Introduction

Glaucoma is traditionally defined as a chronic, progressive optic neuropathy with characteristic optic nerve head (ONH) and retinal nerve fiber layer (RNFL) changes, with or without correlating visual field (VF) defects. However, with the advent of spectral domain optical coherence tomography (SDOCT), another region that has been shown to demonstrate structural changes in glaucoma is the macula. Imaging the inner retinal layers at macula has evolved as a useful modality to diagnose glaucoma.1, 2, 3, 4, 5, 6, 7, 8, 9 Standard automated perimetry (SAP) has been the preferred method to evaluate the defects in the central 24° or 30° VF in glaucoma. The commonly used program to evaluate the central 24° of VF tests the visual sensitivities at 54 points that are 6° apart. However, the macular region (central 5°–10°) is tested using only 8–16 points in this program which are 6° apart. Also, the locations of the ganglion cells stimulated by the central 16 points of the 24° program are farther from the fovea because the ganglion cells in the fovea are displaced.10, 11 The sampling density of central 24° program to estimate visual sensitivity at macula is also inadequate. A study by Schiefer et al12 has shown better detection of VF defects in eyes with suspected glaucoma when closely spaced test points were used. In addition to the 24° program, SAP also has a program to examine the central 10° of the VF with higher resolution. This program evaluates the central 10° of VF using 68 test points that are 2° apart. Traynis et al13 have reported field defects on 10° program in eyes with no defects on 24° VF program.

In addition to the SAP, another relatively newer method of estimating visual sensitivities at retinal points is the microperimetry (MP). MP, which is also known as fundus perimetry, assesses retinal sensitivity while directly examining the ocular fundus. MP results are independent of eye movements and exactly related to the stimulated area.14 Therefore, the VF sensitivities measured by MP is supposed to have better spatial localization.

Although there are multiple studies evaluating the diagnostic ability of inner retinal layers at the macula in glaucoma (as referenced earlier), there are no studies comparing the diagnostic abilities of inner retinal layers with that of the functional evaluation of the macula with either the SAP or MP. The purpose of our study was to compare the abilities of structural (inner retinal layer thickness measured using SDOCT) and functional (visual sensitivities measured using SAP and MP) assessments of macula in diagnosing glaucoma.

Materials and methods

This was a prospective, cross-sectional study conducted at a tertiary eye care center between January 2015 and April 2015 to evaluate the structural and functional changes at macula in glaucoma. Written informed consent was obtained from all participants and the Ethics Committee of LV Prasad Eye Institute approved the study methodology. All methods adhered to the tenets of the Declaration of Helsinki for research involving human subjects.

Inclusion criteria were age ≥18 years, best-corrected visual acuity of 20/40 or better and refractive error within ±5 D sphere and ±3 D cylinder. Exclusion criteria were the presence of any media opacities that affected SDOCT scans and SAP and MP results, and any retinal (including macular) or neurological disease other than glaucoma, that could confound the evaluations. All participants underwent a comprehensive ocular examination, which included a detailed medical history, best-corrected visual acuity measurement, slit-lamp biomicroscopy, Goldmann applanation tonometry, gonioscopy, dilated fundus examination, central VF examination with SAP and MP, and SDOCT imaging with Cirrus high-definition OCT (HD-OCT; Carl Zeiss Meditec Inc., Dublin, CA, USA).

SDOCT examination was performed with Cirrus HD-OCT (software version 6.0). Macular Cube 200 × 200 protocol was used to image the macula. This protocol has been explained in detail previously.15, 16 Ganglion cell analysis (GCA) is a software that measures the ganglion cell-inner plexiform layer (GCIPL) thickness within a 14.13 mm2 elliptical annulus centered on the fovea with an inner vertical radius of 0.5 mm and outer vertical radius of 2 mm, stretched horizontally by 20%. The thickness parameters derived from GCA are the average GCIPL thickness across the entire elliptical annulus and the thickness at six 60° sectors of the elliptical annulus. Scans with signal strength <6 or with segmentation errors were excluded from analysis.

SAP examination of the macula was performed using a Humphrey Field analyzer, model 750i (Zeiss Humphrey Systems, Dublin, CA, USA), with the Swedish interactive threshold algorithm standard 10-2 program. VFs were considered reliable if the fixation losses were ≤20% and false-positive and -negative response rates were ≤15%.

MP examination of the macula was performed using the Macular Integrity Assessment (CenterVue, Padova, Italy). The protocol used was a 10-2 grid that tests the retinal sensitivity in the central 10° of the retina using 68 points, similar to the 10-2 program of SAP. The stimulus size was equal to the Goldmann size III stimulus of SAP. Background luminance was set at 4 apostilb, with a maximum luminance of 1000 apostilb and a stimulus dynamic range of 36 dB. Visual sensitivities were estimated using a 4-2 threshold strategy. MP results were considered reliable if the reliability index was >70% (percentage of control points projected on the optic nerve that were not seen by the subject) and the fixation was stable (>75% of the fixation points were located within a 2° diameter circle centered in the gravitational center of all fixation points).

To evaluate the GCIPL thickness and visual sensitivity at exactly similar locations at macula, we used the structure–function correlation map described previously using the GCIPL sectors of Cirrus HD-OCT and the 10-2 program of SAP and MP.17 This map also considered the displacement of the RGCs at the macula by using equations derived from the histological analysis to approximate the location of the RGCs with each SAP test point.10 Visual sensitivities at all 68 points of SAP and MP were first converted from the dB scale to a linear scale (reciprocal of Lambert scale) using the following formula.

The values from all test points within the VF sectors corresponding to the anatomic sectors described above were then averaged. The average visual sensitivity per sector was converted back to the dB scale for the analysis. This procedure has been explained in detail previously.18 The sectors of SAP were flipped horizontally and vertically to correlate with the sectors of HD-OCT and MP.

Glaucoma group consisted of a consecutive sample of eligible glaucoma patients attending the hospital during the study period. Glaucoma was diagnosed in all these eyes by the treating glaucoma specialist based on the characteristic optic discs findings of neuroretinal rim thinning, notching, or localized or diffuse RNFL defects. We also included a control group that consisted of subjects attending the hospital for a routine eye check for refractive error or were spouses and friends of the recruited patients. Control subjects had no family history of glaucoma, no history of raise in IOP in the past, normal anterior and posterior segments as assessed by experts on clinical examination, and IOP ≤21 mm Hg. Neither VF nor SDOCT results were used in the definition of glaucoma. Both the VF examinations (SAP and MP) and the SDOCT examination were carried out on the same day in all the subjects.

Statistical analysis

Descriptive statistics included mean and standard deviation for normally distributed variables and median and interquartile range (IQR) for non-normally distributed variables. Receiver-operating characteristic (ROC) curves and sensitivity at a fixed specificity of 95% were used to describe the ability of the different structural and functional parameters to discriminate glaucomatous eyes from control eyes. To obtain confidence intervals for area under the ROC curves (AUCs) and sensitivities, a bootstrap resampling procedure was used (n=1000 resamples). As measurements from both eyes of the same subject are likely to be correlated, the standard statistical methods for parameter estimation lead to underestimation of standard errors and to confidence intervals that are too narrow.19 Therefore, the cluster of data for the study subject was considered as the unit of resampling and bias-corrected standard error was calculated during all estimations. This procedure has been used in literature to adjust for the presence of multiple correlated measurements from the same unit.20, 21 Z-test was used to compare the AUCs and tests of proportions were used to compare the sensitivities at 95% specificities of structural and functional parameters.22, 23

Statistical analyses were performed using the commercial software (Stata ver. 12.0; StataCorp, College Station, TX, USA). A P-value <0.05 was considered statically significant.

Results

One hundred and twenty-seven eyes of 80 participants (53 eyes of 31 control subjects and 74 eyes of 49 glaucoma patients) underwent GCIPL imaging with HD-OCT and macular visual sensitivity assessment with SAP and MP. After excluding the poor quality HD-OCT, SAP, and MP tests (shown in the Venn diagram; Figure 1), final analysis included 108 eyes of 72 participants. Table 1 shows the age, GCIPL thickness, SAP, and MP sensitivities of the glaucoma and the control groups. Control subjects were significantly younger than the glaucoma patients. AUCs and sensitivities at fixed specificity, therefore, were adjusted for the differences in the age between the control and the glaucoma groups using covariate adjustment, as proposed by Pepe.24 On evaluating the 24-2 SAP results and the mean deviation (MD) of the 24-2 SAP of the 61 eyes with the glaucoma, 9 had preperimetric, 15 early (MD of better than −6 dB), 16 moderate (MD between −6 and −12 dB), and 21 severe (MD of worse than −12 dB) VF loss.

Figure 1
figure 1

Venn diagram showing the number of eyes excluded from the analysis because of poor test results.

Table 1 Structural and functional characteristics of the participants

The AUCs of structural and functional parameters to differentiate glaucoma from control eyes are shown in Table 2. Among the parameters, inferotemporal, inferior, and superotemporal sectors along with the overall average measurements of GCIPL thickness and visual sensitivities showed the best AUCs to diagnose glaucoma. Table 2 also shows the P-values associated with the comparisons between the structural and functional parameters. Although the AUCs of most GCIPL parameters were better than that of the functional measurements of SAP and MP, none of the differences were statistically significant.

Table 2 Area under the receiver-operating characteristic curvesa of the structural and functional parameters to diagnose glaucoma

Sensitivities at 95% specificity of structural and functional parameters to differentiate glaucoma from control eyes are shown in Table 3. Among the parameters, inferotemporal, inferior, and superotemporal sectors along with the overall average measurements of GCIPL thickness and visual sensitivities showed the best sensitivities to diagnose glaucoma. Table 3 also shows the P-values associated with the comparisons between the structural and functional parameters. Sensitivities of GCIPL parameters were significantly better than that of the corresponding visual sensitivity parameters of SAP and MP.

Table 3 Sensitivity at 95% specificitya of the structural and functional parameters to diagnose glaucoma

Figure 2 shows the ROC curves of the structural and functional parameters for the superotemporal (a), inferotemporal (b), and inferior (c) sectors and overall average (d) measurement.

Figure 2
figure 2

ROC curves showing the diagnostic abilities of ganglion cell-inner plexiform layer (GCIPL) thickness and standard automated perimetry (SAP) and microperimetry (MP) sensitivities of (a) superotemporal sector, (b) inferotemporal sector, (c) inferior sector, and (d) overall average measurement.

Discussion

In this study to compare the diagnostic abilities of structural and functional tests at macula in glaucoma, the GCIPL thickness measurements with HD-OCT performed significantly better than the visual sensitivity measurements by SAP and MP.

Studies comparing the structural and functional evaluation of macula in diagnosing glaucoma are sparse. Ozturk et al25 compared the macular sensitivities with 10-2 SAP and an older version of MP, and the macular retinal thickness with confocal scanning laser ophthalmoscope (CSLO) in a group of glaucoma patients (glaucoma defined based on optic disc changes and VF defects on 24-2 SAP).25 They found that the AUCs of the mean sensitivities of 10-2 SAP and MP in diagnosing glaucoma were 0.75 and 0.65, respectively, and that of CSLO was 0.59. They concluded that the functional defects occur before anatomic defects in glaucoma.25 In another study, Klamann et al26 compared the diagnostic abilities of the mean sensitivity of MP with that of the RNFL thickness of OCT and CSLO in preperimetric glaucoma (defined as glaucomatous optic neuropathy with normal 24-2 SAP results). They, in contrast to the previous study, found that the AUC of the mean sensitivity with MP was 0.59, while that of the RNFL thickness with CSLO and OCT were 0.70 and 0.89, respectively. They concluded that the morphological changes appear earlier than the functional changes in preperimetric glaucoma.26

A few earlier studies comparing SAP with MP have also reported better ability of MP to detect VF defects compared with SAP. In 20 glaucomatous eyes with paracentral VF defects (defects involving the central 10° of VF on a 24° test), Lima et al27 found that the visual sensitivities of MP in the central 10° of VF not only correlated significantly with those from 10-2 SAP but also detected abnormal areas of retinal sensitivity in quadrants with normal SAP values. They also found that 75% of the areas showing abnormal visual sensitivity with MP and normal sensitivity with SAP showed a corresponding reduction in macular thickness on time domain OCT.27 In a previous study with the first version of MP, Orzalesi et al28 had also found reduction in visual sensitivities with MP in areas of localized RNFL defects but with normal SAP results. However, we found no consistent difference between the diagnostic abilities of MP and SAP. The AUCs of all SAP and MP visual sensitivity measurements were comparable. Sensitivity of SAP at 95% specificity was better than that of MP in the superior and superonasal sectors while the same was lesser than that of MP in the inferonasal sector.

An important difference between our study and most of these previous studies is the way cases and controls were defined. It is difficult to define an independent reference standard for glaucoma when comparing structural and functional tests without favoring either one of them.29 Previous studies defined glaucoma based on the presence of ONH changes as well as VF defects on SAP.25, 27 This definition is likely to favor the functional measurements. We defined glaucoma based on the clinical examination of optic disc and RNFL. Our definition was not dependent on the results of SAP or SDOCT. The definition of glaucoma cannot be based on clinical examination of the macula at this point in time. We therefore believed that the definition of glaucoma in our study neither favored the structural nor the functional tests. However, the possibility of our definition favoring the structural test (GCIPL measurement) cannot be ruled out completely as a previous study by Kim et al30 has shown that the diagnostic ability of GCIPL parameters is affected by the topographic profiles of RNFL defects. We however did not evaluate the profile of RNFL defects in our study and therefore cannot rule out this possibility completely. Our study is an initial step in comparing the structural and functional tests at macula in diagnosing glaucoma and should not be extrapolated to the situation of detecting disease in suspects. Future studies should evaluate the structural and functional changes at macula in suspects with high IOP (who would have a higher prior probability of glaucoma) against a normotensive control group.

One of the limitations of the study is the small sample size. The other limitation was the inclusion of two eyes of a few subjects for the analysis. We however accounted for the correlation between two eyes of subjects using validated statistical methods during the analysis. We also ran the analysis with one randomly chosen eye per subject and found the results to be essentially the same. We also did not evaluate the normative database classification provided by SDOCT and its agreement with SAP and MP results in this study. There are two reasons for this. First is that the diagnostic ability of the normative database classification of GCIPL parameters has been reported by previous studies.9, 31 Second is the fact that there are no validated normative database classification systems with the central 10° program of either SAP or MP for diagnosing glaucoma. Future studies should also focus on developing and evaluating different classification systems based on the sensitivity or the probability maps of 10-2 SAP and MP for diagnosing glaucoma.

In conclusion, on evaluating the diagnostic abilities of structural and functional tests at macula in glaucoma, GCIPL thickness measurements with HD-OCT performed significantly better than the visual sensitivity measurements by SAP and MP.