharge. SEM images were obtained using a FEI Nova 2300 field‐emission SEM at an acceleration voltage of 10 kV. The extinction spectra of plasmonic nanostructures were obtained using a Shimadzu UV-1800 spectrophotometer. Fluorescence mappings were recorded using LI-COR Odyssey CLx imaging system. The X-ray diffraction (XRD) patterns of the Fe3O4 nanoparticles were obtained using a Bruker D8-Advance X-ray powder diffractometer using Cu Kα radiation (λ = 1.5406 Å) over the 2θ range 10°–90°.
Mechanical test
The mechanical properties of the microneedle patch were measured with a displacement-force test station (Instron 5583 Electro-mechanical Universal Testing Machine) (Fig. 3a). A microneedle patch was attached to a rigid platform with microneedles facing up. The sensor probe was brought in contact with the microneedles in the vertical direction at a speed of 0.1 mm s−1. The initial distance between the sensor and microneedle tips was set to be 1 cm. Displacement-force measurements were acquired from the point at which the sensor first touched the microneedle tips to the onset of buckling of the microneedles.
Skin penetration efficiency test
Mouse skin after administration of microneedle patch was imaged by digital camera to assess skin penetration efficiency. After administration of microneedle patch for 15 min, mouse ventral skin was stained with trypan blue for 20 min. The mouse was subsequently euthanized, and the skin sample was imaged by digital camera after wiping remaining dye from the skin. In a separate experiment, after 15 min of administration of microneedle patch, the mouse was euthanized, and its ventral skin was isolated carefully. These excised skins were washed with deionized water and then fixed with 4% formalin solution for H&E staining. Histological examination was conducted by an optical microscope under bright-field illumination (Biotek Lionheart FX).
Protein retention test
To access stability of protein bound on the microneedles, the microneedles were coated with BSA–CW800. After washing with PBST and drying, fluorescence images were recorded using LI-COR Odyssey CLx imaging system. Subsequently, microneedles were administered on mouse dorsal skin and left for different periods of time, varying from 15 s to 60 min. Fluorescence images were recorded again using the same parameters in LI-COR Odyssey CLx imaging system. The fluorescence intensity before and after administration was compared to assess the retained fraction of the protein on the microneedle surface.
Assessing in vitro biocompatibility of microneedle patch
To evaluate toxicity of microneedles in vitro, HDFs were selected as a model system. Uncoated microneedles and BSA-coated microneedles were incubated in cell culture medium for 1 h and 16 h, representing short-term and potentially long-term contact with skin tissues. After removing the microneedle patch, HDFs were cultured in the medium for 24 h. Cell viability was quantified using the MTS assay.
Assessing systemic toxicity of microneedle patch
For toxicity assessment in vivo, mice administered with BSA-coated microneedle patch were selected as the treatment group, while mice without administration were used as control group. The representative organs including heart, lung, liver, kidney and spleen in each group were collected and fixed in 4% neutral buffered formalin for histological analysis, which were subsequently processed by The Musculoskeletal Histology and Morphometry Core at Washington University in St Louis for paraffin embedding, sectioning and H&E staining. Histological examination was performed with an optical microscope under bright-field illumination (Biotek Lionheart FX).
Bacterial endotoxin level of microneedle patch
To evaluate bacterial endotoxin level on microneedle surface, especially after blocking with BSA, microneedle patches coated with BSA were first incubated with endotoxin-free water overnight. Subsequently, the bacterial endotoxin level in the water was detected and measured by Kinetic Chromogenic Limulus Amebocyte Lysate (LAL) assay (Thermo Scientific, 88282).
Mouse IL-6 ELISA on microneedle and on microtitre plate
Mouse IL-6 DuoSet ELISA kit (R&D systems, DY406, lot number P195781) was used in the study. Microneedle patches were first placed in a 24-well plate (Corning, 3526) which was clamped on a magnetic plate separator (Luminex Corporation, CN-0269-01) designed to hold the microplate. Microneedle patches were immersed and incubated with 1 ml of capture antibodies (2 µg ml−1 in PBS; R&D Systems, 840171) overnight at room temperature, followed by washing and blocking with 2 ml reagent diluent (1× PBS containing 3% BSA, 0.2 μm filtered). During incubation and washing steps, microneedle patches were tightly attached to the bottom of well. After three washes with PBST, 1 ml of serial diluted standard samples (R&D systems, 840173) were added to different wells and microneedle patches were incubated at room temperature for 2 h. Subsequently, patches were washed and incubated with biotinylated detection antibodies (R&D Systems, 840172, 75 ng ml−1 in reagent diluent) for 2 h, washed again with PBST, and incubated with HRP-labelled streptavidin (R&D Systems, 893975; 40-fold dilution using reagent diluent) for 20 min. One millilitre of substrate solution (1:1 mixture of colour reagent A (H2O2) and colour reagent B (tetramethylbenzidine) (R&D Systems, DY999)) was added to each well and the reaction was stopped by adding 500 µl H2SO4 (2 N) (R&D Systems, DY994) after 20 min. Optical density of each well was determined immediately using a microplate reader set to 450 nm, after removing microneedle patch from each well. IL-6 ELISA on 96-well plate was implemented in the same manner.
Mouse IL-6 FLISA and p-FLISA on microneedle
Mouse IL-6 FLISA was implemented adopting the similar approach as the ELISA described above, except that HRP-labelled streptavidin was replaced with 800CW-labelled streptavidin (LI-COR, 926-32230; 20 ng ml−1 for 20 min). The patches were washed three times each with PBST followed by nanopure water. In case of p-FLISA, 1 ml plasmonic fluor–800CW was added subsequently (extinction ~0.5), incubated for 30 min, and the patches were washed 3 times each with reagent diluent followed by PBST. Patches were imaged using a LI-COR CLx fluorescence imager with the following scanning parameters: laser power ~L2; resolution ~21 μm; channel: 800; height: 0 mm.
Detection of cocaine-specific antibody in an immunized mouse model
Animals and cocaine immunization
Male mice (C57BL/6, Jackson Laboratories 000664), at the age of 5–6 weeks, were purchased from Jackson Laboratories. Mice were housed 4 per cage and allowed to acclimate for 7 d before the experiment in the housing facility. Eleven mice were randomly divided into two groups, vaccination group and control group. For each mouse in vaccination group, cocaine vaccination solution including 100 μg cocaine–BSA conjugate (Fitzgerald Industries, 80-1037) and 2 μg LPS (InvivoGen, vac-3pelps) in 100 μl saline mixed with 100 μl 2% Alhydrogel (InvivoGen, vac-alu-250) was freshly prepared before vaccination. Each mouse was first vaccinated subcutaneously with a total of 200 μl of cocaine vaccine solution on 4 injection sites and boosted with half the total dosage on 2 sites at day 14, 21 and 28, both on the dorsal side. Two weeks after the last boost, blood from both vaccinated and control group was collected via the submandibular vein and the serum was stored at −20 °C.
Detection of anti-cocaine antibody in mouse serum
Ninety-six-well ELISA plates (Thermo Scientific, 15041) were coated with 1 μg ml−1 BSA or cocaine–BSA conjugate in PBS at 4 °C overnight and blocked with 300 μl of Odyssey Blocking Buffer for 1 h. Serum was diluted in PBST at a range of dilution in duplicates and applied on both BSA and cocaine–BSA conjugate coated wells for 1 h at room temperature. After washing with PBST, plate was incubated with a biotin-labelled donkey anti-mouse antibody (R&D systems, BAF018; 1:2,000 in 1% BSA–PBST, 100 μl per well) for another 1 h, followed by HRP-labelled streptavidin (R&D systems, 893975; 40-fold dilution) for 20 min. One-hundred microlitres of substrate solution (1:1 mixture of colour reagent A (H2O2) and colour reagent B (tetramethylbenzidine) (R&D Systems, DY999)) was added to each well and the reaction was stopped with 50 µl H2SO4 (2 N) (R&D Systems, DY994) and absorbance was measured at 450 nm. Within the vaccination group, anti-cocaine–BSA titre of each mouse was defined as three mean + 3σ of the anti-BSA titre. Mice with higher anti-cocaine–BSA antibody titre than the anti-BSA titre were defined as ‘vaccine responders’.
In situ sampling and quantification of cocaine-specific antibody in mouse dermal ISF through microneedles
Studying effect of administration time on antibody detection
Microneedle patches were incubated with BSA (1 μg ml−1 in PBS) in a 24-well plate at room temperature overnight, followed by washing and blocking with Odyssey Blocking Buffer for 1 h. After three washes with PBST and blow drying with nitrogen gas, microneedle patches were administered on the dorsal skin of the mouse (vaccine responder) under anaesthesia. The period of administration varied from 30 s to 5 min. Subsequently, patches were washed with PBST and blocked with Odyssey Blocking Buffer for another 30 min, followed by incubation with biotin-labelled donkey anti-mouse antibody (1:2,000 in 1% BSA–PBST) for 1 h and 800CW-labelled streptavidin (20 ng ml−1) for 20 min. One millilitre of plasmonic fluor–800CW (extinction ~0.5) was subsequently added, incubated for 30 min, and the patches were washed 3 times each with PBST. Patches were imaged with a LI-COR CLx fluorescence imager with the following scanning parameters: laser power ~L2; resolution ~21 μm; channel: 800; height: 0 mm.
Studying the effect of administration location on antibody detection
Eight microneedle patches were incubated with BSA (1 μg ml−1 in PBS) in a 24-well plate at room temperature overnight, followed by washing and blocking with Odyssey Blocking Buffer for 1 h. After three washes with PBST and drying under a stream of nitrogen gas, four microneedle patches were administered on dorsal skin and another four patches were administered on ventral skin of mice for 30 s, simultaneously. Subsequently, patches were washed with PBST and blocked with Odyssey Blocking Buffer for another 30 min, followed by incubation with biotin-labelled donkey anti-mouse antibody (1:2,000 in 1% BSA–PBST) for 1 h and 800CW-labelled streptavidin (20 ng ml−1) for 20 min. One millilitre plasmonic fluor–800CW (extinction ~0.5) was added subsequently, incubated for 30 min, and the patches were washed 3 times each with PBST. Patches were imaged with a LI-COR CLx fluorescence imager with the following scanning parameters: laser power ~L2; resolution ~21 μm; channel: 800; height: 0 mm. The intensity of fluorescence signal represents the amount of anti-BSA antibody binding to the BSA on microneedle surface.
Detection of anti-cocaine antibody through microneedles
Microneedle patches were incubated with BSA or BSA–cocaine conjugate (1 μg ml−1 in PBS) in a 24-well plate at room temperature overnight, followed by washing and blocking with Odyssey Blocking Buffer for 1 h. After washing and drying, two microneedle patches coated with BSA and BSA–cocaine conjugate were simultaneously administered on dorsal skin of mouse under anaesthesia for 30 s. Subsequently, microneedle patches were washed and blocked, followed by incubation with anti-mouse antibody for 1 h and 800CW-labelled streptavidin for 20 min. One millilitre plasmonic fluor–800CW was subsequently added (extinction ~0.5) and incubated for 1 h. Patches were imaged with a LI-COR CLx fluorescence imager using the same scanning parameters as stated above.
Detection and quantification of cytokines in a mouse model of endotoxin shock
Animals and induction of endotoxin shock
Female mice (BALB/C, Jackson Laboratories, 000651), at the age of 5–6 weeks, were purchased from Jackson Laboratories. Mice were housed 3 per cage and allowed to acclimate for 7 days in the housing facility before the microneedle experiment. To induce endotoxin shock, mice received intraperitoneal injection of LPS (InvivoGen, vac-3pelps; 1 mg kg−1), whereas control group mice received intraperitoneal injections of saline solution.
In situ sampling and quantification of mouse IL-6 in dermal ISF through plasmonic fluor-enhanced microneedles
To sample IL-6 in mouse dermal ISF, microneedle patches were pre-functionalized with IL-6 capture antibody, followed by washing with PBST and blocking with reagent diluent (1× PBS containing 3% BSA, 0.2 μm filtered) for 1 h. Microneedle patches were administered on mice ventral skin under anaesthesia for 20 min to obtain baseline IL-6 measurements. Subsequently, mice were injected with LPS and microneedle patches were administered on mice at 1, 2.5 and 4 h after injection in a successive manner. Following 30 min of blocking with reagent diluent (1× PBS containing 3% BSA, 0.2 μm filtered), microneedle patches were exposed to biotinylated detection antibody, streptavidin–800CW and plasmonic fluor as described above. Fluorescence images of microneedle patches were obtained using a LI-COR CLx fluorescence imager using the same scanning parameters described above. Blood was collected via the submandibular vein immediately after the removal of microneedle patch at last time point (4 h after LPS injection) and serum was stored at −20 °C for further analysis. IL-6 ELISA was performed as described above in a 96-well microtitre plate to determine the concentration of IL-6 in mouse serum.
To investigate the relationship between serum and ISF at early time points, we used another three groups of mice (randomly divided from 9 mice), administered microneedle patches and collected blood at earlier time points, namely before LPS injection, and 1 and 2.5 h after LPS injection, respectively. Blood was collected via the submandibular vein right after the removal of microneedle patch at each time point and serum was stored at −20 °C for further analysis. Methods and reagents to determine the concentration of IL-6 in ISF and serum are identical to those mentioned in the previous section. On the basis of the standard curve, concentration of IL-6 in mouse dermal ISF exhibited a gradual increase from before the LPS injection to 2.5 h after LPS injection. The IL-6 concentration in sera also exhibited a similar increase in concentration from before LPS injection to 2.5 h after LPS injection, in agreement with that observed using the microneedle method (Supplementary Fig. 16).
Specificity of the IL-6 test
It is possible that the fluorescence signals on microneedles are generated by other cytokines or chemokines whose concentration also increase after LPS injection. To investigate the specificity of the test, we incubated different concentrations of human IL-6, mouse IFN-γ, mouse TNF, mouse IL-12, mouse CXCL9, mouse CCL19 and mouse CCL20 with mouse IL-6 capture antibody, followed by mouse IL-6 detection antibody and HRP or plasmonic fluor (concentration and source of each protein were listed in Supplementary Table 2). For conventional plate-based ELISA, p-FLISA and microneedle-based p-FLISA, assay methods and reagents were identical to those mentioned above.
ELISA (Supplementary Fig. 17a) and p-FLISA (Supplementary Fig. 17b) implemented in microtitre plates indicate that other mouse cytokines or chemokines and IL-6 from other species do not interfere with the quantification of mouse IL-6. Different concentrations of these interfering proteins exhibited signals similar to those of the blank. In addition, p-FLISA performed on microneedles (Supplementary Fig. 17c) also exhibited weak signals (similar to or lower than the blank) corresponding to interference proteins. Note that the y axis in Supplementary Fig. 17c uses a logarithmic scale.
Detection and quantification of endogenous matricellular protein periostin in periosteum and skin
Animals
PostnKO mice and WT littermate controls on a mixed B6;129 background (Postntm1Jmol, Jackson Laboratories 009067) were a gift from M. F. Rai (Department of Orthopedic Surgery, Washington University in St Louis). Two WT male and two PostnKO male mice were used for the plasmonic fluor-enhanced microneedle detection of periostin at the age of 12 weeks.
Histology and immunohistochemistry
The calvaria of WT and PostnKO mice were processed by the Musculoskeletal Histology and Morphometry Core at Washington University in St Louis for paraffin embedding, sectioning and H&E staining. Unstained tissue slides were acquired from the core for periostin immunostaining. In brief, antigen retrieval was performed in sodium citrate buffer (10 mM sodium citrate, 0.05% Tween 20, pH 6.0) at 50–55 °C overnight. The tissues were then permeabilized in 0.2% Triton X-100 buffer (Sigma Aldrich 9002–93–1) for 10 min, blocked in 2.5% normal horse serum (IMMPRESS HRP Anti-Rabbit IgG kit, Vector Laboratories, MP-7401) for 1 h at room temperature, and incubated with anti-periostin antibody (Sigma Aldrich, HPA012306) with a dilution rate of 1:200 at 4 °C overnight. The endogenous peroxidase was quenched in 0.3% H2O2 for 30 min and the sections were incubated with biotinylated secondary antibody (IMMPRESS HRP Anti-Rabbit IgG kit, Vector Laboratories, MP-7401-15) for 30 min at room temperature and the stain was developed by incubating with metal enhanced DAB substrate (IMMPACT DAB kit, Vector Laboratories, SK-4105). Nuclei were counterstained with haematoxylin (Ricca chemical 3536-16). All washes between steps were performed in either double-distilled water or TNT buffer (0.1 M Tris-HCl, 0.15 M NaCl, 0.05 % Tween).
In situ sampling and quantification of periostin in mouse periosteum and skin through plasmonic fluor-enhanced microneedle
Mouse periostin Duoset ELISA kit (R&D Systems, DY2955, lot number P217047) was used. To sample periostin in mouse periosteum and dermal ISF, microneedle patches were pre-functionalized with periostin capture antibody (R&D Systems, 842318) in a 24-well plate clamped on a magnetic plate separator, followed by washing with PBST and blocking with reagent diluent (1× PBS containing 3% BSA, 0.2 μm filtered). Skin above periosteum the layer was carefully incised using scissors. A microneedle patch was administered on periosteum and another patch was administered on dorsal skin for 10 min, under anaesthesia. This was followed by another 30 min of blocking with reagent diluent (1× PBS containing 3% BSA, 0.2 μm filtered), microneedle patches were exposed to biotinylated detection antibody (R&D Systems, 842319), streptavidin–800CW and plasmonic fluor. Fluorescence maps of microneedle patches were imaged using a LI-COR CLx fluorescence imager using the same scanning parameters as described above. Blood was collected by cardiac puncture and serum was stored at −20 °C for further analysis. Periostin ELISA for mouse serum were implemented in similar approach on 96-well microtitre plate.
Statistics
To analyse the statistical difference between two groups, unpaired one-tailed t-test with Welch’s correction was used. To analyse the statistical difference between each data point in two groups, two-way ANOVA with Sidak’s multiple-comparison test was used. For analysing the statistical difference between two or more groups one-way ANOVA with Tukey’s multiple-comparison test was used. Statistical significance of the data was calculated at 95% (P < 0.05) confidence intervals. All values are expressed as mean ± s.d. GraphPad Prism 8 was used for all statistical analysis. We used four-parameter logistic or polynomial fit to calculate the LOD in the standard curves of bioassays. The LOD is defined as the analyte concentration corresponding to the mean fluorescence intensity of blank plus 3σ. Origin 2016 was used to calculate the LOD.
Reporting Summary
Further information on research design is available in the Nature Research Reporting Summary linked to this article.
Data availability
The main data supporting the results in this study are available within the paper and its Supplementary Information. The raw data are available from figshare at https://doi.org/10.6084/m9.figshare.13331366.
References
Kool, J. et al. Suction blister fluid as potential body fluid for biomarker proteins. Proteomics 7, 3638–3650 (2007).
Müller, A. C. et al. A comparative proteomic study of human skin suction blister fluid from healthy individuals using immunodepletion and iTRAQ labeling. J. Proteome Res. 11, 3715–3727 (2012).
Tran, B. Q. et al. Proteomic characterization of dermal interstitial fluid extracted using a novel microneedle-assisted technique. J. Proteome Res. 17, 479–485 (2018).
Heikenfeld, J. et al. Accessing analytes in biofluids for peripheral biochemical monitoring. Nat. Biotechnol. 37, 407–419 (2019).
He, R. et al. A hydrogel microneedle patch for point-of-care testing based on skin interstitial fluid. Adv. Healthc. Mater. 9, 1901201 (2020).
Wang, Z. et al. Transdermal colorimetric patch for hyperglycemia sensing in diabetic mice. Biomaterials 237, 119782 (2020).
Gromov, P. et al. Tumor interstitial fluid—a treasure trove of cancer biomarkers. Biochim. Biophys. Acta 1834, 2259–2270 (2013).
Yang, B., Fang, X. & Kong, J. In situ sampling and monitoring cell-free DNA of the Epstein–Barr virus from dermal interstitial fluid using wearable microneedle patches. ACS Appl. Mater. Interfaces 11, 38448–38458 (2019).
Al Sulaiman, D. et al. Hydrogel-coated microneedle arrays for minimally invasive sampling and sensing of specific circulating nucleic acids from skin interstitial fluid. ACS Nano 13, 9620–9628 (2019).
McHugh, K. J. et al. Biocompatible near-infrared quantum dots delivered to the skin by microneedle patches record vaccination. Sci. Transl. Med. 11, eaay7162 (2019).
Chang, H. et al. A swellable microneedle patch to rapidly extract skin interstitial fluid for timely metabolic analysis. Adv. Mater. 29, 1702243 (2017).
Kiistala, U. Suction blister device for separation of viable epidermis from dermis. J. Invest Dermatol. 50, 129–137 (1968).
Krogstad, A., Jansson, P. A., Gisslen, P. & Lönnroth, P. Microdialysis methodology for the measurement of dermal interstitial fluid in humans. Br. J. Dermatol. 134, 1005–1012 (1996).
Bodenlenz, M. et al. Open flow microperfusion as a dermal pharmacokinetic approach to evaluate topical bioequivalence. Clin. Pharmacokinet. 56, 91–98 (2017).
Samant, P. P. & Prausnitz, M. R. Mechanisms of sampling interstitial fluid from skin using a microneedle patch. Proc. Natl Acad. Sci. USA 115, 4583–4588 (2018).
Taylor, R. M., Miller, P. R., Ebrahimi, P., Polsky, R. & Baca, J. T. Minimally-invasive, microneedle-array extraction of interstitial fluid for comprehensive biomedical applications: transcriptomics, proteomics, metabolomics, exosome research, and biomarker identification. Lab. Anim. 52, 526–530 (2018).
Muller, D. A., Corrie, S. R., Coffey, J., Young, P. R. & Kendall, M. A. Surface modified microprojection arrays for the selective extraction of the dengue virus NS1 protein as a marker for disease. Anal. Chem. 84, 3262–3268 (2012).
Coffey, J. W., Meliga, S. C., Corrie, S. R. & Kendall, M. A. Dynamic application of microprojection arrays to skin induces circulating protein extravasation for enhanced biomarker capture and detection. Biomaterials 84, 130–143 (2016).
Nedrebø, T., Reed, R. K., Jonsson, R., Berg, A. & Wiig, H. Differential cytokine response in interstitial fluid in skin and serum during experimental inflammation in rats. J. Physiol. 556, 193–202 (2004).
Zhang, X., Chen, G., Bian, F., Cai, L. & Zhao, Y. Encoded microneedle arrays for detection of skin interstitial fluid biomarkers. Adv. Mater. 31, 1902825 (2019).
Coffey, J. W., Corrie, S. R. & Kendall, M. A. Rapid and selective sampling of IgG from skin in less than 1 min using a high surface area wearable immunoassay patch. Biomaterials 170, 49–57 (2018).
Luan, J. et al. Ultrabright fluorescent nanoscale labels for the femtomolar detection of analytes with standard bioassays. Nat. Biomed. Eng. 4, 518–530 (2020).
Dulkeith, E. et al. Gold nanoparticles quench fluorescence by phase induced radiative rate suppression. Nano Lett. 5, 585–589 (2005).
Davis, S. P., Landis, B. J., Adams, Z. H., Allen, M. G. & Prausnitz, M. R. Insertion of microneedles into skin: measurement and prediction of insertion force and needle fracture force. J. Biomech. 37, 1155–1163 (2004).
Kampman, K. M. The treatment of cocaine use disorder. Sci. Adv. 5, eaax1532 (2019).
Shorter, D. & Kosten, T. R. Novel pharmacotherapeutic treatments for cocaine addiction. BMC Med. 9, 119 (2011).
Kinsey, B. M., Kosten, T. R. & Orson, F. M. Anti-cocaine vaccine development. Expert Rev. Vaccines 9, 1109–1114 (2010).
Martell, B. et al. Cocaine vaccine for the treatment of cocaine dependence: a randomized double-blind placebo-controlled efficacy trial. Arch. Gen. Psych. 66, 1116–1123 (2009).
Scott, E. A., Karabin, N. B. & Augsornworawat, P. Overcoming immune dysregulation with immunoengineered nanobiomaterials. Annu. Rev. Biomed. Eng. 19, 57–84 (2017).
Taylor, J., Laudenbach, M., Tucker, A., Jenkins, M. & Pravetoni, M. Hapten-specific naive B cells are biomarkers of vaccine efficacy against drugs of abuse. J. Immunol. Methods 405, 74–86 (2014).
Monteiro-Riviere, N. A., Bristol, D. G., Manning, T. O., Rogers, R. A. & Riviere, J. E. Interspecies and interregional analysis of the comparative histologic thickness and laser Doppler blood flow measurements at five cutaneous sites in nine species. J. Invest. Dermatol. 95, 582-6 (1990).
Davidson, A., Al-Qallaf, B. & Das, D. B. Transdermal drug delivery by coated microneedles: geometry effects on effective skin thickness and drug permeability. Chem. Eng. Res. Des. 86, 1196–1206 (2008).
Copeland, S., Warren, H. S., Lowry, S. F., Calvano, S. E. & Remick, D. Acute inflammatory response to endotoxin in mice and humans. Clin. Diagn. Lab. Immunol. 12, 60–67 (2005).
Hong, Y.-H., Chao, W.-W., Chen, M.-L. & Lin, B.-F. Ethyl acetate extracts of alfalfa (Medicago sativa L.) sprouts inhibit lipopolysaccharide-induced inflammation in vitro and in vivo. J. Biomed. Sci. 16, 64 (2009).
Colnot, C. Skeletal cell fate decisions within periosteum and bone marrow during bone regeneration. J. Bone Miner. Res. 24, 274–282 (2009).
Dwek, J. R. The periosteum: what is it, where is it, and what mimics it in its absence? Skelet. Radiol. 39, 319–323 (2010).
Sakai, D. et al. Remodeling of actin cytoskeleton in mouse periosteal cells under mechanical loading induces periosteal cell proliferation during bone formation. PLoS ONE 6, e24847 (2011).
Moore, S. R., Milz, S. & Knothe Tate, M. L. Periosteal thickness and cellularity in mid-diaphyseal cross‐sections from human femora and tibiae of aged donors. J. Anat. 224, 142–149 (2014).
Merle, B. & Garnero, P. The multiple facets of periostin in bone metabolism. Osteoporos. Int. 23, 1199–1212 (2012).
Kyutoku, M. et al. Role of periostin in cancer progression and metastasis: inhibition of breast cancer progression and metastasis by anti-periostin antibody in a murine model. Int. J. Mol. Med. 28, 181–186 (2011).
Yan, J. et al. Circulating periostin levels increase in association with bone density loss and healing progression during the early phase of hip fracture in Chinese older women. Osteoporos. Int. 28, 2335–2341 (2017).
Bonnet, N., Garnero, P. & Ferrari, S. Special Issue on bone disease mechanisms: periostin action in bone. Mol. Cell. Endocrinol. 432, 75–82 (2015).
Cai, Y. et al. Magnet patterned superparamagnetic Fe3O4/Au core–shell nanoplasmonic sensing array for label-free high throughput cytokine immunoassay. Adv. Healthc. Mater. 8, 1801478 (2019).
Hu, M. et al. Gold nanostructures: engineering their plasmonic properties for biomedical applications. Chem. Soc. Rev. 35, 1084–1094 (2006).
Lee, K.-S. & El-Sayed, M. A. Dependence of the enhanced optical scattering efficiency relative to that of absorption for gold metal nanorods on aspect ratio, size, end-cap shape, and medium refractive index. J. Phys. Chem. B 109, 20331–20338 (2005).
Gole, A. & Murphy, C. J. Azide-derivatized gold nanorods: functional materials for ‘click’ chemistry. Langmuir 24, 266–272 (2008).
Tebbe, M., Kuttner, C., Männel, M., Fery, A. & Chanana, M. Colloidally stable and surfactant-free protein-coated gold nanorods in biological media. ACS Appl. Mater. Interfaces 7, 5984–5991 (2015).
Acknowledgements
We acknowledge support from the National Science Foundation (CBET-1900277) and the National Institutes of Health (R01DE027098, R56DE027924, R01CA141521, R21DA036663, R21CA236652). We thank the Nano Research Facility and Institute of Materials Science and Engineering at Washington University for providing access to electron microscopy facilities, N. Huebsch for providing access to the fluorescence microscope, K. Magee for the help with mouse experiments, Y. Diao for help with digital photographs and M. Shen for inspiring discussions.
Author information
Authors and Affiliations
Contributions
S.S., Z.W. and J.L. conceived the project. S.S., J.S.R., E.L.S., Z.W. and J.L. designed the experiments. Z.W., J.L. and M.Y. fabricated the microneedle patches. Z.W., A.S., L.L., X.Z., J.S.R. and E.L.S. performed the animal experiments. Z.W., J.L., A.S., L.L., Q.Z. and J.J.M. performed the bioassays. J.L. and Q.J. synthesized the plasmonic fluors. P.G. synthesized the magnetic nanoparticles. R.G. performed TEM imaging. P.R. and Y.W. performed SEM imaging. S.C. collected fluorescence images of the microneedles. S.S., J.S.R., E.L.S., J.J.M., Z.W. and J.L. wrote the paper. All authors reviewed and commented on the manuscript.
Corresponding authors
Ethics declarations
Competing interests
The authors declare the following competing financial interests: J.L., J.J.M. and S.S. are inventors on a provisional patent related to plasmonic fluor technology and the technology has been licensed by the Office of Technology Management at Washington University in St Louis to Auragent Bioscience LLC, which is developing plasmonic fluor products. J.L., J.J.M. and S.S. are co-founders and shareholders of Auragent Bioscience LLC. These potential conflicts of interest have been disclosed and are being managed by Washington University in St Louis.
Additional information
Peer review information Peer reviewer reports are available.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
About this article
Cite this article
Wang, Z., Luan, J., Seth, A. et al. Microneedle patch for the ultrasensitive quantification of protein biomarkers in interstitial fluid. Nat Biomed Eng 5, 64–76 (2021). https://doi.org/10.1038/s41551-020-00672-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/s41551-020-00672-y
This article is cited by
-
Long-term monitoring of ultratrace nucleic acids using tetrahedral nanostructure-based NgAgo on wearable microneedles
Nature Communications (2024)
-
Interstitial fluid-based wearable biosensors for minimally invasive healthcare and biomedical applications
Communications Materials (2024)
-
AgNPs-Modified Polylactic Acid Microneedles: Preparation and In Vivo/In Vitro Antimicrobial Studies