Sir,
Diode laser trans-scleral cyclophotocoagulation (DLTSC) is an effective treatment in refractory glaucomas.1, 2, 3 Complications, such as uveitis, pupillary distortion, conjunctival burns, hyphema, chronic hypotony, cystoid macular oedema, retinal detachment, and phthisis bulbi, have been reported.1, 4, 5, 6, 7, 8 Two cases of scleral perforation requiring suturing were described in the published literature.9, 10 We describe a case of scleral perforation during DLTSC in a patient with pre-existing scleral thinning, confirmed with ultrasound biomicroscopy. The leakage stopped after 1 day of patching without the need of suturing.
Case report
A 66-year-old Chinese woman had undergone limbal incision ECCE to both eyes 20 years ago with no intraocular lens implantation. Thereafter, she had been abusing unprescribed topical ophthalmic steroid for more than 20 years, unaware of its potential side effects. She had no other ophthalmic or medical history of note. On April 2004, she suffered a blunt trauma to her left eye while she bumped into a glass door with a resultant 3.5-mm full thickness dehiscence of the limbal cataract wound in the supero-temporal quadrant, with vitreous prolapse. There were peripheral corneal and scleral thinning from 10 to 2 o' clock positions, presumably related to her chronic steroid abuse.
Emergency repair of the corneal wound and anterior vitrectomy were carried out, followed by a repeat pars-plana vitrectomy and anterior vitrectomy 2 weeks later for persistent vitreous incarceration in the corneal wound.
Nevertheless, she suffered from subsequent secondary synechial angle closure glaucoma. Despite maximal antiglaucomatous medications, including four topical medications and oral acetazolamide 500 mg four times a day, her intraocular pressure eventually reached 48 mmHg. Gonioscopy examination revealed severe peripheral anterior synechiae >270° and a cup-to-disc ratio of 0.8. Visual acuity was 4/60 in the left eye at this stage.
DLTSC was performed under retrobulbar anaesthesia using a G-probe (IRIS Medical Instruments, Inc., Mountain View, CA, USA). A total of 14 laser spots each of 1500 mW, 2-s duration, were applied. A popping sound was heard at six points.
While applying the 14th laser application over the thinned sclera at 10 o' clock position, a gush of aqueous was noticed. Inspection under the operating microscope confirmed a round, ‘punched out’ full thickness perforation, 1.2 mm behind the limbus with leakage demonstrated with Siedel test (Figure 1). There were no vessels or pigment around the perforation site. There were no other conjunctival or scleral burns. The G probe was inspected and no black deposits at the tip were observed. The laser procedure was aborted. The eye was padded with antibiotic ointment and crepe bandage. Oral acetazolamide 500 mg was given four times daily to reduce aqueous flow through the perforation.
On the next day, the leakage had stopped, possibly having been plugged by vitreous. The intraocular pressure was 9 mmHg. Topical prednisolone acetate 1% four times daily, topical chloramphenicol 0.5% four times daily, and topical atropine 1% twice daily were prescribed. During the subsequent 3 weeks, the scleral hole healed. Ultrasound biomicroscopy revealed a scleral thickness of 0.437 mm around the perforation site. In the ensuring 3 months, the intraocular pressure gradually increased to 40 mmHg despite antiglaucomatous agents. At the time of writing, the patient was pain-free, and she refused further intervention. Visual acuity was hand movement in the left eye at most recent follow-up. The right eye was aphakic with a best-corrected visual acuity of 1.0. The remote risk of sympathetic ophthalmitis and its symptoms were discussed, and the patient preferred observation at this stage.
Comment
Diode laser of 810 nm wavelength is an effective tool for treating refractory glaucomas.1, 2, 3
To the best of our knowledge, there were two reported cases of scleral perforation due to DLTSC in the published literature.10, 11 The case reported by Gaasterland and Pollack9 was thought to be due to the sharp edge of the probe cutting conjunctival vessels and causing bleeding. Thin adherent debris was then carbonized, allowing the laser tip temperature to rise and causing scleral perforation. The defect required suturing. This led to the redesigning of the laser probe tip. In our case, there was no such carbonized debris seen and we think it is unlikely to be the reason for our perforation.
Sabri and Vernon10 reported a case of scleral perforation using the new contact G-probe. The defect required suturing with two 10-0 vicryl sutures. However, 1 week later, the scleral leak recurred and further suturing was needed.
In our case, with crepe bandage and oral acetazolamide for 1 day, we were able to stop the leakage. During subsequent follow-up, the scleral hole healed and was covered by intact conjunctival epithelium. This demonstrated that suturing may not always be necessary, especially when the perforation is small.
Pre-existing scleral thinning is a common risk factor in the two previously reported cases, and also in our patient. Hence, Sabri and Vernon10 suggested the use of a lower laser power setting (50%), though there is no good proof that such a lower power could prevent perforation and is still as effective.
With heightened awareness of this complication, and appropriate management when it occurs, we believe the risk of scleral perforation and its consequences could be minimized.
References
Mastrobattista JM, Luntz M . Ciliary body ablation: where are we and how did we get here. Surv Ophthalmol 1996; 41: 193–213.
Gupta N, Weinreb RN . Diode laser transscleral cyclo-photocoagulation. J Glaucoma 1997; 6: 426–429.
Threlkeld AB, Johnson MH . Contact transscleral diode cyclophotocoagulation for refractory glaucoma. J Glaucoma 1999; 8: 3–7.
Kosoko O, Gaasterland DE, Pollack IP, Enger CL . Long-term outcome of initial ciliary ablation with contact diode laser transscleral cyclophotocoagulation for severe glaucoma. The Diode Laser Ciliary Ablation Study Group. Ophthalmology 1996; 130: 1294–1302.
Bock CJ, Freedman SF, Buckley EG, Shields MB . Transscleral diode laser cyclophotocoagulation for refractory pediatric glaucomas. J Pediatr Ophthalmol Strabismus 1997; 34: 235–239.
Bloom PA, Tsai JC, Sharma K, Miller MH, Rice NS, Hitchings RA et al. ‘Cyclodiode’: trans-scleral diode laser cyclophotocoagulation in the treatment of advanced refractory glaucoma. Ophthalmology 1997; 104: 1508–1520.
Schlote T, Derse M, Thiel H, Benedikt J . Pupillary distortion after contact transscleral diode laser cyclophotocoagulation. Br J Ophthalmol 2000; 84: 337–339.
Bhola RM, Prasad S, McCormick AG, Rennis IG, Talbot JF, Parsons MA . Pupillary distortion and staphyloma following trans-scleral contact diode laser cyclophotocoagulation: a clinicopathological study of three patients. Eye 2001; 15: 453–457.
Gaasterland DE, Pollack IP . Initial experience with a new method of laser trans-scleral cyclo-photocoagulation for ciliary ablation in severe glaucoma. Trans Am Ophthalmol Soc 1992; 90: 225–246.
Sabri K, Vernon SA . Scleral perforation following trans-scleral cyclodiode. Br J Ophthalmol 1999; 83(4): 502–503.
Schlote T, Derse M, Rassmann K, Nicaeus T, Dietz K, Thiel HJ . Efficacy and safety of contact transscleral diode laser cyclophotocoagulation for advanced glaucoma. J Glaucoma 2001; 10: 294–301.
Author information
Authors and Affiliations
Corresponding author
Additional information
Financial and proprietary interest: Nil
Rights and permissions
About this article
Cite this article
Kwong, Y., Tham, C., Leung, D. et al. Scleral perforation following diode laser trans-scleral cyclophotocoagulation. Eye 20, 1316–1317 (2006). https://doi.org/10.1038/sj.eye.6702179
Published:
Issue Date:
DOI: https://doi.org/10.1038/sj.eye.6702179
This article is cited by
-
Skleraeinschmelzung nach Zyklophotokoagulation
Der Ophthalmologe (2019)
-
Minimising blinding complications of cyclodiode laser in high risk and only eyes
Eye (2007)