Extracorporeal membrane oxygenation (ECMO) has become a rescue therapy for neonates with severe, but potentially reversible, respiratory failure when maximal conventional therapy has failed. Without ECMO, there is a high mortality rate (1,2). Since the first newborn was successfully treated with venoarterial extracorporeal membrane oxygenation (va-ECMO) by Bartlett et al. in 1976 (3), more than 22,000 neonates have been treated with ECMO, of which 15,000 with va-ECMO, with an overall survival rate of 80% (4). Although ECMO has increased survival, the occurrence of hemorrhagic and ischemic cerebral lesions resulting in future neurologic and neurodevelopmental dysfunction are of major concern (5,6). Imaging studies revealed hemorrhagic or ischemic intracranial abnormalities in 10 up to even 52% of the patients (79). These complications justify a continuous search for factors related to intracranial hemorrhage and ischemia and treatment strategies aimed to improve outcome.

Part of the va-ECMO system is a bladderbox, which is situated between the venous cannula and the rollerpump of the extracorporeal system. The bladderbox controls venous drainage of blood from the right atrium of the patient. If the venous drainage becomes inadequate to maintain the established flow of the ECMO system, the bladderbox will give an alarm, slow down the pump, and finally cause an acute interruption of the rollerpump resulting in interruption of blood flow to the patient. If other causes of inadequate venous drainage from the right atrium and consecutive bladderbox alarms, like disturbed position of the cannula, are ruled out, intravascular volume administration is often given to restore the circulation through the ECMO system and the patient. Recently, our group first showed a statistically significant relation between the total number and volume of intravascular volume administration and the occurrence of intracranial hemorrhage (ICH) during va-ECMO treatment (10). We hypothesize that this relation may be explained by fluctuations in cerebral oxygenation and hemodynamics caused by either the bladderbox alarms with a consecutive interruption of the ECMO blood flow to the patient or by intravascular volume administration itself. Both might be a contributing risk factor for intracranial hemorrhage and ischemia. The aim of this study is to describe possible fluctuations in cerebral oxygenation and hemodynamics, as measured by near infrared spectrophotometry (NIRS), caused by simulated bladderbox alarms during va-ECMO in lambs.

METHODS

Study population and ECMO.

In a prospective interventional setting, six healthy lambs were studied for changes in cerebral oxygenation and hemodynamics caused by bladderbox alarms during va-ECMO. The lambs were obtained from local farmers, and the Ethical Committee on Animal Research of the Radboud University of Nijmegen approved the study. The care and handling of the animals were in accordance with the guidelines issued by the National Institutes of Health. Experimental settings were mainly based on earlier research in our group (11). General anesthesia was induced by i.v. administration of midazolam 0.2 mg · kg−1, ketamine 15 mg · kg−1, atropine 0.01 mg · kg−1, and pentobarbital 20 mg · kg−1. Muscle relaxation was obtained with 0.02 mg · kg−1 pancuronium. After endotracheal intubation, mechanical ventilation was started using a Babylog 8000 (Dräger, Lübeck, Germany) to maintain normal blood gas values (pH, 7.30–7.40; Pao2, 10–15 kPa; Paco2, 4.5–5.5 kPa). During mechanical ventilation, anesthesia was maintained by infusion of midazolam (0.2 mg · kg−1 · h−1), ketamine (10 mg · kg−1 · h · −1), sufentanyl (20 μg · kg−1 · h−1), and pancuronium (0.02 mg · kg−1 · h−1). Heart rate (HR) was monitored by needle electrodes on the chest wall. Central temperature was measured with a rectal probe and maintained between 38.5 and 39.5°C. A catheter was placed in the aorta through the right femoral artery for continuous measurement of mean arterial blood pressure (MAP). The right femoral vein was used to place a 7.5-Fr catheter into the inferior caval vein to create a venous access for medication and measurement of central venous pressure (CVP). Both MAP and CVP were represented on one monitor (HP 68S, Hewlett Packard, Boeblingen, Germany). Blood flow in the left common carotic artery (Qcar) was measured using an ultrasonic flowmeter (Transonic Systems Inc, NY, NY). Finally, arterial oxygen saturation (Sao2) and end-tidal CO2 were continuously registered.

Before venoarterial cannulation, the ECMO circuit was primed with full fresh sheep blood. The right common carotid artery was cannulated with an arterial catheter (8–12 Fr) and the right internal jugular vein with a venous catheter 10–12 Fr (both Biomedicus, Medtronics Bio-Medicus, Eden Prairie, MN). Thereafter, va-ECMO was initiated. During the cannulation procedure, a loading dose of heparin was administered (150 IU/kg) followed by continuous i.v. infusion of heparin (100–200 IU · kg−1 · h−1) to maintain the activated clotting time between 200 and 250 s (Hemochron, Edison, NJ).

The ECMO circuit itself consisted of a custom packed ¼-inch flexible polyvinylchloride tubing (Baxter, Uden, The Netherlands), with a silicone reservoir, the bladderbox (Seabrook Medical Systems, Inc, Cincinnati, OH), a hollow fiber membrane oxygenator with integrated heat-exchanger (Medos Hilite 2400LT, Medos, Stolberg Germany), and a rollerpump (Jostra RPM 20–320, Lung, Sweden). An ultrasonic flow meter (Transonic Systems, Ithaca, NY) was placed on the venous site of the ECMO system continuously measuring actual flow in the circuit. The aimed flow rate of the ECMO pump was 100–150 mL · kg−1 · min−1.

NIRS.

Cerebral oxygenation and hemodynamics were studied with NIRS. The NIRS equipment (OXYMON) used was developed by the instrumentation department and the department of physiology of the Radboud University Nijmegen Medical Centre, The Netherlands (12). The technique is based on the spectrophotometric measurement of changes in the absorption properties of Hb in the near infrared region, depending on its oxygenation state, first described by Jöbsis in 1977 (13). The NIRS procedure used in our ECMO center has been described to more extend in earlier studies of our research group (14,15). From an optode, placed frontal in the midline of the skull, three wavelengths of near infrared light (905, 850, and 767 nm) were transmitted through the head and received by two other optodes placed both on the left and right parietotemporal region of the skull, to evaluate the oxygenation and hemodynamics in each hemisphere separately. Concentration changes of oxyhemoglobin (ΔcO2Hb) and deoxyhemoglobin (ΔcHHb) were calculated from changes in light absorption using the modified Lambert-Beer law. We used the Keele absorption matrix and a constant path length multiplier factor of 4.27 times the distance between both optodes (1618). Using a value of 1.05 g · mL−1 for brain specific mass, concentration changes are expressed in μmol · 100 · g−1 (19). ΔcO2Hb and ΔcHHb reflect changes in cerebral oxygen supply, if oxygen consumption remains constant. Differences between ΔcO2Hb and ΔcHHb were calculated and indicated as changes in the Hb oxygenation index (ΔHbD, where ΔHbD = ΔcO2Hb − ΔcHHb in μmol · 100 · g−1). Concentration changes in total Hb (ΔctHb) were calculated as the sum of ΔcO2Hb and ΔcHHb (ΔctHb = ΔcO2Hb + ΔcHHb). ΔHbD represents changes in cerebral blood flow (20), whereas ΔctHb reflects changes in cerebral blood volume (ΔCBV), which are calculated from the formula: ΔCBV = 4 × ΔctHb/0.69 × cHb, expressed in mL · 100 · g−1. In this formula, cHb is the Hb concentration in circulating blood (mmol/L), 0.69 is the cerebral-arterial hematocrit ratio, and 4 is a correction factor. Also, because ctHb is calculated from changes in light absorption using an extinction coefficient based on the tetrahaeme molecule, whereas cHb determination is based on the monohaeme molecule (21).

Experimental procedures.

After the initiation of va-ECMO, we waited for a stable NIRS signal. Once this was established, bladderbox alarms were simulated. During the first experiment (experiment A), the ECMO flow rate of the rollerpump was increased stepwise with 10 mL · kg−1 · min−1 until bladderbox alarms occurred. After 10 s, the ECMO flow rate was decreased to the preexperiment value, and the bladderbox alarms were thereby resolved. Experiment A was repeated three times in each lamb with a minimal time interval of 2 min to obtain a stable NIRS signal before each experiment for at least 30 s.

During the second experiment (experiment B), the venous cannula of the ECMO circuit was partially occluded by a nonocclusive clamp that was developed for this experiment. Bladderbox alarms occurred as result of insufficient venous drainage and after 10 s, the clamp was removed to restore the preexperiment ECMO flow and resolve the bladderbox alarms. Again, experiment B was repeated three times in each lamb with a minimal time interval of 2 min.

During the third experiment (experiment C), bladderbox alarms were simulated similar to experiment A. However, after 10 s, bladderbox alarms were resolved by starting intravascular volume administration with 50 mL of normal saline. To avoid an overload of intravascular volume administration, experiment C was repeated two times in each lamb, with a time interval long enough to obtain at least a 30 s stable NIRS signal.

Arterial blood samples were obtained before and after each experiment to check for changes in pH, Pao2, Paco2, and cHb values.

Data analysis.

All physiologic parameters were registered by a data acquisition system (Poly, Inspector Research System, Amsterdam, The Netherlands) with a sampling frequency of 1 Hz and synchronized with the NIRS data. Baseline values (mean ± SD) of cO2Hb, cHHb, ctHb, and physiologic parameters were determined during a 15-s period during the stable NIRS signal before (t = 0) each single experiments (A, B, and C). Maximal changes in cO2Hb, cHHb, ctHb, and physiologic parameters were searched by a computer analysis during (t = 1) and after resolving (t = 2) the bladderbox alarms as mean values during a 3-s period. Because of the continuous registration, related changes in physiologic data could be described during changes of the NIRS signals. Changes in HbD and CBV were calculated by using the formulas mentioned earlier. All changes were described separately for the left and right hemisphere of the brain. Therefore, ΔHbD is expressed as ΔHbDright and ΔHbDleft, while ΔCBV is expressed as ΔCBVright and ΔCBVleft. Statistical analysis was performed for maximal changes at t = 1 and t = 2 compared with the baseline values by using Wilcoxon signed ranks tests, and p values <0.05 were considered to be statistically significant.

RESULTS

All the three experiments A, B, and C were successfully performed in six lambs, weight (mean ± SD) 5.2 ± 1.1 kg. The flow rate of the ECMO pump aimed for (100–150 mL · kg−1 · min−1) was reached in five lambs. In one lamb, maximal flow rate of the ECMO pump was 90 mL · kg −1 · min −1. As measured by arterial blood gas samples, all the lambs had blood gas values within the normal range before and after each experiment. For each experiment, cHb could be determined and used for calculations of ΔCBV. An overview of mean changes in important physiologic parameters, HbD and CBV, is given in Table 1, while Figure 1 shows the typical pattern of cO2Hb, cHHb, HbD, and CBV during and after bladderbox alarms in one lamb. Baseline HbD and CBV values, calculated during a 15-s period showed a baseline variability with a mean SD of 0.03 ± 0.01 μmol · 100 · g−1 and 0.02 ± 0.01 mL · 100 · g−1, respectively. Significant changes in Sao2 and end-tidal CO2 were not observed.

Table 1 Changes in hemoglobin oxygenation index (HbD), cerebral blood volume (CBV), mean arterial pressure (MAP), blood flow in the left carotic artery (Qcar), heart rate (HR), and central venous pressure (CVP) during and after bladderbox alarms, compared with baseline values
Figure 1
figure 1

Typical pattern of changes in cerebral oxyhemoglobin (ΔcO2Hb), deoxyhemoglobin (ΔcHHb), Hb oxygenation index (ΔHbD), and cerebral blood volume (ΔCBV) during and after bladderbox alarms.

Bladderbox alarms caused a significant and acute decrease in O2Hb and increase in HHb. After the acute decrease, there was a slight increase in O2Hb during the period of bladderbox alarms, followed by an acute increase to values above baseline values when bladderbox alarms were resolved. After the acute increase, HHb did show a further very slight increase during the period of bladderbox alarms as well, followed by a more acute decrease to values lower than baseline values after the bladderbox alarms were resolved.

Bladderbox alarms caused a significant and acute decrease in HbD, CBV, MAP, and Qcar. There was a significant increase in CVP. During the 10 s course of bladderbox alarms, HR increased as did HbD, CBV, MAP, and Qcar; however, just before resolving the bladderbox alarms, values were still lower than preexperiment baseline values. Patterns for all parameters were not different between those bladderbox alarms simulated by increasing the ECMO flow rate and those simulated by partially clamping of the venous cannula. We did not find any significant difference between ΔHbDright and ΔHbDleft or between ΔCBVright and ΔCBVleft. The decrease in HbD when simulating bladderbox alarms ranged from −0.05 to −0.45 μmol · 100 · g−1, while the decrease in CBV ranged from 0 to −0.20 mL · 100 g−1.

When resolving the bladderbox alarms, there was a fast increase in HbD to values significantly higher than baseline values before the experiment. In experiment C, when bladderbox alarms were resolved by the administration of intravascular volume, this increase in HbD (range, 0.09 to 0.84 μmol · 100 · g−1) was significantly larger than in experiments A and B [range, −0.06 to 0.35 μmol · 100 · g−1 (p < 0.05)]. Because normal saline is used as intravascular volume administration to resolve bladderbox alarms in experiment C, this will result in hemodilution (decrease of cHb). Therefore, changes in CBV cannot be reliably calculated. However, in experiments A and B, CBV increased to values significantly higher than baseline values (range, 0 to 0.26 mL · 100 · g−1). MAP and Qcar recovered to baseline values in experiments A and B but increased significantly when intravascular volume was administered (experiment C). CVP tends to decrease after bladderbox alarms are resolved but increases during intravascular volume administration. HR was still increased at the end of each experiment. Again, we did not find any significant difference between ΔHbDright and ΔHbDleft or between ΔCBVright and ΔCBVleft.

DISCUSSION

Because cerebral hemorrhage and ischemia belong to the devastating complications of ECMO treatment, it is important to identify risk factors related to the use of the ECMO technique itself and eliminate these as much as possible. After describing a significant relation between intravascular volume administration during va-ECMO treatment and ICH (10), this is the first study to describe statistically significant changes in cerebral oxygenation and hemodynamics, expressed by ΔHbD and ΔCBV, in relation to bladderbox alarms. These alarms often precede the need for intravascular volume administration.

Bladderbox alarms cause a stop of the ECMO pump, which results in an acute interruption of the arterial circulation and an acute decrease in MAP, Qcar, HbD, and CBV. Although bladderbox alarms persist, several mechanisms might explain observed findings of our study. First, in an attempt to restore the circulation, the body will try to increase cardiac output with an increase in HR. Especially, as lambs in our study have healthy lungs, an increase in cardiac output will result in an increased arterial blood flow, which explains the increase in MAP, Qcar, HbD, and CBV.

When the ECMO pump is interrupted, venous drainage from the right atrium will be reduced and CVP will increase resulting in less venous drainage from the brain. This might contribute to an increase in CBV. Finally, cerebral vasodilatation might occur as compensation for decreased CBF during bladderbox alarms. When bladderbox alarms are resolved, an acute recovery of ECMO flow will occur with a consecutive increase in MAP and Qcar, occurring while the system was just adapting to reduced CBF by vasodilatation. This might, partially, explain why there is an overshoot in HbD and CBV to values higher than baseline values. Once intravascular volume administration is used, there is a further increase of MAP, Qcar, and cerebral blood flow with a consecutive further increase in HbD.

Several authors reported that the majority of cases of ICH occurred within 72 h after the start of ECMO (2224). This might fit with our findings, as the majority of bladderbox alarms occur in the first days of ECMO treatment, when a relatively high ECMO flow rate is required to maintain adequate arterial oxygen saturation in a period that the patients lung is passing a phase of “white out” and the patient is still hemodynamically stabilizing. Next to this, the acute changes in HbD and CBV, related to bladderbox alarms in the first days, occur while cerebral autoregulation (AR) in the patient seems to be disturbed. In normal, healthy infants, AR maintains the cerebral blood flow over a wide range of cerebral perfusion pressures with a contra-regulation response occurring with a delay of 2 s after sudden changes in blood pressure. However, it has been shown that AR is disturbed in severely ill term infants (2528). Additional studies in newborn lambs showed that prolonged hypoxia and/or the ECMO treatment itself significantly disturbs the AR (2932). These studies demonstrated that AR in lambs placed on ECMO was more disturbed in animals that had hypoxia before ECMO compared with animals without preceding hypoxia. Therefore, the hemodynamic changes that occur during bladderbox alarms as demonstrated in our study could even be more profound in the clinical situation of newborns with respiratory and/or circulatory distress.

Large prospective randomized studies in human newborns to investigate the relation between (simulated) bladderbox alarms and the occurrence rate of intracranial lesions during treatment with ECMO are impossible to perform. Therefore, it is important to establish observational intervention studies. The most important limitation of using NIRS, with its variables like HbD and CBV, is the lack of reference values below which global cerebral oxygenation and hemodynamics are compromised and clinically relevant for the individual patient. Findings can be used to describe trends in a group of patients and the relevance can only be suspected when compared with studies where known risk factors for the brain are related to changes in NIRS variables. Baseline variability of HbD has been described to range from −0.12 to + 0.13 μmol · 100 · g−1 (33). With a variability of 0.03 ± 0.01, this was lower in our study. In the same study, a decrease in HbD of >0.3 μmol · 100 · g−1 is likely to be clinically relevant since this was found to correspond with the effect of reducing Sao2 by 12%. Compared with these values, mean changes in HbD of 0.11 to 0.33 μmol · 100 · g−1 and changes for individual lambs in HbD up to 0.45 μmol · 100 · g−1 in experiments A and B and up to 0.84 μmol · 100 · g−1 when intravascular volume administration was used could be important. Fluctuation of HbD, expressed as changes between maximal decrease in HbD, and the consecutive maximal increase in HbD are even more profound. For ΔCBV, changes have been compared with usual physiologic parameters like changes in Paco2 and MAP. During hypercapnia in normoxemic piglets on ECMO, a mean ΔCBV of 0.03 mL · 100 · g−1 · kPa−1 occurred (34). In another study, ΔCBV was calculated in relation to changes in MAP in newborn lambs to be 0.07 mL · 100g−1 · mmHg−1 (35). Compared with these values, a mean change in CBV of 0.11 to 0.13 mL · 100 · g−1 and changes for individual lambs in CBV up to 0.26 mL · 100 · g−1 as found in our study could be important. Changes between the maximal decrease in CBV and consecutive maximal increase in CBV are more profound. Finally, the value of CBV in lambs has been found to be 2.5 mL · 100 · g−1 (36). Changes in CBV found in our study are up to 10%. Besides the magnitude, changes in HbD and CBV occur very acute, which might be an additional risk factor for cerebral lesions by itself. Some authors describe predominance of cerebral lesions in especially the right hemisphere of the brain, whereas in other studies, this predominance, possibly related to ligation of the right carotic artery and right internal jugular vein, could not be confirmed (3743). To study differences between both hemispheres, we used two receiving NIRS optodes. However, changes in HbD and CBV were not different between the left and right hemisphere of the brain.

Cerebral hemorrhage and ischemia are generally considered to be caused by preECMO as well as ECMO-related factors. We showed a significant relation between bladderbox alarms and changes in cerebral oxygenation and hemodynamics. This might play a role in the development of neurologic sequelae. As changes in HbD and CBV are more extended when bladderbox alarms are resolved by administration of normal saline, it would be interesting to further study the effects on cerebral oxygenation and hemodynamics of intravascular volume administration by itself as well. The results of our study are limited to ECMO systems with a rollerpump and a bladderbox and might support the use of centrifugal pumps, systems without a bladderbox or ECMO strategies with reduced bladderbox alarms. However, more research is needed for other systems as well to study, for example, the effects of less-controlled drainage from the right atrium of the patient and cerebral oxygenation and hemodynamics.

On the basis of this study, we conclude that bladderbox alarms during treatment with va-ECMO result in statistically significant fluctuations in cerebral oxygenation and hemodynamics. This might be an additional risk factor for the occurrence of intracranial hemorrhage and ischemia.