Abstract
PULSE-HEIGHT spectra from lithium-drifted p–i–n junctions in silicon, irradiated by γ-rays, have been published1. The photopeaks in these spectra are from about 1/20 to about 1/100 the height of the Compton distribution, rendering useful γ-ray spectroscopy difficult. The photoelectric cross-section varies as Z5, where Z is the atomic number of the absorbing element, the Compton cross-section varying as Z. Thus a high atomic number-absorbing medium will produce more useful γ-ray spectra.