Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

  • Original Article
  • Published:

Superiority of targeted neonatal echocardiography for umbilical venous catheter tip localization: accuracy of a clinician performance model

Abstract

Objectives:

To compare targeted neonatal echocardiography (TnECHO) and antero-posterior (AP) chest radiographs in the localization of the umbilical venous catheter (UVC) tip and to determine the accuracy of UVC tip localization by TnECHO when performed by a trained cohort of pediatric housestaff physicians.

Study design:

Prospective, observational study of consecutive neonates requiring UVC insertion, in a tertiary care center, in an 18-month period. Chest radiographs reporting optimal position of the UVC tip were compared with the TnECHO results of the pediatric cardiologist. The latter was also compared with the results of TnECHO performed by the pediatric housestaff physicians.

Result:

Thirty neonates with birth weight ranging between 270 and 4490 g and gestational age ranging between 24 and 44 weeks were enrolled. Nine patients (27%) required UVC tip repositioning as the cardiologist performed TnECHO revealed sub-optimal tip position despite optimal position on chest radiography. Among them, four had the UVC tip in the right atrium and five in the left atrium. Compared with the cardiologist, the housestaff physicians had reported TnECHO with a high-accuracy rate (area under the receiver operating characteristic curve=0.81).

Conclusion:

TnECHO is superior to chest radiography for identifying malpositioned catheters. TnECHO performed by pediatric housestaff physicians with basic training, demonstrated high-accuracy rates.

This is a preview of subscription content, access via your institution

Access options

Buy this article

Prices may be subject to local taxes which are calculated during checkout

Figure 1
Figure 2
Figure 3

Similar content being viewed by others

References

  1. Megha M, Jain N, Pillai R . Pericardial tamponade in a newborn following umbilical catheter insertion. Ind Pediatr 2011; 48: 404–405.

    Google Scholar 

  2. Alabsi S . Neonatal cardiac tamponade and pleural effusion resolved with chest tube placement. Neonatal Netw 2010; 29: 347–351.

    Article  Google Scholar 

  3. Sehgal A, Cook V, Dunn M . Pericardial effusion associated with an appropriately placed umbilical venous catheter. J Perinatol 2007; 27: 317–319.

    Article  CAS  Google Scholar 

  4. Traen M, Schepens E, Laroche S, van Overmeire B . Cardiac tamponade and pericardial effusion due to venous umbilical catheterization. Acta Paediatr 2005; 94: 626–628.

    Article  Google Scholar 

  5. Van Niekerk M, Kalis NN, Van der Merwe PL . Cardiac tamponade following umbilical vein catheterization in a neonate. S Afr Med J 1998; 88 (Suppl 2): C87–C90.

    PubMed  Google Scholar 

  6. González Diéguez CC, Bárrios Alonso V, Torrealday Taboada H, Tamariz Martel A, Hellín Sanz T, Brito Pérez JM et al. Intracardiac mycetoma induced by central catheterization. An Esp Pediatr 1992; 37: 63–65.

    PubMed  Google Scholar 

  7. Rastogi A, Luken JA, Pildes RS, Chrystof D, LaBranche F . Endocarditis in neonatal intensive care unit. Pediatr Cardiol 1993; 14: 183–186.

    Article  CAS  Google Scholar 

  8. Ades A, Sabble C, Cummings S, Cross R, Martin G . Echocardiographic evaluation of umbilical venous catheter placement. J Perinat 2003; 23: 24–28.

    Article  Google Scholar 

  9. Reval CR, Gonzalea E, Bhat AM, Pearlman SA, Stefano JL . Umbilical venous catheter: evaluation of radiographs to determine position and associated complications of malpositioned umbilical venous catheters. Am J Perinat 1995; 12: 201–204.

    Article  Google Scholar 

  10. Shukla H, Ferrara A . Rapid estimation of insertion length of umbilical catheters in newborn. Am J Dis Child 1986; 140: 786–788.

    CAS  Google Scholar 

  11. Chatel-Meijer MP, Roques-Gineste M, Fries F, Bloom MC, Laborie S, Lelong-Tissier MV et al. Cardiac Tamponade secondary to umbilical venous catheterization accident in premature infant. Arch Fr Pediatr 1992; 49: 373–376.

    CAS  PubMed  Google Scholar 

  12. Diamond LK, Allen FH, Wo T . Erythroblastosis fetalis: treatment with exchange transfusion. N Engl J Med 1951; 244: 39–49.

    Article  CAS  Google Scholar 

  13. Greenberg M, Movahed H, Peterson B, Bejar R . Placement of umbilical venous catheters with use of bedside real time ultrasonography. J Pediat 1995; 126: 633–635.

    Article  CAS  Google Scholar 

  14. George LJ, Deane W, Kirkpatrick E, Searle W, Turner SJ, Peppelbaum A . Umbilical vascular catheters: localization by two-dimensional echocardio/aortography. Pediatr Cardiol 1982; 2: 237–243.

    Article  CAS  Google Scholar 

  15. Nadroo AM . Death as a complication of peripherally inserted central catheter in neonates. J Pediat 2001; 138: 599–601.

    Article  CAS  Google Scholar 

  16. Nadroo AM, Roland BG, Jing L, Robert SG, Holzman LR . Change in upper extremity position caused migration of peripherally inserted central catheter in neonates. Pediatrics 2002; 110: 131–136.

    Article  Google Scholar 

Download references

Acknowledgements

We thank the study participants and their parents. We acknowledge Richard Tucker, Department of Pediatrics, Women and Infants Hospital of Rhode Island for the kind assistance with the statistical analyses in the study and Mandy Zhao, ECHO Technologist, Department of Pediatric Cardiology, New York Methodist Hospital, for her technical assistance with the study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to A M Nadroo.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Pulickal, A., Charlagorla, P., Tume, S. et al. Superiority of targeted neonatal echocardiography for umbilical venous catheter tip localization: accuracy of a clinician performance model. J Perinatol 33, 950–953 (2013). https://doi.org/10.1038/jp.2013.96

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/jp.2013.96

Keywords

This article is cited by

Search

Quick links