Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Original Article
  • Published:

Alarm safety and oxygen saturation targets in the Vermont Oxford Network iNICQ 2015 collaborative

Abstract

Objectives:

To assess progress of neonatal intensive care units (NICUs) participating in the Vermont Oxford Network iNICQ 2015: Alarm Safety Collaborative in achieving Joint Commission 2014 alarm safety goals with respect to oximeters, and to compare patient-level oxygen saturation (SpO2) and oximeter alarm data to local policies.

Study Design:

Prospective multicenter audits in February and August 2015 assessed implementation of policies addressing Joint Commission 2014 Alarm Safety goals, and ascertained SpO2 targets, oximeter alarm settings and compliance with policy-specified SpO2 targets and alarms.

Results:

Eighty-six NICUs completed both audits. Of 13 policies addressing mandated goals, median (interquartile range) 8 (5, 9) policies were implemented at audit 1 and 9 (6, 11) at audit 2 (P=0.004). At audit 1, 28 NICUs had implemented 9 policies versus 47 at audit 2. For 794 infants <31 weeks gestation, <36 weeks postmenstrual age, and on supplemental oxygen, median SpO2 target lower limit was 88% (interquartile range 87%, 90%; range 75% to 94%), upper limit 95% (interquartile range 94%, 96%; range 85% to 100%). High oximeter alarm was set according to local policy for 63% of infants, for whom SpO2 >97% was less frequent than when high alarm was not set to policy (10.1% vs 21.5%, P=0.006).

Conclusions:

Participating NICUs showed significant progress between audits in their implementation of Joint Commission Alarm Safety goals for oximeter monitoring. Oximeter high alarm not set per local policy is associated with increased hyperoxemia in preterm infants. Recommendations to standardize oxygen saturation targets for infants at risk for oxygenation-related outcomes have not been widely adopted.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Purchase on Springer Link

Instant access to full article PDF

Prices may be subject to local taxes which are calculated during checkout

Figure 1

Similar content being viewed by others

References

  1. The Joint Commission [Internet]. Medical device alarm safety in hospitals. The Joint Commission Sentinel Event Alert. 2013; Issue 50. April 8, 2013. Available at http://www.jointcommission.org/assets/1/18/SEA_50_alarms_4_5_13_FINAL1.PDF. Accessed: 7 October 2014.

  2. ECRI Institute. Top 10 Health Technology Hazards for 2013. Health Devices 2013; 41 (11): 342–365. Available at https://www.ecri.org/Documents/Secure/Health_Devices_Top_10_Hazards_2013.pdf. Accessed 7 October 2014.

    Google Scholar 

  3. Association for the Advancement of Medical Instrumentation [Internet]. Alarms Pose Challenges to Healthcare Facilities. Horizons. 2011; Spring 2011. Available at http://www.aami.org/publications/AlarmHorizons/articles/Alarms_Pose_Challenges_5.pdf Accessed 7 October 2014.

  4. The Joint Commission [Internet]. National patient safety goals effective January 1, 2014. The Joint Commission Hospital and Critical Access Hospital Accreditation Program. 2014; Available at http://www.jointcommission.org/assets/1/18/jcp0713_announce_new_nspg.pdf Accessed: 10 October 2016.

  5. Schmidt B, Whyte R, Asztalos E, Moddemann D, Poets C, Rabi Y et al. Effects of targeting higher vs lower arterial oxygen saturations on death or disability in extremely preterm infants. JAMA 2013; 309 (20): 2111–2120.

    Article  CAS  Google Scholar 

  6. Stenson B, The BOOST II United Kingdom A, New Zealand Collaborative Groups. Oxygen saturation and outcomes in preterm infants. NEJM 2013; 368: 2094–2104.

    Article  Google Scholar 

  7. SUPPORT Study Group of the Eunice Kennedy Shriver NICHD Neonatal Research Network. Target ranges of oxygen saturation in extremely preterm infants. NEJM 2010; 362: 1959–1969.

    Article  Google Scholar 

  8. Taylor M, McNicholas C, Nicolay C, Darzi A, Bell D, Reed J . Systematic review of the application of the plan-do-study-act method to improve quality in healthcare. BMJ Qual Saf 2014; 23 (4): 290–298.

    Article  Google Scholar 

  9. Atwater A, Hartmann E, Brown B, Carteaux P, Freeman M, Hegwood P et al. Evaluation and development of potentially better practices for staffing in neonatal intensive care units. Pediatrics 2006; 118 (Suppl 2): S134–S140.

    Article  Google Scholar 

  10. Langley G, Moen R, Nolan K, TW N, Norman C, Provost L . The Improvement Guide: A Practical Approach to Enhancing Organizational Performance, 2nd edn. Jossey-Bass: San Francisco, CA, 2009.

    Google Scholar 

  11. Armbruster J, Schmidt B, Poets C, Bassler D . Nurses' compliance with alarm limits for pulse oximetry: qualitative study. J Perinatol 2010; 30 (8): 531–534.

    Article  CAS  Google Scholar 

  12. Chow L, Wright K, Sola A. CSMC Oxygen Administration Study Group. Can changes in clinical practice decrease the incidence of severe retinopathy of prematurity in very low birth weight infants? Pediatrics 2003; 111 (2): 339–345.

    Article  Google Scholar 

  13. Hagadorn J, Furey A, Nghiem T, Schmid C, Phelps D, Pillers D et al. Achieved versus intended pulse oximeter saturation in infants born less than 28 weeks' gestation: the AVIOx study. Pediatrics 2006; 118 (4): 1574–1582.

    Article  Google Scholar 

  14. Nghiem T, Hagadorn J, Terrin N, Syke S, MacKinnon B, Cole C . Nurse opinions and pulse oximeter saturation target limits for preterm infants. Pediatrics 2008; 121 (5): e1039–e1046.

    Article  Google Scholar 

  15. Sink D, Hope S, Hagadorn J . Nurse:patient ratio and achievement of oxygen saturation goals in premature infants. Arch Dis Child Fetal Neonatal Ed 2011; 96 (2): F93–F98.

    Article  Google Scholar 

  16. Ivers N, Jamtvedt G, Flottorp S, Young J, Odgaard-Jensen J, French S et al. Audit and feedback: effects on professional practice and healthcare outcomes. Cochrane Database Syst Rev 2012; 6: CD000259.

    Google Scholar 

  17. Saugstad O, Aune D . Optimal oxygenation of extremely low birth weight infants: a meta-analysis and systematic review of the oxygen saturation target studies. Neonatology 2014; 105 (1): 55–63.

    Article  CAS  Google Scholar 

  18. Poets C, Roberts R, Schmidt B, Whyte R, Asztalos E, Bader D et al. Association between intermittent hypoxemia or bradycardia and late death or disability in extremely preterm infants. JAMA 2015; 314 (6): 595–603.

    Article  CAS  Google Scholar 

  19. Fierson W. American Academy of Pediatrics Section on O, American Academy of O, American Association for Pediatric O, Strabismus, American Association of Certified O. Screening examination of premature infants for retinopathy of prematurity. Pediatrics 2013; 131 (1): 189–195.

    Article  Google Scholar 

  20. Stenson B, Brocklehurst P, Tarnow-Mordi W . Increased 36-week survival with high oxygen saturation target in extremely preterm infants. N Engl J Med 2011; 364 (17): 1680–1682.

    Article  CAS  Google Scholar 

  21. Schmidt B, Whyte R, Roberts R . Trade-off between lower or higher oxygen saturations for extremely preterm infants: the first benefits of oxygen saturation targeting (BOOST) II trial reports its primary outcome. J Pediatr 2014; 165 (1): 6–8.

    Article  Google Scholar 

  22. Polin R, Bateman D, Sahni R . Pulse oximetry in very low birth weight infants. Clin Perinatol 2014; 41 (4): 1017–1032.

    Article  Google Scholar 

  23. Bancalari E, Claure N . Oxygenation targets and outcomes in premature infants. JAMA 2013; 309 (20): 2161–2162.

    Article  CAS  Google Scholar 

  24. Polin R, Bateman D . Oxygen-saturation targets in preterm infants. N Engl J Med 2013; 368 (22): 2141–2142.

    Article  CAS  Google Scholar 

  25. Bateman D, Polin R . A lower oxygen-saturation target decreases retinopathy of prematurity but increases mortality in premature infants. J Pediatr 2013; 163 (5): 1528–1529.

    Article  Google Scholar 

  26. Stenson B . Oxygen targets for preterm infants. Neonatology 2013; 103 (4): 341–345.

    Article  Google Scholar 

  27. Askie L . Optimal oxygen saturations in preterm infants: a moving target. Curr Opin Pediatr 2013; 25 (2): 188–192.

    Article  Google Scholar 

  28. Saugstad O, Speer C, Halliday H . Oxygen saturation in immature babies: revisited with updated recommendations. Neonatology 2011; 100 (3): 217–218.

    Article  CAS  Google Scholar 

  29. Lakshminrusimha S, Manja V, Mathew B, Suresh G . Oxygen targeting in preterm infants: a physiological interpretation. J Perinatol 2015; 35 (1): 8–15.

    Article  CAS  Google Scholar 

  30. Johnston ED, Boyle B, Juszczak E, King A, Brocklehurst P, Stenson B . Oxygen targeting in preterm infants using the Masimo SET Radical pulse oximeter. Arch Dis Child Fetal Neonatal Ed 2011; 96 (6): F429–F433.

    Article  Google Scholar 

  31. Schmidt B, Roberts R, Whyte R, Asztalos E, Poets C, Rabi Y et al. Impact of study oximeter masking algorithm on titration of oxygen therapy in the Canadian Oxygen Trial. J Pediatr 2014; 165 (4): 666–71 e2.

    Article  CAS  Google Scholar 

  32. Manja V, Lakshminrusimha S, Cook D . Oxygen saturation target range for extremely preterm infants: a systematic review and meta-analysis. JAMA Pediatr 2015; 169 (4): 332–340.

    Article  Google Scholar 

  33. Walsh M, Di Fiore J, Martin R, Marie Gantz M, Carlo W, Finer N . Association of oxygen target and growth status with increased mortality in small for gestational age infants: further analysis of the Surfactant, Positive Pressure and Pulse Oximetry Randomized Trial. JAMA Pediatr 2016; 170 (3): 292–294.

    Article  Google Scholar 

  34. Anderson C, Benitz W, Madan A . Retinopathy of prematurity and pulse oximetry: a national survey of recent practices. J Perinatol 2004; 24 (3): 164–168.

    Article  Google Scholar 

  35. Brocklehurst P [Internet]. Boost-II UK: benefits of oxygen saturation targeting: protocol—which oxygen saturation level should we use for very premature infants? a randomised controlled trial. 2011; Version 3. University of Oxford; Available at https://www.npeu.ox.ac.uk/downloads/files/boost/BOOSTII-Protocol-Version-3-Nov-2011.pdf. Accessed: 28 January 2016.

  36. Ketko A, Martin C, Nemshak M, Niedner M, Vartanian R . Balancing the tension between hyperoxia prevention and alarm fatigue in the NICU. Pediatrics 2015; 136: e496–e504.

    Article  Google Scholar 

  37. Clucas L, Doyle L, Dawson J, Donath S, Davis P . Compliance with alarm limits for pulse oximetry in very preterm infants. Pediatrics 2007; 119 (6): 1056–1060.

    Article  Google Scholar 

  38. Ruiz T, Trzaski J, Sink D, Hagadorn J . Transcribed oxygen saturation vs oximeter recordings in very low birth weight infants. J Perinatol 2014; 34 (2): 130–135.

    Article  CAS  Google Scholar 

  39. American Academy of Pediatrics Section on Ophthalmology, American Academy of Ophthalmology, American Association for Pediatric Ophthalmology and Strabismus, American Association of Certified Orthoptists. Screening examination of premature infants for retinopathy of prematurity. Pediatrics 2013; 131 (1): 189–195.

    Article  Google Scholar 

Download references

Acknowledgements

No external funding was secured for this study.

Author contributions

Dr Hagadorn drafted the initial manuscript; Dr Edwards and Ms Morrow performed statistical analysis; and all authors contributed to conceptualization, design, and analytical approach, reviewed and revised the manuscript, approved the final manuscript as submitted and agree to be responsible for the content accordingly.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to D W Sink.

Ethics declarations

Competing interests

Dr Buus-Frank, Ms Ferrelli, Ms Morrow, Dr Horbar and Dr Soll are employees of the Vermont Oxford Network. Dr Edwards is supported by a grant from the Vermont Oxford Network to the University of Vermont. Drs Hagadorn and Sink were paid consultants to the Vermont Oxford Network for this project.

Additional information

Supplementary Information accompanies the paper on the Journal of Perinatology website

Supplementary information

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hagadorn, J., Sink, D., Buus-Frank, M. et al. Alarm safety and oxygen saturation targets in the Vermont Oxford Network iNICQ 2015 collaborative. J Perinatol 37, 270–276 (2017). https://doi.org/10.1038/jp.2016.219

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/jp.2016.219