Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Article
  • Published:

Transcutaneous bilirubinometer use and practices surrounding jaundice in 150 California newborn intensive care units

Abstract

Objectives

Transcutaneous bilirubin measurements (TcBs) provide a noninvasive method for screening infants for hyperbilirubinemia and have been used extensively in term and late preterm newborns in well baby nurseries, offices, and outpatient clinics. Several studies have also demonstrated the utility of TcBs as a screening tool for infants > 28 weeks’ gestation and their ability to reduce the need for blood sampling. The objectives of this study are to identify how often TcBs are used among California Newborn Intensive Care Units (NICUs) in preterm, late preterm and term infants, and other aspects of jaundice management.

Methods

We conducted a survey on TcB use and practices relating to jaundice management in 150 California NICUs between April and October 2016.

Results

TcB screening is routinely used in 28% (42/150) of NICUs. Only 7% (11/150) of NICUs use TcB in preterm infants < 28 weeks. Practice varied similarly across NICU levels of care. Among the subset of NICUs that responded to questions related to phototherapy and screening practices, prophylactic phototherapy was used in 38% (23/59) and 90% (55/61) screened for glucose-6-phosphate dehydrogenase deficiency based on race, ethnicity, and/or family history.

Conclusion(s)

Despite studies validating the accuracy of TcB in preterm infants > 28 weeks, only 28% of California NICUs routinely use TcB devices. TcB screening in infants < 28 weeks gestation is not widely used and no recommendation can be made in this regard until there is more experience with its application using a standardized protocol in these infants and on a large scale.

This is a preview of subscription content, access via your institution

Access options

Rent or buy this article

Prices vary by article type

from$1.95

to$39.95

Prices may be subject to local taxes which are calculated during checkout

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Engle WD, Jackson GL, Engle NG. Transcutaneous bilirubinometry. Semin Perinatol. 2014;38:438–51.

    Article  Google Scholar 

  2. Wickremasinghe AC, Karon BS, Saenger AK, Cook WJ. Effect of universal neonatal transcutaneous bilirubin screening on blood draws for bilirubin analysis and phototherapy usage. J Perinatol. 2012;32:851–5.

    Article  CAS  Google Scholar 

  3. Quist FK, Bapat R, Kuch-Kunich HK, Ezeanoule K, Keeni S, Thomas R, et al. Clinical utility of transcutaneous bilirubinometer (TcB) in very low birth weight (VLBW) infants. J Perinat Med. 2016;44:933–9.

    Article  CAS  Google Scholar 

  4. Wainer S, Parmar SM, Allegro D, Rabi Y, Lyon ME. Impact of a transcutaneous bilirubinometry program on resource utilization and severe hyperbilirubinemia. Pediatrics. 2012;129:77–86.

    Article  Google Scholar 

  5. American Academy of Pediatrics Subcommittee on H. Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics. 2004;114:297–316.

    Article  Google Scholar 

  6. Maisels MJ, Bhutani VK, Bogen D, Newman TB, Stark AR, Watchko JF. Hyperbilirubinemia in the newborn infant > or = 35 weeks’ gestation: an update with clarifications. Pediatrics. 2009;124:1193–8.

    Article  Google Scholar 

  7. Maisels MJ, Watchko JF, Bhutani VK, Stevenson DK. An approach to the management of hyperbilirubinemia in the preterm infant less than 35 weeks of gestation. J Perinatol. 2012;32:660–4.

    Article  CAS  Google Scholar 

  8. Bhutani VK, Gourley GR, Adler S, Kreamer B, Dalin C, Johnson LH. Noninvasive measurement of total serum bilirubin in a multiracial predischarge newborn population to assess the risk of severe hyperbilirubinemia. Pediatrics. 2000;106:E17.

    Article  CAS  Google Scholar 

  9. Rubaltelli FF, Gourley GR, Loskamp N, Modi N, Roth-Kleiner M, Sender A, et al. Transcutaneous bilirubin measurement: a multicenter evaluation of a new device. Pediatrics. 2001;107:1264–71.

    Article  CAS  Google Scholar 

  10. Nagar G, Vandermeer B, Campbell S, Kumar M. Reliability of transcutaneous bilirubin devices in preterm infants: a systematic review. Pediatrics. 2013;132:871–81.

    Article  Google Scholar 

  11. Chawla D, Jain S, Kaur G, Sinhmar V, Guglani V. Accuracy of transcutaneous bilirubin measurement in preterm low-birth-weight neonates. Eur J Pediatr. 2014;173:173–9.

    Article  CAS  Google Scholar 

  12. Murli L, Thukral A, Sankar MJ, Vishnubhatia S, Deorari AK, Paul VK, et al. Reliability of transcutaneous bilirubinometry from shielded skin in neonates receiving phototherapy: a prospective cohort study. J Perinatol. 2017;37:182–7.

    Article  CAS  Google Scholar 

  13. Nagar G, Vandermeer B, Campbell S, Kumar M. Effect of phototherapy on the reliability of transcutaneous bilirubin devices in term and near-term infants: a systematic review and meta-analysis. Neonatology. 2016;109:203–12.

    Article  CAS  Google Scholar 

  14. American Academy of Pediatrics Committee on F, Newborn. Levels of neonatal care. Pediatrics. 2012;130:587–97.

    Article  Google Scholar 

  15. Maisels MJ, Coffey MP, Kring E. Transcutaneous bilirubin levels in newborns < 35 weeks’ gestation. J Perinatol. 2015;35:739–44.

    Article  CAS  Google Scholar 

  16. Rubio A, Epiard C, Gebus M, Deiber M, Samperiz S, Genty C, et al. Diagnosis accuracy of transcutaneous bilirubinometry in very preterm newborns. Neonatology. 2017;111:1–7.

    Article  CAS  Google Scholar 

  17. Schmidt ET, Wheeler CA, Jackson GL, Engle WD. Evaluation of transcutaneous bilirubinometry in preterm neonates. J Perinatol. 2009;29:564–9.

    Article  CAS  Google Scholar 

  18. Maisels MJ. Transcutaneous bilirubin measurement: does it work in the real world? Pediatrics. 2015;135:364–6.

    Article  Google Scholar 

Download references

Acknowledgements

We want to thank all 150 NICU Directors of California for their support. We want to thank CAN (California Association of Neonatologists) for their support in completing the survey.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Dilip R. Bhatt.

Ethics declarations

Conflict of interest

M. Jeffrey Maisels is a consultant to Draeger Inc (Telford, PA), the supplier of the JM-103 and JM-105 transcutaneous bilirubinometers. The remaining authors declare no conflicts of interest.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bhatt, D.R., Kristensen-Cabrera, A.I., Lee, H.C. et al. Transcutaneous bilirubinometer use and practices surrounding jaundice in 150 California newborn intensive care units. J Perinatol 38, 1532–1535 (2018). https://doi.org/10.1038/s41372-018-0154-3

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/s41372-018-0154-3

This article is cited by

Search

Quick links