Abstract
Purpose
To assess corneal biomechanical changes in eyes undergoing phototherapeutic keratectomy (PTK) and to investigate the relationship of PTK with corneal thickness.
Methods
We examined 36 eyes of 23 patients undergoing PTK for the treatment of granular corneal dystrophy (GCD). Corneal hysteresis (CH) and corneal resistance factor (CRF) were measured with an Ocular Response Analyzer™ before and 3 months after surgery. We also investigated the relationships between corneal biomechanical parameters and central corneal thickness before and after surgery.
Results
The CH was significantly decreased from 10.2±2.2 to 8.7±1.8 mm Hg after PTK (Wilcoxon signed-rank test, P<0.001), as was the CRF, from 10.3±2.0 to 8.5±1.8 mm Hg (P<0.001). We found a significant correlation between both CH and CRF and the central corneal thickness before surgery (Pearson correlation coefficient r=0.37, P=0.03 for CH, r=0.51, P=0.002 for CRF) and 3 months after surgery (r=0.54, P=0.001 for CH, r=0.74, P<0.001 for CRF).
Conclusions
PTK induces a significant decrease in the biomechanical parameters of the cornea. The corneal thickness may play some role in corneal biomechanics even in eyes with GCD.
Similar content being viewed by others
Introduction
The purpose of phototherapeutic keratectomy (PTK) is to remove diseased corneal tissue and increase corneal transparency by surgical laser photoablation. PTK as well as other excimer laser refractive surgery (eg, photorefractive keratectomy (PRK) and laser in situ keratomileusis (LASIK)) can affect the structural integrity of the cornea by the stromal tissue removal and loss of Bowman's membrane, and there is therefore, theoretically, a risk of corneal biomechanical weakening after both surgical procedures. Actually, it has been shown that the corneal biomechanical parameters were significantly decreased after LASIK.1, 2, 3 Although PTK usually requires less laser ablation of the corneal stroma than is needed by other excimer laser surgery, the biomechanics of the cornea before PTK may have already been more affected by histopathological deformation than that of normal eyes before keratorefractive surgery. In addition, the presence of iatrogenic ectasia after PTK has also been documented.4 Thus, it is of clinical importance to assess the corneal biomechanical changes after excimer laser surgery, but these have not so far been studied in the case of PTK. The purpose of this study is twofold: to quantitatively assess the biomechanical changes of the cornea in eyes with granular corneal dystrophy (GCD) that undergo PTK, and to investigate the relationship of the biomechanical parameters with the central corneal thickness in such eyes.
Materials and methods
We prospectively studied 36 eyes of 23 consecutive patients (male, 5; female, 18; 69.6±6.8 years of age, mean±s.d.) (range, 52–81 years) undergoing PTK for the treatment of GCD. The central corneal thickness before PTK was 545.1±41.0 μm (479–616 μm). The intraocular pressure was 15.8±3.8 mm Hg (range, 9–23 mm Hg). Preoperative patient demographics are summarised in Table 1.
Phototherapeutic keratectomy was performed using the VISX Star S2 excimer laser system (VISX Inc., Santa Clara, CA, USA) with the following parameters: wavelength, 193 nm; fluency, 160 mJ/cm2; repetition rate, 10 Hz; ablation zone diameter, 5.0 mm; transition zone, 0.5 mm; and ablation depth, 45–50 μm based on the clinical appearance of the depth of pathology from the preoperative slit-lamp evaluation. We used the transepithelial technique for corneal epithelium removal. Postoperatively, steroidal (0.1% fluorometholone, Flumetholone™ Santen, Osaka, Japan) and antibiotic (levofloxacin, Cravit™ Santen, Osaka, Japan) medications were topically administered four times daily for 1 week, together with a soft contact lens, with the dose being steadily reduced thereafter.
We also measured the biomechanical parameters of the cornea, namely, corneal hysteresis (CH) and corneal resistance factor (CRF), using the Ocular Response Analyzer™ (ORA: Reichert Ophthalmic Instruments, Depew, New York, USA) before and 3 months after surgery.1 This device uses a rapid air impulse to deform the cornea, and the shape changes are monitored by an electro-optical system.1 The air puff induces two applanations, inward and outward, of the cornea. The air deforms the cornea through an initial applanation event (peak 1), then beyond it into a concavity, and gradually subsides, allowing the cornea to rebound through a second applanation (peak 2). CH is calculated as difference between pressures (mm Hg) where the infra-red peaks 1 and 2 occur. The CRF is calculated as a linear function of P1 and P2, based on the results of a large-scale clinical data analysis, according to the manufacturer.5 We carried out this measurement three times to ensure consistent signal quality and to look for consistent signal morphology and measurement values, and the average value was used for statistical analysis according to the manufacturer's instructions. The central corneal thickness was also measured using an ultrasound pachymeter (DGH-500, DGH Technologies, Exton, USA) before and 3 months after surgery. Additionally, to assess the repeatability of the measurements of this device, we made three consecutive measurements of CH and CRF with this device in 18 eyes with granular dystrophy before surgery at the same time of day on 2 days. We evaluated the repeatability of the two measurements by Bland–Altman plots as described previously.6 Informed consent was obtained from all patients. The study adhered to the tenets of the Declaration of Helsinki. Institutional review board approval was not required for this study, because the ORA has been widely used as a non-contact tonometer, which is considered to be a routine part of a postoperative examination in a clinical setting.
All statistical analyses were performed using StatView version 5.0 (SAS, Cary, USA). The results are expressed as mean±SD, and a P-value of less than 0.05 was considered statistically significant.
Results
After PTK, the CH was decreased significantly from 10.2±2.2 mm Hg (mean±SD) (range, 6.2–15.6 mm Hg) to 8.7±1.8 mm Hg (range, 5.1–13.8 mm Hg), (Wilcoxon signed-rank test, P<0.001) (Figure 1); and the CRF also showed a significant decrease from 10.3±2.0 mm Hg (range, 6.3–15.8 mm Hg) to 8.5±1.8 mm Hg (range, 5.0–12.0 mm Hg) (P<0.001) (Figure 1). The central corneal thickness was reduced significantly from 545.1±41.0 μm (range, 479–616 μm) to 485.9±46.0 μm (range, 423–591 μm) (P<0.001). There were weak, but significant, correlations between CH and the central corneal thickness before (Pearson correlation coefficient r=0.37, P=0.03) and 3 months after surgery (r=0.51, P=0.002) (Figure 2), and modest significant correlations between CRF and the central corneal thickness before (r=0.54, P=0.001) and 3 months after surgery (r=0.74, P<0.001) (Figure 3). There was no significant association between patient age and CH (r=−0.11, P=0.54), or between age and CRF (r=−0.21, P=0.23) before surgery. Iatrogenic keratectasia did not occur in any case during the observation period. Bland–Altman plots indicate that the mean difference between the two measurements with ORA (±95% limits of agreement; LoA) was 0.0±0.7 mm Hg (−1.5 to 1.4 mm Hg) for CH, and 0.1±0.7 mm Hg (−1.2 to 1.4 mm Hg) for CRF (Figure 4).
Discussion
In this study, we showed that PTK significantly decreased both CH and CRF, indicating that PTK, which requires surgical removal of the diseased cornea, may induce a significant reduction of the biomechanical characteristics of the cornea just as LASIK can.1, 2, 3 To our knowledge, this is the first report to evaluate the biomechanical changes of the cornea after PTK. This assessment is clinically important for the dissemination of this surgical technique, especially in consideration of the occurrence of iatrogenic keratectasia in post-PTK eyes.4 Moreover, both CH and CRF are significantly correlated with the central corneal thickness before and after PTK, suggesting that the preoperative and postoperative corneal thickness may play some role in these biomechanical characteristics of the cornea even in eyes with GCD. Our results also indicate that CRF may reflect the overall rigidity of the cornea, depending on CCT more correctly than CH does even in eyes with GCD, as evidenced by the higher correlation coefficient of CCT before and after PTK. With regard to CCT, Lu et al7 showed that there was no significant correlation between CH and CCT induced by wearing soft contact lenses during eye closure. Kirwan et al8 also stated that the lowest postoperative CH was recorded in an eye with a postoperative CCT of 517 μm, suggesting that factors in addition to CCT may play a role in corneal biomechanics. On the other hand, Lam et al9 reported that CH was positively associated with CCT in normal eyes. Broman et al10 showed that CH was significantly correlated with CCT with a modest correlation coefficient in patients presenting at a glaucoma clinic. Shah et al11 also showed that CH increased with increasing CCT in normal eyes, but the correlation was moderate. We have already shown that CCT was the most relevant factor affecting CH as well as CRF, suggesting that CCT may play an important role in the biomechanics of the cornea.12
The diseased cornea of granular dystrophy may affect the biomechanical strength of the cornea, possibly as a result of histopathological deterioration in the collagen lamellae13 or in the Bowman's layer.14 Szentmáry et al13 reported that the collagen lamellae in the anterior stroma of eyes with granular dystrophy showed a rather irregular and loose architecture and were interspersed with small vascular inclusions between the collagen fibers. They also showed that Bowman's layer was absent in the laser-treated area, and that the architecture of the subepithelial stromal collagen lamellae was irregular.13 Seitz et al14 stated that the breaks of Bowman's layer were found in 68% of specimens with GCD. Miyata et al4 first described a case of progressive corneal ectasia following PTK for the treatment of band keratopathy. They stated that it occurred even though the residual corneal thickness was greater than 500 μm, which, as the fully acceptable thickness, was usually deemed to avoid iatrogenic ectasia in normal eyes. Considering that the amount of laser ablation (45–50 μm) in PTK was smaller than that in other excimer laser surgery, such as PRK or LASIK, there is some possibility that the biomechanical strength of the cornea may have already been weakened by histopathological deformation. However, this assumption may be contradicted by the results of our present and previous studies that showed that the preoperative biomechanical properties of eyes with GCD (10.2±2.2 mm Hg for CH, 10.3±2.0 mm Hg for CRF, 545.1±41.0 μm) are very similar to those of normal eyes having similar corneal thickness (10.2±1.3 mm Hg for CH, 10.2±1.7 mm Hg for CRF, 540.4±31.0 μm).12 Our findings may support the view that the biomechanical parameters of GCD are not influenced by its histopathological deformation. We need further investigations to confirm whether or not its deformation may influence the biomechanical characteristics.
In this study, we found no significant association between patient age and biomechanical parameters in eyes with GCD, possibly because of the small sample size (n=36) and the narrow range of patient ages (from 52 to 81 years). We have already demonstrated that the eyes of older volunteers have shown a tendency to have a lower CH value.12 Moreover, Kotecha et al15 reported in a study of patients attending the Glaucoma Research Unit that CH was dependent on age. Ortiz et al2 also reported that the CH value was lower in older eyes, and the difference between the youngest age group (9–14 years) and the oldest age group (60–80 years) was statistically significant. Elsheikh et al16 showed using experimental means that the CH area was significantly larger with decreased age. By contrast, Kirwan et al17 reported that CH in children was similar to that reported in adults. We cannot deny the possibility that patient age may play some role in the biomechanical properties of the cornea even in eyes with GCD. Further investigations are required to clarify this point.
It has been shown that the decrease in CH was not statistically different after LASIK and laser-assisted subepithelial keratectomy (LASEK), indicating that LASIK involving a thin 120-μm flap did not induce additional biomechanical change.8 On the other hand, in our preliminary data, LASIK decreased CH and CRF significantly more than PRK (data not shown). This may be due to the fact that LASIK requires not only surgical tissue removal but also flap creation, whereas PRK requires tissue removal only. In addition, LASIK ablates more of the deeper layers of the corneal stroma than PRK, which may contribute to the greater damage to the corneal nerves evident after LASIK because these are transected at a deeper plane of the cornea. In this study, PTK appeared to result in less decrease in CH and CRF than LASIK as in previous studies, possibly because the amount of laser ablation in PTK is smaller than that in LASIK. We should be aware that it is still difficult to directly compare the effects of the different surgical procedures on corneal biomechanics, because these biomechanics were influenced by many factors such as the amount of laser ablation, age, or race.
In light of the increasing number of reports of iatrogenic keratectasia after excimer laser surgery, such as PRK, LASIK, or PTK, it is clinically important to assess the postoperative biomechanical changes of the cornea. The ORA (Reichert Ophthalmic Instruments) has been developed to assess the biomechanical properties of the cornea.1 It has been shown that there were no significant differences of CH or CRF in post-LASIK patients with or without keratectasia.18, 19 The values of CH and CRF may be useful when used in conjunction with other parameters such as aberrometry or the morphology of the signals. As indicated in Figure 4, Bland–Altman plots show that the mean difference between two consecutive measurements with ORA (±95% LoA) was 0.0±0.7 mm Hg (−1.5–1.4 mm Hg) for CH, and 0.1±0.7 mm Hg (−1.2–1.4 mm Hg) for CRF. The ±95% LoA with granular dystrophy was slightly wider than those with normal eyes in our present and previous studies.7, 12 Although the repeatability of the measurement with GCD appeared to be slightly inferior to those with normal eyes, we consider that this device offers reasonable repeatability even in the evaluation of the biomechanical properties with GCD.
The study assessed the biomechanical parameters of the cornea before and 3 months after surgery. It has been shown that it took 3–6 months to develop a clinically apparent ectasia after PTK.4 In light of a considerable number of case reports of late-onset keratectasia,20, 21, 22 it is confidently assumed that biomechanical deformation slowly develops over time, and then progresses rapidly after a certain threshold is passed. In addition, the recurrence of the diseased cornea has been reported in the late postoperative period after PTK.23, 24, 25, 26, 27, 28, 29, 30 More prolonged observation is necessary to ascertain whether iatrogenic keratectasia or the recurrence of the disease after PTK occurs in the late postoperative period.
In summary, we showed that the biomechanical parameters of the cornea were significantly decreased after PTK in eyes with GCD, possibly because of the surgical removal of the diseased corneal tissue. Moreover, these biomechanical parameters were significantly associated with the central corneal thickness, not only before surgery but also after it, suggesting that the corneal thickness may play some role in the corneal biomechanics, even in eyes with a diseased cornea. A further study with greater numbers of patients is required to confirm these preliminarily findings.
References
Luce DA . Determining in vivo biomechanical properties of the cornea with an ocularresponse analyzer. J Cataract Refract Surg 2005; 31: 156–162.
Ortiz D, Piñero D, Shabayek MH, Arnalich-Montiel F, Alió JL . Corneal biomechanical properties in normal, post-laser in situ keratomileusis,and keratoconic eyes. J Cataract Refract Surg 2007; 33: 1371–1375.
Pepose JS, Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ . Changes in corneal biomechanics and intraocular pressure following LASIK usingstatic, dynamic, and noncontact tonometry. Am J Ophthalmol 2007; 143: 39–47.
Miyata K, Takahashi T, Tomidokoro A, Ono K, Oshika T . Iatrogenic keratectasia after phototherapeutic keratectomy. Br J Ophthalmol 2001; 85: 247–248.
Luce DA . Determining in vivo biomechanical properties of the cornea with an ocularresponse analyzer. J Cataract Refract Surg 2005; 31: 156–162.
Bland JM, Altman DG . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307–310.
Lu F, Xu S, Qu J, Shen M, Wang X, Fang H, Wang J . Central corneal thickness and corneal hysteresis during corneal swelling induced by contact lens wear with eye closure. Am J Ophthalmol 2007; 143: 616–622.
Kirwan C, O'Keefe M . Corneal hysteresis using the Reichert ocular response analyser: findings pre- and post-LASIK and LASEK. Acta Ophthalmol 2008; 86: 215–218.
Lam A, Chen D, Chiu R, Chui WS . Comparison of IOP measurements between ORA and GAT in normal Chinese. Optom Vis Sci 2007; 84: 909–914.
Broman AT, Congdon NG, Bandeen-Roche K, Quigley HA . Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure. J Glaucoma 2007; 16: 581–588.
Shah S, Laiquzzaman M, Cunliffe I, Mantry S . The use of the Reichert ocular response analyser to establish the relationship between ocular hysteresis, corneal resistance factor and central corneal thickness in normal eyes. Cont Lens Anterior Eye 2006; 29: 257–262.
Kamiya K, Hagishima M, Fujimura F, Shimizu K . Factors affecting corneal hysteresis in normal eyes. Graefes Arch Clin Exp Ophthalmol 2008; 246: 1491–1494.
Szentmáry N, Seitz B, Langenbucher A, Schlötzer-Schrehardt U, Hofmann-Rummelt C, Naumann GO . Histologic and ultrastructural changes in corneas with granular and macular dystrophy after excimer laser phototherapeutic keratectomy. Cornea 2006; 25: 257–263.
Seitz B, Behrens A, Fischer M, Langenbucher A, Naumann GO . Morphometric analysis of deposits in granular and lattice corneal dystrophy: histopathologic implications for phototherapeutic keratectomy. Cornea 2004; 23: 380–385.
Kotecha A, Elsheikh A, Roberts CR, Zhu H, Garway-Heath DF . Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 2006; 47: 5337–5347.
Elsheikh A, Wang D, Rama P, Campanelli M, Garway-Heath D . Experimental assessment of human corneal hysteresis. Curr Eye Res 2008; 33: 205–213.
Kirwan C, O'Keefe M, Lanigan B . Corneal hysteresis and intraocular pressure measurement in children using the reichert ocular response analyzer. Am J Ophthalmol 2006; 142: 990–992.
Kirwan C, O'Malley D, O'Keefe M . Corneal hysteresis and corneal resistance factor in keratoectasia: findings using the Reichert ocular response analyzer. Ophthalmologica 2008; 222: 334–337.
Kerautret J, Colin J, Touboul D, Roberts C . Biomechanical characteristics of the ectatic cornea. J Cataract Refract Surg 2008; 34: 510–513.
Geggel HS, Talley AR . Delayed onset keratectasia following laser in situ keratomileusis. J Cataract Refract Surg 1999; 25: 582–586.
Leung AT, Rao SK, Lam DS . Delayed onset keratectasia after LASIK. J Cataract Refract Surg 1999; 25: 1036–1037.
Lifshitz T, Levy J, Klemperer I, Levinger S . Late bilateral keratectasia after LASIK in a low myopic patient. J Refract Surg 2005; 21: 494–496.
Hahn TW, Sah WJ, Kim JH . Phototherapeutic keratectomy in nine eyes with superficial corneal diseases. Refract Corneal Surg 1993; 9: S115–S118.
Maclean H, Robinson LP, Wechsler AW, Goh A . Excimer phototherapeutic keratectomy for recurrent granular dystrophy. Aust N Z J Ophthalmol 1996; 24: 127–130.
Rapuano CJ . Excimer laser phototherapeutic keratectomy: long-term results and practical considerations. Cornea 1997; 16: 151–157.
Dinh R, Rapuano CJ, Cohen EJ, Laibson PR . Recurrence of corneal dystrophy after excimer laser phototherapeutic keratectomy. Ophthalmology 1999; 106: 1490–1497.
Stewart OG, Pararajasegaram P, Cazabon J, Morrell AJ . Visual and symptomatic outcome of excimer phototherapeutic keratectomy (PTK) for corneal dystrophies. Eye 2002; 16: 126–131.
Fagerholm P . Phototherapeutic keratectomy: 12 years of experience. Acta Ophthalmol Scand 2003; 81: 19–32.
Chen W, Qu J, Wang Q, Lu F, Barabino S . Automated lamellar keratoplasty for recurrent granular corneal dystrophy after phototherapeutic keratectomy. J Refract Surg 2005; 21: 288–293.
Das S, Langenbucher A, Seitz B . Excimer laser phototherapeutic keratectomy for granular and lattice corneal dystrophy: a comparative study. J Refract Surg 2005; 21: 727–731.
Author information
Authors and Affiliations
Corresponding author
Additional information
The authors do not have any proprietary interest in this work.
Rights and permissions
About this article
Cite this article
Kamiya, K., Shimizu, K. & Ohmoto, F. The changes in corneal biomechanical parameters after phototherapeutic keratectomy in eyes with granular corneal dystrophy. Eye 23, 1790–1795 (2009). https://doi.org/10.1038/eye.2008.373
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/eye.2008.373
Keywords
This article is cited by
-
Assessment of the corneal biomechanical features of granular corneal dystrophy type 2 using dynamic ultra-high-speed Scheimpflug imaging
Graefe's Archive for Clinical and Experimental Ophthalmology (2023)
-
Corneal biomechanical properties in eyes with no previous surgery, with previous penetrating keratoplasty and with deep anterior lamellar keratoplasty
Japanese Journal of Ophthalmology (2013)
-
Corneal hysteresis in eyes undergoing phototherapeutic keratectomy
Eye (2010)
-
Reply to Corneal hysteresis in eyes undergoing phototherapeutic keratectomy
Eye (2010)