Introduction

In recent years, defined energy deposition by focused ultrashort laser pulses has been of increasing interest both in science and for industrial applications, e.g., for precise micro machining.1 For example, pulses of several picoseconds (ps) allow material ablation with low thermal side effects for accurate drilling or cutting.2 However, for some applications, even shorter pulses of several tens of femtoseconds (fs) are attractive because high intensities >1013 W cm2 can be reached by focusing such fs-pulses with a comparatively low energy of only several μJ. When the focus is located within transparent materials such as glasses or tissue, nonlinear absorption processes lead to the formation of a micrometer-sized plasma in the focal volume, initiating a laser-induced optical breakdown.3 Depending on the incident intensity, material modifications such as permanent refractive index changes or disruptions can be realized.4,5

Especially for ophthalmic purposes, the well-localized nonlinear laser-tissue interaction has enabled new surgical techniques such as the well-established fs-laser-assisted in situ keratomileusis, where corneal tissue is cut and ablated for refractive correction.6 In contrast to solid-state materials, however, biological tissue consists primarily of water, and the laser–water interaction significantly contributes to the induced modifications.7 Thus, for sufficient pulse energy, the laser-induced plasma transfers its energy to the surrounding material within several ps, where the temperature rise causes the vaporization of the tissue. Finally, a cavitation bubble is formed, which leads to the local disruption of the transparent tissue.7

New therapeutic approaches aim at the treatment of the central and posterior segment of the eye.8 For example, in fs-laser cataract surgery, ultrashort laser pulses of several μJ are currently used for the rhexis of the lens capsule and for the fragmentation of the cataractous lens tissue to simplify its extraction.9,10 A similar strategy is applied in the concept of fs-laser presbyopia treatment, which aims to restore the elasticity of the lens by inducing laser disruptions in the hardened lens nucleus.11,12,13 Moreover, fs-laser-induced cuts within the vitreous body may reduce vitreous traction and are therefore discussed as a preventive therapy for posterior vitreous detachment.14

When treating the central or posterior segment of the eye, its anatomy requires focusing with a relatively low numerical aperture (NA) of 0.1. Thus, the long propagation distance of the pulse within the ocular medium and the high peak intensities initiate nonlinear side effects.15,16 When the pulse energy exceeds a critical power Pcrit, a dynamical balance between Kerr self-focusing and the defocusing effect of the induced plasma may result in the formation of an elongated filament along the propagation axis.17 Moreover, nonlinear frequency conversion, due to self-phase modulation, self-steepening of the pulse and sharp plasma gradients, produces a broadband supercontinuum (SC).18 All such nonlinear side effects are detrimental because the increased breakdown length reduces the quality of surgical cuts or even harms tissue outside the focal spot, while extended cut patterns of enlarged, persistent disruptions decrease the visual quality of the eye.19 Additionally, retinal safety limits may be affected due to the strong broadband SC generation.

So far, several studies have investigated laser-induced optical breakdown and its side effects such as self-focusing and filamentation in condensed matter.20,21,22 Investigations of the breakdown formation in biological tissue have revealed a strong increase in nonlinear side effects for NA <0.5.23 Thus, the formation of >100 µm streaks inside the tissue of laser-treated rabbit corneas was detected for rather low pulse energies of 1 µJ,15 whereas higher pulse energies even hampered a precise energy deposition due to strong self-focusing in a tissue model.14

To overcome these undesired side effects, varying the pulse duration along the direction of propagation by simultaneous spatial and temporal focusing (SSTF) is a promising approach.24,25 In this concept, the spectral components of the incident ultrashort laser pulse are spatially separated, resulting in a rainbow-like collimated beam, while reducing the local bandwidth within this beam leads to an increased pulse duration. Downstream from the focusing optics, the previously separated spectral components start to overlap. Within the focal volume, the entire bandwidth is recovered, leading to the reconstruction of the initial ultrashort pulse duration. In addition to the minimization of nonlinear effects, the intensity in front of the focus is reduced owing not only to the geometrical change of the beam diameter but also to the change of the pulse duration. Consequently, the nonlinear laser–material interaction is more strongly confined to the focal volume than for conventional focusing.

Until now, the concept of SSTF has been used for various applications, mainly focusing pulses with low energy and high NA at high repetition rates, e.g., in nonlinear fluorescence microscopy.25,26,27 In addition to the advantage of an improved axial resolution, the strong intensity confinement of SSTF enabled the precise ablation of transparent materials and the generation of circular microfluidic channels in glass.28,29 Furthermore, the suitability of SSTF for tissue ablation by laser-induced optical breakdown has recently been shown.30,31

In this paper, we present detailed experimental and numerical investigations of the nonlinear pulse propagation and breakdown formation induced by SSTF and conventional focusing at low repetition rate. Special emphasis is put on possible applications of SSTF in ophthalmic fs-laser surgery. Here, water is a commonly accepted model system for experiments on laser–tissue interaction due to its high content in biological tissue.7 We therefore analyze plasma formation and subsequent disruptions as well as nonlinear side effects such as filamentation in a water cell through pump-probe shadowgraphy. While previous studies of SSTF were frequently based on strongly elliptical beams caused by spectral separation,30,32,33,34 this study emphasizes equivalent focusing parameters such as a comparable NA to compare the nonlinear pulse–material interaction induced by conventional focusing and SSTF. This factor is of special importance because several applications, including ophthalmic laser surgery, require certain focusing conditions, where, for example, the NA is limited by the pupil size of the eye. Beside the analysis of optical breakdown, a measurement of the spectral broadening by SC generation is presented. Finally, we conclude with a detailed discussion of how SSTF drastically improves the conditions for fs-surgery of the lens or the posterior eye segment.

Materials and methods

Experimental realization

To induce and characterize the nonlinear laser-material interaction through conventional focusing and SSTF, a pump-probe set-up is used (Figure 1).

Figure 1
figure 1

Sketch of the experimental pump-probe set-up to induce and observe the optical breakdown in a water cell by simultaneous spatial and temporal focusing.

Pulses are delivered by a laser system consisting of a mode-locked oscillator and a chirped pulse amplification laser (Tsunami & Spitfire; Newport Corporation, Irvine, CA, USA). The system emits pulses with a minimal pulse duration of τ0=50 fs and a bandwidth of Δλ≈30 nm, centered at a wavelength of λ0=800 nm. The output beam diameters are 2wx=5.2 mm and 2wy=5.1 mm (1/e2). The pulse repetition rate of the laser system is set to 200 Hz, which is further reduced to 10 Hz using a synchronized mechanical chopper (MC2000; Thorlabs Inc., Newton, NJ, USA) to rule out interactions between subsequent breakdown events.

Pulses entering the set-up are split into a pump and a probe beam path. Within the pump-beam path, the pulse energy is adjusted using a half-wave plate and a polarization beam splitter cube. In the SSTF configuration, a cylindrical telescope reduces the horizontal beam width by a factor of 5 to an elliptical beam with a width of 2wx0=1.0 mm. Afterward, the elliptical beam is spectrally separated and collimated using a double-pass diffraction grating stretcher. The reflection grating (1500 lines mm−1) and incident angle were chosen to be equal to the compressor grating set-up of the chirped pulse amplification laser system to minimize third-order dispersion. Second-order dispersion induced by the SSTF grating configuration is compensated by pre-chirping the incident laser pulse by optimizing the chirped pulse amplification laser compressor gratings.35

The SSTF grating configuration expands the elliptical beam horizontally to a beam with a width of 2wx=5.4 mm, consisting of spectrally separated beamlets (see inset in Figure 1). Because the vertical beam size is kept at 2wy=5.1 mm, the spectrally separated beam has an approximately circular shape prior to focusing by means of a 90° off-axis parabolic mirror (effective focal length 25.4 mm, NT47-096; Edmund Optics, Barrington, NJ, USA). In accordance with the spectral separation, an autocorrelation of the partially blocked SSTF beam31 yielded a pulse duration of 300 fs within the collimated beam prior to focusing. Moreover, the measurement revealed a minimal pulse duration of 55 fs achievable at the focal spot of the SSTF configuration, similar to the initial pulse duration of the laser system (τ0=50 fs).

For purposes of investigating the optical breakdown induced by conventional focusing, the cylindrical telescope and the grating configuration are bypassed and the fs-pulses are focused directly. The resulting NA of the focusing optics in both regimes is NA≈0.1. The focal spot size is characterized using a high-resolution CMOS (complementary metal oxide semiconductor) camera sensor (pixel pitch 1.54 µm; Pentax Q, Ricoh Imaging Co., Ltd., Tokyo, Japan), which is moved along the propagation direction using a motorized translation stage (M-126 PD1; PI Physik Instrumente GmbH & Co. KG, Karlsruhe, Germany) to detect the xy intensity distribution at the focal plane.

To investigate laser-induced optical breakdown simulating the treatment of the lens or the vitreous body of the eye, pulses are focused 15 mm deep into a cell (704.002-OG; Hellma GmbH & Co. KG, Müllheim, Germany) filled with distilled water. To observe the plasma and disruption formation, the focal area is illuminated by probe pulses, which can be delayed up to 10 ns relative to the pump pulse. Shadowgraphic images are acquired using a 10× microscope objective (NT36-132; Edmund Optics Inc., Barrington, NJ, USA) and a CMOS camera (Thorlabs DCC1645M; Thorlabs Inc., Newton, NJ, USA). The image acquisition rate is synchronized with the chopper frequency (10 Hz).

For a quantitative breakdown analysis, artifacts are reduced by subtracting a reference image with the pump beam path blocked. After brightness normalization, the images are binarized to determine the full-width at half-maximum plasma length and diameter.

To analyze the plasma density, the temporal plasma evolution was observed by shifting the probe-pulse delay in steps of 170 fs to detect the maximum absorbance. When observing the optical breakdown in water, free electrons are the main reason for absorbance within the first few hundred fs after ionization.21,36 Afterward, the absorbance decreases due to recombination and diffusion, whereas a significant amount of electrons (20%–30%) is solvated and trapped by the water molecules due to their high polarity. This trapping causes longer-lasting absorbance because the solvated states exhibit broad optical resonances in the near infrared and decay times >500 ps. Because the losses caused by the reflectivity of the plasma are weak (0.01%) due to the low refractive index contrast of the plasma (<10−3) and the small probe pulse incidence angle, the absorbance of the free and solvated electrons is the main reason for the darkening in the shadowgraphic images.37 Thus, the average plasma density ne directly after the pump pulse incidence can be estimated by neα/σ, where α is the absorption coefficient of the plasma and σ=1.44×10−18 cm2 is the absorption cross-section for an electron collision time τc=10 fs according to the Drude model.38

The SC generated by the pulse within the water cell is collimated and characterized by two CCD (charge coupled device) spectrometers (USB 2000 & USB 4000; Ocean Optics Inc., Dunedin, FL, USA), which allow single shot measurements of the visual (350–850 nm) and near infrared (750–1050 nm) spectra, respectively. For shorter or longer wavelengths, the transmission of the optical components, the limited spectrometer sensitivity and the low SC intensity impede a reliable single-shot detection. At each pulse energy and wavelength range, 100 single-shot measurements are acquired. The measurements are averaged and calibrated by the neutral density and color filter transmission and by the spectrometer sensitivity. After normalization, the measurements of the visual (<800 nm) and infrared (>800 nm) spectra are combined to characterize the entire spectral broadening.

Numerical simulations

In our numerical simulations, we solve the so-called standard model for fs filamentation,17,39 adapting the model to water as the nonlinear medium:16,40

Here, the symbol ^ denotes the usual Fourier transform with respect to time t and transverse spatial coordinates x and y. The main approximation in the derivation of this model is the description of the laser pulse by a scalar forward-running field only, i.e., any backward-scattered fields are negligible. In the experimental study, linearly polarized and moderately focused (NA≈0.1) laser pulses are employed to generate plasma densities ne<5×1020 cm−3 well below the neutral density nnt=3.3×1022 cm−3, which justifies the above assumptions. Linear diffraction and dispersion are rigorously included in our fully space- and time-resolved (3D+1) numerical code, and we use the tabulated values of the complex susceptibility χ(ω) of water. Then, k2=(ω2/c2)(1+χ), and n0 accounts for the refractive index at 800 nm. Third-order nonlinear polarization is included via the Kerr coefficient n2=1.9×10−16 cm2 W−1, while both nonlinear dispersion and higher order nonlinear susceptibilities are neglected.16 Note that our code rigorously accounts for third harmonic generation; however, due to the strong phase mismatch, the third harmonic spectral amplitude stays more than four orders of magnitude below the fundamental one and is not discussed in this paper. Ionization is included via the Keldysh theory,41 where we assume a gap potential of Ui=7 eV. In addition to the intensity-dependent Keldysh ionization rate W(I), the plasma model is governed by avalanche ionization and the quadratic recombination with coefficient a=2×10−24 cm3 fs−1.

To mimic the experimental pulse profiles while keeping the computational time within reasonable limits, the nonlinear propagation is started 2 mm before the linear focus. In the case of conventional focusing, a simple Gaussian pulsed beam with a radius of w0=200 µm and a duration of τ0=50 fs is assumed and is focused by a perfect lens with a focal length f=2 mm (NA≈0.1). In the SSTF configuration, an elliptical pulsed beamlet with wx0=40 µm, wy=200 µm, τ0=50 fs and a spatial chirp rate35 of 4.9 mm fs in the x direction is employed. Note that due to the spatial chirp rate, the resulting intensity profile at z=−2 mm is perfectly symmetrical in the transverse xy plane, with a 1/e2 intensity radius of 200 µm, as for conventional focusing, but the pulse duration is increased by a factor of 5 to 250 fs.

Results and discussion

Experimental results

For conventional focusing, the measurement of the intensity distribution at the focal plane in air yields a circular focal spot diameter of 7 µm (1/e2), as shown in Figure 2a. Regarding SSTF, the size of the spectrally separated beam prior to focusing is equivalent (2wx,y 5 mm), thus leading to a comparable NA and beam propagation toward the focal plane. Near the geometrical focus, the interference of the spectrally separated beamlets leads to the local shortening of the pulse duration. Furthermore, previous studies on SSTF have shown that a tilt of the pulse-front is induced near the focus,26,42 which leads to a sweep of the focal spot across the focal plane in x direction. As the focal spot sweeps in a much faster time than the image acquisition system can resolve, the focal spot appears elliptical, as shown in Figure 2b.

Figure 2
figure 2

Intensity distribution at the focal plane (z=0) for (a) conventional focusing and (b) SSTF configuration. SSTF, simultaneous spatial and temporal focusing.

For a detailed understanding of the breakdown dynamics, Figure 3a–3e presents time-resolved shadowgraphic images of the temporal plasma evolution in water for conventionally focused pulses of 3 µJ. The transmission and the average plasma density of the central breakdown area (marked in Figure 3a–3e) are shown in Figure 3f. Right after the pump pulse incidence (t=0), the absorption of the probe pulse increases, indicating the onset of plasma formation (Figure 3a–3c). The maximum absorbance is reached within 0.5 ps (Figure 3c), in good accordance with previous studies on plasma dynamics in water.36,43 Based on the simple assumption of a homogeneous plasma distribution within the plasma channel with a measured average diameter of 8±1 µm, we estimate an average plasma density of (4±1)×1020 cm−3, according to the reduction in transmission at the location marked in Figure 3. Then, the plasma density decreases due to the recombination and diffusion of the free electrons, while longer-lasting absorbance is caused by solvated electrons.

Figure 3
figure 3

Temporal evolution of the plasma formation induced by conventionally focused pulses of 3 µJ: (ae) present shadowgraphic images for increasing the probe-pulse delay (Δt=0.5 ps), beginning at t=−0.5 ps in a. Pump pulses incident from the left. Black scale bar indicates 100 µm. The white line marks the position where the probe-pulse transmission and plasma density are analyzed and plotted in (f).

Shadowgraphic images of the plasma induced by conventional focusing and SSTF are shown in Figure 4a and 4b. The probe-pulse delay was optimized for maximum average plasma contrast after pump pulse incidence. Figure 4c and 4d illustrates the beginning of the disruptions due to vaporization at t=10 ns after the pump pulse incidence.

Figure 4
figure 4

Shadowgraphic images of the plasma formation (a, b) and subsequent disruptions (c, d), illustrating the strongly confined optical breakdown and reduced filamentation effects in the SSTF configuration (right column, b, d) compared to conventional focusing (left column, a, c ). Pump pulses incident from the left. SSTF, simultaneous spatial and temporal focusing.

Regarding conventional focusing, the threshold for visible plasma formation on the shadowgraphic images is Ethres 0.2 µJ or I 6×1012 W cm−2, which is in good agreement with the breakdown thresholds in water determined by previous studies.44,45 Up to 2 µJ, the plasma volume appears homogeneous and well confined to the focal plane, reaching a length of 350 µm and a diameter of 6 µm (Figure 4a). Thus, disruptions recorded 10 ns after the pump pulse incidence also appear well localized (Figure 4c). Next to the central breakdown, brighter lines display shockwaves, which are emitted as a result of the fast thermal expansion of the breakdown volume.46 Using an intermediate pulse energy of 4 µJ, the contrast of the shadowgraphic plasma images starts to become inhomogeneous for conventional focusing, indicating variations of the plasma density along the focal volume. Due to the inhomogeneous energy deposition, the formation of additional disruptions next to the optical axis is observed. For a high pulse energy of 8 µJ, conventional focusing induces strong filamentation and the formation of several plasma side branches. Furthermore, the central plasma filament even exceeds the image area, with a length of >1 mm. Due to the formation of lateral filaments, the average vertical plasma diameter becomes >13 µm. Additionally, the strongly inhomogeneous contrast within the central plasma filament indicates significant variations in the plasma density. As a consequence, the subsequent disruption breaks up into a multitude of smaller disruption volumes because the plasma density locally drops below the threshold of disruption formation.

For SSTF, the threshold for visible plasma formation is slightly increased to Ethres 0.4 µJ because of the slight increase of the focal pulse duration (55 fs vs. 50 fs) and the elliptically enlarged focal spot (Figure 2b). Because the breakdown area is observed in the vertical yz plane, the horizontal widening of the plasma volume is not imaged in our experiment. However, in contrast to conventional focusing, plasma formation is well localized to the focal plane and does not show the formation of extended filaments, additional side branches or inhomogeneities, even at high pulse energies up to 8 µJ (Figure 4b). The threshold of plasma formation is only exceeded within the focal volume, leading to a more homogeneous energy deposition and reducing the length and the vertical diameter of the plasma volume by a factor of 2. Thus, a plasma channel with a length of <500 µm and an average vertical diameter of 7 µm is detected when focusing pulses with 8 µJ using SSTF. We would like to recall here that in the case of SSTF, the laser pulses are largely pre-chirped (τprobe-pulse >10 ps) to compensate for the dispersion induced by the SSTF set-up, which prevents the time-resolved detection of the maximum absorption and, thus, the estimation of the plasma density.

Figure 4d presents shadowgraphic images of the disruptions within the water cell 10 ns after the incidence of the SSTF pump pulses. Similar to the homogeneous plasma formation, the subsequent disruptions are well confined to the focal plane. In contrast to conventional focusing, there is no splitting of the disruption volume. While the break-up of the disruptions leads to inhomogeneous shockwave emission in conventional focusing, the shockwaves generated by SSTF are more uniform. This uniformity is of special interest regarding fs-surgery because both the bubble formation by vaporization and the emitted shockwaves may damage adjacent cells.7

Numerical analysis

For a detailed understanding of plasma formation in SSTF and conventional focusing, Figure 5 presents snapshots of the simulated space–time x/t intensity profiles (y=0) for linear and nonlinear (Epulse=4 µJ) pulse propagation, illustrating both focusing conditions. For conventional focusing, the linear pulse propagation is symmetrical to the focal plane, showing the well-known converging and diverging behavior (Figure 5a–5c). For higher pulse energies, nonlinear pulse propagation leads to a complex interplay between the Kerr self-focusing and defocusing effects of the induced plasma (Figure 5d–5f). Because the divergence of the pulse is hampered downstream from the focal plane, the intensity is kept trapped within a central region near the optical axis.

Figure 5
figure 5

Snapshots of linear (ac and gi) and nonlinear pulse propagation (df and jl, Epulse=4 µJ) for conventional focusing (upper section) and SSTF (lower section), showing x/t intensity profiles (y=0) at the indicated distances relative to the focus. Color scales are normalized to the individual maximum intensities of the subfigures. SSTF, simultaneous spatial and temporal focusing.

In contrast to conventional focusing, the spatiotemporal coupling mechanism responsible for pulse compression at the focus for the SSTF configuration is clearly visible in Figure 5g–5i (linear regime). Outside the focal plane (z=± 0.2 mm), the spectral components are separated, leading to a pulse duration of 220 fs. At the focal plane (Figure 5h), the spectral overlap leads to a local shortening of the pulse to τ=50 fs. However, the spatiotemporal character of this pulse shortening manifests itself in the abovementioned characteristic pulse front tilt.28,42 At any position in the transverse (x,y) plane, the pulse duration is short (50 fs), whereas the spatially integrated power profile of the pulse still has the initial duration of 250 fs for the single beamlets. In the nonlinear regime (Figure 5j–5l), a strong distortion of the pulse profile is visible. However, the basic features of SSTF survive, leading to a broader intensity distribution before and behind the focus compared to conventional focusing. Although a similar maximum intensity of Imax 24 TW cm−2 is reached at the focal plane (z=0) for both focusing conditions, SSTF strongly suppresses the filamentation effect. Thus, 200 µm downstream from the focal plane, the maximum intensity is already significantly decreased by a factor of >10 compared to conventional focusing.

To illustrate the resulting plasma formation caused by the simulated intensity distributions, Figure 6 presents iso-electron density surface plots at 1019 cm−3 for different pulse energies and both focusing configurations. As in the experimental study on breakdown formation, the plasma channels induced by conventionally focusing low pulse energies ≤2 µJ are well localized. However, self-focusing can be observed even here, shifting the breakdown volume toward the focusing lens. For pulse energies ≥4 µJ, the formation of a central filament strongly increases the length of the induced plasma (Figure 6c and 6d).

Figure 6
figure 6

Simulation of plasma channels induced by conventional focusing (left column, (a–d)) and SSTF (right column, (e–h)) at 1 µJ, 2 µJ, 4 µJ and 8 µJ, respectively. Iso-electron density surface plots are shown at 1019 cm−3. SSTF, simultaneous spatial and temporal focusing.

In SSTF, the formation of a central filament is impeded due to the sharply confined intensity distribution, therefore leading to a well-localized plasma formation in propagation direction (Figure 6e–6h). For example, the plasma channel produced by the 8 µJ SSTF pulse (Figure 6h) has a similar length as the channel produced by the 2 µJ conventional pulse (Figure 6b). At the same time, inspection of the simulation data reveals that the energy deposited by the 8 µJ SSTF pulse is more than three times larger than that deposited by the 2 µJ conventional pulse. The maximum plasma densities achieved are in the range of (2–4)×1020 cm−3, in excellent agreement with the estimates from our experiments. Interestingly, this value does not depend on the pulse energy, provided that the threshold energy for plasma formation is exceeded. The reason for this behavior is the well-known intensity clamping effect in femtosecond filaments:17,39 the maximum plasma densities and maximum intensities are solely determined by an inversion of the sign in the nonlinear induced refractive index change. As a consequence, all pulse intensities, whether conventionally focused or in the SSTF configuration, saturate at approximately 25 TW cm−2.

Because we use ideal focusing conditions in our simulations, the resulting plasma channels are shorter than in the experiments (Figure 4). Moreover, due to the perfectly Gaussian beam shapes and the absence of aberrations, which certainly distort the pulse propagation in the experiment, no multifilamentation occurs in the numerical simulations.47 Consequently, the diameter of the plasma channel does not differ much for conventional focusing and SSTF. However, the experimentally observed overall reduction of the length of the plasma channel by a factor of 2 due to SSTF is nicely reproduced in our simulations.

SC generation

In addition to the plasma formation, the interaction of the intense laser pulse with the nonlinear medium leads to the generation of a broadband SC. The measurement and simulation results of SC induced by conventional focusing and SSTF are compared in Figure 7. For conventional focusing, a significant broadening of the pulse spectrum is already observed for the pulse energies ≤1 µJ. With increasing pulse energy, the generated SC becomes broader and stretches toward shorter wavelengths (Figure 7a). Up to pulse energies of 4 µJ, the maximum blue shift increases. For pulse energies >4 µJ, a cutoff at 400 nm is observed. Further increase of the pulse energy results only in an increased intensity of the broadened spectrum, as similarly reported by Liu et al.,48 even for higher pulse energies up to 200 µJ. Plasma defocusing clamps both the intensity and the wavelength shift to a maximum value because additional plasma filaments are formed due to the modulation instabilities of the laser pulse, as shown in Figure 4a.

Figure 7
figure 7

Experimental (a, b) and numerical (c, d) characterization of the supercontinuum generation in water, induced by conventional focusing (left column) and SSTF (right column). The black dotted lines indicate the input pulse spectra. SSTF, simultaneous spatial and temporal focusing.

Regarding SSTF, the spectral broadening of the incident laser pulse is strongly decreased (Figure 7b). For low pulse energies up to 2 µJ, almost no frequency conversion is observed, and only a slight broadening is detected even for high pulse energies up to 8 µJ.

The simulated spectra reproduce the strong decrease of SC in the SSTF configuration (Figure 7d). However, the spectra obtained for conventional focusing (Figure 7c) are much narrower than are their experimental counterparts. This discrepancy is readily explained by the differing nonlinear interaction lengths, as the experimental pump pulses propagate a longer distance of 15 mm within the water cell. Self-phase modulation as the main driver for SC generation is a cumulative effect, and a shorter channeling range should therefore reduce the SC generation.

In this context, we would like to comment on an interesting observation that once more illustrates certain peculiarities of SSTF pulses. When analyzing the measured or simulated spectra in Figure 7 with respect to plasma formation in Figures 4a, 4b and 6, one may wonder why SC is more pronounced for conventional focusing, even when compared to SSTF pulses with higher energies and comparable plasma length. For example, a 2 µJ pulse with conventional focusing produces a much broader spectrum than an 8 µJ SSTF pulse, though the length of their respective plasma channels is almost the same. The reasons for this initially surprising behavior are the differences in the actual (spatio-)temporal pulse shape. We noted above that the spectral broadening is mainly driven by self-phase modulation, which crucially depends on the pulse duration: short pulses modulate the phase much more strongly than longer pulses. For conventional focusing, the pulse envelope is always short, here 50 fs or even less. In the SSTF configuration, however, the pulse is short only in the vicinity of the focus (see also Figure 5), while slightly before or after the focus, the pulse duration is increased considerably and the self-phase modulation is strongly suppressed.

Conclusions

In conclusion, SSTF was investigated with the aim of improving the nonlinear processing of transparent materials using fs-laser pulses. Certain applications require optical breakdown formation by focused pulses of several μJ and limited NA. Thus, nonlinear side effects such as self-focusing, filamentation, the break-up of disruptions and strong SC generation may reduce the precision and quality of the induced modifications. This drawback is of particular importance, e.g., in ophthalmic surgery: to fragment the cataractous lens tissue prior to extraction or to cut the vitreous body, pulse energies up to 10 µJ are applied, while the NA is limited by the pupil of the eye.9,10,14 Thus, filamentation reduces the quality of surgical cuts or may even harm the nearby capsular bag when using conventional focusing. Moreover, strong white-light generation poses a risk for retinal laser safety.

Here, simultaneous spatial and temporal focusing is applied for improved intensity confinement and a decrease in the nonlinear side effects. Because the ultrashort pulse is only formed at the focal plane in the SSTF configuration, the intensity outside the focal volume is strongly reduced, thus confining the laser–material interaction.

We have demonstrated this phenomenon by evaluating the optical breakdown and nonlinear side effects in a water cell, both experimentally and numerically comparing conventional focusing and SSTF. Regarding conventional focusing, pump-probe shadowgraphy revealed strong nonlinear side effects, such as the formation of additional plasma side branches and extended plasma filaments reaching a length of >1 mm. Thus, the inhomogeneous plasma density caused by the complex interplay of self-focusing and plasma defocusing led to a break-up of the disruptions and strongly fluctuating shockwave emission. Furthermore, self-phase modulation caused a broadband SC reaching <400 nm. For SSTF, the improved intensity confinement in propagation direction significantly reduced the nonlinear side effects. In excellent agreement with the numerical simulations, the length of the plasma channels was reduced by a factor of 2, and no filamentation was observed, even for high pulse energies up to 8 µJ. Because no plasma side branches or axial inhomogeneities were induced, the subsequent disruptions were well confined to the focal volume. Moreover, white light generation through SC could be drastically reduced by SSTF. Only slight spectral broadening was observed experimentally or in the simulations, even for high pulse energies.

Consequently, the confined laser–material interaction and strongly reduced nonlinear side effects make SSTF a versatile technique for the high-precision processing of transparent materials, such as ophthalmic fs-surgery, where the restricted breakdown size and the impeded SC generation ensure precise and gentle treatment.

Contributions

The presented work was carried out in collaboration among all authors. RK, RA and SN designed the experimental methods, and RK performed the experimental investigations and the data analysis. The numerical simulations were performed by JG and SS. RK and SS wrote the manuscript. The interpretation and discussion of the results were supported by JT and AT.