Introduction

Several guidelines include a blood pressure (BP) category to define arterial hypertension.1, 2, 3 BP measured over the brachial artery is usually used for patient evaluation and management, based on its role as a powerful predictor of cardiovascular morbidity and mortality.1, 2, 3, 4, 5, 6 Recent studies have suggested that central BP is more closely associated with cardiovascular events and target organ damage than brachial BP,7, 8, 9, 10, 11 although this idea is still controversial.12, 13 In spite of its emergence as the better index for cardiovascular risk management, a wider implementation of central BP into clinical practice is hampered by the lack of established reference values.

The superiority of central BP over brachial BP in monitoring therapeutic effects and predicting cardiovascular events has been well established using the SphygmoCor technique (AtCor Medical, Sydney, Australia).7, 9, 14, 15 The SphygmoCor device directly records radial pulse waveform by tonometry, and this is converted to a central aortic pressure waveform using a generalized transfer function. Central BP is then estimated by calibration against brachial BP.16, 17 Recently, an automated device for the estimation of central systolic BP has been introduced, namely the Omron HEM-9000AI (Omron Healthcare, Kyoto, Japan).18, 19, 20, 21, 22 The Omron device records radial pulse waveform in an automated manner, and BP corresponding to the second systolic peak is obtained by calibration against brachial BP. The central systolic BP is then estimated using a regression equation.18, 22 Because of the simplicity of these measurements,18, 22 use of the Omron device is now widespread especially in Japan. These noninvasive estimations show close correlation with the invasive measurements of central BP in ascending aorta.18, 21, 22 Furthermore, central systolic BP values estimated by the Omron device are highly correlated with those estimated on a SphygmoCor device.19, 20, 21 However, both devices underestimate central systolic BP, with the SphygmoCor device producing a larger deviation (average, –15 mm Hg) than the Omron device (–2 mm Hg).21 These results indicated that central systolic BP estimated by either the Omron or SphygmoCor device could provide a useful clinical marker in the management of hypertension, although the Omron device apparently requires its own reference values because of technological differences between the systems. Thus, this study was designed to investigate the distribution of central systolic BP values estimated by the Omron device in the Japanese general population.

Methods

Study design and subjects

A cross-sectional study was used to evaluate characteristics relevant to central BP in subjects with various cardiovascular risk factors. We enrolled 10 756 consecutive subjects aged 20 years or more (male=6574; mean age 55.3±12.5 years, range 20–91 years), who underwent a multiphasic health checkup in the Department of Health Checkup, Enshu Hospital, from January 2009 to September 2011. Subjects with overt cardiovascular disease were excluded from the study. We undertook this study in accordance with the principles of the Declaration of Helsinki. The study protocol was approved by the Ethics Committee of Enshu Hospital. All patients gave written informed consent to participate before the start of the study.

Study procedures

Our health checkup program included a routine physical examination, chest X-ray, electrocardiography and laboratory assessment of cardiovascular risk factors. Brachial BP (oscillometer) was measured and central systolic BP was estimated after participants were seated in a chair for 5 min with their backs supported and their arms supported at heart level in an air-conditioned room (HEM-9000AI; Omron Healthcare).

The population was categorized according to age decade (20–29, 30–39, 40–49, 50–59, 60–69 and 70 years) and BP category. To investigate the distribution of central systolic BP values, subjects were excluded if taking medications that may affect BP or arterial stiffness, which is an important determinant of central BP (Figure 1). ‘Population A’ was defined as subjects of both sexes taking no medication for hypertension, dyslipidemia or diabetes mellitus. Subjects with major risk factors were then excluded from ‘population A’ to form a near-normal population. ‘Population B’ was defined as subjects of both sexes without major cardiovascular risk factors other than hypertension (without diabetes, dyslipidemia or current smoking) and not taking antihypertensive medication. ‘Population C’ was defined as subjects with optimal or normal BP and no major cardiovascular risk factors (Figure 1). Classification of brachial BP levels was defined as follows: optimal BP, systolic BP <120 mm Hg and diastolic BP <80 mm Hg; normal BP, systolic BP <130 mm Hg and diastolic BP <85 mm Hg; high normal BP, systolic BP 130–139 mm Hg and/or diastolic BP 85–89 mm Hg; hypertension, systolic BP 140 mm Hg and/or diastolic BP 90 mm Hg.1, 2, 3 Subjects taking antihypertensive medications were classified as hypertension. In this study, diabetes mellitus was defined by a fasting plasma glucose 126 mg dl−1 or by the use of antidiabetic medications, whereas dyslipidemia was defined by low-density lipoprotein cholesterol levels 140 mg dl−1, high-density lipoprotein cholesterol values <40 mg dl−1, triglycerides 150 mg dl−1 or the use of antidyslipidemic medications. The estimated glomerular filtration rate was calculated using the Japanese Society of Nephrology formula.23

Figure 1
figure 1

Flowchart describing the methods for selection of subjects included in populations A–C. BP, blood pressure.

Estimation of central BP

Radial artery pressure waveforms and brachial BP were recorded simultaneously using a fully automated device (HEM-9000AI; Omron Healthcare) to calculate late systolic pressure in the radial artery (SBP2) and estimate central systolic BP. The brachial BP was measured with an oscillometric manometer and the radial pulse waveforms were recorded noninvasively using an applanation tonometer, which consisted of a sensor unit with an array of 40 fine-pitch microtransducer elements and a monitor unit. As one of these 40 sensor elements is automatically selected to obtain the optimal radial arterial pressure waveform, this method is considered to provide better objectivity. Signals of the radial arterial pressure wave were low-pass filtered: first at a cutoff frequency of 105 Hz to remove high-frequency noise, and second at 25 Hz to extract pressure waveforms. The radial arterial waveform obtained with this device is reportedly identical to the simultaneously and invasively measured intra-arterial pulse waveform of the opposite radial artery.18

Inflection points or peaks that corresponded to early and late systolic BP were obtained by multidimensional derivatives of the original pulse pressure waveforms. The maximal systolic pressure and diastolic pressure in the radial artery were calibrated with the brachial systolic BP and diastolic BP, respectively. The late systolic BP in the radial artery (SBP2) was calculated by the following equation:

where P2 and PP indicate the height of the late systolic shoulder/peak pressure and the pulse pressure of the radial arterial pressure contour, respectively. Central systolic BP was estimated using a regression equation with SBP2 as a major independent variable.18, 22 Subjects who showed arrhythmia during this procedure were excluded from this study.

Statistics

All analyses were performed using StatView 5.0 (SAS Institute, Cary, NC, USA). Data in the text and the tables are expressed as mean±s.d. unless otherwise stated. Differences between two means that had a normal distribution were compared by unpaired Student’s t-test or analysis of variance followed by Scheffe’s F-test. A P-value of <0.05 was considered statistically significant.

Results

The baseline characteristics of study subjects are listed in Table 1. Although subjects with overt cardiovascular disease were excluded from this study, the majority of study subjects (n=7570, 70.4%) had at least one of the following cardiovascular risk factors: hypertension, diabetes mellitus, dyslipidemia or a smoking habit. Among 3758 subjects with hypertension, 2040 subjects (54.3%) were taking antihypertensive medication. Similarly, 83.7% and 28.6% of subjects with diabetes and dyslipidemia, respectively, were taking medications. Table 2 presents the characteristics of populations A, B and C. BP was lower in population B than in population A, and lower still in population C. Similarly, the mean age was lowest in population C (Table 2).

Table 1 Characteristics of study subjects
Table 2 Characteristics of subjects in populations A–C

Brachial BP increased with age in populations A and B (Figure 2 and Table 3). Such an increase was also observed in population C, although the impact of increasing age on brachial BP was small in subjects without hypertension, and the BP values were lower in population C than in population A or B.

Figure 2
figure 2

Relationship between age and brachial systolic blood pressure (BP) in (a) population A (n=7348, r=0.329, P <0.0001), (b) population B (n=3760, r=0.411, P <0.0001) and (c) population C (n=2672, r=0.286, P <0.0001).

Table 3 Brachial SBP in each age decade of populations A–C

The estimated central systolic BP was affected by various factors according to the univariate analysis (Table 4). Among them, brachial BP and age correlated closely with central systolic BP. A significant linear correlation was observed between central systolic BP and age in this study (Figure 3), although the effect was minimal in subjects without major cardiovascular risk factors (hypertension, diabetes, dyslipidemia or smoking; population C; Figure 3c). Central systolic BP in each decade was greater than the corresponding brachial systolic BP in subjects aged 30 years in populations A, B and C, whereas there was no difference in subjects 20–29 years of age between central systolic BP and brachial systolic BP (population A, P=0.24; population B, P=0.22; population C, P=0.97; Tables 3 and 5).

Table 4 Relationship between estimated central systolic blood pressure and other variables by univariate analysis
Figure 3
figure 3

Relationship between age and central systolic blood pressure (BP) in (a) population A (n=7348, r=0.420, P <0.0001), (b) population B (n=3760, r=0.481, P <0.0001) and (c) population C (n=2672, r=0.420, P <0.0001).

Table 5 Distribution of central systolic blood pressure values and late systolic pressure values in the radial artery (SBP2) in each age decade of populations A–C

Table 5 lists the mean and 2 s.d. of central systolic BP in each decade of age in populations A, B and C. Values of central systolic BP for individuals without major cardiovascular risk factors are listed in the column marked as ‘population B’ (Table 5), including those with hypertension determined by brachial BP. However, brachial BP is an established and independent cardiovascular risk factor, supporting the concept that individual risk stratification should be made taken into account both brachial and central BP. Central systolic BP was lowest in population C comprising subjects in the optimal and normal brachial BP categories (Table 5).

Central systolic BP was closely correlated with brachial systolic BP (Figure 4) and, thus, was significantly different among subjects from different BP categories defined using brachial BP; central systolic BP was increased across the brachial BP categories in populations A, B and C (Table 6).

Figure 4
figure 4

Relationship between central and brachial systolic blood pressure (BP) in population A (left panel; n=7348, r=0.940, P <0.0001) and population B (right panel; n=3760, r=0.943, P <0.0001). Lines of identity (dotted lines) and regression (solid lines) are shown.

Table 6 Distribution of central and brachial systolic blood pressure values in each brachial blood pressure category (BP category) of populations A–C

Discussion

Despite the general agreement that central BP bears a closer relationship to cardiovascular target organ damage and cardiovascular events than brachial BP,7, 8, 9, 10, 11 its use in clinical practice has been limited because of the lack of reliable reference values. As direct measurement of central BP requires an invasive approach, several techniques for the noninvasive estimation of central BP have been developed.7, 8, 9, 10, 11, 14, 15, 18, 19, 20, 21, 22 Central BPs obtained using different devices, although yielding the same type of information, are not readily comparable between measurements, and each technique seems to require its own reference values as diagnostic thresholds. This study provided important information for considering reference values for central systolic BP estimated using the Omron device by investigating characteristics of these measurements in the Japanese general population. This may be an initial step in establishing central BP as a clinical tool in the diagnosis and management of hypertension and for individual risk evaluation.

Although numerous factors affect central BP, this study identified age and brachial BP as particularly important factors in investigating the distribution of central systolic BP values based on their significance of effect in univariate analysis. Presentation of reference values by age and brachial BP category would be a possible option for practical use, although not validated. Although the increase in central systolic BP with age could be largely attributed to the close relationship between central systolic BP and brachial systolic BP, the age-dependent increase was somewhat greater for central systolic BP than for brachial systolic BP. An age-dependent increase in aortic stiffness would cause premature return of the reflected waves in late systole, and thus an augmentation of central BP.24, 25 This may be one of the reasons for the greater impact of age on central than on brachial BP. The age-dependent increase in central systolic BP may reflect age-dependent increase in the number and/or severity of cardiovascular risk factors in individuals. Indeed, most tested cardiovascular risk factors had a positive impact on central systolic BP. The impact of increasing age on central systolic BP was not prominent in population C and central systolic BP did not significantly increase with age in subjects aged 50 years in this population, further supporting the considerable contribution of age-dependent increases in brachial systolic BP to those in central systolic BP.

The study subjects had no overt cardiovascular disease, although some were under medical treatment for hypertension, diabetes or dyslipidemia. Antihypertensive drugs differentially affect central BP,7, 26 and antidiabetic and antidyslipidemic drugs may alter central BP through modification of arterial stiffness. Thus, subjects under such medication were first excluded from the study population (population A). In the next step of analysis, we investigated the effects of cardiovascular risk factors on central BP values. As previous studies demonstrated, central BP is an independent predictor of target organ damage and cardiovascular events, and a better marker in the management of hypertension than conventional brachial BP.7, 8, 9, 10, 11, 12 Thus, population B was determined along the concept that central BP, but not brachial BP, should be used for risk evaluation. Subjects in this population had neither major cardiovascular risk factors (diabetes, dyslipidemia, smoking) other than hypertension (defined using brachial BP) nor antihypertensive medication. However, excluding subjects with diabetes, dyslipidemia and smoking from population A did not alter the subsequent value of central systolic BP in each age decade or brachial BP category (population B). This does not necessarily indicate that effects of cardiovascular risk factors on central BP are minimal. Indeed, central systolic BP was higher in subjects with hypertension, diabetes, dyslipidemia or smoking habit as compared with those without such cardiovascular risk factors in population A (data not shown). In this study, some subjects with diabetes or dyslipidemia, probably those with a moderate to severe condition, might have been taking relevant medications and, thus, were already excluded from population A. This may, at least partially, contribute to the similar central systolic BP values between populations A and B. Alternatively or additionally, the lack of data on some other cardiovascular risk factors such as family history may contribute to the results. Exclusion of overweight subjects (body mass index >25 kg m−2) or those with metabolic syndrome from population B did not alter the results (data not shown). Although central BP may be superior to brachial BP, brachial BP is already established as an independent and strong predictor of cardiovascular morbidity and mortality.1, 2, 3, 4, 5, 6 Thus, subjects with hypertension based on brachial BP category should be excluded from populations used to investigate characteristics of central systolic BP in near-normal subjects (without major risk factors). In line with this concept, population C consisted of subjects with optimal and normal brachial BP and no major cardiovascular risk factors. The values of central systolic BP in this population were lower than those in population A or B, and the age-dependent increase in central systolic BP was smaller in population C than in population A or B. Thus, a large part of the aging effects on central systolic BP disappeared after excluding subjects with increased brachial BP. Although normal values of central BP should be determined on the basis of the impact of central BP on cardiovascular mortality and morbidity in longitudinal studies, the distribution of central BP values obtained in population C may provide important information for discussing the normal value.

Epidemiological and cohort studies have clearly shown that brachial systolic BP increases with advancing age.1, 2, 3 As for brachial BP, however, classification of BP category is defined regardless of age in most hypertension guidelines,1, 2, 3 resulting in the majority of elderly population defined as having hypertension. A similar phenomenon is observed in the relationship between age and central systolic BP. Although we presented central systolic BP values in each age decade, this does not necessarily mean that the reference value of central systolic BP should be established in each age decade. As in the case for brachial BP, benefits from BP-lowering treatment in various ages and central systolic BP levels should be taken into consideration for future determinations of reference or normal values of central systolic BP. For the moment, in the clinical setting, central systolic BP may be assessed with age and brachial BP category taken into consideration, and then individual cardiovascular risk could be stratified by age, brachial BP and central systolic BP.

In this study, the characteristics of central systolic BP estimated by the Omron device were studied for the following reasons: (1) several devices based on different technologies are available for estimating central BP, and both reference and normal values need to be individually determined for each device; (2) central BP estimated by the SphygmoCor device is reported to be a better predictor of cardiovascular outcome than brachial BP7, 9, 14, 15 and highly correlated with central BP estimated by the Omron device;19, 20, 21 (3) central systolic BP estimated by the Omron device is closely and linearly correlated with simultaneously and invasively measured central aortic pressure;18, 21, 22 and (4) differences between invasively measured and noninvasively estimated central systolic BP are larger using the SphygmoCor device than using the Omron device.21 As the Omron device estimates central systolic BP in an automated manner, estimation of central systolic BP using this device is suitable for the clinical setting and is used widely spread in Japan, thus requiring reference values. Moreover, greater interindividual variability in central systolic BP than in brachial systolic BP throughout brachial BP categories observed in this study supports the clinical significance of the establishment of reference values for central BP besides those for brachial BP.27

The potential limitations of this study include that this was a cross-sectional study performed in a single institution and, at present, outcome data are not available. Second, although we provided the distribution of central BP values in each age decade, how to use these values in a clinical setting (diagnosis and management of hypertension) remains to be discussed. Third, central systolic BP was estimated only once in one subject and, thus, we could not exclude subjects with poor reproducibility of central systolic BP measurements. These points should be noted when interpreting the present data.

In conclusion, this study investigated the characteristics of central systolic BP estimated by the Omron technique and provided the distribution of these values. The present results contribute to individual risk stratification and make an important initial step in the implementation of central BP as a clinical tool in the diagnosis and management of hypertension.