Introduction

Intermittent catheterization (IC) is considered to be the gold standard for the management of Neurological Lower Urinary Tract Dysfunction (NLUTD).1, 2, 3, 4, 5 It is effective in patients with detrusor underactivity or acontractility and in patients with detrusor overactivity if the condition is controlled by pharmacological intervention, such as anticholinergic therapy.6 The relative merits of reusable intermittent catheters or single-use disposable catheters, the use of lubricants or which specific lubricant, and aseptic or clean technique continues to be debated.

Guttmann first introduced IC at the end of the Second World War.7 The technique was further adapted by Lapides in 1972,8 who promoted the technique of clean IC, where patients simply wash their hands before holding the catheter, prior to insertion. Currently, the clean technique remains the most common practice of insertion for IC users in the community,9, 10, 11 and some authors have suggested that using hydrophilic-coated catheters may reduce the rates of bacteriuria associated with the clean technique.12 Clean technique as defined by the EAU guidelines is ‘disposable or cleansed reusable catheters, genitals washed’.13 While clean IC remains common practice, concerns have been raised by studies that show complications associated with this method including, urethral trauma and recurrent urinary tract infections.11, 14, 15, 16, 17 The method of simple hand washing prior to insertion as used with the clean technique, may not be sufficient to remove bacteria in some individuals, leading to an increase of contamination.18

Sterile IC, as introduced by Guttmann and Frankel, significantly reduces the risk of contracting a Urinary Tract Infection (UTI) and/or bacteriuria related to clean IC.13, 19, 20, 21 However, previous authors suggest that sterile IC is not widely accepted by patients in the community, as it is considered expensive and impractical.11, 22 Aseptic technique has been proffered as an alternative that reduces the risks associated with clean IC and is also easier for patients to use than sterile IC. The EAU has recently published guidelines for IC stating aseptic technique should be considered the method of choice.6 This group defines aseptic technique as, ‘the catheters remain sterile, the genitals are disinfected, and disinfecting lubricant is used’. Some authors have raised concerns regarding the use of disinfecting lubricants, while other authors have declared that simple sterile lubrication is sufficient for aseptic technique.22, 23 This paper does not address that topic. However, a method has been developed to maintain the sterility of the catheter, which facilitates the use of aseptic technique by the patient. These catheters incorporate a ‘no-touch sleeve’ that enables the user to grip the catheter through the sleeve, avoiding the need to touch the catheter. Other manufacturers promote the no-touch method by adapting the product wrapping for use during handling. Although a number of clinical studies have been reported comparing results for clean and sterile methods of IC, the aseptic method has not yet been subjected to the same scrutiny; data from large randomized controlled trials testing the method have yet to be published. The authors recognized this gap in the evidence and therefore, initial testing by the following in vitro method was performed to prove the principle that the no-touch sleeve can prevent or reduce the potential contamination of the catheter.

Methods

The test protocol was designed, validated and conducted at an independent testing laboratory. Six different models of catheters from four different manufacturers were tested. Catheters A, B, and C are standard hydrophilic catheters from two different manufacturers. Catheter D is a hydrophilic catheter that can be handled, in part, through the wrapper. Catheters E and F are catheters utilizing a self-contained gel lubricant and permit insertion via a ‘no-touch’ sleeve that covers the entire body of the catheter (Catheter A, EasiCath (Coloplast, Humlebæk, Demark); Catheter B, SpeediCath (Coloplast, Humlebæk, Demark); Catheter C, LoFric (Astra Tech, Mölndal, Sweden); Catheter D, LoFric H2O (Astra Tech, Mölndal, Sweden); Catheter E, Advance (Hollister, Illinois, USA); Catheter F, VIALOG Mobile (Medical Service, Bad Liebenzell, Germany)). Prior to performing the microbial contamination study, an Adverse Substance Screening Test was performed to ensure that catheter materials themselves did not interfere with the bacteria recovery and culturing procedures. The prospective acceptance criterion was 70% recovery. The mean recovery of both test organisms for all six catheter samples was at least 82%; therefore the recovery method was considered valid for all catheter types.

Suspensions of test microorganisms Staphylococcus aureus (S. aureus) and Escherichia coli (E. coli) were prepared according to the laboratory's Standard Operating Procedure Preparation and Maintenance of Microorganism Assay Suspensions. The target suspension level was at least 5.0 × 105–5.0 × 106 colony forming units (CFU) per ml. A technician put on sterile, nonlatex powder-free gloves that were then contaminated on each hand with 0.25 ml of each of the suspensions (total 0.5 ml), equivalent to 1.0 × 105–1.0 × 106 CFU. The contaminated hands were then rubbed together to evenly distribute the inoculum over each glove. Each inoculum's level was verified immediately after the test procedure, by direct sampling of the prepared suspension.

The preparation and insertion procedure for each catheter system was simulated according to the IFU provided with the product; this resulted in a total handling time of 30 s. Where the catheter required lubricating with water that was not contained within the packaging (catheters A and B), sterile water was utilized in order to exclude any possible contamination from the water source. The test procedure was performed in triplicate for each catheter model. A fresh set of gloves were used for each test replicate. Another technician wearing sterile gloves and using sterile scissors cut off the first-half of the catheter (used for insertion into the urethra), dropping it into a sterile receptor. An agar contact plate sample was taken of each of the wrappers, immediately after handling, to verify that microbial translocation from the gloves was actively taking place. Catheters of each sample type that had not been handled in any way were used as negative controls.

The numbers of microorganisms present on the catheters was determined according to the laboratory's bioburden extraction and plate count procedure. Briefly, each test catheter was submerged in 100 ml extraction fluid. Samples were placed on a mechanical shaker for 15 min. Serial log dilutions (100–10−3) of 5 ml aliquots were placed in sterile Petri dishes. All dishes were poured with molten tryptic soy agar and allowed to harden. The plates were inverted and incubated at 30–35°C for a minimum of 48 h. At the conclusion of the incubation period, the number of CFU per plate was counted. Any microorganisms recovered on the test plates were verified as the challenge organism via morphology. The number of microorganisms recovered for each test catheter was calculated.

Results

The average number of microorganisms recovered from the catheters expressed in CFU is shown in Table 1.

Table 1 Microorganisms recovered from catheters

The data were analysed using two different statistical methods. The first is a nonparametric median test. The results of this test indicated that a statistically significant difference exists among the catheters for both S. aureus (P=0.0210), and E. coli (P=0.0078). For both S. aureus and E. coli, the median values observed for catheters A, B, C, and D were greater than or equal to the grand median, whereas the median values for catheters E and F were less than the grand median.

To obtain more insight, the data were normalized using a log10 transformation, entered into a general linear model, and subjected to an a posteriori comparison of the means to detect differences. The statistical method used was a Student Newman Kuels test where the ranked treatment mean values were divided into subsets which are consistent with the hypothesis of no difference (null hypothesis as tested).

Mean values sharing the same letter are considered statistically similar.

Discussion

The advent of IC has directly resulted in decreased mortality and morbidity associated with NLUTD.11, 24, 25 By enabling the maintenance of low intravesical pressure with frequent bladder emptying, IC remains an excellent option for patients' long-term bladder management. Clean technique was hailed as an advancement for IC, because it costs less and is more practicable for the user, compared to the sterile technique. However, infection remains a concern for those using IC,4, 9, 17, 26, 27, 28 and recently respected professional bodies, such as the EAU, have issued guidelines promoting aseptic technique over traditional clean technique. New products have been introduced recently that facilitate the use of aseptic technique, while maintaining ease of use of the product. These closed system products have been subjected to some initial testing and were found to reduce the number of positive cultures obtained from patients using IC, when compared to open systems.26, 29

The handling of catheters A, B, and C in this study was consistent with guidelines for the clean IC method; however, the use of sterile water, instead of regular tap water, to hydrate the hydrophilic catheters that did not provide water within the packaging (catheters A and B) removed a second source of possible contamination. As a result, this study may underestimate the total bioburden that occurs with those catheters that are commonly hydrated with tap water. The handling of catheter D followed the manufacturer's recommendation to use the product's packaging to limit direct contact between the hands and the catheter. However, it is important to note that there are design issues which may limit the ability of the typical patient to use the product in this manner. Ease of use was not addressed in this study.

The challenge organisms were consistently identified on the wrappers for all catheters, confirming that there was active transfer of microorganisms from the tester's hands, during manipulation by the testers that mimicked insertion. There was some variation between the microbial challenge levels for the different catheter types; this was due to performing the test over a number of days. In all cases, the inoculum level presented a significant microbial challenge and did not have an adverse impact on the results.

Conclusion

In laboratory tests, utilizing a handling technique consistent with each catheter's instructions for use, the number of CFUs for both S. aureus and E. coli found on catheters with a ‘no-touch sleeve’ was significantly lower than all the other products. Bacteria recovery from catheter D was less than that from the traditional hydrophilic catheters but still higher than the two with the true no-touch design. Since Catheter D has not been designed specifically to facilitate no-touch insertion, it may be considered more difficult to handle, particularly for patients with limited dexterity. The utilization of catheters providing a no-touch feature is consistent with the recommendations of the EAU for aseptic IC. More importantly, it may provide a long-term benefit to patients requiring IC by reducing the introduction of exogenous bacteria into the urinary tract.