Abstract
Silicon carbide (SiC) is a wide band-gap semiconductor material with many excellent properties, showing great potential in fusion neutron detection. The radiation resistance of 4H-SiC Schottky diode detectors was studied experimentally by carefully analyzing the detectors’ properties before and after deuterium-tritium fusion neutron irradiation with the total fluence of 1.31 × 1014 n/cm2 and 7.29 × 1014 n/cm2 at room temperature. Significant degradation has been observed after neutron irradiation: reverse current increased greatly, over three to thirty fold; Schottky junction was broken down; significant lattice damage was observed at low temperature photoluminescence measurements; the peaks of alpha particle response spectra shifted to lower channels and became wider; the charge collection efficiency (CCE) decreased by about 7.0% and 22.5% at 300 V with neutron irradiation fluence of 1.31 × 1014 n/cm2 and 7.29 × 1014 n/cm2, respectively. Although the degradation exists, the SiC detectors successfully survive intense neutron radiation and show better radiation resistance than silicon detectors.
Similar content being viewed by others
Introduction
Since many giant scientific fusion devices, such as ITER1,2,3, EAST4,5, NIF6,7,8,9, etc. came into use in several countries in the world, the diagnostic of the neutron field in nuclear fusion plasmas has become an interesting and challenging subject. Owing to the high neutron fluence and extreme temperature in fusion devices, some of currently used semiconductor detectors based on silicon and germanium materials cannot satisfy the demands of neutron detection very well10,11. The germanium detectors need to operate in low temperature. The radiation resistance of silicon detectors is not ideal: significant radiation damage has been observed once the irradiation fluence reaching to 1 × 1012 n/cm2, and they are expected to be difficult to operate above neutron fluence of 1 × 1014 n/cm2 12,13. New radiation detectors based on new materials have been developed. Diamond is a good detection medium, with an ultra-high neutron radiation resistance and stable properties at varying temperatures, but the applications of diamond detectors are badly restricted by the tiny dimension and high cost of high-quality diamond materials14,15,16,17,18,19. With the development of semiconductor technologies, silicon carbide (SiC) has been found to be an ideal material for radiation detection with better radiation resistance in intense radiation field and better stability at high temperature than silicon and germanium20,21,22,23,24,25,26,27. In addition, the mature of SiC preparation technology has made it possible to fabricate large high-quality SiC detectors: the largest commercial SiC wafer up to 6 inch in diameter and high-quality epitaxial film over 100 μm in thickness has been successfully fabricated.
The radiation resistance is a key parameter for SiC detectors. In this paper, the radiation resistance of SiC detectors with Schottky diode structure is discussed. The detectors were irradiated at room temperature by deuterium-tritium fusion neutrons with energy of 14 MeV and total neutron fluence of 1.31 × 1014 n/cm2 and 7.29 × 1014 n/cm2 (Table 1). Then their parameters, including the curves of forward I-V, reverse I-V and C-V, the photoluminescence, the alpha particle spectra and the charge collection efficiency (CCE) were investigated. The SiC detectors survived after fast neutron irradiation with fluence over 1014 n/cm2 showing better resistance than silicon detector.
Experimental
Detector fabrication
Three 4H-SiC detectors (#1, #2 and #3) were fabricated with high-quality lightly doped epitaxial 4H-SiC layers grown by chemical vapor deposition (CVD) on commercial 4H-SiC N+ conducting substrate wafers (Φ 4 in. × 350 μm, target nitrogen doping concentration of 1019 cm−3). The epitaxial thickness is 20 μm and the target nitrogen doping concentration is 1 × 1014 cm−3. The front Schottky electrodes were made with Ni, 100 nm in thickness, prepared by thermal vacuum evaporation, and were coated with 2-μm-thick Au. The back ohmic contact electrodes were formed with Ni/Au (100 nm/3μm-thick). A set of multi-floating rings were made around the front contact to protect it from the damage of high voltage. All the three detectors were of Schottky structure, and have a sensitive volume of 1mm × 5 mm × 20 μm and dead layer of Ni/Au (100 nm/2 μm). The SiC diode chips were packaged in ceramic shells and were bonded with Au wires to the electrodes. Figure 1 is schematic diagram of the 4H-SiC detectors.
Neutron radiation
The irradiation was performed at the K600 Neutron Generator in China Institute Atomic Energy (CIAE) in Beijing, China, which can provide a constant fast neutron beam generated by the deuterium-tritium fusion, with an average energy of 14 MeV and a neutron fluence rate of 4–12 × 109 n/cm2s. Two SiC chips (Detector #2 and #3, before being packaged) were irradiated by the fast-neutron with fluence of 1.31 × 1014 n/cm2 and 7.29 × 1014 n/cm2, respectively. The radiation temperature was at 283 K.
Results and Discussion
I-V and C-V characteristics
The front and reverse I-V curves were measured with IWATSU CS-3200C Curve Tracer, and the results are shown in Fig. 2. The detectors were applied with reverse bias voltages in the range of 10 V to 600 V, thus the dark current of the detectors could be expressed by the reverse current. It is found that the dark current (reverse current) of the two detectors being irradiated increased significantly with the increase of neutron fluence. At the low fluence, it increased by three times or more; at the high fluence, it increased even more greatly by more than thirty times.
The forward I-V curve of the un-irradiated diode chip shows the rectification character, but the two chips being irradiated lost that character and the Schottky junction was broken down. The forward I-V characteristics of Schottky barrier diodes can be described by the Bethe’s Thermionic emission theory, in which the effective Richardson’s constant is 146 A cm−2 K−2 for 4H-SiC28,29,30,31,32. According to the forward I-V characteristics and the Bethe equation, the ideality factor was calculated to be 1.14, which indicates the current is dominated by thermionic current. The Schottky barrier height Фb for the Ni/4H-SiC contact was calculated to be 1.7 eV.
Figure 3a shows the C-V curve acquired by Agilent B1500A Semiconductor Parameter Analyzer. Figure 3b shows the curve of 1/C2 vs. V of the detector #1 (un-irradiated), from which the net doping concentration of 4H-SiC epitaxial layer and the built-in voltage of the Schottky diode can be acquired. From Fig. 3, we find the two samples (#2 and #3) being irradiated lost their C-V characters, and the effective doping concentration (Neff) of the 4H-SiC epitaxial layer was 7.9 × 1013 cm−3 and the built-in Vbi potential of the Schottky contact was 1.7 V. The Schottky barrier height can be expressed as
where N c is the effective density of the states in the conduction band of 4H-SiC, here is taken as 1.7 × 1019 cm−3. The barrier height thus was calculated 2.0 eV. The difference between the barrier heights from the I-V and C-V curves are due to the following factors: barrier height obtained from the forward I-V curve was calculated with the current which flows through the Schottky barrier over the entire area where the metal electrode covers, while the one derived from the C-V curve was calculated with the average capacitance related to the whole detector. Besides, the ideality factor we got is deviated from 1, which exposes the spatial inhomogeneity of the surface barrier height.
Photoluminescence
Photoluminescence (PL) experiments were performed to detect defects in the epitaxial material of the SiC detectors. Low temperature photoluminescence (LTPL) spectra were acquired in the spectral region between 380 nm and 800 nm at temperature of 83 K–203 K. A He-Cd laser with a wavelength of 325 nm was used as the excitation light source. Figure 4 shows the integrated pulsed PL spectra of the 20-μm-thick lightly doped epitaxial 4H-SiC layer taken at 83 K, with a nitrogen concentration of about 7.9 × 1013 cm−3.
As indicated in Fig. 4a, the distribution of the PL intensity of the 4H-SiC layer changes remarkably after neutron irradiation. The PL spectrum of the detector #1 is dominated by the near-band-gap nitrogen bound exciton lines (3.15–3.24 eV) and their associated phonon replicas (LO), but for the detector #2 and #3 being irradiated, the luminescence is completely quenched. This might be attributed to the severe lattice damage induced by neutron radiation.
The PL spectra in lower energy are dominated by a broad PL peak, covering green to yellow-green spectral range, peaking at about 2.12 ~ 2.23 eV, and in the spectra of all the samples, the intense and broad vibronic bands can be observed. By reference to the reported broad PL band by Sridhara et al.33, Gao et al34. and Sakai et al.35, we found this broad PL band might be composed by the peaks at 2.10 eV, 2.35 eV and 2.80 eV. The peaks at 2.35 eV and 2.80 eV might be due to the donor-to-acceptor (DAP) transition from the nitrogen donor (0.1 eV below the conduction band) to the deeper and shallower boron acceptors (0.7 eV and 0.3 eV above the valence band)36. The peak at 2.10 eV might be due to the carbon vacancy, which is a candidate for the electron transition, or other unidentified defect level35. The average luminance wavelength shifts with neutron fluence, the higher the neutron fluence is, the shorter the average wavelength would be, indicating some non-radiative defects have been produced by neutron irradiation.
Figure 5a shows the PL of the samples measured at temperature ranging from 84 K to 203 K. Broadband green luminescence is observed in the figure. As the temperature increases, the intensity of the luminescence decreases and the luminescence band becomes broad. This possibly derives from the lattice vibration and lattice scattering. (Fig. 5b) The broadband green luminescence might be due to both the vacancies of carbon and its extended point defects, which would lead to the consistent existence of green luminescence and the variation of the intensity and wavelength at different temperatures.
Alpha particle spectra
The detectors were tested under the irradiation of alpha particles from 239Pu alpha source (Φ10 mm, Eα = 5.157 MeV [73.3%]), 5.144 MeV[16.1%], 5.105 MeV[11.5%], 20000 Bq), provided by Northwest Institute of Nuclear Technology in Xi’an, China. The alpha source and the detectors were enclosed in a vacuum chamber, about 80 mm away from each other. The reverse bias was provided by a PS350 high voltage supply (Stanford research system Inc.) through the Ortec 142B preamplifier in the range of 0 to 300 V. Standard electronic devices, including an Ortec 142B preamplifier (with gain of 20 mV/MeV), an Ortec672 amplifier (with shaping time of 1μs and gain of 50 times) and an Ortec multi-channel analyzer (MCA), were used to record the signals. A laptop was used to obtain the pulse height spectra. The experimental setup is shown in Fig. 6(a).
According to the calculation with SRIM2003 Code37, the dead layer of the SiC detectors, comprised of Ni/Au (100 nm/2 μm), can absorb about 1.00 MeV of the incident alpha particles, leaving about 4.16 MeV kinetic energy penetrating into the active layer. Because the projected range of those remnant alpha particles, about 12.2 μm, is smaller than the sensitive thickness of the SiC detector (20 μm), all the remnant energy of the alpha particles would be deposited in the active layer.
The counts of the alpha particles as a function of channel number are shown in Fig. 6(b). The alpha-particle peaks can be clearly observed. The peak centroid of alpha particles is at Channel 665 for detector 1#, Channel 619 for detector 2# and Channel 516 for detector 3#. The higher neutron fluence is, the lower the peak centroid would be. The width of alpha peaks decreases with the increase of the incident neutron fluence. Fitting the peaks with Gaussian function, we got the FWHMs of the three alpha peaks, which are 391 keV for detector 1#, 384 keV for detector 2# and 270 keV for detector 3#. Excluding the influence of electronic noise (10 keV), static broadening (6.0 keV) and energy straggling of dead layer (180 keV)27, we got the inherent FWHM of 347 keV for detector 1#, 334 keV for detector 2# and 201 keV for detector 3#.
Charge Collection Efficiency
Charge collection efficiency (CCE) is defined as the ratio between the numbers of the electric charges collected by the detectors (Qc) and the total number of electric charges of all the excited carriers (Qg)28,32. Qg is dependent upon the energy of the alpha particles deposited in the detectors’ sensitive volume, about 4.16 MeV. Because all the electronic devices in our alpha detection experiments were kept working stably, the Qg could be determined by the peak centroid of the detector 1# at the reverse bias voltages which could make the detector fully depleted, at Channel 665, and Qc could be determined by the peak centroid of alpha peaks. Figure 7 gives the CCE of the three detectors at different bias voltages with different neutron radiation fluence.
As shown in Fig. 7, the CCE decreases with the increase of the neutron fluence, which is consistent with early researches of other scientists28,32. The 4H-SiC detectors couldn’t work without power supply after neutron irradiation. After the irradiation, the Schottky barrier of the detector was broken down, the 4H-SiC detector would not be depleted even if it is applied with reverse bias voltages within 300 V. The detectors are still alive after neutron irradiation but need high reverse bias voltages to avoid steep decrease of signal intensity. The CCEs of the SiC detectors decrease about 7.0% and 22.5% at 300 V when the neutron fluence reaches to 1.31 × 1014 n/cm2 and 7.29 × 1014 n/cm2, respectively. But the CCE of a silicon detector decreased much more, over 75%, at lower fast-neutron irradiation fluence of 9.98 × 1012 n/cm2 38. It can be concluded that the radiation resistance of SiC detector is much better than silicon detector.
Theoretically, the CCE of the un-irradiated detector is expressed as32
where E ion is the energy of the incident alpha particles, 5157 MeV; d is the depletion width at a given bias; dE/dx is the rate of loss of energy of the implanted alpha particles as they penetrate the 4H-SiC epilayer; R is the projected range of the incident particles with an energy of E ion, 16 μm; L d is the diffusion length of the minority carriers, 6.8 μm; ε and ε r are dielectric and relative dielectric constant of the dead layer, 9.7 and 8.85 × 10−14 F/cm, respectively; q is 1.6 × 10−19 C; N eff is 7.9 × 1013 cm−3.
The CCEs of the SiC detectors being irradiated are expressed as28
where E is the electric field, E = V/d; T is the thickness of the 4H-SiC epitaxial layer, T = 20 μm; λp and λn are the mean free path of the holes and electrons, respectively, \({\lambda }_{p}={\mu }_{p}{\tau }_{p}E,\,{\lambda }_{n}={\mu }_{n}{\tau }_{n}E\), μ is the carrier mobility; τ is the carrier lifetime;. μnτn and μhτh are 2.2 × 10−8 cm2/V and 1.7 × 10−8 cm2/V with the low fluence of 1.31 × 1014cm−2, 1.5 × 10−8 cm2/V and 1.1 × 10−8 cm2/V with the high fluence of 7.29 × 1014cm−2. Compared with the value of μτ (over 1000 × 10−8 cm2/V) for the holes of the detector not being irradiated28, it decreases significantly after neutron irradiation. The decrease of μτ can be atrributed to the neutron irradiation defects, which act as trapping centers for carriers, reducing the probability of the carrier transportation through the SiC material.
The calculation result of the CCEs is plotted in Fig. 7 with open dots. They are well consistent with the experimental results acquired in the alpha particle detection.
Conclusions
We compared the properties and performance of 4H-SiC Schottky diode detectors before and after the irradiation of deuterium-tritium fusion neutrons with total fluence of 1.31 × 1014 n/cm2 and 7.29 × 1014 n/cm2 at room temperature. We found that the 4H-SiC Schottky diode detectors being irradiated survived the intense neutron radiation, and were still effective and could be used in radiation detection even though the detector performance was degraded by the increase of dark current, reduction of CCE, decrease of μτ and movement of alpha peaks’ centroid. The degradation can be attributed to the lattice damage, non-radiative defects and other defects induced by neutron irradiation.
It is known that the silicon detector is hard to operate above neutron fluence of 1 × 1014 n/cm2 and the degradation of its CCE is worse than SiC detector after fast-neutron irradiation12,38. Hence it can be concluded that the 4H-SiC Schottky diode detectors have a better neutron resistance than silicon detector and could be expected to be well used in fusion neutron detection.
References
Shimada, M. et al. Progress in the ITER Physics Basis -Chapter 1: Overview and summary. Nuclear Fusion. 47(6), (2007).
Bertalot, L. et al. Fusion neutron diagnostics on ITER tokamak. J Instrum. 7(04), C04012 (2012).
Johnson, L. C. et al. Neutron diagnostics for ITER[J]. Rev Sci Instrum. 68(1), 569–572 (1997).
Chen, H. et al. Improvement of the laser beam pointing stability for east thomson scattering diagnostic. J FUSION ENERG. 34(1), 9–15 (2015).
Zang, Q. et al. Development of a thomson scattering diagnostic system on east. Plasma Sources Sci Technol. 12(2), 144–148 (2010).
Lindl, J. D. et al. The physics basis for ignition using indirect-drive targets on the national ignition facility. Phys Plasmas. 11(2), 339–491 (2004).
Miller, G. H. The national ignition facility. Opt Eng. 43(12), 21–51 (2001).
Glebov, V. Y. et al. Development of nuclear diagnostics for the national ignition facility. Rev Sci Instrum. 77(10), 11 (2006).
Murphy, T. J. et al. Nuclear diagnostics for the national ignition facility (invited). Rev Sci Instru. 72(1), 773–779 (2001).
Li, Z. Radiation hardness / tolerance of Si sensors / detectors for nuclear and high energy physics experiments. Office of Scientific & Technical Information Technical Reports. (2002).
Lindström, G., Moll, M. & Fretwurst, E. Radiation hardness of silicon detectors – a challenge from high-energy physics. Nucl Instrum Methods Phys Res A. 426(1), 1–15 (1999).
Adam, W. et al. Review of the development of diamond radiation sensors. Nucl Instrum Methods Phys Res A. 434, 131–145 (1999).
Robert, W. Kuckuck, Semiconductor detectors for use in the current mode, Lawrance Radiation Laboratory, UCRL–51011,(1971).
Gabrysch, M. Electronic Properties of Diamond. Uppsala, April 2008 (ISSN: 0349– 8352).
Pan, L. S. & Kania, D. R. Diamond: Electronic Properties and Applications, Kluwer Academic Publishers, Boston (1995).
Kania, D. R. et al. Diamond radiation detectors. Diam Relat Mater. 2(5–7), 1012–1019 (1993).
Liu, L. Y. et al. Polycrystalline chemical-vapor-deposited diamond for fast and ultra-fast neutron detection. Sci China Technol Sci. 55(9), 2640–2645 (2012).
Liu, L. Y. et al. Polycrystalline CVD diamond detector: fast response and high sensitivity with large area. AIP Adv. 4, 017114 (2014).
Liu, L. Y. et al. (2016). Properties comparison between nanosecond x-ray detectors of polycrystalline and single-crystal diamond. Diam Relat Mater. 73, 248–252 (2017).
Wright, N. G. & Horsfall, A. B. SiC sensors: a review. J Phys D Appl Phys 40(20), 6345 (2007).
Ruddy, F.H. Silicon carbide radiation detectors: progress, limitations and future directions. Mater Res Soc Symp. (Spring Meeting, mrss13-1576-ww01-01) (2013).
Ruddy, F. H. et al. Development of a silicon carbide radiation detector. IEEE Trans. Nucl. Sci. 45(3), 536–541 (1998).
Ha, J. H. et al. 4H-SiC PIN-type semiconductor detector for fast neutron detection. Prog Nucl Sci Technol. 2011(1), 237–239 (2011).
Seshadri, S. et al. Demonstration of an SiC neutron detector for high-radiation environments. IEEE Trans. Electron Devices. 46(3), 567–571 (1999).
Giudicea, A. L. et al. Performances of 4H-SiC Schottky diodes as neutron detectors. Nucl Instrum Methods Phys Res A. 583(1), 177–180 (2007).
Wu, J. et al. Feasibility study of a SiC sandwich neutron spectrometer. Nucl Instrum Methods Phys Res A. 708, 72–77 (2013).
Liu, L. Y. et al. (2016). Properties of 4h silicon carbide detectors in the radiation detection of 86 Mev oxygen particles. Diam Relat Mater. 73, 177–181 (2017).
Wu, J. et al. Effect of neutron irradiation on charge collection efficiency in 4H-SiC Schottky diode[J]. Nucl Instrum Methods Phys Res A 735(1), 218–222 (2014).
Moloi, S. J. & Mcpherson, M. Capacitance–voltage behaviour of schottky diodes fabricated on p-type silicon for radiation-hard detectors. Radiat Phys Chem. 85(16), 73–82 (2013).
Verzellesi, G., Vanni, P., Nava, F. & Canali, C. Investigation on the charge collection properties of a 4h-sic schottky diode detector. Nucl Instrum Methods Phys Res A. 476(3), 717–721 (2002).
Sze, S. M. & Ng, K. K. Physics of Semicondutor Devices, 3rd., John Wiley&Sons.
Mannan, M. A. et al. Effect of Z1/2, EH5, and Ci1 deep defects on the performance of n-type 4H-SiC epitaxial layers Schottky detectors: Alpha spectroscopy and deep level transient spectroscopy studies. J Appl Phys 115(22), 224504 (2014).
Sridhara, S. G. et al. Photoluminescence and transport studies of boron in 4H SiC. J Appl Phys 83, 7909–7919 (1998).
Gao, Y. et al. Selective doping of 4H–SiC by codiffusion of aluminum and boron. J Appl Phys. 90(11), 5647–5651 (2001).
Sakai, K. et al. Impurity and defect centers of n-type 4H-SiC single crystals investigated by a photoluminescence and a piezoelectric photo thermal spectroscopies. Solid State Electron. 48(10), 1873–1876 (2004).
Kimoto, T. et al. Nitrogen donors and deep levels in high‐quality 4H–SiC epilayers grown by chemical vapor deposition. Appl Phys Lett. 67(19), 2833–2835 (1995).
Ziegler, J. F. & Biersack, J. P. SRIM-2003.26: The Stopping and Range of Ions in Matter, SRIM.com, Annapolis, MD (2003).
Wu, J. Study of neutron detection based on 4H-SiC detectors, China Academy of Engineering Physics (2014).
Acknowledgements
The research work in this paper is supported by the National Natural Science Foundation of China (Grant Nos 11605140, 11435010, 11275153, and 61504099). The authors want to thank Prof. X.L. Lu, Dr. T. Jiang, Dr. Q. Zhu and Prof. X.H. Ma. at Xidian University for their help of the low temperature PL measurements, Dr. L.W. Lu at Institute of Semiconductors of Chinese Academy of Science for useful discussions and L. Wang for English revision.
Author information
Authors and Affiliations
Contributions
L.Y. Liu and X.P. Ouyang designed the experiment, finished neutron irradiation, finished all the radiation measurements and wrote the main manuscript text. A. Liu, L. Lv and P. Jin carried out some of measurements. S. Bai designed the SiC diodes and analyzed part of P.L. measurement result. All authors reviewed the manuscript.
Corresponding authors
Ethics declarations
Competing Interests
The authors declare that they have no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Liu, L., Liu, A., Bai, S. et al. Radiation Resistance of Silicon Carbide Schottky Diode Detectors in D-T Fusion Neutron Detection. Sci Rep 7, 13376 (2017). https://doi.org/10.1038/s41598-017-13715-3
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-017-13715-3
This article is cited by
-
Isoelectronic aluminum-doped gallium nitride alpha-voltaic cell with efficiency exceeding 4.5%
Communications Materials (2023)
-
Latest developments in room-temperature semiconductor neutron detectors: Prospects and challenges
Science China Physics, Mechanics & Astronomy (2023)
-
Simultaneous measurement of energy spectrum and fluence of neutrons using a diamond detector
Scientific Reports (2022)
-
Thermal neutron irradiation effects on structural and electrical properties of n-type 4H‒SiC
Journal of Materials Science: Materials in Electronics (2020)
-
Recoil-proton track imaging as a new way for neutron spectrometry measurements
Scientific Reports (2018)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.