Background & Summary

Access to data is playing an increasingly central role in research. Indeed, many research fields rely almost entirely on the availability, curation, and sharing of global data sources provided through data repositories, which have nowadays become a vital part of research infrastructures.

In this context the NFFA–EUROPE project (www.nffa.eu), an advanced European distributed research infrastructure dedicated to nanoscience, has, as one of its key tasks, the objective to design effective means of data sharing at the nanoscale in Europe. The project involves twenty European research partners, and more than 150 different experimental/computational instruments and techniques; the volume of scientific data produced by the project activities is aimed to be hosted in the NFFA-EUROPE Information and Data Repository Platform (IDRP)1 being open and fully available to the scientific community well defined data policy. The IDRP is complemented by a set of data analysis services that has been steadily growing since the beginning of the project.

Among the available instruments of the NFFA-EUROPE catalogue, Scanning Electron Microscopes (SEMs) are among the mostly requested ones from users, and are available at ten NFFA-EUROPE sites. SEM is indeed a routinely used characterization technique able to provide information about the samples' topography and composition down to a few nanometers resolution by scanning a focused electron beam onto the sample surface.

The first data analysis achievement is a service offered by the NFFA-EUROPE IDRP consisting of an automatic tool for image recognition to store, classify and label SEM images: we employed a supervised machine learning algorithm using deep convolutional neural networks for recognition of SEM images2. In order to train the network, it was necessary to provide a labeled training set, i.e. a set of SEM images properly classified by humans into categories.

In this paper, we describe the datasets presented in ref. 2, together with the methods employed to classify and validate them: the primary dataset used for neural network training (SEM dataset, Data Citation 1), a detached smaller dataset in which a hierarchical structure has been introduced (Hierarchical dataset, Data Citation 2), and two datasets which contain ~3,000 newly classified images in addition to the SEM dataset, according to different validation criteria (Majority dataset, Data Citation 3, and 100% dataset, Data Citation 4), as described in the Technical Validation section.

The images were generated by the SEM located at the CNR-IOM (Istituto di Officina dei Materiali, www.iom.cnr.it) in Trieste, Italy. In total, close to 100 users have generated roughly 150,000 images over five years at the facility.

From this sample, an initial batch of ~20,000 images was manually classified by nanoscientists into 10 categories which spanned a broad range of nano-objects, from 0D particles, 1D nanowires, 2D films and patterned surfaces, and 3D pillars and microelectromechanical devices.

This classified dataset (Data Citation 1) is intended to be functional to a number of possible implementations. Primarily, it can set the basis for a more complete labeled dataset for supervised machine learning which can be further expanded with more SEM images, thus allowing the development of an even more accurate and precise SEM image recognition software. Some categories can be split into multiple subcategories, as further discussed in the text.

On the other hand, image processing techniques are being increasingly applied to microscope images3,4. This dataset allows such programs to be tested on a large set of images belonging to a particular category for specific purposes.

In general, we are confident that the accuracy, diversity, and hierarchical structure of our SEM datasets can offer additional opportunities to researchers, connecting the image recognition and the nanoscience communities.

The paper is organized as follows: The Methods section describes how we retrieved and classified the SEM images; the Data Records section contains an overview of each dataset associated with this work, including the repository where they are stored, the contained data files, and their formats; in the Technical Validation section, we present the adopted validation procedure and the obtained results. Finally, the Usage Notes section offers a summary of the datasets and of the aims they were designed at.

Methods

Most of the methods described in detail in this section have been presented in our previous publication2.

SEM imaging

All the images collected in the datasets were acquired by nearly 100 scientists and researchers over five years using a ZEISS SUPRA 40 high resolution Field Emission Gun (FEG) SEM, located in the TASC laboratories of the CNR-IOM in Trieste. The microscope can be operated with 0.1–30 kV acceleration voltage and a 4 pA-10 nA probe current with a nominal resolution of 1.3 nm at 15 kV. It is equipped with an Everhart-Thornley Secondary Electrons Detector, a Back-scattered Electrons Detector, and a High efficiency In-lens detector, the latter providing an increased signal-to-noise ratio in image acquisition. The microscope is also equipped with a 4-Channel High-Angle Annular Scanning Transmission ElectronMicroscopy (STEM) detector allowing detection of transmitted electrons and with an Oxford Aztec Energy Dispersive X-ray Spectroscopy (EDS) system and a X-act 10 mm Silicon Drift Detector (SDD) for compositional analysis.

SEM image classification

The SEM images were classified based mainly on the dominant shape and structure of the imaged object. The dimensionality of the nanostructure (0D, 1D, 2D or 3D) was used as a first criterion, being the visual benchmark of the nanostructure itself. In this classification, 0D objects refer to particles which can be either dispersed and individually identified over the image area, or packed together into a clustered structure. 1D objects are referred to as nanowires, rods, or fibres. These structures can be either entangled in bundles or mutually aligned with respect to each other. 2D structures refer to films and coatings on a surface, and include a wide range of cases differing from each other for the surface roughness. 3D structures can refer to pillars or other micro-devices such as microelectromechanical systems (MEMS), typically fabricated by lithography techniques. A schematic view of this classification is illustrated in Fig. 1.

Figure 1: Schematic view of nanostructure classification based on dimensionality from 0D to 3D.
figure 1

(a) Particles (b) Clusters (c) Nanowires (d) Fibres (e) Films (f) Patterned surfaces (g) Pillars (h) MEMS devices.

The dimensionality discussed above provided the main basis for categorization, but was not exhaustive. After investigation of the SEM images database and discussions with SEM scientists, additional categories such as biological samples and tips (these last mostly related to atomic force microscopes) were introduced to enlarge the spectrum of the possible categories. The full set of selected categories is illustrated in Fig. 2, with a representative image for each of them, while the breakdown of the number of images in each category is shown in Table 1. These ten categories cover a broad range of SEM images, and they are applicable to scientists working in different areas of nanoscience. At the same time, they were chosen to be as more distinct from one another as possible.

Figure 2: Representative images for each of the categories chosen for the SEM dataset.
figure 2

(a) Tips (b) Particles (c) Patterned surfaces (d) MEMS devices and electrodes (e) Nanowires (f) Porous sponge (g) Biological (h) Powder (i) Films and coated surfaces (j) Fibres. The dimensionality of nanoscience objects provided the basis for the choice. Other categories, such as Biological and Tips, were added, as they were common images found in the SEM database.

Table 1 The original SEM annotated dataset.

Preliminary degree of hierarchy

A second dataset of 1,038 images (Data Citation 2) was selected to explore a subtree for each category, when appropriate. We warn that this must be considered a preliminary work, due to the very small number of images in ~40% of the subcategory nodes. A summary of this initial work is shown in Table 2. The table must be read from the left column to the right one as root-to-leaf branches. Images in some categories did not present a sufficient variety to be split into subclasses. In these cases, the columns of Level 1 and Level 2 were left empty. However, we do not exclude the possibility to enrich the dataset with new subcategories in the future, when more images will be analyzed.

Table 2 Preliminary hierarchical structure established.

Code availability

The Python script used to generate the Majority dataset (Data Citation 3) and the 100% dataset (Data Citation 4) includes the following steps:

  1. 1

    read the database containing the validation results from the web classifier (sem-classifier.nffa.eu) we developed;

  2. 2

    extract the list of 400 images, distinguishing between the 200 images of the human set and the 200 images of the image recognition set;

  3. 3

    count the validation results by the users for each image;

  4. 4

    collect the images which have been validated by all the scientists and those which have been validated by the absolute majority of them (at least four out of seven);

  5. 5

    create the 100% and the majority datasets, starting from the SEM dataset (Data Citation 1):

    1. a

      to create the Majority Dataset (Data Citation 3), images from the human set which were not validated by the absolute majority of the scientists were removed from the SEM dataset, while those from the image recognition set which were validated by the absolute majority of them were added;

    2. b

      to create the 100% dataset (Data Citation 4), images from the human set which were not validated by all of the scientists were removed from the SEM dataset, while those from the image recognition set which were validated by all of them were added.

  6. 6

    The script can be provided by the authors upon request to Dr. Rossella Aversa (aversa@iom.cnr.it) or Dr. Stefano Cozzini (cozzini@iom.cnr.it).

Data Records

In this section, we provide detailed information on the published datasets, including a breakdown of each of them. In addition to the original SEM dataset, we present a preliminary hierarchical subset and two increased datasets obtained by classifying and validating other ~3,000 images. A schematic description to summarize the main details is reported in Table 3, while validation results are described in the Technical Validation section and summarized in Table 4.

Table 3 Summary of the datasets presented in this paper.
Table 4 Validation results, according to both majority and 100% criteria.

SEM dataset

To guarantee the reproducibility of the convolutional neural network training results results presented in ref. 2, we provide the reference to the original SEM dataset (Data Citation 1) used in the paper. The SEM dataset consists of 18,577 SEM images produced at the CNR-IOM (Trieste, Italy). The images are 1,024×728 pixels jpg files, classified into 10 categories in a folder structure (Table 1). The folders are individually stored as tar archives, for a total size of ~10 GB, and are publicly available for download (Data Citation 1).

Hierarchical dataset

The Hierarchical dataset is composed of 1,038 SEM images produced at CNR-IOM (Trieste, Italy). The images are 1,024×728 pixels jpg files, classified into 10 categories and 26 subcategories (as sub-trees) in a folder-subfolder structure (Table 2). The root folders are individually stored as tar archives, including the subfolders, for a total size of ~1 GB, and are publicly available for download (Data Citation 2). The preliminary results obtained from this dataset have been published in ref. 2.

Majority dataset

The Majority dataset consists of 21,272 SEM images produced at CNR-IOM (Trieste, Italy). The images are 1,024×728 pixels jpg files, classified into 10 categories in a folder structure according to the majority criterion (Table 4): classification labels have been checked by a group of nanoscientists by means of our SEM web classifier (sem-classifier.nffa.eu), and only those images which have been validated by the absolute majority of the group have been included in the dataset. The folders are individually stored as tar archives, for a total size of 11 GB, and are publicly available for download (Data Citation 3).

100% dataset

The 100% dataset is composed of 21,169 SEM images produced at CNR-IOM (Trieste, Italy). The images are 1,024×728 pixels jpg files, classified into 10 categories in a folder structure according to the 100% criterion (Table 4): classification labels have been checked by a group of nanoscience experts by means of our SEM web classifier (sem-classifier.nffa.eu), and only those images which have been validated by all the members of the group have been included in the dataset. The folders are individually stored as tar archives, for a total size of ~10 GB, and are publicly available for download (Data Citation 4).

Technical Validation

In this section, the procedure applied to validate the two increased SEM datasets and the statistical results obtained are presented. Since October 2017, we have labeled other ~3,000 images, thus increasing the dataset size by ~17%. As explained later, the method we applied ensures the same conclusions to be true also on the original SEM dataset (Data Citation 1).

To provide a reliable validation, 200 images (20 for each category) were randomly selected out to form a set (which we call human set) aimed to be judged by a group of 7 scientists, expert in SEM image interpretation. The human set was incremented by picking the same number of images (which we call image recognition set) classified by a neural network we had previously trained (see ref. 2 for details). We chose only image recognition results with a score ≥ 99%. We developed a web site (sem-classifier.nffa.eu), where the scientists were asked to access and validate the label assigned to each of the images or to suggest a different category, being unaware of which set (human or image recognition) they were coming from.

The dataset was validated at two levels: if a label of the human set reached the absolute majority of agreement from scientists, the image was validated (Majority dataset, Data Citation 3). This is a commonly adopted method, used e.g. to validate ImageNet5; alternatively, a more stringent criterion was to validate only the images for which the label reached the agreement of all the scientists in the group (100% dataset, Data Citation 4). By definition, the majority dataset contains the 100% dataset.

The results of the validation analysis are reported in Table 4. Looking at the last two rows, it can be noted that the images validated by the absolute majority of the scientists are almost 80% of the total (311 over 400), 168 (54%) of which come from the human set and 143 (45%) from the image recognition set. On the other side, roughly 50% of images (204 over 400) were validated by all the scientists; also in this case, more than 50% of validated images come from the human set (115 over 204).

It is worth noting that during the validation procedure 48 out of the 200 human labeled images were tagged by nanoscientists as “Not clear”. They should not be considered as misclassified images, rather they can be used to estimate the human uncertainty contribution to the validation error.

Based on the above results, the appropriate error on each of the two validated datasets can be expressed as:

(1) 1 validated tot ( nc - vnc ) tot

The second fraction represents the human uncertainty, where nc (not clear) are all the images tagged as not clear, and vnc (validated not clear) images are those ones which have been tagged as not clear but still validated, thus included in the dataset (i.e., tagged as not clear by the minority of the scientists). Of course, vnc=0 when considering the 100% criterion.

According to this formula, we have

(2) 1 168 200 ( 48 25 ) 200 = 0.045

for the Majority dataset, and

(3) 1 115 200 48 200 = 0.185

for the 100% dataset. These errors, being calculated on a randomly selected set of 200 images labeled by humans, can be safely considered as the validation error on the total datasets.

To conclude, the datasets we present in this paper have a 4.5% validation error according to the most commonly adopted majority criterion, and a 18.5% validation error when a more stringent 100% criterion is applied.

If a label was not validated (according to the first or second criterion), the corresponding image was removed. The additional validated images from the image recognition set were added to the respective datasets (Majority or 100%), to increase the number of examples and eventually enhance the performance of future neural network trainings. Nevertheless, they were not included in the original SEM dataset (Data Citation 1), to guarantee the reproducibility of the scientific results presented in ref. 2.

Usage Notes

We made available the four datasets presented in this paper through the EUDAT-B2SHARE service (https://b2share.eudat.eu/):

  1. 1

    The original SEM dataset(Data Citation 1), to reproduce the results of ref. 2;

  2. 2

    The Hierarchical dataset (Data Citation 2), which should be used carefully, being a preliminary version with few examples;

  3. 3

    The Majority dataset (Data Citation 3), based on the extended SEM dataset. Images from the human set which were not validated by the absolute majority of scientists were removed, while images from the image recognition set validated by the absolute majority of scientists were added, to further increase the dataset size;

  4. 4

    The 100% dataset (Data Citation 4), based on the extended SEM dataset. Images from the human set which were not validated by all of the nanoscientists were removed, while those from the image recognition set which were validated by all of the nanoscientists were added.

When applying machine learning techniques on these datasets, as a general rule, we suggest to use 90% of the images as the training set and 10% as the test set6.

All the datasets contain real SEM images as jpeg files, not as original tif format generated by the instrument. This implies that any scientific metadata associated to the measure is not present. By means of this choice, the authors make clear that the SEM image datasets are appropriate for the purposes of this study and in general for visual object recognition software research. Each dataset is therefore relevant as a whole, being the single images entirely detached from any specific information or scientific detail related to the displayed subject.