Abstract
Considering the advantages and research status of microfluidic chip coupling with mass spectrometry (MS), a microfluidic chip-based multi-channel ionization (MCMCI) for the extraction of untreated compounds in complex matrices without sample pretreatments was developed. Quantitative analysis of human urine spiked with various rhodamine B concentrations was also performed, and good linearity was obtained. Comparing to the macro ionization device, MCMCI significantly improved the integration of ionization source, simplified the operation of such a device, and greatly increased the signal intensity with much lower gas pressure. Comparison of our MCMCI with two and three gas channels indicated that the liquid–liquid extraction process before spraying and after spraying produced similar MS results. Moreover, this MCMCI with three gas channels also implemented simultaneous dual sprays with high DC voltages, the interference of two samples was minor and ion suppression effect was drastically alleviated. Such advantages may easily enable internal calibration for accurate mass measurement. Furthermore, dual extraction can be implemented by integrating such multi-spray configuration, which can improve the extracted signal intensity and sensitivity. These technologies open up new avenues for the application of microfluidic chip coupling with MS.
Similar content being viewed by others
Introduction
Microfluidic chips have been widely applied in various research fields due to their efficient and fast separations, integrating complex sample pretreatment functions and automatic manipulation of small sample volumes1,2,3,4,5. The coupling of microfluidic chips with MS has gained considerable attention in the MS community6,7,8. However, most microfluidic chip-based ionizations only possess one liquid channel for spraying, lacking gas atomizing, desorption function, and many other ionization methods9. Studies on MCMCI are even less10, let alone the studies on extractive electrospray ionization (EESI) and internal calibration in microfluidic chips11,12.
Among the current atmospheric pressure ionization techniques13, dual-channel ionization has been widely applied for EESI14,15,16,17, internal calibration12,18,19, and protein analysis20,21,22. EESI plays a significant role in MS community15, which was similar to the FD-ESI proposed by Shiea et al.23. Based on the theory of liquid–liquid extraction among the colliding micro-droplets from two different spray sources, EESI owns the capability to directly analyze compounds in the solution, gas phase, and other complicated matrices without sample pretreatments14,24. However, optimizing the angle between the two sprayers, which significantly affect the sensitivity and stability of the measurements is difficult25. Despite that the air-flow-assisted EESI ionization source25 was developed to overcome this disadvantage, the complex macrostructure was still difficult to operate and optimize. Comparatively, microfluidic chip-based ionization is compact and integrated, making it particularly facile for coupling with MS and even with portable MS in the future. Lu et al.11. proposed a dual-channel electrospray microchip for solutions to mix at the spray tip, where high voltage was applied on the aqueous sample rather than the organic buffer solution. The relative standard deviation (RSD) of the total ion current (TIC) obtained from mixed spray was rather high.
In addition, dual-channel ionization is an effective approach to realize internal calibration for accurate mass measurements because the dual channels might prevent ion suppression of the two compounds12,18,19. However, numerous methods reported with two independent ESI emitters failed to introduce the analyte and reference solutions into a single mass spectrometry inlet simultaneously because the interference among the electric fields of the adjacent sprayers is severe12,26,27,28. By extending the ion delivery distance29,30 or applying the AC power18, both of them implemented simultaneous dual-channel sprays. However, the macrostructure was complicated and the AC power achieved half-lower signal intensity than the DC power. Ramsey et al.12 reported a microfluidic chip with dual ESI emitters for internal calibration. The chip sprayed two compounds sequentially by dynamic control of the ESI voltages.
In our previous work1,9, a microfluidic dual-channel self-aspiration sonic spray ionization chip was reported. This device enabled dual channels to spray simultaneously without high voltage, but the signal intensity and stability must be further improved. Herein, we report a novel design of MCMCI with high voltage for MS. This MCMCI enabled the extractive solvent to extract samples with the assistance of airflow. Besides, another similar microfluidic chip structure with high DC powers applying on the dual channels implemented to spray two samples simultaneously, which successfully prevented the ion suppression to some extent. These advantages simplified the operation of microfluidic chip-based ionization coupling with MS because optimizing the angle of EESI25, positioning the dual emitters at the MS inlet28, and dynamic controlling the ESI voltages for dual sprays were necessary12,31. Moreover, dual extraction (dual EESI) was realized based on the multi-spray configuration with high DC powers, which can improve the extracted signal intensity and sensitivity. All of these advantages of MCMCI were particularly suitable for coupling with portable MS in the future and may drastically extend the application of microfluidic chip in MS community; such applications are throughput increments32, sensitivity improvements33, and rapid mixing for ion/ion reactions21,22,34.
Results
MCMCI with two gas channels for extraction
The five layers design of microfluidic chip ionization source is presented in Fig. 1 and the whole schematic of four microfluidic chips is shown in Figure S1, ESI†. For MCMCI with two gas channels (the distance between two liquid channels was 10 µm), the two liquid channels were close to each other without a separating gas channel, and the liquids rapidly mixed at the tip of the device as one Taylor cone prior to spraying, which is shown in Fig. 1c. This result was similar to the study of Michel Prudent et al.10, except for the bilateral gas channels. Thus, the extraction mechanism of MCMCI with such configuration may be mainly based on the liquid–liquid extraction prior to spraying (i.e., the tip-mixing EESI mode). To demonstrate that this technique was applicable to detect complex matrices without sample preparation steps, undiluted human urine with 1 µM rhodamine B was tested. Figure 2 shows the MS spectrum of the urine test. The intensity and signal to noise ratio of rhodamine B were very high. For comparison, we also performed the experiment of 1 μM Rhodamine B in methanol and water sprayed by the commercial electrospray ionization with 5 KV high voltage as shown in Figure S2 ESI†. The signal intensity was similar with our on-chip extraction using real biological samples with same conditions, which clearly demonstrate the effeteness of our on-chip configurations.
Figure 3 describes the MS signal of rhodamine B in human urine at low levels. Although the background intensity was high in the full scan shown in Fig. 3a, the rhodamine B intensity shown in Fig. 3b was high enough for analyzing the sample when MS2 was utilized, which convincingly proved that this microfluidic chip possessed the ability to detect low concentration samples. Moreover, to test the stability of this microfluidic chip-based EESI, an experiment of purified water spiked with 6 µM rhodamine B was performed over 10 min (Figure S3, ESI†). The RSD of TIC in this experiment was only 2.3%. These experiments demonstrated that the sensitivity and stability of MCMCI were excellent.
To further demonstrate the characteristics of this extraction function, experiments under the same extraction conditions with various rhodamine B concentrations were performed. The distance between MS inlet and microfluidic chip for experiments was approximately 6 mm. For such illustrative calibration experiments, we diluted the urine with water five times for the following EESI experiments because the urine collected once from one person was not enough, as different concentration solutions for quantitative analysis required large amounts of urine background solvent. If urine was collected from different people or different times, the urine solvent background might be various and affect the quantitative analysis. Figure 4 shows a plot of TIC intensity as a function of the original rhodamine B concentrations in human urine. Good linearity was obtained. Under such experimental conditions with LCQ MS, the current best linear range is approximately 0.5 µM to 20 µM, because the thermo LCQ mass spectrometer with three-dimensional ion trap owned the capacity limitation for high concentration quantitative analysis. For comparision, we also performed the experiment of 1 μM Rhodamine B in undiluted urine and urine diluted by five times sprayed by the commercial electrospray ionization with 5 KV high voltage respectively as show in Figure S4 ESI†. Firstly, the signal was similar no matter the urine was diluted or not; secondly, the signal intensity of Rhodamine B in urine was much worse than in methanol and water solution as shown in Figure S2 ESI†. Thus, we can speculate that the five times dilution had little influence in these experiments and the MCMCI did really improve the signal intensity greatly comparing to commercial ESI.
The comparison experiments of this tip-mixing EESI and macro EESI were also conducted. The structure and position of macro EESI was based on the references14,17, the picture of which is shown in figure S5, ESI†. Two concentrations of rhodamine B in urine (0.1 and 3 µM) were tested by using the macro EESI and tip-mixing EESI. Each concentration was measured for six times, and each time lasted for 1 minute. The experimental conditions were the same for the macro EESI and tip-mixing EESI. The intensities of macro EESI were much lower than the tip-mixing EESI, particularly for high concentrations. The average extracted ion intensities (mass range from 443 to 444) of macro EESI (0.1 and 3 µM) were 3.04E3 and 3.62E3, respectively. However, the average extracted ion intensities (mass range from 443 to 444) of the tip-mixing EESI (0.1 and 3 µM) were 5.38E3 and 1.15E5 separately. Figure 5 shows the MS spectrum of human urine (diluted five times by purified water) spiked with 3 µM rhodamine B. The intensity and signal-to-noise ratio of macro EESI were much worse than the tip-mixing EESI. Considering that the high gas pressure (approximately 10 bar) was usually applied on the macro EESI, we also attempted to increase the gas pressure to improve the signal intensity. Thus, macro EESI was utilized to measure the signal of 15 µM rhodamine B in human urine with high gas pressure. When the experimental conditions were the same as mentioned above, the intensity of 15 µM rhodamine B was 5.42E3. When the gas pressures applied on the two sprayers increased to 10 bar, the signal intensity increased to 3.9E4, which was lower than the intensity of 3 µM from the tip-mixing EESI. Despite much higher gas pressure was applied on the macro EESI, the intensity of the macro EESI was still much lower than microfluidic chip.
MCMCI with three gas channels for extraction and dual sprays with high DC powers
For MCMCI with three gas channels, a separating gas channel was placed between the two liquid channels; thus, the fluids did not mix at the tip of the device before they were sprayed; two liquids formed two isolated Taylor cones at each liquid channel front end, as shown in Fig. 1d. According to such observation and assumption that the two liquid channels in MCMCI with three gas channels served as two isolated electrospray ionizations, the samples were isolated until they were sprayed out. Therefore, the extraction mechanism of MCMCI with such configuration might be mainly based on the theory of liquid–liquid extraction among colliding droplets from two different spray sources (i.e., the droplet-collision EESI mode). The EESI mass spectrum signal of MCMCI with three gas channels is shown in Figure S6, ESI†; the signal intensity (9.7E4) was similar but a little lower than the signal of MCMCI with two gas channels (1.5E5), as shown in Figure S3b, ESI†. These results suggested that for MCMCI-based EESI, tip-mixing EESI based on the liquid–liquid extraction before spraying and droplet-collision EESI based on the liquid–liquid extraction among the colliding droplets after spraying shared very similar MS results.
Based on the isolated spraying assumption, the interferences and ion suppression effect between the two liquid channels might be relieved with high DC powers supplied simultaneously. Figure 6a shows the long-term stability of the two samples spraying at such situation. RSD was only 1.6%, which definitely proved that the MCMCI could spray stably. Moreover, comparing the signal intensities of dual emitters (Fig. 6b) with single emitter (Fig. 6c and d), the signal for reserpine was decreased by a factor of 2.5, whereas rhodamine B remained at similar signal intensity. This finding proved that the interference between two sprays existed but was not severe, as pointed out in the previous study18 where the signal intensity decreased by a factor of 4.7. Thus, our MCMCI avoided the severe interference among the DC electric fields of the adjacent sprayers, which existed in the previous reported devices12,18. Thus, AC power was utilized in the previous study18, but the signal intensity of which was nearly half of the DC nanoESI because of the half cycle’s negative voltage in AC power; furthermore, optimizing the amplitude and waveform of AC power is also complicated18.
To further illustrate the ion suppression effect, peptide GPRP and PEG-400 were utilized in the experiments. Under the same experimental conditions, the MS spectra of peptide GPRP and PEG-400 is illustrated in Fig. 7. The signal intensity of peptide GPRP was submerged when the mixture of PEG-400 and peptide was sprayed by one channel, ion suppression effect existed severely, as shown in Fig. 7a and proved in the previous study18. While the PEG-400 and peptide were sprayed separately in two channels, it could easily obtain comparative signals of PEG-400 and the peptide simultaneously, as demonstrated in Fig. 7b. These results further proved the assumption that samples were isolated until they were sprayed out, and the extraction was occurred after spraying in the MCMCI with three gas channels. Furthermore, such dual spraying with high DC powers supplied simultaneously may have two advantages. First, such configuration is suitable for the application of accurate mass measurements by internal calibration. Second, such configuration may be extended to integrate dual extraction function with simultaneously supplied high voltages on the multi-spray channels (the extraction solvent channels were set at both side of the sample channel), as shown in the following section.
Microfluidic Chip Ionization source for Dual EESI
Microfluidic chip ionization for dual EESI was implemented by integrating the functions of extraction and multi-spray configuration with the above-mentioned high DC powers. Multi-sprays mean three liquid spraying channels were set with two (dual tip-mixing EESI) or four (dual droplet-collision EESI) gas channels, as illustrated in Figures S1c and S1d, respectively. Figure 8a shows the structure of dual tip-mixing EESI. The sample was introduced through the middle liquid channel and the two extractive solvents distributed at both sides of the sample, these three liquids also fused together as one Taylor cone at the tip of the liquid channels. The TIC intensity of 0.1 µM rhodamine B is described in Fig. 8b. Dual tip-mixing EESI significantly improved the extraction sensitivity and signal intensity, which can also be clearly examined in Fig. 8c and d. This dual EESI helps improve the signal intensity, but a larger Taylor cone might cause a slight instability of spraying compared with a smaller one. Dual droplet-collision EESI with three liquid channels and four gas channels was also designed. The experiments were also performed; Figure S7, ESI† also proved that dual droplet-collision EESI could obtain higher signal intensity than single EESI.
Discussion
EESI was implemented in microfluidic chip coupling with MS for the first time. Stable extraction function enabled quantitative analysis of untreated compounds in complex matrices, and the good linearity (R2 = 0.998) showing in Fig. 4 proved that metabolites in biological samples might be quantitatively analyzed without any pretreatments by MCMCI coupling with MS. More importantly, compared with the macro EESI device, microfluidic chip ionization reflects enormous superiority. As demonstrated in Fig. 5, microfluidic chip based tip-mixing extraction obtained higher signal intensity, sensitivity and signal to noise ratio with much lower gas pressure; moreover, optimizing the angle between two sprayers was elided, which significantly simplified the operation of EESI device. MCMCI with three gas channels also enabled extraction in the droplet-collision mode. The mass spectrum of tip-mixing and droplet-collision extraction was similar as shown in Figure S3b and S6, ESI†. In other words, the MCMCI with two and three gas channels both possessed the advantages of microfluidic chip extraction.
Owing to the electric interference between two adjacent sprayers, dual simultaneous electrosprays based on microfluidic chip was barely reported. This study demonstrated the function of spraying two samples simultaneously with high DC voltages. The long-term stability of two samples spraying simutaneously proved that the MCMCI could spray stably, as shown in Fig. 6a, where the RSD was only 1.6%. The alleviation of ion suppression effect made it suitable for internal calibration for accurate mass measurement. MCMCI was free from interference of electric field, it might benefit from the dual spray channels rather than the dual spray tips, which generated two high electric fields.
Dual extraction microfluidic chip ionization was also developed by integrating the functions of extraction and dual simultaneous electrosprays. Such multi-spray configurations were beneficial to further improve signal intensity and sensitivity. This fully embodied the integration characteristics of microfluidic chip.
In summary, compared with the macro ionization device, tip-mixing and droplet-collision extraction based on the microfluidic technology significantly improved the integration of ionization and simplified the operation of such device, and the dual extraction configuration can further improve the performance, making it particularly suitable for coupling with portable MS in the future. This paper mainly introduced the new chip structure, realized the fabrication process and verified its abilities; further application will be conducted in our future works, which we will consider blood samples, sweat samples and oil samples with different analytes, besides the urine samples. These advantages may significantly extend the application of microfluidic chip in MS community; such applications are throughput increments32, sensitivity improvements33, and protein analysis34.
Methods
Materials and Equipment
HPLC-grade methanol and acetic acid were purchased from Merck KGaA (Darmstadt, Germany). PDMS elastomer base and curing agent (Sylgard 184) were purchased from Dow Corning (Midland, MI, USA). SU-8 photoresist was obtained from Microchem Co. (Naton, MA, USA). Samples used in this study were obtained from commercial sources. The urine was collected from one adult directly. All solutions were supplied to the microfluidic chip through short stainless steel tubes embedded in the reservoirs using a pneumatic pressure controller (MFCS, Fluigent, Paris, France). The high voltage generated by a power supply module (Dongwen High Voltage Power Supply Co., Ltd, Tianjin, China) was applied on the stainless steel tube of the extractive solvent. A thermo LCQ MS(Thermo Fisher Scientific Inc., Waltham, MA, SA) was coupled to the microfluidic chip, and MS data were collected by the computer. The detailed schematic of the platform was shown in Figure S8, ESI†. The MCMCI was held by a laboratory-built platform and coupled to the ion trap mass spectrometer. The distance between the microfluidic chip emitter and the MS inlet orifice was adjusted by a xyz-manipulator.
Design of MCMCI
The three layers of MCMCI were first designed by using AutoCAD (Autodesk, Inc., San Rafael, CA, USA). Then, the three photo masks were manufactured by Qingyi Precision Mask Making Co., Ltd (Shenzhen, China). The whole schematic of microfluidic chips is shown in Figure S1, ESI†. The width was 30 µm at the end of all channels. The figure contains a so-called tip-mixing extraction configuration with two gas channels (Figure S1a), a droplet-collision extraction configuration with three gas channels, which was also used for dual sprays simultaneously with high DC powers (Figure S1b), a dual tip-mixing extraction configuration with two gas channels (Figure S1c), and a dual droplet-collision extraction configuration with four gas channels (Figure S1d).
Fabrication of MCMCI
MCMCI was fabricated by polydimethylsiloxane (PDMS) utilizing standard multilayer soft lithography techniques as we have reported for fabricating the microfluidic chip-based sonic spray ionization9. The completely detailed fabrication process is shown in S1, ESI†. Five layers of MCMCI are illustrated in Fig. 1a. The top half chip included convex layer, gas channel layer, and liquid channel layer, whereas the bottom half chip contained gas channel layer and convex layer. Gas channel layers and convex layers were the same in the top and bottom half chips to form a symmetrical structure.
MCMCI for extraction
For tip-mixing and droplet-collision extraction experiments, high voltage (5 kV) was applied on the extractive solvent, which was a mixture of methanol/water/acetic acid (70: 20: 10). The samples were urine (diluted five times by purified water) and purified water spiked separately with rhodamine B. Rhodamine B served as the standard determinant, and the urine served as the real biological solvent background. The pressures on the liquid channels were all 300 mbar, whereas the pressure was all approximately 3.5 bar on the gas channel inlet; such gas pressure was the optimized value for single channel ESI according to our previous work1,9, and thus was selected for each channel in our multiple channel experiments. Under such conditions, the flow rate was approximately 5.6 µL/min, which was similar to the flow rate of 5–10 µL/min that is usually used in EESI.
MCMCI for dual sprays with high DC powers
As shown in Fig. 1d, apart from the droplet-collision extraction, the MCMCI with three gas channels was also developed for and dual sprays with high DC voltages applying simultaneously on two liquid channels, where the two liquids sprayed separately. In the experiments of dual sprays with high DC powers on, two liquid channels, where high voltages (5 kV) were all applied on, introduced two different samples. The pressures on the liquid channels were all 300 mbar, whereas the pressure was all approximately 3.5 bar on the gas channel inlet.
MCMCI for dual EESI
Dual EESI integrates the functions of signle EESI and dual sprays with high DC powers. Thus, the experimental conditions of dual EESI combined the conditions of the above two functions. High voltages (5 kV) were applied on the two extracted solvents, which were distributed at the two sides of the sample liquid channel. The pressures on the liquid channels were also 300 mbar, whereas the pressure was all approximately 3.5 bar on the gas channel inlet.
The coupling of MCMCI with mass spectrometry
The coupling of MCMCI with a thermo LCQ MS is shown in Fig. 1b. The optimized distance between MS inlet and microfluidic chip was approximately 6 mm, which was adjusted by a multi-dimensional manipulator; such optimized distance was determined to be around 5mm-6mm for our expermiment platform for coupling microfluidic ESI with MS in our previous work35. Compared with macro EESI, MCMCI was much easy to couple with MS because the angle between two sprayers was not required to optimize.
Data Availability
All data generated or analysed during this study are included in this published article (and its Supplementary Information files).
References
Yu, C. et al. Three-Dimensional Electro-Sonic Flow Focusing Ionization Microfluidic Chip for Mass Spectrometry. Micromachines-Basel. 6, 1890–1902 (2015).
Sun, X., Kelly, R. T., Tang, K. & Smith, R. D. Ultrasensitive Nanoelectrospray Ionization-Mass Spectrometry Using Poly(Dimethylsiloxane) Microchips with Monolithically Integrated Emitters. Analyst. 135, 2296–2302 (2010).
Chen, W., Yin, X., Mu, J. & Yin, Y. Subfemtomole Level Protein Sequencing by Edman Degradation Carried Out in a Microfluidic Chip. Chem Commun. 2488 (2007).
Qu, H. Y. et al. Stable Microstructured Network for Protein Patterning On a Plastic Microfluidic Channel: Strategy and Characterization of On-Chip Enzyme Microreactors. Anal Chem. 76, 6426–6433 (2004).
Culbertson, C. T., Jacobson, S. C. & Ramsey, J. M. Microchip Devices for High-Efficiency Separations. Anal Chem. 72, 5814–5819 (2000).
Gao, D., Liu, H. X., Jiang, Y. Y. & Lin, J. M. Recent Advances in Microfluidics Combined with Mass Spectrometry: Technologies and Applications. Lab Chip. 13, 3309–3322 (2013).
Koster, S. & Verpoorte, E. A Decade of Microfluidic Analysis Coupled with Electrospray Mass Spectrometry: An Overview. Lab Chip. 7, 1394 (2007).
Xue, Q. F. et al. Multichannel Microchip Electrospray Mass Spectrometry. Anal Chem. 69, 426–430 (1997).
Yu, C. et al. Microfluidic Self-Aspiration Sonic-Spray Ionization Chip with Single and Dual Ionization Channels for Mass Spectrometry. Rsc Adv. 6, 50180–50189 (2016).
Prudent, M., Rossier, J. S., Lion, N. & Girault, H. H. Microfabricated Dual Sprayer for On-Line Mass Tagging of Phosphopeptides. Anal Chem. 80, 2531–2538 (2008).
Lu, Y., Liu, F., Lion, N. & Girault, H. H. Dual-Channel Electrospray Microchip. J am Soc Mass Spectr. 24, 454–457 (2013).
Chambers, A. G. & Ramsey, J. M. Microfluidic Dual Emitter Electrospray Ionization Source for Accurate Mass Measurements. Anal Chem. 84, 1446–1451 (2012).
Monge, M. E., Harris, G. A., Dwivedi, P. & Fernandez, F. M. Mass Spectrometry: Recent Advances in Direct Open Air Surface Sampling/Ionization. Chem Rev. 113, 2269–2308 (2013).
Law, W. S. et al. On the Mechanism of Extractive Electrospray Ionization. Anal Chem. 82, 4494–4500 (2010).
Chen, H., Venter, A. & Cooks, R. G. Extractive Electrospray Ionization for Direct Analysis of Undiluted Urine, Milk and Other Complex Mixtures without Sample Preparation. Chem Commun. 2042 (2006).
Wang, R. et al. On the Mechanism of Extractive Electrospray Ionization (Eesi) in the Dual-Spray Configuration. Anal Bioanal Chem. 402, 2633–2643 (2012).
Luo, M. et al. Extractive Electrospray Ionization Mass Spectrometry for Sensitive Detection of Uranyl Species in Natural Water Samples. Anal Chem. 82, 282–289 (2010).
Li, Y. et al. Induced Dual-Nanospray: A Novel Internal Calibration Method for Convenient and Accurate Mass Measurement. J am Soc Mass Spectr. 24, 1446–1449 (2013).
Jiang, L. & Moini, M. Development of Multi-Esi-Sprayer, Multi-Atmospheric-Pressure-Inlet Mass Spectrometry and its Application to Accurate Mass Measurement Using Time-of-Flight Mass Spectrometry. Anal Chem. 72, 20–24 (2000).
Mortensen, D. N. & Williams, E. R. Ultrafast (1 Μs) Mixing and Fast Protein Folding in Nanodrops Monitored by Mass Spectrometry. J am Chem Soc. 138, 3453–3460 (2016).
Mortensen, D. N. & Williams, E. R. Investigating Protein Folding and Unfolding in Electrospray Nanodrops upon Rapid Mixing Using Theta-Glass Emitters. Anal Chem. 87, 1281–1287 (2015).
Fisher, C. M., Kharlamova, A. & McLuckey, S. A. Affecting Protein Charge State Distributions in Nano-Electrospray Ionization Via in-Spray Solution Mixing Using Theta Capillaries. Anal Chem. 86, 4581–4588 (2014).
Chang, D., Lee, C. & Shiea, J. Detecting Large Biomolecules From High-Salt Solutions by Fused-Droplet Electrospray Ionization Mass Spectrometry. Anal Chem. 74, 2465–2469 (2002).
Zhu, L., Gamez, G., Chen, H., Chingin, K. & Zenobi, R. Rapid Detection of Melamine in Untreated Milk and Wheat Gluten by Ultrasound-Assisted Extractive Electrospray Ionization Mass Spectrometry (Eesi-Ms). Chem Commun. 559–561 (2009).
Sun, J. et al. Development of an Air-Flow-Assisted Extractive Electrospray Ionization Source for Rapid Analysis of Phthalic Acid Esters in Spirits. Rapid Commun Mass Sp. 29, 1711–1716 (2015).
Bowers, J. J., Zimmerman, J. R., Oglesbee, R. A. & McLuckey, S. A. Adjacent Pulsed Nanoelectrospray Ionization Emitters for the Alternating Generation of Ions of Opposite Polarity. Anal Chem. 82, 1147–1150 (2010).
Satomi, Y., Kudo, Y., Sasaki, K., Hase, T. & Takao, T. Accurate Mass Measurement in Nano-Electrospray Ionization Mass Spectrometry by Alternate Switching of High Voltage Between Sample and Reference Sprayers. Rapid Commun Mass Spectrom. 19, 540–546 (2005).
Nepomuceno, A. I. et al. Dual Electrospray Ionization Source for Confident Generation of Accurate Mass Tags Using Liquid Chromatography Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. Anal Chem. 75, 3411–3418 (2003).
Young, N. L. et al. Steady-State Asymmetric Nanospray Dual Ion Source for Accurate Mass Determination within a Chromatographic Separation. Anal Chem. 79, 5711–5718 (2007).
Loo, R., Udseth, H. R. & Smith, R. D. Evidence of Charge Inversion in the Reaction of Singly Charged Anions with Multiply Charged Macroions. Journal of Physical Chemistry. 95, 6412–6415 (1991).
Schneider, B. B. & Covey, T. R. Calibrant Delivery for Mass Spectrometry. J am Soc Mass Spectr. 18, 991–996 (2007).
Schneider, B. B., Douglas, D. J. & Chen, D. D. Y. Multiple Sprayer System for High-Throughput Electrospray Ionization Mass Spectrometry. Rapid commun mass sp. 16, 1982–1990 (2002).
Tang, K., Lin, Y., Matson, D. W., Kim, T. & Smith, R. D. Generation of Multiple Electrosprays Using Microfabricated Emitter Arrays for Improved Mass Spectrometric Sensitivity. Anal Chem. 73, 1658–1663 (2001).
Xia, Y., Liang, X. & McLuckey, S. A. Pulsed Dual Electrospray Ionization for in/in Reactions. J am Soc Mass Spectr. 16, 1750–1756 (2005).
Qian, X. et al. A Reliable and Simple Method for Fabricating a Poly(Dimethylsiloxane) Electrospray Ionization Chip with a Corner-Integrated Emitter. Sensors-Basel. 15, 8931–8944 (2015).
Acknowledgements
This work was supported by the National Natural Science Foundation of China (Grant No. 81201165) and the Shenzhen fundamental research funding (Grant No. JCYJ20160531195439665 and JCYJ20160428182352200).
Author information
Authors and Affiliations
Contributions
Xiang Qian designed the experiments and revised the paper; Cilong Yu performed the experiments and wrote the paper; Fei Tang also performed the experiments and revised the paper; Yan Chen optimized the fabrication process and revised the paper; Quan Yu analyzed the mass spectrometer data; Kai Ni designed the homemade electrical and mechanical setup; Xiaohao Wang organized and revised the paper. All authors were involved in the preparation of this manuscript.
Corresponding authors
Ethics declarations
Competing Interests
The authors declare that they have no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Yu, C., Tang, F., Qian, X. et al. Multi-channel microfluidic chip coupling with mass spectrometry for simultaneous electro-sprays and extraction. Sci Rep 7, 17389 (2017). https://doi.org/10.1038/s41598-017-17764-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-17764-6