Nature Publishing Group, publisher of Nature, and other science journals and reference works NATURE.COM NATURE NEWS NATUREJOBS NATUREEVENTS ABOUT NPG
Help Nature.com site index  
Journal of Perinatology
SEARCH     advanced search my account e-alerts subscribe register
Journal home
Advance online publication
Current issue
Archive
Press releases
For authors
For referees
Contact editorial office
About the journal
For librarians
Subscribe
Advertising
naturereprints
Contact NPG
Customer services
Site features
NPG Subject areas
Access material from all our publications in your subject area:
Biotechnology Biotechnology
Cancer Cancer
Chemistry Chemistry
Dentistry Dentistry
Development Development
Drug Discovery Drug Discovery
Earth Sciences Earth Sciences
Evolution & Ecology Evolution & Ecology
Genetics Genetics
Immunology Immunology
Materials Materials Science
Medical Research Medical Research
Microbiology Microbiology
Molecular Cell Biology Molecular Cell Biology
Neuroscience Neuroscience
Pharmacology Pharmacology
Physics Physics
Browse all publications
 

July/August 2002, Volume 22, Number 5, Pages 360-366

Table of contents    Previous  Abstract  Next   Full text  PDF

Original Article

Reliability of Conventional and New Pulse Oximetry in Neonatal Patients

William W Hay Jr, Donna J Rodden, Shannon M Collins, Diane L Melara, Kathy A Hale and Lucy M Fashaw

Department of Pediatrics, The Neonatal Clinical Research Center, University of Colorado Health Sciences Center, Denver, CO, USA

Correspondence to: William W. Hay, Jr., MD, Box B-195, University of Colorado Health Sciences Center, 4200 East Ninth Avenue, Denver, CO 80262, USA

Abstract

OBJECTIVES: Pulse oximetry is widely used in the NICU, but clinicians often distrust the displayed values during patient motion, i.e., questionable oxygen saturation (SpO2) and pulse rate (PR) values. Masimo Corporation (Irvine, CA) has developed pulse oximetry with claims of resistance to sources of interference. To test this premise, we compared the performance of the Masimo SET pulse oximeter to a conventional device, Nellcor N-200, and then with three other new-generation pulse oximeters, Nellcor N-395, Novametrix MARS, and Philips Viridia 24C.

STUDY DESIGN: We studied 26 nonsedated NICU infants who were on supplemental oxygen and/or mechanical ventilation. ECG heart rate (HR) from a bedside monitor and SpO2 and PR from the two pulse oximeters were captured by a PC for a total of 156 hours. The ECG HR and pulse oximeter spectral waveform were analyzed at alarms for hypoxemia (SpO2£85%) and/or bradycardia (HR£80 bpm). We then compared the performance of the Masimo SET to three other new-generation pulse oximeters, Agilent Viridia 24C, Nellcor N-395, and Novametrix MARS, in a similar population of seven infants for a total of 28 hours. We added to the test criteria the ability of the various pulse oximeters to track acute changes in HR.

RESULTS: Compared with Nellcor, Masimo SET had 86% fewer false alarms, which also were shorter in duration, resulting in 92% less total alarm time. Masimo SET also identified nearly all bradycardias versus 14% for the Nellcor. Compared with the new-generation pulse oximeters, false desaturations, data drop-outs, and false bradycardias were lowest for Masimo SET, as was the capture of true desaturations and bradycardias. Notably, the new-generation devices differed greatly in their ability to detect changes in HR (i.e., the frequency of frozen PR during times of ECG HR change was 0, 6, 11, and 46 for Masimo, Nellcor, Philips, and Novametrix, respectively).

CONCLUSIONS: Masimo SET pulse oximetry recorded markedly fewer false SpO2 and PR alarms and identified more true hypoxic and bradycardic events than either conventional or other new-generation pulse oximeters. Masimo SET also most closely reflected the ECG rate irrespective of accelerations or decelerations in HR.

SPECULATION: Routine use of Masimo SET pulse oximetry in the NICU could improve clinician confidence in the parameter leading to more judicious titration of oxygen with possible reductions in hypoxic (e.g., pulmonary hypertension) and hyperoxic (e.g., retinopathy of prematurity) pathology. Additionally, a more trustworthy technology should equate with fewer confirmatory arterial blood gas analyses (less blood loss), and faster weaning from the mechanical ventilation (less chronic lung disease). Journal of Perinatology (2002) 22, 360-366 doi:10.1038/sj.jp.7210740

July/August 2002, Volume 22, Number 5, Pages 360-366

Table of contents    Previous  Abstract  Next   Full text  PDF

Privacy Policy © 2002 Nature Publishing Group