Introduction

Generalized diseases of the retina such as, for example, retinitis pigmentosa are detected by standardized Ganzfeld electroretinography (ERG) with high sensitivity, even in early stages of the disease.1,2,3 In localized retinal diseases, for example, in macular diseases, the Ganzfeld ERG is a less useful technique, as its global response is only affected in proportion to the size of the altered retinal area. The pattern ERG (PERG) improved at that point our ability to examine localized retinal diseases.4,5 However, attempts to obtain ERG responses from small retinal areas never overcame the main problem of small signal-to-noise ratio.6,7 The multifocal ERG allows topographic mapping of retinal function of about 50° of the central retina.8,9 A stimulus array is presented on a monitor and the stimulated areas are changing independently alternating between black and white. The patient has to fixate on a small fixation spot presented in the middle of the stimulation matrix during the whole examination. A major problem of this technique is that no sufficient control of fixation can be performed by the examiner. Therefore, in patients with fixation problems, results may be unreliable, such as in patients with reduced compliance, in children or patients with eccentric fixation. To overcome these difficulties, Kondo et al 10 used an infrared fundus camera combined with a stimulus array of 19 light-emitting diodes (LEDs), allowing the performance of a multifocal ERG under simultaneous fundus control. Seeliger et al 11 showed the application of scanning laser ophthalmoscope-evoked mfERG (SLO-mfERG) measurements in animals, using a green laser (515 nm). In many animals, the only way to obtain satisfactory results exists in using a short-wave laser because of the lack of long-wavelength cones.11

In this study, we used an SLO-mfERG to examine patients with macular hole (Figure 1) and normal individuals. This new technique offers the possibility to generate a stimulus array with a helium–neon laser under simultaneous control of fixation.12,13,14,15,16,17 The purpose of this study was to assess the feasibility of this method to measure focal retinal function or dysfunction under simultaneous control of fixation.

Figure 1
figure 1

Fundus photography of a patient with a macular hole (Gass II).

Materials and methods

The scanning laser ophthalmoscope (Rodenstock; München, Germany) was used both as a stimulus and trigger device as well as a fundus-monitoring unit. To record an mf ERG, a confocal scanning laser ophthalmoscope is connected to an mfERG device (RETIscan, Roland Consult, Wiesbaden, Germany). The visual stimulus, consisting of a matrix of hexagonal areas flickering concurrently according to a short m-sequence, is displayed directly onto the retina covering a visual angle of about 12° or a diameter of 24°, respectively (Figure 2).18,19 It consists of a matrix of 61 hexagonal elements and is generated by a helium-neon laser (632.8 nm) modulated via an acousto-optic modulator. The matrix has been presented according to a scaling factor 1:4, so that the hexagonal size would be accordingly scaled with eccentricity in order to evoke focal responses of approximately the same amplitude per stimulus element in the response arrays of normal individuals.20 With this setting the outermost hexagonal areas were approximately four times larger than the central one. The elements change from red to black or vice versa according to a complete cycle of a binary m-sequence.21 The contrast is about 90%.22 The Roland system uses ‘short-m-sequences’ and a complete measurement cycle consists of 511 steps lasting 40.8 sec (511 × 80 ms). Scan time is 20 ms prohalf-frame (50 half-frames/s, 25 frames/s). The examined persons were asked to gaze at a centrally positioned cross. If loss of fixation occurred, the measurement would be interrupted. In cases of unstable fixation the measurement cycle was discarded. The signal of each hexagonal field is calculated by cross-correlation analysis between the m-sequence and the response. The signals are bandpass filtered (10–100 Hz) and amplified by a two-channel amplifier with a common mode reaction of >110 dB and a sensitivity of 1–20 μV/div. Simultaneous funduscopy and control of fixation is guaranteed by illumination with an infrared diode laser (730 nm).

Figure 2
figure 2

The stimulus pattern on the posterior pole of the eye: the figure displayed was derived from the screen that is used for real-time fundus monitoring, where the fundus picture and the stimulation matrix appear in black and white. Actually, the stimulation pattern generated by the helium–neon laser consists of red light.

Patients and controls

In all, 10 patients (ages 46–77 years, mean 64 years) with unilateral diagnosis of an idiopathic macular hole were compared with 19 healthy volunteers (ages 18 to 63 years, mean 40 years) contributing to the study with one eye each. Normal eyes of patients did not serve as controls. Altogether 29 persons were examined with SLO-mfERG using a scaling factor of 1:4. Ophthalmic examination included slit-lamp examination, measurement of visual acuity, funduscopy, and fundus photography. Macular holes were diagnosed by biomicroscopy using a 78 dioptre lens (Volk) and by optical coherence tomography (OCT). Two patients with a macular hole stage II according to Gass showed central fixation, while eight patients with macular hole stage Gass III demonstrated slightly paracentral fixation, but still in the central hexagonal element. Patients with more eccentric fixation were excluded from the study. The visual acuity range was 0.05–0.4.

The recordings were analysed by using concentic ring grouping with evaluation of amplitude density and by displaying the plots. ERGs retrieved from the individual hexagonal areas would also be plotted. Special traces could be selected for detailed data analysis. Amplitude density (nV/deg2) represents the amplitudes for each grouping adjusted for the distance angle of the stimulus element, giving a more accurate view of the actual response amplitudes of each group.23 To construct the two and three-dimensional plots the mean of amplitude density was used.9

The criteria for the normal controls included in the study were an age of over 18 years, visual acuity of 1.0 with or without correction and contact lens tolerance for both groups. Exclusion criteria for both groups were media opacities, refractive errors of more than 3 dioptres, a history of ocular trauma, diabetic retinopathy, age-related macular degeneration, or glaucoma. The pupils were not dilated. Refractive errors would be compensated from an integrated system of the SLO. This, generally described, consists of movable mirrors that automatically can be moved, shifting therefore the light beam. Jet contact lens electrodes (Universa SA, La Chaud-de-Fonds, Switzerland) were used. The reference electrode was placed in line with ocular canthi behind the orbital ridge. This research adhered to the tenets of the Declaration of Helsinki and informed consent was obtained from all persons.

Results

To quantify the electrophysiological results, the responses were grouped by retinal eccentricity in five ring areas. To summarize the results, boxplots showing the 25th and 75th percentile of amplitude density were calculated for both groups of examined persons (healthy individuals and patients with macular holes) (Figure 3). Fifth, 25th, 75th, 95th percentiles for the normative data and the patients' individual responses can be seen in Table 1. The mean and median values as well as the standard deviation (SD), the interquartile range, and the coefficient of variation of the central hexagonal element and the rings were calculated for each group (Table 2). Our data were tested for and were found inconsistent with the normality assumption (using the Kolmogorov–Smirnov test). The significance level was set at P=0.05 and the Mann–Whitney test was applied. Two- and three-dimensional plots of amplitudes P1 both for the healthy individuals as well as for the patients were derived and compared (Figures 4a,b and 5a,b).

Figure 3
figure 3

Bold lines in the boxes indicate median, upper and lower edges of the boxes indicate 25th and 75th percentiles accordingly. Upper/lower ends of bars indicate highest/lowest data that is not an outlier. Circles on the graph represent ‘outliers’. Outliers are cases with values between 1,5 and 3 box lengths from the upper or lower edge of the box. The box length is the interquartile range. Small stars on the graph represent ‘extremes’. The latter are cases with values more than three box lengths from the upper or lower edge of the box.

Table 1 Individual responses of the patients and the 5th, 25th, 75th, and 95th percentiles for the normative data
Table 2 Amplitude density (nV/deg2): statistical values of normal individuals and patients with macular holes
Figure 4
figure 4

(a) Two-dimensional presentation of amplitudes P1 of a healthy person. (b) Three-dimensional presentation (side view) of amplitudes P1 of a healthy person showing the central peak.

Figure 5
figure 5

(a) Two-dimensional presentation of amplitudes P1 of a patient with macular hole. (b) Three-dimensional presentation (side view) of amplitudes P1 of a patient with macular hole showing the reduction of the central peak.

The amplitudes of the first ring area (central hexagonal element) were found statistically significantly higher in the group of healthy individuals (P<0.05), M (normals)=198.6, SD=74.3; M (macular holes)=104.8, SD=54.0. The amplitudes and latencies of rings 2–5 areas showed no significant differences (Figures 6a,b and 7a,b, Table 2).

Figure 6
figure 6

(a) Plots of amplitudes P1 from the 61 hexagonal elements of a healthy person. The number of the hexagonal elements is given at the top of the curves while at the bottom appears the amplitude of each field (nV). (b) Concentric ring grouping of a healthy person showing the amplitudes and latencies for each ring. In the first columns is given the amplitude density in nV/deg2. In the second and third column are the summarized amplitudes P1 and N1 accordingly per ring grouped area in μV. In the fourth and fifth columns are the latencies P1 and N1 for the summarized curves in ms.

Figure 7
figure 7

(a) Plots of amplitudes P1 from the 61 hexagonal elements of a patient with macular hole. The number of the hexagonal elements is given at the top of the curves while at the bottom appears the amplitude of each field (nV). No recognizable amplitudes in the fields 14, 22, 25, and 48 may be caused by shifts because of microsaccades. (b) Concentric ring grouping of a patient with macular hole showing the amplitudes and latencies for each ring. In the first column is given the amplitude density in nV/deg2. In the second and third columns are the summarised amplitudes P1 and N1 accordingly per ring grouped area in μV. In the fourth and fifth columns are the latencies P1 and N1 for the summarised curves in ms.

The coefficient of variation (or percentage error) measures the relative scatter of data with respect to the mean and therefore the imprecision in our estimates. It is calculated as 100 (SD)/(mean value of set). In our data set, the coefficient of variation of the response density of the central hexagonal element is relatively higher centrally in the patients with a macular hole as compared to the healthy individuals.

Discussion

The Ganzfeld ERG produces a generalized response because of the diffuse stimulation of the retina and therefore neglects the effect of limited lesions. Conventional, monitor stimulated mfERG is a technique capable of measuring retinal function of the posterior pole with high spatial resolution. By multifocal stimulation and by the simultaneous measurement of all the focal responses, a strict correlation of any given signal to a corresponding field exists.24,25,26 By using short m-sequences, the quality of each measurement cycle can be evaluated separately, and any cycle in which significant artefacts occur can be discarded. The SLO-system we have employed is also a significant technical advance. It allows the direct projection of the laser-generated stimulus pattern onto the retina. Moreover, the presence of an infrared diode laser allows simultaneous funduscopy, control of fixation, and imaging of the hexagonal matrix during the whole examination. This is important for all patients with instability of fixation. The importance of an appropriate fixation for a reliable stimulation–response correlation analysis has been lately shown using mfERG recordings and SLO-microperimetry.27,28 Moreover, in a recent study where mfPERG recordings were obtained with an eye tracker from normal individuals, 49% of the recordings were acquired, while fixation was inadequate.29 Loss of fixation would lead to an artificial correlation of the recorded amplitudes to the expected measured retinal areas. The SLO allows continuous monitoring of fixation. The patient can be encouraged and epochs in which control is poor can be discarded. Ultimately, software that compensates for eye movements could be introduced.

Although the main advantage of the mfERG is its spatial resolution, in accordance with Parks et al,30 who concluded that repeatability of mfERG can be further enhanced by forming regional groups, concentric rings were grouped around the innermost hexagon of the stimulation array. In concentric macular diseases such as macular holes, an averaging of concentric rings allows a better comparability of responses obtained from the central retina to those derived from more eccentric parts.30 As known from the conventional mfERG, response density decreases markedly from the foveal region towards the periphery. This distribution of response density approximately corresponds to the anatomical occurrence of cone distribution of the posterior pole.18,19,20,31 The control group in this study was not age matched, which could have an influence on the results because it is known that an age-related effect exists. Other studies showed a decrease of amplitudes in older patients.32,33

In patients with macular hole, a significant reduction of response density could be demonstrated in the centre of the stimulus array. Retinal dysfunction leads to profound amplitude reduction of the central hexagonal element in the SLO-mfERG results of these patients. In the areas corresponding to the elevated margins adjacent to the hole, no significant reduction could be seen in the retinal electrical response. As expected, the strength of the recorded amplitudes of the more peripheral areas does not differ significantly in these patient in comparison with normals.34,35,36,37,38

The higher coefficient of variation of amplitudes of the central hexagonal element in patients with macular holes (Table 2) could be explained because of loss of fixation in these patients. However, it raised questions about the measurements' reliability. Fixation loss (through foveal displacement) shifts the central hexagonal element to a more peripheral retinal area, and for a short but variable time period destroys correlation between recorded amplitude and field location. Part of the response attributed to the inner ring will also be generated by the fovea and vice versa, which would be otherwise covered by the central hexagonal element. Our measurements were made under simultaneous fundus visualization and direct fixation control, and still the coefficient of variation implies that such results contain significant errors. Not only in patients but also in healthy individuals, the fixation can pose a real problem. When performing the examination with the monitor version, these possibilities of fixation control are not available. Eyetracking, among other efforts, to solve this problem cannot be a reliable solution, because absence of eye movement does not absolutely mean proper fixation of the eye to the presented target (fixation cross). Thus, when the above-mentioned precautions are not available, it is important to remember that scaling may introduce significant difficulties in obtaining reliable data.11,29,39

While fixation loss cannot be prevented, the value of our fixation monitoring system depends on the possibility of continuous visualization of fixation during the examination. Fixation depending on concentration, motivation, and retinal function can be, on the one hand, positively affected and on the other objectively evaluated by using the SLO direct fundus control method in order to record mfERGs. In conclusion, it seems that the SLO-mfERG is a useful technique to investigate retinal and in particular macular function under control of fixation, and further studies will be helpful to improve this diagnostic tool.