Introduction

It is widely accepted among vitreoretinal surgeons that removal of the internal limiting membrane (ILM) during surgery for tractional maculopathies improves both functional and anatomical success1, 2, 3, 4 and may help to decrease the incidence of recurrences of the disease.4 Peeling of the ILM is a challenging procedure and requires surgical skills and experience. In order to facilitate visualization of this transparent structure and to reduce excessive and prolonged manipulation on the retina, staining substances have been introduced over the last decade. The first dye suggested was the cyanine dye indocyanine green (ICG)5 in the year 2000. ICG became a subject of debate because of its potential toxicity, most likely related to its chemical instability and a photosensitizing effect that has been demonstrated experimentally,6, 7, 8, 9 which may lead to functional deficits in clinical practice.10, 11 A few years later, the triarylmethane dye brilliant blue became commercially available. On the basis of clinical and experimental studies published so far, brilliant blue seems to provide an improved safety profile compared with ICG.12, 13, 14, 15, 16 Very recently a comparative study evaluating the microstructure of the inner and outer retina and visual function after macular hole surgery using brilliant blue or indocyanine green revealed that postoperative visual acuity and retinal sensitivity were better in eyes after BBG-assisted vitrectomy.17

However, regardless which dyes the surgeon may choose for the surgical intervention, it has been reported that the stained ILM provides an increased rigidity and can be peeled off more easily and in larger fragments as compared with the unstained tissue.15

Atomic Force Microscopy (AFM) is a well established examination technique that has been used to measure the thickness and rigidity of human ocular basement membranes such as the unstained ILM.18, 19 The aim of the present investigation was to assess and quantify in vitro changes of the ILM of the human retina after staining with either brilliant blue or ICG with and without subsequent illumination using a standard surgical light (PENTA LUX x 50, Ophthalmologische Systeme GmbH Fritz Ruck, Eschweiler, Germany) source compared with unstained ILM using AFM.

Methods

The study was approved by the local Institutional Review Board.

Tissue preparation and handling

Specimens of unstained human ILM were obtained during vitreoretinal procedures to treat macular holes and epiretinal membranes. Vitreoretinal surgery consisted of a standard 23-gauge vitrectomy, induction of a posterior vitreous detachment, and removal of epiretinal membranes if present using a pair of end-gripping 23-gauge forceps. This was followed by a peeling of the unstained ILM (surgeon AK), which was immediately placed in BSS and handed over to a technician (RS) for further processing.

Immediately after removal from the eye, the ILM specimen was placed on a glass slide and divided into five parts. Each piece was placed inside of a BSS droplet on a microscope slide (SuperFrost Plus, Gerhard Menzel GmbH, Braunschweig, Germany) that had been coated with poly-L-lysine. BSS was removed slowly while the membrane was carefully smoothed on the glass using a glass needle avoiding dehydration of the specimen during preparation. A circle was then drawn around each ILM fragment with a PAP pen. Then, a drop of brilliant blue 0.025% (Brilliant Peel; Fluoron GmbH, Neu-Ulm, Germany) or indocyanine green 0.05% was placed on two fragments each. ICG dye was prepared as follows: 25 mg of dry ICG substance (Pulsion, Munich, Germany) was first dissolved in 5 ml sterile water as recommended by the manufacturer. One microliter of this 0.5% solution was then diluted with 9 ml BSS plus resulting in a final concentration of 0.05%. Excessive dye was then rinsed off after 1 min. One of these specimens was then illuminated using a standard light source used for vitreoretinal surgery (PENTA LUX x 50, Ophthalmologische Systeme GmbH Fritz Ruck) for one additional minute. The remaining unstained fragment served as a control. Therefore, atomic force microscopy could be performed in five fragments of each specimen: an untreated and unstained control, one after staining with brilliant blue, one after staining with brilliant blue and subsequent illumination, one after staining with ICG, and one after staining with ICG and subsequent illumination. All in all, 20 ILM specimens were analyzed for each dye, with the retinal side (BBG [n=8]; ICG (n=11)) and the vitreal side (BBG (n=12); ICG (n=9)) being evaluated separately.

ILM specimens were also stained for laminin after AFM measurements in order to determine the orientation (vitreal or retinal side) on the glass slide. For this purpose, the ILM samples were first washed three times with 0.01 molar phosphate buffered saline (PBS). In a second step, a droplet of 120 μl 4% paraformaldehyde was placed onto the membrane for about 15 min. After another washing step, 120 μl laminin labeling antibody 1 : 100 (Polyclonal Rabbit Anti-Laminin, Dako Deutschland GmbH, Hamburg, Germany) was added, and the sample was stored in a humidity box at 8 °C for 24 h. Thereafter, a third washing step was performed followed by the staining with 100 μl cy3 fluorescent dye-coupled secondary antibody 1 : 100 (Donkey anti-Rabbit IgG (H+L)-Cy3, Dianova GmbH, Hamburg, Germany) that was applied for 45 min. All samples were rinsed again in PBS, and 20 μl of mounting medium (Fluorescence Mounting Medium, Dako Deutschland GmbH) was applied onto the ILM, and then covered with a round cover slip and allowed to solidify at room temperature for about 3 h.

All ILM specimens were also evaluated using a fluorescence microscope. All fluorescence imaging experiments were performed using an Axiovert 25 inversion fluorescence microscope (Carl Zeiss AG; Oberkochen, Germany). As a light source, a mercury vapor lamp (HBO 50; Osram, Munich, Germany) was used. Using the excitation filter ET545/25x (Chroma; Taoyuan Hsien, Taiwan), the 546 nm peak in the mercury spectrum was selected. The light was reflected by the dichroic T565LP (Chroma) through the objective (Fluar 10x/0.50; Carl Zeiss AG) onto the sample and excited the cy3 dye used for labeling. The emitted light passed the dichroic, passed the emission filter ET605/70 m (Chroma) and was detected using a CCD camera (Infinity 2; BFI Optilas; Dietzenbach, Germany) and processed using a computer. The camera was controlled by the software ‘Infinity Capture 5.0.0’ (Lumenera, Leipzig, Germany).

Atomic force microscopy measurements

All AFM imaging and force indentation experiments were performed using a Nanowizard II Atomic Force Microscope (JPK Instruments, Berlin, Germany), which was mounted on an inverted microscope. For imaging and stiffness measurements with Nanowizard II AFM, standard silicon nitride triangular cantilevers (MLCT, Veeco Instruments, Plainview, NY, USA) with integrated sharp silicon–nitride pyramidal tips (nominal spring constant k=0.1 N/m, apex angle α35°) were used. Before every measurement, the spring constant value was determined using the thermal fluctuation method. The samples were imaged in contact mode with a scan rate of 0.6 Hz while kept in PBS at room temperature. Two scan regions of 10 × 10 μm were chosen on the sample. Stiffness was determined using AFM in a force spectroscopy mode. Briefly, indentations were made over an 8 × 8 point grid at a rate of one load/unload cycle per 2.4 s with a maximal load of 5 nN. The force curves were plotted with an approaching speed of 3 μm/s, a retraction speed of 7 μm/s, and a data rate of 5000 Hz. Assessment of the dynamic indentation modulus E of each force-indentation curve was plotted off-line using a data processing software (JPK Instruments AG). Therefore, the height data of the scan regions were calculated to tip-sample separation and plotted against indentation force. After fitting the Sneddon model19 to the data, the resulting elasticity modulus was averaged for each scan region.

The stiffness of an elastic material is quantified by Young’s modulus (tensile modulus, elastic modulus, modulus of elasticity), a material property used to characterize materials. The SI unit of Young’s modulus is the pascal (Pa or N m2).

Identification of ILM sample orientation

During the measurements it became necessary to determine the orientation of the sample on the glass slide. Preliminary measurements on ILM specimens stained during vitreoretinal surgery within the human eye and removed for analysis confirmed that there is a difference in elasticity on the retinal and on the vitreal side of the ILM as described in the literature.20, 21 As a consequence, a reliable orientation of the ILM specimens for each measurement is crucial in order to obtain reliable results. In our study, AFM imaging was used to clearly identify the rough retinal surface and the smooth vitreal surface as topographic landmarks (Figure 1a).

Figure 1
figure 1

(a) AFM contact mode images with a size of 80 × 20 μm of both ILM sides. The smooth vitreal side can be clearly distinguished from the rough and undulated retinal side (roughness retinal: 9075 nm, vitreal: 2672 nm). The deflection signal is shown in units of nm. (b) ILM stained for laminin and imaged using fluorescence microscopy (magnification: × 10). The smooth vitreal and rough retinal side can be determined easily. The ILM sample is partially flipped over, and hence the difference of both sides can be appreciated.

Additional tests for correct tissue orientation were performed using a protocol for staining for laminin and consequent fluorescence microscopy (Figure 1b). This was performed after the AFM measurements had been obtained to exclude potential effects of the illumination by the microscope on the result.

Results

Staining of the ILM-samples resulted in a statistically significant increase in tissue rigidity compared with controls (Figure 2). Of note, the retinal surface of all specimens, both control and stained tissue, was about fivefold stiffer than the vitreal surface. Compared with unstained controls of ILM, the stiffness of the tissue significantly increased with the use of brilliant blue or ICG for staining. This was true both for the vitreal surface (BBG: P<0.001; ICG: P<0.01) as well as the retinal surface (BBG: P<0.01; ICG: P<0.01). The difference in stiffness comparing the retinal versus the vitreal surface seen in the controls remained unchanged after staining of the tissue and subsequent illumination. Calculated for each dye, there was a trend to a further increase in stiffness after illumination of the stained tissue, but this effect did not reach statistically significant levels (Figure 2, Table 1). Comparing the two dyes after illumination, there were statistical differences seen depending on which side was measured (BBL versus ICGL measured at the vitreal side, Table 1).

Figure 2
figure 2

Averages of all map-results of ILM are plotted with respect to the measured side (vitreal: blue; retinal: red), with error bars showing the SE of the mean. Indicated P-values hold a significant level of 5% and are calculated from a t-test. BB, brilliant blue; BBL, brilliant blue and illumination; ICG, indocyanine green; ICGL, indocyanine green and illumination; WT, control. Differences between control (WT) and BBL and ICGL were also significant both for the retinal as well as the vitreal surface.

Table 1 Significant levels comparing dyes and controls with regard to the increase of stiffness measured at the retinal and vitreal surface of the ILM

Relative elasticity changes and the effect of Illumination

Comparing the Young’s moduli in relation to the corresponding control sample, we noted that except for ILM samples after brilliant blue staining and illumination (BBL) the relative elasticity change of the vitreal side was more pronounced compared with the retinal side (Figure 3, Table 2). Of note, values for ICG with and without illumination were higher compared with brilliant blue in ILM specimens. In addition, except for BB/BBL, the relative elasticity of stained specimens increased by 1.2-fold following illumination (retinal surface: 1.18±0.02 for BB/BBL, 1.23±0.04 for ICG/ICGL; vitreal side: 0.69±0.05 for BB/BBL and 1.21±0.01 for ICG/ICGL). This implies an additional stiffening effect of the tissue following illumination, which was not statistically significant. The gradient was around 1.2 for all dyes and tissues (except for vitreal BBL) (Table 3).

Figure 3
figure 3

Relative elasticity changes of the ILM measured on the vitreal and retinal side of the tissue. BB, brilliant blue; BBL, brilliant blue and illumination; ICG, indocyanine green; ICGL, indocyanine green and illumination; WT, control.

Table 2 Relative elasticities: each Young’s modulus is divided by its respective control value
Table 3 Gradients between the dye-specific lightexposed and unexposed samples

Discussion

The ILM of the retina is a multilaminar structure mainly composed of type IV collagen, laminin, and fibronectin and essentially represents the basement membrane of Müller cells. Although the inner vitreal surface of the ILM appears smooth, the ILM has a rough, undulated outer retinal surface, a finding being most remarkable at the posterior pole. This morphological difference helps to identify the respective side of the ILM in histological specimens as performed in the present study and helps to correlate ILM samples with regard to their localization. The ILM is adherent to the type II collagen fibers of the vitreal cortex,22 helping vitreal tractional forces to be transmitted to the retinal surface, contributing to the pathogenesis of several tractive maculopathies such as macular pucker, macular hole, or vitreoretinal traction syndrome. Therefore, ILM removal is considered to be an important step to successfully treat these conditions and prevent recurrences in order to remove all tractional forces adherent to the ILM.

Of note, the ILM is a very delicate structure that varies in thickness, being thicker posteriorly (0.5–3.2 μm) and thinner at the fovea and the disc (0.01–0.10 μm).23 The ILM is a surprisingly rigid structure. Previous AFM studies showed that the mechanical strength is in the mPa range, very similar to the articular cartilage and about 1000-fold stronger than cell layers.18, 24, 25 Our results underline previous reports20 describing that the average stiffness of the ILM is over fivefold higher at the retinal side as compared with the vitreal side. This difference nicely explains the fact that the ILM curls up towards its vitreal side after peeling. The high mechanical strength of the natural, unstained ILM gives the possibility to grasp the ILM using an end-gripping forceps and to peel off mechanically the ILM from the underlying nerve fiber layer.

As unaided, unstained ILM peeling requires great surgical skills. Staining substances were introduced during the last years that greatly facilitate the visualization and removal of the ILM during vitreomacular surgeries. Both dyes in clinical use, ICG and BBG, provide selective staining properties and a good enough contrast to identify and peel off the ILM. The advantage of ILM staining is obvious. Better visualization is supposed to result in an easier removal without damage to the underlying nerve fiber layer and consequently better functional results postoperatively.

However, it is a commonly reported intraoperative observation that stained ILM can be peeled of more easily and in larger fragments as compared with the unstained ILM. On the basis of the present study, we hypothesize that there are specific tissue dye interactions that increase the stiffness of ILM, and we conclude that the observed change in stiffness is a very likely explanation for the easier removal of ILM after staining. The effect appears more pronounced when using ICG as compared with BBG. Interestingly, illumination using a standard surgical light source did not contribute a significant effect to even further increase the stiffness of the tissue. However, except for brilliant blue, illumination increased the stiffness in all examined tissues and irrespective of the measured surface in a very consistent manner by 1.2-fold. Interestingly, the change of elasticity was more pronounced at the vitreal side compared with the retinal side. This may be explained by the presence of collagen remnants adherent to the vitreal side of the specimens that further increase the stiffening effect after staining using the dye.

One may argue that the easier removal after staining is merely related to the better visualization of the ILM. Better visualization undoubtedly helps to identify the ILM and initiate the peeling at the correct plane of dissection, but does not explain the altered mechanical properties as such. Alternatively, one may suggest that the adhesion of the stained ILM is decreased, implying that there is an alteration of the cleavage plane or some kind of loosening of the ILM from the underlying nerve fiber layer as an effect of the dye applied. The issue of the alteration of the cleavage plane was extensively discussed in the literature as a possible effect of ICG-related phototoxicity.10, 26 Under the light of the present study and its results, this hypothesis must be challenged as BB and ICG reveal quite similar changes concerning their mechanical properties, and no phototoxicity is known for BB (see below).

However, the underlying mechanism for the observations described in the present study remains hypothetical. Looking at the published literature there are some studies describing an alteration of the mechanical properties of the ILM after staining: For ICG, Wollensak et al.27, 28reported an increased stiffness of the ICG-stained ILM following 3 min of illumination of the stained post-mortem porcine retina, showing a significant increase in ultimate force by 45% and a decrease in ultimate elongation by 24%. Interestingly, no such effect was seen without subsequent illumination, suggesting a light-dependent process. They, therefore, concluded that the stiffening effect of ICG combined with light is related to a photosensitizing effect of ICG by the formation of a triplet state of the ICG molecule and reactive oxygen species (type I reaction of photooxidation).27, 29 Reactive oxygen species then lead to photooxidative damage of cells and physical cross-linking of collagen fibers such as the type IV collagen of the ILM.27, 28, 30 However, in our own experiments using AFM in fresh human tissue, an increase in stiffness was seen also for ICG without subsequent illumination maybe because the specimens were not kept in darkness as described for Wollensak’s experimental setting.27 It is widely accepted that the potential toxic effects of ICG10, 11, 31, 32 as described clinically are very likely related to photosensitizing properties of ICG as described above and the decomposition products of the ICG molecule after illumination.7, 8, 33

Importantly, other additional relevant tissue–dye interactions seem to have a role as a causative effect for the increase in tissue stiffness, as increased rigidity was also seen for BBG in our experiments. In contrast to ICG, BBG, a triarylmethane dye, has no known photosensitizing properties. This observation is of relevance as histological evaluations of ICG-stained ILM specimens obtained during macular hole surgery revealed the striking presence of retinal debris and large cellular fragments adherent to the retinal surface of the ILM.10 As such extensive cellular components had not been seen before the use of ICG in macular surgery, it was hypothesized that the presence of cellular material was the result of an ICG-induced alteration of the cleavage plane from the retinal surface of the ILM to the innermost retinal layers.26 Although large retinal cellular fragments were noted after ICG staining, cellular retinal fragments in a lesser degree were also seen when ICG had not been used for staining and fragments appeared only in areas of folded ILM in specimens obtained during surgery for epimacular membranes.34 It was suggested that this observation may be related to the increased rigidity of a multilayered structure such as an epimacular membrane compared with the bare ILM, and the presence of retinal debris was interpreted as the result of a mechanical trauma during peeling.34 The presence of sub-ILM changes in terms of cellular fragments was confirmed by a recent study by Hiscott and colleagues35 suggesting that the plane of separation during ILM peeling for epiretinal membrane surgery may be altered. They concluded that this alteration was not related to the use of the dye, but should be interpreted as a result of the ERM formation or of the ERM-inducing pathology such as the modulation of GFAP within Müller cells or the continuity between components of epimacular membranes and the retina through pores of the ILM, which may increase the adhesion forces between these cells and the ILM.35, 36 In addition, one may hypothesize that an epiretinal membrane, which is peeled off along with the underlying ILM, is stiffer compared with the ILM alone as removed during macular hole surgery. However, when taking together the described cellular interactions and our observations of increased rigidity in the stained ILM, regardless which dye is used, it is quite comprehensible that there is an increased likelihood to alter the cleavage plane during vitreoretinal surgeries both for epimacular membranes and macular holes to the inner retinal layers, because the rigidity is altered either by the presence of epimacular tissue or by the use of certain vital dyes.

The present investigation has some limitations. One needs to consider that AFM provides measurements of the surface of the tissue with a limited penetration. It is known that both ICG and BB do not penetrate the ILM, the staining effect is limited to the tissue surface. Therefore, a stiffening effect as measured by AFM might be more pronounced at the surface of the stained tissue. In addition, AFM measurements as performed herein do not represent direct measurements of the changes of the tensile strength of the ILM. Nevertheless, as the ILM represents an irregularly structured meshwork of various fibers, one may transfer our findings and conclude that an increase in the stiffness measured at the surface correlates to an increase in the tensile strength, which would not hold for regularly structured tissues.

Against the background of the present study, it will be interesting to evaluate potential influences of different dye concentrations and exposure times. In addition, potential effects of age or sex of the patients as well as of other underlying diseases may be evaluated in a larger sample size.