Introduction

Glaucoma is a group of eye disorders that cause optic neuropathy with characteristic cupping of the optic nerve and visual field (VF) changes. If left untreated, glaucoma can lead to visual impairment and blindness. The Vision Loss Expert Group of the Global Burden of Disease Study has reported that glaucoma is the most common cause of irreversible blindness in the world together with age-related macular degeneration in their meta-analysis [1]. The only treatment proven successful to stop development of glaucoma and reduce progression of the disease is by lowering intraocular pressure (IOP) [2,3,4,5]. Once diagnosed, patients need to be on lifelong treatment and follow-up.

Monitoring of glaucoma requires measurement of IOP and assessment of functional and structural parameters of optic nerve. Any change in patients’ treatment plan depends on the level of IOP (and response to the medication) and deterioration of VF and/or optic nerve and/or structural assessments of the retina. However, all measurements are subjected to fluctuation and patients require frequent follow-up to determine whether a change has occurred.

A high IOP is a major risk factor for progression of glaucoma [6,7,8]. Measurement of IOP is usually taken in the clinic, but studies have shown more than 50–75% of IOP peaks occur outside clinic hours [9, 10]. IOP diurnal profile as well as peaks can be assessed by monitoring the IOP for 24 h but this needs the patients to be admitted for repeated measurements. This practice is time and resource consuming, which is rarely feasible in routine practice.

VF progression is usually detected by looking through a series of VF examinations obtained over a period of time, possibly several years or more [11]. It has been proposed that it would take 6 years to detect a relatively fast progression rate of 1 dB per year if only one VF test was performed per year [12]. Confirmation of deterioration can be reduced if repeat examinations are performed in a shorter period i.e., VF progression can be seen in 2 years if three examinations were performed each year [12]. In order to diagnose true disease progression from measurement variability inherent in the VF data, the authors recommended six VF tests to be performed in the first 2 years after diagnosis to help in determining rate of progression in a glaucoma patient.

Currently, patients need to travel to the hospital to have their IOP and VFs checked by the eye healthcare professionals. Despite the potential benefits of increased testing frequency, current healthcare systems do not have capacity for repeated, multiple testing within a year. In addition the need for frequent visits to hospital is also an inconvenience to patients and carers.

Innovations of medical devices may help overcome some of the monitoring challenges with frequent measurements of IOP and VF. Increasing availability of home monitoring devices may play an important role to detect uncontrolled or progressive disease, and also reassure patients that the glaucoma is well controlled. Home monitoring of IOP and VFs can give more data on measurements and help to better monitor the glaucoma.

Home intraocular pressure monitoring

Various devices has been designed to measure 24-h IOP such as integration of IOP sensors into intraocular lenses after cataract extraction surgery [13, 14], implantable IOP sensor placed either in the ciliary sulcus or the capsular bag [15, 16] and contact lens sensor [17]. However, the implantable devices have only been used in very few human studies with preliminary results, and would be available only in people undergoing intraocular surgery, while IOP readings of the contact lens sensor are in units rather than mmHg with no available conversion method yet.

A potential and practical option to measure 24-h IOP is to measure patients’ own IOP using a tonometer designed for self-tonometry. Rebound tonometry developed by Icare (Finland Oy) has been used since 2003 and has demonstrated to have high agreement with the standard tonometry, Goldmann applanation tonometry (GAT) [18]. These devices are easy to use and do not require the use of topical anesthesia making them potentially ideal for home monitoring. Rebound tonometry measures IOP using the impact/induction principle [19]. The IOP is quantified by measuring the deceleration of a magnetized disposable probe as it rebounds from the surface of the cornea [20].

A new Icare tonometry, Icare HOME (TA022; Icare Finland Oy) is currently available for home IOP monitoring (Fig. 1). Icare HOME is a handheld rebound tonometer that has an automatic-side recognition and positioning assistant for the correct alignment of the tonometer [21] compared with its predecessor. During measurement, Icare Home is placed at 4–8 mm from the cornea where the inbuilt software acquires six readings. The highest and lowest IOP readings are eliminated, leaving four readings which are be averaged and displayed on the device. If reliable readings are not obtained, an error bar is displayed and patients will need to repeat the measurements. The Icare HOME stores the final IOP with other information such as date, time, laterality of the eye, and measurement quality into its built-in memory module [22]. Icare HOME can be linked to the patient’s smartphone and data can be uploaded and retrieved from the cloud allowing remote access to the data. Home tonometry can be used by the patient or by another person.

Fig. 1
figure 1

Icare® HOME tonometer (TA022; Icare Finland Oy)

Icare HOME showed good repeatability in a study by Valero et al., who reported an intra-class correlation value of 0.872 of three repeated measurements by patients, although no coefficients of repeatability (more informative) were not estimated. There is moderate agreement between Icare HOME and GAT in glaucoma patients (Table 1) [20,21,22,23,24,25,26,27,28,29,30,31,32]. Most patients are able to measure their own IOP at home after given appropriate training and instruction [24, 26]. However, high intraindividual variation in diurnal IOP was seen across consecutive days as subjects exhibit different diurnal curve from day to day [27,28,29].

Table 1 Comparison between iCare HOME tonometry (by patients) and Goldmann applanation tonometry in adult patients with glaucoma. Study characteristics

Home visual field monitoring

Recently, there are several tablet-based devices and head-mounted displays that have been developed to measure VF away from clinical setting [23]. These devices could be potentially performed by patients without direct supervision of a trained eye care health professional with prior training making it possible for patients with glaucoma to monitor their VF at home. Tablet-based devices, such as Moorfields Motion Displacement Test (MMDT) and Melbourne Rapid Field (MRF), are currently available for use, while head-mounted displays, such as IMO [33] and Virtual Eye [34], are still in the preliminary stages of development.

Moorfields Motion Displacement Test (MMDT)

MMDT is computer-based software program for VFs testing in the community [35]. It is designed to run on a laptop computer. The test uses vertical white line stimuli at 31 test locations which spatially corresponds to Humphrey Visual Field Analyzer (HFA), Swedish interactive threshold algorithm (SITA) 24-2 [Carl Zeiss Meditec, Dublin, CA]) test locations. These stimuli are presented continuously throughout the test, against a gray background. The test locations were scaled by estimates of retinal ganglion cell density and selected using the Garway-Heath anatomical map [36, 37].

One eye is tested at a time using enhanced suprathreshold screening algorithm (ESTA) 99.5 program. ESTA 99.5 program presents displacements at the 95 centile, according to normative values derived in the UK from a population familiar with modern technology [35]. Patients place their head in a collapsible chin and forehead rest mounted on the laptop and set at 30 cm from the computer screen where the eyes can be aligned directly to a central white spot on the screen. Patients are instructed to look steadily at the central white spot throughout the test and press the computer mouse or spacebar every time they see a line move on the screen. Test is performed in a dim illumination room of approximately 85 lux. Average time need to complete the VF test using MMDT is about 5 min for both eyes. This test is designed to overcome the effects of cataract and uncorrected refractive error; therefore patients can perform MMDT without any refractive correction.

An index which measures the VF defect, the global probability of true damage (GPTD) can be calculated by summation of the probability of true damage derived from each of the 31 locations. The GPTD is expressed in quotient of 100 (0–100%), where higher GPTD values represent a greater probability of damage. GPTD value ranges from 0.31 to 31 [35,36,37,38]. Test reliability is determined by false-positive (FP) response rate. False positive is recorded when responses are <180 ms from the time the stimulus presented. False negative is usually not recorded as patients with glaucoma have fluctuating results and high false-negative results [38]. The recommended pass–fail and reliability criteria are a GPTD ≥ 3, and FP response rate ≤ 15% [35]. The test repeatability of the MMDT was reported as “good” by Loughman et al. using LoA plots and coefficient of repeatability [35].

The GPTD index had an AUC of 0.930 (95% CI [0.893–0.967]) for discriminating glaucoma from nonglaucomatous eyes with a sensitivity of 88.5% and specificity of 85% in a population-based setting [38]. Severity of glaucoma was mild in 17, moderate in 27, and advanced in 34 participants. It is difficult to compare with the standard HFA results as MMDT is a suprathreshold test and there are no similar global indices, such as mean deviation (MD) or pattern standard deviation. Test is easy to administer but may be challenging to some computer naive persons. This is probably due to strain in using the computer mouse as a response button for an extended period of the time during the test. However, performance between older, less educated computer naive subjects was similar to young computer-familiar subjects [35].

Melbourne Rapid Fields (MRF)

MRF (GLANCE Optical Pty Ltd, Melbourne, Australia) is a tangent perimeter application developed on iPad tablet platform (Apple, Cupertino, California, USA). As iPad tablets are portable and commonly available, it allows VF testing outside of clinic and in remote areas. Furthermore, it raises a possibility to be used by patients with glaucoma to perform VF tests at home unsupervised with minimal training [39,40,41,42,43].

Test should be performed in a quiet, dimly lit room with screen brightness automatically adjusted to maximal hardware brightness by the app at the start of the testing procedure. One eye is tested at a time and the fellow eye is patched. Patient is seated comfortably at a table with an iPad tablet placed on a typing stand accompanied by a Bluetooth keyboard. Patients view the 9.7-in. iPad tablet device (iPad 3 or later) at 33 cm wearing their normal reading glasses measured by a fixed piece of string from the iPad screen to bridge of patient’s nose at the start of the test. Head position is maintained throughout the test.

The testing grid consist of 66 test locations placed in a radial pattern covering 34° horizontal and 25° vertical of the VF. At the start of the test, blind spot will be located, and a stimulus will be presented about 8–10 times in this location throughout the test using central fixation. Testing will then occur in two phases: an initial central field test (36 points tested with fixation in the center of the screen) followed by a peripheral field test (30 test points) where patient fixate at each corner of the iPad screen sequentially to increase target eccentricity [39]. Location of the blind spot is not implemented during peripheral testing. FP and false-negative checks are presented throughout the test. FP checks are performed by using a interspacing periods (1000–1400 ms) throughout the test during which no stimulus is presented on the screen and recorded when patients gives a respond during this period [39].

A voice commands (in English language) generated by the tablet is provided to guide the patients throughout the test. Patients respond to the stimulus is recorded using the spacebar on the bluetooth keyboard. This method was selected because smear of finger marks on the screen can make the screen dirty and obscure a stimulus. It also provides a better tactile feedback to patients. Testing may take ~4–6 min, depending on the nature of the field loss and reliability of the patient. Average test duration for the MRF was shown to be statistically significant faster (5.7 ± 0.1 min) compared with the average HFA SITA Standard 24-2 test time (6.3 ± 0.1 min) [39].

MD and pattern deviation can be calculated from the MRF. MD was calculated from the average point-wise deviations using the age-adjusted expectation returned from the analysis of 17,390 thresholds adjusted for grid and size effect [39]. While, PD was calculated using a standard formulae as the average residual after allowing for the patient’s MD [39]. In terms of perimetric results, MD and PD results from MRF were in agreement and correlate with MD from HFA SITA Standard 24-2 (MD ICC = 0.93, PD ICC = 0.86) [40, 41]. Test–retest reliability was also shown to be comparable with HFA SITA Standard 24-2 with ICC of 0.93 for MD and 0.89 for PD. However, MRF recorded a statistically higher percentage of fixation losses (36 ± 4%) compared with the HFA (6 ± 1%) which may be due to free-space viewing corrupting the blind-spot monitor [40]. Bland-Altman analysis showed a small bias of 1.4 dB, with less negative MD returned by MRF [41]. In terms of scotoma detection using the cluster criteria the MRF on repeat testing detected 31/36 of abnormal hemifields in manifest glaucoma patients on the second test [44].

MRF looks promising as a portable tangent perimeter but further research is required before it can be widely implemented for home monitoring of VF. The current studies were conducted in a controlled environment with strict control of viewing distance and environment. Proper instructions and training to use the device unsupervised at home will be needed for the patients and to achieve reliable VF results.

It is common practice to use imaging technologies to monitor patients with glaucoma and detection of structural progression often influences clinical decisions. There are attempts to develop portable low-cost optical coherence tomography that, theoretically, could further support the goal of home monitoring [45, 46].

Conclusion

Detection of glaucoma progression requires regular IOP and VF monitoring in the eye outpatient clinics. However, these visits are time consuming, expensive, and difficult for patients. With the invention of newer “smart” technology, IOP devices, and VF tablet applications allows home monitoring of glaucoma providing abundant priceless data, which could provide a better understanding about the effects of treatment to patients and improve patients care. Home monitoring may also increase patients’ engagement and help in the patients’ adherence to their medication. Apart from that, home monitoring may also lead to more affordable diagnostic tests and available to individual with limited access to healthcare. Future research may be directed to validate the current applications and devices and invention of wearable device for home monitoring of glaucoma.