Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

Measurement of axial length in the calculation of intraocular lens power

Abstract

Purpose Accurate measurement of ocular axial length is essential for accurate intraocular lens (IOL) power calculation. Although it is common practice to average several axial length measurements to improve accuracy, it has been suggested that a single high-quality A-scan ultrasonographic measurement is adequately accurate owing to the high test-retest reliability of A-scan biometry. The aim of this study was to compare the accuracy of a single high-quality A-scan measurement with that of the average of three acceptable measurements in the calculation of IOL power.

Method We studied 103 eyes of 103 patients who underwent cataract-IOL surgery. All these patients underwent pre-operative ocular biometry, a standardised extracapsular cataract extraction with posterior chamber IOL implantation, and clinical refraction between 10 and 14 weeks post-operatively.

Results There was no statistically significant difference between the two study groups in measured axial length (p < 0.01), calculated emmetropic IOL power (p < 0.05) or the prediction of post-operative refraction (p<0.99).

Conclusion The use of a single high-quality axial length measurement was as accurate as the mean of three acceptable axial length measurements in the calculation of IOL power.

References

  1. 1

    Ossoinig KC . How to obtain maximum measuring accuracy with a standardised A scan. Doc Ophthalmol 1984;38:197–216.

    Google Scholar 

  2. 2

    Strobel J . Die Berchnung der Brachkrafte von intraokularen Linsen. Fortschr Ophthalmol 1985;82:165–7.

    CAS  PubMed  Google Scholar 

  3. 3

    Butcher JM, O'Brien C . The reproducibility of biometry and keratometry measurements. Eye 1991;5:708–11.

    Article  Google Scholar 

  4. 4

    Sander R, Kraff MC . A comparison of the Digital Biometry Ruler-300 and Echo oculometer 3000. Am Intraocul Implant Soc J 1982;8:365–9.

    Article  Google Scholar 

  5. 5

    Giers U, Epple C . Comparison of A-scan device accuracy. J Cataract Refract Surg 1990;16:235–42.

    CAS  Article  Google Scholar 

  6. 6

    Sander DR, Retzlaff J, Kraff MC . Comparison of the SRK II formula and other second generation formulae. J Cataract Refract Surg 1988;14:136–41.

    Article  Google Scholar 

  7. 7

    Richards SC, Olsen RJ, Richards WL . Factors associated with poor predictability by IOL calculation formulae. Arch Ophthalmol 1985;103:515–8.

    CAS  Article  Google Scholar 

  8. 8

    Holladay JT, Praeger TC, Ruiz RJ, Lewis JW, Rosenthal H . Improving the predictability of IOL power calculations. Arch Ophthalmol 1986;104:539–41.

    CAS  Article  Google Scholar 

  9. 9

    Holladay JT, Praeger TC, Chandler TY, Musgrove KH . A three-part system for refining IOL power calculations. J Cataract Refract Surg 1988;14:17–24.

    CAS  Article  Google Scholar 

Download references

Author information

Affiliations

Authors

Corresponding author

Correspondence to P Sunder Raj.

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Raj, P., Ilango, B. & Watson, A. Measurement of axial length in the calculation of intraocular lens power. Eye 12, 227–229 (1998). https://doi.org/10.1038/eye.1998.54

Download citation

Keywords

  • Axial length
  • Biometry
  • Cataract surgery
  • Intraocular lens
  • Power calculation

Further reading

Search

Quick links