Abstract
Microsphere-assisted microscopy utilizing a microsphere in immediate proximity of the specimen boosts the imaging resolution mainly as a result of an increase in the effective numerical aperture of the system.
Light microscopes are one of the most widely used tools for sample inspection in life and material sciences. Their achievable spatial resolution, however, is fundamentally limited to ~λ/(2 NA) due to the diffraction of light waves, in which λ is the wavelength of light and NA=n Sinθ is the numerical aperture of the system where θ is the maximum collection half-angle of light by the lens. Improving microscopy resolution beyond this limit or even enhancing the resolution of a given system to reach this limit without significant modifications would be of significant interest for various applications.
It has been shown in the past decade that micron-scale dielectric spheres and cylinders can be used as supplementary lenses to improve microscopy resolution, a technique termed “microsphere-assisted microscopy” (MAM)1. MAM is a simple, yet efficient approach in which a microsphere (MS) is placed in the immediate vicinity of a specimen, as schematically shown in Fig. 1a. Focusing the objective lens on the virtual image formed underneath the specimen by the MS enables microscopy with higher resolution and magnification compared to using the objective lens without the MS (θ2 > θ1 in Fig. 1a). MAM can be performed with low-index (n~1.5) MSs2 as well as high-index (n > 1.9) MSs3 placed in a background medium with a lower refractive index than that of the microsphere4. The latter approach would allow fabrication of novel resolution-improving optical devices5. MAM is a versatile technique and can be incorporated with different microscopes, such as fluorescent6,7, confocal8, and two-photon9 setups. MAM can also be extended to interference and digital holographic microscopies for 3D imaging with enhanced resolution. This can be simply achieved in an interferometric arrangement, by placing the MS within the working distance of either a conventional objective lens10,11,12,13,14,15 or a dedicated interferometric objective lens (Fig. 1b)16,17.
MAM has well matched up with interferometric-based approaches, such that both have benefited from each other’s features. Although interferometric arrangements have significantly higher axial resolution compared to conventional microscopy, their lateral resolution would still be restricted by the diffraction limit. However, MAM combined with interferometric setups has shown promising results in producing 3D label-free images with enhanced lateral resolution. Additionally, MAM has shown its potential to be easily integrated with other resolution-enhancing schemes, such as structured and oblique illumination, to achieve even further resolution enhancements in 3D microscopy systems18,19. Microsphere-assisted interference microscopy provides an inexpensive and non-destructive imaging technique for 3D surface metrology17,20 and microsphere-assisted digital holographic microscopy (DHM) brings forward a cost-effective and easy-to-implement high-quality quantitative phase imaging approach in transmission and reflection modes21. The advantages of MAM over other resolution enhancement techniques for DHM are further featured;22,23 in MAM, due to the symmetric structure of the MS, the resolution enhancement is achieved simultaneously in all spatial directions, which is extremely important in real-time imaging of specimens with rapid dynamics12,24. On the other hand, MAM benefits from DHM’s advantageous features. Adding a spherical lens to a conventional microscope introduces non-negligible spherical aberrations1. It also brings about a curved deformation in the measured phase in interference microscopy, which in turn further sacrifices the effective field-of-view (FoV)13. However, in DHM, thanks to the phase compensation possibility, the effect of any possible contaminations and aberrations in the optical train on the final data can be removed. To this end, phase information of the reference hologram (taken for no-sample state in the setup) is subtracted from the object phase during the numerical reconstruction process25. Another challenging issue in MAM that can be less important in microsphere-assisted DHM is precise positioning of the MS and in some cases preserving the relative sample-to-microsphere distance26 during the imaging. However, in DHM, it is consistently possible to be numerically focused on the image plane post-recording of the digital hologram to achieve the best possible imaging quality27. Nevertheless, similar to all kinds of MAM, microsphere-assisted interferometric-based approaches also suffer from the limited FoV, which still needs to be addressed in a sustainable manner.
The open questions and challenges in MAM have been listed in a recent review1; among them were the exact mechanism for resolution enhancement and a robust measurement of the achieved resolution. The enhancement of the numerical aperture was listed as the main contributing factor in resolution enhancement in MAM. Other factors included evanescent wave collection, photonic nanojet effects, resonant effects, and substrate and specimen-specific effects.
In an article published in this issue28, the investigation team performed numerical simulation using finite element method (FEM) to investigate the resolution enhancement mechanism in a microsphere-enhanced interferometry setup29,30. Although previous simulation works31,32 have been done, their model involves rigorous treatment of light scattering at the specimen’s surface and modeling the high NA objective lenses used for illumination and imaging. They considered full 3D conical Kӧhler illumination as well as conical imaging of the scattered waves by the MS. The resolution enhancement and magnification were studied with respect to the NA of the objective lens. Their case study was done for a 5-μm-diameter sphere with refractive index of 1.5 placed on a silicon sinusoidal phase grating with 25 nm peak-to-valley amplitude and 13.2 μm period. A 100× (0.9 NA) objective lens and monochromatic light at a non-resonant (440 nm) and resonant (480.76 nm) wavelength was considered; more simulation detail is provided in their previous work33. Their results confirmed that the enhancement of the NA is the main factor in resolution enhancement in MAM. No significant difference between the resonant and non-resonant wavelengths was observed.
The proposed method can be extended to conventional and confocal microscopies to help better understand and shed more light on MAM. Their approach can also be used to investigate parameters influencing MAM and find the optimum MS in terms of the refractive index, size, and index of the background medium34 for a given application.
References
Darafsheh, A. Microsphere-assisted microscopy. J. Appl. Phys. 131, 031102 (2022).
Wang, Z. B. et al. Optical virtual imaging at 50 nm lateral resolution with a white-light nanoscope. Nat. Commun. 2, 218 (2011).
Darafsheh, A. et al. Optical super-resolution by high-index liquid-immersed microspheres. Appl. Phys. Lett. 101, 141128 (2012).
Darafsheh, A. Optical super-resolution and periodical focusing effects by dielectric microspheres. PhD thesis, University of North Carolina at Charlotte, Charlotte (2013).
Darafsheh, A. Photonic nanojets and their applications. J. Phys. Photonics 3, 022001 (2021).
Yang, H. et al. Super-resolution biological microscopy using virtual imaging by a microsphere nanoscope. Small 10, 1712–1718 (2014).
Darafsheh, A. et al. Optical super-resolution imaging by high-index microspheres embedded in elastomers. Opt. Lett. 40, 5–8 (2015).
Darafsheh, A. et al. Advantages of microsphere-assisted super-resolution imaging technique over solid immersion lens and confocal microscopies. Appl. Phys. Lett. 104, 061117 (2014).
Forouhesh Tehrani, K. et al. Resolution enhancement of 2-photon microscopy using high-refractive index microspheres. Proc. SPIE 10498, 1049833 (2018).
O’Connor, T., Anand, A. & Javidi, B. Field-portable microsphere-assisted high resolution digital holographic microscopy in compact and 3D-printed Mach-Zehnder Interferometer. OSA Contin. 3, 1013–1020 (2020).
Wang, W. C. et al. Microsphere-assisted Fabry–Perot interferometry: proof of concept. Appl. Opt. 61, 5442–5448 (2022).
Wang, Y. X. et al. Resolution enhancement phase-contrast imaging by microsphere digital holography. Opt. Commun. 366, 81–87 (2016).
Perrin, S. et al. Microsphere-assisted phase-shifting profilometry. Appl. Opt. 56, 7249–7255 (2017).
Abbasian, V., Rasouli, S. & Moradi, A. R. Microsphere-assisted self-referencing digital holographic microscopy in transmission mode. J. Opt. 21, 045301 (2019).
Wang, F. F. et al. Three-dimensional super-resolution morphology by near-field assisted white-light interferometry. Sci. Rep. 6, 24703 (2016).
Aakhte, M. et al. Microsphere-assisted super-resolved Mirau digital holographic microscopy for cell identification. Appl. Opt. 56, D8–D13 (2017).
Kassamakov, I. et al. 3D super-resolution optical profiling using microsphere enhanced mirau interferometry. Sci. Rep. 7, 3683 (2017).
Xie, Z. Y. et al. 3D super-resolution reconstruction using microsphere-assisted structured illumination microscopy. IEEE Photonics Technol. Lett. 31, 1783–1786 (2019).
Abbasian, V. et al. Super-resolved microsphere-assisted Mirau digital holography by oblique illumination. J. Opt. 20, 065301 (2018).
Hüser, L. & Lehmann, P. Microsphere-assisted interferometry with high numerical apertures for 3D topography measurements. Appl. Opt. 59, 1695–1702 (2020).
Abbasian, V. et al. Digital holographic microscopy for 3D surface characterization of polymeric nanocomposites. Ultramicroscopy 185, 72–80 (2018).
Micó, V. et al. Resolution enhancement in quantitative phase microscopy. Adv. Opt. Photonics 11, 135–214 (2019).
Gao, P. & Yuan, C. J. Resolution enhancement of digital holographic microscopy via synthetic aperture: a review. Light Adv. Manuf. 3, 105–120 (2022).
Chen, L. W. et al. Microsphere enhanced optical imaging and patterning: from physics to applications. Appl. Phys. Rev. 6, 021304 (2019).
Anand, A., Chhaniwal, V. K. & Javidi, B. Real-time digital holographic microscopy for phase contrast 3D imaging of dynamic phenomena. J. Disp. Technol. 6, 500–505 (2010).
Darafsheh, A. Comment on ‘Super-resolution microscopy by movable thin-films with embedded microspheres: resolution analysis’ [Ann. Phys. (Berlin) 527, 513 (2015)]. Ann. Phys. 528, 898–900 (2016).
Kim, M. K. Digital Holographic Microscopy: Principles, Techniques, and Applications (Springer, 2011).
Pahl, T. et al. FEM-based modeling of microsphere-enhanced interferometry. Light Adv. Manuf. 3, 49 (2022).
Hüser, L. & Lehmann, P. Microsphere-assisted interference microscopy for resolution enhancement. tm. - Technisches Mess. 88, 311–318 (2021).
Hüser, L. et al. Microsphere assistance in interference microscopy with high numerical aperture objective lenses. J. Optical Microsyst. 2, 044501 (2022).
Sundaram, V. M. & Wen, S. B. Analysis of deep sub-micron resolution in microsphere based imaging. Appl. Phys. Lett. 105, 204102 (2014).
Hoang, T. X. et al. Focusing and imaging in microsphere-based microscopy. Opt. Express 23, 12337–12353 (2015).
Pahl, T. et al. 3D modeling of coherence scanning interferometry on 2D surfaces using FEM. Opt. Express 28, 39807–39826 (2020).
Darafsheh, A. Influence of the background medium on imaging performance of microsphere-assisted super-resolution microscopy. Opt. Lett. 42, 735–738 (2017).
Acknowledgements
We would like to dedicate this work to the memory of Kian Pirfalak (2013–2022), a student from Izeh, Iran, who was passionate to become an engineer.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Darafsheh, A., Abbasian, V. Dielectric microspheres enhance microscopy resolution mainly due to increasing the effective numerical aperture. Light Sci Appl 12, 22 (2023). https://doi.org/10.1038/s41377-022-01056-4
Published:
DOI: https://doi.org/10.1038/s41377-022-01056-4
This article is cited by
-
A dataset of digital holograms of normal and thalassemic cells
Scientific Data (2024)