Abstract
The measurement of the microwave field is crucial for many developments in microwave technology and related applications. However, measuring microwave fields with high sensitivity and spatial resolution under ambient conditions remains elusive. In this work, we propose and experimentally demonstrate a scheme to measure both the strength and orientation of the microwave magnetic field by utilizing the quantum coherent dynamics of nitrogen vacancy centres in diamond. An angular resolution of 5.7 mrad and a sensitivity of 1.0 μT Hz−1/2 are achieved at a microwave frequency of 2.6000 GHz, and the microwave magnetic field vectors generated by a copper wire are precisely reconstructed. The solid-state microwave magnetometry with high resolution and wide frequency range that can work under ambient conditions proposed here enables unique potential applications over other state-of-art microwave magnetometry.
Similar content being viewed by others
Introduction
Recent advantages in microwave (MW) technology have led to tremendous developments in communications, high-speed electronics and magnetic resonance. Most of these developments rely on the discovery of nanoscale, high-frequency MW processes1,2,3 and materials4,5,6. Measurement of MW electromagnetic field with high spatial resolution is crucial but has been a long-standing problem. Measurements with high sensitivity and angular resolution have been achieved for MW electric field7,8. It is not so for the measurement of MW magnetic field vector, despite its importance, for example, in MW antenna test9 and the study of the MW response of materials to the magnetic field5. Although various approaches, such as coplanar-waveguide-type probe10,11,12,13,14, superconducting quantum interference devices15, cold atoms16,17 or spin hall effect films18, have been applied to measure the MW magnetic field, the sensor size, the extreme measurement conditions required or the lack of vector measurement ability strongly limit their applications.
In this work, we demonstrate a high-resolution vectorial MW magnetometry based on nitrogen vacancy (NV) centres in diamond. The NV centre is a defect that consists of a substitutional nitrogen atom and an adjacent vacancy in the carbon lattice of diamond. The ground spin triplet state of negatively charged NV centres can be initialized and measured by laser illumination and fluorescence intensity under laser illumination, respectively. Owing to its good spin properties, the NV centre has been shown to be a perfect solid-state quantum system for sensing magnetic fields with nanoscale resolution and high sensitivity under ambient conditions19,20. In previous works, NV centres have been utilized in the measurement of oscillating magnetic fields with frequencies ranging from kilohertz19 to megahertz21,22. Here we propose and then demonstrate experimentally a different scheme based on NV centres to realize the measurement of the vector of the MW magnetic field. The angular resolution and sensitivity are then analysed.
Results
Principles of vector MW magnetometry
The main idea of the MW magnetometry is based on the Rabi oscillation of a solid-state spin driven under resonant MW magnetic field. The Hamiltonian of the NV’s electron spin is H0=ΔSz2+γB0Sz+γ Bmw(t)·S, where Δ=2.87 GHz is the zero field splitting, γ=2.8025 MHz Gauss−1 is the gyromagnetic ratio of the electron spin, S and Sz are the associated spin operator and its z component, and Bmw(t)=Bmw·cos(2πft) is the MW magnetic field. We selectively address the spin operation on transition of ms=0↔−1 (denoted as |0› and |−1›, respectively) with a transition frequency of ω0=Δ−γB0 (Fig. 1b). Then the system can be treated similarly as spin 1/2. The resonance condition ω0=f can be fulfilled by adjusting B0. The component of the vector Bmw that perpendicular to Sz, denoted as Bmwp, can be treated as the sum of two circular polarized field with opposite rotating directions. The left-hand rotating circular polarized field B1 with an amplitude of directly drives the spin rotation (Fig. 1c). The Hamiltonian H0 can be simplified in the rotation frame as Hr=γB1Sx (Fig. 1d), which results in the spin state oscillating between |0› and |−1› with Rabi frequency Ωr=γB1. In the Bloch sphere representation, the spin state vector rotates around B1 by an angle of α=2πγB1τ after a sensing time τ.
Decoherence of the NV spin states needs to be taken into account in an actual case. After considering the interaction between NV and the 14N nuclear spin and the noise due to 13C nuclear spin bath, we have the whole Hamiltonian H=H0+S·A·I +QIz2+bSz, where I is the 14N nuclear spin operator, A is the hyperfine coupling tensor and b denotes the effective fluctuation magnetic field of the noise23,24. There are two factors need to be considered. One is the off resonance due to the hyperfine interaction with the nitrogen nuclear spin. The probability on |0› under off resonance MW driving is written as25
where is the Rabi frequency under off-resonance frequency detuning δω=ω−f. In existence of 14N nuclear spin, the electron spin resonance transition line splits into three peaks with frequencies f+mIA, where mI=0,±1 denotes the spin state of 14N nuclear spin (Fig. 2a). As a result, there are two frequencies Ωfast and Ωslow in the Rabi oscillation curve fulfilling the relation of Ωfast2=Ωslow2+A2. Ωslow is the Rabi frequency of the on-resonance peak and indicates B1 value. The other factor is the inhomogeneous line broadening effect caused by the spin bath noise that dominates the signal decay. The broadened line shape can be characterized by adding a Gaussian type noise to the electron spin23,24,25:
where kI is the relative height of each peak and T2* is the electron spin dephasing time. The final signal can be obtained from integration of all the off-resonance contributions in the three peaks:
The magnetic field vector can be reconstructed by making use of NV centres with four different orientations (Fig. 3a), [111], [–111], [1–11], [–1–11] (labelled as 1, 2, 3, 4). The static magnetic field can be aligned subsequently to each of the four orientations, and then the set of NV centres of the orientation is selectively driven and its Rabi oscillation is measured to deduce the relevant projected MW magnetic field (i=1, 2, 3 and 4), which is related to Bmwp, as shown in Fig. 1c, by
The four relations in equation (4) are over-complete but theoretically consistent in determination all the three components of Bmw=(Bx, By, Bz). Actually, any three of them should be sufficient, except for the exceptionally few singular points such as Bmwp being along [100], [010] or [001] (Supplementary Note 1). Hence the sensor can be a nanosized diamond with three or more appropriate NV centres. For the sample with NV centres of all the four orientations used here, we can make use of the over-completeness by applying Maximum-likelihood estimation method to the data to obtain the best estimation of the magnetic vector. The likelihood can be written as
Maximizing ℓ will find the optimum value of the vector according to Bayes’ theorem (Supplementary Note 4). The MW magnetic field can be uniquely determined in this way.
Experimental demonstration of the magnetometry
The probe head of the setup is shown in Fig. 1a. For the ease of demonstration, we use a chemical vapour deposition synthesized, type IIa diamond plate of 50 μm in thickness from Element Six, rather than a nanodiamond that contains NV centres as the magnetometer. The average distance between each NV centre is less than 100 nm, and there are ~20 NV centres in the laser spot. The static magnetic field B0 is aligned to the axis of each NV centre group to avoid ground-state mixing. The MW electronic circuits are designed to be 50 Ω impedance and the total power reflection is reduced to S11<−10 dB, so that a highly linearly polarized MW magnetic field is generated. As the relative permeability of diamond crystal is ~1.0000, the magnetic field inside the diamond should be the same as that outside of it. In our demonstration, we fix the applied static field to B0≈94.34 Gauss, and the corresponding frequency for the MW field is f=2.6000 GHz.
The pulse sequence and the measured Rabi oscillation results are shown in Fig. 2b. Figure 2c shows an example of Rabi oscillations under the MW source power of 0.13 mW. By fitting the curve, we derive the projected MW magnetic field strength. Bmwp=(2.106±0.002) × 10−4 T (see Supplementary Note 2 and 3 for details). The fitting result is confirmed by the Fourier transformation of the data in the frequency domain (Fig. 2d). By varying the MW source power, we find that proportionality of the projected MW magnetic field strength deduced from the measured Rabi oscillation to the square root of the MW source power holds in a wide Bmwp range (Fig. 2e and Supplementary Fig. 1).
Further demonstration to show the angular resolution was performed by using the vector MW magnetometry to measure in a series of space locations the strength and orientation of the linearly polarized MW magnetic field generated by a copper wire of ~5 mm in length. The copper wire is much longer than the distance between the copper wire and the diamond surface, and so can be treated as indefinitely in length. The MW magnetic near field has a magnitude inversely proportional to the distance to the wire and an orientation along the tangent as Ampere’s circuital law. We scan the laser spot in a line to measure Rabi Oscillations. The scanning line is divided into 150 points with a step of ~130 nm and several micrometres below the diamond surface (Fig. 3b). By controlling the orientation of the static magnetic field, we measure the Rabi oscillations of the NV centres of all the four groups in each point. In prior, the diamond is put at an angle of 0.37 rad with the copper wire to avoid the singular point (Supplementary Fig. 2). The experiment results are shown in Fig. 3c (see Supplementary Table 1 for the fitting parameters). All the three components of the vector Bmw are then deduced by utilizing maximum-likelihood estimation method, and so the orientation of the vector is obtained (Supplementary Note 5). The orientation of Bmw versus position is shown in Fig. 3d, together with theoretically predicted lines. The experiment result of ϕ and θ perfectly matches the theoretical predictions of the MW magnetic field. The minimum angular resolution achieved is δω=5.7 mRad, which can be further improved by increasing the detecting time or photon collection efficiency26,27.
Discussion
In conclusion, we have proposed a vector MW magnetometry scheme using NV centres in diamonds with high sensitivity and spatial resolution and under ambient conditions, and then carried out experimental demonstration of the amplitude and angular sensitivity with the measurement of the MW magnetic field vector generated by an infinitely long copper wire. In our demonstration, the spatial resolution is only diffraction limited, which gives a sensor size of ~230 nm (see Methods section). The capability of the vector MW magnetometry based on the NV centre is not limited by what we demonstrate here. By adopting wide-field imaging technology, a two-dimensional reconstruction of the MW magnetic field can be realized28,29. Based on solid-state spins of NV centre in diamond, the sensor size can be reduced to ten nanometre scale by a well-treated nanodiamond30. Also the working frequency can be extended to sub-terahertz band continuously by an external static field 8.6 T (ref. 31) and reach terahertz under 35 T, which is available in the current technology.
Our solid-state MW vector magnetometry will directly find its application in testing the MW electronics and materials. Owing to long-term stability of the NV centre, it can also serve for the purpose to stabilize the amplitude of the MW. Moreover, by combining the NV centres with a scanning probe microscopy, the NV centre-based scanning near-field vector MW magnetometry can be used in studying antenna radiation as a non-destructive measurement9. Finally, as the sensor can be downsized to nanometre scale, our scheme may promote the ferromagnetic resonance imaging32 to the nanoscale.
Methods
Experiment setup
Our setup is based on our homebuilt confocal microscopy. A temperature-stabilized diode laser (CNI MGL-III-532) is used for the initialization and readout of quantum states. The laser beam passes through a double-pass acoustic optical modulator (Crystal Technology 3200-121) to reach an isolation of about 60 dB. Then it enters a single-mode fibre acting as a spatial filter. A high numerical aperture (NA) objective lens (Olympus UPLSAPO 100XO) is used for the laser focusing and fluorescence collection. A high-resolution scanner (Asylum research) and fast steering mirror (Newport FSM-300) is used for the laser scanning. The magnetic field is applied by a permanent magnet on a three-dimensional translation stage. The fluorescence passes through a band-pass filter with 650–775 nm and 20 μm pinhole to the APD (Perkin Elemer SPCM-AQRH-14), which limit the optical detection size to be diffraction limited. The MW generated by a MW source (Agilent N5181B) is cut to square wave pulse by an MW PIN switch (CMCS0947A-C1, 3 ns rising time). An amplifier (Mini-circuits ZHL-16 W-43+) is used to amplify the MW signal.
Thermal drift
The experimental measurements in this demonstration last for a month, so the long-term stability of the whole setup is crucial. First, we put the whole probe head in a chamber with temperature stability of ±50 mK to reduce mechanical thermal drift, which enables us to focus on the same point for a long time without any drift. Second, we put some diamond nanocrystal with NV centres on the diamond surface for locating and carrying out position feedback. Finally, the most important thing is to keep all the MW circuit stable to ±0.5 K and keep a constant output power, so that the MW magnetic field amplitude remains stable during the measurement.
Sensor size and spatial resolution
In our demonstration, the diamond sample is rather big, and the spatial resolution is the pinhole size divided by the magnification of the objective lens, with the low limit being the Abbe diffraction limit, which can be written as:
With our band-pass filter wavelength, λ~650 nm, we obtain d~230 nm. However, a well-selected nanosized diamond could be used to reach a nanoscale spatial resolution.
Additional information
How to cite this article: Wang, P. et al. High-resolution vector microwave magnetometry based on solid-state spins in diamond. Nat. Commun. 6:6631 doi: 10.1038/ncomms7631 (2015).
References
Hen-Sum Philip, W., Frank, D. J., Solomon, P. M., Wann, C. H. J. & Welser, J. J. Nanoscale CMOS. Proc. IEEE 87, 537–570 (1999) .
Bernstein, K. et al. High-performance cmos variability in the 65-nm regime and beyond. IBM J. Res. Dev. 50, 433–449 (2006) .
Calhoun, B. H. et al. Digital circuit design challenges and opportunities in the era of nanoscale cmos. Proc. IEEE 96, 343–365 (2008) .
Duan, X., Huang, Y., Cui, Y., Wang, J. & Lieber, C. M. Indium phosphide nanowires as building blocks for nanoscale electronic and optoelectronic devices. Nature 409, 66–69 (2001) .
Yen, T. J. et al. Terahertz magnetic response from artificial materials. Science 303, 1494–1496 (2004) .
Schurig, D. et al. Metamaterial electromagnetic cloak at microwave frequencies. Science 314, 977–980 (2006) .
Sedlacek, J. A. et al. Microwave electrometry with Rydberg atoms in a vapour cell using bright atomic resonances. Nat. Phys. 8, 819–824 (2012) .
Sedlacek, J. A., Schwettmann, A., Kbler, H. & Shaffer, J. P. Atom-based vector microwave electrometry using rubidium rydberg atoms in a vapor cell. Phys. Rev. Lett. 111, 063001 (2013) .
Yaghjian, A. D. An overview of near-field antenna measurements. IEEE Trans. Antennas Propag. 34, 30–45 (1986) .
Serov, R., Richardson, M. C. & Burtyn, P. Probe for the measurement of magnetic fields with sub-nanosecond resolution. Rev. Sci. Instrum. 46, 886–888 (1975) .
Gao, Y. & Wolff, I. A new miniature magnetic field probe for measuring three-dimensional fields in planar high-frequency circuits. IEEE Trans. Microw. Theor. 44, 911–918 (1996) .
Kohei, S., Chun-Ping, C., Zhewang, M., Anada, T. & Thomas, D. W. P. Non-contacting electric and magnetic field probe for measuring EM fields on microwave planar circuits. In Microwave Conference, 2007. APMC 2007. Asia-Pacific 1–4IEEE (2007) .
Yingjie, G. & Wolff, I. A miniature magnetic field probe for measuring fields in planar high-frequency circuits. In IEEE MTT-S 1159–1162 (1995) .
Dubois, T. et al. Near-field electromagnetic characterization and perturbation of logic circuits. IEEE Trans. Instrum. Meas. 57, 2398–2404 (2008) .
Black, R. C. et al. Imaging radio frequency fields using a scanning squid microscope. Appl. Phys. Lett. 66, 1267–1269 (1995) .
Böhi, P., Riedel, M. F., Hänsch, T. W. & Treutlein, P. Imaging of microwave fields using ultracold atoms. Appl. Phys. Lett. 97, 051101 (2010) .
Ockeloen, C. F., Schmied, R., Riedel, M. F. & Treutlein, P. Quantum metrology with a scanning probe atom interferometer. Phys. Rev. Lett. 111, 143001 (2013) .
Vlaminck, V. et al. Mapping microwave field distributions via the spin hall effect. Appl. Phys. Lett. 101, 252406 (2012) .
Maze, J. R. et al. Nanoscale magnetic sensing with an individual electronic spin in diamond. Nature 455, 644–647 (2008) .
Balasubramanian, G. et al. Nanoscale imaging magnetometry with diamond spins under ambient conditions. Nature 455, 648–651 (2008) .
de Lange, G. et al. Single-spin magnetometry with multipulse sensing sequences. Phys. Rev. Lett. 106, 080802 (2011) .
Loretz, M., Rosskopf, T. & Degen, C. L. Radio–frequency magnetometry using a single electron spin. Phys. Rev. Lett. 110, 017602 (2013) .
Hanson, R., Dobrovitski, V. V., Feiguin, A. E., Gywat, O. & Awschalom, D. D. Coherent dynamics of a single spin interacting with an adjustable spin bath. Science 330, 60–63 (2008) .
Huang, P. et al. Observation of an anomalous decoherence effect in a quantum bath at room temperature. Nat. Commun. 2, 570 (2011) .
Schweiger, A. & Jeschke, G. Principles of Pulse Electron Paramagnetic Resonance Oxford Univ. (2001) .
Maletinsky, P. et al. A robust scanning diamond sensor for nanoscale imaging with single nitrogen-vacancy centres. Nat. Nanotech. 7, 320–324 (2012) .
Hadden, J. P. et al. Strongly enhanced photon collection from diamond defect centers under microfabricated integrated solid immersion lenses. Appl. Phys. Lett. 97, 241901 (2010) .
Steinert, S. et al. High sensitivity magnetic imaging using an array of spins in diamond. Rev. Sci. Instrum. 81, 043705 (2010) .
Pham, L. M. et al. Magnetic field imaging with nitrogen-vacancy ensembles. New J. Phys. 13, 045021 (2011) .
Chang, Y.-R. et al. Mass production and dynamic imaging of fluorescent nanodiamonds. Nat. Nanotech. 3, 284–288 (2008) .
Takahashi, S., Hanson, R., van Tol, J., Sherwin, M. S. & Awschalom, D. D. Quenching spin decoherence in diamond through spin bath polarization. Phys. Rev. Lett. 101, 047601 (2008) .
Lee, I. et al. Nanoscale scanning probe ferromagnetic resonance imaging using localized modes. Nature 466, 845–848 (2010) .
Acknowledgements
We thank Professor Fei Xue and Professor Xidong Wu for useful discussions. This work was supported by the National Key Basic Research Program of China (Grant No. 2013CB921800), the National Natural Science Foundation of China (Grant Nos.11227901, 31470835, 91021005 and 11275183) and the CAS, and the Fundamental Research Funds for the Central Universities.
Author information
Authors and Affiliations
Contributions
P.W., Z.Y., F.S. prepared the setup and performed the experiments; P.W., X.R. built the microwave circuits; X.X. and M.W. designed and fabricated the microwave delivery device; J.D. supervised the setup and experiments. All authors discussed the results and participated in writing the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Supplementary information
Supplementary Information
Supplementary Figures 1-2, Supplementary Table 1, Supplementary Notes 1-5 and Supplementary References (PDF 299 kb)
Rights and permissions
This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
About this article
Cite this article
Wang, P., Yuan, Z., Huang, P. et al. High-resolution vector microwave magnetometry based on solid-state spins in diamond. Nat Commun 6, 6631 (2015). https://doi.org/10.1038/ncomms7631
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/ncomms7631
This article is cited by
-
Vector Magnetometry Using a Single NV–13C Complex in Diamond
Journal of Applied Spectroscopy (2023)
-
Quantum-assisted distortion-free audio signal sensing
Nature Communications (2022)
-
A spin–orbit torque device for sensing three-dimensional magnetic fields
Nature Electronics (2021)
-
System for the remote control and imaging of MW fields for spin manipulation in NV centers in diamond
Scientific Reports (2020)
-
A CMOS-integrated quantum sensor based on nitrogen–vacancy centres
Nature Electronics (2019)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.