Abstract
Single-molecule Förster Resonance Energy Transfer (smFRET) is a powerful technique capable of resolving both relative and absolute distances within and between structurally dynamic biomolecules. High instrument costs, and a lack of open-source hardware and acquisition software have limited smFRET’s broad application by non-specialists. Here, we present the smfBox, a cost-effective confocal smFRET platform, providing detailed build instructions, open-source acquisition software, and full validation, thereby democratising smFRET for the wider scientific community.
Introduction
FRET is a photophysical process which results in the transfer of excitation energy from a donor fluorophore to an acceptor chromophore1. The efficiency of this transfer process scales inversely with the sixth power of the distance between the two chromophores. Therefore, by measuring the FRET efficiency (e.g. by observing the emission of the two fluorophores under excitation of the donor), spatial information can be determined in the 3–10 nm range, making FRET a ‘spectroscopic ruler’2 well matched to the dimensions of biomolecules such as nucleic acids and proteins3. In ensemble measurements this can be used to detect on-off/relative distance changes such as binding and cleaving in bimolecular interactions, or conformational changes (e.g. opening and closing) in unimolecular processes. At the single-molecule level FRET is sensitive to heterogeneous subpopulations, can measure kinetics of processes at equilibrium4, and as demonstrated by a recent inter-laboratory benchmarking study5, absolute FRET efficiencies can be used to infer precise distances for biomolecular structure determination6,7,8,9,10,11.
Two experimental formats are commonly employed to obtain smFRET data: a confocal approach, in which individual molecules are detected as they diffuse through a confocal volume3; and a total internal reflection fluorescence (TIRF) microscopy approach12, in which individual molecules are immobilised on a glass coverslip and excited by an evanescence field. The two approaches are largely complementary, together allowing the interrogation of biomolecular dynamics at timescales spanning twelve orders of magnitude: from the picosecond–millisecond (confocal); and millisecond–hours (TIRF) (see ref. 13 for a recent review). Confocal experiments are often the first smFRET experiments performed on a new biomolecular system, as this is the more straightforward approach, not requiring surface immobilisation (which can be non-trivial). Such experiments can reveal different biomolecular conformations present in (dynamic) equilibrium, and the corresponding rates of conformational transitions, at timescales relevant to key cellular processes such as protein (un)folding14, transcription15 and DNA replication and repair16,17. Confocal experiments are also the approach of choice for generating multiple smFRET restraints for integrative structural modelling6,7,8,9,10,11, due to the simpler sample preparation, fast data acquisition, and higher time resolution.
Despite the many advantages of smFRET, it is currently rarely used outside specialist labs, largely due to the high costs of commercial instruments and lack of self-build, easy to use alternatives. To address this, here we provide detailed build instructions, parts lists, and open-source acquisition software, to enable a broad range of scientists to perform confocal smFRET experiments, on a validated, self-built, robust and economic instrument.
Results and discussion
Here we present the smfBox18, a cost-effective confocal-based platform capable of measuring the FRET efficiency between dye pairs on freely diffusing single molecules, using variable alternating laser excitation (ALEX)19 for verification of correct dye stoichiometry and the determination of accurate FRET correction factors5,20,21. The smfBox (Fig. 1a, b) is constructed from readily-available optics and optomechanical components, replacing an expensive microscope body with machined anodised-aluminium, which forms a light-tight box housing the excitation dichroic, objective, lenses, and pinhole (see Supplementary Note 1, Supplementary Movie 1 and online18 for the complete parts list, and animated building and alignment protocols). When assembled, the smfBox is sufficiently light-tight to allow safe and effective operation under ambient light conditions, as a Class I laser product (eliminating the need for user laser-safety training). The smfBox can be operated with either customisable LabVIEW acquisition software, or a stand-alone user interface written in C++. Both versions of the software provide all the necessary functionality for setting up the microscope (alignment and focusing) and recording data (Fig. 1c and Supplementary Notes 2 and 3). The raw data, comprising photon arrival times and detector ID, are saved in the open-source photon-HDF5 data format (Supplementary Note 4)22, and can subsequently be analysed either with the FRETBursts python module23 using the Jupyter Notebooks provided (Supplementary Note 6 and online18), or with the GUI-based MATLAB package PAM24.
a Schematic of the smfBox with parts labelled according to Supplementary Tables 1, 2 and 4. Lasers are collimated (L1), cropped (Iris) and steered by two mirrors (M1, M2) onto a 10:90 beam splitter (BS1). 10% of the beam is focused on to a photodiode for continuous power measurement and alternation cycle monitoring. 90% of the beam is directed via dichroic mirror 1 (DC1) to the objective. Light from the back reflection is reflected by DC1 and BS1 onto a CCD camera for accurate focusing. Fluorescence emission from the sample passes through the excitation dichroic (DC1) and is focused onto a pinhole (P1) to remove out of focus light, before being split by colour (DC2) onto one of two avalanche photodiodes (APD0, APD1). b 3D model of the completed smfBox, with the front panel of the microscope body removed. A CAD file of the assembled instrument is provided (Supplementary Data 1) along with an animated build sequence (Supplementary Movie 1). c A flowchart of the smfBox platform showing the functionality of the acquisition software. d Schematic: single molecules diffuse through a confocal volume (residence time ~1 ms) constructed by focusing the lasers into a near-diffraction-limited spot and using a pinhole to section the emission light in a thin focal plane. Lasers (515 nm and 638 nm) are alternated (20 kHz) to ensure multiple excitations of the donor and acceptor dyes for each molecule. e A typical time trace for an smFRET experiment. Fluorescently-labelled molecules diffuse through the confocal volume emitting bursts of fluorescence. Individual photons generated by emission from the donor under donor excitation (DD—green) are recorded on APD0. APD1 records photons emitted by the acceptor either under donor excitation (DA—red) or direct acceptor excitation (AA—purple). f 2D ES histogram showing uncorrected FRET efficiency (E*) and stoichiometry (S). Donor-only molecules appear with low E* but high S, and acceptor-only molecules appear with low S. Doubly labelled molecules appear with intermediate S.
The smfBox detects both donor and acceptor emission from single molecules freely diffusing through a confocal spot under alternating laser excitation (Fig. 1d, e). Emission under green excitation is used to determine the FRET efficiency (E; Supplementary Equation 1), whilst the response to the red laser confirms the presence of an active acceptor on the molecule, and allows the calculation of the stoichiometry parameter (S; Supplementary Eq. 2) and all correction parameters required for accurate FRET determination (Supplementary Eqs. 3–5)5,20,21. The precise ALEX cycle of the smfBox can be fully customised, allowing for faster or slower cycles, periodic acceptor excitation (PAX)25, or an asymmetric ALEX scheme, which we show can reduce the width of FRET histograms, thereby increasing the resolution of different FRET species (Supplementary Note 7). Furthermore, the design of the smfBox includes a 10:90 beam splitter in the excitation path, directing excitation light to a photodetector (Fig. 1a) to allow for precise monitoring of both laser powers in real time during the experiment, which can be saved into the HDF5 file. Our default optimal values of the laser powers, iris diameter, and ALEX cycle are provided for this setup (see methods).
To test the performance of the smfBox we measured the FRET efficiencies of three DNA standards (Fig. 2a), which were recently characterised by multiple labs using a range of commercial and home-built microscopes5. Using the published correction procedures implemented in our open-source python analysis (Jupyter notebooks - Supplementary Note 6), we obtained data in excellent agreement with those from the other labs in the blind study. This provides both an excellent validation of the smfBox, but also a useful diagnostic for users to test their own builds of this instrument, as the successful reproduction of these data means that all hardware, acquisition and analysis software must be working correctly.
a Fully corrected FRET efficiency histograms of three doubly labelled DNA standards (1a, 1b and 1c, cartoons with dye accessible volumes; for sequences see Supplementary Note 5) measured using the smfBox (grey). Vertical black lines and curves show Gaussian fits of our data, E = 0.17 ± 0.07, E = 0.57 ± 0.1, E = 0.77 ± 0.07 (mean ± sd), compared to the results from 20 other labs as part of a multi-lab benchmarking study5 (red crosses—Supplementary Note 6). b Proximity ratio (uncorrected FRET efficiency) histograms of a DNA hairpin at indicated salt concentrations (see Supplementary Note 5). c Salt dependent rates for hairpin opening (kopen) and closing (kclose) determined by dynamic photon distribution analysis (dPDA)4 (mean ± SD, n = 2 with >1000 molecules per technical repeat at each [NaCl])—Source data are provided in a Source Data file. d Proximity ratio histograms of High-, Mid- and Low-FRET hairpins (at 300 mM NaCl). Data (grey) were fit using dPDA (black) to a two-state model, comprising a closed population (blue), open population (orange) and interconverting dynamic population (yellow). e Plot of rates determined from dPDA of nine data sets for each hairpin, each containing 2000 molecules, quoting the mean and standard deviation across the data sets, with the mean chi-squared of the fits plotted to the right.
Furthermore, we have demonstrated the capability of the smfBox to recover rates of molecular conformational dynamics, using DNA hairpins as a test system (Fig. 2b). DNA hairpins have been shown to interconvert between a closed and open state, with rates that are dependent on NaCl concentration26,27, within time scales accessible to smFRET experiments (see Supplementary Note 8). First, we reproduced data for a hairpin used in a recent study26, using dynamic photon distribution analysis (dPDA)4 to determine opening and closing rates for a series of NaCl concentrations (Fig. 2c). Next, we analysed two further hairpins (Fig. 2d), identical in DNA sequence, but with a smaller and greater inter-fluorophore distance in the closed state, to test the effects of the magnitude of FRET efficiency changes on the precision of the recovered kinetic parameters. As might be expected, the rates of interconversion for the hairpin with a higher-FRET-efficiency closed state could be determined more precisely, whereas analysis of the lower-FRET hairpin produced more variable results (Fig. 2e). In cases where the static (low-FRET and high-FRET) species have considerable overlap with each other (and therefore with the dynamic population) the dPDA model is less well constrained, leading to a greater variation in the values of recovered rates for a given sample size. These results have implications for the optimal positioning of FRET dyes when designing dynamic experiments.
Whilst the microscope we describe is built for ALEX confocal smFRET, its modular design makes for easy expansion to several related techniques. Without any additional hardware, the smfBox is capable of fluorescence correlation spectroscopy (FCS—using a single continuous-wave laser) and fluorescence cross-correlation spectroscopy (FCCS—using two lasers). We demonstrate this capability by determining the diffusion constant for a duplex DNA (Supplementary Note 10). The addition of one or more pulsed lasers and time-correlated single-photon counting (TCSPC) electronics will enable fluorescence lifetime correlation spectroscopy (FLCS) and pulsed-interleaved excitation (PIE) experiments28. The further addition of polarisation filters and two additional APDs would constitute a full multi-parameter fluorescence detection (MFD) setup29. Furthermore, the addition of an XY-stage to the Z-positioning stage already included would facilitate any number of imaging techniques by scanning the sample. A list of recommended components for such expanded applications can be found in Supplementary Note 9.
In conclusion, we have provided all necessary instructions and software required for the construction and operation of the smfBox, a cost-effective and open-source smFRET microscope with competitive capabilities18. We demonstrate the smfBox can determine absolute FRET efficiencies with the same accuracy as other instruments used by the community5, and can recover biomolecular interconversion kinetics in the range ~50–500 s−1, in agreement with previous studies26. We have shown that an asymmetric ALEX duty cycle can reduce the width of smFRET histograms, increasing the resolution of different FRET species. Finally, we have experimentally assessed the ability to determine kinetic rates of interconversion using dPDA, for systems with differing magnitudes of FRET efficiency changes, providing useful information for the design of dynamic smFRET experiments. We anticipate that our low-cost open-source approach will ease the adoption of smFRET by the wider scientific community.
Methods
The smfBox
Full details of the construction and operation of the smfBox are described in Supplementary Note 1, Supplementary Figs. 1–14, Supplementary Movie 1 and online18. Briefly, the smfBox alternates two lasers (515 nm–222 μW, and 635 nm–68 μW, Omicron LuxX plus lasers, powers measured immediately before the excitation dichroic) by TTL-controlled modulation of electronic shutters. The beams are coupled into a single-mode fibre before being collimated (to 10 mm) and cropped by an iris (to 5 mm), then directed into a custom built anodised-aluminium microscope body (see Supplementary Figures 1–14, and Supplementary Data 1 and 2 for technical drawings and a CAD file of the assembled instrument). A dichroic mirror (Chroma ZT532/640 rpc 3 mm) directs the beam into an objective (Olympus UPLSAPO ×60 NA = 1.35 oil immersion), and the same objective collects the emission, which is focussed onto a 20 μm pinhole and split (Chroma NC395323—T640lpxr) to two avalanche photodiodes (SPCM-AQRH-14 and SPCM-NIR-14, Excelitas), where photon arrival times are recorded by a national instruments card (PCIe-6353).
Accurate FRET experiments
Three duplex DNA constructs (referred to as 1a, 1b and 1c) labelled with Atto550 (donor) and Atto647N (acceptor), and were provided by the Hugel Lab as part of the blind, multi-lab FRET study5 (for sequences see Supplementary Note 5). DNAs were diluted to approximately 100 pM in observation buffer 1 (20 mM MgCl2, 5 mM NaCl, 5 mM Tris, pH 7.5), and ~50 μl placed on a coverslip passivated with 1 mg/ml BSA, and data were acquired by the smfBox. Analysis was done in Anaconda 5.3.0, with Jupyter Notebooks using the FRETBursts python module23 (version 0.6.5). Background in each channel was estimated by means of an exponential fit of inter-photon delays. Bursts were identified using an all photon sliding window algorithm previously described23,30 with L = 10 and F = 45 for both channels, and background was subtracted. Spectral cross talk factors were found by combining data from all standards and extracting bursts with a stoichiometry >0.95 as the donor only population and <0.175 for the acceptor only population (see Fig. 1f) to calculate α and δ, respectively. A dual channel burst search (DCBS) was then used to extract doubly-labelled bursts from each 30-min acquisition, and used to find E and S with single Gaussian fits. Combined data from all three oligos were then plotted together and fitted to obtain γ and β (Supplementary Equation 5). Corrected FRET efficiencies of all doubly-labelled bursts were then obtained using all four correction parameters as previously described5. See Supplementary Note 7 for more details. Jupyter Notebooks and raw data (HDF5 files) are available on the smfBox github18.
Hairpin dynamics
DNA hairpins were made from a self-complementary oligonucleotide labelled with Cy3B annealed to a short oligonucleotide with Atto647N at different positions (see Supplementary Note 5). Oligonucleotides were purchased from LGC Biosearch (UK), with internal amino modified–dT bases which were labelled with Cy3B and Atto647N NHS esters purchased from GE Healthcare (US) and ATTO-TEC (Germany), respectively. Labelled DNAs were purified via polyacrylamide gel electrophoresis and annealed in annealing buffer (50 mM NaCl, 10 mM Tris HCl, pH 7.5, 1 mM EDTA), by heating to 95 °C (5 min) followed by overnight cooling. Hairpins were then diluted to approximately 100 pM in Hairpin buffer (Tris 50 mM pH 7.5, BSA 0.1 mg/ml, EDTA 1 mM, Glycerol 5%, DTT 1 mM) with additional NaCl where specified, placed into a chamber made of two coverslips and a silicone gasket (to enable >2 h acquisitions with no sample evaporation, therefore maintaining a constant salt concentration), and data were acquired by the smfBox. Data were analysed using the MATLAB software package PAM24. Bursts were selected using a sliding window dual channel burst search, with a 50 photon threshold and a 500 μs window size. Doubly-labelled bursts were selected between 0.2 and 0.85 S, and bursts were cut into 0.5, 1, and 1.5 ms lengths. To access the precision of the kinetic parameters, acquisitions were split into subsets of 2000 molecules before further analysis. Dynamic PDA4 was then used to fit a two-state model to the data using the histogram library method implemented in PAM. The raw data for the hairpin experiments (HDF5 files) are available on the smfBox github18.
Reporting summary
Further information on research design is available in the Nature Research Reporting Summary linked to this article.
Data availability
As keen proponents of open science and open data, we have made the photon arrival time data (photon-HDF5 files) which support the findings of this study publicly available from Zenodo, https://zenodo.org/record/3625603#.XxmB3ChKhm9. This includes data pertaining to: Fig. 2a—Jnotebooks/definitiveset/1ax.hdf5; Fig. 2b, c—Hairpin salt data/1/…; Fig. 2d, e—Hairpin hi-mid-low fret data. The individual rate constant determinations for Fig. 2c are available in the Source Data file. Full build instructions for the smfBox, including animations, are available through our GitHub https://craggslab.github.io/smfBox/18. Source data are provided with this paper.
Code availability
All new acquisition and analysis software are accessible through the Craggs Lab Github (https://craggslab.github.io/smfBox/)18.
References
Förster, T. Zwischenmolekulare Energiewanderung und Fluoreszenz. Ann. Phys. 437, 55–75 (1948).
Stryer, L. & Haugland, R. P. Energy transfer: a spectroscopic ruler. Proc. Natl Acad. Sci. U. S. A. 58, 719–726 (1967).
Lerner, E. et al. Toward dynamic structural biology: two decades of single-molecule Förster resonance energy transfer. Science 359, eaan1133 (2018).
Kalinin, S., Valeri, A., Antonik, M., Felekyan, S. & Seidel, C. A. M. Detection of structural dynamics by FRET: a photon distribution and fluorescence lifetime analysis of systems with multiple states. J. Phys. Chem. B 114, 7983–7995 (2010).
Hellenkamp, B. et al. Precision and accuracy of single-molecule FRET measurements—a multi-laboratory benchmark study. Nat. Methods 15, 669 (2018).
Hellenkamp, B., Wortmann, P., Kandzia, F., Zacharias, M. & Hugel, T. Multi-domain structure and correlated dynamics determined by self-consistent FRET networks. Nat. Methods 14, 174–180 (2017).
Craggs, T. D. et al. Substrate conformational dynamics facilitate structure-specific recognition of gapped DNA by DNA polymerase. Nucleic Acids Res. 47, 10788–10800 (2019).
Tsytlonok, M. et al. Dynamic anticipation by Cdk2/Cyclin A-bound p27 mediates signal integration in cell cycle regulation. Nat. Commun. 10, 1676 (2019).
Nagy, J. et al. Complete architecture of the archaeal RNA polymerase open complex from single-molecule FRET and NPS. Nat. Commun. 6, 6161 (2015).
LeBlanc, S. J. et al. Coordinated protein and DNA conformational changes govern mismatch repair initiation by MutS. Nucleic Acids Res. 46, 10782–10795 (2018).
Kalinin, S. et al. A toolkit and benchmark study for FRET-restrained high-precision structural modeling. Nat. Methods 9, 1218–1225 (2012).
Roy, R., Hohng, S. & Ha, T. A practical guide to single-molecule FRET. Nat. Methods 5, 507–516 (2008).
Lerner, E. et al. The FRET-based structural dynamics challenge - community contributions to consistent and open science practices. Preprint at https://arxiv.org/abs/2006.03091v1 (2020).
Orte, A., Craggs, T. D., White, S. S., Jackson, S. E. & Klenerman, D. Evidence of an intermediate and parallel pathways in protein unfolding from single-molecule fluorescence. J. Am. Chem. Soc. 130, 7898–7907 (2008).
Robb, N. C. et al. The transcription bubble of the RNA polymerase-promoter open complex exhibits conformational heterogeneity and millisecond-scale dynamics: Implications for transcription start-site selection. J. Mol. Biol. 425, 875–885 (2013).
Hohlbein, J. et al. Conformational landscapes of DNA polymerase I and mutator derivatives establish fidelity checkpoints for nucleotide insertion. Nat. Commun. 4, 2131 (2013).
Bennet, I. A. et al. Regional conformational flexibility couples substrate specificity and scissile phosphate diester selectivity in human flap endonuclease 1. Nucleic Acids Res. 46, 5618–5633 (2018).
Ambrose, B., Steele, E.M. smfBox—Craggs Lab Github. https://craggslab.github.io/smfBox/ (2019).
Kapanidis, A. N. et al. Fluorescence-aided molecule sorting: analysis of structure and interactions by alternating-laser excitation of single molecules. Proc. Natl Acad. Sci. 101, 8936–8941 (2004).
Lee, N. K. et al. Accurate FRET measurements within single diffusing biomolecules using alternating-laser excitation. Biophys. J. 88, 2939–2953 (2005).
Hohlbein, J., Craggs, T. D. & Cordes, T. Alternating-laser excitation: single-molecule FRET and beyond. Chem. Soc. Rev. 43, 1156–1171 (2014).
Ingargiola, A., Laurence, T., Boutelle, R., Weiss, S. & Michalet, X. Photon-HDF5: open data format and computational tools for timestamp-based single-molecule experiments. Proc. SPIE—Int. Soc. Opt. Eng. 9714, 971405 (2016).
Ingargiola, A., Lerner, E., Chung, S., Weiss, S. & Michalet, X. FRETBursts: an open source toolkit for analysis of freely-diffusing single-molecule FRET. PLoS ONE 11, e0160716 (2016).
Schrimpf, W., Barth, A., Hendrix, J. & Lamb, D. C. PAM: a framework for integrated analysis of imaging, single-molecule, and ensemble fluorescence data. Biophys. J. 114, 1518–1528 (2018).
Doose, S., Heilemann, M., Michalet, X., Weiss, S. & Kapanidis, A. N. Periodic acceptor excitation spectroscopy of single molecules. Eur. Biophys. J. 36, 669–674 (2007).
Farooq, S. & Hohlbein, J. Camera-based single-molecule FRET detection with improved time resolution. Phys. Chem. Chem. Phys. 17, 27862–27872 (2015).
Tsukanov, R. et al. Detailed study of DNA hairpin dynamics using single-molecule fluorescence assisted by DNA origami. J. Phys. Chem. B 117, 11932–11942 (2013).
Müller, B. K., Zaychikov, E., Bräuchle, C. & Lamb, D. C. Pulsed interleaved excitation. Biophys. J. 89, 3508–3522 (2005).
Widengren, J. et al. Single-molecule detection and identification of multiple species by multiparameter fluorescence detection. Anal. Chem. 78, 2039–2050 (2006).
Nir, E. et al. Shot-noise limited single-molecule fret histograms: comparison between theory and experiments. J. Phys. Chem. B 110, 22103–22124 (2006).
Acknowledgements
The authors thank Gary Turner for refining the design and producing the aluminium pieces for the smfBox, Anna van den Boom for help with AutoCAD and part checking, Lachlan Whitehead for image rendering in Fig. 1b, and Alison Twelvetrees for critical reading and helpful discussions. We are grateful for funding from the following sources: EPSRC studentships (B.A. and E.S.); STFC (B.C.B., M.L.M.-F. and T.D.C); EU Erasmus + (M.A.); The University of Sheffield Start-up grant (T.D.C.); the Royal Society (RGS\R2\180405, T.D.C), and the BBSRC (BB/T008032/1, T.D.C.).
Author information
Authors and Affiliations
Contributions
T.D.C. designed the research; B.A., J.B., J.C., J.S. and T.D.C. built the instrument; B.A. and M.W. performed all validation experiments and data analysis. B.A., J.B., J.C., E.S. and B.C.B. wrote acquisition and analysis software. J.B. and M.A. performed CAD design and rendering. M.L.M.-F., A.C. and T.D.C. provided supervision. B.A., J.B., J.C. and T.D.C. wrote the manuscript with edits from all authors.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Peer review information Nature Communications thanks Siân Culley and the other, anonymous, reviewer(s) for their contribution to the peer review of this work. Peer reviewer reports are available.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Source data
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ambrose, B., Baxter, J.M., Cully, J. et al. The smfBox is an open-source platform for single-molecule FRET. Nat Commun 11, 5641 (2020). https://doi.org/10.1038/s41467-020-19468-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41467-020-19468-4
This article is cited by
-
Reliability and accuracy of single-molecule FRET studies for characterization of structural dynamics and distances in proteins
Nature Methods (2023)
-
Open microscopy in the life sciences: quo vadis?
Nature Methods (2022)
-
Instruments of change for academic tool development
Nature Physics (2021)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.