Abstract
Study Design: Forty-two consecutive patients undergoing dynamic ABC plated one-level ACF utilizing iliac crest autograft (38 patients) and fibula allografts (four patients) were evaluated. The unique ABC slotted plate design allows for up to 10 mm of cephalad and 10 mm of caudad plate migration.
Objectives: To evaluate the incidence and etiology of complications in forty-two patients undergoing anterior cervical dynamic ABC plating (Aesculap, Tuttlingen, Germany), during one level anterior corpectomy with fusion (ACF).
Setting: New York, USA.
Methods: Serial dynamic X-ray and 2 Dimensional CAT Scan (2D-CT) studies, obtained 3, 6, and up to 12 months postoperatively, in 42 patients documented the presence of fusion or complications including plate or graft extrusion or pseudarthrosis.
Results: Four (9.5%) of 42 patients developed postoperative plate or graft-related complications during the average follow up interval of 34 months. One patient, with a plate/graft extrusion, required a second two level ACF with posterior wiring and fusion (PWF). Two patients with pseudarthroses and one patient with a delayed iliac crest strut fracture required secondary PWF.
Conclusions: Effective arthrodesis and a low incidence of complications following one level ACF performed utilizing dynamic ABC plates were attributed to reduced stress shielding and greater graft compression afforded by the unique plate design. Applying dynamic ABC plates for one level ACF was biomechanically advantageous with low morbidity.
Similar content being viewed by others
Introduction
Plate-related complications following anterior cervical procedures for contiguous disc disease combined with spondylosis or ossification of the posterior longitudinal ligament demonstrating retrovertebral extension, include graft or plate extrusion, fractures, and pseudarthrosis.1,2,3,4,5,6,7,8,9,10,11,12 To address complications the dynamic ABC plates were developed allowing for up to 10 mm of cephalad and 10 mm of caudad migration, reducing stress shielding and allowing for graft settling. One level anterior cervical corpectomy with fusion (ACF) performed with autologous iliac crest (38 patients) or fibula allografts (4 patients) were performed utilizing 42 dynamic ABC plates. Fusion or graft and plate-related failures and pseudarthrosis were documented on serial postoperative dynamic X-ray and 2D-CT studies obtained 3, 6, and up to 12 months postoperatively.
Materials and methods
Dynamic plates
Dynamic ABC plates have a unique slotted design allowing up to 10 mm of cephalad and 10 mm of caudad plate migration. The screw heads lock into the slots but allow the plate to slide over the screw heads. Variable angle screws toggle up to 34°, facilitating screw application.
Dynamic plates were applied in 42 patients with two level disc disease, spondylosis, and/or ossification of the posterior longitudinal ligament with significant retrovertebral extension who required single level corpectomy with fusion (Figure 1) (Table 1). Patients averaged 44 years of age including 23 males and 19 females. The Nurick Myelopathy Scale was employed to grade preoperative and postoperative neurological status. This Scale includes Grades 0–V: Grade 0: radiculopathy, Grade I: mild myelopathy, Grade II: mild–moderate myelopathy, Grade III: moderate myelopathy, Grade IV: moderate–severe myelopathy, Grade V: severe myelopathy.4,5,6,7,8 Preoperatively, all patients' exhibited moderate myeloradiculopathy (average Nurick Grade 2.9), while postoperatively they all improved to a level of mild residual radiculopathy/mild myelopathy (average Nurick Grade of 0.4).
Single level anterior corpectomy with fusion (ACF) in 42 patients employed iliac crest autografts in 38, three having fresh frozen fibula allografts implanted for morbid obesity and one for technical reasons; deep wound, shallow vertebral width (Table 2, Case 3). Nineteen patients weighted over 200 lbs, the average weight being 193 lbs (range 140–320). Iliac crest grafts averaged 36 mm in length; the plate was typically 46 mm long. Surgery averaged 3 h in duration for 22 patients. Patients were discharged an average of 3.5 days postoperatively, and were followed an average of 34 months.
Single level ACF included partial resection and partial perforation of the respective cephalad and caudad vertebral end plates with removal of the intervening discs and vertebral body; ie C5–C7. Non-reversed iliac crest autografts were applied in all patients except four who received fresh frozen fibula struts.
Fusion and plate-related complications were documented on static X-ray, dynamic X-rays, and CT studies (2-D in particular). Serial examinations were obtained from 2–6 months postoperatively, or until fusion or plate-related failures were confirmed at up to 12 months.5,13 Static X-ray and 2D-CT fusion criteria included the lack of bony lucency, and the presence of bony trabeculation, while dynamic films confirmed lack of active motion (<1 mm) between contiguous spinous processes.
Of note, 5 of 15 similar patients undergoing fixed-plated single level ACF developed plate and/or graft-related failures; one plate extrusion (6 weeks postoperatively), two pseudarthroses (6 months postoperatively), and two delayed iliac crest strut fractures (1 and 2 years postoperatively). Fixed-plated patients were followed an average of 44 months.
Results
Dynamic plate series
Fusion was confirmed on both dynamic radiographs and 2D-CT studies in 38 of 42 patients an average of 3.7 months postoperatively (Table 1) (Figures 2 and 3). The average postoperative cephalad plate migration was 5.3 mm (range 3–10 mm) and the average caudad migration was 5.5 mm (range 2–10 mm).
Four (9.5%) of 42 patients exhibited postoperative complications (Tables 1 and 2). A morbidly obese male 2 pack/day smoker, developed a graft/plate extrusion 2 weeks following a C5–C7 ACF utilizing a fibula strut allograft (Table 2, Case 1). Fusion was demonstrated 3 months following a secondary ACF C5–T1 with C4–T1 posterior wiring and fusion (PWF) and halo application. Two patients had single level ACF performed with fibula strut allografts, one with morbid obesity, and the other, for a technically difficult C2–C4 ACF (Table 2, Case 3). Here, pseudarthroses occurred respectively 6 months and 1.5 years postoperatively necessitating secondary C4–T1 PWF with fusion 4 and 5 months later. The fourth patient had a C5–C7 ACF and demonstrated fusion on a 2D-CT scan 3 months postoperatively. He immediately returned to driving a truck against medical advice. Three months later, he presented with neck pain and a mid-iliac crest strut fracture (Table 2, Case 4) (Figure 4). He failed to fuse after wearing a Cervico-Thoracic Orthosis (CTO) for 6 months (at his request), required a secondary C4–T1 PWF, and fused 4 months following the secondary fusion (Figures 5 and 6).
Discussion
Failures associated with non-plated anterior cervical procedures prompted the application of different types of anterior cervical plates.1,2,3,4,5,6,7,8,9,10,11,12 Emery et al found a 16% pseudarthrosis rate following 1–2 level non-plated anterior diskectomy with fusion (ADF) or single level ACF.2 Epstein noted a 24% pseudarthrosis rate in non-plated 1–4 level ADF.14
The impact of fixed plates on single level ADF remains controversial.1,10,11,14,15,16,17,18,19 Caspar et al found a 9% incidence requiring secondary surgery in 210 patients with non-plated 1–2 level ADF, but only a 2% rate for those with 146 similarly plated procedures.16 Anterior cervical plates increased fusion rates for 1–2 level ADF up to 97.9%15 and 100%1 in some series, while decreasing pseudarthrosis, graft, and plate extrusion rates.11,16,18 Biomechanical data confirmed that single level ADF plates provided load-sharing rather than load-shielding and facilitated fusion.20
Dynamic plates appear to provide an added benefit for multilevel constructs.1,13 Saunders documented three plate/graft failures in 31 four level ACF.21 Using fixed plates for multilevel ACF alone, Vaccaro et al demonstrated a 9% failure rate for two level ACF, and 50% incidence for three level ACF.22 Performing 2–4 level ACF with a posterior tension band (posterior wiring and fusion), Epstein also found that three of 22 non-plated, and five of 38 fixed plated cases failed.6,7,8 In 26 dynamic plated multilevel ACF/PWF, only one failure was observed, resulting in a partial pseudarthrosis necessitating a secondary PWF.8
The 9.5% dynamic ABC failure rate following single level ACF in 42 patients appears high, but below our initial experience with fixed plates. The biomechanical advantages of dynamic plates include reduced stress shielding and greater graft compression. However, two factors which contributed significantly to increased morbidity include performing only anterior constructs in the morbidly obese and utilizing less effective fibula strut allografts. Construct failures would be reduced in the morbidly obese patient if simultaneous posterior fusions were performed. Di Angelo et al, looking at the in vitro biomechanics of load transfer through multilevel strut-grafts using a comparable single level ACF model, determined that multilevel anterior cervical plates reversed graft loads and excessively loaded anterior grafts in extension, promoting graft pistoning.23 Simultaneous application of a posterior tension band in obese individuals would presumably limit this reversal and thus reduce construct failures. The fibula allograft remains suboptimal in single level ACF when performed without a simultaneous posterior tension band. Contributory factors against utilizing allograft fibula include the shallow graft/vertebral-end plate base and the tendency for hard cortical fibula grafts to piston through the inferior vertebra. Also a longer period is required for such grafts to incorporate. Alternatively, iliac crest autografts with the combination of cortical and cancellous bone, provide a wider graft/vertebral base reducing the tendency for inferior pistoning and a more rapid incorporation. In morbidly obese patients, if donor site morbidity is to be avoided, a fibula strut allograft in association with a single level ACF should be accompanied by a PWF. The described advantages of dynamic plating systems should further contribute to an overall reduction in morbidity. Dynamic plate failures and their relationship to patient-related co-morbidities require further evaluation in a larger series of patients.
References
Connolly PJ, Esses SI & Kostuik JP . Anterior cervical fusion: outcome analysis of patients fused with and without anterior cervical plates. J Spinal Disord 1996; 9: 202–206.
Emery SE, Bohlman HH, Bolesta MJ & Jones PK . Anterior cervical decompression and arthrodesis for the treatment of cervical spondylotic myelopathy. Two to seventeen year follow up. J Bone Joint Surg (AM) 1998; 80: 941–951.
Epstein NE . Advanced cervical spondylosis with ossification into the posterior longitudinal ligament and resultant neurologic sequelae. J Spinal Disord 1996; 9: 477–484.
Epstein NE . Vertebral body fractures following extensive anterior cervical surgical procedures for ossification of the posterior longitudinal ligament. Neuro-Orthopaedics 1997; 21: 1–11.
Epstein NE . Evaluation and treatment of clinical instability associated with pseudarthrosis after anterior cervical surgery for ossification of the posterior longitudinal ligament. Surg Neurol 1998; 49: 246–252.
Epstein NE . Circumferential surgery for the management of ossification of the posterior longitudinal ligament. J Spinal Disord 1998; 11: 200–207.
Epstein NE . The value of anterior cervical plating in preventing vertebral fracture and graft extrusion following multilevel anterior cervical corpectomy with posterior wiring/fusion: indications, results, and complications. J Spinal Disord 2000; 13: 9–15.
Epstein NE . The efficacy of anterior dynamic plates in complex cervical surgery. J Spinal Disord 2002; 15: 221–227.
Grubb MR et al. Biomechanical evaluation of anterior cervical spine stabilization. Spine 1998; 23: 886–892.
MacDonald RL et al. Multilevel anterior cervical corpectomy and fibular allograft fusion for cervical myelopathy. J Neurosurg 1997; 86: 990–997.
McAfee PC et al. One-stage anterior cervical decompression and posterior stabilization. A study of one hundred patients with a minimum of two years follow-up. J Bone Jt Surg Am 1995; 77: 1791–1800.
Richman JD et al. Biomechanical evaluation of cervical spine stabilization methods using a porcine model. Spine (US) 1995; 20: 2192–2197.
Wang JC et al. The effect of cervical plating on single-level anterior cervical discectomy and fusion. J Spinal Disord (US) 1999; 12: 467–471.
Epstein NE . Anterior cervical diskectomy and fusion without plate instrumentation in 178 patients. J Spinal Disord 2000; 13: 1–8.
Bose B . Anterior cervical fusion using Caspar plating: analysis of results and review of the literature. Surg Neurol (US) 1998; 49: 25–31.
Caspar W, Geisler FH, Pitzen T & Johnson TA . Anterior cervical plate stabilization in one- and two-level degenerative disease: overtreatment or benefit? J Spinal Disord 1998; 11: 1–11.
Geisler FH, Caspar W, Pitzen T & Johnson TA . Reoperation in patients after anterior cervical plate stabilization in degenerative disease. Spine 1998; 23: 911–920.
Kostuik JP, Connolly PJ, Esses SI & Suh P . Anterior cervical plate fixation with the titanium hollow screw plate system. Spine 1993; 18: 1273–1278.
Lowery GL & McDonough RF . The significance of hardware failure in anterior cervical plate fixation. Patients with 2–7 year follow up. Spine 1998; 23: 181–186.
Rapoff AJ et al. Anterior cervical graft and plate load sharing. J Spinal Disord (US) 1999; 12: 45–49.
Saunders RL . Four-level cervical corpectomy. Spine (US) 1009; 23: 2455–2461.
Vaccaro AR et al. Early failure of long segment anterior cervical plate fixation. J Spinal Disord (US) 1998; 1195: 410–415.
Di Angelo DJ et al. Anterior cervical plating reverses load transfer through multilevel strut-grafts. Spine (US) 2000; 25: 783–795.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Epstein, N. Anterior cervical dynamic ABC plating with single level corpectomy and fusion in forty-two patients. Spinal Cord 41, 153–158 (2003). https://doi.org/10.1038/sj.sc.3101418
Published:
Issue Date:
DOI: https://doi.org/10.1038/sj.sc.3101418
Keywords
This article is cited by
-
Treatment of cervical OPLL by cervical anterior fusion using autologous vertebral bone grafts
Acta Neurochirurgica (2009)
-
4- and 5-level anterior fusions of the cervical spine: review of literature and clinical results
European Spine Journal (2007)
-
Dynamic anterior cervical plates for multilevel anterior corpectomy and fusion with simultaneous posterior wiring and fusion: efficacy and outcomes
Spinal Cord (2006)
-
Effect of constrained posterior screw and rod systems for primary stability: Biomechanical in vitro comparison of various instrumentations in a single-level corpectomy model
European Spine Journal (2005)