Introduction

Quantitative analysis of fundal lesions is important in many areas of ophthalmology including the assessment of lesion size for laser therapy in photodynamic therapy and also for the monitoring of potentially progressive lesions like choroidal naevi. Digital acquisition systems are increasingly employed in fundus photography. The instant availability of the images on the computer monitor allows instant interpretation leading to earlier diagnosis of fundal abnormalities. Many digital systems offer a built-in measuring tool allowing quantitative analysis of linear and spatial dimensions of posterior pole features. We are not aware of any study that has evaluated the reliability of such measuring tools on a digital system. In this study, we investigated the measuring tool in our digital system by comparison with a standard technique employed in the analogue film system.

Materials and methods

A Topcon TR501A fundus camera equipped with a high-resolution Kodak Megaplus monochrome digital camera as well as a conventional film camera back was used.1 The digital system used was Topcon Imagenet 2000. During routine fluorescein angiography studies on 10 patients, with refractive errors of less than 4D, images were captured using the digital system as well as film. All images were captured with a 35° field setting on the camera. This produces the same magnified view both on film as well as on the monitor. However, in our model of Topcon camera, the field setting on the capture software has to be manually selected. This gives an opportunity to test the accuracy of the measuring tool if the field setting on the software did not correspond to that of the camera.

In the first five patients (group 1), the field settings on the camera and in the capture software were set at 35°. In the next five patients (group 2), the settings were mismatched so that the camera was set at 35° while the capture software was set at 50°. In total, 10 film images, each with a corresponding digital image from 10 patients, were used for the study. In each image, 10 wide-ranging linear dimensions were selected using easily identifiable landmarks.2 For each linear dimension chosen, the distance was measured on the film from the negative using a transparent reticule with 0.1 mm intervals and divided by 2.5 to correct for the camera's magnification.3 The corresponding linear dimension on the digital image was measured using the software's measuring tool. Measurements taken with the two methods were compared using linear correlation as well as a graphical method for agreement and bias.4,5

Results

A total of 10 film/digital pairs of images were used for comparison. In each pair, 10 linear dimensions were chosen giving a total of 100 comparisons. The magnitude of the dimensions selected ranged from 0.6 to 10.4 mm.

In group 1, where the camera's field setting and the capture software's setting were 35°, the correlation between measurements obtained on film and those obtained digitally are shown in Figure 1. The correlation coefficient was 0.99 indicating a high correlation between the two methods of measurement. Agreement between analogue measurements and digital measurements is shown in Figure 2. On average, the film measurements were 0.32 mm less than digital. The 2SD limits of agreement were (−0.88 to +0.24), indicating high level of agreement between analogue and digital. However, it is clear from the agreement plot in Figure 2 that agreement between the two methods decreased with increasing dimensions.

Figure 1
figure 1

Group 1—correlation measurements.

Figure 2
figure 2

Group 1—agreement between analogue and digital measurements.

In group 2, where there was a discrepancy between the settings of the fundus and the digital camera (35° for the fundus and 50° for the digital), the correlation coefficient was also high (0.98), but from the agreement plot (Figure 3) analogue measurements were 1.25 mm less than digital measurements on average. The 2SD limits of agreement were −2.91 to +0.40 mm. The incorrect setting gave rise to considerable bias in the digital measurements. There was also a tendency for the measurements to be less accurate with increasing distances measured.

Figure 3
figure 3

Group 2—agreement between analogue and digital measurements.

Discussion

Linear dimensions measured using the measuring tool on the Imagenet 2000 software showed a good level of agreement with those estimated using the conventional reticule on film negative method and should be reliable when used in clinical practice. Although larger dimensions were associated with less agreement, the deviations were small as shown by the 2SD limits of within 1 mm.

In this study, we compared images taken by the same camera using different capturing media so the refractive errors of the patients were not relevant. To ensure good image quality, we used patients with low refractive errors.6,7

The graphical method of assessing agreement used in this study merely compared the digital system to a standard analogue technique. It does not imply that the analogue technique is more accurate than the digital system. It is possible that there could have been some operator error involved in either technique. Without resorting to doing a repeatability type evaluation of what was essentially a length measuring procedure, the high level of correlation (0.99) between the two types of measurements indicates that there is unlikely to be any significant random operator-error as a lower correlation coefficient would be obtained.

From the agreement plots, the bias was negligible when the capture software was set correctly at 35°. However, in group 2 when the capture software was set incorrectly to 50°, a significant bias in the measurement was introduced. This problem can occur in older cameras such as the one used in our study. When using such a camera, the field of view shown on the monitor screen of the image captured corresponds to the field setting on the camera and not to the field setting on the capture software. The measuring tool calibration adopted by the software depends on the field setting on the software at the time of capture. It is an easy pitfall to capture at the wrong field setting, which could give rise to erroneous measurements. Fortunately, with many newer camera models and newer imaging software, when the field setting is altered on the camera, the connectivity functions between the camera and the imaging software ensure that the software capture settings are altered automatically. This is an important feature to look for especially when using imaging software, which is not supplied by the camera manufacturer.