Abstract
Accurate quantification of physical activity energy expenditure is a key part of the effort to understand disorders of energy metabolism. The Actiheart, a combined heart rate (HR) and movement sensor, is designed to assess physical activity in populations.
Objective:
To examine aspects of Actiheart reliability and validity in mechanical settings and during walking and running.
Methods:
In eight Actiheart units, technical reliability (coefficients of variation, CV) and validity for movement were assessed with sinusoid accelerations (0.1–20 m/s2) and for HR by simulated R-wave impulses (25–250 bpm). Agreement between Actiheart and ECG was determined during rest and treadmill locomotion (3.2–12.1 km/h). Walking and running intensity (in J/min/kg) was assessed with indirect calorimetry in 11 men and nine women (26–50 y, 20–29 kg/m2) and modelled from movement, HR, and movement+HR by multiple linear regression, adjusting for sex.
Results:
Median intrainstrument CV was 0.5 and 0.03% for movement and HR, respectively. Corresponding interinstrument CV values were 5.7 and 0.03% with some evidence of heteroscedasticity for movement. The linear relationship between movement and acceleration was strong (R2=0.99, P<0.001). Simulated R-waves were detected within 1 bpm from 30 to 250 bpm. The 95% limits of agreement between Actiheart and ECG were −4.2 to 4.3 bpm. Correlations with intensity were generally high (R2>0.84, P<0.001) but significantly highest when combining HR and movement (SEE<1 MET).
Conclusions:
The Actiheart is technically reliable and valid. Walking and running intensity may be estimated accurately but further studies are needed to assess validity in other activities and during free-living.
Sponsorship:
The study received financial support from the Wellcome Trust and SB was supported by a scholarship from Unilever, UK.
This is a preview of subscription content, access via your institution
Access options
Subscribe to this journal
Receive 12 print issues and online access
$259.00 per year
only $21.58 per issue
Buy this article
- Purchase on SpringerLink
- Instant access to full article PDF
Prices may be subject to local taxes which are calculated during checkout
Similar content being viewed by others
References
Ainslie P, Reilly T & Westerterp K (2003): Estimating human energy expenditure: a review of techniques with particular reference to doubly labelled water. Sports Med. 33, 683–698.
Albanes D, Conway JM, Taylor PR, Moe PW & Judd J (1990): Validation and comparison of eight physical activity questionnaires. Epidemiology 1, 65–71.
Alfonzo-Gonzalez G, Doucet E, Almeras N, Bouchard C & Tremblay A (2004): Estimation of daily energy needs with the FAO/WHO/UNU 1985 procedures in adults: comparison to whole-body indirect calorimetry measurements. Eur. J. Clin. Nutr. 58, 1125–1131.
Andrews RB (1971): Net heart rate as a substitute for respiratory calorimetry. Am. J. Clin. Nutr. 24, 1139–1147.
Atkinson G & Nevill AM (1998): Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med. 26, 217–238.
Avons P, Garthwaite P, Davies HL, Murgatroyd PR & James WP (1988): Approaches to estimating physical activity in the community: calorimetric validation of actometers and heart rate monitoring. Eur. J. Clin. Nutr. 42, 185–196.
Bassett Jr DR, Ainsworth BE, Swartz AM, Strath SJ, O'Brien WL & King GA (2000): Validity of four motion sensors in measuring moderate intensity physical activity. Med. Sci. Sports Exerc. 32, S471–S480.
Bland JM & Altman DG (1986): Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1, 307–310.
Boudet G & Chamoux A (2000): Heart rate monitors and abnormal heart rhythm detection. Arch. Physiol. Biochem. 108, 371–379.
Brage S, Brage N, Franks PW, Ekelund U, Wong MY, Andersen LB Froberg K & Wareham N (2004): Branched equation modeling of simultaneous accelerometry and heart rate monitoring improves estimate of directly measured physical activity energy expenditure. J. Appl. Physiol. 96, 343–351.
Brage S, Brage N, Wedderkopp N & Froberg K (2003a): Reliability and validity of the Computer Science and Applications accelerometer in a mechanical setting. Meas. Phys. Edu. Exerc. Sci. 7, 101–119.
Brage S, Wedderkopp N, Andersen LB & Froberg K (2003b): Influence of step frequency on movement intensity predictions with the CSA accelerometer: a field validation study in children. Ped. Exerc. Sci. 15, 277–287.
Brage S, Wedderkopp N, Franks PW, Andersen LB & Froberg K (2003c): Reexamination of validity and reliability of the CSA monitor in walking and running. Med Sci. Sports Exerc. 35, 1447–1454.
Cavagna GA, Franzetti P, Heglund NC & Willems P (1988): The determinants of the step frequency in running, trotting and hopping in man and other vertebrates. J. Physiol. (Lond) 399, 81–92.
Cavagna GA & Kaneko M (1977): Mechanical work and efficiency in level walking and running. J. Physiol. (Lond) 268, 647–681.
Cavagna GA, Thys H & Zamboni A (1976): The sources of external work in level walking and running. J. Physiol. (Lond) 262, 639–657.
Ceesay SM, Prentice AM, Day KC, Murgatroyd PR, Goldberg GR, Scott W & Spurr GB (1989): The use of heart rate monitoring in the estimation of energy expenditure: a validation study using indirect whole-body calorimetry. Br. J. Nutr. 61, 175–186.
Cole CR, Blackstone EH, Pashkow FJ, Snader CE & Lauer MS (1999): Heart-rate recovery immediately after exercise as a predictor of mortality. N. Engl. J. Med. 341, 1351–1357.
Ekelund U, Aman J & Westerterp K (2003): Is the ArteACC Index a valid indicator of free-living physical activity in adolescents? Obes. Res. 11, 793–801.
Freedson PS, Melanson E & Sirard J (1998): Calibration of the Computer Science and Applications, Inc. accelerometer. Med. Sci. Sports Exerc. 30, 777–781.
Godsen R, Carroll T & Stone S (1991): How well does Polar Vantage XL heart rate monitor estimate actual heart rate? Med. Sci. Sports Exerc. 23, S14.
Haskell WL, Yee MC, Evans A & Irby PJ (1993): Simultaneous measurement of heart rate and body motion to quantitate physical activity. Med. Sci. Sports Exerc. 25, 109–115.
Hausswirth C, Bigard AX & Le Chevalier JM (1997): The Cosmed K4 telemetry system as an accurate device for oxygen uptake measurements during exercise. Int. J. Sports Med. 18, 449–453.
Haymes EM & Byrnes WC (1993): Walking and running energy expenditure estimated by Caltrac and indirect calorimetry. Med. Sci. Sports Exerc. 25, 1365–1369.
Hendelman D, Miller K, Baggett C, Debold E & Freedson P (2000): Validity of accelerometry for the assessment of moderate intensity physical activity in the field. Med. Sci. Sports Exerc. 32, S442–S449.
Hiilloskorpi HK, Pasanen ME, Fogelholm MG, Laukkanen RM & Manttari AT (2003): Use of heart rate to predict energy expenditure from low to high activity levels. Int. J. Sports Med. 24, 332–336.
Jakicic JM, Winters C, Lagally K, Ho J, Robertson RJ & Wing RR (1999): The accuracy of the TriTrac-R3D accelerometer to estimate energy expenditure. Med. Sci. Sports Exerc. 31, 747–754.
Karvonen J, Chwalbinska-Monet J & Säynäjäkangas S (1984): Comparison of heart rates measured by ECG and microcomputer. Phys. Sports Med. 12, 65–69.
King GA, Torres N, Potter C, Brooks TJ & Coleman KJ (2004): Comparison of activity monitors to estimate energy cost of treadmill exercise. Med. Sci. Sports Exerc. 36, 1244–1251.
Kriska A (2000): Physical activity and the prevention of type 2 diabetes mellitus: how much for how long? Sports Med. 29, 147–151.
LaMonte MJ & Ainsworth BE (2001): Quantifying energy expenditure and physical activity in the context of dose response. Med. Sci. Sports Exerc. 33, S370–S378.
Leger L & Thivierge M (1988): Heart rate monitors: validity, stability, and functionality. Phys. Sports Med. 16, 143–151.
Li R, Deurenberg P & Hautvast JG (1993): A critical evaluation of heart rate monitoring to assess energy expenditure in individuals. Am. J. Clin. Nutr. 58, 602–607.
Livingstone MB, Prentice AM, Coward WA, Ceesay SM, Strain JJ, McKenna PG, Nevin GB, Barker ME & Hickey RJ (1990): Simultaneous measurement of free-living energy expenditure by the doubly labeled water method and heart-rate monitoring. Am. J. Clin. Nutr. 52, 59–65.
Luke A, Maki KC, Barkey N, Cooper R & McGee D (1997): Simultaneous monitoring of heart rate and motion to assess energy expenditure. Med. Sci. Sports Exerc. 29, 144–148.
Masse LC, Heesch KC, Fulton JE & Watson KB (2002): Raters' objectivity in using the compendium of physical activities to code physical activity diaries. Meas. Phys. Edu. Exerc. Sci. 6, 207–224.
Mazur W, Rivera JM, Khoury AF, Basu AG, Perez-Verdia A, Marks GF, Chang SM, Olmos L, Quinones MA & Zoghbi WA (2003): Prognostic value of exercise echocardiography: validation of a new risk index combining echocardiographic, treadmill, and exercise electrocardiographic parameters. J. Am. Soc. Echocardiogr. 16, 318–325.
McLaughlin JE, King GA, Howley ET, Bassett Jr DR & Ainsworth BE (2001): Validation of the COSMED K4 b2 portable metabolic system. Int. J. Sports Med. 22, 280–284.
Montoye HJ, Washburn R, Servais S, Ertl A, Webster JG & Nagle FJ (1983): Estimation of energy expenditure by a portable accelerometer. Med. Sci. Sports Exerc. 15, 403–407.
Moon JK & Butte NF (1996): Combined heart rate and activity improve estimates of oxygen consumption and carbon dioxide production rates. J. Appl. Physiol. 81, 1754–1761.
Mora S, Redberg RF, Cui Y, Whiteman MK, Flaws JA, Sharrett AR & Blumenthal RS (2003): Ability of exercise testing to predict cardiovascular and all-cause death in asymptomatic women: a 20-year follow-up of the lipid research clinics prevalence study. JAMA 290, 1600–1607.
Nichols JF, Morgan CG, Sarkin JA, Sallis JF & Calfas KJ (1999): Validity, reliability, and calibration of the Tritrac accelerometer as a measure of physical activity. Med. Sci. Sports Exerc. 31, 908–912.
Paffenbarger Jr RS, Brand RJ, Sholtz RI & Jung DL (1978): Energy expenditure, cigarette smoking, and blood pressure level as related to death from specific diseases. Am. J. Epidemiol. 108, 12–18.
Rennie K, Rowsell T, Jebb SA, Holburn D & Wareham NJ (2000): A combined heart rate and movement sensor: proof of concept and preliminary testing study. Eur. J. Clin. Nutr. 54, 409–414.
Rennie KL, Hennings SJ, Mitchell J & Wareham NJ (2001): Estimating energy expenditure by heart-rate monitoring without individual calibration. Med. Sci. Sports Exerc. 33, 939–945.
Rennie KL & Wareham NJ (1998): The validation of physical activity instruments for measuring energy expenditure: problems and pitfalls. Public Health Nutr. 1, 265–271.
Rowell LB (2004): Ideas about control of skeletal and cardiac muscle blood flow (1876–2003): cycles of revision and new vision. J. Appl. Physiol. 97, 384–392.
Saltin B, Lindgarde F, Houston M, Horlin R, Nygaard E & Gad P (1979): Physical training and glucose tolerance in middle-aged men with chemical diabetes. Diabetes 28 (Suppl 1), 30–32.
Seaward BL, Sleamaker RH, McAuliffe T & Clapp III JF (1990): The precision and accuracy of a portable heart rate monitor. Biomed. Instrum. Technol. 24, 37–41.
Sekioka K, Takaba H & Nakano T (1998): Parallel recording of physical activity on commercial Holter recorders. Front. Med. Biol. Eng. 8, 253–268.
Singel DJ & Stamler JS (2004): Blood traffic control. Nature 430, 297.
Speakman JR (1998): The history and theory of the doubly labeled water technique. Am. J. Clin. Nutr. 68, 932S–938S.
Strath SJ, Bassett Jr DR & Swartz AM (2003): Comparison of MTI accelerometer cut-points for predicting time spent in physical activity. Int. J. Sports Med. 24, 298–303.
Strath SJ, Bassett Jr DR, Thompson DL & Swartz AM (2002): Validity of the simultaneous heart rate-motion sensor technique for measuring energy expenditure. Med. Sci. Sports Exerc. 34, 888–894.
Strath SJ, Bassett Jr DR, Swartz AM & Thompson DL (2001): Simultaneous heart rate-motion sensor technique to estimate energy expenditure. Med. Sci. Sports Exerc. 33, 2118–2123.
Strath SJ, Swartz AM, Bassett DR, O'Brien WL, King GA & Ainsworth BE (2000): Evaluation of heart rate as a method for assessing moderate intensity physical activity. Med. Sci. Sports Exerc. 32, S465–S470.
Tanaka H, Monahan KD & Seals DR (2001): Age-predicted maximal heart rate revisited. J. Am. Coll. Cardiol. 37, 153–156.
Tate DF, Jackvony EH & Wing RR (2003): Effects of Internet behavioral counseling on weight loss in adults at risk for type 2 diabetes: a randomized trial. JAMA 289, 1833–1836.
Terbizan DJ, Dolezal BA & Albano C (2002): Validity of seven commercially available heart rate monitors. Meas. Phys. Edu. Exerc. Sci. 6, 243–247.
Treuth MS, Adolph AL & Butte NF (1998): Energy expenditure in children predicted from heart rate and activity calibrated against respiration calorimetry. Am. J. Physiol. 275, E12–E18.
Treuth MS, Schmitz K, Catellier DJ, McMurray RG, Murray DM, Almeida MJ, Going S, Norman JE & Pate R (2004): Defining accelerometer thresholds for activity intensities in adolescent girls. Med. Sci. Sports Exerc. 36, 1259–1266.
Uth N, Sorensen H, Overgaard K & Pedersen PK (2004): Estimation of VO2max from the ratio between HRmax and HRrest—the Heart Rate Ratio Method. Eur. J Appl. Physiol. 91, 111–115.
Wareham NJ, Jakes RW, Rennie KL, Mitchell J, Hennings S & Day NE (2002): Validity and repeatability of the EPIC-Norfolk Physical Activity Questionnaire. Int. J. Epidemiol. 31, 168–174.
Wareham NJ & Rennie KL (1998): The assessment of physical activity in individuals and populations: why try to be more precise about how physical activity is assessed? Int. J. Obes. Relat. Metab. Disord. 22 (Suppl 2), S30–S38.
Weir JB (1949): New methods for calculating metabolic rate with special reference to protein metabolism. J. Physiol. 109, 1–9.
Whipp BJ, Higgenbotham MB & Cobb FC (1996): Estimating exercise stroke volume from asymptotic oxygen pulse in humans. J. Appl. Physiol. 81, 2674–2679.
Wickstrom G & Bendix T (2000): The ‘Hawthorne effect’—what did the original Hawthorne studies actually show? Scand. J. Work Environ. Health 26, 363–367.
Acknowledgements
We like to thank the study participants for their time and effort. Mrs S Hennings, Mrs J Mitchell, Mrs S Emms, and Mrs E Lucia-Rolfe are acknowledged for assistance with data collection. The use of commercial names in this manuscript is for information purposes only and does not imply endorsement.
Author information
Authors and Affiliations
Corresponding author
Additional information
Guarantors: S Brage and NJ Wareham.
Contributors: SB and NJW designed the study and SB was responsible for data collection. SB and NB conducted the data analysis, the results of which were discussed and interpreted by all authors. SB drafted the manuscript, which was reviewed by all authors.
Conflict of interest: None to declare.
Rights and permissions
About this article
Cite this article
Brage, S., Brage, N., Franks, P. et al. Reliability and validity of the combined heart rate and movement sensor Actiheart. Eur J Clin Nutr 59, 561–570 (2005). https://doi.org/10.1038/sj.ejcn.1602118
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1038/sj.ejcn.1602118
Keywords
This article is cited by
-
Timing of physical activity in relation to liver fat content and insulin resistance
Diabetologia (2023)
-
Resistance to thyroid hormone induced tachycardia in RTHα syndrome
Nature Communications (2023)
-
Joint associations between objectively measured physical activity volume and intensity with body fatness: the Fenland study
International Journal of Obesity (2022)
-
Longitudinal cardio-respiratory fitness prediction through wearables in free-living environments
npj Digital Medicine (2022)
-
Intensity of occupational physical activity in blue-collar workers: do self-reported rating and device-worn measurements agree?
European Journal of Applied Physiology (2022)