Introduction

Accurate blood pressure measurement is the cornerstone of the diagnosis and treatment of hypertension. Contrary to expectation, the most preferable method cannot be settled solely because it greatly depends on the condition of the patient, the environment of the measurement space, and the measurement device. Furthermore, a combination of blood pressure information as well as regular, continual measurement of blood pressure is also fundamental for appropriate management of blood pressure. In this review, we provide an overview of the current topics on blood pressure measurement, including automated office blood pressure (AOBP) and validation of blood pressure measurement devices, in relation to the recommendations in the most recent guidelines [1,2,3,4] on hypertension.

Conventional “in-office” blood pressure

Conventional blood pressure, which is measured at an office or in a screening setting by a sphygmomanometer, has been used as the gold standard of blood pressure measurement techniques regardless of the guidelines based on cumulative evidence and common use in clinical practice. Although variations in measurement conditions still exist, office blood pressure should be measured after a few minutes [1]—3 [4], 5 [2], or >5 [3] min—of rest, twice [1, 3] or three times [2, 4] and then these values should be averaged with special conditions, as summarized in Table 1.

Table 1 Office blood pressure measurement conditions using an automated device.

AOBP was initially defined as office blood pressure measured by an automated device for multiple readings (three times or more [5]), recorded automatically with the patient under rest in an undisturbed quiet room in the absence of an observer [5, 6]. Five minutes of rest in the unattended setting before AOBP measurement has been further used in many recent studies (Table 1) [7,8,9]. Stults et al. raised concern about inhabitation of AOBP utilization in real-world settings [10]; the AOBP including antecedent rest time requires >10 min per person, which includes a few minutes for preparation and clean up. The Hypertension Canada 2018 Guidelines stated that AOBP may be performed with no special period of rest [11]. However, as we discussed in detail elsewhere [8], this statement can be applied almost exclusively for measurement using BpTRU (BpTRU Medical Devices, Coquitlam, Canada; the company ceased operations in 2017) which usually performs six consecutive readings and averages two to six readings. The discard of the first reading serves as the antecedent rest time, and it is fair to prepare a certain rest time before the first measurement—5 min would be acceptable regarding conventional measurement—when we average all the measurement values. Whereas, a conventional office blood pressure measurement also requires more than a few minutesʼ antecedent rest when faithfully based on recent guidelines [1,2,3,4]. Although the antecedent rest time might not be mandatory when systolic blood pressure level is <130 mmHg [12], the amendment of the measurement protocol according to the estimated values makes it difficult to evaluate and compare the measurements among individuals. Furthermore, the consumption of office space in addition to long measurement time reduces the feasibility of AOBP [8]. We need at least one extra separate room with a person in charge to manage the AOBP measurement and the cost of the measurement is generally not reimbursed [8].

Recent studies reported that average levels of AOBP were similar to the 24-h ambulatory blood pressure in a population [9, 13]. However, we cannot conclude that AOBP can alter the out-of-office measurement because the variation in differences by individuals should not be ignored. Our research group recently reported that there was a marked variation in individuals between self-measured home blood pressure and AOBP values, while average levels are almost identical, as systolic blood pressure difference was 0.9 mmHg, with a 95% agreement limit of −34.0 to 35.8 mmHg (home blood pressure was measured in the morning; adjusted correlation coefficients, 0.14; 95% confidence intervals (CI), 0.03–0.25) [8]. These low correlations and wide range of difference indicate that we cannot estimate out-of-office blood pressure values by AOBP measurements in individuals and vice versa, and AOBP cannot be used as an alternative to home blood pressure [8].

Besides the white-coat observer effect by the presence of medical staff [6], AOBP measurement is performed at a clinic or in a screening setting where participants cannot be fully relaxed like at their own home. Such a clinic effect still affects in-office readings even based on AOBP [8]. Home blood pressure has no clinic effect regardless of self-measurement or observer measurement; however, participants can be affected by the observer effect during the latter observer measurement of home blood pressure. The environmental condition can be cross-classified as shown in Fig. 1 [14]. Meanwhile, blood pressure might be increased by arousal or by arm-cuff inflation [15]. In this respect, nighttime ambulatory blood pressure values during sleep, when measurements are free from arousal and alerting reactions, would have great prognostic ability for cardiovascular complications [16]. Much still remains to be done for the assessment of the wide variety of blood pressure conditions.

Fig. 1: Cross-classification of blood pressure measurement conditions and effects.
figure 1

Out-of-office blood pressure measurements are presented in the left panels (A, C), with in-office measurements in the right panels (B, D). During the measurements, medical staff are in attendance (A, B) during home visit/home healthcare/surveillance (A) or not in attendance (C, D). Out-of-office blood pressure includes, but is not limited to, home measurements. Ambulatory blood pressure monitoring (overlaid domain) is performed at various locations, mostly out-of-office settings. For automated office blood pressure (D), medical staff are not in attendance but supervise the measurement procedure. Reproduced from Asayama et al. [14].

Ambulatory and home blood pressure

Self-measured home blood pressure and ambulatory blood pressure monitoring are categorized as out-of-office blood pressure, and they have more prognostic ability than conventional in-office blood pressure [16,17,18]. Beyond its greater predictive power, home blood pressure can substantially refine conventional in-office blood pressure-based risk stratification, particularly among those whose in-office blood pressure levels are assumed to carry no or tiny increased cardiovascular risk [19]; the utility of risk stratification was also reported based on ambulatory blood pressure values [20]. Our research group recently performed participant-level meta-analysis based on 11,135 adults from Europe, Asia, and South America who measured 24-h ambulatory blood pressure [16]. Per 20 mmHg nighttime systolic blood pressure increments, the multivariable-adjusted hazard ratio for total mortality was 1.23 (95% CI, 1.17–1.28) and for cardiovascular events was 1.36 (95% CI, 1.30–1.43). Results were essentially similar when 24-h ambulatory measurements were assessed. Furthermore, even after adjustment for any of the other systolic blood pressure indexes, e.g., daytime blood pressure, or in-office conventional blood pressure or AOBP, both 24-h and nighttime systolic blood pressures were significantly associated with mortality and composite cardiovascular outcome. Although the usefulness of in-office blood pressure should not be underestimated with respect to identifying white-coat and masked hypertension [19, 21], the usefulness of out-of-office blood pressure as a single source of information would be unwavering. Meanwhile, it is fair to note that we defined AOBP in that analysis [16] as the mean of the ambulatory recordings during the first recording hour because the monitors in that period were applied in a medical environment. The measurement did not follow the aforementioned AOBP criteria, and there is room for investigating whether ambulatory blood pressure is superior to AOBP under the defined measurement condition [5, 6].

Based on measurements under well-defined conditions, we reported the current status of self-measured home blood pressure among Japanese five cohorts [22]. According to the recent Japanese guidelines (JSH 2019) [1], participants in the Modern DAtabase on Self-measured home blood pressure (MDAS) measured morning home blood pressure within 1 h of waking before breakfast and taking antihypertensive medication if any, after urination, and after a few minutes of rest in a sitting position. They measured evening home blood pressure just before going to bed after a few minutes of rest in a sitting position. The measurement period was 2001–2018, and they used two home blood pressure measurement devices made using the same measurement technology (HEM-705IT and HEM-7080IC, Omron Healthcare Co. Ltd, Kyoto, Japan) except the earlier periods of Ohasama (2001–2017) in which another validated device from the same manufacturer (HEM-747IC-N) was used. Aggregate summary data per classified cell of sex, age group in 5-year intervals, and antihypertensive drug treatment were collected and integrated, and mean blood pressure values in each cell were calculated by a meta-analysis approach. Morning home blood pressure was consistently higher than evening blood pressure regardless of age group, sex, and antihypertensive treatment (Fig. 2), and that among participants receiving antihypertensive medication was higher than those without medication. The average diastolic blood pressure was almost identical among those ≥80 years of age, regardless of sex and antihypertensive treatment status, in both the morning (73.7–75.8 mmHg) and evening (67.1–68.9 mmHg) measurements. Home measurement values were grossly homogenous among the study population by heterogeneity tests, suggesting sufficient representativeness of these data on recent home blood pressure levels among Japanese.

Fig. 2: The self-measured home blood pressure among Japanese five cohorts in each 5-year age category.
figure 2

The mean systolic (A, B) and diastolic (C, D) home blood pressure measured in the morning (A, C) and evening (B, D). Error bars indicate one-sided 95% confidence intervals. The horizontal auxiliary lines, which represent blood pressure level, are aligned in height on the adjacent panels. Heterogeneity tests among the cohorts were not significant (corrected P ≥ 0.063), with the exception of treated men aged 70–74 years (corrected P = 0.0030) in the morning measurement and of untreated men aged 60–64 years (corrected P = 0.0091) and 75–79 years (corrected P = 0.033). Reproduced from Asayama et al. [22].

Measurement devices

As marketed medical tools, manufacturers of the blood pressure measurement devices have to follow domestic standards and obtain approval from governmental organizations in each country [23]. In the majority of cases, however, blood pressure is measured under diverse clinical settings and such government agencies have little responsibility for the validation of accuracy in each setting. Validation studies are therefore provided according to the measurement settings, including age group, anthropometric condition, and with consideration of those having a special condition such as pregnancy, children, or treated by hemodialysis. For example, the MDAS coworkers performed a validation study of their device for a population aged around 85 years old; this resulted in the first publication for guaranteeing a home blood pressure measurement device for oldest-olds [24]. The Association for the Advancement of Medical Instrumentation (AAMI) and American National Standards Institute (ANSI) revised the ANSI/AAMI standard to adopt International Organization for Standardization (ISO) standards (ISO 81060-2:2013) [25], and the European Society of Hypertension (ESH) [26] recently followed the updated ISO standards (ISO 81060-2:2019) [27]. It should be noted that the ISO standards have been published under the full responsibility of the ISO committees, although these recent standards are sometimes called by different name in different organizations which contributed to the standards, e.g., ANSI/AAMI/ISO [28] or AAMI/ESH/ISO [26].

According to a recent report, <15% of blood pressure measurement devices have undergone independent validation testing among >3300 devices worldwide from >450 manufacturers [29]. In the online marketplace available from Australia on December 2018, only 18.3% of the 278 unique upper arm-cuff home blood pressure measurement devices for sale had passed international validation protocols, and the prevalence of validated wristband-wearable devices among 532 unique ones on the market was zero [30]. Several consortiums have tried addressing the blood pressure validation issues. The Accuracy in Measurement of Blood Pressure (AIM-BP) Collaborative is focused on advocating for and implementing optimal blood pressure measurement practice globally by complimentary existing efforts [31]. The activity of AIM-BP is in connection with the Lancet Commission on Hypertension Group [32], the World Hypertension League, and related organizations [33], and major objectives of the Collaborative include “host a webpage on the World Hypertension League website that summarizes critical information and resources” [31]. Such summarization of blood pressure measurement information has been conducted by several organizations, e.g., dabl Educational Trust [34] and Medaval [35]. Despite the great efforts, people have been facing difficulty providing information continuously as dabl Educational Trust is no longer actively updated [29], and to make and update the summary under a fair and accurate fashion is also tough partly because interpretation of publications regarding validation of blood pressure device can differ. In the Science and Technology for Regional Innovation and Development in Europe BP initiative which was founded in 2018 by international experts in blood pressure measurement, the board members independently review publications on blood pressure variability based on the concurrent knowledge and advancement of validation procedures, and they judge whether the device is acceptable for defined measurement situation such as in the general population, children, etc. [36]. In contrast, our group organized a working group within the Japanese Society of Hypertension (JSH), and, in 2016, started to collect information on blood pressure measurement devices marketed in Japan (http://www.jpnsh.jp/com_ac_wg1.html). In this listing updated annually, we refrain from scoring the devices; instead, we ask the device manufacturers via several public consortiums to provide related data in detail, and we disseminate the information without essentially processing it. Although this activity is mainly targeted toward Japanese, this local activity is also compatible with the global practice in AIM-BP [31].

Out-of-office blood pressure in guidelines

In the 2017 ACC/AHA Guidelines, out-of-office blood pressure measurements were “recommend to confirm the diagnosis of hypertension and for titration of blood pressure-lowering medication, in conjunction with telehealth counseling or clinical interventions” [3]. Similarly, wider use of out-of-office blood pressure measurement is “an option to confirm the diagnosis of hypertension” by the 2018 ESC/ESH Guidelines [2]. As Jones [37] summarized, diagnosis based on in-office blood pressure is still an essential blood pressure management decision in the US and European guidelines. However, in the 2018 ESC/ESH Guidelines, out-of-office BP, particularly home blood pressure, is recognized as a tool for detection of white-coat and masked hypertension as well as for monitoring blood pressure control [2]. AOBP theoretically enhances the in-office blood pressure measurement condition [37] if practitioners are able to prepare the environment of AOBP.

A unique vision of the JSH regarding out-of-office blood pressure is worth introducing. Out-of-office blood pressure is “often considered to have clinical values comparable to, or greater than, that of office blood pressure” [1]. The difference in the nature of blood pressure information is summarized in Table 2. Furthermore, the home blood pressure-based six-graded classification is provided based on a recent outcome-driven meta-analysis [38]. Although there is currently no randomized trial to guarantee the classification based on home blood pressure, the JSH 2019 Guidelines certainly met the expectations in the applied country. Home blood pressure is widely prevalent in Japan, e.g., 77% of patients with hypertension and 40% of individuals possess a home blood pressure measurement device, and ~40 million home blood pressure devices are estimated to be in use in Japan [1, 39]. A lot of Japanese routinely measure home blood pressure and therefore doctors, healthcare providers as well as people themselves have required the classification based on home blood pressure. Home blood pressure is further used as the prioritized diagnosis information of hypertension when an office blood pressure-based diagnosis differs [1]. To suit the clinical practice, it is reasonable that the JSH 2019 Guidelines differ from those from the other regions [2, 3]. Similar to the relationship between conventional office and AOBP, home blood pressure is emphasized more than ambulatory blood pressure because ambulatory monitoring is expensive and certain mental/physical stresses in a person are invoked during the measurement [40], although the advantages of ambulatory measurement are stated in the JSH 2019 Guidelines [1].

Table 2 Characteristics of blood pressure information.

As a successful example, an emerging topic first comes to fruition as specific guidelines, and is integrated to broad guidelines several years later. The initial guidelines for self-measured home blood pressure were published in Japan in 2003 [41] and revised in 2012 [42], and the unique utility of home blood pressure as the prioritized diagnostic tool was emphasized in the JSH 2014 [43], which continues to the most recent JSH 2019, as abovementioned [1]. Meanwhile, a joint policy statement on home blood pressure was recently published by the AHA and the American Medical Association [44]. In the statement [44], the advantages and recommended use of home blood pressure were described in detail. What should be noted by those outside the USA is that, although not applicable to all regions (states), a few private and commercial payers and the domestic insurance agencies (Medicaid and Medicare) have started to support self-measured home blood pressure monitoring since 2020. Pickering et al. [45], some of whom were also coauthors of the statement [44], called for action on home blood pressure measurement and emphasized the importance of reimbursement of the measurement. Ambivalent situations remain in Japan with respect to repeated petitions for reimbursement of home blood pressure measurement that have been declined by the government because the measurement is already in widespread use. We understand the severe financial situation; however, introducing the reimbursement shall enhance the support for home blood pressure measurement among people, particularly those with special needs, to improve the population health of society. Furthermore, our cost analysis revealed that the introduction of home blood pressure measurement for hypertension treatment would be a very effective method of reducing costs of hypertension treatment and treatment years for hypertension [46]. The US action of introducing reimbursement can be as a pioneering step for other countries, and we expect to expand the application in the near future.

The International Society of Hypertension (ISH) recently published the Hypertension Practice Guidelines (2020 ISH Guidelines) [4]. In contrast to developed countries, out-of-office blood pressure measurement “may not be feasible in low resource settings” [4]. Nevertheless, the 2020 ISH Guidelines also emphasize the importance of out-of-office blood pressure measurement for the accurate diagnosis of hypertension and for treatment decisions when the measurement would be possible and available [4]. The criteria are consistent with the aforementioned guidelines as shown in Table 3, and the usefulness of the combination of office and out-of-office blood pressure measurements to identify people with white-coat hypertension and those with masked hypertension is also stated in detail. The 2020 ISH Guidelines [4] therefore guarantee the application of out-of-office blood pressure, and call for action on every effort worldwide including out-of-office blood pressure measurement, somewhat like the previous US statement [44].

Table 3 Criteria for hypertension based on office, ambulatory, and home blood pressure measurement.

Importance of accurate measurement of blood pressure for diagnosing and treating hypertension

For diagnosing and treating hypertension, accurate blood pressure measurement as well as reliable data sources are essential. However, recently emerged reports in relation to coronavirus disease-2019 (COVID-19) have unveiled the current inconvenient situation. In our recent review on the association between hypertension and COVID-19 available as of June 7, 2020 [47], none of the nine available epidemiological studies concerning the association between hypertension and COVID-19 stated the diagnostic methodology used or the criteria for hypertension, such as in-office or out-of-office blood pressure, or self-reported or testimony of a family in serious situations. The huge variation in data sources makes the evaluation of reported incidence and risk of hypertension inaccurate [47]. We must keep tackling appropriate assessment of blood pressure during peacetime, in usual daily routines, so that we can provide timely and accurate evidence on hypertension under any situation, including an unexpected pandemic.