Abstract
Objective
No diagnostic gold standard for keratoconus in children and adolescents exists. Our objective was investigating the diagnostic accuracy of various indices for keratoconus (KC) detection in paediatric eyes.
Methods
All retrievable data of significance from 432 normal right paediatric eyes and 48 eyes of paediatric KC and forme fruste KC (FFKC), imaged by use of a rotating Scheimpflug camera (Oculyzer II, Pentacam HR) between December 2013 and October 2018 at Watany Eye Hospitals, Cairo, Egypt, including Scheimpflug images data, were collected. The area under the receiver operating characteristic curve (AUROC) was calculated for different indices in this retrospective descriptive study.
Results
All 36 tested indices showed discriminative power differentiating KC and FFKC from normal corneas (AUROC P-value <0.05), except AC volume, AC angle, and horizontal decentrations of the steepest and thinnest points. The 32 indices showed variable degrees of diagnostic accuracy. The highest AUROC was that of the corneal assessment index from the relational thickness and other OCULUS values (CAIRO 8). Only 8 indices showed non-inferiority to it, namely, Ambrosio’s relational thickness maximum (ART max) and avg, the pachymetric progression index maximum (PPI max) and avg, the back elevation from the best-fit toric ellipsoid (BE BFTE), the KC index (KI), the topographic KC indices (TKC), and the index of height decentration (IHD) (P > 0.05).
Conclusions
The 8 most useful rotating Scheimpflug imaging indices for KC detection in paediatric eyes are CAIRO 8 followed by ART max and avg, PPI max and avg, BE BFTE, KI, TKC, and IHD.
Similar content being viewed by others
Introduction
Keratoconus (KC) is an ectatic corneal disorder characterised by progressive thinning of the cornea and structural weakening, resulting in corneal protrusion, irregular astigmatism, and reduced vision [1, 2].
KC usually starts at puberty. In the paediatric age group, KC is often more advanced at the time of diagnosis and the progression of the disease is more rapid, with a higher need for corneal grafting because treatment by corneal crosslinking, the Athens protocol, intracorneal rings, or ring segments must be performed at earlier stages. Moreover, visual impairment can affect social and educational development in children [3]. Therefore, the detection of progressive KC in its early stages is necessary to prevent such consequences [4].
Few papers have discussed the epidemiology of paediatric KC, but a Lebanese study reported an incidence of 0.53% in children aged 14 years or younger who were diagnosed in a tertiary referral eye centre [5]. Based on the American Academy of Ophthalmology’s Intelligence Research in Sight Registry (IRIS), the prevalence of KC is 0.16% in the paediatric population [6].
The clinical manifestation of KC in children is somewhat different from that seen in adults. The cone is reported to be more centrally located in paediatric cases and, therefore, irregular astigmatism is less evident [7]. Clinically, the diagnosis of frank KC is straightforward. However, when it comes to subclinical, form fruste, suspect, or borderline cases, there is widespread ambiguity. The implications of tomographic indices are of the highest value and impact in these cases before the KC is frankly manifest [8,9,10,11]. The Scheimpflug camera system with its tomographic indices enables the assessment of both the anterior and posterior corneal surfaces [12]. The detection of posterior corneal surface abnormalities in clinically normal patients was a breakthrough in the diagnosis and monitoring of the disease [13].
As the minimum age approved for corneal refractive surgery is 18 years, corneal tomography is not commonly done in children and adolescents, so this age group has no well-established diagnostic gold standard [14]. We therefore published the normative data in paediatric age groups [15]. The current study aims to evaluate the accuracy and to compare the different indices of KC detection in paediatric eyes using the rotating Scheimpflug camera (Oculyzer II, Pentacam HR).
Materials and methods
This is a retrospective observational study conducted at Watany Eye Hospitals, Cairo, Egypt. The database of cases examined in the time interval between December 2013 and October 2018 by Allegro Oculyzer II (WaveLight, GmbH, Erlangen, Germany), using the current software version 1.20r20 equivalent to Pentacam HR, was searched for paediatric cases between 3 and 18 years old [16]. The candidates with bad scan quality, previous ocular trauma, surgeries, or corneal pathology other than ectasia were excluded. All retrievable data of significance from the remaining 432 normal paediatric right eyes and 48 eyes of paediatric KC (corneas with slit-lamp signs as Vogt striae or Fleisher ring or frank tomographic diagnosis) or forme fruste KC (FFKC) (fellow eye of a frank KC but with no abnormal findings by either slit-lamp examinations or corneal tomography) were collected. The study adhered to the tenets of the Declaration of Helsinki and was approved by the Ethics Committee of Watany Eye Hospitals and waived from consent taking as it is a retrospective study including no intervention other than the usual care offered.
The investigated indices were
-
A.
Anterior Chamber Parameters:
-
1.
Internal anterior chamber depth (ACD)
-
2.
Anterior chamber volume at 10-mm diameter (ACV)
-
3.
Maximum anterior chamber angle in the horizontal meridian (ACA)
-
B.
Curvature-based Indices:
-
1.
Index of height asymmetry (IHA)
-
2.
Index of height decentration (IHD)
-
3.
Index of surface variance (ISV)
-
4.
Index of vertical asymmetry (IVA)
-
5.
Central keratoconus index (CKI)
-
6.
Keratoconus index (KI)
-
7.
Mean curvature power of the cornea within the central 3 mm circle expressed in dioptres (K mean)
-
8.
Keratometric power of the flat meridian (K1)
-
9.
Keratometric power of the steep meridian (K2)
-
10.
Keratometry of the steepest point of the front surface (K max front)
-
11.
Steepest point of the front surface keratometry displacement in the x-axis (K max front x)
-
12.
Steepest point of the front surface keratometry displacement in the y-axis (K max front y)
-
13.
Keratometric astigmatism (absolute value)
-
14.
True net power at the corneal apex (TNP apex)
-
15.
Tilt by Fourier analysis (Tilt)
-
16.
Mean eccentricity in the central 30 degrees by Fourier analysis (Ecc Sph)
-
17.
High-order irregularity by Fourier analysis (other than spherical power and asphericity, tilt and astigmatism) (Irregularity)
-
C.
Elevation-based Indices:
-
1.
Front elevation from the best-fit sphere (FE from BFS)
-
2.
Front elevation from the best-fit toric ellipsoid (FE from BFTE)
-
3.
Back elevation from the best-fit sphere (BE from BFS)
-
4.
Back elevation from the best-fit toric ellipsoid (BE from BFTE)
-
D.
Pachymetry-based Indices:
-
1.
Corneal thickness at the apex (Pachy apex)
-
2.
Corneal thickness at the point corresponding to the pupil centre (Pachy pupil)
-
3.
Corneal thickness at the thinnest point (TP)
-
4.
Minimum pachymetric progression index (PPI min)
-
5.
Average pachymetric progression index (PPI avg)
-
6.
Maximum pachymetric progression index (PPI max)
-
7.
Thinnest point displacement at the y-axes (TP y)
-
8.
Thinnest point displacement at the x-axes (TP x)
-
E.
Combined Indices:
-
1.
Topographic keratoconus indices (TKC)
-
2.
Ambrosio’s relational thickness average (ART avg)
-
3.
Ambrosio’s relational thickness maximum (ART max)
-
4.
Corneal assessment index from the relational thickness and other OCULUS values analysed for 8 mm zone (CAIRO 8) [17]
The diagnostic accuracy of an index based on the area under the receiver operating characteristics (AUROC) was stratified as either excellent (>0.9), good (0.8 to 0.9), fair (0.7 to 0.8), or poor (0.6 to 0.7). An AUROC < 0.6 is considered a “failure” and should not be used for diagnosis [18].
Statistical analysis
Data were collected and verified, and the CAIRO 8 index was calculated. Statistical analyses were performed using MedCalc Statistical Software version 19.2.1 (MedCalc Software Ltd., Ostend, Belgium). All the retrievable data from patients fitting the inclusion and exclusion criteria gave a significantly larger sample size than that of most similar studies [19], and a post hoc analysis showed that the sample size was adequate to assess the accuracy (AUROC significantly different from the null hypothesis (0.5) for all indices of interest “PPI avg and max, ART avg and max, FE and BE from BFS [19] and most of the other indices. The following tests were performed: calculation of the mean, 95% confidence interval (CI), chi-square test, the AUROC and its binomial exact confidence interval, the AUROC comparison with the null hypothesis and with the AUROC of the most accurate index using the DeLong method [20], the best cut-off value, and the sensitivity and specificity for each index. P-values were considered statistically significant if less than 0.05.
Results
The study comprised 432 normal eyes, 8 FFKC eyes, and 40 frank KC eyes. KC and FFKC were more commonly prevalent in older age groups: the mean age was 13.5 years (95% CI: 13.2 to 13.8) in normal eyes and 15.3 years (14.7 to 15.9) in KC and FFKC eyes. There was no statistically significant sex predilection (Table 1).
The highest AUROC was that of CAIRO 8. The remaining 35 indices had numerically lower AUROCs, with 27 of them proving to be statistically inferior to CAIRO 8. Four indices had AUROC < 0.6 and failed to discriminate KC and FFKC from normal corneas (AUROC P-value >0.05): AC volume, AC angle, and the horizontal decentrations of the steepest point and of the thinnest point (Table 2 and Fig. 1). The 27 indices comprised 7 indices with excellent AUROC, 10 indices with good AUROC, 4 indices with fair AUROC, 2 indices with poor AUROC, and the previously mentioned 4 indices failing to discriminate the cases (Table 2 and Fig. 1). The sensitivities and specificities of all the tomographic indices are illustrated in Table 2 Fig. 2. Also, AUROC values of all the tomographic indices in patients below and above 12 years old were compared in Fig. 3.
Discussion
Paediatric KC is considered a dilemma in comparison with the adult disease, due to under-diagnosis, poor compliance with contact lens use, and modifications in treatment patterns [21].
Rotating Scheimpflug imaging is a useful tool for unveiling the disease in adults. However, there are not enough studies of paediatric KC [22] and there is a need to investigate the accuracy and the cut-off values of each index in KC diagnosis. Our study therefore evaluated these issues, revealing that the most useful rotating Scheimpflug imaging indices for KC detection in paediatric eyes were CAIRO 8, ART max and avg, PPI max and avg, BE BFTE, KI, TKC, and IHD respectively. On the other hand, four indices (ACV, ACA, K Max Front x, and TP x) did not show sufficient discriminative power between FFKC, KC, and normal corneas and can be excluded from the screening analysis.
The CAIRO 8 index is calculated by regression analysis that combines the highly sensitive ART max with the highly specific anterior elevation [17]. It was initially used for adult corneas and has now been validated in this study for paediatric corneas.
In the current study, where we analysed only data obtained from good quality scans with an analysable area of at least 9 mm2, analysis of elevation and pachymetry indices showed that CAIRO 8 had the highest accuracy (AUROC = 0.979) followed by ART max (AUROC = 0.967).
Our findings also support the importance of another combined index, namely ART max, which is calculated as the ratio between the thinnest point and the PPI max [23].
The results show that the pachymetric indices, namely PPI max and PPI avg, come next to the combined indices in terms of higher sensitivity in paediatric eyes (AUROC = 0.966 and 0.961, respectively). Further analysis of pachymetric indices revealed that thickness at TP had a higher AUROC than the pachy apex and pachy pupil. This can be attributed to the relative sensitivity to fixation of both pachy apex and pachy pupil due to the potential error of their localisation, especially in children [24]. On the other hand, the determination of the TP is not related to a certain location but rather done by scanning all available points and choosing the thinnest of them irrespective of its location. Likewise, the TP value had a higher AUROC than the AUROC of its vertical and horizontal decentrations (AUROC = 0.909, 0.621, and 0.553, respectively).
Regarding the elevation parameters, we concluded that the most accurate indices were BE BFTE (AUROC = 0.953) followed by BE, FE BFTE, and FE (AUROC = 0.947, 0.928, and 0.92, respectively). The elevations of the posterior surface, being the site of primary subclinical tomographic changes, have slightly better AUROCs than those of the anterior surface that could be partially neutralised by the epithelium [25,26,27,28].
In the paediatric population, curvature indices regain their diagnostic importance and compete with elevation indices [19, 29]. The KI, IHD, and TKC indices were among those with excellent AUROC (0.952, 0.947, and 0.942, respectively).
Although KC in the paediatric age was reported to be mainly central, the KI had higher AUROC than CKI, following the general rule in adult KC shown by Orucoglu et al. [30]. As shown by the high AUROC of IHD and KI in our study, vertical asymmetry seems prevalent in paediatric cases.
K max front was more accurate than its vertical and horizontal decentrations (AUROC = 0.877 versus 0.747, 0.558) and the K mean, (AUROC = 0.877 versus 0.727).
Aslankurt et al. compared 52 eyes of children with Down syndrome to 60 eyes of normal children regarding early topographic changes. They found that ISV had the highest sensitivity for subclinical KC (AUROC = 0.887) [24]. In the current study, although the ISV AUROC was 0.875, which is good and comparable to the results of Aslankurt, it was still significantly inferior to CAIRO 8 (P = 0.001). The ISV reflects the variation of the corneal radii of curvature compared with the mean value. Therefore, it is simply a non-specific expression of the surface curvature irregularity, including non-ectatic high-corneal astigmatism [18, 31].
In children, the corneal ectasia may have enough amplitude to affect the central ACD (AUROC = 0.704), but not yet enough widespread corneal involvement to affect the AC volume and the AC angle (0.582 and 0.561, respectively).
To the best of our knowledge, this is the first study to investigate the accuracy of different Pentacam indices in the detection of KC in a wide range of ages of paediatric eyes and with a relatively large sample size. The limitation of this study is its retrospective nature. Therefore, we recommend future prospective longitudinal studies for those indices. On the other hand, we also recommend further study of paediatric corneas biomechanics.
In conclusion, for diagnosis of paediatric KC or even FFKC, we recommend relying on some indices, namely CAIRO 8, ART max and avg, PPI max and avg, BFTE, KI, TKC, and IHD, rather than the other indices.
Summary
What was known before:
-
Pentacam is a reliable tool for early detection of KC in adults.
-
Paediatric age group has no well-established diagnostic gold-standard of corneal ectasia.
What this study adds:
-
The current study evaluated the accuracy of different tomographic indices of KC in paediatric eyes and found that Pentacam can accurately diagnose KC in paediatric age group
-
We recommended the use of the CAIRO 8, ART max and avg, PPI max and avg, BFTE, KI, TKC, and IHD indices.
Data availability
The datasets generated during and analysed during the current study are available from the corresponding author on reasonable request.
References
Loukovitis E, Kozeis N, Gatzioufas Z, Kozei A, Tsotridou E, Stoila M, et al. The proteins of keratoconus: a literature Review exploring their contribution to the pathophysiology of the disease. Adv Ther 2019;36:2205–22.
Khaled ML, Helwa I, Drewry M, Seremwe M, Estes A, Liu Y. Molecular and histopathological changes associated with keratoconus. Biomed Res Int. 2017;2017:7803029.
Olivo-Payne A, Abdala-Figuerola A, Hernandez-Bogantes E, Pedro-Aguilar L, Chan E, Godefrooij D. Optimal management of pediatric keratoconus: challenges and solutions. Clin Ophthalmol. 2019;13:1183–91.
Léoni-Mesplié S, Mortemousque B, Touboul D, Malet F, Praud D, Mesplié N, et al. Scalability and severity of keratoconus in children. Am J Ophthalmol. 2012;154:56–62.e1.
El-Khoury S, Abdelmassih Y, Hamade A, Slim E, Cherfan CG, Chelala E, et al. Pediatric keratoconus in a tertiary referral center: Incidence, presentation, risk factors, and treatment. J Refract Surg. 2016;32:534–41.
Moshirfar M, Heiland MB, Rosen DB, Ronquillo YC, Hoopes PC. Keratoconus screening in elementary school children. Ophthalmol Ther. 2019;8:367–71.
Soeters N, Van Der Valk R, Tahzib NG. Corneal cross-linking for treatment of progressive keratoconus in various age groups. J Refract Surg. 2014;30:454–60.
Omar IAN. Keratoconus screening among myopic children. Clin Ophthalmol. 2019;13:1909–12.
Salomão M, Hoffling-Lima AL, Lopes BT, Belin MW, Sena NB, Dawson DG, et al. Recent developments in keratoconus diagnosis. Expert Rev Ophthalmol. 2018;13:329–41.
Belin MW, Villavicencio OF, Ambrósio RR Jr. Tomographic parameters for the detection of keratoconus: suggestions for screening and treatment parameters. Eye Contact Lens. 2014;40:326–30.
Lopes BT, Ramos IC, Dawson DG, Belin MW, Ambrósio R Jr. Detection of ectatic corneal diseases based on pentacam. Z Med Phys. 2016;26:136–42.
Faria-Correia F, Ambrósio R. Clinical applications of the Scheimpflug principle in ophthalmology. Rev Bras Oftalmol. 2016;75:160–5.
Değirmenci C, Palamar M, İsmayilova N, Eğrilmez S, Yağcı A. Topographic evaluation of unilateral keratoconus patients. Turk J Ophthalmol. 2019;49:117–22.
Copeland RA, Afshari N. Principles and practice of cornea. Vol 1. Jaybee Brothers, India; 2013;1204.
Hashem AO, Roshdy MM, Wahba SS, Saleh MI, Elkitkat RS. Normative values of various pentacam HR parameters for pediatric corneas. Cornea. 2020;39:1394–400.
WHO (2007) Pediatric age categories to be used in differentiating between listing on a model essential medicines list for children. Available at http://archives.who.int/eml/expcom/children/Items/PositionPaperAgeGroups.pdf (Accessed 2 June 2020)
Roshdy MM, Wahba SS, Fikry RR. New corneal assessment index from the relational thickness and other OCULUS values (CAIRO Index). Clin Ophthalmol. 2018;12:1527–32.
Motlagh MN, Moshirfar M, Murri MS, Skanchy DF, Momeni-Moghaddam H, Ronquillo YC, et al. Pentacam corneal tomography for screening of refractive surgery candidates: a review of the literature, part I. Med Hypothesis Disco Innov Ophthalmol. 2019;8:177–203.
Wahba SS, Roshdy MM, Elkitkat RS, Naguib KM. Rotating Scheimpflug imaging indices in different grades of keratoconus. J Ophthalmol 2016;2016:6392472.
DeLong ER, DeLong DM, Clarke-Pearson DL. Comparing the areas under two or more receiver characteristic curves: a nonparametric approach. Biometrics. 1988;44:837–45.
Kankariya VP, Kymionis GD, Diakonis VF, Yoo SH. Management of pediatric keratoconus - evolving role of corneal collagen cross-linking: an update. Indian J Ophthalmol. 2013;61:435–40.
Matheus V, Arnaldo G, Camila Z, Bruna F, Rosane C, Andre O, et al. Corneal evaluation in healthy Brazilian children using a Scheimpflug topography system. J Clin Exp Ophthalmol. 2017;8;648–54
Ambrósio R, Caiado AL, Guerra FP, Louzada R, Sinha RA, Luz A, et al. Novel pachymetric parameters based on corneal tomography for diagnosing keratoconus. J Refract Surg. 2011;27:753–58. Epub 2011 Jul 29
Aslankurt M, Aslan L, Aksoy A, Kurt MM, ÖzdemIr M. Evaluation of early corneal topographic changes in children with Down syndrome. Turkish J Med Sci. 2013;43:810–14.
Labiris G, Giarmoukakis A, Gatzioufas Z, Sideroudi H, Kozobolis V, Seitz B. Diagnostic capacity of the keratoconus match index and keratoconus match probability in subclinical keratoconus. J Cataract Refract Surg. 2014;40:999–1005.
Chan TCY, Liu D, Yu M, Jhanji V. Longitudinal evaluation of posterior corneal elevation after laser refractive surgery using swept-source optical coherence tomography. Ophthalmology. 2015;122:687–92.
Wang B, Zhang Z, Naidu RK, Chu R, Dai J, Qu X, et al. Comparison of the change in posterior corneal elevation and corneal biomechanical parameters after small incision lenticule extraction and femtosecond laser-assisted LASIK for high myopia correction. Cont Lens Anterior Eye. 2016;39:191–6.
Saad A, Gatinel D. Screening for keratoconus: lack of a scientific base regarding the global consensus? Cornea. 2015;34:e33–5.
Epstein RL, Chiu YL, Epstein GL, Pentacam HR. criteria for curvature change in keratoconus and postoperative LASIK ectasia. J Refract Surg. 2012;28:890–4.
Orucoglu F, Toker E. Comparative analysis of anterior segment parameters in normal and keratoconus eyes generated by Scheimpflug tomography. J Ophthalmol. 2015;2015:925414.
Shetty R, Rao H, Khamar P, Sainani K, Vunnava K, Jayadev C, et al. Keratoconus screening indices and their diagnostic ability to distinguish normal from ectatic corneas. Am J Ophthalmol. 2017;181:140–8.
Acknowledgements
No financial grants or funding to be declared.
Funding
Open access funding provided by The Science, Technology & Innovation Funding Authority (STDF) in cooperation with The Egyptian Knowledge Bank (EKB).
Author information
Authors and Affiliations
Contributions
AOH was responsible for designing the review protocol, writing the protocol and report, conducting the search, screening potentially eligible studies, extracting and analysing data, interpreting results, updating reference lists. BFA was responsible for designing the review protocol and screening potentially eligible studies. He contributed to writing the report, extracting, and analysing data, interpreting results, and creating’Summary of findings’ tables. MMS conducted the statistical analyses and contributed to the design of the review protocol, writing the report, arbitrating potentially eligible studies, extracting, and analysing data and interpreting results. AIE contributed to data extraction and provided feedback on the report. SSW provided feedback on the report
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Hashem, A.O., Aziz, B.F., Wahba, S.S. et al. Diagnostic accuracy of different keratoconus detection indices of pentacam in paediatric eyes. Eye 37, 1130–1138 (2023). https://doi.org/10.1038/s41433-022-02070-x