Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Article
  • Published:

Body composition, energy expenditure and physical activity

Lack of agreement of in vivo raw bioimpedance measurements obtained from two single and multi-frequency bioelectrical impedance devices

Abstract

Background

It is important for highly active individuals to accurately assess their hydration level. Bioelectrical impedance (BIA) can potentially meet these needs but its validity in active individuals is not well established.

Methods

We compared whole-body bioimpedance measurements obtained from multi-frequency bioelectrical impedance spectroscopy (BIS, Xitron 4200) at a 50 kHz frequency with those determined by a phase-sensitive single-frequency device (SF-BIA, BIA-101, RJL/Akern Systems) in two populations: active adults and elite athletes.

Results

One hundred twenty-six participants, including active males involved in recreational sports (N = 25, 20–39 yr) and elite athletes (females: N = 26, 18–35 yr; males: N = 75, 18–38 yr) participated in this study. Reactance (Xc), Resistance (R), Impedance (Z), and phase angle (PhA) were obtained by BIS and SF-BIA. Small but significant differences (R: −9.91 ± 15.09 Ω; Xc: −0.97 ± 2.56 Ω; Z: −9.96 ± 15.18 Ω; PhA: 0.12 ± 0.2°) were observed between the bioimpedance equipment in all measured variables (p < 0.05) though differences were within the devices’ technical error of measurements. Device-specific values were highly (p < 0.0001) correlated [R2 ranged from 0.881 (Xc) to 0.833 (R)], but slopes and intercepts were different (p < 0.0001) from 1 and 0, respectively. Relatively large limits of agreement were observed for R (−40 to 21 Ω), Xc (−6 to 4 Ω), PhA (−0.4 to 0.5°), and impedance (−40 to 20 Ω).

Conclusion:

Bioimpedance measurements from the current single- and multi-frequency devices should not be used interchangeably. The of lack of agreement between devices was observed in determining individual values of R, Xc, Z and PhA of highly active populations possibly due to methodological and biological factors.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Fig. 1

Similar content being viewed by others

References

  1. Gatterer H, Schenk K, Laninschegg L, Schlemmer P, Lukaski H, Burtscher M. Bioimpedance identifies body fluid loss after exercise in the heat: a pilot study with body cooling. PLoS One. 2014;9:e109729.

    Article  Google Scholar 

  2. Lukaski HC. Evolution of bioimpedance: a circuitous journey from estimation of physiological function to assessment of body composition and a return to clinical research. Eur J Clin Nutr. 2013;67(Suppl 1):S2–9.

    Article  Google Scholar 

  3. Earthman CP. Body composition tools for assessment of adult malnutrition at the bedside: a tutorial on research considerations and clinical applications. JPEN J Parenter Enter Nutr. 2015;39:787–822.

    Article  Google Scholar 

  4. Nescolarde L, Yanguas J, Terricabras J, Lukaski H, Alomar X, Rosell-Ferrer J, et al. Detection of muscle gap by L-BIA in muscle injuries: clinical prognosis. Physiol Meas. 2017;38:L1–L9.

    Article  CAS  Google Scholar 

  5. Nescolarde L, Yanguas J, Lukaski H, Rodas G, Rosell-Ferrer J. Localized BIA identifies structural and pathophysiological changes in soft tissue after post-traumatic injuries in soccer players. Conf Proc: Annu Int Conf IEEE Eng Med Biol Soc IEEE Eng Med Biol Soc Annu Conf. 2014;2014:3743–6.

    CAS  Google Scholar 

  6. Lukaski HC, Kyle UG, Kondrup J. Assessment of adult malnutrition and prognosis with bioelectrical impedance analysis: phase angle and impedance ratio. Curr Opin Clin Nutr Metab Care. 2017;20:330–9.

    Article  Google Scholar 

  7. XITRON. TECHNOLOGIES ISD, CA, USA. HYDRA ECF/ICF (Model 4200). Bio-Impedance Spectrum Analyzer. For measuring intracellular and extracellular fluid volumes. OPERATING MANUAL. REVISION 1.03. 2007;7–13.

  8. Gonzalez MC, Barbosa-Silva TG, Bielemann RM, Gallagher D, Heymsfield SB. Phase angle and its determinants in healthy subjects: influence of body composition. Am J Clin Nutr. 2016;103:712–6.

    Article  CAS  Google Scholar 

  9. Genton L, Norman K, Spoerri A, Pichard C, Karsegard VL, Herrmann FR, et al. Bioimpedance-derived phase angle and mortality among older people. Rejuvenation Res. 2017;20:118–24.

    Article  Google Scholar 

  10. Tanabe RF, de Azevedo ZM, Fonseca VM, Peixoto MV, dos Anjos LA, Gaspar-Elsas MI, et al. Distribution of bioelectrical impedance vector values in multi-ethnic infants and pre-school children. Clin Nutr. 2012;31:144–8.

    Article  Google Scholar 

  11. Food and Nutrition Technical Report. Human energy requirements. 2001;35–52.

  12. World Medical Association. Declaration of Helsinki - ethical principles for medical research involving human subjects. WMJ. 2008;54:122–5.

    Google Scholar 

  13. Carter JEL, Heath BH. Somatotyping: development and applications. Cambridge: Cambridge University Press; 1990.

    Google Scholar 

  14. Marfell-Jones M, Olds T, Stewart A, Carter JEL. International standards for anthropometric assessment. Sydney: UNSW Press; 2006.

    Google Scholar 

  15. Santos DA, Gobbo LA, Matias CN, Petroski EL, Goncalves EM, Cyrino ES, et al. Body composition in taller individuals using DXA: a validation study for athletic and non-athletic populations. J Sports Sci. 2013;31:405–13.

    Article  Google Scholar 

  16. Matias CN, Judice PB, Santos DA, Magalhaes JP, Minderico CS, Fields DA, et al. Suitability of bioelectrical based methods to assess water compartments in recreational and elite athletes. J Am Coll Nutr. 2016;35:413–21.

    Article  CAS  Google Scholar 

  17. Piccoli A, Rossi B, Pillon L, Bucciante G. A new method for monitoring body fluid variation by bioimpedance analysis: the RXc graph. Kidney Int. 1994;46:534–9.

    Article  CAS  Google Scholar 

  18. Piccoli A, Nigrelli S, Caberlotto A, Bottazzo S, Rossi B, Pillon L, et al. Bivariate normal values of the bioelectrical impedance vector in adult and elderly populations. Am J Clin Nutr. 1995;61:269–70.

    Article  CAS  Google Scholar 

  19. Lukaski HC, Johnson PE, Bolonchuk WW, Lykken GI. Assessment of fat-free mass using bioelectrical impedance measurements of the human body. Am J Clin Nutr. 1985;41:810–7.

    Article  CAS  Google Scholar 

  20. LaForgia J, Gunn S, Withers RT. Body composition: validity of segmental bioelectrical impedance analysis. Asia Pac J Clin Nutr. 2008;17:586–91.

    PubMed  Google Scholar 

  21. Van Loan M, Mayclin P. Bioelectrical impedance analysis: is it a reliable estimator of lean body mass and total body water. Human Biol; Int Rec Res. 1987;59:299–309.

    Google Scholar 

  22. van Marken Lichtenbelt WD, Westerterp KR, Wouters L, Luijendijk SC. Validation of bioelectrical-impedance measurements as a method to estimate body-water compartments. Am J Clin Nutr. 1994;60:159–66.

    Article  Google Scholar 

  23. Armstrong LE, Kenefick RW, Castellani JW, Riebe D, Kavouras SA, Kuznicki JT, et al. Bioimpedance spectroscopy technique: intra-, extracellular, and total body water. Med Sci Sports Exerc. 1997;29:1657–63.

    Article  CAS  Google Scholar 

  24. Ellis KJ, Wong WW. Human hydrometry: comparison of multifrequency bioelectrical impedance with 2H2O and bromine dilution. J Appl Physiol. 1998;85:1056–62.

    Article  CAS  Google Scholar 

  25. Moon JR, Tobkin SE, Roberts MD, Dalbo VJ, Kerksick CM, Bemben MG, et al. Total body water estimations in healthy men and women using bioimpedance spectroscopy: a deuterium oxide comparison. Nutr Metab (Lond). 2008;5:7.

    Article  Google Scholar 

  26. Piccoli A, Pastori G, Guizzo M, Rebeschini M, Naso A, Cascone C. Equivalence of information from single versus multiple frequency bioimpedance vector analysis in hemodialysis. Kidney Int. 2005;67:301–13.

    Article  Google Scholar 

  27. Piccoli A, Pastori G, Codognotto M, Paoli A. Equivalence of information from single frequency v. bioimpedance spectroscopy in bodybuilders. Br J Nutr. 2007;97:182–92.

    Article  CAS  Google Scholar 

  28. Ward LC, Byrne NM, Rutter K, Hennoste L, Hills AP, Cornish BH, et al. Reliability of multiple frequency bioelectrical impedance analysis: an intermachine comparison. Am J Human Biol: Off J Human Biol Counc. 1997;9:63–72.

    Article  CAS  Google Scholar 

  29. Bolton MP, Ward LC, Khan A, Campbell I, Nightingale P, Dewit O, et al. Sources of error in bioimpedance spectroscopy. Physiol Meas. 1998;19:235–45.

    Article  CAS  Google Scholar 

  30. Buendía RS, Fernando, Gil-Pita Roberto. Experimental validation of a method for removing the capacitive leakage artifact from electrical bioimpedance spectroscopy measurements. Meas Sci Technol. 2010;21:115802. https://doi.org/10.1088/0957-0233/21/11/115802.

    Article  Google Scholar 

  31. Heitmann BL, Swinburn BA, Carmichael H, Rowley K, Plank L, McDermott R, et al. Are there ethnic differences in the association between body weight and resistance, measured by bioelectrical impedance? Int J Obes. 1997;21:1085.

    Article  CAS  Google Scholar 

  32. Micheli ML, Pagani L, Marella M, Gulisano M, Piccoli A, Angelini F, et al. Bioimpedance and impedance vector patterns as predictors of league level in male soccer players. Int J Sports Physiol Perform. 2014;9:532–9.

    Article  Google Scholar 

  33. Hui D, Bansal S, Morgado M, Dev R, Chisholm G, Bruera E. Phase angle for prognostication of survival in patients with advanced cancer: preliminary findings. Cancer. 2014;120:2207–14.

    Article  Google Scholar 

  34. Brantlov S, Ward LC, Jodal L, Rittig S, Lange A. Critical factors and their impact on bioelectrical impedance analysis in children: a review. J Med Eng & Technol. 2017;41:22–35.

    Article  Google Scholar 

Download references

Acknowledgements

We express our gratitude to all the participants involved in this study.

Author contributions

AMS conceptualize and designed the study. DAS, CNM and CLN acquired the data, performed the data analysis and interpretation. EM, HL and LBS revised the manuscript critically for important intellectual content. All authors contributed to the final approval of the version to be submitted.

Funding

This work was supported by the Portuguese Foundation for Science and Technology (Grant: PTDC/DES/69495/2006 and PTDC/DES/098963/2008) and by European Hydration Institute. DAS is supported by a scholarship from the Portuguese Foundation for Science and Technology (grant: SFRH/BPD/92462/2013).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Analiza M. Silva.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Silva, A.M., Matias, C.N., Nunes, C.L. et al. Lack of agreement of in vivo raw bioimpedance measurements obtained from two single and multi-frequency bioelectrical impedance devices. Eur J Clin Nutr 73, 1077–1083 (2019). https://doi.org/10.1038/s41430-018-0355-z

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/s41430-018-0355-z

This article is cited by

Search

Quick links