During PAV, the applied positive airway pressure is servocontrolled throughout each spontaneous inspiration such that it increases in proportion to a weighted sum of the airflow and inspired volume of spontaneous breathing (1–4). During expiration, the airway pressure returns to a preset positive end-expiratory pressure level. Thus, the inspiratory change in airway pressure has two components: a resistive unloading component that increases the airway pressure in proportion to inspiratory airflow (5), and an elastic unloading component that additionally increases the airway pressure in proportion to the inspiratory volume (6). These two components decrease resistive and elastic work of breathing in proportion to their gain settings (7).

To accomplish the proportional changes in airway pressure, the ventilator uses airflow and volume signals from a pneumotachograph as reference to generate the pressure waveforms. The pneumotachograph is mounted between the ventilator circuit and the endotracheal tube. This sensor provides clear, artifact-free signals in most cases. However, in neonates and particularly in small premature infants, the use of a pneumotachograph may be associated with a significant additional dead space load. Moreover, its signal may include a substantial and varying endotracheal tube leak component. This mimics inspiratory airflow and renders the signal useless for generating pressure waveforms in proportion to changes in lung volume. A small flow sensor is prone to malfunction whenever condensed water, secretions, or other particles obstruct its flow channels. Furthermore, the airflow signal as measured at the proximal airway may lag behind the diaphragmatic effort in the presence of substantial chest wall distortion. In this situation, the rise in ventilator pressure during PAV may be unduly delayed.

This study evaluates the applicability of RIP signals as substitutes for pneumotachograph signals to provide input information for volume- or flow-PAV. Inductance plethysmography derives respiratory volume changes noninvasively from inductance coils wrapped around the chest and abdomen (8–10). This system does not add any respiratory dead space, and the signal is not influenced by endotracheal tube leaks. However, it may be prone to body motion artifacts (11).

We hypothesized that RIP signals can replace pneumotachography as input signal to the ventilator during PAV. As markers for assessing the effectiveness of PAV with the different input signals, we measured arterial blood gases, tidal volume, minute ventilation, airway and esophageal pressure changes, and indices of thoracoabdominal motion asynchrony. Piglets and adult rabbits with normal and meconium-injured lungs were studied to test the system across different configurations and sizes of chest and abdomen, across different breathing patterns, and with normal and impaired pulmonary mechanics.

METHODS

Animal model.

Ten newborn piglets <10 d old (weight, 3120 ± 603 g [mean ± SD]) and five young adult New Zealand White rabbits (3140 ± 891 g) were anesthetized with 2 mg/kg xylazine and 20 mg/kg ketamine intramuscularly. Anesthesia was maintained by a continuous i.v. sodium pentothal infusion of 3–5 mg/kg/h. This dosage suppressed gross body movements but not spontaneous breathing efforts. Intravenous infusion fluids were administered at 3 mL/kg/h. Tracheostomy was performed on each animal, and the trachea was cannulated with an endotracheal tube (length, 150 mm; inner diameter 3.0 mm for piglets, 2.5 mm for rabbits), secured in place by a peritracheal ligature. An arterial catheter was inserted into the femoral artery (piglets) or the right carotid artery (rabbits) for intermittent determination of arterial blood gases (Ciba-Corning Diagnostics Corp., Medfield, MA, U.S.A.). A fluid-filled 8F infant feeding tube was inserted orally for esophageal manometry. The tip of the catheter was first advanced into the stomach and then withdrawn while observing the on-line pressure tracing until the largest negative inspiratory pressure deflections occurred. The catheter was flushed intermittently to maintain patency and to prevent gas bubbles from accumulating. The pressure transducer (Sorenson Transpac Transducers, Abbott Critical Care Systems, North Chicago, IL, U.S.A.; precision higher than ± 0.01 kPa, amplitude-frequency response flat to 20 Hz) was calibrated with a water manometer as reference. The animals were then placed in a right decubitus position and connected to a ventilator. They received 100% O2 throughout the study.

The transducer bands for RIP (Respigraph Noninvasive Monitoring Systems, Miami Beach, FL, U.S.A.) were cloth of 2.5 cm width with embedded sinusoid coils of Teflon-insulated wire. The rib cage band was wrapped around the chest wall with its lower rim 1 cm above the manubrium sterni. The upper rim of the abdominal band was placed 2 cm below the costal margin. The position of the bands was secured with tape. We started the Respigraph and used its built-in qualitative diagnostic calibration procedure to set the relative electrical gains of the rib cage and abdominal transducers (9, 12). The protocol was then commenced for the study on normal animals.

To induce lung injury, two doses of 2 mL/kg freshly prepared 20% human meconium in sterile saline solution were introduced deep into the trachea while the animals were rotated from side to side to assure even distribution. Between doses, the animals were manually ventilated with a bag. Meconium solution that did not distribute into the periphery of the lung was removed by repeated endotracheal suctioning. Conventional time-cycled, pressure-limited mechanical ventilation was then initiated for stabilization. Starting 30 min after meconium application during conventional ventilation, blood gases were obtained at 10-min intervals. The peak inflation pressure was adjusted if the Paco2 was outside a target range of 30–50 mm Hg. Additional doses of 0.5–1 mL/kg meconium solution were given until the Pao2 values no longer improved between blood sampling intervals. A maximum of up to 3 mL/kg of additional meconium solution was required in some animals. The degree of lung injury was judged stable enough for the purpose of the experiment when changes in Pao2 and Paco2 levels were <15% and did not show any consistent trend for at least 30 min. The protocol was then repeated to study the animals with injured lungs. The Respigraph was recalibrated immediately before restarting the protocol.

Protocol.

The animals were exposed to 10-min (normal lungs) or 20-min (injured lungs) epochs of PAV with three types of settings:1) elastic unloading, 2) resistive unloading, and 3) combined unloading. Each type was applied three times with different input signals to the ventilator:1) the pneumotachograph signal, 2) the AB, and 3) the Sum. The gain of the flow-pressure and the volume-pressure feedback, i.e. the degree of unloading, was adjusted to achieve equal peak inspiratory pressures within each mode. The end-expiratory pressure was set to 0 before lung injury and increased to 0.2 kPa thereafter. At the end of each test period, airflow, airway pressure, esophageal pressure, and the inductance signals were recorded over at least 10 consecutive respiratory cycles. Arterial blood samples were then obtained. Before the next test setting, a sidehole of 1 mm in diameter was opened in the endotracheal tube connector for a period of five respiratory cycles to mimic an endotracheal tube leak (Fig. 1). This was done to document the different consequences of a leak on ventilator pressure waveform when using the pneumotachograph or the inductance plethysmography signals as input. To allow for sequence effects, all test settings were applied in random order determined by choosing from a set of nine sealed envelopes that contained the mode assignment and the input signal type to the ventilator. To preclude potential carryover effects, at least 5 min of CPAP (normal lungs) or conventional mechanical ventilation (postmeconium lungs) were interposed between the test runs.

Fig 1
figure 1

Volume-PAV (elastic unloading) in a piglet with meconium-injured lungs. The airway pressure (Paw) waveforms were generated on the basis of a pneumotachographic signal (left) or from an AB (right). An endotracheal tube leak was opened at the beginning of the third breath. The airway pressure waveforms remain unaffected by the leak when based on the inductance signal (right). They are distorted and out of proportion to the volume entering the lung when based on the pneumotachographic signal (left). Upper ventilator pressure and inspiratory time limits were in place to prevent excessive and prolonged rise in airway pressure during leaks. V, volume of spontaneous breathing as measured by a pneumotachograph mounted between the endotracheal tube and the Y connector of the respirator circuit;A.U., arbitrary units.

This protocol was approved by the University of Miami Institutional Review Board.

Ventilator.

A STEPHANIE Infant Ventilator (F. Stephan Medizintechnik GmbH, Gackenbach and Dresden, Germany) was used throughout the study. Technical details of this device have been described elsewhere (13). It is a servocontrolled system with sensors for airway pressure and airflow near the endotracheal tube. The flow sensor is a pneumotachometer with a 1.1 kPa/L/s resistance at 5 L/min flow and a dead space of 0.9 mL. In addition to the conventional modes of mechanical ventilation, the STEPHANIE Infant Ventilator provided elastic unloading, resistive unloading, or a combination of these modalities. During the latter modes, the ventilator receives the flow signal of spontaneous breathing as input to a closed-loop rapid feedback circuitry that generates a pressure at the endotracheal tube connector as output. This pressure is proportional to the inspiratory flow signal (resistive unloading), or the integral of the flow signal, i.e. the volume signal (elastic unloading), or the sum of the two (combined resistive and elastic unloading). The proportions between the outlet pressure and the flow or volume signals can be adjusted as gains on continuous scales in kPa/L/s or kPa/mL, respectively. For the purpose of this study, the ventilator was modified so that the user could substitute the flow signal of the pneumotachograph for other analog electrical signals, such as the plethysmograph signals, as input to the feedback circuitry. The pneumotachograph remained in place for recording purposes when the other driving signals were used.

Data acquisition and processing.

The analog signals of airflow, airway pressure, esophageal pressure, and inductance plethysmography were digitized at a rate of 100 Hz and recorded on disk using data acquisition software (DATAQ Instruments, Inc., Akron, OH, U.S.A.). All the recorded signals were displayed on-line on the computer screen during the experiments. Tidal volume was obtained by digital integration of flow. The TCD ratio was calculated as the ratio between the sum of the absolute values of rib cage and abdominal displacement (entering a negative chest wall movement as positive) divided by the true sum of those values (8). The TCD increases above 1.0 with chest wall distortion. A delay (lag) between the start of inspiratory airflow and the beginning of the outward movement of the chest wall from its resting position was measured in degrees of the total respiratory cycle.

Statistics.

Data are given as mean ± SD. Repeated measures ANOVA was used to compare data obtained with the three input signals under investigation (pneumotachograph signal, AB, and Sum from inductance plethysmography). Bonferroni's method was applied to isolate differences among the three signal types. Significance was accepted for p < 0.05. To provide a measure of the reproducibility of the measurements, coefficients of variation were calculated for the within animal data variability from CPAP periods interposed between the PAV episodes.

RESULTS

Pulmonary compliance and resistance as measured before starting the protocol were 22 ± 9 mL/kPa/kg and 6.8 ± 2.4 kPa/L/s, respectively, in piglets; 23 ± 7 mL/kPa/kg and 5.2 ± 0.6 kPa/L/s, respectively, in rabbits. Five piglets died during meconium application. Postmeconium pulmonary compliance and resistance were 8 ± 3 mL/kPa/kg and 7.9 ± 4.0 kPa/L/s, respectively, in the remaining piglets; 10 ± 3 mL/kPa/kg and 7.0 ± 0.8 kPa/L/s, respectively, in rabbits. The reproducibility of the physiologic measures as assessed during baseline CPAP periods is given in Table 1.

Table 1 Coefficients of variation for the within animals data variability, calculated from CPAP periods between the PAV episodes

The resistive unloading part of the protocol could not be completed in any of the animals after meconium injury because arterial oxygenation could not be maintained with the low end-inspiratory airway pressures that occurred with resistive unloading alone. The animals maintained a regular breathing pattern during all other test settings before and after meconium injury. Their arterial blood gases were not significantly different among the three input signals. With few exceptions, the outcome variables reflecting ventilation and the pressures generated by the ventilator were also similar with the different input signals (data on elastic unloading are given in Tables 2 and 3, data on resistive and combined unloading are available from the authors on request). The few statistically significant differences between settings are small in size and may be regarded as clinically insignificant. None of these small differences were consistent in both species or in both the normal and injured lung situations. Chest wall distortion as measured by the TCD ratio and by the lag time was small in all normal animals. It increased moderately after meconium aspiration. Significant differences in these indices between input signals did not occur during any of the PAV test runs (Tables 2 and 3). There was never a consistent time lag of measurable size between the onset of inspiratory airflow as measured by the pneumotachograph and the onset of the inspiratory outward movement of the abdomen.

Table 2 Comparison of gas exchange, ventilatory variables, and indices of chest wall distortion in normal animals during volume-PAV generated from a pneumotachograph signal or from RIP
Table 3 Comparison of gas exchange, ventilatory variables, and indices of chest wall distortion in meconium-injured animals during volume-PAV generated from a pneumotachograph signal or from RIP

When a leak around the endotracheal tube developed, airway pressure waveforms generated from a pneumotachograph signal became distorted and out of proportion to the true volume and airflow changes of spontaneous breathing. In contrast, pressure waveforms generated from plethysmography signals were not affected by leaks around the endotracheal tube (Fig. 1).

Visual inspection of the ventilator pressure waveforms did not show qualitative differences between input signals. Plots of tidal volume versus ventilator pressure generated with plethysmography signals as feedback showed a linear relationship (Fig. 2). However, ventilator pressure waveforms driven by the plethysmographic Sum became asynchronous to the animals' breathing pattern when two conditions coincided: chest wall distortion and inadequate calibration of the plethysmograph. In the presence of chest wall distortion and an inappropriately high gain of the RC relative to the AB, the plethysmographic Sum may show a decrease or remain flat during initial inspiration instead of showing a rise. The decline of Sum during expiration may then be delayed because the rib cage signal becomes positive during expiration. This, in turn, delays the rise in airway pressure at the beginning of inspiration and retards its decline around the transition from inspiration to expiration (Fig. 3). Such signs of inappropriate relative calibration of the rib cage and abdominal sensors occurred regularly after dislocation of the bands during gross body movements or after major changes in body position. They were not noted initially after the plethysmography device had passed the standard calibration procedure.

Fig 2
figure 2

Ten consecutive breaths during volume-PAV (elastic unloading). The ventilator generated the inspiratory increase in airway pressure with the AB. Tidal volume was measured by pneumotachography at the endotracheal tube. Note that airway pressure is in proportion to tidal volume during inspiration. Loops turn clockwise. Tracing for a piglet with meconium-injured lungs breathing at 96 breaths/min.

Fig 3
figure 3

Volume-PAV mode based on an inappropriately calibrated Sum in a piglet with meconium-injured lungs. Sum is out of phase with the spontaneous respiratory effort because the RC component has been assigned an inappropriately high gain. As there is inward movement of the rib cage during initial inspiration, the rise of Sum is delayed. This, in turn, causes the airway pressure to lag behind the inspiratory effort. Because the RC becomes positive during expiration, airway pressure continues to rise after inspiration is completed, thus delaying expiration. Pes, esophageal pressure;V, airflow.

DISCUSSION

A PAV device generates immediate airway pressure changes proportional to an input signal of flow, volume, or both. These signals are usually obtained by pneumotachography. This study shows that inductance plethysmography can provide a reliable input signal. Arterial blood gases, ventilation, and ventilator pressure changes were similar when the ventilator was driven by the pneumotachography signal compared with the use of plethysmography.

To produce PAV, plethysmography is superior to pneumotachography in at least two respects: it avoids an additional dead space and an overestimation of tidal airflow and volume in the presence of an endotracheal tube leak. Both these aspects are of particular concern in low birth weight infants. Their tidal volumes are small relative to any added dead space such as that of a flow sensor in the airway. They also usually have an endotracheal tube leak, because tightly fitting endotracheal tubes are avoided to prevent laryngeal and tracheal injury. We selected two small animal species for their physiologic characteristics to render the study relevant for infants undergoing neonatal intensive care. The body weight range of our animals covers the weight range of newborn infants. The animals' rapid respiratory rates and small tidal volumes represent typical breathing patterns of term and preterm infants.

PAV driven by a single AB was of sufficient quality to allow for regular breathing, effective ventilation, and gas exchange. This may have important implications for patients without an endotracheal tube on noninvasive ventilation. The usually large and changing leak during noninvasive ventilation would not impair the performance characteristics of a PAV system based on inductance plethysmography. With a single abdominal band, the setup for PAV for those patients would also be simplified compared with the use of other input signals. The standard relative calibration of the two inductance signals was shown to be easily lost during body movements. This calibration, however is redundant when only the abdominal sensor is used. We initially expected a further advantage of the AB in that it might detect the inspiratory effort earlier than a pneumotachograph in the presence of chest wall distortion. This, however, could not be corroborated in our study, possibly because the chest wall in our animals was relatively stable and the induced degree of chest wall distortion was not high enough.

Our study shows that the Sum may not be useful under clinical conditions. An appropriate calibration of the relative weights of the RC and AB components of the plethysmographic Sum would be essential in the presence of any degree of chest wall distortion. Otherwise, asynchrony between the airway pressure waveforms and the spontaneous respiratory efforts will reduce the effectiveness of the system. With Sum-driven PAV, there was a trend toward lower tidal volumes and decreased minute ventilation in meconium-injured animals of both species (Table 3). This may be related to inadequate calibration of the plethysmograph during the study. A good calibration, however, is sometimes difficult to obtain, and frequent recalibration may become necessary with a variety of conditions, such as change in sleep state, body position, and dislocation of an inductance plethysmography band (11). Our study was performed on anesthetized animals that were mostly motionless, with quiet or dyspneic, single-posture breathing. This provided an optimal standardized situation for the inductance system. In clinical practice, additional limitations to the use of plethysmography may arise as bands may change location with frequent gross body movements in active newborn babies. This may compromise the effectiveness of plethysmography for this application and requires further study. It should, however, be less of a safety concern because the PAV software makes provision for gross artifacts on the input signal by limiting peak airway pressures and inflation times, and by providing backup conventional ventilation when no reasonable signal of spontaneous breathing is received.

In conclusion: RIP signals, in particular those from the abdomen alone, can be used to drive a respirator in the proportional assist modes. This offers advantages to infant ventilation as it avoids problems associated with the use of pneumotachographs, such as obstruction of the sensor by secretions or condensed water, additional dead space, and effects of varying endotracheal tube leaks.