Introduction

Biosensors are analytical devices that convert a biological response into an electrical signal. They are normally composed of two main components: the sensing device and the sensing molecule, that is, the chemical or biological recognition elements. As a functional hybrid system, the main challenge is to design a system that couples the unique features of the biorecognition event with sensitive signal detection and amplification. Many researchers have reported the potential utilization of nanostructure-based biosensors in the detection of biological species featuring direct electrical transduction.

Nanowire-based biosensors, in particular, are promising candidates for the transduction of analytical signals in solution into electrical signals, and ultrasensitive sensing has been demonstrated for ions,1 molecules,2 proteins,3, 4, 5, 6 DNA7 and viruses.8 Nanowires are fabricated either via bottom-up or top-down processes. However, bottom-up fabrication has thus far prevented the commercialization of such nanowire sensors mainly due its incompatibility with high-volume manufacturing. In contrast, various top-down methods entail a high level of non-standard processing complexity and high process variation.5, 9, 10, 11 In addition, both methodologies produce three-dimensional nanowires that inherently suffer from a high density of surface states and non-uniform doping distributions that entail degradation in device performance.12, 13, 14 The development of nanofabrication tools and techniques, such as electron beam lithography, focused ion beam and nanoimprint lithography, has propelled the momentum of functional nanostructure-based biosensor fabrication. Altghough this rich inventory of nanofabrication methods has been available for many years, the methods still lack the capability for on-chip integration of nanostructure-based biosensors with an allied signal-processing circuitry and a fluid-handling module using standard microfabrication technology. A novel, double-gate nanowire field effect transistor for biosensing was recently reported.15 The device was patterned with 193 nm lithography, and specific sensing of 1.34 nmol l−1 of anti-avian influenza was demonstrated.

In this work, we present an electrostatically formed nanowire (EFN) biosensor capable of specific, label-free and real-time detection of femtomolar protein concentrations. The EFN biosensor is composed of a nanowire-like conducting channel that is not physically fabricated, but rather, electrostatically formed post fabrication. Consequently, fabrication requires standard, integrated-circuit processes, thus eliminating the need of sophisticated nanoscale fabrication. Competitive performance coupled with standard silicon processing makes the EFN biosensor a viable technology for various real-time diagnostic applications.

The EFN biosensor resembles the silicon-on-insulator (SOI) G4-FET developed in 200216, 17, 18 and is presented schematically in Figure 1a. The biosensor is composed of an accumulation-type transistor with four, independent gates: a backgate VGb, a front-gate/reference electrode (VREF) and two additional lateral junction gates (VGj1 and VGj2) located on both sides of the conducting channel. The nanoscale channel is induced and controlled by depletion regions formed around it by the four surrounding gates. In this way, the size, shape and location of the channel in the bulk SOI are determined. Figures 1b and c illustrate an EFN biosensor with voltages VGj1=VGj2 that provide wide and narrow channels (nanowire dimension), respectively; the channel vertical position is controlled by VGb and VREF as demonstrated in Figures 1d and e. Figure 1f shows a scanning electron microscope top image of the EFN.

Figure 1
figure 1

Schematic illustration of the EFN biosensor with antibody/analyte complexes bonded to the gate dielectric. (a) Note the n++ and the n+ implants for the source/drain and channel regions, respectively. The junction areas are p+ doped to allow efficient reverse biasing of the p-n junctions that electrostatically induce and control the width of the channel, using the electrical wiring shown in the inset. (be) Demonstrate the electrostatic control of the size, shape and location of the conducting channel. (b) Wide channel; (c) Reverse-bias forming a narrow channel; (d) Biasing of the top reference electrode and the backgate move the channel close to the biological event or (e) away from the noise centers at the Si/SiO2 interface. (f) Scanning electron microscope top image of the EFN device. The red rectangle in the middle of the device is the opening of the passivation layer where the modified gate dielectric is exposed to the analytes.

The EFN is conceptually different from both the conventional silicon nanowire (SiNW) and the double-gate nanowire. A SiNW is a three-dimensional structure in which the confinement is chemical (compositional), that is, it is due to a material (for example, air or SiO2) with different properties surrounding the wire. The confinement potential of the double-gate nanowire is chemical on the SiO2 side and electrostatic on the silicon side (this is an ordinary confinement potential of an inverted metal-oxide-silicon capacitor). On the contrary, in the EFN device, the confinement is fully electrostatic; that is, it is due to the external applied bias rather than chemistry. This fact implies that the interplay between sensing, charge carrier distribution and the confinement potential is inherently different in comparison to a SiNW or a double-gate nanowire (see the Supplementary Information for device simulations of the EFN parabolic potential vs the step-like potential in the SiNW and the double-gate nanowire). Additionally, degradation in the performance of SiNW-based devices due to metal catalysts and non-uniform dopant distribution19, 20, 21, 22 does not exist in EFN biosensors because EFNs are electrostatically shaped inside single-crystalline silicon.

The EFN configuration allows the channel to be sufficiently removed from the Si/SiO2 interface,23 which is the dominant noise source in SiNWs24, 25, 26, 27, 28, 29 as well as in inversion-type planar devices30 (and in the double-gate nanowire, where the conduction is due to the formation of an inversion layer at the silicon/SiO2 interface). It was recently demonstrated that the low-frequency noise originating from charge trapping/detrapping at the Si/SiO2 interface is the prevailing noise source in wet environments, as the noise-level dependency on ion concentration and pH level is negligible.31, 32, 33 This EFN degree of freedom improves the signal-to-noise ratio, which is crucial for biosensing applications because the signal-to-noise ratio level will determine the sensitivity limit of the biosensor.34 Moreover, the ability to control the vertical position of the channel compromises between the need for maximum proximity of the biological events to the conducting channel on the one hand, and the need for high signal-to-noise ratio on the other. Figures 1d and e illustrate the control of the vertical location of the EFN nanowire using solely the appropriate VGb and VREF. In Figure 1d, the channel is closer to the biological reaction and the noise centers (trap states that exchange electrons with the conducting channel) at the Si/SiO2 interface, whereas in Figure 1e, the channel is further removed from the reacting surface and from the Si/SiO2 interface.

As the confinement potential in the EFN is controlled electrostatically, the channel position can be changed in real time, and in this manner, a further increase in device sensitivity can be achieved (as examined in the Discussion section below). To summarize, the EFN allows post-fabrication determination of the channel size and shape, as well as the channel lateral and z-locations. In contrast, these parameters are predetermined and fixed in both SiNW and double-gate nanowires. We note that one of the advantages in biological detection using NWs is attributed to their large surface-to-volume ratio.3, 4, 5, 6, 7, 8 We demonstrate below that the sensitive sensing with the EFN biosensor is governed by the downscaling of the sensor active area to dimensions similar to that of the analytes. It should be noted that the EFN-enhanced sensitivity is not due to additional biasing as was argued for the dual-gated SiNW.35 An enhanced sensitivity was achieved with the dual-gated SiNW, once the shift in backgate threshold voltage was recorded (due to the ratio between the top dielectric and the buried dielectric). In the present case, only the shift in the front threshold voltage is measured, and hence the enhanced sensitivity is not due to the presence of additional gates.

Experimental procedures

Silicon fabrication

Silicon-on-insulator wafers measuring 6″ in diameter were used (SOITEC, Bernin, France). The thickness of the SOI layer was 260 nm, with boron doping of 1.6 × 1014 cm−3 (resistivity of 13–22 Ω cm). The thickness of the buried oxide was 1 μm. The fabrication of the devices was as follows: the devices were isolated from one another using isolation trenches. Successive implants for the channel region, the source/drain regions and junction regions were performed. The SiO2 gate dielectric of 50 Å in thickness was grown via low-pressure chemical vapor deposition (LPCVD). The channel doping (As) was in the range of 1.6 × 1017 cm−3. The EFN biosensor is partially depleted for the given channel doping and SOI thickness. The doping (As) at the source and drain regions was in the range of 5 × 1019 cm−3, and the junction doping (B) was in the range of 5 × 1019 cm−3. The source/drain and channel implants ensure an n-type accumulation device. The implant steps were followed by 100 nm SiO2 plasma-enhanced chemical vapor deposition (PECVD) for the interlayer dielectric and opening of the contacts. Ti/Al/TiN was sputtered and patterned for interconnections followed by a 4500 Å passivation layer of plasma-enhanced chemical vapor deposition nitride. The last step of the process was the opening of the passivation above the EFN biosensor-active region. This step was performed with a dry etch followed by a final wet etch. Owing to the EFN geometry, two channel lengths need to be defined. The full channel length (L) is defined as the lithographic length between the source and drain, and the effective length (Leff) is defined as the length of the junction gate (Figure 1a). All devices had a full channel length of 10 μm and an effective length of 7 μm. The channel width (W) for all devices was 400 nm. See the Supplementary Information for schematic illustrations of the silicon processing.

Chemical modifications

Prior to modification, the devices were rinsed with EtOH (100%), dried using N2 and activated with UV/O3 (UVOCS T10X10, Montgomeryville, PA, USA) for 30 min. Immediately after activation, the devices were dipped for 10 min in 1% APTMS (3-aminopropyltrimethoxysilane, Sigma, Rehovot, Israel), in EtOH/DI (95:5) water solution, rinsed and sonicated for 10 min in 100% EtOH, dried in N2 and annealed for 10 min at 110 °C. The binding of either antibodies (anti-troponin and anti-PSA) to the APTMS was performed using a single-step preparation by dipping the APTMS-activated device for 2 h in 0.1 × MES buffer pH 4.7 supplemented with 5 mM 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide (Pierce, Rockford, IL, USA) and 1 mg ml−1 polyclonal rabbit anti-troponin or 1 mg ml−1 polyclonal rabbit anti-PSA for the anti-troponin and anti-PSA devices, respectively (Abcam, Cambridge, MA, USA). The dies were washed twice with 0.1 × MES buffer pH 4.7, then once in phosphate buffer pH 6. The devices were measured immediately or stored under N2 until use.

Electrical connections

Electrical measurements of current vs applied reference-electrode voltage (IDS-VREF) and current vs time (IDS-t) of the EFN biosensor in a wet environment were performed with a custom-made flow cell. The measurements were performed with a homemade Ag/AgCl reference electrode and 0.01 × diluted commercial phosphate-buffered saline working solution with 10 mM sodium phosphate and 150 mM NaCl at pH 7.4 (Pierce). The electrical connectivity of the devices is illustrated in the inset of Figure 1a. Five voltages were applied to the devices: drain-source voltage (VDS), voltage at the reference electrode (VREF), back gate voltage (VGb) and two voltages at the junctions (VGj1, VGj2). All measurements were performed with VGj=VGj1=VGj2. The pH sensitivity of EFN biosensors was measured with alternating injections of 0.01 × phosphate-buffered saline solutions at pH 7 and 9. All the IDS-t measurements were performed with VDS=50 mV. The low value of VDS was applied to guarantee the condition of VDSVGj. This condition provides a conducting channel with homogenous diameter throughout the length of the channel.

Results

Figure 2a shows IDS-VREF measurements of the EFN biosensor modified with cardiac troponin (anti-cTnI) for various junction gate voltages: VGj=0, −0.5, −1, −1.5 and −2 V, in which the backgate voltage (VGb) was −7 V for all devices. Note that IDS decreases as VGj becomes more negative due to the presence of p-n junctions on both sides of the channel formed by the boron and arsenic dopants. The p-n junctions control the effective width, Weff, of the conducting channel; reverse bias (negative VGj) increases the depletion regions on both sides of the channel and hence reduces its width (IDS decrease), whereas forward bias (positive VGj) reduces the depletion areas and consequently increases the channel width (IDS increase). The sensitivity of IDS to variations in VREF is characterized by the transistor gain (gm=dIDS/dVREF), which is extracted from the IDS-VREF measurements (Figure 2a). The shape of the gm curve shows the expected single-peak signature for a device operation in the linear regime. The gain decreases as VGj becomes more negative (reversed bias), which is a direct consequence of the increase in the depletion regions at the junctions that reduce IDS and gm. Finally, the shift in gm towards higher values of VREF as VGj is more negative, reflecting the shift in the front threshold voltage (VTf) due to a decrease in the front channel potential (ψsf).18 The value of VREF (for a given VGj) that provides the highest gain was selected for the IDS-t measurements for both the pH-sensitivity measurements and the measurements for analyte detection.

Figure 2
figure 2

Electrical characterization of the EFN biosensor. (a) IDS vs VREF and gain (gm) curves of the EFN biosensor measured for different lateral gate voltages (VGj=0, −0.5, −1, −1.5 and −2 V). Note that as the junctions are more reverse biased, the IDS and the maximum value of gm are smaller because the effective width of the channel decreases. The measurements were performed in 0.01 × phosphate-buffered saline buffer solution at pH 7. The channel is doped with 1.6 × 1017 cm−3 of (As), the source/drain regions are doped with 5 × 1019 cm−3 (As), and the junction areas are doped with 5 × 1019 cm−3 (B). The lithographic width and length of the conducting channel are 400 nm and 7 μm, respectively. (b) Solution-related leakage currents in EFN. Leakage current at the reference electrode vs applied reference voltage for different biasing of junction gates. (c) Small hysteresis between forward and backward sweeps. (d) pH sensitivity of EFN biosensor. The modulation of IDS due to alternating injections of pH 7 and pH 9 solutions is presented; the resulting pH sensitivity is 43 mV dec−1.

Solution-related leakage currents of an EFN modified with anti-cTnI are presented in Figure 2b. Note that the solution-junction gate-leakage current (IGj) was measured for VGj=−2 V and 0.5 V for VREF range of −2 to 2 V. It is shown that IGj is not dependent upon VREF. As expected, the solution leakage current is greater when the PN junctions are forward-biased. However, the solution reference electrode leakage current (IREF) exhibits strong dependence upon VREF, where negative VREF implies an increase in IREF. Finally, IREF is greater when the PN junctions are forward-biased, as expected. Drift currents (IDS) were on the order of several mV for the initial 10 h. Figure 2c presents the hysteresis between forward sweep and backward sweep. The small hysteresis between the two slopes is indicative of small, defect-induced charge trapping.

Figure 2d shows the sensitivity to pH of the EFN biosensor modified with anti-cTnI for VGj=0 V (VDS=50 mV, VREF=1.4 V, VGb=−7 V). Alternating injections of pH 7 and 9 solutions were performed, and as expected, higher pH entails smaller IDS as the biologically modified gate dielectric is more negatively charged. The gain of the EFN biosensor was measured with a VREF=−10 mV calibration pulse, and in this manner, the shift in the front threshold voltage (due to the change in pH) was extracted. The pH sensitivity of the device was calculated to be 43 mV dec−1, where the maximum theoretical value for the pH sensitivity of the SiO2 gate dielectric at this range of pH is 47–50 mV dec−1.22 A real-time, specific and label-free detection of 10 pg ml−1 (340 fM) cTnI is demonstrated in Figure 3a. The measurement was performed with narrow Weff (VGj=−2 V), and VREF was selected to provide the highest gm according to the IDS-VREF plot. VDS and VGb were set to 50 mV and −7 V, respectively. The analyte injection occurred at t=0, where a sharp decrease in the current is recorded. This effect is in agreement with the isoelectric point of cTnI (pI=5.2–5.4) that, in conjunction with the pH of the solution (pH=7.4), suggests that the troponin is negatively charged, and hence the current decrease is measured for the n-type channel. Non-specificity (with identical conditions) was also demonstrated for a device modified only with APTMS.

Figure 3
figure 3

Specific, label-free and real-time detection with an EFN biosensor. The measurements were performed in a buffer solution of 0.01 × phosphate-buffered saline at pH 7. (a) Response of IDS to injection of 10 pg ml−1 cTnI at t=0 s. The device was biased as follows: VDS=50 mV, VGb=−7 V, VGj=−2 VREF=1.6 V. A decrease in IDS of 15% is clearly observed following the cTnI binding. The decrease in current is expected for the given pI (5.2–5.4) of cTnI and the pH of the solution. Nonspecificity is also demonstrated because no change in current is apparent following injection of 10 pg ml−1 of cTnI to a device modified solely with APTMS (at t=0 s). (b) Normalized source-drain currents of six EFN biosensors subjected to 10 pg ml−1 cTnI injections at t=0 s and biased with VDS and VGb of 50 mV and −7 V, respectively. Each device was biased with a different VGj: 0 V, −0.5 V, −1 V, −1.5 V and two devices with −2 V. For VGj=0 V, no modulation in current is recorded, whereas for VGj of −0.5 V and −1 V, only a minute current change is noticeable. Substantial current decreases of 10 and 15% are clearly observed for VGj=−1.5 V and for the two devices with VGj=−2 V, respectively. (c) Normalized IDS vs time for the EFN dose response at cTnI concentrations of 10 pg ml−1, 10 ng/ l−1, 10 μg ml−1. The applied voltages were VDS=50 mV, VGb=−7 V, VGj=−1.5 (Weff=55 nm) and VREF=1.6 V. (d) Real-time, specific, and label-free detection of 3 fM PSA was obtained for VGj=−2 V (Weff=25 nm), VDS=50 mV, VREF=1.5 V, VGb=−7 V. Note a shift of 30 mV in threshold voltage. Nonspecificity is also demonstrated.

Figure 3b presents the plots of normalized source-drain current vs time of six devices pre- and post-injection of 10 pg ml−1 cTnI. Each device was subjected to a different VGj to study the effect of the channel width upon threshold shift due to label-free and specific binding of cTnI. Four devices were operated with VGj=0, −0.5, −1 and −1.5 V, and two devices with VGj=−2 V. For each device, VREF was selected to provide the largest gm according to IDS-VREF plots, and VDS and VGb were set to 50 mV and −7 V, respectively. The gm for each plot was measured experimentally, and in conjunction with the shift in current, the shift in the front threshold voltage (ΔVTf) due to the injection of cTnI was extracted. No modulation of current was recorded for VGj=0 V, whereas only a minute decrease in current of 2% was recorded for VGj=−0.5 and −1 V, which provides ΔVTf of 7 mV. However, a substantial decrease in current of 10% was demonstrated for VGj=−1.5 V (ΔVTf of 30 mV), and a more pronounced decrease in current of 15% was demonstrated for the two devices subjected to VGj=−2 V, which entails ΔVTf of 73 mV and ΔVTf of 65 mV.

Figure 3c presents the EFN dose-response for cTnI concentrations of 10 pg ml−1, 10 ng ml−1 and 10 μg ml−1 for VGj=−1.5 V. Another EFN biosensor was modified with APTMS and anti-PSA, and the following voltages were applied: VGj=−2 V (Weff=25 nm), VDS=50 mV, VREF=1.5 V and VGb=−7 V. PSA at a concentration of 100 fg ml−1 (3 fM) was injected, and a real-time decrease in the current was evident. The decrease in current, with a corresponding, calculated threshold-potential shift of ΔVTf=30 mV, is consistent with the pI of PSA (5.8) and the pH of the solution (7.4) and suggests that the PSA is negatively charged (Figure 3d). Nonspecificity was verified by repeating the experiment with an identical device that was modified only with APTMS.

Discussion

A three-dimensional device simulator (Synopsys TCAD Sentaurus, Mountain View, CA, USA) was used to calculate Weff as a function of VGj; Weff was estimated using the FWHM of the electron density of the conducting channel at the z-plane (vertical axis), where this density attains its maximum width (Figures 4a and b). We note that Weff decreases from 190 nm for VGj=0 V to 25 nm for VGj=−2 V and that for Weff >80 nm (VGj=−1 V), the current modulation upon binding of troponin is negligible. However, for Weff <than 80 nm, the effect of the troponin binding is clearly observed. An even smaller EFN can be realized in the following way. Adequate thermal treatment will drive the boron dopants of the junction gates towards the center of the conducting channel and in this way form the PN junctions closer to the channel center. (See the Supplementary Information for simulations of 10 nm EFN).

Figure 4
figure 4

Simulations and modeling of the EFN biosensor. (a) The variation of the front threshold voltage, ΔVTf, and the calculated effective channel width, Weff (right axis), as a function of VGj. The simulated cross-section of the electron density in the channel center at the middle of the channel (x=L/2) is shown (top) for each value of VGj. Note the large increase in ΔVTf with the formation of the nanoscale channel (smaller Weff). A significant change in ΔVTf is observed when Weff <80 nm, and the induced channel is approaching the top surface, as is schematically shown above in Figure 1d. (b) The simulated profile of the normalized electron density in the middle of the channel (x=L/2) for each value of VGj; a channel width of 25 nm at FWHM is obtained for VGj=−2 V. (c) The variation of the average potential in the SOI region, ΔψSOI, and Weff (red line, right axis) calculated using the Synopsys TCAD Sentaurus simulator for different VGj. The simulated device includes only the inorganic components of the EFN biosensor, and the length of the device is set to 3 μm rather than 7 μm. The change ΔψSOI is calculated as the difference between a device with and without a fixed charge (on top of the gate dielectric) of 1019 cm−3 representing the analyte (see main text). The simulations support our observation that larger VGj (and consequently smaller Weff) implies greater ΔψSOI and enhanced detection sensitivity. (d) Calculated charge density (Q*cTnI) as a function of VGj (this is the troponin charge density directly above the conducting nanowire). Note that as the gate voltage VGj is decreased from 0 to −2 V and the corresponding channel width and vertical location change, the charge density of the cTnI layer over the conducting nanowire increases by one order of magnitude. This effect suggests that as Weff decreases, there is a probability of obtaining an analyte distribution that provides a greater effective charge density (the charge density just above the conducting nanowire), and hence, an enhanced device sensitivity.

The inverse correlation between Weff and the difference between the average potential of the depleted/accumulated SOI region pre- and post-analyte binding, ΔψSOI, which is correlative to ΔVTf, was simulated and presented in Figure 4c. The EFN device was simulated with parameters identical to those of the measured devices (apart from the length of the device, which was set to 3 μm to reduce computation time). A 1000 nm3 SiO2 cube having a fixed charge of 1019 cm−3 was used to simulate the presence of a single-charged protein analyte, and VGj was changed from 0 to −1.8 V in steps of 0.2 V. Two simulations were performed for each VGj, one with the localized charge in the channel center and the other one without any charge, and the average potential in the SOI region (ψSOI) was extracted for each simulation. The difference in ψSOI with and without charge was calculated for each value of VGj. When no voltage is applied at the junctions (VGj=0 V and Weff=190 nm), the shift in ψSOI is 2 mV (Weff=190 nm), whereas for VGj=−1.8 V (Weff=37 nm), the shift is ΔψSOI=20 mV. Note that, as expected, a decrease in Weff entails a greater shift in ψSOI (See the Supplementary Information).

Further modeling was used to extract the analyte charge density above the nanowire. Analytical modeling of the threshold voltage of a G4 transistor was recently reported.16 When the metal gate of the G4 transistor is removed and the gate dielectric is exposed to an electrolyte and a reference electrode, the metal work function (φm) is replaced in the following manner36: φm/q=Eref−ψ0mol, where Eref is the reference electrode potential relative to vacuum, ψ0 is the double layer potential at the solution–oxide interface and χmol is the dipole potential of the solution. For an EFN biosensor, we assume that the organic layers (APTMS, anti-cTnI and cTnI) can be modeled as a system of serially connected capacitors. Finally, we assume that the potential drops across the ATPMS layer, the anti-cTnI layer and ψ0 did not change following the binding event. It was previously demonstrated− that using the above assumptions, the shift in front threshold voltage (ΔVTf) is37:

where QcTnI, Q*cTnI, CcTnI, ɛcTnI and tcTnI are the overall charge, the charge density per unit volume (that is, the troponin charge density directly above the conducting nanowire), the capacitance, the permittivity and the thickness of the cTnI layer, respectively (the Debye screening is neglected because the measurements were performed with 0.01 × phosphate-buffered saline, which ensures a Debye length of 7 nm). We assume a sparse cTnI layer (as full coverage of the APTMS layer and antibodies is not anticipated) with ɛcTnI=80 and tcTnI=10 nm. The value of Q*cTnI is calculated for each VGj and presented in Figure 4d. Note that as Weff decreases (or as VGj is more reverse-biased), there is a probability of obtaining an analyte distribution that provides greater charge density above the induced nanowire and consequently an enhanced device sensitivity. Finally, considering a typical dimension of 10 nm of a single cTnI molecule and 30% surface coverage of the cTnI layer, then the charge of a single cTnI molecule is 5 e, which is consistent with the theoretical calculations based on the typical Henderson–Hasselbalch equation (using software such as SEQMOL or PyMOL). Finally, the EFN detection sensitivity can be even further enhanced if a ’lateral voltage scan’ mode is applied in the following manner. The two side-gate voltages VGjs are continuously adjusted in such a way that the nanowire sweeps the channel from one p+ region to the other (See the Supplementary Information and Supplementary Movie 1). In this way, the source-drain current is modified each time the EFN passes in close proximity to an analyte. This mode is similar to sensing using a NW array.

We have presented an electrostatically formed biosensor possessing a promising immunodetection performance in terms of real-time, specific and label-free detection with fM sensitivity. The conducting nanowire of the EFN biosensor is electrostatically defined post fabrication, which permits standard, CMOS-compatible silicon processing with extremely relaxed design rules. It was demonstrated that the electrostatically defined shape and position of the nanowire govern the detection sensitivity, that is, smaller the width of the nanowire, greater is the shift in the EFN threshold voltage. The EFN competitive sensing performance yields an inexpensive, disposable and ultrasensitive platform for various sensing applications.