Introduction

Diffusion weighted imaging (DWI) is a widely utilized magnetic resonance imaging (MRI) technique that provides in vivo information on the random Brownian motion of free water and is quantified by the calculation of apparent diffusion coefficients (ADCs). The contrast in the resulting images results from differing signal loss from hydrogen proton movement1,2. DWI has extensively been used in tumor grading3,4, early detection or differential diagnosis of benign and malignant tumors5,6,7,8, distinguishing viable and necrotic tissues9,10,11,12, assessment of organ function13, prediction of pathologic outcomes14,15 and evaluation of tumor response to therapies16,17,18,19.

Unfortunately, DWI is sensitive to inhomogeneity of the magnetic field and vulnerable to artefacts with long echo train time20, especially for when used on small experimental animals. In previous studies the uniformity of the magnetic field was generally improved either by reducing diameter of the coil or by increasing the magnetic field intensity; however, most high quality anatomical images were more concentrated in solid organs and few in hollow organs21,22. Moreover, no relatively detailed studies exist comparing methods for improving DWI. In addition, the poor inhomogeneity of the magnetic field at the air-tissue interface produces severe susceptibility artefacts1,20,23, with higher magnetic field intensity resulting in more severe artefacts. Some researchers have attempted immersing experimental animals in liquid to isolate such air-tissue interface effects and achieved interesting results24. However, if the liquid is flowing, the imaging position of experimental animals and coil size are significantly restricted, limiting further improvement in DWI quality. It is also difficult to achieve constant temperature for liquids, which proves a great challenge for the longitudinal observation and scanning duration of experimental animals.

On the other hand, diffusion-weighted images with different b values reflect the different restricted degree of water molecules movement. This signal loss of water molecules can be obtained quantitatively with two images of different b values10. Due to the serious deformation, artefacts and poor co-registration of diffusion-weighted images with different b values, the accuracy of the ADC value is inevitably affected. Therefore the extension of some research results is limited due to the poor measurement repeatability of ADC values25.

To solve the above problems, our research chose the alginate-water mixture together with ultrasonic coupling medium for shaping and isolating the mouse from surrounding air for improved DW imaging. And by comparing the quality of diffusion-weighted images and ADC values of different organs between the alginate and the control group, the improved diffusion-weighted images for the quantitative research of ADC values of experiment animals were obtained. We found the mouse model of orthotopic gastric cancer to be useful in demonstrating the efficacy of this imaging modification in identifying and clarifying internal disease relevant to human conditions.

Results

Orthotopic implantation

Two nude mice died before the first MR scan. Tumours were detected in the stomach of all fourteen remaining nude mice by MRI (red arrows, Figure 1), especially in DW imaging. The subsequent six MR scans were well-tolerated by all fourteen mice.

Figure 1
figure 1

MR images of gastric cancer for orthotopic mouse model.

(A) T2 axial image of orthotopic gastric cancer (red arrows), (B–C) T1 axial image and DW images (with alginate mould) of the same slice. (D–E) T2 coronal and sagittal image of the orthotopic tumour.

DW imaging with and without the mould

The container and mouse with and without the alginate mould are shown in Figure 2. The DW images obtained from the two groups with and without the alginate mould, scanned from 7 to 42 days after orthotopic implantation, are shown in Figure 3. The scoring system utilized is shown in Table 1. The artefact and distortion of DW images were greatly reduced in the alginate mould group and the two observing radiologists were in agreement concerning image rating (kappa coefficient = 0.84, p < 0.01). In the group imaged with the alginate mould, the average score of the DW images in outline, distortion and artefacts were 2.71, 2.92 and 2.90, while the average score of the control group were 1.53, 1.51 and 2.16, respectively (Table 2). From these results, we determined that distortion of DW images was best inhibited in the alginate mould group. As the total score of the mould and control groups were 8.53 and 5.20, respectively, the imaging quality of the alginate mould group was significantly improved (p < 0.001). The total score distributions in the two groups are shown in Figure 4. For both DWI and T2 images, the signal-to-noise ratios (SNR) of the tumour between the two groups were not significantly different (p = 0.934 for DWI and p = 0.948 for T2 sequence) (Supplemental Table 1).

Table 1 Diffusion weighted imaging quality scoring system
Table 2 Scoring for DW images
Figure 2
figure 2

Alginate mould preparation.

(A) The mould for pouring alginate powder mixed with warm water to form the imaging mould. (B) The solid alginate-mouse chamber 30–60 s after liquid alginate was poured around the mouse. (C) The alginate-mouse chamber shaped for the wrist coil.

Figure 3
figure 3

Diffusion weighted images of mouse with gastric cancer from 7 to 42 days after orthotopic implantation.

Figure 4
figure 4

Stacked percentage bar plot for the total image quality scores for the control and alginate mould groups.

Score 8–9 represented high-quality images; Score 6–7 represented moderate-quality images; Score 3–5 represented poor-quality images.

ADC stability using the alginate mould

The ADC values and coefficient of variation (CV) in the normal liver of each mouse in both the mould and control groups are provided in Table 3. The mould group exhibited much less variation (CV = 0.07) compared with those obtained without use of the mould (CV = 0.27), p < 0.001 (Figure 5).

Table 3 Normal liver ADC value (×10−3 mm2/sec)
Figure 5
figure 5

Box plot for the coefficient of variation of ADC in normal liver.

The ADC of orthotopic gastric cancer in vivo

From the above results, we determined that the mould group provided not only clearer and less distorted images, but also measurements with high repeatability. Therefore, we used data obtained from the alginate mould group to assess ADC changes of the orthotopic tumour. We used alginate moulding in the serial monitoring of tumour growth by DWI of orthotopic nude gastric cancer. All 7 orthotopic transplanted nude mice survived subsequent seven serial MR scans. The maximum tumour areas increased rapidly during the first 28 days after which growth decelerated (Figure 6A). However, ADC values gradually decreased within 10–30 days after transplantation and these values increased slightly with tumour growth (Figure 6B). No significant correlation was detected between ADC and the maximum area of the orthotopic tumour (r = 0.06, P = 0.733).

Figure 6
figure 6

Longitudinal ADC value and the maximum orthotopic tumour area.

(A) Longitudinal ADC value of gastric cancer for each mouse. (B) The average ADC value and average maximum area of the orthopic tumours for all seven mice in each group.

Discussion

Restricted by sequence design, DWI images exhibit obvious deformation and artefacts, especially for small experimental animals. In addition, the accuracy of ADC value measurement can also be affected. Our study attempted to improve the disadvantages of DWI using homemade scanning mould material. Commonly used dental alginate with superior water solubility, fast solidification, high safety and appropriate flexibility was applied to pre-positioned experimental animals26,27. Our goal in using such an alginate mould was to suppress DWI image deformation and artefacts, estimate DWI image quality and the impact of ADC values and subsequently choose improved DWI images for ADC value quantitative research.

In the subjective evaluation of DWI imaging in our study, alginate embedding reduced the density difference at the mouse-medium interface compared to a mouse-air interface. The uniformity of the magnetic field around the mice was indirectly improved and image distortion was suppressed. The DWI deformation occurred primarily due to poor uniformity of the surrounding magnetic field20,28. Image distortion was inhibited to some extent by increasing the magnetic field intensity or reducing the inside diameter of the coil, especially for deep solid organs5,29,30. However, such methods had no inhibitory effects on artefacts caused by subject conditions which usually occurred between regions of significant differences in magnetic susceptibility (animal - air interface), of the chemical shift region (water - lipid interface) and physical movement region31,32; therefore, DWI image artefacts of superficial parts of the body were still evident. The susceptibility artifacts were somewhat reduced when mice were scanned in water in a previous study24, but the particular vessel used for placing mice in water required the animals to be scanned with a coil with large inside diameter, limiting further enhancement in image quality.

Our study took advantage of the fluidic character of the alginate/water combination to effectively reduce the tissue-air interface, which minimized the difference in magnetic susceptibility between adjacent objects and lessened the surrounding artefacts. Secondly, the alginate mould could be easily shaped to fit different scanning coils, improving the flexibility of its use. In addition, the alginate mould was inexpensive, easily acquired and made. It is also more convenient to use for maintaining animal body temperature during imaging than immersing the animal in water. The alginate mould could enhance the tolerance of the mice to the experimental conditions and increase the feasibility of extended observation durations.

Our imaging results were similar to previous clinical DWI research in the objective evaluation of DWI image quality. If motion artefacts are suppressed, higher intensities are revealed in the ROI and with higher signal-to-noise ratio (SNR) and we found that the alginate mould suppressed deformation, magnetic susceptibility artefacts and improved the SNR of images32,33. Secondly, coil load is another factor that could affect the SNR of an image. By using the alginate mould to contain the animal, the volume of the imaged object could approach the optimal value of the coil, which would enhance capture of signal changes and improve signal intensity. In conclusion, our study shows that distortion and artefacts of DWI were suppressed and the repeatability of ADC values was improved by use of the described alginate mould in a mouse model of orthotopic gastric cancer.

Methods

Cell line and orthotopic stomach tumour implantation

All procedures were performed according to the protocol approved by our institutional review board. A human gastric cancer cell line, BGC-823, was maintained in RPMI 1640 media (Beijing Solarbio Science & Technology Co., Ltd, Beijing, China) supplemented with 10% fetal bovine serum (FBS, Gibco, New York, USA) at 37 ± 2°C, pH 7.2, in a humidified atmosphere of 95% air and 5% CO2. Cells were passaged and expanded by trypsinization followed by replating every 2–3 days. The culture media was changed every 2 days. For mouse inoculation, cells in log-phase growth were harvested by trypsinization and medium containing 10% FBS was added. Following centrifugation for 5 min at 1000 rpm, cells were resuspended in phosphate-buffered saline (PBS) and kept at 37 ± 2°C until inoculation.

Anesthetized female athymic 7 week old BALB/c nude mice, maintained under SPF (Specified Pathogen Free) conditions in our hospital's animal facilities, received 5.0 × 106 BGC-823 cells suspended in a volume of 0.2 ml sterile DPBS by subcutaneous injection below the dorsal flank. The mice were observed for the following 3 weeks, after which tumour incidence was determined. Mice were then sacrificed and tumours were removed. The tumour parenchyma was cut into small pieces less than 1 mm3.

Under anaesthesia of 0.5% amobarbital in 0.1 ml, a small median abdominal incision was made for each of sixteen mice and then purse-string sutures were performed on the anterior wall of stomach. The sixteen mice were randomly divided into two groups, an alginate group and control group, in this study. The surgeon made a serosal incision within the purse of the anterior wall and a small piece of tumour was implanted under the subserosal layer. The serosa and purse were both sutured to prevent the tumour from entering the peritoneal cavity. The stomach was then returned to the peritoneal cavity and the abdominal wall was closed with absorbable suture.

MR imaging

Alginate-mouse chamber and nude preparation

As a prerequisite, we developed a mouse-imaging chamber adapted to our wrist joint coil, enabling the head of the mouse to remain outside the chamber for breathing. The alginate (Zhermack, Badia Polesine, Italy) used in our study is a common type often used by dentists for dental impressions. It is commercially available as a ready-to-use powder and solidifies within 2–3 min when mixed with water to a rubbery consistency. We used approximately 150% of the amount of warm water (40–42°C) to reduce viscosity of the mould for improved flow ability and an increased time span until solidification. We then positioned the mouse into the chamber as follows:

For the alginate group, the nude mice were anesthetized by 0.5% amobarbital in 0.1 ml and then overlaid with coupling medium commonly used in ultrasonography in our hospital. We positioned the anesthetized mouse in the chamber with the head outside through the air vent and then filled the chamber with alginate/water mixture. Finally, the chamber was immobilized in the wrist coil. Control group mice were prepared for MRI scanning with the same procedure as the alginate group, but without the use of coupling medium and alginate filling.

Imaging procedures

To avoid incisional infection of the stomach, the first MR imaging was performed on the seventh day after orthotopic implantation. The 14 total nude mice underwent MR imaging at 7, 10, 14, 21, 28, 35 and 42 days after orthotopic implantation. MRI scanning was performed on a clinical 3T scanner (Discovery MR750 3.0T, GE Healthcare, Milwaukee, WI, USA) using an 8-channel HD Wrist Array coil for the mouse abdomen. A total of 7 scans were performed at each time point for each animal for a total of 49 obtained images for each of the control and alginate groups. Prior to DW imaging, all mice underwent examination with routine gastric MR protocols to localize the tumours and provide morphologic details. The MR sequencing parameters are detailed in Table 4.

Table 4 MRI scanning parameters

Image analysis and scoring system of DWI imaging quality

All DW images were analysed independently by two experienced radiologists (X.Y.Z. and L.T., with 6 and 10 years' experience in MR, respectively), working on a workstation (AW4.4; GE Medical Systems, Waukesha, WI, USA) and using a scoring system based upon the following. The total score equalled the sum score of the three indexes shown in Table 1. Score 8–9 represented high-quality images; Score 6–7 represented moderate-quality images; Score 3–5 represented poor-quality images.

ROI designation and quantification

The regions of interest (ROI) were drawn by the consensus between two radiologists.

Liver: ROIs were manually drawn cross images of the liver with an area of approximately 5 mm3 excluding the vessels and bile ducts. The ADC of each ROI was recorded and an average ADC was obtained (Figure 7).

Figure 7
figure 7

ROI in the orthotpic gastric cancer and normal liver.

(A–B) ROI in the orthotopic tumour and its ADC value. (C–D) ROI in the normal liver and its ADC value.

Tumour: After identification of an orthotopic gastric tumour, ROIs were manually drawn with the same areas of the tumour in T2 images. The ADC of each ROI was recorded and an average ADC was obtained for each lesion. The longest diameter and the vertical diameter of the tumour were also recorded. We used the product of these two diameters as the maximum area of the tumour (Figure 7).

Statistical analysis

All statistical analysis was conducted with SPSS statistical software (Version 15.0. Chicago IL, USA). Continuous data were expressed as median and range or mean ± SD. Coefficient of variation (CV) was used to assess the degree of variation. A two-tailed student's t-test was used to compare values between the alginate and control groups. One-way analysis of variance (ANOVA) was utilized to detect differences in CV of hepatic ADC value between animals across within the alginate and non-alginate groups. A Mann–Whitney U test was used to compare the sum score for image quality between the alginate and control groups. A bar chart was drawn to display the distribution of the sum score for image quality. The changing trends of ADC values and the maximum area of orthotopic tumours were illustrated using a scatter diagram. The Kappa coefficient was calculated to measure the inter-rater agreement (0–0.20 as slight, 0.21–0.40 as fair, 0.41–0.60 as moderate, 0.61–0.80 as substantial and 0.81–1 as perfect agreement). The correlation between ADC value and the maximum area of tumour was assessed with Pearson's correlation coefficient. A p value < 0.05 indicated statistical significance.