Introduction

Because of the stable fluorescence and long coherence time of its spin state, the negatively charged nitrogen vacancy (NV) center in diamond has been studied extensively over the past decade. This defect has the potential to be used for quantum computation,1,2,3,4 nanoscale metrology5,6,7,8 and biological imaging.9,10,11 To further extend the study of the interaction between a multi-NV center and the nanoscale sensing with the NV center, it is necessary to detect and control the NV center spin-state dynamics with high spatial resolution.7,12,13,14 Therefore, many optical super-resolution microscopy techniques have been developed to detect single NV centers.13,14,15,16,17Among these methods, stimulated emission depletion (STED) microscopy12,18,19,20 is one of the most promising. This method utilizes a doughnut-shaped laser to produce the position-dependent stimulated emission, which changes the fluorescence signal. With STED, the electron spin resonance signals of NV centers have been detected with a resolution lower than the diffraction limit.12,21,22

A new super-resolution microscopy technique was recently developed by Han et al.15 The authors replaced the stimulated excitation of STED with the dark-state pumping of NV centers. The dark state was later proven to be the neutral charge NV center (NV0) by other groups.23,24,25 The charge-state conversion results in a change of the local field in diamond26 and the spectral diffusion of the NV center.24,27 For high-fidelity quantum manipulation, the charge state should be well controlled.28 Based on the mechanism of charge-state conversion and super-resolution microscopy,25,29,30 we demonstrated the optical manipulation of the charge state of an NV center with subdiffraction resolution. By changing the duration and power of laser pulses, we optimized the charge-state depletion (CSD) microscopy to improve the resolution to 4.1 nm without oil immersion lens. With the charge-state manipulation, we were able to detect the electron spin-state dynamics of NV centers with subdiffraction resolution, which can be used for nanoscale sensing of electromagnetic fields and biological molecules. Because charge-state conversion changes the local field in diamond,2,26 the CSD microscopy method can potentially be used for the study of spin-state quantum coherent dynamics and the interaction between coupled spin systems in diamond.2,13,31

Materials and methods

The NV center in diamond consists of a substitutional nitrogen atom and an adjacent vacancy. Two charge states are usually observed: NV0 and NV (Figure 1a). Both NV and NV0 are stable in the absence of optical excitation.15,32 The electron spin of the NV ground state can be optically initialized and detected.1 The coherence time of the spin state could exceed several milliseconds.33 However, the NV0 charge state does not exhibit such properties. As observed in Figure 1c, the photons of NV0 and NV are different in wavelength. Therefore, we are able to distinguish the two charge states by measuring the NV center fluorescence. The conversion between these two charge states can be observed as the fluorescence on and off processes, which can be used for reversible saturable optical fluorescence transition microscopy.

Figure 1
figure 1

(a) A sketch depicting the charge-state conversion between NV and NV0. (b) Schematics of the experimental configuration. The lasers and fluorescence emission were combined and split using DM1 (edge wavelength 605 nm), DM2 (edge wavelength 536.8 nm) and DM3 (edge wavelength 658.8 nm). The fluorescence of NV was detected by APD2 with a long-pass filter (LP, edge wavelength 668.9 nm), and the fluorescence of NV0 was detected by APD 1 with a short-pass filter (SP, edge wavelength 599 nm). Two phase masks were used to produce the doughnut-shaped laser beams. The 637-nm and 532-nm lasers were used to initialize or switch the charge state, and the 589-nm laser is for the detection of the charge state. (c) Fluorescence spectrum of the NV center ensemble excited by a 532-nm laser at temperature approximately 250 K. The charge states of NV are marked by different ZPLs. The PSB of each charge state presents photons at longer wavelength. (d) Charge-state conversion process of NV ensemble pumped by a 532-nm laser. The charge states were initialized using a 637-nm laser. (e and f) Power dependence of the steady charge state population and charge-state conversion rates (γ=γr+γi). All of the lasers were Gaussian beam lasers. The charge-state conversion rates were obtained by fitting the results of NV fluorescence in d with Equation (3). APD, avalanche photodiode; DM, long-pass dichroic mirrors; NV, nitrogen vacancy; PSB, phonon sideband; ZPL, zero phonon line.

The samples used in this experiment were 〈100〉 surface diamond plates with nitrogen concentrations lower than 5 ppb. The NV centers were produced by N+ ion implantation and subsequent annealing. The optical system used to locate and detect the NV center is shown in Figure 1b. Laser pulses with various wavelengths and beam shapes were switched using acoustic optical modulators and then focused on the sample through an objective (numerical aperture=0.9). The fluorescence emission of NV0 and NV was split using a long-pass dichroic mirror and was further filtered using a short-pass filter and a long-pass filter, respectively.

Results and discussion

Optical manipulation of the charge state

Various methods have been developed to control the charge states of NV centers,25,27,30,34,35,36,37 with the most convenient method being high-speed optical manipulation. To demonstrate photon-induced charge-state conversion between NV0 and NV, the NV center ensemble fluorescence emission of both NV0 and NV charge states was detected. As demonstrated in Figure 1d, the fluorescence intensities of the two charge states changed in opposite manners when the NV center was pumped with a 0.68-mW, 532-nm laser. The fluorescence changing rates of the two charge states were almost the same (approximately 2.7 µs−1 in Figure 1d). In contrast, the excited state lifetimes of NV and NV0 were both approximately 13 ns in our experiments. As the fluorescence intensity is directly proportion to the population of charge state, the fluorescence changing rate can be simply presented as the charge-state conversion rate. This result indicates that the charge state can be switched from NV0 to NV using a 532-nm laser.

Both the ionization (NV to NV0) and recharging (NV0 to NV) processes can be excited using the same laser. The steady-state population is determined by both the ionization and recharging rates, which are significantly changed by the wavelength and power of the laser. We measured the charge-state population using the single-shot charge-state readout method.25,30 In our experiment, the highest population of NV (approximately 75%) was obtained by excitation with a 532-nm laser, while the lowest population of NV (approximately 5%) was obtained using a 637-nm laser (Figure 1e). Therefore, these two lasers were selected to initialize and change the charge-state population of the NV centers. The fidelity of charge-state initialization would be higher than that in Han's work,15 where a 473/592-nm laser was used to initialize the NV center charge state. Previous experiments demonstrated that the charge-state conversion is mainly a two-photon process with a laser wavelength longer than 500 nm.25,30,36 The NV is first pumped to the excited state of NV or NV0; then, an electron is released to the conduction band during ionization or captured from the valance band during recharging.24 The nonlinear power dependence of the conversion rates in Figure 1f was obtained by repeating the measurement of Figure 1d with different laser powers. The results demonstrate that the charge-state conversions with 532 and 637 nm are two-photon processes (see Supplementary Information for details).23,25,30

Simply, we can write the rate equations during charge-state conversion as

where ρ0 and ρ are the population of NV0 and NV, respectively, and γr and γi are the recharging and ionization conversion rates, respectively. If only the photons of NV are detected, the fluorescence intensity can be written as

where γ=γr+γi is the experimentally measured fluorescence changing rate, as observed in Figure 1d. is the steady-state NV population with laser pumping, where ρ−,in is the initial NV population. And, τ is the laser duration. Therefore, the high-spatial-resolution charge-state manipulation requires a position-dependent charge-state conversion rate γ(r) or a position-dependent steady-state population ρ−,st(r). According to the experimental results in Figure 1e, the steady-state populations change little with the power of 532-nm and 637-nm lasers. In contrast, the charge-state conversion rates increase with the power of the laser (Figure 1f). Hence, a doughnut-shaped laser beam is used to produce the position-dependent charge-state conversion rate. The power intensity at the beam center approaches zero. We considered two NV centers, one at the beam center with position r0 and the other at r0r. Because γ(r0)=0, the fluorescence difference between the two NV centers is . Therefore, for any Δr, the fluorescence difference is saturable and can be increased with the power and duration of the laser.

CSD microscopy

Next, we will outline the procedure for subdiffraction CSD microscopy imaging. Three different laser pulses were used in the experiment, as shown in Figure 2a and 2b. First, a Gaussian beam (G) laser was used to initialize the charge state in the focus region. Second, a doughnut-shaped (D) laser pulse was applied. The wavelength of the D laser was different from that of the G laser. This laser was able to deplete the NV0 or NV charge state except for the NV at the beam center. Finally, the charge state was detected with a 0.1-mW, 589-nm laser. The duration of the 589-nm laser was 5 µs. The microscopy image was obtained by detecting the fluorescence of NV and scanning the sample with a piezo stage. All of the lasers were circularly polarized to obtain the same conversion rates for the NV centers with different symmetry axes.

Figure 2
figure 2

(a and b) Laser sequences and example images for the high-resolution charge-state ionization and recharging manipulation, respectively. (c and d) The resolution (presented by FWHM) improved by increasing the D laser power or duration. The solid lines are the fits with Equation (6). The duration of the D lasers in c is 40 µs. The D laser in d is a 34-mW, 532-nm laser for iCSD and a 22-mW, 637-nm laser for rCSD. The inserts in c and d are the images of iCSD with different powers and rCSD with different durations, respectively. The scale bar is 100 nm. (e) The profiles of confocal (blue star dots) and CSD images. In our experiment, the best rCSD resolution was approximately 28.6 nm (22-mW, 637-nm doughnut laser with 160-µs duration), shown as red square dots. And the best iCSD resolution was 4.1 nm (48-mW, 532-nm doughnut laser with 200-µs resolution), shown as the green circular dots. The inset shows the iCSD image of the best resolution. The scale bar is 30 nm. FWHM, full width at half maximum; iCSD, ionization-charge-state depletion; rCSD, recharging-charge-state depletion.

For high-resolution ionization manipulation (change NV center to a high NV0 population state), we used a 637-nm G laser and a 532-nm D laser (Figure 2a). The 637-nm laser initialized the NV with an NV0 population of approximately 95%. The 532-nm D laser converted NV0 into NV with a position-dependent conversion rate. A typical microscopy image is presented in Figure 2a. NV centers appear as dark spots in the image. In contrast, using a 637-nm D laser and a 532-nm G laser, we achieved higher solution recharging manipulation (change NV center to a high NV population state) (Figure 2b). The NV center would be initialized to NV with a probability of approximately 75% and converted into NV0 with a position-dependent rate. NV centers appear as bright points in the fluorescence microscope image. To distinguish the imaging using different laser sequences, the CSD microscopy using a 637-nm D laser can be termed as recharging-CSD (rCSD), and the 532-nm D laser CSD is termed as ionization-CSD (iCSD).

The effective point-spread-function of the CSD image is represented by h(r)=hdet(r)ρ(r), where hdet(r) is the detection point-spread-function. For the two-photon charge-state conversion, the charge-state conversion rate should be quadratically dependent on the power of the laser. We simply write the conversion rate as γαI2 (see details in Supplementary Information), where I is the power of laser. The simplest function to depict the detection point-spread-function and the doughnut conversion laser intensity is the standing wavefunction:38,39

For simplicity, we assume (the width of the detection Gaussian beam laser approximates the width of the D laser). The resolution of CSD can be approximated by a Taylor series:

where , which indicates that increasing the power and duration of the D laser could improve the CSD resolution, as observed in Figure 2c and 2d. In addition, the microscopy resolution is theoretically unlimited. For the same power, the charge-state conversion rate of the 532-nm laser is higher than that of the 637-nm laser, as observed in Figure 1f, which might result from the wavelength-dependent absorption cross-section of the NV center at room temperature.25,36 Therefore, the resolution of iCSD is better than that of rCSD for the same power and duration of the D laser.

In reality, the improvement of the CSD resolution remains limited by the power of the laser and stability of the piezo stage. Figure 2e shows the best resolution in our experiment. The inset presents the iCSD image with the FWHMs (full widths at half maximum) in two directions, approximately 4.1 nm and 7.8 nm, respectively. As the drift of the piezo stage is estimated to be less than 0.15 nm s−1, the effect of drift on the resolution is approximately 1 nm. The difference between the two axes might be caused by the inhomogeneous intensity distribution of the doughnut laser beam center or the drift of the stage.22 To further improve the spatial resolution, a high numerical aperture oil objective could be used in the future. Such an objective would produce a higher charge-state conversion rate gradient than a dry objective.

Similar to the STED method, CSD can resolve the location of high-density NV centers better than confocal microscopy (Figure 3a and 3b). The lasers for charge-state conversion of CSD microscopy are much weaker than the lasers in STED microscopy with NV.19,21 Furthermore, our experiments demonstrate that an optimized G laser duration would also improve the spatial resolution without changing the D laser duration(see Supplementary Information for details). Therefore, the durations of both the G laser and D laser should be carefully adjusted to improve the imaging quality.

Figure 3
figure 3

(a and b) Confocal microscopy and rCSD microscopy images of the same area in the diamond sample. The confocal microscopy images were obtained by pumping and detecting NV with a 532-nm Gaussian beam laser. The rCSD microscopy images were obtained using method shown in Figure 2b. The D laser was a 22-mW, 637-nm laser with 50-µs duration. The dwell time of every scanning step was 50 ms. (c) The energy levels structure of NV. The inset shows the details of the triplet ground state. (d) The pulse sequence used to detect the ODMR signals. The signal of ODMR was detected by a 0.7-mW, 532-nm laser with 300-ns duration. (e and f) The ODMR signals of two NV centers. The resonant frequencies are marked with pink (NV A) and green (NV B) bars. With the magnetic field in e, the resonant frequencies were ω+,A=2.894 GHz, ω_,A=2.852 GHz, ω+,B=2.908 GHz, ω_,B=2.837 GHz. In f, the resonant frequencies were ω+,A=2.914 GHz, ω_,A=2.845 GHz, ω+,B=2.975 GHz, ω_,B=2.774 GHz. NV, nitrogen vacancy; ODMR, optically detected magnetic resonance; rCSD, recharging-charge-state depletion.

Spin-state detection

We have demonstrated that the NV center charge state can be controlled using CSD microscopy, which enables us to detect the spin state of NV with subdiffraction spatial resolution. The energy levels of the NV triplet ground state, as shown in Figure 3c, can be changed by the magnetic field and temperature. The two resonant frequencies of the spin-state transition can be detected using the optically detected magnetic resonance (ODMR) method. To compare the difference between CSD ODMR and confocal ODMR, we detected the ODMR signals of two adjacent NV centers with different symmetry axes, indicated as NV A and NV B in Figure 3b. The pulse sequences are shown in Figure 3d. The 532-nm Gaussian beam laser was used for both initialization and detection of the NV center. Using the 532-nm laser for detection will not affect the spatial resolution of rCSD ODMR, as the contrast of the ODMR signal between different NV centers will only be determined by the charge-state population before the microwave is applied. The resonant signals with different magnetic field were measured (Figure 3e and 3f). The confocal ODMR in the focus area did not distinguish the electron spin resonance signals of the different NV centers. In contrast, using rCSD ODMR, the resonant frequencies of each NV were separately determined. Only the NV centers at the beam center of the D laser will remain to be NV and contribute to the resonant signals of rCSD ODMR.

In Figure 3e and 3f, the noise of rCSD ODMR appears to be higher than that for confocal ODMR with the same acquisition time. This effect is caused by the long duration of the D laser, which compresses the time for fluorescence detection. Even though we did not detect iCSD ODMR, we expected that this method could also separate the signals for the different NV centers. The iCSD ODMR would require one more laser beam than rCSD ODMR, as the 637-nm Gaussian beam is not suitable for use for detecting the fluorescence of NV. Here, the magnetic field was supplied by a permanent magnet with a low magnetic field gradient. Supposing the magnetic field at NV A was the same as that at NV B, we calculated the magnetic field vector in Figure 3e40 to have magnitude B1≈25.5G and direction θ1,A≈73°, θ1,B≈60°. Here, θ represents the angle between the magnetic field and the symmetry axis of the NV center. Another magnetic field in Figure 3f was B2≈48.8G, θ2,A≈76°, θ2,B≈42°. The CSD ODMR results revealed the change in both the magnitude and direction of the magnetic field.

Although STED ODMR is also able to detect resonant frequencies of NV centers with subdiffraction resolution,12 the principles of STED ODMR and CSD ODMR are different. In the STED method, the resonant microwave pulses affect all of the NV centers; however, only one NV center's fluorescence is detected. Unlike STED ODMR, the CSD ODMR high-resolution resonant signal detection relies on charge-state manipulation. If an NV center is at NV0 charge state, no ODMR signal will be detected. The difference between STED ODMR and CSD ODMR might emerge in the results of coupled NV centers. For NV A and NV B in Figure 3, the resonant frequencies are different. The spin states of these two NV centers can be separately controlled using different frequencies of microwave pulses even without the help of CSD microscopy. However, NV C and NV D in Figure 3b have the same resonant frequency (Figure 4a).

Figure 4
figure 4

(a) rCSD ODMR signals of NV C and NV D in Figure 3. The different ODMR results for different magnetic fields revealed that the two NV centers had the same symmetry axis. (b) The electron spin-state Rabi oscillation using rCSD. (c) The Ramsey fringes of the NV spin states using the rCSD method. NV, nitrogen vacancy; ODMR, optically detected magnetic resonance; rCSD, recharging-charge-state depletion.

To selectively detect the electron spin-state dynamics of each NV center, we used the CSD method. The Rabi oscillation of an individual NV center was measured using the rCSD method, as shown in Figure 4b. The individual Ramsey fringe signals were also detected to obtain spin-state coherence information (Figure 4c). Both of the NV centers exhibited short spin coherence times, which might result from the local implantation damage.41 As the distance between these two NV centers was larger than 100 nm, the effect of dipole interaction was not observed. The results demonstrate that the spin-state information of NV can be detected with subdiffraction spatial resolution using CSD microscopy.

The charge-state conversion will destroy the information on the NV spin state. However, recent experiments have demonstrated that the charge-state conversion of NV centers could induce a detectable change in the local field in diamond,2,26 which might affect the dynamics of the spin system that is coupled to the NV center. Subsequently, we expect that CSD could enable us to study the NV center spin-state dynamics by changing the charge state of adjacent NV centers in future work. We can use different D lasers to change the adjacent NV to NV or NV0 and maintain the charge state of NV at the beam center to be NV. This proposal requires us to maintain the spin state of NV at the beam center while changing the charge state of other NVs. However, for the lasers used in this experiment, the charge-state conversion is usually much slower than the NV spin-state polarization.30 The spin state of NV might be polarized to ms=0 before we can change the charge state of an adjacent NV center. To solve this problem, one method would be to combine CSD microscopy with spin-RESOLFT microscopy.14 Another method would be to use a short wavelength laser (e.g., 405 nm) as the D laser. Because the NV center will not be pumped to the excited state of NV during a single-photon charge-state conversion process,25,30 the spin state of NV would not be changed before the charge state is switched.

Conclusions

In summary, we demonstrated subdiffraction charge-state manipulation of the NV centers in diamond. Laser beams with different wavelengths and shapes were used to control and detect the charge state. The best spatial resolution achieved in our experiment is approximately 4.1 nm, which can be further improved. Using CSD microscopy, we measured the resonance signal and coherent dynamics of a NV center spin state with subdiffraction spatial resolution, which could be used for nanoscale sensing. In the future, we expect that high-resolution NV charge-state manipulation could help to control the spin-state dynamics of NVs and to switch interactions between NV centers.