Objectives
We report the operative details and surgical outcomes of 39 consecutive transconjunctival sutureless vitrectomies using the TSV 25™ system.
Methods
We retrospectively reviewed the medical records of 35 patients (39 eyes) who underwent surgery using TSV 25™ at the Department of Ophthalmology, Waterford Regional Hospital between March 2003 and February 2005. Preoperative, intraoperative and postoperative details were recorded and analysed. We also performed a systematic review of our surgical technique.
Results
Mean (±SD; range) duration of surgery was 32.85 (17.54; 5–85) min. Methods of anaesthesia included retrobulbar anaesthesia (31; 79.5%) and general anaesthesia (8; 20.5%). The mean (±SD; range) length of stay was 1.53 (1.25; 0–4) days. Mean (±SD; range) visual acuities (logMAR) were −1.335 (1.15; 0.0 to −3.0) and −0.803 (1.03; 0.1 to −3.0) preoperatively and postoperatively, respectively. The mean (±SD) intraocular pressures were 14.69 (4.64) mmHg and 11.56 (9.03) mmHg preoperatively and at the first dressing, respectively. Intraoperative complications included blockage of the TSV™ cutter tip during attempted clearance of organized intragel haemorrhage, thus requiring conversion to conventional trans pars plana vitrectomy (1; 2.6%). No other intraoperative complication arose. Complications in the postoperative period included: reactivation of dormant chorioretinal toxoplasmosis (1; 2.6%); postoperative posterior segment haemorrhage (3; 7.7%); hyphaema (1; 2.6%); transient postoperative hypotony (10; 25.6%).
Conclusions
With appropriate case selection, and with appreciation of this technique's limitations, transconjunctival sutureless vitrectomy is a safe procedure.
Similar content being viewed by others
Introduction
The use of 25 gauge transconjunctival sutureless vitrectomy (TSV 25™) has many advantages over conventional trans pars plana vitrectomy. First, the self-sealing transconjunctival sclerotomies obviate the need for conjunctival peritomy, with a consequential reduction of operating time and surgically induced trauma at the sclerotomy sites. Discomfort from conjunctival sutures is also eliminated using the TSV 25 system, as is postoperative astigmatism arising from scleral sutures. A less cited, but very important, advantage of the TSV 25 system rests on the absence of subconjunctival fibrosis and adhesions postoperatively, thus rendering any subsequent vitreoretinal and/or filtration surgery easier. Finally, transconjunctival sutureless vitrectomy results in a more aesthetically pleasing eye in the immediate postoperative period. We report on the experience of the first 39 consecutive cases of vitreoretinal surgery using the TSV 25 system in our unit, and discuss the outcomes thereof and make recommendations so that other vitreoretinal surgeons can avoid pitfalls with this novel and innovative technique.
Materials and methods
We retrospectively reviewed the medical records of 35 patients (39 eyes) who consecutively underwent trans pars plana vitrectomy using the TSV 25 system at the Department of Ophthalmology at Waterford Regional Hospital between March 2003 and February 2005. Data retrieved may be classed as preoperative, intraoperative and postoperative. Preoperative details recorded included: intraocular pressure (IOP) and visual acuity (VA); indications for surgery; ophthalmic details, including previous intraocular surgery in the study eyes. Intraoperative data included: method of anaesthesia; operating time; complications (if any). Postoperative details included IOP and VA recordings, as well as the clinical course. VA was typically recorded in the medical records as Snellen visual acuity, but this was converted to the logMAR scale using a conversion chart1 to aid with statistical analysis.
Surgical technique
A Bausch and Lomb Millennium™ TSV 25™ System for transconjunctival standard vitrectomy was used in all cases. We created three ports through the pars plana, in the superotemporal, superonasal, and inferotemporal quadrants, respectively. No conjunctival peritomy was performed and an entry site alignment system was used to maintain alignment between the conjunctival and scleral entry sites. Microcannulae were used to facilitate this, which consist of thin-walled polyamide tubes 3.6 mm in length with an inner/outer diameter of 0.57/0.62 mm and were inserted into the eye using a trocar. We displaced the conjunctiva laterally when inserting the microcannulae and sutures were not required to hold these in place. The first microcannula was inserted inferotemporally to provide access for the infusion cannula (which does not need to be sutured in place). We placed the other microcannulae more inferiorly than typical, at 10 and 2 O’clock instead of at the conventional 11 and 1 O’clock positions, thus allowing a better view of superior vitreous and retina intraoperatively. The surgeries were performed using 25 gauge instrumentation, introduced into the eyes via the preplaced microcannulae. Postoperatively the microcannulae were removed and the conjunctiva gently manipulated over the sclerotomy sites to prevent leakage.
Results
Mean (±standard deviation (SD)) age of patients at the time of surgery was 67.63 (13.31) years and the male to female ratio was 22 : 17. Indications for surgery are given in Table 1. In all, 31 (79.5%) surgeries were performed using retrobulbar anaesthesia, and the remaining eight (20.5%) were performed under general anaesthesia. The mean duration (±SD) of surgery was 32.85 (17.54) min, with a range of 5–85 min. Seven (17.95%) cases were performed on a daycase basis and the mean length of hospital stay was 1.53 (1.25) days with a range of 0–4 days.
Mean (±SD; range) visual acuities (logMAR) were −1.335 (1.15; 0.0–−3.0) and −0.803 (1.03; 0.1 to −3.0) preoperatively and postoperatively, respectively. The mean (±SD) intraocular pressures were 14.69 (4.64) and 11.56 (9.03) mmHg preoperatively and at the first dressing, respectively. In all, 10 eyes (25.6%) suffered transient postoperative hypotony (IOP<5 mmHg) for at least one postoperative day; however, by day 4 the intraocular pressure had normalized in all of these cases.
One intraoperative complication arose, where TSV™ cutter tip became blocked during attempted clearance of organized intragel haemorrhage, necessitating enlargement of the sclerotomy and conversion to conventional the trans pars plana vitrectomy. Of note, no TSV 25™ sclerotomy sites required suturing in the postoperative period due to leakage. Complications in the postoperative period included: reactivation of dormant chorioretinal toxoplasmosis (1; 2.6%); postoperative posterior segment haemorrhage (3; 7.7%); hyphaema (1; 2.6%); transient postoperative hypotony (10; 25.6%).
Discussion
Transconjunctival sutureless vitrectomy was first described by Fujii et al2, 3 in 2002. Since then there have been a paucity of articles describing and exploring the potential indications for this technique, as well as surgical complications.1, 2, 3, 4, 5, 6, 7, 8 We report the first 39 consecutive transconjunctival vitrectomies performed at Waterford Regional Hospital, discuss our experience with this relatively new technique, and furnish the reader with some surgical pointers to avoid potential complications with this technology.
In our series, we encountered no complication with the TSV 25™ system, which can be attributed to the use of this method over conventional trans pars plana vitrectomy, other than a transient hypotony. We defined postoperative hypotony as an intraocular pressure of less than or equal to 5 mmHg on the first postoperative day. The prevalence of this complication was high (10/39; 25.6%) presumably due to the sutureless nature of this technique and consequent slow leakage from the entry ports. Risks associated with transient postoperative hypotony include an increased risk of endophthalmitis (due to the facilitation of intraocular access to microorganisms) and an increased risk of choroidal detachment. None of these eyes however, suffered choroidal detachments or any adverse sequelae and intraocular pressure had normalized by day 4 postoperatively in all cases. Of note, the prevalence of postoperative posterior segment haemorrhage following TSV 25™ vitrectomy for haemorrhage clearance, was also high (3/10; 30%) and may be related to the postoperative hypotony. To try and reduce the incidence of postoperative hypotony, we recommend that the surgeon displace the conjunctiva laterally when inserting the microcannulae so that the anatomical recoil of the displaced conjunctiva will ensure that the sclerotomies are adequately covered in the immediate postoperative period, thus reducing the risk of leakage and/or infection (see Table 2). In any case, no complications resulted in long-term sequelae, which compromised the vision.
As with all new and innovative techniques, there are potential advantages and disadvantages to consider. Advantages of transconjunctival sutureless vitrectomy include a reduction in operating time, thus enhancing theatre efficiency. Another advantage of the TSV 25™ system includes less subconjunctival adhesion formation postoperatively, thus facilitating any subsequent vitreoretinal surgery or filtration surgery, if required. Further, the introduction of instruments through a microcannula should, in theory, be associated with a lower risk of entry site breaks, because shearing of the vitreous base upon repeated introduction of surgical instruments through a conventional sclerotomy is avoided. Finally, transconjunctival sutureless vitrectomy is associated with a more aesthetically pleasing appearance of the eye in the immediate postoperative period when compared with the appearance following traditional pars plana vitrectomy.
The disadvantages of TSV 25™ vitrectomy include an increased incidence of postoperative hypotony, thus necessitating daily review of the patient and possibly affecting the proportion of patients suitable for daycase surgery. Also, the instruments for TSV 25 vitrectomy are very small and malleable, resulting in limited maneuverability of the eye. Further, the small size of the cutting port and a potentially weaker aspiration may lead to difficulty with clearance of dense vitreous haemorrhage and with induction of posterior vitreous detachment, respectively. Moreover, there is a theoretically increased risk of endophthalmitis following transconjunctival sutureless vitrectomy because of the transient hypotony with consequential facilitation of intraocular access to microorganisms. And finally, there is an inevitable learning curve inherent with any new surgical technique, which will represent additional psychological stress to the surgeon and surgical assistants.
We have compiled a list of surgical pointers to aid those contemplating transconjunctival sutureless vitrectomy, so that potential pitfalls can be avoided (Table 2).
In summary, we believe that the TSV 25™ vitrectomy system, or similar transconjunctival vitrectomy methods, represent a significant advance in vitreoretinal surgery, and has a role to play. TSV 25™ vitrectomy is a safe and effective procedure for a wide variety of vitreoretinal indications and we believe that vitreoretinal surgeons should become acquainted with this new technology, and comfortable with it, at the very least for simple internal posterior segment procedures or surgeries. In the context of appropriate case selection, and with full appreciation of the limitations of transconjunctival sutureless vitrectomy, transconjunctival sutureless vitrectomy does not compromise the surgical or visual outcome of internal posterior segment surgery.
References
Holladay JT . Guest Editorial. J Cat Refract Surg 2004; 30 (2): 287–290.
Fujii GY, de Juan Jr E, Humayun MS, Pieramici DJ, Chang TS, Awh C et al. A new 25-guage instrument system for transconjunctival sutureless vitrectomy surgery. Ophthalmology 2002; 109: 1807–1813.
Fujii GY, de Juan Jr E, Humayun MS, Chang TS, Pieramici DJ, Barnes A et al. Initial experience using the transconjunctival sutureless vitrectomy system for vitreoretinal surgery. Ophthalmology 2002; 109: 1814–1820.
Ibarra MS, Hermel M, Prenner JL, Hassan TS . Longer-term outcomes of transconjunctival sutureless 25-gauge vitrectomy. Am J Ophthalmol 2005; 139 (5): 831–836.
Hilton GF, Josephberg RG, Halperin LS, Madreperla SA, Brinton DA, Lee SS et al. Office-based sutureless transconjunctival pars plana vitrectomy. Retina 2002; 22 (6): 725–732.
Eckardt C . Transconjunctival sutureless 23-gauge vitrectomy. Retina 2005; 25 (2): 208–211.
Lee HK, Kim CY, Kwon OW, Kim EK, Lee SC, Seong GJ et al. Removal of dense posterior capsule opacification after congenital cataract extraction using the transconjunctival sutureless vitrectomy system. J Cataract Refract Surg 2004; 30 (8): 1626–1628.
Shah VA, Gupta SK, Chalam KV . Management of vitreous loss during cataract surgery under topical anaesthesia with transconjunctival vitrectomy system. Eur J Ophthalmol 2003; 13 (8): 693–696.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
O' Reilly, P., Beatty, S. Transconjunctival sutureless vitrectomy: initial experience and surgical tips. Eye 21, 518–521 (2007). https://doi.org/10.1038/sj.eye.6702255
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/sj.eye.6702255
Keywords
This article is cited by
-
Intraocular pressure outcomes after 23-G vitreoretinal surgery with two different transconjunctival sutureless sclerotomy techniques: vertical versus tunnel entry
International Ophthalmology (2022)
-
25th RCOphth Congress, President's Session paper: 25 years of progress in vitreoretinal surgery
Eye (2014)
-
A comparative study of tissue glue and vicryl suture for conjunctival and scleral closure in conventional 20-gauge vitrectomy
Eye (2009)
-
23-Gauge transconjunctival sutureless pars plana vitrectomy: results of a prospective study
Eye (2009)
-
Transconjunctival sutureless vitrectomy: comment on O'Reilly and Beatty
Eye (2007)