Introduction

Several methods have been used to elucidate the corneal tissue biomechanics. The tensile strip extensiometry,1, 2, 3, 4, 5 involves applying a tensile force to strips of cornea held between two grips. Corneal biomechanical properties have also been measured by pressurizing the cornea within an artificial anterior chamber that uses internal pressure as the stress and by measuring the strain by means of marks on the cornea (bulge testing method).6, 7, 8, 9, 10, 11 Biomechanical measurements of excised donor human corneas report a plastic deformation of the cornea in vitro.5, 12 No attempt has been made until now to plastically deform the human cornea in vivo or to alter the collagen fibers of the stroma for corrective purposes.

Materials and methods

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional review board of the Hospital de la Cruz Roja in Barcelona and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. All applicable international, national, and/or institutional guidelines for the care and use of animals were followed.

Informed consent was obtained from all individual participants included in the study. Additional informed consent was obtained from all individual participants for whom identifying information is included in this article.

To plastically deform the living cornea we applied a vacuum to the cornea of eight rabbits and five human eyes with a novel device. The number of rabbits was selected based on accepted sample size calculators. The rabbits, three female and five male, were randomly selected from an albino colony. The small sample size was considered sufficient because a substantial effect was expected. The rabbits were 7 to 10 weeks old at the time of treatment. The novel device consists of a chamber of 11 mm in diameter. The chamber is radially divided in four interconnected sub-chambers. One side of the cylindrical chamber is closed, while the other side of the chamber is open, exposing the sub-chamber openings, Figure 1. When the open side is placed in contact with the cornea, the cornea closes and seals the sub-chamber openings. A vacuum of up to 760 mm-Hg was then applied to the chamber, exerting a force on the cornea. A vacuum of 760 mm-Hg means that the atmospheric pressure was removed from the front of the cornea, therefore a pressure of 760 mm Hg plus the intraocular pressure pushed the cornea outwards at the location of the chambers.

Figure 1
figure 1

The device used to alter the shape of the cornea consists of a vacuum chamber with openings that produce a force when placed on the cornea.

The rabbits were first treated with the device. The chamber was applied to their right eye and vacuum maintained for 5–10 min. The corneal topography of the eyes was measured pre-treatment and 3 days, three months, and 1 year post- treatment with a Placido based topographer, (Atlas 995; Carl Zeiss Meditec, Dublin, CA, USA). All measurements were blind in that they were made without knowledge of the particular rabbit and its prior readings. Pre-treatment and post- treatment inter-ocular differences in central corneal power were compared using the Student’s t-test of paired comparisons. The left, untreated eye was used as a control. The eyes were examined at the time of the topography when basic parameters of the eyes were also measured. After the conclusion of the one year post-treatment, rabbit 8 was also subjected to the same procedure at a pressure of 3 atm for 10 min to acquire further safety information. Rabbit corneas were sent for histological or micro-morphological analysis.

Human subjects were then selected based on the criteria described next. Subjects were 30 to 50 years of age. Their eyes were myopic in a range of 2 to 20 Diopters (D). No ocular disease was present. Only subjects with 1 D or less of astigmatism were included. A total of 5 subjects met the above criteria. Subjective refractions were performed by the same experienced refractionist, who had no knowledge of the study. Automated refractions were also used for objective confirmation of the refractive change. Visual acuity was recorded.

To detect potential safety issues of the procedure, the following exams were performed: Slit lamp exam, recording of Intraocular Pressure (IOP), pupillary reflex, direct and indirect ophthalmoscopy, Optical Coherence Tomography (OCT), and Confocal Laser Scanning Microscopy (CLSM).

We then confirmed, as described in results below, that the application of vacuum was also innocuous on the human eyes by applying vacuum at progressively higher pressures, of up to 760 mm Hg, and for longer durations, up to a five minute period. Thereafter, we applied a vacuum of 760 mm Hg for 5 min to the five left eyes of the volunteers. All treatments were done at a room temperature of 25 °C under topical anesthesia. Figure 2 is a photograph of a patient’s eye during the procedure.

Figure 2
figure 2

Application of the device to a patient’s left eye showing the chamber and vacuum line. The speculum is not necessary, but a convenience.

The refractions, visual acuity tests, and exams except OCT were done pre-treatment and 1 day, 2 months, and 2 years post-treatment. The treatment and examination were completed on each subject’s eyes before proceeding to the next subject. Therefore, if an adverse effect was detected, the study could be aborted.

Results

We report the changes in anatomy observed after the treatment. Corneal topography maps showed that the radius of curvature in the center of the cornea (in a zone of 3 mm of diameter) increased after treatment and decreased in the periphery. The corneal central radius of the intact rabbit eye was compared with the radius three days after treatment. The three-day change in the radius of the treated eye was found statistically significant. Pre-treatment radius of r=6.50 mm or 51.92 D, standard deviation, σ=0.80 D versus third day post-treatment radius of r=6.59 mm or 51.19 D, standard deviation, σ=0.67 D, p=0.006 (Table 1a). In Tables 1a, 1b, 1c and 1d, radius and power are the mean of the maximum and minimum values in the central cornea. The flattening of the cornea in diopters for each rabbit is shown in bar graphs in Supplementary Figure A. All corneas flattened after the treatment. About half of the corneas flattened about 1 D; one flattened about 1.75 D and three flattened about 0.25 D. We wanted to evaluate the long-term stability of the change in the radius of curvature without interference from the natural variation of over 10 D that occurs in the corneas of young rabbits. To accomplish that we compared the average radii of the right eye (OD) and left eyes (OS) three months after treatment and one year after treatment. All radii are shown in Table 1b. This comparison uses the non-treated (left) eye as the control. The pre-treatment difference between the average radius of the right eye (6.50 mm or 51.87 D) and the left eye (6.52 mm or 51.75 D) was not statistically significant. The same inter-ocular difference was statistically significant 3 months and 1 year after treatment. The comparison at the third month showed the following statistics: OD r=7.66 mm or 44.10 D, σ=0.70 D versus OS r=7.57 mm or 44.60 D, σ=0.50 D, p=0.026. The comparison at about one year post-treatment showed: OD r=8.23 mm or 41.01 D, σ=0.64 D versus OS r=8.13 mm or 41.50 D, σ=0.68 D, p=0.015. Corneal flattening was therefore stable and no significant regression of the effect was observed after three months. The flattening of rabbits’ corneas regressed to half a diopter three months post-treatment, and remained stable thereafter. This estimation of refractive change in rabbits is based on the correlation between the power of a corneal optical zone of 3 mm of diameter and refractive power.13 The rabbits tolerated the treatment well with topical anesthetic eye drops only. No injury was observed to the epithelium or Descemet layers with the slit lamp. The retinas also retained their normal appearance as seen with a direct and indirect ophthalmoscope. Histological analysis of the corneas did not evidence any Descemet tears, break in the collagen fibers, or any remarkable alteration. Supplementary Figure B is a histological slide of a corneal cross section of rabbit 8.

Table 1a Pre-treatment and Post-treatment corneal radii and power (rabbit)
Table 1b Post-treatment corneal radii and power at three months and a year (rabbit)
Table 1c Pre-treatment and Post-treatment OS corneal radii, power, refraction and corrected visual acuity (human)
Table 1d Post-treatment OS corneal radii, power, refraction and corrected visual acuity at two months and two years (human)

The treated human eyes evidenced no significant corneal epithelium damage, no increase in intraocular pressure, no decrease in corrected visual acuity and normal slit lamp appearance of the cornea and anterior chamber with no increased cell and flare. Ophthalmoscopy did not evidence change or abnormality and pupillary reflex was present. As the vacuum was increased up to 760 mm-Hg, the eyes were examined again. This time we included Optical Coherence Tomography (OCT) to provide some objective data of potential optic nerve and macula changes. None was observed. Eyes treated at full vacuum showed similar tolerance to the treatment and changes in corneal curvature. Two eyes of about −3 D flattened about 0.5 D. Two eyes of about −6 D flattened 1 D and 1.5 D. A highly myopic eye of −20 D flattened 5 D. The flattening was demonstrated by the corneal topography and by subjective and objective refraction methods. The curvature of the treated corneas displayed the same revolution regularity as before treatment. The topographical maps showed less curvature in the treated central cornea with increasing and then decreasing curvature towards the periphery. Astigmatism remained essentially unchanged in all subjects. Spherical refractions remained spherical. Figure 3 is a pair of representative topographic maps of the cornea of subject 5 before and after treatment. The flattening of the human corneas remained unchanged 2 months after treatment. Human eyes have been followed for 2 years, showing little or no regression in their subjective and objective refraction. Corrected distance visual acuity (CDVA) in the decimal scale ranged between 0.3 and 1.5 before treatment and 0.5 and 2 after treatment. CDVA was equal or better after treatment for every eye. See Tables 1c and 1d that list the mean radii and power, spherical equivalent refraction, and CDVA for all subjects.

Figure 3
figure 3

Topographic maps (in diopters) of left cornea of subject 5 pre-treatment (left) and one day post-treatment (right) with their corresponding color key.

The corneas were examined pre and post-treatment with a slit lamp and CLSM, which allowed visualization of corneal cells. The cornea thickness was also measured with a standard ultrasonic pachymeter. The eyes had a normal appearance when tested 24 h after treatment. The collagen fibers of the stroma appeared transparent and intact after treatment. CLSM revealed no changes in shape of endothelial cells, characteristic of dead cells. Supplementary Figure C is a CLSM image of the endothelial cells of subject 2 one day post-treatment. There was no significant change in cell density after 2 years. The thickness of the cornea was reduced by an average of 7%. The intraocular pressure pre-treatment and post-treatment was within normal range in all cases and no trend was present.

Discussion

The human and rabbit cornea are visco-elastic materials that continually deform or creep when strained beyond the elastic range. Such a deformation has been observed in strips of human and rabbit cornea in vitro.12 The rabbit and human cornea have substantial similarities in their elastic-plastic characteristics.5, 12 Anecdotal evidence such as permanent refractive anomalies after forceps delivery suggested that the human cornea can undergo plastic deformation.14

Stressing the cornea locally by the application of a vacuum produces a permanent deformation of the collagen fibers of the living cornea. This is the first time that collagen fibers have been permanently altered in vivo in a controlled manner. Numerous models have been developed to account for corneal deformation after refractive procedures, but they do not extend to plastic deformation.15 The corneal deformation described here changes corneal curvature, and therefore eye refraction. All corneas flattened and myopia was reduced in all human subjects. The extent of flattening of the cornea can be controlled by the shape and number of the device chambers, the level of vacuum, the duration and repetition of the procedure and the temperature, alone or in combination.

The corneal temperature is around 35 °C at 25 °C room temperature. By adjusting ambient temperature from 40 to 80 °C, the corneal temperature will vary from 37 to 42 °C.16, 17 For a given plastic strain of the collagen fibers in the rabbit sclera, the higher the temperature, the greater the strain.18 The rabbit sclera at 39.5 °C permanently strains 1.5% with a stress pressure of only 100 mm Hg.19 Assuming that the human cornea behaves the same way, elevating the temperature of the cornea could be an easy way to increase and control the flattening effect and the myopia corrected.

The reduction in the thickness of the central cornea was not unexpected. The cornea (as most materials) thins when strained. This effect can be permanent and has been observed even at near physiological intraocular pressures.20 Stressed human corneas have a Poisson ratio of about 0.5; their thickness reduces by 50% of what they strain.21 The central cornea strained radially as a result of peripheral vacuum stress in that direction because the central cornea and limbus were opposing that stress. Although it is undesirable that the cornea strains beyond the elastic range in the radial direction, it is unavoidable with a few large sub-chambers, needed for increased stress.

Myopia affects about one quarter of the western adult population, and more than one half of the population in some Asian countries.22 Myopia appears to be a feedback process amenable to control.23 Preventive methods are rarely used, and depend on the application of accurate models of myopia. Several surgical methods have been tried to correct myopia, but none is simple, or free of risks. Current surgical methods to correct myopia achieve their goal by invasively modifying the corneal curvature. Methods which are used, have been used or are being developed include: radial keratotomy24 (radial incisions made in the periphery of the cornea), excimer laser photorefractive keratectomy (PRK)25, 26 (laser ablation of the anterior surface of the cornea), laser assisted in situ keratomileusis or (LASIK),27 Small Incision Lenticule Extraction (SMILE), implantation of intra ocular ring or lens28 and potentially, customized Corneal Collagen crosslinking (CXL). We call this new procedure 'Pneumatic Keratology.' The vacuum force flattens the central part of the cornea by distending the stressed areas where vacuum is applied, in a similar manner as radial keratotomy incisions extend the cornea perpendicularly to the length of the incisions.24

The duration and pressure applied corrected a fraction of the human subjects’ myopia, up to 5 D in one case, and close to 1 diopter (average 0.73 D) in the rabbits. The flattening of rabbits’ corneas regressed to half a diopter three months later, and remained stable thereafter. The plastic property of the cornea could be used to correct not only myopia, but hyperopia and other refractive conditions. Our experiments show that the same vacuum force has different effects depending on the degree of the subject’s myopia. The implication is that the higher the myopia, the less the ocular rigidity. Researchers have hypothesized that myopes have a weaker sclera prone to elongation.29 The weakened sclera of the rabbit exhibited a continuous elongation.19 Although our experiments were done on the cornea, it is known that sclera and cornea are very similar anatomically and mechanically.21, 29, 30, 31 Our results, therefore, support the view that myopic eyes have weaker corneoscleral envelopes, and that their weakness is proportional to their myopia.

The way we produced a flattening of the cornea by the application of a vacuum force was by using openings elongated in the radial direction. An elongated opening produces a bulging of the cornea when the vacuum is applied, as depicted in Figure 4. The curvature of the deformed cornea in the elongated direction is necessarily less than in the perpendicular, short direction because α<β. If the curved sections of the deformed cornea in the elongated and short directions in Figure 4 are approximated by circular segments, it is evident that the short direction strains more than the long direction. If the elongated openings are arranged radially, as with the device described here, the cornea can be made to strain elastically in the radial direction, while plastically in the circumferential direction. The chambers produced a bulging of the cornea not in the center but in the periphery. That is the result intended and achieved here with the device. The foregoing analysis is consistent with the analysis of stress in the spheroidal sclera. Greene reported that stress in the direction of the higher curvature in the equator of the sclera is greater than in the direction with less curvature.32 Although we preferred elongated openings, other openings also produce a net flattening of the cornea, as we have confirmed experimentally. Treatment of cadaver eyes with different openings and the variations in procedure described above has demonstrated greater flattening of the cornea that corrects up to 8 D of myopia. These experiments will be published in a separate report when completed.

Figure 4
figure 4

The strain (b'-b)/b caused by an elongated opening in the short direction of length b is greater than the strain (a'-a)/a in the long direction a. Since the curvature of the arc of length b' is greater than the arc of length a' (because a>b), then b'/b>a'/a, which implies that (b'-b)/b>(a'-a)/a.

The maximum vacuum that can be applied to the eye is limited by the pressure outside the eye, normally the atmospheric pressure, or about 760 mm-Hg. We have shown that such a pressure is sufficient to plastically deform the human cornea. In vitro data from Hoeltzel et al5 confirms that the rabbit cornea subjected to an equivalent intraocular pressure of about 760 mm-Hg permanently strains by 6%, while the human cornea subjected to only 400 mm-Hg permanently strains by about 1%. Permanent strain of the human cornea in vitro has been observed even under physiological range stress.20

Conclusions

The plastic, and therefore permanent, deformation of the cornea was obtained by straining the cornea with a vacuum force. A vacuum chamber with radial openings flattens the cornea. A flatter cornea corrects or reduces myopia.

The ability to modify collagen tissue in a controlled manner has implications in organs beyond the eye. Tendons, cartilage, heart valves and blood vessels for example, conceivably could be custom modified for corrective purposes.