Abstract
Aim The study was carried out to see if levels of nitrous oxide in dental theatres and community dental clinics were being controlled in accordance with the Control of Substances Hazardous to Health Regulations.
Setting A multi-centre study looked at exposure levels in a dental teaching hospital and two community dental clinics in the South Wales area between 1997 and 1998.
Methods A MIRAN infra-red spectrophotometer was used to measure static levels of nitrous oxide during general anaesthesia and conscious sedation. NIOSH method 6600 was used to collect personal samples of the individual administering the anaesthetic.
Results The results showed compliance with the regulations when averaged out over an 8 hour time weighted average. However, over short periods of time peak concentrations of up to 1190 ppm were observed through static sampling, and up to 734 ppm through personal sampling.
Conclusions The results highlighted the need to provide adequate control measures such as anaesthetic gas scavenging, to reduce occupational exposure to dental staff in dental operating theatres and community dental clinics, therefore providing a safe working environment.
Similar content being viewed by others
Main
Nitrous oxide is still widely used in both dental theatres and community dental clinics. Many of these areas, especially community dental clinics, were not designed for the use of anaesthetic agents and therefore do not have the necessary ventilation and anaesthetic gas scavenging available to them. This may result in dentists and dental assistants being exposed to levels above the occupational exposure limit of 100 ppm over an 8 hour time weighted average (TWA) set by the Health Services Advisory Committee (HSAC)1 which came into force on the 1 January 1996.
There is concern relating to high occupational exposures since nitrous oxide is known to exert a number of effects.
Megaloblastic changes are reported in individuals with prolonged clinical exposures2 and interference with DNA synthesis has been reported for patients exposed for 2–4 hours.3
Although N2O is teratogenic in rats at high doses,4 evidence of reproductive effects in humans is equivocal. Early studies on spontaneous abortion in operating theatre staff5,6,7 as well as dated and recent surveys of dental assistants8,9 or midwives10 do not permit accurate risk estimation. There is no evidence of increased risk of malignancy in exposed personnel.11
Environmental monitoring
In a dental hospital, a total of 12 static monitoring sessions were carried out in the dental operating theatres and 7 in the recovery room. The sterile dental theatre was used for surgical procedures during the mornings and the minor procedures theatre was used in the afternoons with the majority of patients being paediatric cases. A total of 11 personal samples were taken from anaesthetists in theatre and 6 from staff in the recovery room.
The sterile dental theatre and the minor procedures theatre were adjoining rooms and access to the recovery room was via the minor procedures theatre.
Prior to the monitoring, the level of ventilation and the provision of anaesthetic gas scavenging was established. Air change measurements were carried out in both theatres and the recovery room using method MDHS 7312 as approved by the Health and Safety Executive. At the time of monitoring there was limited gas scavenging in the theatre areas but no provision of anaesthetic gas scavenging in the recovery room, the number of air changes in these areas were marginally below the recommended minimum level of 15 air changes per hour.13 No air change measurements were carried out in the community clinics because of time and financial constraints.
Patients were anaesthetised in the operating theatre. During monitoring in the sterile theatre, nasal masks were used, anaesthetic induction was by a combination of intravenous drugs and nitrous oxide, although on a few occasions only gaseous inductions were carried out. A circle system was used in the sterile theatre where procedures had a longer duration.
In the minor procedures theatre laryngeal masks were used in general, although on a number of occasions a facemask was used if the procedure was expected to last a very short time. It was observed that, as with the sterile theatre, in general anaesthetic induction was carried out by a combination of intravenous drugs and anaesthetic gas, notably nitrous oxide.
Static monitoring was carried out using a Foxboro MIRAN infrared spectrophotometer. This was positioned in each of the areas to obtain a sample of the atmosphere in the working environment. Each monitoring session was carried out for the duration that nitrous oxide was being used.
Personal sampling was carried out using a Negretti personal sampling pump attached to a 1 litre gas reservoir bag, following the National Institute of Occupational Safety and Health (NIOSH) analytical method 6600.14 The pump and bag were attached to the dentist where relative analgesia was being used, and were on the anaesthetist where general anaesthesia was being undertaken.
A total of four static monitoring sessions were carried out in community dental clinics: one during a general anaesthetic list and three during conscious sedation. No active anaesthetic gas scavenging was present and no provision was made for increased ventilation in these areas. A total of six personal samples were taken during the monitoring in the community dental clinics: three during conscious sedation and three during general anaesthetics.
Results
In theatres for five of the sessions monitored the static TWA environmental level over the session period exceeded the OES level although this was only the case for one session in the recovery room (see Table 1). Only one personal sample exceeded the OES over the time monitored.
A community dental clinic (Clinic 1) was monitored which did not have increased ventilation or the provision of anaesthetic gas scavenging. During this time six patients were treated using conscious sedation. The TWA over 180 minutes was 277 ppm although a peak level of 700 ppm was recorded. Personal sampling was also carried out in the community dental clinic (Clinic 1). Three Personal samples were taken from the dentist carrying out the conscious sedation. The mean level was calculated at 19 ppm range (0–40ppm), see Table 4.
A clinic run within the dental hospital (Clinic 2) was also monitored on two occasions. This was an open plan treatment area with many different work areas. On the occasions monitored only one patient was being treated under conscious sedation, the first static monitoring result shows the TWA over 240 minutes as 81 ppm with a peak level of 650 ppm. The second result shows a TWA over 60 minutes of 20 ppm with a peak level of 42 ppm. No personal sampling was undertaken in this area.
In a community dental clinic (Clinic 3), run on a weekly basis within a hospital, during a general anaesthetic list with 13 children being treated. The peak static level recorded of nitrous oxide was 560 ppm, with an average level of 177 ppm being calculated over the duration of the 130 minutes monitoring time. Although this was a community clinic, it was based within a hospital environment and therefore an anaesthetist was present to administer the general anaesthetic. Personal samples from the anaesthetist show a mean level of 299 ppm range (20–734 ppm).
The results in both the dental theatres monitored and the community clinics show that high levels of nitrous oxide are present during the dental procedures being carried out. Personal exposure levels measured show that levels are high in the community clinic using general anaesthetics.
Discussion
The survey results indicate that staff exposures in operating theatres and community dental clinics were high on some occasions.
It was observed from the environmental monitoring undertaken, that where dental procedures are being carried out on children there is a very high throughput over a relatively short period of time. The procedures carried out often only require the child to be anaesthetised for very short periods of time. This leads to large amounts of waste anaesthetic gas being present in the working atmosphere of the operating theatre or treatment room.
This is highlighted in the results. Three different areas were monitored, the first being a sterile dental theatre where more lengthy procedures are carried out. This theatre is used during the morning with an average of 3 persons being treated. The afternoon operating session takes the form of a clinic with a high throughput of young children being anaesthetised for the removal of teeth. Up to 13 children may be treated in any one afternoon in this theatre.
It is recognised that the use of nitrous oxide in the operating theatre environment is in decline,15,16,17,18 however it is still very much used in the dental environment. The possibility of using xenon gas as an anaesthetic gas has also been suggested although its cost precludes its widespread use.19
A recent report by the General Dental Council20 highlighted that both conscious sedation and general anaesthesia must only be carried out in areas where suitable equipment and adequate facilities are available and only by an individual specifically trained to the requirements of the Royal College of Anaesthetists. Standards for General Anaesthesia in dentistry, published by the Royal College of Anaesthetists, questions the need for such a high proportion of procedures carried out under general anaesthesia in the UK21 but the report fails to point out the importance of controlling exposure to anaesthetic agents in areas carrying out general anaesthesia.
The provision of anaesthetic gas scavenging is an essential element in the control of exposure. The introduction of gas scavenging can reduce the amount of pollution from waste anaesthetic gases by as much as 90% or more.22,23 Where necessary anaesthetic gas scavenging should be installed to ensure that levels of waste anaesthetic are within the standards set by the Control of Substances Hazardous to Health (COSHH) Regulations 1999.24
Active anaesthetic gas scavenging systems should comply with the British Standard 6834: 1987 'Active Anaesthetic Gas Scavenging Systems'. There are specific pieces of equipment that deal with this issue in a dental environment, with manufacturers including McRea Engineering and Anaesthetic Pollution Control (APC). These systems are designed to provide effective scavenging in both a theatre and dental clinic environment.
It is also important that finances are made available to dental clinics to enable them to ensure that where anaesthetic agents are being used, adequate ventilation is provided.
Tests should also be carried out to ensure there is the correct number of air changes in the area using anaesthetic agents. Theatre and recovery areas should have 15 air changes per hour.6 Areas using conscious sedation should have a minimum of 6–7 air changes per hour.1 Where control measures are in place it is essential that they be maintained every 14 months in conjunction with the COSHH Regulations. It is also recommended that where nitrous oxide is being used that this also should be monitored on an annual basis.
References
Health Services Advisory Committee. Anaesthetic Agents: Controlling Exposure under COSHH 1995; HMSO 071 761043 8.
Shaw A D S, Morgan M . Nitrous oxide: time to stop laughing? Anaesthesia 1998; 53: 213–215.
Nunn J F, Chanarin I, Tanner A G, Owen E R T C . Megaloblastic bone marrow changes after repeated nitrous oxide anaesthesia. Br J Anaesthesia 1986; 58: 1469–1470.
Keeling P A, Rocke D A, Ninn J F, Monk S J, Lumb M J, Halsey M J . Folinic acid protection against nitrous oxide teratogenicity in the rat. Br J Anaesthesia Year 1986; 58: 528–534.
Cohen E N, Brown B W, Bruce D L, Cascorbi H F, Corbett T H, Jones T W, Whitcher C E . Occupational disease among operating room personnel: a national study. Anaesthesiology 1974; 41: 321–340.
Ericson H A, Kallen A J B . Hospitalisation for miscarriage and delivery outcome among Swedish nurses working in operating rooms 1973-1978. Anaesthesia Analgesia 1985; 64: 981–988.
Hemminki K, Kyyronen P, Lindbohn M L . Spontaneous abortions and malformations in the offspring of nurses exposed to anaesthetic gases, cytotoxic drugs and other potential hazards in hospitals based on registered information of outcome. J Epidemiol Community Health 1985; 39: 141–147.
Cohen E N, Brown B W, Wu M L et al. Occupational disease in dentistry and chronic exposure to trace anaesthetic gases. J Am Dent Assoc 1980; 101: 21–31.
Rowland A J, Baird D D, Shore D L, Weinberg C R, Savitz D A, Wilcox A J . Nitrous oxide and spontaneous abortion in female assistants. Am J Epidemiol 1995; 141: 531–8.
Ahlborg G J R, Axelsson G, Bodin L . Shift Work, Nitrous Oxide Exposure and Subfertility among Swedish midwives. Int J Epidemiol 1996; 25: 783–790.
Doll R, Peto R . Mortality among doctors in different occupations. Br Med J 1977; 1: 1433–6.
Methods for the Determination of Hazardous Substances: 73, measurement of air change rates in factories and offices. 1992. Croner's Handbook of Occupational Hygiene.
Health Building Note 26: Operating Department. NHS Estates 1991.
National Institute of Occupational Safety and Health. Analytical Method 6600.
Yagiela J A . Health Hazards and Nitrous Oxide: A time for reappraisal. Anaesthetic Progress 1991; 38: 1–11.
Donaldson D, Meechan J G . The hazards of chronic exposure to nitrous oxide: an update. Br Dent J 1995; 178: 95–100.
Worthington L M, Flynn P J, Strunin L . Death in the dental chair: an avoidable catastrophe? 1998; 80: 131–132.
Brownlie G S, Walters F J M . Should we still be using nitrous oxide? Current Anaesthesia and Critical Care 1994; 5: 109–114.
Dingley J, Ivanova-Stoilova T M, Grundler S, Wall T . Xenon: recent developments. Anaesthesia 1999; 54: 335–346.
General Dental Council. Maintaining Standards: Guidance to Dentists on Professional and Personal Conduct. 1997.
Royal College of Anaesthetists. Standards and Guidelines for General Anaesthesia for Dentistry. February 1999.
Gray W M . Scavenging equipment. Br J Anaesthesia 1985; 57: 543–549.
Oulton J L . Operating room venting of trace concentrations of inhalation anaesthetic agents. Can Med Assoc J Year 1977: 116: 1148–1151.
Control of Substances Hazardous to Health (COSHH) Regulations: Approved Codes of Practice. HMSO 1994.
Author information
Authors and Affiliations
Additional information
Refereed paper
Rights and permissions
About this article
Cite this article
Henderson, K., Matthews, I. Environmental monitoring of nitrous oxide during dental anaesthesia. Br Dent J 188, 617–619 (2000). https://doi.org/10.1038/sj.bdj.4800556
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/sj.bdj.4800556
This article is cited by
-
Nitrous oxide occupational exposure in conscious sedation procedures in dental ambulatories: a pilot retrospective observational study in an Italian pediatric hospital
BMC Anesthesiology (2019)
-
A study of the effectiveness of oral midazolam sedation for orthodontic extraction of permanent teeth in children: a prospective, randomised, controlled, crossover trial
British Dental Journal (2002)