Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

A high-impedance detector-array glove for magnetic resonance imaging of the hand

A Publisher Correction to this article was published on 29 August 2018

This article has been updated

Abstract

Densely packed resonant structures used for magnetic resonance imaging (MRI), such as nuclear magnetic resonance phased array detectors, suffer from resonant inductive coupling, which restricts the coil design to fixed geometries, imposes performance limitations and narrows the scope of MRI experiments to motionless subjects. Here, we report the design of high-impedance detectors, and the fabrication and performance of a wearable detector array for MRI of the hand, that cloak themselves from electrodynamic interactions with neighbouring elements. We experimentally verified that the detectors do not suffer from the signal-to-noise degradation mechanisms typically observed with the use of traditional low-impedance elements. The detectors are adaptive and can accommodate movement, providing access to the imaging of soft-tissue biomechanics with unprecedented flexibility. The design of the wearable detector glove exemplifies the potential of high-impedance detectors in enabling a wide range of applications that are not well suited to traditional coil designs.

This is a preview of subscription content, access via your institution

Access options

Rent or buy this article

Prices vary by article type

from$1.95

to$39.95

Prices may be subject to local taxes which are calculated during checkout

Fig. 1: Design of NMR coil elements.
Fig. 2: Evaluation of signal coupling between three coil elements placed side by side.
Fig. 3: Evaluation of SNR degradation due to coupling between neighbouring coil elements as a function of coil overlap.
Fig. 4: Wearable glove coil design and images.
Fig. 5: Performance of the HIC glove coil compared with three state-of-the-art LIC phased arrays.

Similar content being viewed by others

Change history

  • 29 August 2018

    Owing to a technical error, this Article was originally published with an incorrect published online date of ‘4 May 2018’; it should have been ‘7 May 2018’. This has now been corrected.

References

  1. Bloch, F., Nicodemus, D. & Staub, H. A. Quantitative determination of the magnetic moment of the neutron in units of the proton moment. Phys. Rev. 74, 1025 (1948).

    Article  Google Scholar 

  2. Lauterbur, P. C. Image formation by induced local interactions: examples employing nuclear magnetic resonance. Nature 242, 190–191 (1973).

    Article  CAS  Google Scholar 

  3. Mansfield, P. & Grannell, P. K. NMR ‘diffraction’ in solids? J. Phys. C Solid State 6, L422–L426 (1973).

    Article  CAS  Google Scholar 

  4. Roemer, P. B. et al. The NMR phased array. Magn. Reson. Med. 16, 192–225 (1990).

    Article  PubMed  CAS  Google Scholar 

  5. Wang, J., Reykowski, A. & Dickas, J. Calculation of the signal-to-noise ratio for simple surface coils and arrays of coils. IEEE Trans. Biomed. Eng. 42, 908–917 (1995).

    Article  PubMed  CAS  Google Scholar 

  6. Ocali, O. & Atalar, E. Ultimate intrinsic signal-to-noise ratio in MRI. Magn. Reson. Med. 39, 462–473 (1998).

    Article  PubMed  CAS  Google Scholar 

  7. Schnell, W., Renz, W., Vester, M. & Ermert, H. Ultimate signal-to-noise ratio of surface and body antennas for magnetic resonance imaging. IEEE Trans. Antenn. Propag. 48, 418–428 (2000).

    Article  Google Scholar 

  8. Ohliger, M. A., Grant, A. K. & Sodickson, D. K. Ultimate intrinsic signal-to-noise ratio for parallel MRI: electromagnetic field considerations. Magn. Reson. Med. 50, 1018–1030 (2003).

    Article  PubMed  Google Scholar 

  9. Wiesinger, F., Boesiger, P. & Pruessmann, K. P. Electrodynamics and ultimate SNR in parallel MR imaging. Magn. Reson. Med. 52, 376–390 (2004).

    Article  PubMed  Google Scholar 

  10. Lattanzi, R. et al. Performance evaluation of a 32-element head array with respect to the ultimate intrinsic SNR. NMR Biomed. 23, 142–151 (2010).

    PubMed  PubMed Central  Google Scholar 

  11. Lattanzi, R. & Sodickson, D. K. Ideal current patterns yielding optimal signal-to-noise ratio and specific absorption rate in magnetic resonance imaging: computational methods and physical insights. Magn. Reson. Med. 68, 286–304 (2012).

    Article  PubMed  Google Scholar 

  12. Wiggins, G. C. et al. 96-Channel receive-only head coil for 3 Tesla: design optimization and evaluation. Magn. Reson. Med. 62, 754–762 (2009).

    Article  PubMed  PubMed Central  Google Scholar 

  13. Schmitt, M. et al. A 128-channel receive-only cardiac coil for highly accelerated cardiac MRI at 3 Tesla. Magn. Reson. Med. 59, 1431–1439 (2008).

    Article  PubMed  PubMed Central  Google Scholar 

  14. Fujita, H., Zheng, T., Yang, X., Finnerty, M. J. & Handa, S. RF surface receive array coils: the art of an LC circuit. J. Magn. Reson. Imaging 38, 12–25 (2013).

    Article  PubMed  Google Scholar 

  15. Kurs, A. et al. Wireless power transfer via strongly coupled magnetic resonances. Science 317, 83–86 (2007).

    Article  PubMed  CAS  Google Scholar 

  16. Tierney, B. & Grbic, A. Planar shielded-loop resonators for wireless non-radiative power transfer. In IEEE Antennas and Propagation Soc. Int. Symp. 842–843 (IEEE, 2014).

  17. Gonord, P., Kan, S. & Leroy-Willig, A. Parallel-plate split-conductor surface coil: analysis and design. Magn. Reson. Med. 6, 353–358 (1988).

    Article  PubMed  CAS  Google Scholar 

  18. Serfaty, S., Haziza, N., Darrasse, L. & Kan, S. Multi-turn split-conductor transmission-line resonators. Magn. Reson. Med. 38, 687–689 (1997).

    Article  PubMed  CAS  Google Scholar 

  19. Frass-Kriegl, R. et al. Multi-turn multi-gap transmission line resonators: concept, design and first implementation at 4.7T and 7T. J. Magn. Reson. 273, 65–72 (2016).

    Article  PubMed  CAS  Google Scholar 

  20. Corea, J. R. et al. Screen-printed flexible MRI receive coils. Nat. Commun. 7, 10839 (2016).

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  21. Vasanawala, A. et al. Development and clinical implementation of very light weight and highly flexible AIR technology arrays. In Proc. Intl. Soc. Magn. Reson. Med. 0755 (ISMRM, 2017).

  22. Stormont, R. et al. Reimagining Flexible Coil Technology 69–71 (SIGNA, 2017).

  23. Stengaard, A. Planar quadrature coil design using shielded-loop resonators. J. Magn. Reson. 125, 84–91 (1997).

    Article  Google Scholar 

  24. Sodickson, D. K. & Manning, W. J. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn. Reson. Med. 38, 591–603 (1997).

    Article  PubMed  CAS  Google Scholar 

  25. Pruessmann, K., Weiger, M., Scheidegger, M. B. & Boesiger, P. SENSE: sensitivity encoding for fast MRI. Magn. Reson. Med. 42, 952–962 (1999).

    Article  PubMed  CAS  Google Scholar 

  26. Griswold, M. et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn. Reson. Med. 47, 1202–1210 (2002).

    Article  PubMed  Google Scholar 

  27. Larkman, D. J. et al. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited. J. Magn. Reson. Imaging 13, 313–317 (2001).

    Article  PubMed  CAS  Google Scholar 

  28. Setsompop, K. et al. Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty. Magn. Reson. Med. 67, 1210–1224 (2012).

    Article  PubMed  Google Scholar 

  29. Schnall, M. D., Harihara Subramanian, V., Leigh, J. S. & Chance, B. A new double-tuned probed for concurrent 1H and 31P NMR. J. Magn. Reson. 65, 122–129 (1985).

    CAS  Google Scholar 

  30. Avdievich, N. I. & Hetherington, H. P. 4 T Actively detuneable double-tuned 1H/31P head volume coil and four-channel 31P phased array for human brain spectroscopy. J. Magn. Reson. 186, 341–346 (2007).

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  31. Brown, R. et al. Design of a nested eight-channel sodium and four-channel proton coil for 7T knee imaging. Magn. Reson. Med. 70, 259–268 (2013).

    Article  PubMed  CAS  Google Scholar 

  32. Shajan, G. et al. Three-layered radio frequency coil arrangement for sodium MRI of the human brain at 9.4 Tesla. Magn. Reson. Med. 75, 906–916 (2016).

    Article  PubMed  CAS  Google Scholar 

  33. Kriegl, R. et al. Novel inductive decoupling technique for flexible transceiver arrays of monolithic transmission line resonators. Magn. Reson. Med. 73, 1669–1681 (2015).

    Article  PubMed  Google Scholar 

  34. Hosseinezhadian, S. et al. A flexible transceiver array for 7 T cardiac MRI: first imaging experiments. In Proc. Eur. Soc. Magnetic Resonance in Medicine and Biology 315 (ESMRMB, 2017).

  35. Keil, B. et al. A 64-channel 3T array coil for accelerated brain MRI. Magn. Reson. Med. 70, 248–258 (2013).

    Article  PubMed  Google Scholar 

  36. Noeske, R., Seifert, F., Rhein, K. H. & Rinneberg, H. Human cardiac imaging at 3T using phased array coils. Magn. Reson. Med. 44, 978–982 (2000).

    Article  PubMed  CAS  Google Scholar 

  37. Chung, S., Kim, D., Breton, E. & Axel, L. Rapid B1 + mapping using a preconditioning RF pulse with TurboFLASH readout. Magn. Reson. Med. 64, 439–446 (2010).

    PubMed  PubMed Central  Google Scholar 

  38. Kellman, P. & McVeigh, E. R. Image reconstruction in SNR units: a general method for SNR measurement. Magn. Reson. Med. 54, 1439–1447 (2005).

    Article  PubMed  PubMed Central  Google Scholar 

  39. Winkelmann, S., Schaeffter, T., Koehler, T., Eggers, H. & Doessel, O. An optimal radial profile order based on the Golden Ratio for time-resolved MRI. IEEE Trans. Med. Imaging 26, 68–76 (2007).

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

This work was performed under the rubric of the Center for Advanced Imaging Innovation and Research (www.cai2r.net), a NIBIB Biomedical Technology Resource Center (NIH P41 EB017183). We thank R. Brown for critically reading the manuscript, M. Vester for many valuable discussions and Z. Yu for help during the experiments.

Author information

Authors and Affiliations

Authors

Contributions

B.Z. built the coils and interfaces. B.Z. and M.A.C. designed the experiments and collected the data. B.Z., D.K.S. and M.A.C. analysed the results and wrote the manuscript.

Corresponding authors

Correspondence to Bei Zhang or Martijn A. Cloos.

Ethics declarations

Competing interests

NYU has filed provisional patent applications directed to this technology.

Additional information

Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary information

Supplementary Information

Supplementary figures and video captions.

Reporting Summary

Supplementary Video 1

Magnetic resonance imaging of joint kinematics while playing piano.

Supplementary Video 2

Magnetic resonance imaging of joint kinematics when grabbing objects.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhang, B., Sodickson, D.K. & Cloos, M.A. A high-impedance detector-array glove for magnetic resonance imaging of the hand. Nat Biomed Eng 2, 570–577 (2018). https://doi.org/10.1038/s41551-018-0233-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/s41551-018-0233-y

This article is cited by

Search

Quick links

Nature Briefing

Sign up for the Nature Briefing newsletter — what matters in science, free to your inbox daily.

Get the most important science stories of the day, free in your inbox. Sign up for Nature Briefing