Introduction

Magnetic Resonance Imaging (MRI) is characterised by its flexible contrast with strong dependencies on multiple diverse tissue parameters, thus enabling sensitization of measurements to a wide range of properties in healthy tissue and pathology. One key limitation, however, is the intrinsic macroscopic resolution of MRI, while attempting to probe microscopic tissue features. Every voxel is composed of a variety of microstructural compartments with different properties and interactions. There is therefore a growing interest in quantitative MRI in which multiple measurements are combined with a suitable model to facilitate estimating microscopic parameters of interest1. Two main families of techniques in quantitative MRI are diffusion Magnetic Resonance Imaging (dMRI)2 and MR relaxometry.

The classical approach is to quantify one parameter or family of tissue parameters from one modality at a time. This has the benefits of limiting the complexity of the modelling required and reducing the amount of data that must be acquired, which is often an over-riding concern for in-vivo studies, particularly if clinical applicability is sought. However, joint analyses attempting to quantify multiple tissue parameters from multiple modalities at the same time have great potential, particularly if diverse physical processes are involved.

Diffusion MRI is a powerful modality for probing the microstructural architecture of biological tissues in-vivo2,3. The applied diffusion encoding gradients render this technique sensitive to the random motion of water molecules within the sample—thus informing on the underlying microstructure4. Varying the diffusion encoding strength (b-value) and direction (b-vector) extracts information at different length scales and orientations, hence enabling the estimation of microstructural compartment properties and neural fibre orientations5,6,7. Recent work has highlighted the limitations of microstructure imaging from dMRI alone. For example, disentangling intra- and extra-axonal compartment properties in brain white matter is degenerate and ill-posed8. An emerging trend is therefore to combine dMRI with other contrasts to disentangle multi-compartmental effects9,10.

Recent work8 highlighted the limitations of microstructural imaging from dMRI alone. For example, disentangling intra- and extra-axonal compartment properties in brain white matter is degenerate and ill-posed. An emerging trend is the combination of dMRI with other contrasts to disentangle multi-compartmental effects9,10.

Relaxometry exploits the inherent sensitivity of MRI to the biochemical environment of tissue, captured by transverse (T2 and T2*) and longitudinal (T1) relaxation times. T1 processes describe the recovery of longitudinal magnetization while T2 processes describe the loss of transverse magnetization due to dephasing effects originating from interactions at the molecular level. Additional dephasing—caused mainly by field inhomogeneities, differences in susceptibility, and chemical shift effects—are described by T2′, contributing to the shorter characteristic T2* time. Thus, changes in T1, T2 and T2*directly reflect the biochemical environment in tissue. Relaxometry has been fundamental to enabling quantitative measurements of tissue properties in health and disease and has been proven in a wide variety of tissues.

Biophysically-linked tissue parameters which have been extracted from in-vivo relaxometry or dMRI data include but are not limited to: axon density9, myelin fraction11,12, deoxyhemoglobin concentration, iron concentration13,14 and biliary fibrosis15. While novel relaxometry techniques such as MR Fingerprinting16 or steady-state gradient echo based methods17,18,19,20 are available, the accepted gold-standard techniques are still inversion-recovery (IR) sequences for T1 mapping, and repetitions with multiple echo times for T2/T2* mapping. IR sequences sample the T1 domain by varying the inversion time (TI) between the inversion pulse and the acquisition, while the acquisition of multiple echoes samples the time between excitation and acquisition (TE)13. Repeated gradient-echo experiments allow T2* quantification, whereas repeated spin-echo (SE) experiments estimate T2 - since the refocusing pulse eliminates the previously discussed sources of additional transverse magnetisation decay (i.e. T2′).

Joint relaxometry-diffusion experiments have been conducted successfully in the context of nuclear magnetic resonance spectroscopy, improving the ability the distinguish different compartments21,22,23. The recent study by Almeida et al.23 samples the three-dimensional space T1-T2-diffusion, and adds the shape of the diffusion encoding tensor to bval/bvec as an additional dimension within the diffusion parameters. These techniques have recently been extended to imaging, e.g. Kim et al.24 and Benjamini et al.25 illustrated increased separation of compartments in injured spinal cord tissue by sampling the T2-diffusion and T1-T2-diffusion spaces respectively.

In-vivo application of joint relaxometry-diffusion studies has been hampered by the prohibitively long acquisition times. However, recent in-vivo studies by e.g. Tax et al.26, and Veraart et al.10 demonstrated joint T2*/T2-diffusion experiments and showed great promise. Additionally Arazany and Assaf27, and DeSantis et al.28,29 have published joint T1-diffusion studies. However, these studies share three significant limitations:

  1. 1.

    no in-vivo study to date has sampled both longitudinal and transverse relaxation times

  2. 2.

    every acquisition parameter combination is sampled in a separate acquisition (acquisition is defined in the following by their separate preparation time: for T1 the required TI, for T2* the required time until the desired TE)

  3. 3.

    all the mentioned studies reported scan times in excess of 60 minutes

Only challenge (2) has been addressed so far in the context of joint T1-diffusion experiments by the introduction of slice-shuffling (SS) approaches by Ordidge et al.30 and Wu et al.31. In this approach the slice acquisition order is changed from volume-to-volume to achieve sampling of every spatial location with diffusion weighting at every inversion time TI. The nested structure of this approach makes use of all required preparation times for the acquisition of different slices. However, while limiting the preparation idle time, current SS techniques are not optimally efficient: the intrinsic link between the number of inversion times to the number of slices effectively oversamples the TI range for many geometry choices.

Attempts to sample transverse relaxation time are currently done by acquiring multiple repeats of the dMRI or SS acquisition with varying TE.

Sampling schemes for the described acquisitions are graphically depicted in Supporting Fig. S1.

In this study, we introduce an integrated efficient acquisition technique which can address the three stated challenges and allow efficient sampling of the three-dimensional acquisition parameter space (here T1-T2*-diffusion). This is achieved by nesting the sampling of all required acquisition parameter combinations into one acquisition such that all required preparation times are used. We call it ZEBRA - short for Z-location shuffling, multiple Echos and B-interleaving for Relaxometry-diffusion Acquisitions - in recognition of the intrinsically striped appearance of slice stacks with echo time, b-value and TI all varying. The proposed method is enabled by two novel methodological elements:

  • (I) A multi-echo approach, combining a spin echo with subsequent gradient echoes allows simultaneous exploration of the T2*-diffusion space.

  • (II) Efficient sampling of the inversion recovery curve by interleaving the diffusion encodings in super-blocks, which allows more diverse diffusion weightings to be sampled in the time saved by avoiding over-sampling the TI dimension.

We demonstrate the resulting integrated T1-T2*-Diffusion acquisition in vivo on adult brain data. The obtained quantitative maps show the potential of this technique and test-retest variability experiments demonstrate its robustness. Finally, multi-dimensional scatter plots of the obtained tissue parameters give a first hint at the substantial information. All performed experiments are on healthy volunteers and were planned and performed to establish the robustness as well as the possibilities introduced with the novel ZEBRA method. The acquisition method can be obtained from the authors providing that vendor research agreements are in-place. All post-processing steps and exemplary data is freely available.

Experiments

The novel ZEBRA sequence as described in the methods section was implemented on a clinical Philips Achieva-Tx 3T scanner (Release 3.2). The presented experiments were obtained on a clinical 3T Philips Achieva scanner using the 32ch adult head coil. All methods were carried out in accordance with relevant guidelines and regulations, healthy volunteer scanning was approved by the Riverside Research Ethics Committee (REC 01/11/12) and informed written consent was obtained prior to imaging. The data was obtained using a 32channel adult head coil. An adiabatic non-slice selective inversion pre-pulse32 was employed for the global inversion.

Validation experiment

Phantom validation was performed by imaging an in-house built spherical phantom filled with a 2% agarose solution and primarily focused on the goal to comparing the tissue parameter estimates obtained via the ZEBRA approach to the conventional approaches employed for estimating T2*, T1 and Apparent Diffusion Coefficient (ADC).

Reference T2* values were found via a multi-echo gradient echo sequence (MEGE) sequence. A gold-standard IR sequence was employed for T1 measurements and a conventional SE dMRI sequence was acquired in order to estimate ADC. All acquisitions used single-shot Echo Planar Imaging (EPI), with a fixed Field-of-view (FOV) of 260 × 260 mm2, resolution 3 mm3. The MEGE, dMRI and ZEBRA acquisitions had \({N}_{s}=28\) slices, except for IR, where only one slice was acquired to keep the imaging time acceptable. See Table 1 for all acquisition parameter details. The ZEBRA data was processed individually to match the equivalent processing of the individually acquired data sets: the longest inversion time at b = 0 to estimate the T2* (matching the MEGE data), the longest inversion time at the shortest echo time with all four b-values to estimate the ADC value (matching the dMRI data) and finally all inversion times for b = 0 at the shortest echo time to estimate T1 (matching the individual IR experiments).

Table 1 Performed experiment parameters.

Cramer-Rao Lower Bound Simulations

To further evaluate the super-block strategy regarding the relation between super-block length and robustness of tissue parameter estimation, simulation experiments were performed. For this, the Cramer-Rao Lower Bound (CRLB)17 was used on a slice level to predict the minimum obtainable variance of the model parameters given the performed Nv independent measurements and a fixed noise level.

By considering the signal model \(g({\bf{x}},\theta )\), where \({\bf{x}}=[{\alpha }_{1},{\alpha }_{2},\ldots ,{\alpha }_{{N}_{v}}]\) are the user controlled parameter vector αi = [bvali, TIi, TEi] and \({\rm{\Theta }}=[{\theta }_{1},\ldots ,{\theta }_{M}]\) are the model parameters. These include T1, T2*, proton density (PD) and inversion efficiency (IE): \(\theta =[PD,T1,T{2}^{\ast },ADC,IE]\). For independent measurements following a Gaussian noise distribution with standard deviation σ, the CRLB is calculated as \({({\bf{J}}^{\prime} {\rm{\Sigma }}{\bf{J}})}^{-1}\), where J is the jacobian of the signal model obtained as

$$[\begin{array}{c}\frac{\delta {g}_{1}}{\delta {\theta }_{j}}\,\cdots \,\frac{\delta {g}_{1}}{\delta {\theta }_{M}}\\ \cdots \cdots \cdots \cdots \cdots \\ \frac{\delta {g}_{{N}_{v}}}{\delta {\theta }_{j}}\,\cdots \,\frac{\delta {g}_{{N}_{v}}}{\delta {\theta }_{M}}\end{array}].$$
(1)

Finally, each parameter is weighted by the inverse square of the assumed tissue parameter value \({{\rm{\Theta }}}_{0}\) and summed to obtained the precision p.

The used model here was the signal decay equation in Eq. 3, \({{\rm{\Theta }}}_{0}\) was chosen as

$$[PD=1000,T1=1500\,ms,T{2}^{\ast }=200\,\,ms,ADC=0.001\,\,m{m}^{2}\,{s}^{-1},IE=2].$$
(2)

Other acquisition parameters included \({N}_{s}=40\), TR = 10 s and SNR = 200. Considered TIs are \([50:TR/{N}_{s}:TR-50]\) and b-values between \(0\,m{m}^{-2}\,s\) and \(3000\,m{m}^{-2}\,s\) sampled in steps of \(250\,m{m}^{-2}\,s\) were included. To evaluate specifically the super-block length, the TE was kept constant at TE = 60 ms and for the final evaluation only [S0, T1 and IE] were included. T2* could be neglected due to the fixed TE and ADC could be neglected due to the fixed b-value range. Each simulated set of measurements thus varied in length, b-value and employed TIs as follows: The TIs were chosen as a regularly spaced subset of the considered TIs \([50:TR/{N}_{s}:TR-50]\). For each scenario, both the obtained precision p and the number of required volumes Nv (and thus required time) were evaluated. The goals therefore are twofold: first, to quantitatively establishing the influence of the super block length for a range of b-values and second, to illustrate how the flexible approach for combining different super-block length within one acquisition can help standardise the obtained variance across all considered b-values.

Super-block length experiment

To inform the right choice of super-block length, another experiment (E2) was performed. Here, an IR-DRMI sequence using SS, but without the proposed interleaving was used to acquire an highly sampled inversion curve data (28 points along the inversion recovery curve between 57 and 6500 ms) on a single healthy adult brain. From this dataset, subsets with lower TI sampling density were extracted by choosing the super-block length to be \({N}_{i}=1,\,\mathrm{..},\,10\) and the respective T1 maps calculated.

Test-retest experiment

A healthy adult was scanned twice with ZEBRA (experiment E3) using the acquisition parameters as specified in Table 1 to compare test-retest variability. The described fitting algorithm was employed on both datasets, a brain mask calculated (MRTRIX3) and any bulk motion corrected on a volume level using FLIRT33. The ADC, T2* and T1 values of all voxels within the brain mask were analysed.

Healthy adult experiments

To illustrate the versatility of the proposed approach, two different ZEBRA parametrizations were employed on a total of four adult volunteers:

Two adults were scanned with a protocol sampling only four diffusion encodings \({N}_{d}=4\), corresponding to one direction on b = 0, b = 333, b = 667 and b = 1000 at \({N}_{e}=5\) echo times with one super-block with an interleave factor of \({N}_{i}=4\), resulting in \({N}_{TI}={N}_{s}/{N}_{i}=7\) TIs. Further acquisition parameters are given in Table 1 E4.

Another 2 adults were scanned with a high angular resolution diffusion imaging (HARDI) multi-shell protocol, composed of 3 shells, with 24 directions (b2600), 8 directions (b1000), 6 directions (b500) and 4 b0 volumes. These directions were chosen to maximize electrostatic repulsion on each shell34. These in total \({N}_{d}=42\) directions were scanned in 6 super-blocks, each composed of \({N}_{i}=7\) encodings. This interleave factor was chosen based on the CRLB simulations for the highest considered b-value and led to \({N}_{TI}={N}_{s}/{N}_{i}=4\) TI points sampled. The number of echos was fixed as \({N}_{e}=4\). Further acquisition parameters are given in Table 1 E5.

Results

Validation result

The results from the phantom experiment acquired with the different sequence variations together with the acquisition times are given in Table 2. The obtained tissue parameters with ZEBRA are very close to those obtained using the conventional methods. The times for the conventional inversion recovery sequence are for one individual slice only, times for dMRI, MEGE and ZEBRA are for the whole volume.

Table 2 Results phantom experiments.

For the entire ‘diffusion encoding-TI-TE space’ presented here (5 TEs, 7 TIs and 4 diffusion encodings), the reduction achieved by ZEBRA compared with the current most efficient state-of-the art technique (slice-shuffling) equals 20. This results from a 5-fold acceleration by packing the 5 TEs in subsequent echos and by the achieved 4-fold acceleration due to interleaving \({N}_{i}=4\) diffusion encodings. All considered conventional and ZEBRA sampling schemes are shown in Fig. 1a.

Figure 1
figure 1

(a) Achieved sampling of the acquisition parameter space with conventional methods and ZEBRA. (b) CRLB results for b-values between 0 and 3000, sampled at 250 mm−2 s and \({N}_{i}=1,\mathrm{..},10\) with overlayed contour lines. (c) Results from adult experiment E2. The obtained T1 values for white matter, grey matter and cerebro spinal fluid are shown for interleaved setting between \({N}_{i}=1,\,\mathrm{..},\,10\).

CRLB Simulation results

The results of the CRLB simulations are given in Fig. 1b. The CRLB for the considered bval-Ni grid (where the number of interleaved bval/bvec combinations Ni defines acceleration) is displayed using a color code: Yellow being the highest variance and thus lowest precision and blue the lowest variance and thus highest precision. The isolines illustrate the required interleave factor across considered b-values to obtain similar variance. For a variance level of 0.1 for example, the highest considered bval of \(b=3000\,\,m{m}^{-2}\,s\) could thus only be accelerated by factor 3, while a \(b=0\,\,m{m}^{-2}\,s\) could be acquired with \({N}_{i}=7\) to achieve similar precision in the estimation of [PD,T1,IE] (both marked with a cross in Fig. 1b). Each interleave factor Ni results in a different acquisition time, where \({N}_{i}=1\) corresponds to no acceleration and \({N}_{i}=10\) to a 10-fold acceleration.

Superblock length experiments

The results from imaging experiment (E2) are depicted in Fig. 1c. Example T1 values plotted as a function of super-block size, Ni, for cerebro spinal fluid, grey matter and white matter. These illustrate stable TI values of <2% difference between \({N}_{i}=1\) and \({N}_{i}=10\) for TIs in the grey matter voxels and <5% difference for white matter. Only the values in the cerebro spinal fluid decline by about 12% for \({N}_{i} > 6\). A mid-brain slice from the same experiment is shown in Supporting Fig. S2. On visual inspection, the results are stable for \({N}_{i}=1,\,\mathrm{..},\,10\) corresponding to up to 10-fold acceleration.

Test-retest experiments

The Bland-Altman plots from the test-retest adult experiment in Fig. 2 illustrate the ADC, T2* and T1 for a random subset of all voxels in the brain mask. They show a high degree of reproducibility with \({r}^{2}=0.95\), 0.96 and 0.89 respectively. The relatively high spread of the T1 results is a consequence of subvoxel motion between the scans leading to increased errors, mainly in the CSF.

Figure 2
figure 2

Bland-Altman plots from the adult test-retest experiment for ADC, T2* and T1 values.

Adult brain results

The data sets are obtained from the scanner in acquisition order - thus displaying when reformated to coronal plane slices with different TIs and b-values within the same volume. This is depicted in Supporting Fig. S3 - showing the coronal view of transverse acquired slice stacks for the first superblock (10 volumes) of the first echo. Supporting Fig. S3b shows exemplary slices with their corresponding TIs and b-values illustrating in more detail the achieved densely packed data content.

A mid-brain slice from a dataset, subsampled over all TEs, TIs and b-values is depicted in Fig. 3, depicting the sampling scheme (a) together with data along the three planes in (b-d).

Figure 3
figure 3

A selected mid-brain slice from an in-vivo three-dimensional data set is depicted. In (a) the obtained sampling in the three-dimensional acquisition parameter space is shown together with three potential cuts to visualize the data content. These cuts are sub sampled to facilitate the illustration. (b) The TI-b plane at second echo time and (c) the TE-TI plane at b = 0, (d) the TE-b plane at longest TI, are depicted.

Exemplary obtained T2*, T1, inversion efficiency and ADC maps for subject 1 together with a zoom into the optic radiation is depicted in Fig. 4. The total acquisition time for this dataset was 2:42 min. The estimates of the joint fit are within the expected tissue parameter ranges. No processing or correction for motion and distortion was performed to minimize additional steps from the ZEBRA acquisition to the final parameters.

Figure 4
figure 4

The obtained T2*, T1, ADC and inversion efficiency maps are shown at three axial locations in the brain together with a zoom. Note that no post-processing was performed to illustrate the ZEBRA results - particularly no distortion correction.

The T1, T2*, ADC and IE values obtained from the adult ZEBRA dataset are displayed in scatter plots in Fig. 5 for all voxels within the brain mask. The multi-dimensional tissue space is presented for each possible pairing of two acquisition parameters e.g. T1-T2* (top row, left side), T1-ADC (tow row, middle), T1-IE (top row, right side), T2*-ADC (bottom row, left side), IE-ADC (bottom row, middle) and T2*-IE (bottom row, right side). In each case a third tissue parameter is indicated by colour coding the data points. Thus there are two different colored plots for each tissue parameter paring. The scatter plots show the complex and rich inter-relationships between the different tissue parameter. Concretely, focusing e.g on the plot specifying the correlation between TI (x axis) ADC (y axis) and IE (color), displayed in large in Supporting Fig. S4. Three regions in the plot, illustrated in circles with continuous, dashed and dotted lines illustrate the different relations between the parameters: Both region 1 (continuous) and region 2 (dashed) show linear linear dependence between T1 and ADC, but with different slope. The additional T2* (left side) and IE (right side) information shows similar T2* for the entire region, but IE separate the region into a part with lower IE (left of the black arrow) and one with higher IE (right of the arrow). Region 3 (dotted line) draws the attention to a region with similar T1, T2* and IE but a spread in ADC - illustrating the benefit of fitting all these parameters.

Figure 5
figure 5

Scatterplots of all voxels within the brain mask illustrating the relations between the four discussed tissue parameters T1, T2*, ADC and IE. First row left to right: ADC-T1, color by IE, ADC-T1, color by T2*, T1-T2*, color by IE and T1-T2*, color by ADC. Second row left to right: T1-IE, color by T2*, T1-IE, color byADC, ADC-T2*, color by T1 and ADC-T2*, color by IE.

Results from the second adult protocol (E5) are depicted in Fig. 6. The described HARDI-ZEBRA scan resulted in a total of \(42\cdot 4\cdot 4=672\) volumes acquired in a total scan time of 22:40 (TR = 8 s). This corresponds to an acceleration of \({N}_{e}{N}_{i}=28\), compared to the next-fastest slice-shuffling approach for combined diffusion-T1 analysis acquired in separate acquisitions with different echo times to sample the TE acquisition parameter space.

Figure 6
figure 6

The acquired HARDI-ZEBRA data is shown together with some first preliminary analysis results. (a) The five first volumes are shown as acquired in transverse and in reformatted coronal planes for the first three echos. This represents about 2% of the total acquired samples. The characteristic ZEBRA look in the reformatted coronal plane is due to b-value interleaving. (b) Illustrates the obtained ODFs overlayed over the \(\ell =0\) spherical harmonics term for both short and long TIs as well all five echo times.

The Fig. 6 illustrates the HARDI-ZEBRA experiment. The first five volumes for the first three echos are depicted - corresponding roughly to 2% of the acquired samples. Numerous analysis techniques to exploit the data are possible, only two examples are given with (b) directional T2* maps and (c) TI and TE dependant orientation distribution function (ODF) results, obtained with MRTrix3.

Discussion

This paper presents ZEBRA - an integrated acquisition for quantitative MRI of multiple parameters, maximizing efficiency by optimally sharing preparation periods.

The datasets generated are available from the corresponding author on request together with the employed post-processing and fitting scripts.

Flexibility and combination

The flexible nature of the proposed ZEBRA approach allows the acquisition parameters (interleaving factor, number of super-blocks, number of acquired echoes) to be tuned to the expected SNR and range of variation in the tissue parameters to be estimated. Using the acquisition parameters as presented allowed for a 20-fold/28-fold decrease in imaging time compared to the separate acquisition of T1-diffusion with SS and T2*-diffusion data.

The proposed T1-diffusion sampling combines effortlessly with acceleration techniques such as SENSE and Half-scan as the proposed conceptual change is independent of the individual EPI read-out. Combination with simultaneous slice imaging techniques is also straightforward providing further increases in coverage without increasing acquisition time.

The obtained ZEBRA data was also used to obtain the tissue parameter inversion efficiency, and thus was shown to be able to provide a window on magnetization transfer effects. However, further investigation of this parameter is beyond the scope of this paper but can be of interest for future studies17.

Additional acceleration by reduction of thermal load

An additional benefit is that both changes, mixing the diffusion weighting shot-by-shot and adding additional echoes, can decrease the thermal demands on the gradient system, allowing more efficient operation. This is, however, beyond the scope of this study.

Considerations regarding acquisition parameters

The T2*-diffusion sampling by multiple echoes leads to additional important considerations for the EPI read-out. Reducing the first echo time to ensure that sufficient signal is maintained to sample the T2* decay curve is in direct competition with high-b value gradient preparations and low sense factors.

Further considerations include the sampling of the TE-TI space: More echoes, as required for higher sensitivity to the longer T2* range, lead to a less dense TI spacing. This forms an important part of the sampling design, such as, for example, be counteracted by changing the employed interleaving factor.

Motion robustness

The interleaving of different b-values results in the spreading of the slices forming one volume for same bval/bvec and TI combination across multiple acquired slice stacks. This can lead to more challenging motion patterns within any given volume with a particular weighting. However, as the entire dataset with all acquisition parameter settings forms the input for the multi-parametric modelling, any motion during the acquisition needs to be dealt with independent of interleaving. The more complex motion patterns after sorting the data to individual volumes might contribute to a need for motion correction on the slice level rather then volume level, such as has been used for fetal applications of diffusion MRI (e.g.35) and more recently for adult brain imaging studies36. The multi-echo sampling of the required data for T2* mapping acquires all TE samples within 200 ms and thus effectively freezes motion.

Limitations of this study

All experiments and simulations were limited to regularly spaced and interleaved bval/bvec combinations. The sequence however, allows for free distribution of samples to the inversion curve. Further investigation in this direction could be beneficial and will be part of future work.

The focus of this study lies on the presentation of the obtained highly efficient novel acquisition, not on the processing and modelling. The presented joint fits are merely first attempts to explore the obtained data. While the second performed adult scan demonstrated the ability of ZEBRA to deliver a multi-shell HARDI dataset, the directional information and the multi-compartmental origins of the signal were ignored in model (3). This was a deliberate choice to simplify the multiparametric signal fitting. Future work can focus on advanced multi-dimensional modelling tailored to the rich datasets acquired with the proposed ZEBRA technique.

Outlook

Future work will focus on deploying the proposed ZEBRA acquisition for in depth brain studies. In particular, this novel accelerated acquisition ideally lends itself to bespoke multi-compartment models that can disentangle myelin and axonal compartments. Furthermore, ZEBRA has potential for applications outside the brain, such as kidneys, liver or prostate studies.

Conclusion

The proposed ZEBRA acquisition combines several novel elements to allow highly efficient integrated sampling of the multi-dimensional acquisition parameter space TE-TI-bval/bvec to obtain quantitative data in the T2*-T1-diffusion domain. Its robustness was shown on phantom and adult brain experiments, the analysis of the several dimensions per voxel shows the independence of the tissue parameters and thus the benefits of this higher dimensional approach. It answers two key challenges, scanning efficiency and consistent, simultaneously acquired data and thus facilitates in the future the wider use of joint relaxometry-diffusion techniques both for research applications and clinical use.

Methods

Interleaving diffusion encodings

The conventional IR experiment consists of a slice selective inversion pulse, a delay time defining the TI and a single- or multi-shot read-out as illustrated in Fig. 7a for the case of single shot EPI. SS techniques use a non-selective inversion pulse followed by a stack of single shot EPI slices excited in such a way that any preparation time required to achieve a specific TI in one slice is used to sample another slice at a shorter TI. This leads to one continuous acquisition and no waiting delay times as illustrated in Fig. 7b. If there are Ns slices, then the whole procedure must be repeated Ns times with the slice firing order sequentially shifted to achieve a set of complete stacks. Each slice is also diffusion weighted and they require every slice (\(s=1,\,\mathrm{..},\,{N}_{s}\)) to be sampled with every TI for every diffusion encoding (\(d=1,\,\mathrm{..},\,{N}_{d}\)), where each d signifies one combination of bval/bvec. The total acquisition thus consists of \({N}_{s}\times {N}_{d}\) volumes, parametrized by volume number v with \(v=1,\,\mathrm{..},\,{N}_{s}\times {N}_{d}\). This approach is extremely inefficient as the inversion recovery curve is generally massively oversampled: typical values of TR = 6 s and \({N}_{s}=40\) lead to data points at 150 msec intervals (Fig. 7b, right column) with the minimum time determined by the EPI shot time. More appropriate, sparser sampling is not possible due to the limitations of the traditional ‘one volume – one diffusion encoding’ scheme where every slice is sampled with the same diffusion encoding before the next encoding is chosen.

Figure 7
figure 7

The structure of one single inversion recovery experiment is illustrated in (a), consisting of a global non-selective inversion pulse, a delay time defining the obtained IR contrast and the single-shot EPI module. The IR-dMRI sequence is composed of these blocks as illustrated in (b). Thereby each column represents one time sequence within the TR, each row stands for a novel repeat (volume). The schematic slice-shuffled acquisition without diffusion interleaving is illustrated for two bvec/bval combinations (highlighted as different shades of blue) together with the resulting sampling of the inversion curve using this conventional approach (right column). (c) The schematic slice and diffusion shuffled acquisition schema is illustrated for four bvec/bval combinations resulting in the interleaved sampling of 4 different diffusion encodings along the inversion recovery curve (right column).

Our approach breaks with this traditional paradigm by changing the diffusion encoding per slice37. By interleaving Ni different diffusion encodings within the TR (Fig. 7c, where \({N}_{i}=4\)), multiple diffusion encoding combinations can be sampled within a single recovery curve (Fig. 7c, right column). Instead of repeating the same diffusion encoding with all Ns slice shuffles, this technique only requires \({N}_{s}/{N}_{i}\) repeats. To combine the needs to sample every slice with every diffusion encoding and to achieve full diffusion encoding volumes per TI, a rolling approach is chosen: The diffusion encoding alternates between the chosen Ni encodings within every volumes and the start of this permutation is incremented by one for subsequent volumes. Figure 8a,b show the chosen diffusion encoding (a) and inversion times (b) in the s vs. v space. The four colors in Fig. 8(a) represent four separate bval/bvec combinations illustrating the shifted permutations from volume to volume. The grey levels in Fig. 8(b) represent all chosen TIs and their incrementation from volume to volume. This leads, as illustrated in Fig. 8c (TI vs. v) to a constant mapping of d to TI. The sampling in the T1-diffusion domain, shown in Fig. 8d is thus on an oblique lattice with flexible control of sampling density.

Figure 8
figure 8

The chosen mappings showing (a) the employed diffusion encodings and (b) the inversion times mapping from volumes v to geometric location s. In (c), the resulting mapping of diffusion encoding in the volume v vs. inversion time TI plane is shown. (d) Shows the acquired sampling in the TI vs. d plane.

The order of the chosen Ni encodings is not relevant for the ability to obtain accelerated IR-dMRI scans, it was, however, chosen such that the low-b volumes are maximally spread over time to enhance the potential for further motion correction37.

Super-block design

We use the term super-block to refer to each sequence of Ns volumes. While all super-blocks are equal in length, they differ in the number of included interleaved diffusion weightings Ni for this block. While the same TIs on the recovery curve are sampled in every super-block, varying the length Ni - and thus the number of combined diffusion encodings – allows different sampling densities for each super-block. For these IR-dMRI blocks, the length Ni, and the slice number Ns define the number of points on the recovery curve per diffusion encoding as \({N}_{TI}={N}_{s}/{N}_{i}\). Superblock Ni can for example be selected to balance the declining SNR at higher diffusion weightings, recognizing that lower b-values can allow robust T1 estimation with fewer points on the recovery curve.

Finally, super-blocks without a pre-inversion - termed dMRI super-block - can be included into the sequence to achieve additional dMRI samples at a maximum data rate – these would, however, not be suitable for the simple combined modelling approaches. The sampling points for those equals \({N}_{TI}=1\) and the length of these dMRI blocks equals \({N}_{v}={N}_{i}\).

The final total acquisition will be made up by Nb (variable length) super-blocks of total length \({\sum }_{b}^{{N}_{b}}\,{N}_{v}^{b}\).

The choice of \({N}_{i}=4\) is shown in Fig. 7c. Additional choices with more details are depicted in Supporting Fig. S5 for \({N}_{i}=8,4,2\), illustrating the interleaved sampling, the obtained density as well as the required number of volumes.

Integrated T1-T2*-Diffusion sampling

The required echo times for T2/T2* estimation are typically obtained by including a delay between excitation and read-out to vary the echo time as illustrated in Fig. 9a.

Figure 9
figure 9

Conventional T2-Diffusion sampling in separate scans is illustrated in (a) together with the resulting delays. (b) Illustrates the proposed joint T2*-Diffusion sampling. Finally, (c) depicts the transverse magnetization schematically, showing the T2 decay in green and the T2* decay in orange.

Especially for the sampling of the higher TEs, these idle times render the experiment inefficient. ZEBRA employs a multi-echo approach to efficiently use all preparation times. Instead of repeating the diffusion preparation in different acquisitions by varying the TE, one acquisition employs one diffusion preparation to sample multiple TEs. Therefore, SE-EPI sequence was extended to a multi-echo sequence by adding several EPI read-out blocks with minimal spacing after the initial diffusion preparation and SE-read-out block as illustrated in Fig. 9b. The d-TE space is thus sampled simultaneously with the subsequent echoes. The extension from the spin-echo sequence to a double echo SE-GE sequence was previously exploited for dynamic distortion correction38. The resulting magnetization is illustrated in Fig. 9c.

Post-processing

The data is sorted to individual diffusion-encoding volumes at the obtained \({N}_{s}/{N}_{i}\) TIs and all TEs. Each resulting volume is thus parametrized by [bvec bval TE and TI] and the set of all volumes forms the input for any further modelling step. In the following, the data in every voxel was fitted to a joint signal model to obtain the most basic tissue parameters PD, T1, T2* and ADC using the following equation

$$S=PD(1-IE\,{\exp }^{-(TI/TE)/T1}+{\exp }^{-(TR/T1)})\,{\exp }^{-bADC}\,{\exp }^{-(TE/T{2}^{\ast })}.$$
(3)

Inversion efficiency (IE) is is also estimated in order to account for incomplete inversion (full inversion corresponds to \(IE=2\)). It is known that a bi-exponential longitudinal relaxation is present in tissues with strong magnetization transfer effects, such as white matter39. Magnetization transfer effects are expected to alter the estimated values of both PD and IE.

Tissue parameter estimates of PD, IE, T1, ADC and T2* were obtained based on a least-square criteria between Eq. 3 and data sampled via the proposed ZEBRA acquisition. For each voxel individually, tissue parameter estimates were found using Matlab 2014b Levenberg-Marquart algorithm, with a function tolerance value of 10−9 and a fixed initial guess set to PD = 4000 ms, IE = 2, T1 = 1000 ms, T2* = 200 ms, ADC = 0.003 mm2s−1.