Introduction

The introduction of both high and slow speed dental handpieces has revolutionised dental treatment, allowing for significant improvements in restorative dentistry and patient care. Their use has, however, been accompanied by significant controversies regarding their potential for cross-infection1,2,3 and the methods available for their effective disinfection or sterilisation.4,5,6

Although autoclavable handpieces became more readily available in the 1980s, the main method for decontaminating handpieces continued to be disinfection (of the external surfaces) rather than sterilisation.7 In the 1990s, particularly following the highly publicised case of HIV transmission in a Florida dental practice, attention was drawn to the potential for cross-infection by dental handpieces. Laboratory studies demonstrated aspiration of both dye and microorganisms into the air and water lines and the handpiece chamber, and their subsequent recovery following re-use of the high speed handpiece.1,8

Concerns over handpiece contamination relate to the accumulation of particulate matter (débris from restorative materials and dental hard tissues),9 microorganisms from the oral cavity (bacteria, fungi and viruses),2,10 microorganisms from the water and airlines11,12 and collection of human tissue (blood, saliva)1 within the handpiece chamber, turbine blades, gearing, air and water lines. Dental handpieces are extremely difficult to clean, inspect and sterilise due to the small size and length of lumens, intricate working parts (which require lubrication) and their inability to be readily dismantled. Current recommendations are that dental handpieces should be sterilised between patients.13,14,15 For sterilisation to be effective it should be performed following cleaning which, if inadequately performed, will compromise the sterilisation process.16 Efficient cleaning and lubrication are also important determinants of handpiece functionality and longevity.17 The adherence of dental practitioners to the recommendations for sterilising dental handpieces has, to date, been investigated using postal questionnaires with varying response rates.18,19,20 In order to address the shortcomings of previous questionnaire surveys, we have previously described a methodology for a large observational study of decontamination of dental instruments in general dental practice.21 The aim of the present paper is to report in detail the methods used for decontamination of dental handpieces in general dental practices and the procedures in place for purchase and maintenance of handpieces, as recorded using an observational method.

Materials and methods

Survey methodology

The survey was conducted according to the previously described method.21 The study population comprised all general dental practitioners in Scotland with a National Health Service (NHS) list number. The list was used for randomly selecting practitioners to survey. A total of 184 surgeries were surveyed, with usable data obtained from 179 surgeries.

Data collection

Each surgery was surveyed by an infection control/decontamination expert and an experienced dental practitioner. The surveying team interviewed both the dental practitioner and dental nurse. This survey was concerned with collecting data relating to dental handpieces, in particular: the selection, procurement, cleaning, lubrication, sterilisation, maintenance and replacement of air turbines and air rotors. The decontamination processes undertaken by the surgery nurse were viewed directly by a member of the survey team. All data were recorded onto data collection forms prepared for automated reading.21 The survey visits ran from January 2003 until the end of March 2004.

Technical requirements and guidance

The information recorded on the data collection forms was based on a number of technical requirements and guidelines including SHTM 2030,22 HTM 2030,23 the Glennie Framework,24 the Medical Devices Regulations 2002,25 the Health and Safety at Work Act 197426 and Management of Health and Safety at Work Regulations 1992.27

Results

Dental handpiece purchasing and maintenance

Of the 179 dental surgeries surveyed, 95% (n = 170) had no written policy for purchasing handpieces. The member of surgery staff with responsibility for handpiece purchase was the principal/partner of the practice in 97% (n = 174) of surgeries, 102 (59%) of whom did not seek information on handpiece decontamination before purchase. In the vast majority of surgeries (96%; n = 172), all new handpieces were decontaminated before first use.

The most common air turbine brand held by the dental surgeries surveyed was Kavo (Table 1). Most surgeries (77%; n = 138) had purchased a new air turbine handpiece within the last 12 months and the majority of surgeries (62%; n = 111) had serviced their air turbines within the last 12 months. The most common brand of slow speed (contra-angle) and straight handpiece was NSK (Table 1). Most surgeries (88%; n = 157) had no written maintenance policy for handpieces and only a minority (27%; n = 49) performed their own handpiece maintenance.

Table 1 Types of air turbines, electric motors and handpieces used in surgeries

Handpiece cleaning

Virtually all surgeries cleaned handpieces before disinfection or autoclaving (99%; n = 177). The most common method of handpiece cleaning used was to wipe the external surface with a cloth impregnated with disinfectant (Table 2). In a small number of surgeries (3%; n = 5), irrigation of the internal lumen of the handpiece was undertaken. Following cleaning, the majority of surgeries (92%; n = 165) inspected the external surface of the handpiece for cleanliness and 25% (n = 44) of surgeries dried the handpiece before sterilisation, usually by wiping with a paper towel or leaving to air dry in the surgery.

Table 2 Methods reported for cleaning handpieces

Lubrication of handpieces

Staff in some surgeries (20%; n = 36) lubricated their handpieces following use and before cleaning. Most lubricated handpieces after cleaning but before sterilisation (91%; n = 162), although a number (24%; n = 42) lubricated again after sterilisation. Most surgeries used the handpiece manufacturers' recommended lubricant, none of which were water soluble (source: manufacturers' material safety data sheets and websites).28,29,30

Handpiece autoclaving

The policy in virtually all of dental surgeries (97%; n = 174) was to autoclave all handpieces after use, with most (89%; n = 160) reprocessing handpieces in a bowl and instrument steriliser. Staff in some surgeries (21%; n = 38) were placing their handpieces in paper pouches before sterilisation in a bowl and instrument steriliser.

Storage and use of handpieces

Within the surgeries most handpieces were stored in a dry and secure location, although a number (31%; n = 55) were stored on open work surfaces. Before use, dentists in most surgeries (71%; n = 127) discharged their handpieces to dissipate excess lubricant and some (31%; n = 55) discharged their handpieces for several seconds after use on each patient to reduce the likelihood of aspiration of patient material into the handpiece. A minority of surgeries (20%; n = 36) had a dedicated handpiece for surgical procedures.

Discussion

The results of this survey provide a valuable insight into handpiece decontamination in general dental practice in Scotland. The data were obtained by observing and interviewing dentists and dental nurses in a large number of practices, as opposed to data collection using postal questionnaires.18,19,20 It was reassuring to find that the vast majority of dental surgeries now autoclave dental handpieces after use. Before the intense media coverage of handpiece sterilisation in the early 1990s, large numbers of dentists disinfected rather than sterilised their handpieces. Studies of dental practitioners in England, reported in 1995, indicated that between 46% (response rate of 53% of 500 practitioners)18 and 77% (response rate of 74% from 312 practitioners)19 of practitioners were sterilising handpieces between patients and our data suggest that compliance with this policy has continued to improve. Nevertheless, a more recent survey of 700 GDPs in the Republic of Ireland (response rate of 40%) reported that 53% did not autoclave their handpiece between patients, indicating continued reluctance in some areas to embrace the need for handpiece sterilisation.20

Dental handpieces present a particular problem for decontamination because they have both external and internal surfaces that become contaminated during clinical use. When the air and water delivered to the handpiece during operation are switched off, the temporary negative pressure induced is likely to result in retraction of contaminated fluid and air along the air and water lines into the handpiece.1,2 As a result, the lumen and the turbine blades of the handpiece become contaminated with oral secretions, which may include restorative materials and biological debris such as microorganisms, saliva, blood and tissue debris. In addition, the air and water lines, turbine and drive shafts can be contaminated from the source of the compressed air and water supply. Dental handpieces are used for a wide variety of dental interventions, some of which can be classified as critical,15 and as such handpieces should be sterile when used. Guidance from a number of organisations has recommended that all dental handpieces should be sterilised between patients.13,14,15

In order to facilitate handpiece sterilisation and improve the longevity of the handpiece, a critical control point in the decontamination process is effective cleaning.31 At present there do not appear to be any validated cleaning processes for dental handpieces. Further challenges of handpiece decontamination include the necessity for air removal from the lumens to allow steam penetration, necessitating use of a forced air removal stage during the sterilisation cycle, ie using a vacuum steriliser.13,31,32 The wrapping of dental handpieces and processing in a non-vacuum steriliser will also impede the penetration of steam into the device.32 This practice was identified in a number of practices in this survey. At present, vacuum steam sterilisers are not widely used in dental practice in the UK.33

The stage in the decontamination cycle at which handpiece lubrication was undertaken varied considerably, suggesting some confusion in this area. The handpiece manufacturers recommend lubricating after cleaning and before sterilisation, although the data from the current study indicates that non-water soluble lubricant is quite commonly supplied by the manufacturers. The use of non-water soluble lubricants can impair steam penetration into the surfaces of the handpiece and there is the possibility that such oil-based lubricants could prevent the steam from killing bacterial spores.4,16,31

Thus, there are many technological challenges to overcome before general practitioners have the facility to introduce processes that will reliably deliver sterile handpieces. In the interim, it appears that cleaning the external surface and autoclaving handpieces after each patient is a regular feature of infection control processes in general dental practice.

In conclusion, using a robust, observational method of data collection, the policy in the majority of dental surgeries examined is to autoclave handpieces between patients as a routine element of the infection control protocol. This is a very positive development in recent years and one which significantly enhances patient safety. However, a number of technological issues remain to be resolved before manufacturers' claims that handpieces are sterile can be substantiated. Foremost among these is a validated cleaning process coupled with appropriate lubrication of the internal components, development of which are the responsibility of handpiece manufacturers. An effectively cleaned handpiece would then require to be sterilised under vacuum conditions to satisfy criteria for such devices to be labelled as sterile. Use of a vacuum steriliser would have additional advantages of allowing handpieces to be wrapped before and during sterilisation, thereby preventing the possibility of environmental recontamination and clearly distinguishing processed from unprocessed instruments. However, the widespread introduction of vacuum steam sterilisers into general dental practices remains a contentious issue due to increased capital, revenue costs and technical requirements for periodic testing.