Introduction

Plate-related complications following anterior cervical procedures for contiguous disc disease combined with spondylosis or ossification of the posterior longitudinal ligament demonstrating retrovertebral extension, include graft or plate extrusion, fractures, and pseudarthrosis.1,2,3,4,5,6,7,8,9,10,11,12 To address complications the dynamic ABC plates were developed allowing for up to 10 mm of cephalad and 10 mm of caudad migration, reducing stress shielding and allowing for graft settling. One level anterior cervical corpectomy with fusion (ACF) performed with autologous iliac crest (38 patients) or fibula allografts (4 patients) were performed utilizing 42 dynamic ABC plates. Fusion or graft and plate-related failures and pseudarthrosis were documented on serial postoperative dynamic X-ray and 2D-CT studies obtained 3, 6, and up to 12 months postoperatively.

Materials and methods

Dynamic plates

Dynamic ABC plates have a unique slotted design allowing up to 10 mm of cephalad and 10 mm of caudad plate migration. The screw heads lock into the slots but allow the plate to slide over the screw heads. Variable angle screws toggle up to 34°, facilitating screw application.

Dynamic plates were applied in 42 patients with two level disc disease, spondylosis, and/or ossification of the posterior longitudinal ligament with significant retrovertebral extension who required single level corpectomy with fusion (Figure 1) (Table 1). Patients averaged 44 years of age including 23 males and 19 females. The Nurick Myelopathy Scale was employed to grade preoperative and postoperative neurological status. This Scale includes Grades 0–V: Grade 0: radiculopathy, Grade I: mild myelopathy, Grade II: mild–moderate myelopathy, Grade III: moderate myelopathy, Grade IV: moderate–severe myelopathy, Grade V: severe myelopathy.4,5,6,7,8 Preoperatively, all patients' exhibited moderate myeloradiculopathy (average Nurick Grade 2.9), while postoperatively they all improved to a level of mild residual radiculopathy/mild myelopathy (average Nurick Grade of 0.4).

Figure 1
figure 1

Transaxial non-contrast CT examination obtained at the mid-C6 vertebral body level where OPLL nearly fills the left side of the spinal canal (double arrows)

Table 1 Dynamic plated patients undergoing single level ACF

Single level anterior corpectomy with fusion (ACF) in 42 patients employed iliac crest autografts in 38, three having fresh frozen fibula allografts implanted for morbid obesity and one for technical reasons; deep wound, shallow vertebral width (Table 2, Case 3). Nineteen patients weighted over 200 lbs, the average weight being 193 lbs (range 140–320). Iliac crest grafts averaged 36 mm in length; the plate was typically 46 mm long. Surgery averaged 3 h in duration for 22 patients. Patients were discharged an average of 3.5 days postoperatively, and were followed an average of 34 months.

Table 2 Four dynamic plate-related failures following 1 level ACF

Single level ACF included partial resection and partial perforation of the respective cephalad and caudad vertebral end plates with removal of the intervening discs and vertebral body; ie C5–C7. Non-reversed iliac crest autografts were applied in all patients except four who received fresh frozen fibula struts.

Fusion and plate-related complications were documented on static X-ray, dynamic X-rays, and CT studies (2-D in particular). Serial examinations were obtained from 2–6 months postoperatively, or until fusion or plate-related failures were confirmed at up to 12 months.5,13 Static X-ray and 2D-CT fusion criteria included the lack of bony lucency, and the presence of bony trabeculation, while dynamic films confirmed lack of active motion (<1 mm) between contiguous spinous processes.

Of note, 5 of 15 similar patients undergoing fixed-plated single level ACF developed plate and/or graft-related failures; one plate extrusion (6 weeks postoperatively), two pseudarthroses (6 months postoperatively), and two delayed iliac crest strut fractures (1 and 2 years postoperatively). Fixed-plated patients were followed an average of 44 months.

Results

Dynamic plate series

Fusion was confirmed on both dynamic radiographs and 2D-CT studies in 38 of 42 patients an average of 3.7 months postoperatively (Table 1) (Figures 2 and 3). The average postoperative cephalad plate migration was 5.3 mm (range 3–10 mm) and the average caudad migration was 5.5 mm (range 2–10 mm).

Figure 2
figure 2

Case 1: Lateral extension X-ray obtained 6 months postoperatively demonstrating classical cephalad migration (7 mm) and caudad migration (5 mm) (white arrows) following a C4–C6 ACF. The lack of lucency and presence of bony trabeculation at the graft/vertebral body interface was consistent with fusion (curved arrows)

Figure 3
figure 3

Case 2: Postoperative midline sagittal 2-D CT scan at 6 months confirmed adequate graft/fusion cephalad and caudad (long arrows), while confirming both cephalad and caudad migration of the plate. Note, superior extension of the plate impinging on the inferior aspect of C3 (short arrow)

Four (9.5%) of 42 patients exhibited postoperative complications (Tables 1 and 2). A morbidly obese male 2 pack/day smoker, developed a graft/plate extrusion 2 weeks following a C5–C7 ACF utilizing a fibula strut allograft (Table 2, Case 1). Fusion was demonstrated 3 months following a secondary ACF C5–T1 with C4–T1 posterior wiring and fusion (PWF) and halo application. Two patients had single level ACF performed with fibula strut allografts, one with morbid obesity, and the other, for a technically difficult C2–C4 ACF (Table 2, Case 3). Here, pseudarthroses occurred respectively 6 months and 1.5 years postoperatively necessitating secondary C4–T1 PWF with fusion 4 and 5 months later. The fourth patient had a C5–C7 ACF and demonstrated fusion on a 2D-CT scan 3 months postoperatively. He immediately returned to driving a truck against medical advice. Three months later, he presented with neck pain and a mid-iliac crest strut fracture (Table 2, Case 4) (Figure 4). He failed to fuse after wearing a Cervico-Thoracic Orthosis (CTO) for 6 months (at his request), required a secondary C4–T1 PWF, and fused 4 months following the secondary fusion (Figures 5 and 6).

Figure 4
figure 4

Case 3: Postoperative coronal CT at 6 months demonstrating a mid-lower spontaneous iliac crest strut fracture (arrow) following a C5–C7 ACF

Figure 5
figure 5

Case 3: Lateral plain radiograph obtained immediately following posterior wiring and fusion from C3–T1 addressing the spontaneous C5–C7 iliac crest strut fracture demonstrated in Figure 4. Note clear linear lucency in the mid-iliac crest strut graft anteriorly (arrow). Note the kyphotic deformity accompanying this construct, with superior impingement of the ABC plate on the anterior/inferior C4 vertebral body (arrows)

Figure 6
figure 6

Case 3: Postoperative 2-D CT study at 6 months obtained from the same Case 3 in Figures 4 and 5 revealing progressive fusion of the iliac crest strut fracture (white arrow). The posterior wiring and fusion construct performed with a fibula strut allograft (black arrow) individually wired to the spinous processes shows multiple autogenous bony fragments contained within an inductive conductive matrix. Multiplanar reconstruction demonstrates laterally fused fragments

Discussion

Failures associated with non-plated anterior cervical procedures prompted the application of different types of anterior cervical plates.1,2,3,4,5,6,7,8,9,10,11,12 Emery et al found a 16% pseudarthrosis rate following 1–2 level non-plated anterior diskectomy with fusion (ADF) or single level ACF.2 Epstein noted a 24% pseudarthrosis rate in non-plated 1–4 level ADF.14

The impact of fixed plates on single level ADF remains controversial.1,10,11,14,15,16,17,18,19 Caspar et al found a 9% incidence requiring secondary surgery in 210 patients with non-plated 1–2 level ADF, but only a 2% rate for those with 146 similarly plated procedures.16 Anterior cervical plates increased fusion rates for 1–2 level ADF up to 97.9%15 and 100%1 in some series, while decreasing pseudarthrosis, graft, and plate extrusion rates.11,16,18 Biomechanical data confirmed that single level ADF plates provided load-sharing rather than load-shielding and facilitated fusion.20

Dynamic plates appear to provide an added benefit for multilevel constructs.1,13 Saunders documented three plate/graft failures in 31 four level ACF.21 Using fixed plates for multilevel ACF alone, Vaccaro et al demonstrated a 9% failure rate for two level ACF, and 50% incidence for three level ACF.22 Performing 2–4 level ACF with a posterior tension band (posterior wiring and fusion), Epstein also found that three of 22 non-plated, and five of 38 fixed plated cases failed.6,7,8 In 26 dynamic plated multilevel ACF/PWF, only one failure was observed, resulting in a partial pseudarthrosis necessitating a secondary PWF.8

The 9.5% dynamic ABC failure rate following single level ACF in 42 patients appears high, but below our initial experience with fixed plates. The biomechanical advantages of dynamic plates include reduced stress shielding and greater graft compression. However, two factors which contributed significantly to increased morbidity include performing only anterior constructs in the morbidly obese and utilizing less effective fibula strut allografts. Construct failures would be reduced in the morbidly obese patient if simultaneous posterior fusions were performed. Di Angelo et al, looking at the in vitro biomechanics of load transfer through multilevel strut-grafts using a comparable single level ACF model, determined that multilevel anterior cervical plates reversed graft loads and excessively loaded anterior grafts in extension, promoting graft pistoning.23 Simultaneous application of a posterior tension band in obese individuals would presumably limit this reversal and thus reduce construct failures. The fibula allograft remains suboptimal in single level ACF when performed without a simultaneous posterior tension band. Contributory factors against utilizing allograft fibula include the shallow graft/vertebral-end plate base and the tendency for hard cortical fibula grafts to piston through the inferior vertebra. Also a longer period is required for such grafts to incorporate. Alternatively, iliac crest autografts with the combination of cortical and cancellous bone, provide a wider graft/vertebral base reducing the tendency for inferior pistoning and a more rapid incorporation. In morbidly obese patients, if donor site morbidity is to be avoided, a fibula strut allograft in association with a single level ACF should be accompanied by a PWF. The described advantages of dynamic plating systems should further contribute to an overall reduction in morbidity. Dynamic plate failures and their relationship to patient-related co-morbidities require further evaluation in a larger series of patients.