Childhood asthma resembles a complex syndrome and is of major public health relevance because of its high prevalence (1,2,3). Exhaled nitric oxide (eNO) serves as a surrogate measure of allergic airway inflammation. It is correlated with the expression of inducible NO synthase in airways of atopics, especially individuals with atopic asthma (4,5). In both the clinical and the research settings, eNO aids in discriminating different asthma phenotypes, particularly the atopic vs. the nonatopic form (6,7). Increased in atopic or asthmatic patients and reduced under corticosteroid treatment, eNO has assisted with diagnosis and clinical guidance for pediatric and adult patients alike (8).

The single-breath online method for eNO measurement (eNOsb) is the gold standard for older children, adolescents, and adults according to guidelines by the European Respiratory Society (ERS) and the American Thoracic Society (ATS; ref. 9). Standardization is necessary for the collection of acceptable and comparable data both within one center and between several centers. For eNOsb measurements, standardization primarily involves the expiratory target flow of 50 ml/s, as eNO is inversely related to flow (9,10,11,12,13). Furthermore, standardization requires the following criteria: inhalation to total lung capacity during the inspiratory maneuver, inspired NO concentration of <5 parts per billion (ppb), exhalation time of ≥4 s for children <12 y and ≥6 s if older, duration of expiratory plateau at target flow of ≥2 s, expiratory pressure level of 5–20 cm H2O for velum closure, calibration, hardware, and measurement techniques (9). Performing tests successfully according to guidelines is strongly related to age (9,14,15). Children often find it demanding to perform acceptable measurements, especially at preschool age. Lack of cooperation and a relatively short attention span further complicate matters (9). A method based on quiet tidal breathing might be easier to perform satisfactorily. This is comparable to pulmonary function tests at this age. Here, measurements based on quiet tidal breathing usually result in higher proportions of children being able to perform the maneuver satisfactorily than spirometry as a forced expiratory maneuver, as illustrated in a review on lung function tests in preschool children with cystic fibrosis (16). Until now, multiple-breath eNO (eNOmb) measurements have mostly been performed in infants and toddlers (17,18,19,20,21). Despite requiring less cooperation beyond infancy, criteria for standardization of eNOmb measurements are not yet included in the current ERS/ATS guidelines due to methodological issues (9). Omitting the need for flow control, it may be possible to overcome this difficult hurdle during eNO measurements according to ERS/ATS standards in young children.

Within this article, we hypothesized that eNOmb might constitute an alternative to the current gold standard of eNOsb measurements at school age and set out to close the gap of missing evidence for (i) feasibility and quality control, (ii) correlation, and (iii) accuracy of eNOmb in comparison with eNOsb measurements. We did so in study participants of the Bern Infant Lung Development (BILD) cohort (20,22) who were enrolled in follow-up at the age of 5–6 y and, to also assess comparability at higher eNO values, in children around 11 y of age with suspected asthma presenting for routine work-up in the asthma outpatient clinic of the University Children’s Hospital in Bern, Switzerland.

Results

Feasibility and Quality Control of eNOmb Measurements as Compared With the Gold Standard of eNOsb Measurements

Online eNOmb and eNOsb measurements were performed in 120 children; 99 were BILD cohort study participants and 21 were children with suspected asthma. Table 1 displays demographic characteristics and results of eNO measurements separately for cohort study participants and subjects with suspected asthma.

Table 1 Demographic characteristics and eNO results of the whole study population

Figure 1 presents numbers of acceptable eNO measurements in the whole study population as well as separately for BILD cohort study participants and subjects with suspected asthma. Numbers of eNO measurements excluded from analyses due to quality control and ATS/ERS standards are also displayed (9). We obtained acceptable eNOmb measurements and eNOsb measurements in the whole study population in 82 (68.3%) and 81 (67.5%) children, respectively (P = 0.89). Among those with acceptable eNOmb measurements, 17 (20.7%) were unable to perform acceptable eNOsb measurements (n = 14 BILD cohort study participants). Among those able to perform eNOsb measurements according to standards, 16 (19.8%) were not able to acceptably perform eNOmb measurements (n = 13 BILD cohort study participants). For BILD cohort participants, we found a trend toward a higher proportion of boys (P = 0.07) and toward higher body weight (P = 0.10) among children unable to perform acceptable eNOsb measurements. The trend for a higher proportion of boys among BILD cohort study participants was also found for children not able to produce acceptable pairs of eNOmb and eNOsb measurements (P = 0.09). There were no other differences between children who were able or unable to perform each measurement in an acceptable way, or children who were deemed to have pairs of acceptable tests.

Figure 1
figure 1

Numbers of acceptable eNO measurements and of data sets excluded from analyses. The figure displays numbers (proportion) of acceptable eNOmb and eNOsb measurements in the whole study population and per BILD cohort participants and subjects with suspected asthma on the left. Numbers of individual measurements excluded from analyses are shown on the right together with the respective criteria. Specific numbers for each exclusion criterion are given in parentheses; due to overlap they can add up to numbers higher than total number excluded. BILD, Bern Infant Lung Development; eNO, exhaled nitric oxide; eNOmb, eNO derived from multiple-breath measurement; eNOsb, eNO derived from single-breath measurement; ppb, parts per billion.

Correlation of eNOmb Measurements With the Gold Standard of eNOsb Measurements

To compare the two techniques, we analyzed data pairs with both acceptable eNOmb and eNOsb measurements. Table 2 displays geometric means (SD) for ln-transformed results of acceptable eNOmb50 (eNOmb extrapolated to the standard flow of 50 ml/s) and eNOsb, as well as of NO output derived from multiple-breath measurements (V′NOmb) and NO output derived from single-breath measurements (V′NOsb) together with their respective intra- and between-measurement coefficients of variation. Both on a ln–ln and a log–log scale, online eNOmb50 values were correlated with results of eNOsb measurements. As displayed in Figure 2a , this correlation was significant. The same significant correlation was found when we analyzed V′NOmb compared with V′NOsb ( Figure 2b ). In contrast to eNOmb50, correlation of raw eNOmb results with eNOsb results on either a ln–ln or log–log scale was inferior (slope 0.75, P < 0.0001, r2 = 0.67).

Table 2 Results for pairs of acceptable eNOmb and eNOsb measurements (n = 65)
Figure 2
figure 2

Correlation between different eNO measurement techniques. The figure displays the correlation (a) between eNOsb and eNOmb50 (slope 0.90, r2 = 0.87, P < 0.0001) and (b) between V′NOsb and V′NOmb (slope 0.90, r2 = 0.857, P < 0.0001) on a log–log scale. In both a and b, gray dots indicate results from subjects with suspected asthma, black dots indicate results from participants of the BILD cohort, and lines represent regression lines. eNO, exhaled nitric oxide; eNOmb, eNO derived from multiple-breath measurement; eNOmb50, eNOmb additionally extrapolated to flow of 50 ml/s; eNOsb, eNO derived from single-breath measurement; ppb, parts per billion; V′NOmb, NO output derived from multiple-breath measurement; V′NOsb, NO output derived from single-breath measurement.

Accuracy of eNOmb Measurements as Compared With the Gold Standard of eNOsb Measurements

In a last analysis, we wanted to assess the accuracy of eNOmb50 as compared with the gold standard of eNOsb measurements. Figure 3a,b displays Bland–Altman plots for this comparison and of V′NOmb with V′NOsb, correspondingly. As differences between the two methods were normally distributed, these are presented as raw data on a linear scale. Means of either eNO or V′NO values were right-skewed and means of paired measurements are thus presented on a log-scale, however, without data transformation.

Figure 3
figure 3

Bland–Altman plots of different eNO measurement techniques. The figure displays Bland–Altman plots, (a) means of eNOmb50 and eNOsb per measurement on the x axis (log-scale) plotted against the differences of eNOsb–eNOmb50 per measurement on the y axis, and (b) means of V′NOmb and V′NOsb per measurement on the x axis (log-scale) plotted against the differences of V′NOsb–V′NOmb per measurement on the y axis. In both a and b, solid thick lines represent mean difference, dashed lines represent upper (mean difference + 1.96 SD of differences) and lower (mean difference – 1.96 SD of differences) limits of agreement, gray dots indicate results from subjects with suspected asthma, and black dots indicate results from participants of the BILD cohort. eNO, exhaled nitric oxide; eNOmb, eNO derived from multiple-breath measurement; eNOmb50, eNOmb additionally extrapolated to flow of 50 ml/s; eNOsb, eNO derived from single-breath measurement; ppb, parts per billion; V′NOmb, NO output derived from multiple-breath measurement; V′NOsb, NO output derived from single-breath measurement.

We found that eNOmb50 measurements systemically resulted in higher values as compared with eNOsb measurements with upper and lower limits of agreement (LOAs) of 4.0 and −5.3 ppb, respectively, and a total mean difference of −0.7 ppb across the whole data range. If means of paired eNO measurements were >20.0 ppb (n = 11), these numbers were 6.4 and −5.0 ppb, respectively, with a total mean difference of 0.7 ppb, and if restricted to means of paired eNO measurements of ≤20 ppb (n = 54), these numbers were 3.3 and −5.1 ppb, respectively, with a total mean difference of 0.9 ppb. Also V′NOmb measurements resulted in systemically higher values as compared with V′NOsb measurements. Across the whole data range, upper and lower LOAs were 12.8 and −17.0 nl/s with a total mean difference of −2.1 nl/s. If restricted to means of paired eNO measurements >20.0 ppb, these numbers were 21.7 and −16.0 nl/s with a total mean difference of 2.8 nl/s, and if means of paired eNO measurements were ≤20.0 ppb, these numbers were 10.2 and −16.4 nl/s with a total mean difference of −3.1 nl/s, respectively.

Upper and lower LOAs for the ratios of difference of paired eNO measurements over mean of paired eNO measurements for the whole data range were 52.7% and −65.4%, with a total mean of −6.3%. Restricted to means of paired eNO measurements >20.0 ppb, these numbers were 23.4% and −16.7% with a mean of 3.3%, and if means of paired eNO measurements were ≤20.0 ppb, these numbers were 55.3% and −71.9% with a mean of −8.3%.

Discussion

In summary, we found that eNOmb measurements are at least as feasible as eNOsb measurements with all data being collected under strict quality control and the latter measured according to currently available ERS/ATS standards (9). Results from eNOmb measurements were highly correlated with the gold standard within the range of assessed eNO values. Furthermore, irrespective of resulting in slightly higher eNO values and some outliers, eNOmb measurements demonstrated agreement with measurements using the single-breath technique. Thus, eNOmb measurements may constitute an alternative to the more difficult single-breath measurements at school age or around school entry, and especially if children are unable to perform eNOsb measurements according to current ERS/ATS guidelines due to a lack of cooperation.

The method used to measure eNOmb in this study is a tidal breathing technique with uncontrolled flow rate, thus not requiring continuous adaptation of exhalation resistance or automatic flow rate control. As yet, no standardized tidal breathing method to measure eNO in children exists that is recommended in current guidelines for the clinical setting (9). Our aim was to offer an alternative to eNOsb measurements, which are often difficult to perform within the acceptability criteria, especially in preschool children (9). Age-related difficulties were documented in several studies, such as by Napier et al. regarding eNO-related methodological issues in 61 children aged 4–6 y, younger than our study participants (15). These were further demonstrated in a study by Ito et al. in 165 asthmatic children, who, with a median age of 7 y, were older than the majority of our study population (14). Here, study participants who were naive with regard to eNO measurements displayed age-related success rates: those younger than 8 y were less successful than older children (14).

For lung function tests, acceptable performance is age-dependent and particularly low at preschool age. This is particularly the case if one compares proportions of acceptable data derived from spirometry with those from other pulmonary function tests based on quiet tidal breathing maneuvers. This is illustrated in a review of lung function tests in children with cystic fibrosis at preschool age and also depicted in the current ATS/ERS guidelines for pulmonary function testing in preschool children (16,23). As compared with techniques not based on forced expiratory maneuvers, the feasibility of spirometry is lowest for this age group, with the majority of studies demonstrating success rates of <50% for children aged 2 y and ≥50% for older children. In contrast, measurements based on tidal breathing (such as the forced-oscillation technique or multiple-breath washouts) revealed higher success rates of ≥80% after the age of 4 y (16). Hence, simplifying measurements is important for the clinical setting; however, it is necessary to do so without losing value in the diagnostic process. This underlines the importance and the need for easy-to-perform techniques in the age group of young children.

The single-breath method to measure eNO levels is well established for asthma but also for diseases other than asthma and in research (24,25). In this context, different devices for eNOsb measurements have been compared, also with regard to their applicability. Kalliola et al. compared the use of a portable device for eNOsb measurements to a stationary device in 55 children with a mean age of 5.7 y (26). That study revealed poor accuracy at low eNO levels and lower success rates (73%) for the portable device as compared with the stationary device (93%; ref. 26). Their proportion of successfully performed eNOsb measurements among children with a mean age comparable to the majority of our study population, the BILD cohort study participants, was generally higher than ours (67.5% for eNOsb). However, it remains unclear whether study participants in the study by Kalliola et al. were also naive to eNO measurements. Furthermore, their study was designed to compare two devices for eNOsb measurements. The already mentioned study by Ito et al. has compared different exhalation times (6 s vs. 10 s) for eNOsb measurements with the same portable device, with a duration of 6 s demonstrating better feasibility in children <8 y (92.0% vs. 60%) while still in accordance with guidelines (14). Their proportion of acceptable measurements in children naive to eNO measurements was again higher than ours; however, in the study by Ito et al. it remains unclear whether study participants were also naive to spirometry. There are studies that, similar to ours, compare the gold standard with a potentially easier eNO measurement technique (18,27). Buchvald et al. sampled eNO using an online tidal breathing method at a fixed flow rate (controlled eNO) and in mixed expired air in 16 patients with asthma 7–14 y old and 51 preschoolers (14 healthy, age 2–5 y) and compared these techniques with eNOsb among school-age children only. They could show that in preschoolers controlled eNO measurement was feasible and showed better agreement with eNOsb measurement than eNO measurement in mixed expired air among schoolchildren. However, the tidal breathing technique was different, the children in whom techniques were directly compared were older than ours, and it is unlikely that they were naive to eNO measurements or spirometry (18). Hadjikoumi et al. sampled eNO in mixed expired air in 25 healthy children, mean age 11.5 y, and 20 asthmatics, mean age 12 y, thus older than our study population (27). Success rates were not reported, neither technique demonstrated good correlation, and disease groups were less correctly identified by eNO in mixed expired air as compared with eNOsb measurements (27).

To the best of our knowledge, this is the first study that compared and assessed feasibility as well as quality control, correlation, and accuracy of online eNOmb with uncontrolled flow rate with the current gold standard at school age. For this, we used a very simple method to linearly extrapolate eNOmb results to the recommended target flow for eNOsb measurements. All data were collected in a setup with a number of methodological strengths. Measurements were always performed in the same order and with the same equipment. Thus, the identical careful approach for regular calibration was used (9). Furthermore, we also applied extensive quality control during measurements and during later analyses for comparison of the two techniques. Our eNOsb measurements completely adhere to the latest ATS/ERS standards (9). To compare the simpler eNOmb method with the gold standard, we adopted a conservative approach to assessing its acceptability. Furthermore, all study participants performed eNO measurements for the first time and were also naive with regard to spirometry.

With participants of the population-based BILD cohort study as the majority of the study population, the study may be biased toward a well-educated middle-class population (22). The children who were seen for the first time in our asthma outpatient clinic were included to also cover higher eNO levels in our analyses. They reflect a pediatric population referred to a major center for work-up of suspected asthma and were older than the BILD cohort study participants. However, the focus of our study was not to assess differences in feasibility of the two measurement techniques comparing healthy and children with suspected asthma or comparing different age groups or to assess factors influencing feasibility of assessed measurement techniques. For this, the study would have been underpowered anyway. Our aim was to compare two eNO measurement techniques with regard to their feasibility in direct comparison to each other, to assess their correlation, and the accuracy in relation to the current gold standard. The limitations of our study have to be kept in mind before our findings can be extrapolated to other populations. Furthermore, the simple linear method to extrapolate eNOmb results to the recommended eNOsb target flow may be an oversimplification of the complex mechanisms underlying different eNO levels in health and disease. More elaborate methods that account for the complexity upstream to what we measure as eNO have been published (10). As eNO is flow dependent in a nonlinear fashion, these models aim at modeling NO production in the lung and shedding light on factors influencing NO production dependent and independent of flow (10,13,28). However, they lack the simplicity and ease of use of the multiple-breath method with uncontrolled flow rate, especially for young children.

Conclusion

If online eNOmb measurements are performed adhering to strict quality control, they seem to be as feasible as the current gold standard of eNOsb measurements and to produce similar results. Despite their significant correlation, the wide range of LOAs and systematically slightly higher levels of eNOmb in comparison to eNOsb measurements hampers eNOmb use in research, where exact values across the whole range are warranted. Furthermore, other techniques such as measurements of the fraction of eNO in exhaled air collected in NO-inert bags (offline fraction of eNO measurement) have been shown to be suitable for use in research and clinical use, especially for infants (29). Multiple-breath eNO measurements are less dependent on cooperation and might constitute a promising alternative, particularly for younger children at preschool age, when lack of cooperation represents a crucial problem, especially in the clinical setting. Differences in eNO levels for discrimination between disease groups are greater by far than the differences between the different measurement techniques in our study, especially at higher eNO levels. However, before eNOmb measurements may be used in daily clinical practice or as an alternative technique at preschool age, further testing is required.

Methods

Study Population

Data were primarily collected in participants during follow-up of an ongoing prospective birth cohort of unselected, healthy subjects, the BILD cohort (22). For initial recruitment, the following inclusion criteria were applied: white ethnicity; term delivery; no severe maternal health problems; no maternal drug abuse other than nicotine; and no known major birth defects or perinatal disease of the newborn, such as respiratory distress, airway malformation, or other major respiratory diseases diagnosed after birth (22). To include higher eNO values in the analyses, subjects attending the pulmonary outpatient clinic of the University Children’s Hospital of Bern, Switzerland, for the first time due to suspected asthma were additionally recruited. Children were eligible only if they had never performed spirometry or either of the two eNO measurement techniques (eNOsb and eNOmb) before, except for the infant lung function measurements of participants of the BILD cohort (20). This was ascertained before the measurement by personal interview. Otherwise, exclusion criteria for eNO measurements according to ERS/ATS standards applied (9).

The study was approved by both the Ethics Committee of the Canton of Bern, Switzerland (Kantonale Ethikkommission Bern), and the Research Ethics Committee of University Hospital Bern, Switzerland (Inselspital). All caregivers provided written informed consent for this study.

NO Measurements

Measurements were always performed in the same order: we aimed for three eNOmb measurements followed by three acceptable eNOsb measurements. For both online eNOsb and eNOmb methods, eNO was measured with a rapid-response chemoluminescence analyzer (CLD 88 sp; EcoMedics, Duernten, Switzerland). Flow was recorded using an ultrasonic flow meter (Spiroson; EcoMedics). For signal collection, a software package (WBreath 3.28; ndd Medical Technologies, Zurich, Switzerland) was used and tidal flows and derived volumes were converted to body temperature and ambient pressure, and saturated with water vapor conditions. We measured eNOmb breath by breath during the third quartile of expiration and calculated mean eNO, correcting for expiratory flow (V′E) without flow control as described previously (20,30). For both eNOmb and eNOsb measurements, main outcome parameters were eNO and NO output (V′NO, eNO concentration multiplied by corresponding expiratory flow, V′E). Results for eNOmb were extrapolated to a flow of 50 ml/s (eNOmb50) to allow comparison with the gold standard using the following formula (rule of proportions):

Quality Control of NO Measurements

(i) We performed sensor calibration tests before each measurement. (ii) Contamination by ambient NO was avoided by always using NO-free air for inspiration in the case of both measurement techniques. (iii) For eNOmb measurement, we aimed for 20 quiet tidal breathing maneuvers. (iv) From each of the multiple-breath measurements, mean eNOmb measurement was calculated only from breaths that were within 10% agreement regarding volume. (v) Total mean eNOmb value was then calculated from only two to three measurements that agreed within 10% or from two within 5% (ref. 9). Online eNOsb was measured and calculated according to current ATS/ERS standards (9).

Statistical Analyses

Due to the eNO data being right-skewed, we ln-transformed results to normalize their distribution and calculated geometric means and their differences. We measured the correlation of eNOsb with eNOmb50 as well as of V′NOsb with V′NOmb by linear regression analyses to calculate r2 for log-transformed data to normalize their distribution. Results were identical after ln-transformation. Means of normally distributed data were compared by t-test, proportions were compared by the binomial test of proportion, and the Wilcoxon rank-sum test was used for nonnormally distributed data. A P value <0.05 was considered significant.

We calculated intra- and between-measurement coefficients of variation (%). Intra-measurement coefficients of variation for eNOmb were computed both for eNO and V′NO as the ratio of the standard deviation of acceptable breaths over the mean of the whole measurement. Intra-measurement coefficients of variation for eNOsb could be calculated for eNO only as the standard deviation over the mean eNO during the expiratory plateau at target flow.

Accuracy of raw results from eNOmb50 as compared with eNOsb was determined using the Bland–Altman method by plotting differences of paired measurements against means of paired measurements and by calculating upper and lower LOAs as mean difference of paired measurements ± SD of difference of paired measurements (31). We further calculated upper and lower LOAs for the ratios of difference of paired measurements over mean of paired measurements (%). These analyses were performed over the whole range of values and additionally for means of paired eNO measurements >20.0 and ≤20.0 ppb, separately. Descriptive statistics and linear regression analyses were performed using STATA 11 for Windows (STATA Corporation, College Station, TX).

Statement of Financial Support

This work was supported by the Swiss National Science Foundation (grant no. 3200-B0-12099 to O.F, P.L., N.P., and U.F.), the European Respiratory Society (Long-Term Research Fellowship 675 to O.F.), and the Austrian, German and Swiss Pediatric Respiratory Society (Training Scholarship 2009 to O.F.).